OPERATION MANUAL
SINGLE-PHASE, THREE-WIRE
OUTPUT DRIVER
PCR-LA Series
2P03-PCR-LA
Use of Operation Manual
Please read through and understand this Operation Manual before operating the product. After reading, always keep the manual nearby so that you may refer to it as
needed. When moving the product to another location, be sure to bring the manual as
well.
If you find any incorrectly arranged or missing pages in this manual, they will be
replaced. If the manual gets lost or soiled, a new copy can be provided for a fee. In
either case, please contact Kikusui distributor/agent, and provide the “Kikusui Part
No.” given on cover.
This manual has been prepared with the utmost care; however, if you have any questions, or note any errors or omissions, please contact Kikusui distributor/agent.
Reproduction and reprinting of this operation manual, whole or partially, without our
permission is prohibited.
Both unit specifications and manual contents are subject to change without notice.
© 2005-2008 Copyright Kikusui Electronics Corporation
Safety Symbols
For the safe use and safe maintenance of this product, the
following symbols are used throughout this manual and on
the product. Understand the meanings of the symbols and
observe the instructions they indicate (the choice of symbols used depends on the products).
OR
DANGER
WARNING
CAUTION
Indicates that a high voltage (over 1000 V) is used here.
Touching the part causes a possibly fatal electric shock.
If physical contact is required by your work, start work
only after you make sure that no voltage is output here.
Indicates an imminently hazardous situation which, if
ignored, will result in death or serious injury.
Indicates a potentially hazardous situation which, if
ignored, could result in death or serious injury.
Indicates a potentially hazardous situation which, if
ignored, may result in damage to the product and other
property.
Shows that the act indicated is prohibited.
Is placed before the sign “DANGER,” “WARNING,” or
“CAUTION” to emphasize these. When this symbol is
marked on the product, see the relevant sections in this
manual.
Indicates a protective conductor terminal.
Indicates a chassis (frame) terminal.
2P03-PCR-LA
Safety Symbols I
Arrangement of this manual
This Operation Manual is made up of the following sections.
Chapter 1 General
Provides an overview of the Single-phase, Three-wire Output Driver
and describes its features. Also denotes parts names and descriptions.
Chapter 2 Precautions and Preparation for Use
Describes the precautions and preparation information that must be
understood for use of the Single-phase, Three-wire Output Driver.
Chapter 3 Operation Method
Describes the operations and functions of the 2P03-PCR-LA Singlephase, Three-wire Output Driver.
Chapter 4 Using the Remote Controller
Describes the functions and operation methods for single-phase,
three-wire operation of the PCR-LA power supplies when the Singlephase, Three-wire Output Driver is used together with the Remote
Controller (RC04-PCR-LA).
Chapter 5 Using the GPIB/RS-232C Interface
Describes the functions and messages available when the Singlephase, Three-wire Output Driver and GPIB Interface (IB03-PCR-LA)
or RS-232C Interface are used in combination for single-phase, threewire operation of the PCR-LA power supplies.
Chapter 6 Maintenance
Describes the maintenance procedures for the Single-phase, Threewire Output Driver.
Chapter 7 Specifications
Lists the specifications.
II Arrangement of this manual
2P03-PCR-LA
Contents
Safety Symbols - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - I
Chapter 1 General
1.1
1.2
1.3
1.4
1-1
Outline - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Features - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Part Names and Descriptions - - - - - - - - - - - - - - - - - - - - - - The Applicable ROM Version of the PCR-LA - - - - - - - - - - -
Chapter 2 Precautions and Preparation for Use
2-1
2.1
2.2
2.3
Check at Unpacking - - - - - - - - - - - - - - - - - - - - - - - - - - - - Handling Precautions - - - - - - - - - - - - - - - - - - - - - - - - - - - Functional Restrictions during
Single-phase, Three-wire Operation - - - - - - - - - - - - - - - - - 2.4 Combination with Other Options - - - - - - - - - - - - - - - - - - - 2.5 Single-phase, Three-wire Output Driver Installation Method- 2.5.1 Installing the Phase Cards in a Slot- - - - - - - - - - - - - - - 2.5.2 Connecting the Drive-signal Cable - - - - - - - - - - - - - - - 2.6 PCR-LA AC Power Supply Connection- - - - - - - - - - - - - - - 2.6.1 Connecting the OUTPUT Terminal Board- - - - - - - - - - 2.6.2 Connecting the SENSING Terminal Board - - - - - - - - - Chapter 3 Operation Method
3.1
3.2
4.1
2-1
2-2
2-3
2-3
2-4
2-4
2-5
2-7
2-8
2-8
3-1
Operation Check - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Basic Operation Required for
Single-phase, Three-wire Operation - - - - - - - - - - - - - - - - - 3.3 Switching between Phase Voltage and Line Voltage Displays 3.4 Output Voltage Setting - - - - - - - - - - - - - - - - - - - - - - - - - - 3.5 Limit Value Setting - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3.6 Setting the Current or Power Display Mode - - - - - - - - - - - - 3.7 Memory Function- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3.8 Key-lock Function - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3.9 Self-test Function - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3.10 Functions Restricted during
Single-phase, Three-wire Operation - - - - - - - - - - - - - - - - - 3.11 Special Waveform Output - - - - - - - - - - - - - - - - - - - - - - - - Chapter 4 Using the Remote Controller
1-1
1-1
1-2
1-3
3-1
3-2
3-4
3-4
3-5
3-5
3-6
3-6
3-6
3-7
3-8
4-1
Switching between Phase Voltage and Line Voltage Displays - 4-2
2P03-PCR-LA
Contents III
4.2
4.3
4.4
4.5
Displaying the V-phase Current and Voltage - - - - - - - - - - - Switching to the Current or Power Display Mode - - - - - - - - Sequence Operation - - - - - - - - - - - - - - - - - - - - - - - - - - - - Harmonic Current Analysis Function - - - - - - - - - - - - - - - - -
Chapter 5 Using the GPIB/RS-232C Interface
5.1
5.2
5.3
5.4
4-3
4-3
4-3
4-4
5-1
Phase Voltage/Line Voltage Display Messages- - - - - - - - - - Output Voltage Setting Messages - - - - - - - - - - - - - - - - - - - Output Measurement Messages - - - - - - - - - - - - - - - - - - - - Harmonic Current Analysis Messages - - - - - - - - - - - - - - - - -
5-2
5-3
5-5
5-7
Chapter 6 Maintenance
6-1
Chapter 7 Specifications
7-1
7.1
7.2
Specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-1
Dimensions - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 7-2
IV Contents
2P03-PCR-LA
Chapter 1
General
Provides an overview of the Single-phase, Three-wire Output Driver
and describes its features. Also denotes parts names and descriptions.
