Aquatec® Pure Bidet
en Shower toilet
User Manual
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
©2017 Invacare Corporation
All rights reserved. Republication, duplication or modification in whole or in part is prohibited
without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks
are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
Contents
1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 Symbols in this user manual . . . . . . . . . . . . . . . . . . . . . .
1.3 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.4 Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.5 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.6 Service life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Safety information on electromagnetic compatibility. . . . .
2.3 Labels and symbols on the product . . . . . . . . . . . . . . . . .
3 Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Functional description of the self-cleaning
mechanism . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.1 Installation / Assembly . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Assembling concealed cistern . . . . . . . . . . . . . . . . . . . . .
4.3 Trial operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.4 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.2 Using the unit control panel and remote control . . . . . . .
5.3 Using the shower toilet . . . . . . . . . . . . . . . . . . . . . . . . . .
6 Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1 Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.1 Maintenance and servicing . . . . . . . . . . . . . . . . . . . . . . .
7.2 Cleaning and care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.3 Freeze prevention during winter . . . . . . . . . . . . . . . . . . .
7.4 Changing batteries of remote control. . . . . . . . . . . . . . . . 26
4
4
4
4
4
5
5
6
6
7
8
9
9
10
11
11
14
15
16
18
18
18
19
24
24
25
25
25
26
8 After Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.1 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Re-use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
28
28
28
28
9 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
9.1 Identifying and repairing faults . . . . . . . . . . . . . . . . . . . . 29
10 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.2 Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10.3 Environmental Parameters . . . . . . . . . . . . . . . . . . . . . . .
31
31
32
32
Aquatec® Pure Bidet
1 General
Useful tips, recommendations and information
to ensure efficient and smooth operation.
1.1 General information
This product complies with the directive
93/42/EEC for medical products. The launch
date for this product is specified in the CE
declaration of conformity.
This user manual contains information and notes on the
proper and safe use of the bidet.
Please read this user manual carefully before using the
product. Please heed all the notes, particularly the safety
information, and act accordingly.
The position specifications (A, B, C, ...) in the handling
instructions always refer to the preceding figure.
1.2 Symbols in this user manual
In this user manual, warnings are marked with symbols. The
title next to the warning symbol indicates the level of danger.
DANGER
Indicates an immediate danger that will result
in serious injury or death if not avoided.
WARNING
Indicates a dangerous situation which could
lead to death or serious injury if not avoided.
CAUTION
Indicates a dangerous situation which could
lead to minor injuries if not avoided.
IMPORTANT
Indicates a dangerous situation which could
lead to damage if not avoided.
4
Manufacturer
1.3 Warranty
We provide a manufacturer’s warranty for the product
in accordance with our General Terms and Conditions of
Business in the respective countries. Warranty claims
can only be made through the provider from whom the
appliance was obtained.
1.4 Compliance
Quality is fundamental to the company’s operation, working
within the discipline of ISO 13485.
The product meets the requirements of EN 1717 (Protection
against pollution of potable water installations and general
requirements of devices to prevent pollution by backflow).
Invacare is continuously working towards ensuring that the
company’s impact on the environment, locally and globally,
is reduced to a minimum.
•
•
We comply with the current environmental legislation
(e.g. WEEE and RoHS directives).
We only use REACH compliant materials and
components.
1590852-C
General
•
We are compliant with the Water Regulations Advisory
Scheme (WRAS).
For further information please contact Invacare in your
country (for addresses, see back page of this manual).
1.5 Intended use
This product is intended exclusively as an aid for going to the
toilet in domestic settings. Any other use is prohibited. The
product may be used without an assistant only by persons
who are physically and mentally able to safely deal with it.
WARNING!
Risk of injury
Any other or incorrect use could lead to
hazardous situations.
– Do not use the product as a transfer aid.
– Never stand on the product.
Indications
Toilet attachments with cleansing function are necessary
with illnesses or disabilities that due to considerable
functional deficits of the upper extremities, comparable to
being without hands or arms, make it impossible for the
patient to clean the genital area the usual way on his or her
own after use of the toilet.
