VISION TOF-Cuff NMT
ECG 7 LEADS
NIBP Non Invasive Blood Pressure
ST SEGMENT
SPO2 PULSE OXIMETRY
RESPIRATION
TEMPERATURE (2)
ARRHYTHMIA Detection:
- Asystolia
- Bigeminy
- Ventricular Tachycardia
- Tachycardia
- Trigeminy
- Ventricular Rhythm
- Bradycardia
- Couple of PVC
- Run of PVC
NMT
Neuro Muscular Transmission
Easy monitoring through BP cuff
with built-in electrodes
Vision TOF-Cuff
Technical Data
GENERAL
Dimensions:
310 (W) x 210 (L) x 260 (H) mm
Weight: 7 kg
Power Supply: 85-264 V; 47-63 Hz
AC input power: 45 to 100 VA
Battery: 4 hours autonomy
Built-in charger
Environmental conditions:
Operating temperature:
0 ºC to 40 ºC; 32 ºF to 104 ºF
Relative humidity:
5% to 90% non-condensing
Display:
TFT colour display.
Up to seven (7) traces on screen.
Size: 211 x 158 mm
Diagonal 264 mm / 10.4 inches
Resolution: 640 x 480 pixels
Sweep speed : 12.5 / 25 / 50 mm/s
Recorder: 4 channel (optional).
Safety Standard:
Complies with IEC 601 standards.
Patient leakage current <100 µA
Class I, CF and BF.
Measurement procedure:
Oscillometric method.
Cuffs: Different sizes
Measurement time:
Typical: 25 to 45 s
Maximum: 120 s (Adult)
90 s (Neonatal)
Adult / Child:
Range: 10 to 250 mmHg
Overpressure limit: 300 mmHg
Neonate:
Range: 10 to 130 mmHg
Overpressure limit: 150 mmHg
Pulse rate:
Range: 40 to 250 bpm
Resolution: 1 bpm
Accuracy: ±3 %
Operation Mode
Manual
Automatic 1 to 120 min.
(minute by minute)
Amplifier: Fully isolated, defibrillator
protected > 5 kV; recovery time < 5 s;
RF noise and monitoring filters.
Leads:
I, II, III, V, aVL, aVF, aVR.
Freeze and memory:
ST Segment. Frozen ECG,
12 seconds for analysis and
measurement.
Heart rate:
Range: 20 to 300 bpm
Accuracy: ±2 %
CMRR:
> 115 dB in monitor mode
> 90 dB in diagnostic mode
Signal features:
Diagnostic mode: 0.05 to 100 Hz
Monitor mode: 0.5 to 25 Hz
O. R. mode: 0.5 to 25 Hz
Programmable gain: 1 to 8 mV
Pacemaker detection and rejection:
Pulse duration: 0.1 to 2 ms
Pulse amplitude: ±2 to ±700 mV
Pulse Oximetry SpO2
Respiration
Amplifier: Fully isolated, defibrillator
protected > 5 kV.
Measurement:
Impedance Method
Range: 0 to 150 rpm
Resolution: 1 rpm
Accuracy: ± 2 rpm
Apnoea Alarm.
Signal features:
Maximum impedance: 2 kOhms
Sensitivity range: 0.2 to 15 Ohms
Excitation current: < 400 µA
Frequency: 62.5 kHz
Sweep speed: 6.25 mm/s
Temperature
Amplifier: Fully isolated, defibrillator
protected > 5 kV.
Automatic rejection against ESIS
through real time signal analysis.
Amplifier: Fully isolated, defibrillator
protected > 5 kV.
Improved movement artifact
rejection algorithm.
Saturation measurement:
Range: 0 to 100 %
Resolution: 1 %
Accuracy: 70 to 100%: ±2 %
Pulse rate:
Range: 30 to 250 bpm
Resolution: 1 bpm
Accuracy: ±3 %
Display:
Graphical peripheral pulse.
Method
TOF (Train of Four)
ST (Single Twitch)
PTC (Post Tetanic Count)
Mode · Autonomous relaxation monitoring program during surgery
· Manual Dedicated Key
· Cyclic TOF: 30s or 1, 2, 5, 8, 10, 15 min.
ST: 1 sec. and 10 sec.
Stimulation Current:
1 to 60 mA, Adult
1 to 35 mA, Child
1 to 15 mA, Neonate
(steps of 1 mA)
Pulse Width: 100, 200, 300 µs
Voltage:
500 V (max.)
Pulse Energy: 50 mJ
Graphic Trend of last 50 stimulations
TOF
Intensity bar of 4 pulse responses
Numeric data in % or count
ST
Numeric data in %
PTC
Figure of Post Tetanic Count
Phone: +34 913 042 011
www.rgb-medical.com
Temperature channels: 2
Temperature measurement:
Range: 15 to 45 ºC; 59 to 113 ºF
Resolution: 0.1 ºC; 0.2 ºF
Accuracy: ±0.1 ºC; ±0.2 ºF
Use of oesophageal, rectal or skin
probes.
Ordering Information
NMT Neuro-Muscular Transmission
RGB Medical Devices
Alfonso Gómez, 42
28037 Madrid SPAIN
Use: Alarm limits, gains and scales
settings directly on screen.
Non-Invasive Blood Pressure
ECG
Display
Alarms: Optical and acoustic for all
parameters. Three priority levels;
clinical, technical and procedural.
Trends:
Up to 72 hours for all parameters;
graphical and tabular display.
Standard Interface:
RS-232-c serial channel for
Central Station
1800 NMT:
8051
8151
8011
8111
4701
4040
4020
6110
6235
6225
6220
6245
NMT / ECG / HR / Resp / SpO2 / NIBP / TEMP(2)
ECG main cable, 5-lead
ECG patient cable, 5-lead
ECG main cable, 3-lead
ECG patient cable, 3-lead
SpO2 extension cable
SpO2 finger sensor, adult
SpO2 finger sensor, pediatric
NIBP extension tube
NMT / NIBP cuff, adult (23-31cm)
NMT / NIBP cuff, pediatric (13-18cm)
NMT / NIBP cuff, infant (8-13cm)
NMT / NIBP cuff, big adult (31-42cm)
2000 Temperature probe, skin
2010 Temperature probe, central
Specifications subject to change without notice