1.1
Outline
This device (2P03-PCR-LA) is an option for operating the PCR-LA
Series AC power supplies in single-phase, three-wire output.
Use of the Single-phase, Three-wire Output Driver allows the outputs
of two PCR-LA AC power supplies to be connected in series for use
as single-phase, three-wire system power supplies.
1.2
Features
Use of this device allows the PCR-LA AC power supplies to provide
the following enhanced functions, in addition to the PCR-LA power
supply’s standard functions.
Single-phase Three-wire Output
Use of this device allows two PCR-LA AC power supplies to generate
single-phase, three-wire outputs, in which the U-phase unit of the two
power supplies acts as a master and their V-phase unit act as a slave.
Setting of Either Line Voltage or Phase Voltage
The line voltage can also be set in addition to the phase voltage.
Line Voltage Measurement
Connecting the SENSING terminal board at the terminal box of the
PCR-LA AC power supply to a line voltage measurement point
allows measurement of the line voltage.
Measurements of the Total Power, Total Apparent
Power, and Total Power Factor of Single-phase,
Three-wire Output
Use of this output driver allows for the display of total power and
total apparent power obtained through measurements using the two
PCR-LA AC power supplies. The total power factor of the two phases
can also be calculated from these total values.
2P03-PCR-LA
General 1-1
Measurements of the apparent power and power factor require one of
the RS-232C Interface, the optional Remote Controller (RC04-PCRLA), or the GPIB Interface (IB03-PCR-LA).
1.3
Part Names and Descriptions
Drive-signal connector
Used to connect a drivesignal cable
Mounting holes
Used to fix this device to the PCR-LA
power supply with screws
Fig.1-1
1-2 General
2P03-PCR-LA
2P03-PCR-LA
1.4
The Applicable ROM Version of the
PCR-LA
This Operation Manual applies to cases in which the driver is connected to PCR-LA AC power supplies with ROM of
version 3.35 or later.
When using above options with the PCR-LA in single-phase threewire operation, the PCR-LA must be required with the ROM version
of 3.35 or later. If the ROM version of the PCR-LA is 3.34 or previous version, the PCR-LA is required for the ROM update.
To check the ROM version of the PCR-LA, refer to the operation
manual of the PCR-LA series.
In case, the PCR-LA needs the ROM update, contact your Kikusui
agent or distributor.
When making an inquiry about the product, please provide us with
the following information:
• Model name of the PCR-LA power supplies
• ROM version of the PCR-LA power supplies
• Serial numbers and revision numbers of the PCR-LA power
supplies (indicated at the lower rear of the products)
• Model name of this driver
2P03-PCR-LA
General 1-3
1-4 General
2P03-PCR-LA
Chapter 2 Precautions and Preparation for Use
Describes the precautions and preparation information that must be
understood for use of the Single-phase, Three-wire Output Driver.
2.1
Check at Unpacking
Upon reception of the product, confirm that the package contains the
necessary accessories and that the device and accessories have not
been damaged during transportation.
If the device is damaged or any accessory is missing, notify Kikusui
distributor/agent.
The U- and V-phase cards have
the indications “U” and “V” on
the panel part of each card.
U- and V-phase cards (one each, total of two cards)
M3 x 6 mm
Card-mounting screws (2 pcs. for each card, total of 4 pcs.)
[M3-101-001]
Drive-signal cable (1 pc.)
[91-87-5670]
Fig.2-1
2P03-PCR-LA
Operation Manual (1 copy)
[Z1-002-912]
Items Contained in the Package
Precautions and Preparation for Use 2-1
2.2
Handling Precautions
Handling of the U- and V-phase Cards
Always observe the following precautions when handling each phase
card, as its PCB is not protected. Otherwise, a problem may occur.
