Dräger PulmoVista® 500
D-91-2010
PulmoVista® 500 is an Electrical Impedance Tomograph which
has been specially designed for use in clinical routine. Data is
continuously displayed in the form of images, waveforms and
parameters. Simply put, PulmoVista 500 lets you visualize the
distribution of ventilation.
AMBIENT CONDITIONS
During operation
Temperature (device)
Temperature (electrode belt and cables)
Atmospheric pressure
Relative humidity
5 to 40 °C (41 to 104 °F)
5 to 45 °C (41 to 113 °F)
700 to 1060 hPa (10.15 to 15.37 psi)
20 to 95%, without condensation
During storage and transportation
Temperature
Atmospheric pressure
Relative humidity
-20 to 40 °C (-4 to 104 °F)
500 to 1060 hPa (7.25 to 15.37 psi)
20 to 90%, without condensation
Frame rate
Frame rate with the option ADAP
Cut-off frequency for low pass filter
Upper and lower cut-off frequencies for
the band pass filter
10, 15, 20, or 30 frames per second
10, 15, 20, 30, 40, or 50 frames per second
10 to 300/min
D-25283-2009
SETTINGS
30 to 300/min
PulmoVista 500
02 |
DRÄGER PULMOVISTA® 500
PERFORMANCE CHARACTERISTICS
EIT measurement
Number of electrodes
Feed current amplitude
Feed current frequency
16 electrodes plus 1 reference electrode
80 to 90% of maximum patient auxiliary current conforming to IEC 60601-1 (3rd edition)
80 to 130 kHz
Display unit (Medical Cockpit Infinity C500)
Resolution
Contrast ratio min. 50
Viewing angle
1440 x 900 pixels
min. 500 : 1
130°
OPERATING DATA
Mains supply
Mains supply nominal voltage
and frequency range
Mains supply characteristics
Current consumption
at 230 V
at 100 V
Power consumption
maximum during operation
typically during operation
maximum when device is switched off, but
charging batteries
Integrated battery
Type
Fuse
Time bridged following a mains power failure
with new and fully charged internal battery
100 V to 240 V, 50/60 Hz
mains supply must conform to clause 4.10.2 of IEC 60601-1 (3rd edition)
and especially be overvoltage category II or lower according to IEC 60664-1
max. 0.6 A
max. 1.3 A
125 W
approx. 80 W
40 W
2 x 12 V lead-acid battery
F20AL 32VDC
min. 5 minutes (typically 10 minutes)
Charging
Charging time (fully discharged batteries)
min. 12 hours
Sound pressure level (for free-field measurement
over a reflecting surface)
max. 45 dB(A)
Dimensions (W x H x D)
PulmoVista 500
incl. trolley
600 mm x 1400 mm x 750 mm
(23.62 inch x 55.12 inch x 29.53 inch)
Weight
PulmoVista 500 including trolley at maximum
44 kg (97 lbs)
Materials used
Electrode belt
Patient cable
Trunk cable
silicone rubber, conductive silicone rubber, stainless steel, gold-plated brass
plastics (thermoplastic polyurethane (TPU), polyamide (PA), polyurethane (PUR), polypropylene
(PP), thermoplastic elastomer (TPE), polybutylene terephthalate (PBT))
plastics (polyamide (PA), thermoplastic polyurethane (TPU), polyurethane (PUR))
DRÄGER PULMOVISTA® 500
| 03
CLASSIFICATION
Applied parts
Mode of operation
Protection against electric shock regarding
External power supply
Applied part
Protection against harmful ingress of water
Microenvironments of pollution
Electromagnetic compatibility (EMC)
(conforming to European Directive 89/336/EEC)
Biocompatibility of applied parts
Classification as per EC Directive 93/42/EEC
Annex IX
applied parts are: electrode belt, reference electrode, patient cable, trunk cable
continuous
protection class I conforming to EN 60601-1
type BF conforming to IEC 60601-1
IPX1 (dripping water: vertically falling drops) conforming to IEC 60529
level 2 conforming to IEC 60601-1
tested in accordance with IEC 60601-1-2
tested in accordance with ISO 10993 for intact skin and an application duration of <24 hours
II a
COMMUNICATION INTERFACES ON MEDICAL COCKPIT INFINIT Y ® C500
Digital outputs
USB ports 1 and 2 (one on each side panel)
Digital inputs
RS 232 (9-pin) connectors 1 (on the back panel)
RS 232 (9-pin) connectors 3 (on the back panel)
RS 232 ports are electrically isolated from the
equipment electronics (test voltage 1500 V)
OPTION ADAP
The option ADAP (Advanced Data Analysis Package)
extends the functionality of the basic EIT software
with the following functions:
– Patient data entry
– Data recording
– Event marking
– Data review
– File handling
– Higher frame rate
– Filter setting Band pass
– Manual adjustment of the operating
frequency
Communication interface
passive USB storage devices only
MEDIBUS connection to ventilator
reserved for future use
LIST OF ACCESSORIES
Name/Description
Order-No.
Trunk cable
Patient cable, size S
8420048
8420029
Patient cable, size M
Patient cable, size L
Patient cable, size XL
Patient cable, size XXL
Electrode belt, size S
Electrode belt, size M
Electrode belt, size L
Electrode belt, size XL
Electrode belt, size XXL
ECG electrode (pack of 500)
Medibus cable (male/female)
Medibus cable (female/female)
Retrofit kit ADAP
8420047
8420035
8420271
8420273
8420059
8420058
8420057
8420056
8420055
4527750
8306488
8416326
8420006