1.6 Service life
If the product is used as intended and in accordance with
the safety regulations, the expected product life is 5 years.
The effective service life can vary, however, depending on
the frequency and intensity of use.
1590852-C
5
Aquatec® Pure Bidet
2 Safety
2.1 Safety information
6
DANGER!
Electric shock hazard
– Do not open or repair the product. In case of
damages, consult your authorized dealer.
– Do not use the product while bathing and do
not pour or spray water on the unit.
– Always unplug the appliance during cleaning.
– Do not place or store product where it can fall
or be pulled into a tub or sink and do not place
or drop product into water or other liquid.
– Do not reach for a product that has fallen into
water. Unplug immediately.
– Do not use damaged or loose power cord.
– Damaged power cords must be replaced by the
manufacturer or an authorized dealer.
– Do not touch the power plug with wet hands.
– Never operate this product if it has a damaged
cord or plug, if it is not working properly, if
it has been dropped or damaged, or dropped
into water
– Never drop or insert any object into any
opening or hose.
– Do not cover up the warm air outlet or insert
hands or other objects.
– Keep the power cord away from heated
surfaces.
– Only connect the product to an outlet
corresponding to the rated voltage of the unit.
– Before removing the unit from the toilet bowl,
disconnect the power plug first.
WARNING!
Risk of injury
– For assembly only use the fastening materials
included within the scope of delivery.
– Do not make any unauthorized changes or
alterations to the product.
– Never block the air openings of the product
and keep the air openings free of lint, hair,
and the like.
– Never use while sleeping or drowsy.
– Do not operate where aerosol(spray) products
are being used or where oxygen is being
administered.
– Make sure that children do not play with the
product.
– Make sure that persons (including children)
with reduced physical, sensory or mental
capabilities use the product without close
supervision only if they are able to safely deal
with the product.
– Do not use high pressure or steam cleaners.
– Keep the unit away from fires or inflammable
substances.
– Do not connect product to industrial water
supplies.
– Do not put any lighted cigarette or burning
material in the toilet.
– Never stand on or place objects on the product.
1590852-C
Safety
IMPORTANT!
Risk of damage to the product
– Do not slam down the lid.
– Do not lean on the lid.
– Do not sit on the lid.
– When not in use for a long period of time,
disconnect the power plug and drain the water
tank.
– Do not twist or pull the connection hoses.
2.2 Safety information on electromagnetic
compatibility
This product was successfully tested to international
standards for its electromagnetic compatibility. However,
electromagnetic fields such as those generated by radio
and television transmitters, radio equipment and mobile
telephones, can influence the operation of products with
electrical drives. The electronic used in this product can
also cause weak electromagnetic interference but this lies
below the statutory limits. You should therefore observe
the following warnings:
1590852-C
CAUTION!
Electromagnetic interference
Wireless communications equipment such
as WLAN routers, mobile phones, cordless
telephones and their base stations or
walkie-talkies can affect this equipment.
– Make sure that such devices are kept in a
distance of at least 2 m away from the product.
– If other electrically operated medical products
cause a interference or are themselves
disrupted, switch off your Invacare® product
and contact your dealer.
Guidance and manufacturer's declaration relating to
electromagnetic emission and immunity are available
on request from Invacare® in your country.
7
Aquatec® Pure Bidet
2.3 Labels and symbols on the product
Read user manual
The identification label is applied to the rear side of the
product and contains following information:
WRAS approved product
B
A
M
L
K
J
I
H
C
D
E
F
G
A
Device designation
H
Type of protection
B
Manufacturer
I
Class II equipment
Note (disposal)
J
Nominal voltage /
rated power
D
HMI number
K
Maximum user weight
Production date
E
Interval of use: Max. L
30 minutes continuous
use followed by a 5
minutes break.