■ Never touch any of the electronic parts installed in
the PCB.
■ Never handle the card under conditions in which
static electricity may accumulate.
■ After unpacking the Single-phase, Three-wire Output Driver carton, promptly install each phase card
in a PCR-LA power supply.
■ When storing the phase cards, always take antielectrostatic measures such as storing them in the
bag in which they were packaged.
■ Do not drop a card or subject it to other impact.
■ Do not place a card where it could be exposed to
water or other liquid.
Handling of the Drive-signal Cable
■ Do not damage the cable.
■ Do not pull, bend, or apply any other stress to the
cable.
2-2 Precautions and Preparation for Use
2P03-PCR-LA
2.3
Functional Restrictions during
Single-phase, Three-wire Operation
Use of the Single-phase, Three-wire Output Driver allows the PCRLA AC power supplies to generate single-phase, three-wire outputs.
However, this partially restricts the power supply functions. The following functions are not available when the device is installed in the
power supplies.
• DC mode
• AC + DC mode
• Sensing
• Regulation adjustment
• Average voltage value measurement
• Average current value measurement
NOTE
2.4
• The power line abnormality simulation function can be used.
However, some of the specifications differ from those
achieved in single-phase operation.
Combination with Other Options
The PCR-LA Series AC power supplies have various options in addition to the Single-phase, Three-wire Output Driver. However, the following options can be used only in the PCR-LA power supply
(master) in which the U-phase card has been installed. These options
do not function in a slave unit. (Note that PD03M-PCR-LA can also
be installed in the V-phase power supply.)
Table 2-1 Options Available in the Master Unit
*
Option name
Model name
Remote Controller
RC04-PCR-LA
GPIB Interface
IB03-PCR-LA
Parallel Operation Driver (for the master)
PD03M-PCR-LA*
When this device is combined with the Parallel Operation Driver to perform parallel operation in a single-phase, three-wire system, the number of power supplies to be operated in parallel is up to three for each
phase.
2P03-PCR-LA
Precautions and Preparation for Use 2-3
2.5
Single-phase, Three-wire Output Driver
Installation Method
The Single-phase, Three-wire Output Driver can be installed by
inserting the U- and V-phase cards into SLOT3 or SLOT4 of the two
PCR-LA power supplies, respectively.
2.5.1 Installing the Phase Cards in a Slot
CAUTION • Before installing a phase card in a slot, be sure to turn
the PCR-LA AC power supply POWER switch OFF.
1. Unscrew the screws that fasten the SLOT3 or SLOT4 cover
to remove it.
The slots are located at the upper rear part of the PCR-LA power
supplies.
Keep the removed cover and screws to ensure that they are not lost.
2. Hold each phase card by the panel.
3. Orient the phase card so that the parts-mounted side of the
PCB is at the right, and place the PCB in the slot grooves.
4. Carefully insert the card into the slot so that the PCB does
not come out of the grooves.
5. Fully insert the card, and confirm that it cannot be pulled out
easily.
6. Fix the card to the PCR-LA power supply using the screws
provided.
Fig.2-2
Installing a Phase Card
This completes installation of the phase card.
NOTE
• When a Single-phase, Three-wire Output Driver card has
been inserted into the slot of a PCR-LA power supply, that
2-4 Precautions and Preparation for Use
2P03-PCR-LA
power supply can be used only in a single-phase, three-wire
system. To use the power supply in a single-phase system, the
phase card must be removed from the slot.
2.5.2 Connecting the Drive-signal Cable
CAUTION • Before connecting a drive-signal cable, always turn
the PCR-LA power supply POWER switches OFF.
1. Place the three PCR-LA power supplies as close to each
other as possible.
Arrange them so that no stress is applied to the drive-signal cable.
A PCR-LA power supply with the U-phase card installed plays
the role of the master unit that controls the other two power supplies. It is recommended that the PCR-LA power supply with the
installed U-phase card be placed where it is easily accessible for
operations.
2. Using the drive-signal cable, connect the U- and V-phase
cards.
Open the tabs of the J1 connectors of the U- and V-phase cards,
orient the connector of a drive-signal cable to a J1 connector, and
insert it into the J1 connector.
In such a case, insert the cable connector into the J1 connector so
that the tabs are closed to lock the cable securely.
The J1 connector has two connection ports, each of which operates the same.
The J1 connectors of the AC power supplies located at the right
and left sides have only one cable connected to the slot. Do not
connect anything to the other empty slot.
Connect the shields at both ends of the drive-signal cable to a
card-fixing screw.
Shield
Connect the crimping terminal of
the shield to a card-fixing screw.
Fig.2-3
2P03-PCR-LA
Drive-signal Cable
Precautions and Preparation for Use 2-5
This completes connection of the drive-signal cable.
CAUTION • Moving a PCR-LA power supply with a drive-signal
cable connected to the phase card may result in
breakage of the cable or connector.
Before moving the PCR-LA power supply, be sure to
disconnect the drive-signal cable(s).
2-6 Precautions and Preparation for Use
2P03-PCR-LA
2.6
PCR-LA AC Power Supply Connection
This section describes the procedure for connection of the outputs
between the PCR-LA AC power supplies.
