Incubators
PERFECTLY COORDINATED. PERFECTLY CONTROLLED.
INCUBATOR I
CO2 INCUBATOR INCOmed
Tel.
0320 INCUBATOR
266 171 ICP
COMPRESSOR-COOLED
E-mail Laboratorium@dijkstra.net
www.dijkstra.net
PELTIER COOLED INCUBATOR IPP
STORAGE COOLED INCUBATOR IPS
100% ATMOSAFE. MADE IN GERMANY.
|| www.atmosafe.net
www.memmert.com
www.memmert.com
www.atmosafe.net
www.dijkstra.net
2
Stable.
Safe.
Sensitive.
Memmert incubators for microbiology.
Energy efficient, precise, 100% AtmoSAFE.
Even slight temperature deviations in the working chamber
of an incubator may cause a test to fail. For this reason,
the heating and control system of Memmert incubators
are perfectly adapted to each other. During heating up and
cooling down as well as in running operation, all appliances
precisely keep the desired parameters within the smallest
tolerance limits. Not only at one measuring point, but in
the entire working chamber. Each individual Memmert
incubator is tested according to the strict requirements
of DIN 12 880: 2007-05 and is equipped with a maximum
of safety functions. Each individual Memmert incubator
is 100% AtmoSAFE.
I NHALT S V E R Z E I C H N I S
INCUBATOR I
PAGE 4 TO 5
TECHNICAL DATA
PAGE 6 TO 7
Microbiological tests, colony counts, virology, toxicology
CO2 INCUBATOR INCOmed
PAGE 8 TO 9
TECHNICAL DATA
PAGE 10 TO 11
Cultivation of cells or tissue, in-vitro fertilisation, gene expression
COMPRESSOR-COOLED INCUBATOR ICP
PAGE 12 TO 13
TECHNICAL DATA
PAGE 14 TO 15
Microbiological tests, colony counts, virology, toxicology,
cultivation above and below room temperature, alternate stability tests
PELTIER COOLED INCUBATOR IPP
PAGE 16 TO 17
TECHNICAL DATA
PAGE 18 TO 19
Protein crystallography, microbiological tests, colony counts,
virology, toxicology, cultivation above and below room
temperature, alternate stability tests
STORAGE COOLED INCUBATOR IPS
PAGE 20 TO 21
TECHNICAL DATA
PAGE 22 TO 23
Microbiological tests, cultivation above and below room temperature
OPTIONS AND ACCESSORIES
PAGE 24 + 25
Available for all products of Generation 2012 and 2003
GENERATION 2012
Features of model variants SingleDISPLAY and TwinDISPLAY
AtmoCONTROL software
PAGE 26 + 27
3
4
30 TO 750 L ITRES
N ATU RAL CONVECT ION
FORCED A IR CIRCULAT ION
Incubator IN and IF with SingleDISPLAY
Incubator INplus and IFplus with TwinDISPLAY
Natural convection or forced air circulation
AtmoCONTROL software
Model sizes:
30 / 55 / 75 / 110 / 160 / 260 / 450 / 750
+30 °C to +80 °C
INCUBATOR I Memmert incubators I are at home in the world
of research, medicine, pharmaceutics and food technology. Organic
chamber loads require gentle heating. For this reason, the heating
and control system are especially optimised for low temperatures
of up to +80 °C. To prevent temperature overshoots, temperature
is increased within a very narrow control range and kept exactly at
the setpoint value. As required, the models IN with natural convection or IF with forced air circulation are available.
OPT I MUM UNI F OR MI T Y
S AF E I N C U BAT I O N
As little air circulation as possible in the incubator
Forced air convection may destroy the protective layer from moist air that is generated during
incubation over the samples. This would lead to dehydration of the culture. In a Memmert
incubator, the perfect combination of all-round surface heating and temperature control system
ensures that incubation generally takes place without forced air circulation. Provided the chamber
is fully loaded and forced air circulation is required, it can be precisely adjusted in 10% steps
from 0 to 100 %.
Sterilisation
The chamber of the incubators INplus and IFplus, including all installations and sensors can be
sterilised at +160 °C in a 4-hour programme to guarantee optimum hygiene.
Fresh air is preheated
Temperature deviations caused by fresh air can influence sample characteristics or prolong
drying. In Memmert incubators, the fresh air is therefore fed through a pre-heating chamber
and seamlessly introduced into the working chamber.
100
90
80
70
Pre-warmed air
entering working
chamber
60
50
40
30
20
Temperature [°C]
Pre-heating chamber below
the working chamber
Air supply from outside
Intended use as a medical device:
The intended use of incubators INplus/IFplus is warming of rinsing
solutions and infusions. IF and IFplus are also accredited for warming
non-sterile cloths and blankets.