F
Serial number
Type B applied part
G
Conformity mark
C
8
M
1590852-C
Components
3 Components
3.1 Components
The following components are included within the scope
of delivery:
L
A
B
A
Lid
B
Remote control receiver
C
Control panel
D
Outlet of warm air dryer
E
Seat sensor
F
Cleansing nozzle
G
Bidet nozzle
H
Heated seat
I
LED Night-light
J
Inlet adapter
K
Remote control receiver
L
Power plug
C
D
E
F
G
K
H
J
I
1590852-C
9
Aquatec® Pure Bidet
3.2 Functional description of the self-cleaning
mechanism
B
The self-cleaning mechanism of the showering and cleansing
nozzle is automatically activated before and after each
function / use.
F
A
H
I
C
G
D
E
A
Catch plate
B
Brackets (2x)
C
Fixation bolts (2x)
D
Washers (2x)
E
Packing nuts (2x)
F
PVC Flexible hose (for main unit)
G
T-valve
H
Batteries for remote control
I
Remote control and holder
10
The showering and cleansing nozzle is automatically
retracted and rinsed with fresh water for approx. 5 seconds.
Since the shower and cleansing nozzle is at this time within
the device, water cannot be splashed when the lid is open.
When pressing a function key, the nozzle moves out again
and is ready for use.
When the device had been in standby-mode or had
been re-connected to the power supply, the self-cleaning
mechanism is activated once.
1590852-C
Setup
4 Setup
4.1 Installation / Assembly
A
Assembling the shower toilet
The Aquatec® Pure Bidet fits standard and
norm-WC’s.
C
WARNING!
Risk of being scalded by hot water
– Only connect the shower toilet to a cold water
supply.
WARNING!
– Check all the accessory parts are in place.
– Do not connect the power until installation is
finalized.
B
1.
2.
Unfasten the fixation nuts B of the existing toilet seat
A underneath the toilet bowl.
Remove the toilet seat and keep it with the fixation
parts B and C for later use.
The shower toilet is suitable for all types of water
supply connection with a flushing cistern (top,
bottom, sideways).
If the T-valve does not fit onto the water supply,
adapter pieces are necessary.
A
1.
2.
1590852-C
Close the shut off valve A clockwise.
Drain toilet water tank completely by flushing.
11
Aquatec® Pure Bidet
A
A
B
C
1.
Press lever A to detach the holder plate B from the
bidet.
A
B
1.
2.
Detach the water supply hose A on the part of the
shut off valve B.
Mount the T-valve C between shut off valve and water
supply hose.
If part A is a rigid pipe, it must be replaced by a
flexible hose.
1.
Insert the supplied packing nuts A into the fixing holes
on the toilet bowl.
One pair of standard packing nuts with a diameter of
14 mm are included in the package as an accessory
part. If the fixing hole diameter on the toilet bowl
is less than 14 mm other suitable packing nuts and
washers need to be used.
12
1590852-C
Setup
B
CAUTION!
– Make sure that the grooves underneath the
shower toilet are precisely guided in the
recesses of the mounting plate.
A
C
D
1.
2.
3.
Put the washers D on both packing nuts.
Set in place the brackets B on the catch plate C.
Fasten the catch plate with the fixation bolts A through
the brackets in both packing nuts.
B
A
1.
C
B
1.
2.
Screw the PVC flexible hose A clockwise on the T-valve
B.
Attaching remote control holder
A
A
Screw the PVC flexible hose A clockwise on the inlet
adapter B of the shower toilet.
Slide the shower toilet to the catch plate C installed.
1.
1590852-C
Attach the remote control holder with double stick tape
A on the wall in a suitable location.
13
Aquatec® Pure Bidet
Preparing the remote control
A
Installing water flow over flush-mounted housing
B
A
C
1.
2.
3.
4.
Remove battery cover A from remote control.
Insert batteries B. Ensure correct polarity.
Close battery cover of remote control.
Place remote control D into its holder C.
4.2 Assembling concealed cistern
Installing water flow over angle valve
1.
2.
Connect angle valve and bidet with the PVC Flexible
hose (accessory).
Turn the water tap on.
1.
2.
3.
4.
5.
6.
7.
8.
IMPORTANT!
– Check the connection for leaks and re-fit if
necessary.
