Sapphire Autoclave - Rodwell Autoclave

Sapphire Autoclave
Specifications for a Sapphire 330, 440 & 550 E & DS model
Design Features
 330, 440 & 550 litres
 MP25 Controller
 Square Chamber
† Automatic mechanically hoisted sliding door.
† Fixed Seal
† Space-saving design
† Accelerated water cooling by means of a water jacket.
Optional Accessories
 † Loading Trolley
 † Range of cycle recording systems
 † Drain Line Condenser for plastic drains
 † Scale prevention systems
 † Liquid Ring Vacuum air removal system
 † Air Ballasting option
 † Additional shelves & baskets
 † Steam Jacket drying option
 † Air removal HEPA filter (for CAT3 environments)
 † Anti Lime scale system
Capacities and dimensions (approx)
Model
Sapphire 330
Sapphire 440
Sapphire 550
Autoclave Weight
650 kg
750 kg
850 kg
Export Case Size
1140w x 2200h x 1690d
1140w x 2200h x 1690d
1140w x 2200h x 1690d
Overall Dimensions
980w x 1930h x 1530d
980w x 1930h x 1530d
980w x 1930h x 1530d
Chamber Dimensions
670w x 670h x 745d
670w x 670h x 936d
670w x 670h x 1099d
Chamber Capacity
330 Litres
440 Litres
550 litres
Water Consumption (Steam 45 litres
generator / typical per cycle)
45 litres
45 litres
1L Media Bottles
72
96
108
2L Media Bottles
40
48
54
500ml Media Bottles
98
140
154
NHS Discard Boxes
H1012xW320xL320 (8)
H250xW320xL320 (8)
H250xW320xL320 (12)
Trolley
H671xW625xL1723
H671xW625xL1723
H671xW625xL1723
Sapphire 330
Sapphire 440
Not to scale
Sapphire 550
Human Scale
Safe Design Operating Limits
Model
Sapphire 330
Sapphire 440
Sapphire 550
Max working temperature
138°C
138°C
138°C
Min working temperature
7°C
7°C
7°C
Max ambient temperature
28°C
28°C
28°C
Min ambient temperature
11°C
11°C
11°C
Max hydraulic pressure
4.14 Bar
4.14 Bar
4.14 Bar
Max working pressure
2.62 Bar
2.62 Bar
2.62 Bar
Max working vacuum
Minus 1 Bar
Minus 1 Bar
Minus 1 Bar
Boiler hydraulic pressure
6 Bar
6 Bar
6 Bar
Boiler working pressure
4 Bar
4 Bar
4 Bar
Site Requirements (E Version)
Electrical site requirement: Switched RCD required
Model
Sapphire 330
Sapphire 440
Sapphire 550
Drain Requirements
150°C 35mmØ Copper
150°C 35mmØ Copper
150°C 35mmØ Copper
Voltage
415v
415v
415v
Phase Requirement
3ph/N 50/60 Hz
3ph/N 50/60 Hz
3ph/N 50/60 Hz
Power Consumption
18kw
24kw
24kw
Fuse Rating
40 amp
45 amp
45 amp
Site Requirements (DS Version)
Electrical site requirement: Switched RCD required
Model
Sapphire 330
Sapphire 440
Sapphire 550
Drain Requirements
150°C 35mmØ Copper
150°C 35mmØ Copper
150°C 35mmØ Copper
Mains Steam Supply
55 lb p/h @ 2.67 Bar
70 lb p/h @ 2.67 Bar
70 lb p/h @ 2.67 Bar
Voltage
415v
415v
415v
Phase Requirement
3ph/N 50/60 Hz
3ph/N 50/60 Hz
3ph/N 50/60 Hz
Power Consumption
1.2kw
1.2kw
1.2kw
Fuse Rating
6amp
6amp
6amp
Option Requirements
 Air Supply at 38 psi, 7 cfm required for Air Ballasting option
 Steam Jacket drying option requires liquid ring vacuum option
 Drain requirement of 150°C, 35Ømm copper pipe at floor level.
General Requirements
 † Air supply at 100psi 2cfm for pneumatic sliding door (+13amps for DS machines).
 Sodium Chloride must not exceed a level greater than 500 ppm.
 † Use regular mains water only. DO NOT USE distilled or RO water
 DS Version requires a mains steam supply of 38psi 18kw 55lbs/hour (330) & 38psi 24kw 75lbs/hour (440 & 550).
Typical Installation
Complying Standards
 PED Cat 2 Module B+D Notified Body: British Engineering No. 0040
 Harmonised Standards applied: EMC BS EN 61326
 Community directives applied : Pressure Equipment Directive PED 2014/68/EU (formerly PED 97/23/EC)
 Designed in accordance with PD5500 (formerly BS5500) & BS2646
 Chamber, Door & Jacket constructed of stainless steel BS-EN 10028-7:316
 ISO 9001: 2008
 CE Marked
Information is subject to change to suit different options added. This information is correct as of 7th September 2017, E & OE
The Rodwell Autoclave Company, Bentalls, Pipps Hill Industrial Estate, Basildon, Essex SS14 3SD, United Kingdom
Tel: +44 (0)1268 286646 Fax: +44 (0)1268 287799
E-mail: autoclavesales@rodwellgroup.co.uk
Online: www.rodwell-autoclave.com