APPLICATIONS
Battery Testing
Current Amplitude
Current in Circuits
Voltammetry
• New Battery Development
N600-HJ201B
This instrument is a low cost battery charge discharge unit ideal
for testing various types of batteries. It is suitable for the
development and evaluation of new batteries in a research
environment.
Specialized multi-channel battery charge discharge systems
can easily be formed by combining multiple units of HJ-201B
and multi-channel intermittent recorders. Various sizes of racks
are available for this purpose. Sophisticated data analysis is
also possible by interfacing the auxiliary outputs of the HJ-201B
to a computer system via A/D converters.
Note: We can produce a charge/discharge system to meet
specific requirements. Contact us at 1-800-799-NADA for more
information.
Voltage
Discharge - Range
Discharge - Setting Range
Discharge - Presetting
Discharge - Accuracy
Charge - Range
Charge - Setting Range
Charge - Presetting
Discharge - Range
0 minute ~ 19 days 23 hours 50 minutes
8, 9, 10, 11
Rest after Disch. - Range & Pre.
Same as discharge
Minus -1, 0, 1, 2, 3, 4, 5, 6, 7,
a 12 step rotary switch
a 10 turn vernier dial
< 10mV
0V to 20V
a 20 step rotary switch
< 10mV
Current
Discharge - Current Ranges
1A, 100mA, 10mA, 1mA,
Discharge - Setting
a 5 step rotary switch
Discharge - Accuracy
Charge - Current Ranges
Charge - Setting
Charge - Accuracy
Discharge - Presetting
Charge - Range & Presetting
Rest after Charge - Range & Pre.
Modes
0 to 20
a 10 turn veriner dial
Charge - Accuracy
Timer
Minus -1V to 11V
100µA
a 10 turn vernier dial
< 0.5% full scale
1A, 100mA, 10mA, 1mA,
100uA
a 5 step rotary switch
a 10 turn vernier dial
< 0.5% full scale
Rear View of the Unit
6-digit digital thumbwheel switch
Same as discharge
Same as discharge
a. Constant Current & Voltage Discharge
b. Rest after Discharge
c. Constant Current & Voltage Charge
Switching
d. Rest after Discharge
T: Automatic switching from one mode to
another
V, T: Automatic switching from one mode to
Display
Output - Current
Output - Voltage
Auxiliary Analog Output - Current
another
Day, Hour, Minute, Discharge,
Discharge Rest, Charge, Charge Rest,
Stand-By, Operation C/D Cycle Counter
10V output for every range full scale
I/O conversion ratio is switched with a toggle
switch. Ratio: 2:1 , 1:1, 1:2
10V output for every range; +:Discharge,
-:Charge
Auxiliary Analog Output - Voltage I/O conversion ratio is 2:1
Inputs
I+: Battery + terminal I-: Battery - terminal
Standard Accessories
Alligator Clips: 1.2meters; Earth Cord: 3 meters
Power Requirement
Physical Dimension
Weight
V+: Battery sensor + terminal V-: Battery
AC120V, 60Hz
290mm x 195mm x 330mm (WxHxD)
8.4 kg