For the input and output connections of a single PCR-LA power supply, always read through “Chapter 2 Installation and Preparation for
Use” in the PCR-LA AC Power Supply Operation Manual.
WARNING • Before carrying out connection work, be sure to cut
off the power supply from the switchboard.
PCR–LA
V-phase power supply
U-phase power supply
OUTPUT
G N L SENSING
N L
OUTPUT
G N L SENSING
N L
Sensing cable
Output (load) cable
Load
G
Fig.2-4
NOTE
L2
N
L1
Output Connections
• Because the operation check described in the next chapter is
conducted under a no-load condition, it is still not necessary
to connect a load at this point.
2P03-PCR-LA
Precautions and Preparation for Use 2-7
2.6.1 Connecting the OUTPUT Terminal Board
To generate single-phase, three-wire outputs (using the two PCR-LA
AC power supplies), connect the outputs of these power supplies as
shown below.
PCR-LA series (two units)
L1
Load
L
PCR-LA
U-phase
power supply
N
N
L
PCR-LA
V-phase
power supply
N
Fig.2-5
NOTE
L2
Connecting OUTPUT Terminal Boards
• If the neutral point is omitted, there are cases in which it will
not be possible to produce the rated power.
2.6.2 Connecting the SENSING Terminal Board
The SENSING terminal boards are used to measure line voltage. For
this measurement, connect the SENSING and OUTPUT terminal
boards as specified in Table 2-2.
Table 2-2 Connections of the SENSING Terminal Boards
SENSING terminal boards
OUTPUT terminal boards
L of the U-phase power supply
L of the U-phase power supply
N of the U-phase power supply
N of the V-phase power supply
2-8 Precautions and Preparation for Use
2P03-PCR-LA
Chapter 3
Operation Method
Describes the operations and functions of the 2P03-PCR-LA Singlephase, Three-wire Output Driver.
This Operation Manual describes the procedures for use of the functions available in single-phase, three-wire operation that differ from
those available when a PCR-LA power supply is used in single-phase
operation. For information not provided in this manual, see the PCRLA AC Power Supply Operation Manual. Moreover, before reading
the 2P03-PCR-LA Operation Manual, always read through the PCRLA AC Power Supply Operation Manual.
NOTE
3.1
• Unless otherwise specified, the operation method described in
this chapter applies to the U-phase power supply.
Operation Check
Before conducting an operation check of single-phase, three-wire
operation, confirm the connections of each PCR-LA power supply by
referring to Chapter 2 “Precautions and Preparation for Use”.
CAUTION • To protect a load, carry out an operation check without connecting a load. If a load has been connected
to the PCR-LA power supplies, disconnect it.
Operation Check Procedure
Principally, check the following items in the same way as in “2.7 Operation Check” in the PCR-LA AC Power Supply Operation Manual.
• POWER switch ON
• Voltage setting
• OUTPUT ON/OFF
• Voltage-range switching
For operations specific to single-phase, three-wire operation, see “3.2 Basic
Operation Required for Single-phase, Three-wire Operation” onward.
NOTE
• Even if one power supply is not connected or an error occurs
in either of the two power supplies, neither of the PCR-LA
power supplies operate.
2P03-PCR-LA
Operation Method 3-1
3.2
Basic Operation Required for
Single-phase, Three-wire Operation
The basic operation for single-phase, three-wire operation is as
described below.
POWER-switch operation
• When turning the POWER switches ON, turn the POWER switch
of the V-phase power supply ON first, followed by the U-phase
power supply POWER switch. Alternatively, turn the POWER
switches of both two power supplies ON simultaneously.
• When turning ON (or OFF) both PCR-LA power supply POWER
switches, complete both turn-ONs within three seconds.
NOTE
• If the turn-ON procedure for the POWER switches is
improper, the AC power supplies may not start up. In such a
case, turn OFF all POWER switches and then turn them ON
again by the correct procedure.
Operation and Indications of the Control Panel
• During single-phase, three-wire operation, the U-phase power supply acts as the master to control the V-power supply. Thus, to
change a setting, use the U-phase power supply control panel. Values set through the U-phase power supply control panel will be set
to the V-phase power supply in the same way.
• In phase voltage setting, the voltage concerned can be set from the
V-phase power supply control panel. This setting changes only the
voltage of the V-phase power supply.
DESCRIPTION
• Difference in Display between Single-phase and
Single-phase, Three-wire Operations
When this device is installed in the PCR-LA power supplies to
perform single-phase, three-wire operation, part of the control
panel display differs from that provided for single-phase operation.
(1) Indications common to each phase
• The
symbol and the letter “U” or “V” at the left of
the voltage display area indicate that a PCR-LA AC
power supply has been set to the relevant phase and that
the voltage reading shows the phase voltage.
• The letters “U-V” at the left of the voltage display area indi-
3-2 Operation Method
2P03-PCR-LA
cate that the voltage reading shows the line voltage of the
power supplies concerned (U-phase power supply only.)
(2) Display on the V-phase power supply control panel
• This control panels display the phase (V) indication,
voltage and current values, and LOAD level meter indication only.
(3) U-phase power supply control panel display
• During display of the power, the
symbol and the letters “U” and “V” may all appear at the left of the current
display area. This indicates the total power of the singlephase, three-wire.