T E C H N I C AL DATA
OV ERV I EW
INCUBATORS I
INCUBATORS I
according to 12 880: 2007-05, EN 61010-1 (IEC 61010-1), 61010-2-010
Standard equipment
Interior:
Stainless steel, material 1.4301 (ASTM 304) with all-round
deep-drawn ribs to integrate the large-area heating with
ceramic-metal sheath
Stainless steel grids
(size 30 and 55: 1, size 75 to 750: 2)
Housing:
Textured stainless steel, rear zinc-plated steel, intuitively
operated SingleDISPLAY or TwinDISPLAY (TFT colour display)
with Touchscreen; inner glass door, outside fully insulated
stainless steel door (from size 450 two leaves )
56
(93)
A
C
292
256
Internals:
F
D
(93)
39
E
exhaust air
Ø 57
Fresh air:
Admixture of pre-heated fresh air by electronically
adjusted airflap
Connection:
Mains cable with plug
Installation:
4 feet; size 450 and 750 on lockable castors
Interfaces:
Ethernet
USB
33
B
6
(only TwinDISPLAY)
fresh air
LAN
Model sizes/Description
Stainless steel interior
Textured stainless steel
exterior
Further data
Packing data
Volume
30
55
75
110
160
260
450
750
approx. l
32
53
74
108
161
256
449
749
Width
(A)
mm
400
400
400
560
560
640
1040
1040
Height
(B)
mm
320
400
560
480
720
800
720
1200
Depth (less 39 mm for fan)
(C)
mm
250
330
330
400
400
500
600
600
9
8
14
Stainless steel grids (standard equipment)
number
1
Max. number of grids
number
Max. loading per grid
kg
Max. loading of chamber
kg
60
80
120
175
210
mm
585
585
585
745
745
824
1224
1224
3
2
4
6
5
8
30
300
Width
(D)
Height (size 450, 750 with castors)
(E)
mm
707
787
947
867
1107
1186
1247
1726
Depth (with door handle), door handle + 56 mm (F)
mm
434
514
514
584
584
684
784
784
approx. W
800
1000
1250
1400
1600
1700
1800
2000
Power consumption at 230 V/115 V, 50/60 Hz
Working temperature range
°C
min. 5 K (IN/INplus) 10 K (IF/IFplus) above ambient temperature up to +80
Setting temperature range
°C
+20 to +80
Setting accuracy
K
0.1
Net weight
approx. kg
44
55
65
74
96
110
170
215
Gross weight (packed in carton)
approx. kg
55
67
78
88
112
166
240
290
Width
approx. cm
69
70
70
85
85
93
133
133
Height
approx. cm
96
104
110
114
136
153
145
192
Depth
approx. cm
66
73
73
79
79
92
105
105
IN30
IN55
IN75
IN110
IN160
IN260
IN450
IN750
Order No. Incubators
I
= Incubator
N = Natural convection
F
= Forced convection
plus = Model with TwinDISPLAY
IN30plus IN55plus IN75plus IN110plus IN160plus IN260plus IN450plus IN750plus
IF30
IF55
IF30plus
IF55plus
IF75
IF110
IF160
IF260
IF450
IF750
IF75plus IF110plus IF160plus IF260plus IF450plus IF750plus
T EC HNI C AL DATA
Optionen
30
55
75
110
Interior lighting (up to size 260: 15 W, sizes 450/750: 2 x 15 W)
I N C U BATO R S I
260
450
750
260
450
750
R0
Interior socket can only be ordered with limited temperature range up to max. +70 °C,
ampacity 230 V/2.2 A, can be switched off with the On/Off switch, cannot be
switched individually, moisture tight IP68 (only with SingleDISPLAY)
Entry port, 23 mm clear diameter, for introducing connections at the side,
can be closed by flap and silicone stopper, standard positions
left centre/centre
left centre top
right centre/centre
right centre top
Other port, 23 mm clear diameter, can be closed by flap,
in special positions (please, state location)
left
right
rear
Other port, 14 mm clear diameter, can be closed by flap,
in special positions in the back wall (please, state location)
Other port, 38 mm clear diameter, can be closed by flap,
in special positions in the back wall (please, state location)
Other port, 57 mm clear diameter, can be closed by flap,
in special positions in the back wall (please, state location)
Other port, 100 mm clear diameter, can be closed by flap,
in special positions in the back wall (please, state location)
Other port, 120 mm clear diameter, can be closed by flap,
in special positions in the back wall (please, state location)
4 – 20 mA current loop interface (0 to 90 °C 4 to 20 mA)
Temperature controller, actual value
Temperature of a Pt100 sensor positioned flexibly in chamber
(max. 1 SingleDISPLAY, max. 3 TwinDISPLAY)
R3
F0
F1
F2
F3
F4
F5
F6
D6
F7
F8
F9
D7
V3
V6
Fan speed monitoring – optional only for IFplus
V4
Works calibration certificate for 3 temperatures: +37 °C, +52 °C, +70 °C
Works calibration certificate (measuring point chamber centre) at +37 °C
standard equipment
Accessories
160
OV ERV I EW
D00126
30
55
75
110
160
Stainless steel grids (standard equipment)
E28884
E20164
E20165
E28891
E20182
Perforated stainless steel shelves
B29727
B03916
B00325
B29725
B00328
Stainless steel slide-in drip tray, 15 mm rim (may affect the temperature distribution)
E02070
E02072
E02073
E29726
E02075
Stainless steel bottom drip tray, 15 mm rim (may affect the temperature distribution)
B04356
B04358
B04359
B29722
B04362
Wall bracket (tubular frame for wall mounting)
B29755
Guarantee extension by 1 year
B29756
B29757
GA1Q5
B29758
B29759
–
GA2Q5
7
8
M E D I C A L DEV I C E C L ASS IIA
A LL-ROUND H E AT ING ON 6 S ID E S
CO2 Incubator INCOmed
“Celsius” standard software
Model sizes:
108 / 153 / 246
+20 °C to +50 °C
Humidity 88 to 97 % rh
CO2 content up to 10 %
Humidity 40 to 97 % rh (humidity module)
CO2 content up to 20 % (CO2 module)
O2 content between 1 and 20 % (O2 module)
CO2 INCUBATOR INCOmed For cell cultivation and especially for in-vitro fertilisation, the precision and reliability of CO2 incubators are of crucial importance. During cultivation, the slightest
deviation in the CO2 atmosphere, temperature or humidity can influence cell development. For this reason, Memmert has subjected
its CO2 incubators to a comprehensive evaluation process for their
recognition as medical devices. The classification as class IIa medical
device confirms that all Memmert CO2 incubators INCOmed comply
with the basic safety requirements of the European Medical Devices
Directive 93/42/EEC. The interior chamber including all installations
and sensors can be sterilised at +160 °C in a 4-hour programme.
7 MODULES
S T ER I LI S AT I ON C Y C LE
AC T I V E HUMI D I T Y C O N T R O L
Customised models for every application
As much function as needed, as much customisation as possible. Put together your
own customised INCOmed from 7 additional modules.
•
COMFORT MODULE: Two gas connections with quick release connectors,
automatic switch-over between gas bottles (a combination of the comfort
module and the O2 module is not possible, since the N2 introduction takes
place via a second gas bottle connection)
•
HYGIENE MODULE: Electropolished, seamless laser-welded chamber
•
COMMUNICATION MODULE: USB interface, „Celsius“ standard software
for programming and protocol logging, ring memory, printer port
•
CO2 MODULE: Extended CO2 range from 0 to 20 %
•
O2 MODULE: Control of oxygen concentration by introducing nitrogen, adjustment
range from 1 % to 20 % O2 (combination of comfort module or Premium module
and O2 module not possible, since N2 is introduced via a second gas bottle connection)
For applications with a set O2 value of less than 10 %, the humidity module is highly
recommended
•
PREMIUM MODULE: Includes comfort, hygiene, communication and CO2 module
•
HUMIDITY MODULE: Active microprocessor humidification and dehumidification control (40 - 97 % rh). Recommended for applications with set O2 values of less than 10 %
Homogeneity in the chamber
Heating the working chamber from all six sides along with the electronic humidity control
system and turbulence-free ventilation is decisive for temperature and humidity distribution.
An aluminium thermal conduction layer supports homogeneity in the chamber and serves as a
heat accumulator if there is a temporary power failure.
Short recovery times thanks to active humidity control
The INCOmed standard model features a humidity limiting system to reduce water tray generated
maximum relative humidity inside the chamber from 97 % down to 88 %. To achieve an increase
in usable volume, optimum hygiene and short recovery times after opening the door, the optional
humidity module, an active humidification system with an adjustment range of 40 % to 97 % rh
can be integrated, introducing sterile hot steam into the air stream.