IMPORTANT!
– Check the connection for leaks and re-fit
if necessary.
9.
14
Remove the push plate from the flush-mounted housing.
Remove the rear cover plate to obtain a direct
perspective on the connection.
Turn the water tap off.
Remove connecting sleeve D and replace by T-piece E.
Connect T-piece with PVC Flexible hose C.
Connect PVC Flexible hose with tube A using nut B.
Connect tube A with pure bidet.
Turn the water tap on.
Cut-out inner cover plate F, mounting plate G and
push plate H according to the figure above taking into
account the indicated dimensions [mm].
1590852-C
Setup
10. Re-attach the inner cover plate and the mounting plate
inside the flush-mounted housing.
11. Close flush-mounted housing with the push plate.
Connecting the power plug
WARNING!
– This product must be wired in accordance
with the latest edition of BS 7671 requirement
for electrical installations (UK) or relevant
equivalent wiring standards for your
location/country.
Following parts are needed for the assembly of the
flush-mounted water flow but are not included in
the scope of delivery:
• Tube A
• Nut B
Available at your specialised dealer.
IMPORTANT!
If this product is used in a bathroom or shower
room, the supplied plug must be cut off and
the connection to the power supply must be
made via switched fused spur and RCD. No
other appliance should share the spur with this
product. The work must be carried out by a
qualified electrician.
4.3 Trial operation
Leakage test
1.
2.
Read and follow the safety information regarding
electrical safety, → 2.1 Safety information, page 6 .
Connect power plug to a suitable power source.
WARNING!
Electric shock hazard
– Do not turn over or move the unit after power
connection.
A
1.
2.
Open the shut off valve A completely by turning
counterclockwise.
Check for any water leakage on the connection hoses,
tighten if necessary.
1590852-C
3.
If the power indication light is not turned on after the
power plug connection, press any button except HI/LO,
FR/BK buttons of the remote control, then power goes
on.
15
Aquatec® Pure Bidet
4.4 Disassembly
Checking the key functions
Follow the below instructions if in need of detaching the
unit for removal.
1.
2.
3.
4.
B
Disconnect the power plug.
Close the water supply by turning the shut off valve
clockwise.
Drain the toilet water tank completely by flushing.
Disconnect the PVC flexible hose from the T-valve.
A
C
1.
2.
Place one hand on the seat sensor A.
With the other hand press “CLEANSING / STOP” on the
unit control panel B or “CLEANSING” or “BIDET” on
the remote control.
•
16
Water sprays out of the nozzle. Cover the nozzle
with your free hand C to prevent water spattering
out.
1590852-C
Setup
5.
Lift the device slightly and pull it forwards out of the
holder.
A
6.
Unfasten the drain plug A underneath the unit and
drain the water completely out of the water tank.
7. Disconnect the PVC flexible hose and T-valve.
8. Connect the water-supply hose from the water tank to
the shut off valve.
9. Open the shut off valve by turning counterclockwise.
10. Check the connections for any leakages, tighten if
necessary.
11. Reinstall the original seat and lid to the toilet.
1590852-C
17
Aquatec® Pure Bidet
5 Usage
Overview of key functions on remote control
Pos.
5.1 Safety information
WARNING!
– Before usage read and follow the safety
information, → 2.1 Safety information, page 6 .
5.2 Using the unit control panel and remote
control
O
N
P
Q
M
L
C
R
K
I
D
I
E
Key name
Function
A
WATER / DRY
Display of the current water
pressure or the temperature
of the warm air dryer
B
STOP
Stops all active functions
C
CLEANSING/MOVE
Cleansing function is
performed
D
BIDET/MOVE
Bidet function is performed
E
DRY
Warm air dryer function
begins to operate
F
AUTO
Automatic cleaning function
starts
G
SEAT
Adjusting seat temperature
H
WATER
Adjusting water temperature
I
FRONT / BACK
Change nozzle position
J
HIGH / LOW
Adjusting warm air dryer
temperature
B
Adjusting water pressure
T
U
18
F T
J S
HIP BATH
Hip bath cleansing is
performed
G A S J H
1590852-C
Usage
Overview of key functions on remote control
Overview of key functions on control panel
Pos.