• During a self-test, the
symbol and the letter “U” or
“V” may appear at the left of the voltage display area.
This indicates the self-test result for the AC power supply of the displayed phase.
2P03-PCR-LA
Operation Method 3-3
3.3
Switching between Phase Voltage and
Line Voltage Displays
The phase voltage display and line voltage display modes for display
are provided for single-phase, three-wire operation. The display mode
will be backed up and, when the POWER switch is turned ON, the
display mode used immediately before the POWER switch was
turned OFF will be selected.
1. Press the ESC key to select the Home Position.
2. Press the PHASE (SHIFT, 2) key.
Each time the PHASE key is pressed, the display mode is
switched between phase voltage display and line voltage display.
Phase Voltage Display Mode
• The
symbol and letter “U” or “V” at the left of the voltage display area indicate that the phase voltage display mode is selected.
Line Voltage Display Mode
• The letters “U-V” at the left of the voltage display area indicate that
the line voltage display mode is selected.
• Setting the voltage in the line voltage display mode allows line
voltage setting.
• The line voltage display mode disables the measurement of output
power.
• In the line voltage display mode, the V-phase voltage cannot be set
separately.
3.4
Output Voltage Setting
• Setting the voltage value to the U-phase power supply causes the
voltage of V-phase to be the same.
• Setting a voltage value to the V-phase power supply causes only the
V-phase voltage to change. However, changing the U-phase voltage
afterward causes the V-phase voltage to change to the same value as
the U-phase voltage.
• Setting of the V-phase voltage requires selection of the phase voltage display mode.
To set the voltage using the numeric keys in the OUTPUT ON condition, select the SET mode on the U-phase power supply.
3-4 Operation Method
2P03-PCR-LA
• Voltage setting in the phase voltage display mode sets the phase
voltage; set the voltage to 1/2 of the voltage required between lines.
• Voltage setting in the line voltage display mode sets the line voltage; set the voltage required between lines and leave as is.
• For the voltage setting methods and others, see the PCR-LA AC
Power Supply Operation Manual.
3.5
Limit Value Setting
• Voltage and frequency limit values should be set through the Uphase power supply control panel. The values set on the U-phase
power supply apply to V-phase.
• A voltage limit value is a limit on the phase voltage; a line voltage
limit operates at a value twice the phase voltage limit.
• The current limit value should be set on a phase basis.
3.6
Setting the Current or Power Display
Mode
• Each time the I MODE (SHIFT, I) key is pressed, the current and
power display modes change as follows:
Display of the U-phase power supply
RMS
PEAK
W (U-phase power)
W (total power of the
single-phase, three-wire)
Display of the V-phase power supply
RMS
PEAK
W (V-phase power)
W (V-phase power)
If the output power exceeds 10 kW, “E3” appears on the control
panel.
Example: For output power of 12 000 W, the control panel displays
the following.
W
Fig.3-1
2P03-PCR-LA
Example of 12 000-W Output Power Display
Operation Method 3-5
3.7
Memory Function
• The memory function is available only in the phase voltage display
mode.
• The U-phase power supply allows the voltage and frequency to be
stored in the memory.
• The V-phase power supply allows only the voltage to be stored in
the memory.
• The voltage read from the U-phase power supply memory is also
set to V-phase.
3.8
Key-lock Function
• The key-lock function operates in all phases separately.
3.9
Self-test Function
For single-phase, three-wire operation, the self-test is also conducted
from the U-phase power supply. It will be performed on both two AC
power supplies.
• The self-test operating procedure is as described in the “Alarm-type
checking procedure (SELF TEST)” in “4.6.1 Steps to be Taken in
the Event of an Alarm” in the PCR-LA AC Power Supply Operation Manual.
• If an alarm occurs, the U-phase power supply always generates an
intermittent buzzer tone and displays an “ALARM” indication.
• Some types of alarms also cause an intermittent buzzer tone to be
generated from and an “ALARM” indication to be displayed on the
V-phase power supply in which an alarm has occurred, in addition
to the U-phase power supply.
• Turning JOG causes the numbers in “Ad.” and “No.” to change, in
addition to the letter “U” or “V” at the left of the voltage display
area. Each letter indicates that the relevant phase power supply is
under self-testing.
• When the three power supplies are operated individually in parallel,
the numbers in “Ad.” for these power supplies are as shown in
Table 3-1.
• The numbers in “Ad.” shown in Table 3-1 are displayed on all of the
3-6 Operation Method
2P03-PCR-LA
U- and V-phase power supplies in the same way the letter “U” or
“V” at the left of the voltage display area determines which “Ad.”
number corresponds to which phase.
Table 3-1 Ad. Numbers for AC Power Supplies Operated in
Single-phase, Three-wire Output and in Parallel
PCR-LA power supply
model name
Master
Slave 1
Slave 2
PCR2000LA
1–2
11 – 12
21 – 22
PCR4000LA
1–4
11 – 14
21 – 24
PCR6000LA
1–6
11 – 16
21 – 26
*
U- or V-phase
The words “Master” and “Slave” in Table 3-1 represent the relationship
between the master and slave units in parallel connection.
When this device is combined with the Parallel Operation Driver to
perform parallel operation in a single-phase, three-wire system, the
number of power supplies to be operated in parallel is up to three for
each phase.