9
10
T E C H N I C AL DATA
OV ERV I EW
CO 2 -INCUBATORS INCOme d
CO2-INCUBATORS INCOmed
with automatic sterilisation
according to 12 880: 2007-05, EN 61010-1 (IEC 61010-1), 61010-2-010
supply cable
75
Mains cable with plug
Installation:
4 adjustable feet
Interfaces:
Optional with Communication
resp. Premium module
printer interface
opening
for test
purposes
USB
inner
glass door
Pressure
balancing
pipe
B
Connection:
25
CO2 resp. N2
In2
12
Textured stainless steel, rear zinc-plated steel,
aesthetic functional glass-stainless steel operating
panel with multifunction display and input module;
fully insulated heated stainless steel door and
inner glass door
38
202
Perforated stainless steel shelves (size 108: 2,
sizes 153/246: 3, stainless steel water dishes
(sizes 108/153: 1 (full width), size 246: 2 (half width)
Housing:
F
C
CO2
In1
8 mm (minimum height)
E
Internals:
D
A
75
76
Standard equipment
Interior:
Stainless steel, material 1.4301 (ASTM 304),
deep-drawn
parallel interface
USB interface
adjusting screw
optional
Ethernet
LAN
Model sizes/Description
Stainless steel interior
Volume
153
246
approx. l
108
153
246
Width
(A)
mm
560
480
640
Height
(B)
mm
480
640
640
Depth (less 25 mm for fan)
(C)
Provision for grids or shelves Half width / Full width
Stainless steel exterior
108
mm
400
500
600
number
–/4
–/6
2x6/6
Width
(D)
mm
710
630
790
Height (variable through adjustable feet)
(E)
mm
778
938
938
Depth (without door handle, depth of door handle 38 mm)
(F)
mm
550
650
750
Fully insulated, heated stainless steel door
왏
Extra internal glass door
왏
Ventilation
Uniform atmosphere and temperature distribution through enclosed
non-turbulent ventilation system, fully covered by the sterilisation process
왏
Temperature
Electronic microprocessor temperature controller with Pt100 and auto-diagnostic system
왏
Temperature sensors Pt100 Class A in 4-wire circuit for uninterrupted
operation on failure of one Pt1100 with warning indication
Temperature range
(during sterilisation the temperature is fixed at +160 °C – set value)
double
°C
from +20 to +50
(at least 8 K above ambient)
Temperature fluctuations with time (to DIN 12 880: 2007-05)
K
≤ ± 0.1
Temperature variation in chamber at +37 °C (to DIN 12 880: 2007-05)
K
≤ ± 0.3
Sterilisation
STERICard for automatic chamber sterilisation cycle 4 h at +160 °C
(not for sterilising the load)
왏
CO2
Digital electronic CO2 control with autozero, NDIR system, with auto-diagnostic system
and acoustic fault indication, barometric pressure compensation
왏
Humidity
Setting accuracy
% CO2
0.1
Adjustment range
% CO2
0 to 10
Standard water dishes
Monitor
왏
Capacitive humidity sensor (sterilisable)
number
1
2
Adjustable humidity limit control (88 – 97 %) incl. digital indication and auto-diagnostic
system with visual and acoustic fault indication (air supply via sterile filter) ensures rapid
reaching of set humidity and short recovery times while avoiding condensate formation
왏
Microprocessor temperature monitor acting as overtemperature protection
(protection class 3.1), with Pt100 incorporating fault diagnostics with visual
and audible alarm
왏
Digital over- and undertemperature monitor
왏
Temperature monitoring band automatically linked to the setpoint (ASF)
왏
T EC HNI C AL DATA
Model sizes/Description
OV ERV I EW
C O 2 -I NC UBATO R S I N C O m e d
108
153
Relay for reliable heating cut-off in case of fault
왏
Mechanical temperature limiter (TB)
왏
Audible alarm: Over- and undertemperature, over-CO2 and empty gas cylinder,
open door, underhumidity and empty water tank (with optional Humidity module)
왏
Timer functions
Real-time/weekly programmer with group function (e.g. Monday - Friday)
왏
Setup
Calibration (no separate PC required), Temperature: 3-point calibration on controller, CO2:
3-point calibration at 5 %, 7 % and 10 %, Auto-zero-function of NDIR CO2-sensor after
every sterilisation and cyclically every 24 h, Humidity: 2-point calibration at 20 % and 90 %
왏
Setting of language for dialogue and display D / UK / E / F / I
왏
Monitor
Further data
Power consumption at 230 V, 50/60 Hz
approx. W
1000
Standard accessories
Perforated stainless steel shelves (full width)
number
2
Stainless steel water dishes, 40 mm high
number
246
1500
2000
3
1 (full width)
2 (half width)
왏
Works calibration certificate (measuring point chamber centre at +37 °C)
Packing data
Net weight/Gross weight (packed in carton)
approx. kg
70/78
80/96
110/125
Width/Height/Depth
approx. cm
82/97/67
75/114/84
93/114/93
INC108med
INC153med
INC246med
108
153
246
Order No. CO2-Incubators
Options
Comfort module: two gas connections with quick release connectors, automatic switch-over of gas cylinders
T1
Hygiene module: electropolished interior, seamlessly welded by laser
T2
Communication module: USB interface, “Celsius” standard software for
programming and documentation, ring memory, printer interface
T3
CO2 module: extended CO2 range from 0 to 20 %
T4
Premium module: includes Comfort, Hygiene, Communication and CO2 module
T5
Humidity module: active microprocessor control for humidifying and dehumidifying (40 – 97 % rh), incl. digital indication
and auto-diagnostic system ensures even more rapid reaching of set humidity and very short recovery times while avoiding
condensate formation; recommended for applications with set O2 values of less than 10 %. Humidity supply with distilled
water (from an external tank) by a self-priming pump (standard humidity limit control and water dishes are omitted)
O2 module: control of oxygen concentration by N2 inlet; adjustment range 1 % up to 20 % O2; setting accuracy 0,1 %.
For applications with a set O2-value of less than 10 %, the humidity module is highly recommended.
Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight,
can be closed by silicone stopper, at the back (please, state location)
K7
T6
F7
4-part partitioning of interior with 4-part gas baffle (replacement of 3 full-width shelves by 6 half-width shelves)
–
K4
Door with lock (safety-lock)
B8
Stacking version for 2 units of equal size (bottom unit modification)
G3
Potential-free contact (24 V/2 A) with socket to NAMUR NE 28
for external monitoring (setpoint of temperature and CO2 is reached)
H5
Works calibration certificate for 5 %, 7 % and 10 % CO2 (measured at +37 °C)
D00106
Start-up of INCOmed incubators and brief training (D, A, CH only), through MEMMERT service, not subject to discount
Accessories
Additional perforated stainless steel shelf, full width
11
K9
108
153
246
B00325
B00321
B03813
Additional perforated stainless steel shelf, half width
–
B02742
Additional stainless steel grid, full width
E20165
E2
E20166
20166
20
E03492
Additional water dish
B02787
B0
02784
02784
02
B02784
B02786
Subframe (622 mm high)
B02792
B0
02732
02732
02
B02732
B02793
Subframe (130 mm high for 2 stacked incubators)
B02794
B0
02740
02740
02
B02740
B02795
HEPA-filter for chamber according to EN 1822, packed in sterile condition, incl. fixing unit
B04459
STERICard (additional or as replacement) for automatic chamber sterilisation cycle (not for sterilising load)
E04337
Pressure reducing valve to DIN 8546, incl. gas cylinder monitor
E02087
CO2 connection set, hose with coupling and clamp
B03881
Central water supply, only in combination with humidity module
B04712
Guarantee extension by 1 year
GA2Q5
12
55 TO 750 L ITRES
AI R JAC K E T S YS T E M
Compressor-cooled incubator ICP*
with TwinDISPLAY
AtmoCONTROL software
Model size: 55
-5 °C to +60 °C
Model sizes: 110 / 260 / 450 / 750
-12 °C to +60 °C
COMPRESSOR-COOLED INCUBATOR ICP Ideal at
temperatures around zero and below! If rapid and precise alternation between heating up and cooling down times in ramp operation is required, cooled incubators with compressor cooling prove
to be in peak form – yet still work extraordinarily quiet. Due to the
finely adjusted control technology, temperatures exactly reach the
set point values without energy-intensive bursts of power.
* available as Generation 2012 appliance from April 1, 2013
OPT I MUM UNI F OR MI T Y
R API D HEAT I NG A N D C O O L I N G
Completely enclosed working chamber
Cooling and heating units are situated outside the working chamber inside the air jacket
temperature control system surrounding the entire chamber interior ensuring quick and precise
temperature control. The motor-driven forced air circulation, adjustable in 10 % steps via the
ControlCOCKPIT ensures optimum temperature distribution.
ICP air jacket temperature control system
The sun is rising
•
•
Optional thermally decoupled internal illumination
Realistic day-night simulation by programming function for light,
temperature and ventilation
°C
37°C
loop
20°C
15°C
t
Day/night simulation programme function
Integrated energy saving function
The cooling unit works extremely energy-efficiently, as there is no continuous heating
against cooling. An intelligent DEFROST function enables defrosting as required.
13
14
T E C H N I C AL DATA
OV ERV I EW
COMPRES S OR-COOLED INCUBATORS ICP
COMPRESSOR-COOLED INCUBATOR ICP
according to 12 880: 2007-05, EN 61010-1 (IEC 61010-1), 61010-2-010
F
D
Standard equipment
Interior:
Stainless steel, material 1.4301 (ASTM 304)
Connection:
Mains cable with plug
Installation:
on lockable castors
256
Textured stainless steel, rear zinc-plated steel, intuitively
operated TwinDISPLAY (TFT colour display) with Touchscreen;
fully insulated stainless steel door (from size 450 two leaves)
33
B
Stainless steel grids (size 55: 1 grid, sizes 110 to 750: 2 grids)
Housing:
56
C
(93)
E
Internals:
A
(93)
Interfaces:
USB
Ethernet
135
LAN
Model sizes/Description
Stainless steel interior
Textured stainless
steel exterior
Volume
55
110
260
450
750
approx. l
53
108
256
449
749
Width
(A)
mm
400
560
640
1040
1040
Height
(B)
mm
400
480
800
720
1200
Depth (less 33 mm for fan)
(C)
400
500
600
600
8
14
mm
330
Stainless steel grids (standard equipment)
number
1
Max. number of grids
number
4
Max. loading per grids
kg
Max. loading of chamber
kg
80
175
2
5
9
30
300
Width
(D)
mm
585
745
824
1224
1224
Height (with castors)
(E)
mm
1153
1233
1552
1613
1950
Depth (without door handle), door handle + 56 mm
(F)
mm
514
584
684
784
784
approx. W
500
500
700
750
1200
Further data
Power consumption at 230/115 V, 50/60 Hz
Setting
Setting temperature range
°C
-12 bis +60 (ICP 55 -5 bis +60)
Setting accuracy
K
0.1
Packing data
Net weight
approx. kg
88
109
153
217
249
Gross weight (packed in carton)
approx. kg
104
127
209
287
324
Width
approx. cm
70
85
93
133
133
Height
approx. cm
142
150
181
188
221
Depth
approx. cm
73
79
92
105
105
ICP55
ICP110
ICP260
ICP450
ICP750
Order No. Compressor-cooled Incubators
available from April 1, 2013
T EC HNI C AL DATA
OV ERV I EW
Options
Chamber modification for the application of reinforced perforated stainless steel shelves or stainless steel grids
(bearing rails mounted in the working chamber) – includes replacement of 2 standard grids by 2 reinforced grids
Interior lighting (programme-segment-dependent switching on/off by process controller,
for example day/night simulation) flourescent lamps at the back – Thermally isolated illumination
box with insulating glass window and reflectors
no. of lamps/W
Illumination approx. Lux.