Pos.
Key name
Function
S
HIGH / LOW
Display of current water
temperature or seat
temperature
U
BATTERY SIGN
Illuminates when batteries
need to be replaced
Overview of key functions on control panel
Pos.
Key name
Function
K
CLEANSING/STOP
Device turns off
L
POWER SAVING
Power-saving mode activates
M
LIGHT
Night-light activates
N
LIGHT SENSOR
When night-light function is
activated, bidet automatically
senses toilet light and
night-light LED switches on
O
POWER
Illuminates when unit is
switched on
P
LIGHT
Illuminates when night-light
is activated
Key name
Function
Q
AUTO
Illuminates when automatic
power-saving mode is
activated
R
SUPER
Illuminates when super
power-saving mode is
activated
5.3 Using the shower toilet
Perform manual cleansing
A sensor is located inside the toilet seat, that is only
activated by skin contact.
1.
If plugged-in, the device is turned on and the "POWER"
LED O illuminates on the control panel.
•
2.
3.
Set yourself centered on the toilet seat.
Press the CLEANSING/MOVE button C or BIDET/MOVE
button D.
•
4.
CLEANSING respectively BIDET-function is
performed.
Press again CLEANSING/MOVE button or BIDET/MOVE
button during the cleansing process.
•
1590852-C
If the "POWER" LED does not illuminate, press any
button on the remote control (except HIGH/LOW
or FRONT/BACK buttons).
The cleansing nozzle moves back and forth. Nozzle
stops by pressing the button again.
19
Aquatec® Pure Bidet
5.
Press STOP button B on the remote control.
•
6.
Shower procedure stops.
Press AUTO button F on the remote control.
•
Press DRY button E on the remote control.
•
7.
1.
•
Warm air dryer is activated. Dryer function stops
by pressing the “STOP” button.
Press CLEANSING/STOP button K and hold for more
than 4 seconds.
•
The device is turned off — All displays switch off
on the control panel.
CLEANSING/MOVE and BIDET/MOVE function
operates for 90 seconds. Hip bath function operates
for 3 minutes. DRY function operates for 3 minutes.
The functions are ending with an audible beep.
Before and after each operation, the self nozzle
cleaning function runs and sanitizes the nozzle head.
•
2.
Press STOP button B on the remote control.
•
3.
The 3rd level for water pressure, the 2nd level for
warm air dryer temperature and the last setting of
nozzle position are automatically set-up.
The Sensor forces to set the water- and seat
temperature to “Medium”, in case the temperature
were set up on the “High” position.
A buzzer sounds when the auto cleansing function
finishes.
The automatic cleansing process is interrupted.
Press AUTO button F again.
•
The previous automatic cleansing process is
resumed.
Automatic cleansing function: Move cleansing and
warm air dryer run for 1 minute each.
Perform automatic cleansing
The automatic cleansing function is a default function for
people with disabilities, children or the elderly, in which a
complete cleansing cycle is automatically executed.
20
Setting water pressure
The water pressure can be set in three steps (indicated by
first, middle and last LED):
1590852-C
Usage
1.
2.
Press CLEANSING/MOVE button C on the remote
control.
Press HIGH respectively LOW button J on the remote
control.
•
If the cleaning functions are repeatedly started, the
water temperature drops gradually. To regain the
desired water temperature, suspend the cleaning
function for three minutes. The last value of the
water temperature is automatically saved by the
device.
The water pressure of the cleaning nozzle is raised
or lowered by one step and displayed on the LED
A.
The last set value of the water pressure is
automatically saved by the device.
Setting seat temperature
The seat temperature can be set in four steps:
Setting nozzle position
1.
Press FRONT respectively BACK button I during
cleansing function on the remote control.
•
The cleansing nozzles move forwards respectively
backwards.
When the cleansing process is complete and you
move away from the seat, the nozzle moves back
into the last setting of the nozzle position.