3.10 Functions Restricted during
Single-phase, Three-wire Operation
During single-phase, three-wire operation, the PCR-LA power supply
functions will be partially restricted. The following functions are not
available when the Single-phase, Three-wire Output Driver is
installed in PCR-LA AC power supplies.
• DC mode
• AC + DC mode
• Sensing
• Regulation adjustment
• Average voltage value measurement
• Average current value measurement
NOTE
• The power line abnormality simulation function available in
single-phase, three-wire operation differs partially in its specifications from that available in single-phase operation. For
more information on this, see “Chapter 11 Specifications” of
the PCR-LA AC Power Supply Operation Manual.
2P03-PCR-LA
Operation Method 3-7
3.11 Special Waveform Output
Use of the Remote Controller (RC04-PCR-LA) allows output of the
peak-clipped waveform of the line voltage and the phase voltage
waveform.
The setting range is 1.25 to 1.40.
If a value outside this setting range is set, no waveform will be output.
Line voltage waveform
U to V
Fig.3-2
3-8 Operation Method
Phase voltage waveform
U to N
V to N
“N” indicates neutral.
Peak-Clipped Waveform
2P03-PCR-LA
Chapter 4 Using the Remote Controller
Describes the functions and operation methods for single-phase,
three-wire operation of the PCR-LA power supplies when the Singlephase, Three-wire Output Driver is used together with the Remote
Controller (RC04-PCR-LA).
The Remote Controller is installed in the U-phase power supply.
This manual describes the procedures for the single-phase, three-wire
operation available functions that differ from those available when a
PCR-LA AC power supply and Remote Controller are operated in a
single-phase system. For information not provided in this manual, see
the RC04-PCR-LA Operation Manual. Moreover, before reading the
2P03-PCR-LA Operation Manual, always read through the RC04PCR-LA Operation Manual.
2P03-PCR-LA
Using the Remote Controller 4-1
4.1
Switching between Phase Voltage and
Line Voltage Displays
The phase voltage display and line voltage display modes are provided for single-phase, three-wire operation. The display mode will
be backed up, and when the POWER switch is turned ON, the display
mode used immediately before the POWER switch was turned OFF
will be selected.
1. Press the ESC key to select the Home Position.
2. Press the PHASE (SHIFT, 2) key.
This procedure alternately switches between the phase voltage
display and line voltage display.
Phase Voltage Display Mode
5 0 . 0 0Hz
F RQ
0 . 00A
I r ms
0 . 2V
V r ms
• The
symbol at the left of the voltage display area on the LCD
indicates that the phase voltage display mode is selected.
Line Voltage Display Mode
5 0 . 0 0Hz
F RQ
0 . 00A
I r ms
0 . 2V
V r ms
• The symbol at the left of the voltage display area on the LCD
indicates that the line voltage display mode is selected.
• Setting the voltage in the line voltage display mode allows line
voltage setting.
• The line voltage display mode disables measurement of output
power.
4-2 Using the Remote Controller
2P03-PCR-LA
4.2
Displaying the V-phase Current and Voltage
1. Press the ESC key to select the Home Position.
2. Press the MENU key. Each time this key is pressed, the display changes as follows.
For the phase voltage display mode:
U-phase current and voltage
V-phase current and voltage
Total power, total VA, and total power factor of the two phases
For the line voltage display mode:
U-to-V phase voltage
Power, VA, and power factor (no measured value is displayed)
4.3
Switching to the Current or Power Display Mode
• Each time the SHIFT key is pressed followed by the F3 key in the
Home Position, the current and power display mode changes as follows:
RMS
4.4
Ip
Iph
UWATT (U-phase power)
WATT (total power of the two phases)
Sequence Operation
In the sequence setting mode, items available in the DC or AC + DC
mode cannot be set. In addition, AC voltage (Vac) applies phase voltage setting.
2P03-PCR-LA
Using the Remote Controller 4-3
4.5
Harmonic Current Analysis Function
The Remote Controller (RC04-PCR-LA) connected to the U-phase
power supply allows harmonics in a load current to be analyzed in
each of the U and V phases.
For this operation, add the following step between steps 2 and 3 in
“4.4.3 Harmonic Current Analysis Function” in the RC04-PCR-LA
Operation Manual.
2a. Select the phase that performs harmonic current analysis.
Cu r r e n t
U
F1
F2
H a r mo n i c
V
F3
F4
F5
• Pressing the F1 key selects phase U and pressing the F3 key
selects phase V.
Thereafter, follow the procedure from step 3 onward as specified in
“4.4.3 Harmonic Current Analysis Function” in the RC04-PCR-LA
Operation Manual.
• The letter “U” or “V,” indicating the selected phase, appears at
the left of the first order component of the harmonic current
analysis result indicated on the display.
4-4 Using the Remote Controller
2P03-PCR-LA
Chapter 5 Using the GPIB/RS-232C Interface
Describes the functions and messages available when the Singlephase, Three-wire Output Driver and GPIB Interface (IB03-PCR-LA)
or RS-232C Interface are used in combination for single-phase, threewire operation of the PCR-LA power supplies.