Interior socket, ampacity 230 V/2.2 A, can be switched off with the On/Off switch,
cannot be switched individually, moisture tight IP68
Entry port, 23 mm clear diameter, for introducing connections at the side,
can be closed by flap, standard positions
left centre/centre
left centre top
right centre/centre
right centre top
Entry port (silicone), 40 mm clear diameter, for introducing connections, moisture tight, can be closed by silicone
stopper, at the back (please, state location)
4 – 20 mA current loop interface (-20 to +70 °C 4 to 20 mA)
Temperature controller, actual value
Temperature of a Pt100 sensor positioned flexibly in chamber (max. 3 TwinDISPLAY)
55
C OMPR ES S OR -C OOLED I N C U BATO R S I C P
110
–
–
Reinforced stainless steel grid, max. loading 60 kg (size 450 and 750 only in connection with option K1)
Perforated stainless steel shelf
Reinforced perforated stainless steel shelf, max. loading 60 kg (size 450 and 750 only in connection with option K1)
Stainless steel slide-in drip tray, 15 mm rim (may affect the temperature distribution)
– cannot be used in connection with option K1
Stainless steel bottom drip tray, 15 mm rim (may affect the temperature distribution)
– cannot be used in connection with option K1
750
K1
–
R2
6/15
1000
6/18
2500
6/18
1000
6/30
2500
450
750
R3
F0
F1
F2
F3
F7
V3
V6
Fan speed monitoring
Stainless steel grid (standard equipment)
450
–
V4
Works calibration certificate for 3 temperatures: 0 °C, +37 °C, +60 °C
Works calibration certificate (measuring point chamber centre) at +10 °C and +37 °C standard equipment
Accessories
260
D00130
55
110
260
E20164
E20165
E28891
E20182
–
E29767
E29766
E26696
B03916
B00325
B29725
B00328
–
B29777
B29724
B31120
E02072
E02073
E29726
E02075
B04358
B04359
B29722
B04362
15
16
30 TO 750 L ITRES
P ELTI ER TECH NOLOGY
Peltier cooled incubator IPP* with SingleDISPLAY
Peltier cooled incubator IPPplus* with TwinDISPLAY
AtmoCONTROL software
Model sizes:
30 / 55 / 110 / 260 / 750
0 °C to +70 °C
PELTIER COOLED INCUBATOR IPP Heating and cooling
seamlessly with one system thanks to Peltier technology. In this
respect, cooled incubators IPP not only contribute to climate protection, but it also achieves an additional decrease in operating
costs of up 90 % compared to compressor technology. This perfect
development from the environmentally friendly and energy-saving
heating/cooling technology by Memmert convinces by outstanding
control precision and extremely small fluctuations.
* available as Generation 2012 appliance from November 1, 2012
QUI ET
V I B R AT I ON-F R EE
EC OLOGI C AL
ECONOMICAL
Extremely quiet and vibration-free
The fact that no compressor is required saves space and brings peace and quiet to the
laboratory. As Peltier cooled incubators IPP are almost vibration-free, they can also be
applied in entomology. If defined humidity is also required, an alternative would be the
constant climate chamber HPP, which is also equipped with Peltier technology.
No condensation in the interior chamber
Due to the closed Peltier cooling system, no outside air is exchanged. Physically derived, unavoidable formation of condensation during the cooling process does not take place in the interior
chamber but on the outside heat sink. In addition, the in the Peltier elements integrated fans
ensure a rapid transport of energy as well as an optimal temperature distribution.
Energy-saving heating/cooling technology combination
In contrast to compressor systems, Peltier technology is particularly economical at temperatures
close to the ambient temperature, since energy is only required during heating or cooling.
Therefore heating and cooling function are particularly precisely adjusted to each other.
Comparison compressor technology 250 Watt
and Peltier technology
200 Watt
Reduction in energy
consumption up to 90 %
150 Watt
100 Watt
Compressor technology
50 Watt
Peltier technology
0 Watt
5 °C
20 °C
30 °C
60 °C
17
18
T E C H N I C AL DATA
OV ERV I EW
PE LT IE R COOLED INCUBATORS IPP
PELTIER COOLED INCUBATOR IPP
according to 12 880: 2007-05 , EN 61010-1 (IEC 61010-1), 61010-2-010
F
D
A
Double doors:
Outside stainless steel, fully insulated, inside glass
(size 750 two-leaves)
Connection:
Mains cable with plug
Installation:
4 feet; size 750 on castors
352
Textured stainless steel, rear zinc-plated steel, intuitively
operated SingleDISPLAY or TwinDISPLAY (TFT colour display)
with Touchscreen
Interfaces:
Ethernet
USB
33
(only TwinDISPLAY)
Number of Peltier elements in the rear:
Sizes 30 to 55: 1
Size 110: 2
Size 260: 3
Size 750: 6
LAN
Model sizes/Description
Stainless steel interior
Textured stainless steel
exterior
Further data
Packing data
10
E
Housing:
B
Stainless steel grids
(sizes 30 and 55: 1, sizes 110 to 750: 2)
256
Standard equipment
Interior:
Stainless steel, material 1.4301 (ASTM 304), deep-drawn
Internals:
56
C
(93)
319
(93)
Volume
30
55
110
260
750
approx. l
32
53
108
256
749
Width
(A)
mm
400
400
560
640
1040
Height
(B)
mm
320
400
480
800
1200
Depth (less 10 mm for fan – Peltier)
(C)
mm
250
330
400
500
600
Stainless steel grids (standard equipment)
number
1
Max. number of grids
number
Max. loading per grid
kg
Max. loading of chamber
kg
60
80
175
mm
585
585
745
824
1224
3
2
4
5
9
14
30
300
Width
(D)
Height (size 750 with castors)
(E)
mm
707
787
867
1186
1726
Depth (without door handle), door handle + 56 mm
(F)
mm
524
604
674
774
874
approx. W
125
175
350
525
1050
Power consumption at 230/115 V, 50/60 Hz
Working temperature range without light
Working temperature range with light
°C
°C
0 to +70
+10 to +40
Setting temperature range
°C
0 to +70
Setting accuracy
K
0,1
Net weight
approx. kg
51
62
80
115
234
Gross weight (packed in carton)
approx. kg
62
74
94
171
309
Width
approx. cm
69
70
83
93
133
Height
approx. cm
96
104
114
153
192
Depth
approx. cm
66
73
79
92
105
IPP30
IPP55
IPP110
IPP260
IPP750
IPP30plus
IPP55plus
IPP110plus
IPP260plus
IPP750plus
Order No. Peltier cooled Incubators
IPP = Peltier cooled Incubators
plus = Model with TwinDISPLAY
available from November 1, 2012
T EC HNI C AL DATA
Optionen
OV ERV I EW
30
Chamber modification for the application of reinforced perforated stainless steel shelves or stainless steel grids
(bearing rails mounted in the working chamber) – includes replacement of 2 standard grids by 2 reinforced grids
Light module cold white 5.500 Kelvin: light strips arranged on the side walls of the interior, 10 strips for model 110,
14 for model 260, illumination strength 10.000 Lux, programme-controlled dimming from 0 to 100 % (in 10 % steps),
ramp programming in combination with temperature (only with TwinDISPLAY)