Setting water temperature
The water temperature can be set in four steps:
1.
Press WATER button H repeatedly on the remote
control, until the desired temperature is reached.
•
1590852-C
The water temperature is displayed on the LED
display above.
1.
Press SEAT button G repeatedly on the remote control,
until the desired temperature is reached.
•
The seat temperature is displayed on the LED
display above.
The last value of the seat temperature is
automatically saved by the device.
The seat is equipped with an internal combustion protection
system. If the seat temperature is set to "High" and you
remain longer than 30 minutes on the seat, the seat
temperature drops automatically. As soon as you leave the
seat, the temperature returns to the setpoint.
Setting warm air temperature
The warm air temperature can be set in five steps:
21
Aquatec® Pure Bidet
Automatic power saving mode
1.
1.
2.
Press DRY button E on the remote control.
Press HIGH respectively LOW button J on the remote
control.
•
Press POWER SAVING button L repeatedly, until “AUTO”
LED Q illuminates.
•
Auto power saving function is temporarily released,
when sitting on the seat. Auto power saving function
runs again, when moving away from the seat.
The temperature of the warm air dryer is raised or
lowered by one step and displayed on the LED A.
The last value of the warm air dryer temperature is
automatically saved by the device.
2.
Press POWER SAVING button twice.
•
Turning on/off night-light LED
1.
Press LIGHT button M on the units control panel.
•
The night-light is activated when it gets dark in the
room.
When the night light is activated, the device
automatically recognizes the toilet light and the night
light is turned off. The User can conveniently bidet
without toilet light. The power consumption of the
night-light is very low.
2.
Power saving mode deactivates and “AUTO” LED
turns off.
Super power saving mode
Bidet senses toilet luminosity for itself and shows excellent
power saving function.
1.
Press POWER SAVING button L repeatedly, until
“SUPER” LED R illuminates.
•
Water and seat temperature maintain normal
temperature.
Power saving mode is temporarily released when
toilet light turns on. It returns to normal mode from
power saving mode even if occupying the seat in
dark condition.
Press LIGHT button again to switch off the night-light.
2.
Press POWER SAVING button once.
•
22
Seat temperature is automatically set at 30 °C
Power saving mode deactivates and “SUPER” LED
turns off.
1590852-C
Usage
Perform hip bath function
1.
Press HIP BATH button T on the remote control.
The hip bath function operates for 3 minutes (The
nozzle moves slightly forwards and backwards).The
lowest level of water pressure and the highest level
of water temperature is pre-set.
2.
3.
Press HIP BATH button again to stop the nozzle moving
forwards and backwards.
Press STOP button to cancel the hip bath function.
1590852-C
23
Aquatec® Pure Bidet
6 Transport
6.1 Safety information
IMPORTANT!
Risk of damage to the product
– Only transport the product in dry areas at
temperatures between 0 °C and 40 °C, air
pressure of 795 hPa - 1060 hPa and humidity
of 30% - 75%.
– Do not expose the product to direct sunlight.
24
1590852-C
Maintenance
7 Maintenance
IMPORTANT!
Hygiene is particularly important in a sensitive
area like the toilet.
– Keep the product clean and wash it regularly.
7.1 Maintenance and servicing
If the cleaning and safety notices are observed, the product
will be maintenance free.
IMPORTANT!
All cleaning agents and disinfectants used must
be effective, compatible with one another and
must protect the materials they are used to clean.
– For further information on decontamination
in Healthcare Environments, please refer
to ‘The National Institute for Clinical
Excellence’ guidelines on Infection Control
www.nice.org.uk/CG139 and your local infection
control policy.
– Do not dry-clean the device.
– Clean the product by hand.
– Do not use abrasive cleaning agents.
WARNING!
– Check the product regularly for damage and to
ensure that it is securely assembled.
– In the event of damage and/or necessary
repairs, contact the dealer immediately.
– The product must be checked by an authorized
dealer before every re-use or after 24
months at the latest (see Manufacturer‘s
recommendation for rehab products of the
SPECTARIS industrial association).