When operating the single-phase, three-wire system PCR-LA AC
power supplies using the GPIB or RS-232C feature, use the GPIB
Interface installed in or the RS-232C Interface provided for the Uphase AC power supply.
This manual describes the functions available in single-phase, threewire operation that differ from those available when a PCR-LA AC
power supply and GPIB Interface or RS-232C Interface are operated
in a single-phase system. For information not provided in this manual, see the PCR-LA AC Power Supply Operation Manual. Moreover,
before reading the 2P03-PCR-LA Operation Manual, be sure to read
through the PCR-LA AC Power Supply Operation Manual.
2P03-PCR-LA
Using the GPIB/RS-232C Interface 5-1
5.1
Phase Voltage/Line Voltage Display Messages
The phase voltage and line voltage display modes for display are provided for single-phase, three-wire operation. The display mode will
be backed up and, when the POWER switch is turned ON, the display
mode used immediately before the POWER switch was turned OFF
will be selected.
VPHASE
Switches the voltage value display method to the phase voltage display mode.
Phase voltage setting is available in the phase voltage display mode.
Measurements of the power, apparent power, and power factor can be
performed in the phase voltage display mode.
Program message
• Syntax
Command message: VPHASE
VLINE
Switches the voltage value display method to the line voltage display
mode.
Line voltage setting is available in the line voltage display mode.
Program message
• Syntax
Command message: VLINE
5-2 Using the GPIB/RS-232C Interface
2P03-PCR-LA
5.2
Output Voltage Setting Messages
VSET/ACVSET
Sets or inquires about the phase voltage.
The VSET message and ACVSET message function in the same way.
In output voltage setting, all phase voltages are set to the same value.
These messages are enabled in the phase voltage display mode.
Program message
• Syntax
Command message: VSET <NR2>
ACVSET <NR2>
Query message:
VSET?
ACVSET?
• Program data
Same as that of single-phase operation
Response message
Returns a phase voltage set value in response to VSET?/
ACVSET?
LINEVSET
Sets or inquires about the line voltage.
This message is enabled in the line voltage display mode.
Program message
• Syntax
Command message: LINEVSET <NR2>
Query message:
LINEVSET?
• Program data
Twice the voltage set by the VSET/ACVSET message can be set.
Response message
Returns a line voltage set value in response to LINEVSET?
2P03-PCR-LA
Using the GPIB/RS-232C Interface 5-3
UVSET
Sets or inquires about the U-phase voltage.
This message is enabled in the phase voltage display mode.
Program message
• Syntax
Command message: UVSET <NR2>
Query message:
UVSET?
• Program data
Same as that of the VSET/ACVSET message
Response message
Returns a U-phase voltage set value in response to UVSET?
VVSET
Sets or inquires about the V-phase voltage.
This message is enabled in the phase voltage display mode.
Program message
• Syntax
Command message: VVSET <NR2>
Query message:
VVSET?
• Program data
Same as that of the VSET/ACVSET message
Response message
Returns a V-phase voltage set value in response to VVSET?
5-4 Using the GPIB/RS-232C Interface
2P03-PCR-LA
5.3
Output Measurement Messages
VOUT?
Inquires about a measured voltage value.
Program message
• Syntax
Query message:
VOUT?
Response message
Returns U- and V-phase voltages, delimited by a comma (,), in the
phase voltage display mode; returns U-V voltage in the line voltage display mode
Example: A response message for VOUT? in the phase voltage display mode is
115.3, 115.4
V-phase voltage
U-phase voltage
Example: A response message for VOUT? in the line voltage display mode is
200.0
U-to-V line voltage
IOUT?
Inquires about a measured current value.
Program message
• Syntax
Query message:
IOUT?
Response message
Returns U- and V-phase currents, delimited by a comma (,)
Example: A response message for IOUT? is
10.0, 10.1
V-phase current
U-phase current
2P03-PCR-LA
Using the GPIB/RS-232C Interface 5-5
WATT?
Inquires about a measured power value.
This message is enabled in the phase voltage display mode.
Program message
• Syntax
Query message:
WATT?
Response message
Returns the U- and V-phase power values and the total power,
delimited by a comma (,)
Example: A response message for WATT? is
200.0, 200.1, 400.1
Total power
V-phase power
U-phase power
PF?
Inquires about a measured power factor.
This message is enabled in the phase voltage display mode.
Program message
• Syntax
Query message:
PF?
Response message
Returns U- and V-phase power factors and the total power factor,
delimited by a comma (,)
Example: A response message for PF? is
0.60, 0.62, 0.61
Total power factor
V-phase power factor
U-phase power factor
The total power factor is obtained by PFtotal = Wtotal / VAtotal.
VA?
Inquires about a measured apparent power value.
This message is enabled in the phase voltage display mode.
Program message
5-6 Using the GPIB/RS-232C Interface
2P03-PCR-LA
• Syntax
Query message:
VA?
Response message
Returns U- and V-phase apparent power values and the total
apparent power, delimited by a comma (,)
Example: A response message for VA? is
333.3, 322.7, 656.0
Total apparent power
V-phase apparent power
U-phase apparent power
5.4
Harmonic Current Analysis Messages
Harmonic current analysis can be conducted in the same way as in
single-phase operation. However, it is necessary to use a message corresponding to the relevant phase to read out analyzed data in each
phase.