Light module cold white 5.500 Kelvin + warm white 2.700 Kelvin: LED light strips (5 resp. 7 alternating cold white light
strips and 5 resp. 7 warm white light strips) on the side walls of the interior, illumination strength 10.000 Lux,
programme-controlled dimming from 0 to 100 % (in 10 % steps), ramp programming in combination with temperature
(only with TwinDISPLAY)
Interior socket, ampacity 230 V/2.2 A, can be switched off with the On/Off switch, cannot be switched individually,
moisture tight IP68
Entry port, 23 mm clear diameter, for introducing connections,
can be closed by flap, standard positions
left centre/centre
(not for appliances with light module)
left centre top
right centre/centre
right centre top
Entry port, 23 mm clear diameter for introducing connections at the side,
can be closed by flap (please, state location)
left
right
rear
PELT I ER C OOLED I N C U BATO R S I P P
55
110
K1
–
T7
–
–
T8
–
R3
F0
F1
F2
F3
F4
F5
F6
D6
Other port, 38 mm clear diameter, can be closed by flap, in special positions in the back wall (please, state location)
F7
4 – 20 mA)
Temperature controller, actual value
Temperature of a Pt100 sensor positioned flexibly in chamber (max. 1 SingleDISPLAY, max. 3 TwinDISPLAY)
Works calibration certificate for 3 temperatures: +5 °C, +37 °C, +60 °C
Works calibration certificate (measuring point chamber centre) at +10 °C and +37 °C standard equipment
Accessories
Stainless steel grid (standard equipment)
V3
V6
D00129
30
55
110
260
750
E28884
E20164
E20165
E28891
E20182
E29767
E29766
E26696
B03916
B0
916
B03916
B00325
B29725
B00328
B29777
B29724
B31120
Reinforced stainless steel grid, max. loading 60 kg (from size 750 only in connection with option K1)
Perforated stainless steel shelf
–
B29727
Reinforced perforated stainless steel shelf, max. loading 60 kg (from size 750 only in connection with option K1)
Stainless steel slide-in drip tray, 15 mm rim (may affect the temperature distribution)
– cannot be used in connection with option K1
Stainless steel bottom drip tray, 15 mm rim (may affect the temperature distribution)
– cannot be used in connection with option K1
Guarantee extension by 1 year
750
–
Other port, 14 mm clear diameter, can be closed by flap, in special positions in the back wall (please, state location)
4 – 20 mA current loop interface (-10 to +80 °C
260
19
–
E02070
E02072
E0
E02072
E02073
E29726
E02075
B04356
B04358
B0
B04358
B04359
B29722
B04362
GA1Q5
GA2Q5
20
PE LT I E R
EC O L O G I C AL
ECONOMICA L
Storage cooled incubator IPS*
with SingleDISPLAY
AtmoCONTROL software
Model sizes:
260 / 750
+14 °C to +45 °C
STORAGE COOLED INCUBATOR IPS Save energy and
reduce the strain on the climate at the same time! If microbiological cultures, BOB5 samples, drinks containers or cosmetics need
to be stored over a long period at constant temperatures, storage
cooled incubators IPS with energy-efficient Peltier technology are
the perfect choice: absolute reliability, precision, durability and
eco-friendliness.
* available as Generation 2012 appliance as November 1, 2012
LONG-T ER M S TOR AGE
C OOLI NG
I N C U BAT I O N
Considerable potential for savings
in acquisition and operating costs
Temperature changes are not always necessary for long-term storage or incubating. So why
design heating, cooling and controlling systems for rapid heating up and cooling down times?
The performance of the IPS was tailor-made for permanent operation at constant temperatures
close to room temperature. The advantage: Acquisition costs and operating costs are considerably
reduced in comparison to conventional cooled incubators with compressor technology, as well as
to a large Peltier-cooled incubator.
Ideal for high ambient temperatures
Thanks to Peltier elements integrated for cooling the working chamber, the chamber load
won‘t break into sweat even at high ambient temperatures. Constant and precise incubation at
room temperature is guaranteed.
Low in vibration and durable for
absolutely safe long-term storage
Like the cooled incubator IPP, the IPS offers all the advantages of Peltier technology to the user.
Its interior chamber that is completely insulated from the environment minimises the risk of
drying out of the samples. It is practically noise-free and not only reduces stress on the chamber
load but also soothes the nerves of employees thanks to its quiet operation.
Glimpse into a Memmert storage
incubator:
Peltier elements guarantee perfect
climate inside the chamber.
21
22
T E C H N I C AL DATA
OV ERV I EW
S TORAGE COOLED INCUBATORS IPS
STORAGE COOLED INCUBATOR IPS
according to 12 880: 2007-05, EN 61010-1 (IEC 61010-1), 61010-2-010
855
D
(93)
(93)
A
Connection:
Mains cable with plug
Installation:
4 feet; size 450 and 750 on castors
Ethernet
135
Interfaces:
368
162
Outside stainless steel, fully insulated, inside glass
(size 750 two-leaves)
650
Double doors:
LAN
Model sizes/Description
Stainless steel interior
Textured stainless
steel exterior
Further data
Packing data
10
(162)
Textured stainless steel, rear zinc-plated steel,
intuitively operated SingleDISPLAY (TFT colour
display) with Touchscreen
E
2 stainless steel grids
Housing:
B
Internals:
256
Standard equipment
Interior:
Stainless steel, material 1.4301 (ASTM 304),
deep-drawn
56
C
Volume
260
750
approx. l
256
749
Width
(A)
mm
640
1040
Height
(B)
mm
800
1200
Depth (less 10 mm for fan – Peltier)
(C)
mm
500
600
Stainless steel grids (standard equipment)
number
Max. number of grids
number
Max. loading per grid
kg
30
Max. loading of chamber
kg
300
Width
(D)
Height (size 750 with castors)
Depth (without door handle), door handle + 56 mm
Power consumption at 230/115 V, 50/60 Hz
2
1
14
9
mm
824
1224
(E)
mm
1186
1726
(F)
mm
774
874
approx. W
525
1050
Working temperature range/Setting temperature range
°C
+14 to +45
Setting accuracy
K
0.1
Net weight
approx. kg
110
206
Gross weight (packed in carton)
approx. kg
166
281
Width
approx. cm
93
133
Height
approx. cm
153
192
Depth
approx. cm
92
105
IPS260
IPS750
Order No. Storage cooled Incubators
available from November 1, 2012
T EC HNI C AL DATA
OV ERV I EW
Optionen
Chamber modification for the application of reinforced perforated stainless steel shelves or stainless steel grids
(bearing rails mounted in the working chamber) – includes replacement of 2 standard grids by 2 reinforced grids
Interior socket, ampacity 230 V/2.2A, can be switched off with the On/Off switch, cannot be switched individually, moisture tight IP68
S TOR AGE C OOLED I N C U BATO R S I P S
260
750
–
K1
R3
Entry port, 23 mm clear diameter, for introducing connections at the side,