7.2 Cleaning and care
WARNING!
Electric shock hazard
– Do not touch the power socket with wet hands.
– Disconnect the power plug from the wall outlet
before cleaning the unit.
– Make sure that no water splashes to the plug
or the wall outlet.
– Observe the safety information for electrical
safety, → 2.1 Safety information, page 6 .
1590852-C
Cleaning the shower toilet
1.
2.
Clean the unit with moist, well-wrung soft cloth. For
excessive stains, use soft cloth moistened with mild
detergent solution
Wipe out the device with a dry cloth.
25
Aquatec® Pure Bidet
Cleaning the water strainer
1.
Press “CLEANSING / STOP” and “POWER SAVING”
buttons simultaneously on the control panel.
•
2.
3.
B
Clean the nozzle manually with a soft brush.
Press “CLEANSING / STOP” button.
•
A
C
The nozzle A comes out and stops.
The nozzle returns to its storage position and the
self cleaning function is automatically activated.
Disinfecting the shower toilet
1.
2.
3.
4.
5.
6.
7.
Close the water supply by turning the shut off valve
clockwise.
Detach the PVC flexible hose A from the inlet adapter
B.
Detach the strainer C from the inlet adapter with a
pointed object.
Thoroughly clean the strainer with a soft brush.
Reinsert the strainer on the inlet adapter.
Attach the PVC flexible hose to the inlet adapter.
Open shut off valve fully by turning counterclockwise.
Cleaning the nozzles
1.
2.
3.
Disinfect the product by wiping down all generally
accessible surfaces with commercially available
disinfectants.
Thoroughly wipe clean the treated surfaces with a damp
cloth.
Thoroughly dry the surfaces with a clean, dry cloth.
7.3 Freeze prevention during winter
1.
2.
Connect the power plug and check if the power unit
is on.
Press WATER on the remote control, until the water
temperature is set to "Low" or higher,
see 5.3 Using the shower toilet, page 19.
7.4 Changing batteries of remote control
The battery life cycle depends on the frequency of use.
A
26
1590852-C
Maintenance
CAUTION!
Risk of remote control explosion
– Only use batteries of type AA (LR6) 1.5 V.
– Insert batteries according to the +/- indicator
in the remote control.
– Do not use old and new batteries together.
1.
Change used batteries into new ones when “BATTERY
SIGN” LED blinks on the remote control, see Preparing
the remote control, page 14.
1590852-C
27
Aquatec® Pure Bidet
8 After Use
8.3 Disposal
8.1 Storage
The disposal and recycling of used products and packaging
must be conducted in accordance with the applicable legal
regulations.
IMPORTANT!
Risk of damage to the product
– Store the product in a dry area at a temperature
of between 0 °C and 40 °C, air pressure of 795
hPa - 1060 hPa and humidity of 30% - 75%.
– Do not store the product near heat sources.
– Keep the product out of direct sunlight.
The manufacturer can handle the correct and complete
disposal of the device.
This product contains batteries. Old batteries may contain
hazardous substances which could be a danger to the
environment and public health.
Dispose of batteries according to your local regulations.
IMPORTANT!
– If not in use for a long time, drain the water
tank completely, → 4.4 Disassembly, page 16.
8.2 Re-use
The product is suitable for re-use. The number of re-uses
depends on the frequency of use and the manner in which
the product was used.
Before re-use, the product should undergo hygienic
treatment and be checked to ensure it is in proper and safe
working order. In need of repair contact authorized dealer.
28
1590852-C
Troubleshooting
9 Troubleshooting
9.1 Identifying and repairing faults
Fault
Possible cause(s)
Remedy
Unit does not operate at all
Power is off
Press any button to check except “HIGH”,
“LOW”, “FRONT” or “BACK” buttons of
the remote control.
Power plug is not connected
There is a power failure
CLEANSING, BIDET, DRY, AUTO do not
operate
Seat is covered
Make sure that there is nothing on
the seat surface. (The device is only
activated when the user's skin touches
the seat sensor.)
