PLV®·102b
PORTABLE VOLUME VENTILATOR
,
£
RESPIRONICS
PLV-102b
When you need a portable volume ventilator capable of ventilating a wide range of patients, you need the
PLV-102b. The PLV-102b uses a proven microprocessor-controlled, piston-driven ventilator platform to deliver
volume ventilation in a range of settings. Its features address the needs of many patients who require volume
ventilation and portability or transport in the sub-acute care, nursing home or home setting.
FEATURES
Portable· The PLV-102b is well suited for transport and portable applications due to its small size and light
weight. The ventilator is capable of operating from any of three power sources: 120 volts AC, an internal 12 volt
DC battery and an external 12 volt DC battery, all of which are automatically switched by the PLV-102b.
Controls • The PLV-102b provides three modes: Control and Control-Assist, each of which can be used wit~
optional Sigh, and the SIMV mode. The sensitivity of the PLV-102b is adjustable from -6 to + 18 cm H20, which allows
for use with PEEP up to 20 cm H20. The sensitivity is automatically and continuously corrected for barometriJ
pressure changes.
I
Alarms and Displays· Alarms are provided for Low Pressure, High Pressure, Apnea, Inverse I:E ratio, Increase
In11l1rlllflJ rum IlIlU runur OUUI UU iOUUOOI TIIOI 0 UI U UIOU Oill UlrrUi Dill UIUlIlI UUUUU IU UUUIUI III lruUUIOUllUUlIllY
Large, backlit LeDs display Tidal Volume, Respiratory Rate, I:E Ratio and Inspiratory Flow Rate.
POWER SOURCES
FRONT PANEL CONTROLS
Mode Selection:
CONTROL
CONTROL + SIGH
ASSIST/CONTROL
ASSIST/CONTROL + SIGH
SIMV
Tidal Volume:
0.05 to 0.20 +/- 0.02 liters
0.20 to 3.00 liters +/- 10%
Sigh Volume:
150% of Set Tidal Volume
Sigh Frequency:
1 Sigh Every 100 Tidal Breaths
Breath Rate:
2 to 35 BPM +/- 0.5
36 to 40 BPM +/- 2
Inspiratory Flow Rate:
10 to 120 LPM +/- 10%
Battery Test Switch:
Test Internal/External Battery
Main Power Switch:
On, Off/Recharge
DC Circuit Breaker:
Push/Pull Actuation
AC Circuit Breaker:
Push/Pull Actuation
Airway Pressure Limit:
5 to 100 cm H20 +/- 5
Sensitivity:
+18 cm H20 +/- 1
to at least -6 cm H20
Low Pressure Alarm:
2 to 40 cm H20 +/- 2
41 to 50 cm H20 +/- 5
Manual Sigh:
Push to Activate One Breath
30 Second Alarm Silence:
Push to Activate
FRONT PANEL VISUAL INDICATORS
120 VAC 50/60 Hz
12 VDC Internal Battery
12 VDC External Battery
PHYSICAL DIMENSIONS
Overall Dimensions:
9" H x 12.25" W x 12.' 5" D
(22.9 cm x 31.1 cm x 1.1 cm)
Weight:
28.2 lb. (12.8 kg)
ALARMS
Low Pressure:
Visual and Continuo s Audible
High Pressure:
Intermittent Audible
Apnea:
Visual and Continuo s Audible
Inverse I:E Ratio:
Visual Warning
Increase Inspiratory Flow:
Visual Warning
Low Internal Battery:
Visual and Continuo s Audible
Low External Battery:
Visual and Continuo s Audible
Reverse Battery Connection:
Continuous Audible
Power Failure:
Continuous Audible
Unit Malfunction:
Intermittent Audible
Power Source Change:
Continuous Audible
ORDERING INFORMATION
Manual Sigh:
Yellow Indicator Light
Description
Part Number
Tidal Volume:
Digital Display
PLV-l02b Ventilator
1003111
Rate: Patient Breaths Per Minute:
Digital Display
I:E Ratio:
Digital Display
Inspiratory Flow Rate:
Digital Display
Increase Inspiratory Flow Rate:
Red Indicator Light
120V AC Power:
Green Indicator Light
Internal Battery Source:
Amber Indicator Light
External Battery Source:
Yellow Indicator Light
Patient Airway Pressure:
-10 to +100 cm H20
Assist/Spontaneous:
Green Indicator Light
15-Second Delay (Low Pressure Alarm):
Green Indicator Light
30-Second Alarm Silence:
Yellow Indicator/Push Button
Alarm Code Display (in Front Window):
Digital Display
!&1
Global Headquarters:
Pittsburgh, PennsylvaniaUSA·
412-131-2100
Customer Service: 1-800-345-6443
RESPIRONICS@
Respironics Europe: +33-(0) 1-55-60- 9-80
www.resplronlcs.com
Respironics Asia Pacific: +852-234-3
PLV-102b
of U.S. patents
ventilators
are the subiscts
#4,493,614,
#4,617,637.
#4.58
2-18
967,
~IIUUIlI~rp~IIDlligua AIIDlur~lgrrp!l~mlllU I!! r~QI!I~r~DIr!O~rrr!rl01r(~!plrUIU
CAUTION:
©2000
U. S. federal
Respironics. Inc.
law restricts
this device
to sale by or on the order
of a physici
n.