can be closed by flap, standard positions
left centre/centre
left centre top
right centre/centre
right centre top
F0
F1
F2
F3
left
right
rear
F4
F5
F6
Other port, 23 mm clear diameter, can be closed by flap,in special positions (please, state location)
Entry port, 14 mm clear diameter, can be closed by flap, in special positions in the back wall (please, state location)
D6
Entry port, 38 mm clear diameter, can be closed by flap, in special positions in the back wall (please, state location)
F7
4 – 20 mA current loop interface (0 to +70 °C
4 – 20 mA)
Temperature controller, actual value
Temperature of a Pt100 sensor positioned flexibly in chamber (SingleDISPLAY)
Works calibration certificate for a freely selectable temperature value
Works calibration certificate (measuring point chamber centre) at +18 °C and +25 °C standard equipment
Accessories
V3
V6
D00131
260
750
Stainless steel grid (standard equipment)
E28891
E20182
Reinforced stainless steel grid, max. loading 60 kg (size 750 only in connection with option K1)
E29766
E26696
Perforated stainless steel shelf
B29725
B00328
Reinforced perforated stainless steel shelf, max. loading 60 kg (size 750 only in connection with option K1)
B29724
B31120
E29726
E02075
B29722
B04362
Stainless steel slide-in drip tray, 15 mm rim (may affect the temperature distribution)
– cannot be used in connection with option K1
Stainless steel bottom drip tray, 15 mm rim (may affect the temperature distribution)
– cannot be used in connection with option K1
Guarantee extension by 1 year
GA2Q5
23
24
S PE C I A L EQ U I P M EN T – F O R ALL A PPLIA NCES
SPECIAL EQUIPMENT – GENERATION 2012
Options – available for all appliances
30
55
75
110
Door with lock (safety lock) standard on SN/SF and SNplus/SFplus 450 and 750
260
450
750
B6
Door hinged on the left
B8
Potential-free contact (24 V/2 A) with socket to NAMUR NE 28
for external monitoring (indicates when setpoint is reached)
Potential-free contact for combination error message
(e.g. supply failure, sensor fault, fuse)
Potential-free contact (24 V/2 A) with socket to NAMUR NE 28, for signal
generation, controlled by programme segment, for freely selectable functions
to be activated (e.g. activation of audible and visual signals, exhaust motors,
fans, stirrers, etc. (only for units with TwinDISPLAY)max. 2 contacts
on 1-phase appliances; max. 4 contacts on 3-phase appliances
2 contacts
4 contacts
Process-dependent door lock
(only for units with TwinDISPLAY)
–
H5
H6
H72
H74
D4
Door-open-recognition (only for units with TwinDISPLAY)
V5
Flexible Pt100 for positioning in chamber or in load with socket, 4-pin,
according to NAMUR NE 28, for external temperature recording (load temperature)
max. 3 sensors
Flexible Pt100 temperature sensor, positioned flexibly in chamber or load,
for local temperature measurement (up to 3 additional sensors are possible).
The measured temperature can, if required, be indicated on the display, recorded
in the integral ring store, and can be documented via the AtmoCONTROL software
MobileALERT, notification by SMS in case of any error or alarm of the device.
Requires option H6 “floating contact for alarm”
Temperature restriction (for UN/UF/UNplus/UFplus)
Temperatures: +60, +70, +80, +95, +100, +120, +160, +180, +200, +220
or +250°C (Please, indicate upon ordering)
Accessories – available for all appliances
160
H4
H8
C3
A8
30
55
75
110
160
USB-Ethernet adapter
E06192
Ethernet connection cable 5 m for computer interface
E06189
USB User-ID stick (with User-ID licence): Oven-linked authorisation licence
(User-ID-programme) on Memory-stick, prevents undesired manipulation by
unauthorised third parties. When reordering please specify serial number.
(Only for units with TwinDISPLAY).
USB stick with documentation software AtmoCONTROL and operation manual
for products with SingleDISPLAY, standard for appliances with TwinDISPLAY
260
450
750
E29778
E29780
Set of height adjustable feet (4 pcs)
B29768
Stacking set (4 pcs) for stacking of appliances of same size
(not for models 160, 260, 450 and 750)
Plug-in tube extension (outer diam. 60,3 mm, inner 57 mm), straight, for exhaust
air ducting (if necessary for connection by hose), only models U, I, /S
Plug-in tube extension (outer diam. 60,3 mm, inner 57 mm), angled, for exhaust
air ducting (if necessary for connection by hose), only models U, I, /S
Flush-fit unit (stainless steel frame covering gap between oven and wall opening),
with air slots – technical clarification required
Flush-fit unit (stainless steel frame covering gap between oven and wall opening),
without air slots – technical clarification required
Subframe, adjustable in height, incl. stacking set
(size 30 to 75: height 600 mm, size 110 to 450: height 500 mm)
Subframe, on castors, incl. stacking set
(size 30 to 75: height 660 mm, size 110 to 160: height 560 mm)
B29728
B29730
B29732
B29734
B29736
B29738
B29740
B29742
B29729
B29731
B29733
B29735
B29737
B29739
B29741
B29743
B29751
B29753
–
Castor frame (2-part), height 140 mm
IQ/OQ/PQ check list with works test data for chamber
as support for validation by customer
IQ/OQ/PQ check list with works test data for one free-selectable temperature
value incl. temperature distribution survey for 27 measuring points (9 for size 30)
to DIN 12 880: 2007-05
External measuring instrument with sensors for daylight and UV-light
(product information on demand)
Ditto with additional measuring head for temperature and humidity measurement
(product information on demand)
B29744
–
B29718
B29719
B29745
B29747
B29749
B29746
B29748
B29750
B29762
B29763
B29764
D00124
D00125
D00127
B04713
B04714
–
B29765
–
S PEC I AL EQUI PMENT – F OR AL L A P P L I A N C E S
SPECIAL EQUIPMENT – GENERATION 2003
Options – available for all appliances
Sizes: 200 / 400 / 500 / 600 / 700 / 800 108 / 153 / 246 256
Interface Ethernet instead of USB including software
W4
RS232 interface instead of USB
W6
Computer interface RS485 (for networking a max. of 16 ovens) instead of RS232
V2
Door with lock (safety lock – standard on sterilisers of size 700 and 800 –
not available for VO, VOcool, TTC/CTC
Interior socket, ampacity 230 V/2.2 A, can be switched off with the On/Off switch,
cannot be switched individually, moisture tight IP68
for ICP models
not switchable
switchable with on/off switch in front panel
Flexible Pt100 for positioning in chamber or in load with socket, 4-pin,
according to NAMUR NE 28, for external temperature recording (load temperature)
Additional Pt100 temperature sensor, positioned flexibly in chamber or load,
for local temperature measurement (up to 3 additional sensors are possible).
