Fully Automatic Infrared Moisture Analyzer for Daily

MA37
Fully Automatic Infrared Moisture Analyzer
for Daily Routine Operation
Advantages
– High-speed measurements
– Effortless cleaning
– User-friendly operation
– Display of process status
Product Description
The MA37 is the new Sartorius basic model for moisture determination. It works faster and is easier to use than
its predecessor MA35. In addition, it features a compact footprint and delivers highly accurate results. Moreover,
its BetterClean design also allows for effortless cleaning, and the status light indicates the current process status.
The MA37 is ideal for operations characterized by consistently routine measurement tasks.
Applications
The MA37 moisture analyzer can be used for quick and reliable
determination of the moisture content of liquid, pasty and
solid substances using the thermogravimetric method. Typical
applications of the MA37 include the analysis of foods, beverages,
pharmaceuticals, chemicals, paper materials and environmental
protection products. For daily recurring moisture analysis tasks,
such as in production or incoming goods inspection, the MA37 is
perfect!
Performance
Fully automatic endpoint determination eliminates the
cumbersome capture of termination criteria. The moisture
analyzer monitors the drying process and stops the measurement
once the sample weight is constant. The built-in weighing system
with 1 mg resolution is specifically optimized for use in higher
temperature ranges and delivers the required measurement
accuracy. Two powerful metal tube heating elements heat the
samples in the MA37. These heating elements are fast, extremely
rugged and durable. Compared to glass heating lamps, e.g.
infrared lamps or halogen heaters, they are especially resistant
to dirt and vibrations.
Moisture Analyzer
HUBERLAB. AG
Industriestrasse 123
4147 Aesch
T 061 717 99 77
F 061 711 93 42
www.huberlab.ch
info@huberlab.ch
Technical Specifications
Max. weighing capacity
Repeatability, average
Readability
Typical sample quantity
Result display
Temperature range and settings
Sample heating
Heating programs
Shutoff parameter
Access to sample chamber
Measuring program
Memory for data storage
Operator guidance features
Language selection
Status light
Sample inspection
Draft shield
Cleaning
Log printout
Monitoring of inspection and testing
equipment
Data interface
70 g
for initial sample weight approx. > 1 g: ± 0.2 %
for initial sample weight approx. > 5 g: ± 0.05 %
1 mg, 0.01 %
5 – 15 g
Moisture content in %M and g | dry matter in %S and g | ATRO in %M/S
40°C–160°C, stand-by temperature selectable from 40 – 100°C in increments of 1°C
Infrared metal heating elements
Standard drying, gentle drying
Optional:
– Fully automatic
– Semiautomatic mg (1 – 50 mg | 5 – 300 sec.)
– Semiautomatic % (0.1 – 5.0 % | 5 – 300 sec.)
– Timer settings (02:00 – 99.59 min.)
– Manual
Removable hood with wide opening angle, SoftClose mechanism
1 program saved in a non-volatile memory (freely selectable method parameters)
Results are saved until the start of the next measurement
Intuitive user interface, including touch screen and easy-to-understand menu guidance
English, German, French, Spanish, Italian, Russian, Chinese, Polish, Japanese, Portuguese,
Turkish
Displays the status “analysis running / START”, “analysis completed / STOP” or “analysis error”
LED-illuminated sample chamber, inspection window with grid above the hood
Integrated draft shield
Removable hood, inspection window with grid and sample chamber plate for easy cleaning
in the dishwasher
– Printout using the optional, internal printer YDP30; alternatively, the printer YDP20-OCE
can be used with an adapter (YCC03-D09)
- GLP-compliant, inalterable standard for results printout | paper-saving short record
External calibration using optional calibration weights
Frequency
Power consumption
Temperature range
Mini USB
– Automatic detection of the Sartorius printers YDP30 and YDP40
– Direct data transfer to Microsoft® Windows programs without any additional software
– Programmable data output interval
– Data transfer protocols SBI, table format, text format
50/60 Hz
Max. 640 VA
10°C – 30°C
Housing dimensions (W + D + H)
Weight
215 + 400 + 210 mm
Approx. 6.2 kg
Available Models
MA37 115V/230V
Automatic voltage detection 115V/230V
Accessories
6965542
6906940
6906941
YCW452-AC-02
YDP30
YDP40
YCC03-D09
Disposable sample pans, 80 pcs., aluminum,
d 90 mm
Glass fiber pad for analysis of pasty and
fatty samples, hard quality, 80 pcs.,
d 90 mm
Glass fiber pad for analysis of liquid and
fatty samples, soft quality, 200 pcs.,
d 90 mm
External calibration weight, 50 g (E2) with
DKD certificate
Standard printer
Premium GLP laboratory printer
Adapter cable for connecting the printer
YDP20-OCE
Equipment Supplied
Moisture analyzer, power cord, user manual, aluminum sample
pans, glass fiber filter
Moisture Analyzer
3