Remote control does not operate at all
Batteries in the remote control have
been exhausted
Replace batteries with new ones,
→ 4.1 Installation / Assembly, page 11.
Malfunction of the remote control
Batteries are empty or too weak
Replace batteries with new ones,
→ 4.1 Installation / Assembly, page 11.
Remote control out of reach of the
device
No water spray
Low water pressure
1590852-C
The shut off valve is closed
Open shut off valve.
There is water supply outage
Restore water supply.
The water pressure level is set to "LOW"
Increase water pressure level,
→ 5.3 Using the shower toilet, page 19.
The water strainer is clogged
Clean water strainer, → 7.2 Cleaning and
care, page 25.
Nozzle head is dirty
Clean nozzle, →7.2 Cleaning and care,
page 25.
29
Aquatec® Pure Bidet
Fault
Possible cause(s)
Remedy
Low water temperature
The water temperature level is turned
off or set to ”LOW”
Turn on or increase water temperature
level,
→ 5.3 Using the shower toilet, page 19.
Seat is not warm
The warm seat temperature level is
turned off or set to “LOW”
Turn on or increase seat temperature
level,
→ 5.3 Using the shower toilet, page 19.
Dryer air is not warm
The warm air dryer temperature level is
turned off or set to “LOW”
Turn on or increase warm air dryer
temperature level,
→ 5.3 Using the shower toilet, page 19.
30
1590852-C
Technical data
10 Technical data
10.1 General data
Product depth
156 mm
Product weight
3.9 kg
Cleansing functions
Power supply
220 – 240 V AC, 50/60 Hz
Max. power consumption
1725 – 2053 W
Intermittent (periodic
operation)
85%, 30 min. ON / 5 min.
OFF
Type of protection, entire
device
IP X4
Length of power cord
1.4 m
Battery type remote control
2 x AA (LR6) / 1.5 V
Type water supply
connection
Direct connection to the
water supply
Ø Connection at shower
toilet
1/2”
Temperature adjustment
Room temperature, 35 °C,
38 °C, 40 °C
Ø T-piece
1/2”- 3/8” - 1/2”
Heater capacity
55 – 65 W
Ø Connection hose
1/2” cranked to 3/8”
Safety devices
Thermal fuse
Length connection hose
500 mm
Drying functions
Water supply pressure
79 – 735 kPa
Temperature adjustment
Water supply temperature
3 – 25 °C
Room temperature – 50 °C
(5 levels)
Operating temperature
3 – 40 °C
Heater capacity
195 – 232 W
Product width
389 mm
Safety devices
Thermal fuse, Bi-metal
Product length
515 mm
1590852-C
1)
Water flow rate CLEANSING
0.7 l/min
Water flow rate BIDET
0.85 l/min
Water pressure adjustment
3 levels
Temperature adjustment
Normal water supply
temperature,
34 °C, 37 °C, 40 °C
Heater capacity
1670 – 1987 W
Safety devices
Thermal fuse, Thermostat
Warm seat
Protected against the penetration of water that is splashed
against the housing from all directions (warm seat only).
1)
31
Aquatec® Pure Bidet
10.2 Materials
Seat
PP
Lid
PP
All product components are corrosion resistant and Latex
free.
10.3 Environmental Parameters
Operating conditions
Ambient temperature
3 - 40 °C
Relative humidity
15 % - 93 %, non-condensing
Atmospheric pressure
795 - 1060 hPa
Storage and shipping conditions
Ambient temperature
0 - 40 °C
Relative humidity
30 % - 75 %, non-condensing
Atmospheric pressure
795 - 1060 hPa
32
1590852-C
Notes
Notes
Notes
Invacare® distributors:
United Kingdom:
Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
Tel: (44) (0) 1656 776 222
Fax: (44) (0) 1656 776 220
uk@invacare.com
www.invacare.co.uk
Aquatec Operations GmbH
Alemannenstrasse 10
D-88316 Isny
Germany
1590852-C
2017-12-12
*1590852C*
Making Life’s Experiences Possible®