The measured temperature can, if required, be indicated on the multifunction
display, recorded in the integral ring store, and can be documented via the
“Celsius” software or on an attached printer.
not available for VO, VOcool, TTC and CTC)
Potential-free contact (24 V/2 A) with socket, according to NAMUR NE 28
for external monitoring (indicates when setpoint is reached)
Ditto, according to NAMUR NE 28 for combination error message
(e.g. supply failure, sensor fault, fuse)
Ditto, triple, for signal generation, controlled by programme segment
for a total of 3 freely selected functions to be activated
(e.g. acoustic and visual signals, exhaust motors, fans, stirrers etc.)
(not available with interior lighting)
Accessories – available for all appliances
USB connection cable for computer interface
Parallel/USB converter cable with integrated power supply unit to connect
HP printers with USB interface to MEMMERT units
Documentation package consisting of parallel USB converter cable including
PCL3-compatible HP colour inkjet printer with USB interface (HP OfficeJet 6000
or successor) for direct connection of printer to Memmert unit
Temperature profile write/read unit for programming via PC,
for writing to and reading from the chip card, up to 40 ramps
Additional chip card, blank, formatted (32 kB MEMoryCard XL for a maximum
of 40 ramps). Not available for INCOmed models.
Oven-linked authorisation card (User-ID-Card) prevents undesired manipulation
by unauthorised third parties. When reordering please specify serial number
Software conforming to FDA ”Celsius FDA Edition“.
Meets the requirements for the use of electronically stored data sets and electronic
signatures as laid down in Regulation 21 CFR Part 11 of the US Food and Drug
Administration (FDA). Base licence for the control of one unit.
Integration per additional unit (up to max.15 units)
into an already existent FDA-software licence (E05019)
IQ check list with works test data for chamber
as support for validation by customer
OQ check list with works test data for one free-selectable temperature
value incl. temperature distribution survey for 27 measuring points to
DIN 12 880: 2007-05 as support for validation by customer
OQ check list with works test data for one free-selectable temperature
value incl. temperature distribution survey for 5 measuring points to
DIN 12 880: 2007-05 as support for validation by customer (VO and VOcool only)
External measuring instrument with sensors for daylight and UV-light
(product information on demand)
Ditto with additional measuring head for temperature and humidity measurement
(product information on demand)
B6
R3
R4
H4
H8
H5
H6
H7
Sizes: 200 / 400 / 500 / 600 / 700 / 800 108 / 153 / 246 256
E03643
E05300
B04432
E05284
E04004
E04159
E05019
FDAQ4
D00103
D00104
D00117
B04713
B04714
25
Model variations of Generation 2012
SingleDISPLAY
ControlCOCKPIT with one TFT display
TwinDISPLAY
ControlCOCKPIT with two TFT displays
AVAILABLE APPLIANCES
AVAILABLE APPLIANCES
UN / UF / IN / IF / SN / SF / IPP / IPS
UNplus / UFplus / UNpa / INplus / IFplus / SNplus / SFplus
IPPplus / ICP / HPP / ICH
One high-resolution TFT colour display with touch-sensitive
buttons for selection of functions
Two high-resolution TFT colour displays with touch-sensitive
buttons for selection of functions
Available parameters on the ControlCOCKPIT: Temperature
(Celsius or Fahrenheit), fan speed, exhaust air flap position,
programme time
Available parameters on the ControlCOCKPIT: Temperature
(Celsius or Fahrenheit), fan speed, exhaust air flap position,
programme time, relative humidity, illumination, CO2
One temperature sensor Pt100 DIN class A in a 4-wire circuit
Two Pt100 sensors DIN class A in a 4-wire circuit for mutual
monitoring, taking over functions in case of an error
HeatBALANCE function for application specific adjustment of heat
output distribution (balance) between the upper and lower heating
groups in an adjustment range between -50 % and +50 %
ControlCOCKPIT with USB port for uploading programmes,
reading out protocol logs, activating the User-ID function
Displaying of already logged protocol data on the ControlCOCKPIT
(max 10,000 values correspond to approx. 1 week)
Ethernet interface on the rear of the appliance for reading
out the protocol log
Ethernet interface on the rear of the appliance for reading out the
protocol log and for uploading and implementing programmes and
for online logging
Double overtemperature protection: Electronic temperature
monitoring with freely adjustable monitoring temperature,
mechanical temperature limiter TB acc. to DIN 12 880.
Multiple overtemperature protection: Electronic temperature
monitoring TWW/TWB (protection class 3.1 or 2 resp. 3.3 for
units with active cooling) and mechanical temperature limiter TB
(protection class 1) acc. to DIN 12 880, AutoSAFETY automatically
adjusts to the set value within a freely adjustable tolerance range.
Setting individual MIN / MAX values for over/undertemperature
alarm and also for all other parameters such as relative
humidity, CO2
Structured stainless steel housing, scratch-resistant, robust and durable; rear of zinc-plated steel
High-temperature connectors on the rear of the appliance for single-phase power
connection according to country specific systems and IEC standards
Internal data logger with a storage capacity of at least 10 years
German, English, French, Spanish language settings available on the ControlCOCKPIT
Digital timer, adjustable between 1 minute and 99 days, 23 hours
The SetpointWAIT function guarantees that the process time does not start until the set temperature is reached at
all measuring points – optional for temperature values recorded by the freely positionable Pt100 sensors inside the chamber.
Adjustment of three calibration values for temperature and additional appliance
specific parameters directly at the ControlCOCKPIT (e. g. relative humidity)
27
Software Generation 2012
AtmoCONTROL
The innovative control and logging software
Parameters such as temperature and humidity as well as the process time can be set directly at the ControlCOCKPIT of
Generation 2012 appliances. Ramp programming is done via the control and logging software AtmoCONTROL, which
features a completely new software design.
Drag, drop & go!
Numerical and graphic programming of complex processes is a thing of the past. Today, programming is done via
AtmoCONTROL by means of the mouse or touchpad on your notebook. Even the most complex ramp programmes are
created within minutes. Simply drag & drop the graphical symbols for the desired parameters to the input field and
change the values according to your wishes with a mouse click.
Programming functions for appliances with SingleDISPLAY and TwinDISPLAY
•
•
•
•
•
Reading out, managing and organising the data logger
Saving the log memory in various formats
Online monitoring of up to 32 connected appliances
Optical alarms when the alarm limits individually set at the ControlCOCKPIT are exceeded
Automatic alarm to one or several e-mail addresses
Additional programming functions for appliances with TwinDISPLAY
•
•
•
•
•
Intuitive programming and archiving of ramps and programme sequences
Synchronous visualisation of the created programme sequence during programming
Application-specific repeat functions (Loops) can be inserted within a temperature control programme in any place
Simple creation of repeating weekly programmes
Programming, managing, and transferring programmes via Ethernet or USB stick