This section briefly describes the messages for each phase, and shows
the syntax only. The data format and set value of the program data on
each message and its response message are the same as those applied
in single-phase operation. For more information on these, see “9.4.4
Harmonic Current Analysis” of the PCR-LA AC Power Supply Operation Manual.
UCURHARMA?
Inquires about the value representing the harmonics component (in
current) of each degree in the U-phase.
Program message
• Syntax
Query message: UCURHARMA {<NR1>|ODD|EVEN|LOW|HIGH}?
VCURHARMA?
Inquires about the value representing the harmonics component (in
current) of each degree in the V-phase.
Program message
• Syntax
Query message: VCURHARMA {<NR1>|ODD|EVEN|LOW|HIGH}?
2P03-PCR-LA
Using the GPIB/RS-232C Interface 5-7
UCURHARMP?
Inquires about the percentage of a harmonic current value in the Uphase, obtained by regarding the current value of the basic wave component as 100 %.
Program message
• Syntax
Query message: UCURHARMP {<NR1>|ODD|EVEN|LOW|HIGH}?
VCURHARMP?
Inquires about the percentage of a harmonic current value in the Vphase, obtained by regarding the current value of the basic wave component as 100 %.
Program message
• Syntax
Query message: VCURHARMP {<NR1>|ODD|EVEN|LOW|HIGH}?
5-8 Using the GPIB/RS-232C Interface
2P03-PCR-LA
Chapter 6
Maintenance
Describes the maintenance procedures for the Single-phase, Threewire Output Driver.
Before Requesting a Repair
If a problem occurs in the 2P03-PCR-LA, check that problem in
accordance with Table 6-1. In addition, remove the Single-phase,
Three-wire Output Driver from the PCR-LA power supply concerned, and check whether the power supply alone operates normally.
If the power supply does not recover to normal status, contact Kikusui
distributor/agent to request repairs.
2P03-PCR-LA
Maintenance 6-1
Table 6-1 Troubleshooting Check Sheet
Problem
Check item
The PCR-LA power
supply display panel
displays “Err2” or
nothing, or retains
the version-number
indication.
Check whether each
of the two phase
cards is properly
inserted into a slot.
NO
Check whether the
drive-signal cable is
properly connected.
NO
Improper drive-cable
connection
(See 2.5.2 .)
NO
The PCR-LA power supplies cannot be operated
in a single-phase, threewire system unless both
of the two PCR-LA
power supply POWER
switches are turned ON.
YES
The phase voltage display
mode sets the phase voltage; the line voltage display mode sets the line
voltage.
NO
The cable connection to
the SENSING terminal
board is improper. This
disables line voltage measurement (see 2.6.2 ).
Check whether both
of the two PCR-LA
power supply
POWER switches
are ON.
Voltage different
Check whether the
from the set voltage display mode is coris generated (twice rect.
or 1/2).
The voltage display
becomes abnormal
in the line voltage
display mode.
6-2 Maintenance
Check whether the
SENSING terminal
board is properly
connected.
Results
Possible causes
Improper phase-card
installation
(See 2.5.1 .)
2P03-PCR-LA
Chapter 7
Specifications
Lists the specifications.
7.1
Specifications
The specifications given in this section are for comprehensive performance with the Single-phase, Three-wire Output Driver installed in
the PCR-LA AC power supplies to configure single-phase, three-wire
system power supplies. Other specifications comply with those of the
PCR-LA Series AC power supplies.
Input/output power capacity
(total power of single-phase,
three-wire)
[Power capacity of one PCR-LA power
supply] x 2
Input current capacity
[Current capacity of one PCR-LA power
supply] x 2
Output current capacity (phase
current for each phase)
[Current capacity of one PCR-LA power
supply] x 1
Phase difference of the output
phase voltage *1
Within 180° ± (0.4° + 5 μs)
Within (180° ± (0.4° + ƒ0 x 1.8 x 10-3°)
“ƒ0” indicates the output frequency.) *2
*1
*2
Phase difference between output voltages (phase voltages) when each
phase is checked from the neutral point in the condition in which no
phase difference is varied (180° between phases)
Examples of the results in which the noted expression was converted
into an angle at a specific frequency
Within 180° ± 0.5° (for 60-Hz output)
Within 180° ± 1.2° (for 400-Hz output)
2P03-PCR-LA
Specifications 7-1
Power line Abnormality Simulation (Specifications Different from
Those Achieved in Single-Phase Operation)
Item
T1
T3
Resolution
Setting accuracy
0 deg–360 deg
1 deg
±1.5 ms
x 1 ms
0 ms–999 ms
1 ms
±1 ms
x1
0 ms–999 ms
1 ms
± (0.1 % + 1 ms)
V (T3)
7.2
Setting range
x 1 deg
0 to rated voltage
(phase voltage setting)
Dimensions
MAX145
MAX20
2P03-PCR-LA
121
J1
1PHASE-3WIRE DRIVER
121
25
Unit: mm
Fig.7-1
7-2 Specifications
2P03-PCR-LA Outer Dimensions
2P03-PCR-LA
SINGLE-PHASE, THREE-WIRE OUTPUT DRIVER 2P03-PCR-LA OPERATION MANUAL