Specifications
Specifications
ACCURACY
Accuracies apply for 1 year 19°C to 25°C. Temperature coefficient outside these limits is <0.1 x
quoted range accuracy per °C.
DC Volts
Range
Accuracy
Resolution
100mV
0.06% ± 3 dig.*
10uV
1000mV
0.04% ± 2 dig.
100uV
10V
0.06% ± 2dig.
1mV
100V
0.06% ± 2 dig.
10mV
1000V
0.06% ± 2 dig.
100mV
* after null
Input Impedance:
10M //<100pF, except for Vdc plus Vac measurement when the
1M ac attenuator is in parallel with the 10M dc attenuator.
Maximum Input:
1kV DC or AC peak, any range.
NMR:
>60dB at 50/60Hz.
1k
>90dB at DC/50Hz/60Hz.
Unbalanced CMR:
Maximum Allowable AC Voltage While Measuring DC Voltage
Range
Max AC
Range
Max AC
The table shows the maximum AC voltage
100mV, 1000mV
6V
10V
80V
(45Hz-50kHz) that can be applied without
100V
200V
1000V
1000V
affecting the accuracy of the DC measurement.
AC Volts (True RMS)
Range
Accuracy
45Hz - 10kHz
100mV
10kHz - 20kHz
1% ± 20 dig.
1000mV
10V
0.2% ± 20 dig
0.2% ± 20 dig
100V
750V
Resolution
20kHz - 50kHz
-
10µV
1% ± 50 dig
100µV
1% ± 80 dig
1mV
1% ± 80 dig
10mV
-
100mV
Accuracy specifications apply for readings between 1,000 and 12,000 counts. Additional error at
crest factor = 3 typically 0.2%.
Extended frequency performance
(typical):
100mV range <-1dB at 50kHz; 1000mV, 10V and
100V ranges <-1dB at 100kHz.
Input Impedance:
1M //<100pF
Maximum Input:
750V rms, 1000V peak; any range.
1k
>60dB at DC/50Hz/60Hz
Unbalanced CMR:
3
(AC + DC) Voltage Accuracy
Total measurement error will not exceed the sum of the separate ac and dc accuracy
specifications plus 1 display count.
Resistance
Range
Accuracy
Resolution
100
0.1% ± 3 dig.
10m
1000
0.08% ± 2 dig.
100m
10k
0.09% ± 2 dig.
1
100k
0.09% ± 2 dig.
10
1000k
0.12% ± 2 dig.
100
10M
0.5% ± 2 dig.
1k
20M
0.5% ± 2 dig.
10k
Maximum Input:
300V DC or AC rms, any range.
Maximum Open Circuit Voltage:
4V
DC Current
Range
Accuracy
Resolution
1mA
0.1% ± 3 dig.
100nA
100mA
0.1% ± 3 dig.
10uA
10A
0.3% ± 3 dig. to 1A
1mA
10A
1.0% ±3 dig. to 5A
1mA
10A
3% ± 10 dig to 10A
1mA
Maximum Input:
mA ranges - 500mA DC or AC rms, 250V, fuse protected.
10A range - 10A DC or AC rms, 250V, fuse protected.
Typical Voltage Burden:
mA ranges - <250mV
10A range - <500mV
AC Current (True RMS)
Range
Accuracy (45Hz - 10kHz)
Resolution
1mA
0.35% ± 20dig.
100nA
100mA
0.35% ± 20 dig.
10uA
10A
0.5% ± 20 dig. to 1A
1mA
10A
1.2% ± 20 dig to 5A
1mA
10A
3% ± 20 dig. to 10A
1mA
Accuracy specifications apply for readings between 1,000 and 12,000 counts. Additional error at
crest factory = 3 typically 0.2%.
Maximum Input:
mA ranges - 500mA DC or AC rms, 250V, fuse protected.
10A
Typical Voltage Burden:
mA ranges - <250mV
10A
4
range - 10A DC or AC rms, 250V, fuse protected.
range - <500mV
Frequency
Range
Accuracy
100Hz
1000Hz
Resolution
0.01Hz
0.01% ± 1 dig.
0.1Hz
10kHz
1Hz
100kHz
10Hz
Range:
10Hz to 100kHz
Input sensitivity:
Better than 30m Vrms (100mV range); better than 10% of range
for all other Vac and Iac ranges.
Capacitance
Range
Accuracy
Resolution
10nF
10pF
100nF
100pF
1uF
2% ± 5 dig.
10uF
100uF
1nF
10nF
5% ± 5 dig.
100nF
Continuity and Diode Test
Continuity:
1000
Diode Test:
Test current approximately 0.5mA; displays voltages up to 1.2V.
Maximum Open Circuit
Voltage:
4V
Maximum Input:
300V DC or AC rms, any range.
range selected; audible tone sounds for impedance <10 .
DISPLAY
Display Type:
High contrast LCD. Main display 4½ digits 17mm high, secondary
display 5 digits 10mm high.
Scale Length:
4¼ digits (12000 counts) in most modes.
Annunciators:
LCD annunciators for all ranges, functions and program modes.
Reading Rate:
Varies with function, maximum 4/sec.
Overrange:
Display flashes 12000 if input too great for range.
Overflow:
Displays -Or- if calculated result overflows display.
COMPUTING FUNCTIONS
Null (Relative)
!Null:
Stores current reading and subtracts it from future readings.
Additional non-volatile function for nulling test lead resistance.
Hold:
Reading is frozen
T-Hold (Touch & Hold):
Reading is frozen when stable.
dB:
Displays measurement in dBm relative to 600 !or other user-entered
impedance.
5
AC plus DC:
The RMS value of the ac plus dc parts of the signal is calculated and
displayed.
% Deviation:
Displays % deviation from entered reference value.
Ax+B:
Linear scaling of results, with offset.
Limits:
Reading displayed with HI, LO, or PASS with respect to user-defined
high and low limits.
Min/Max:
Minimum and maximum reading stored.
Power:
Calculates V2/R and displays in Watts with respect to a user-defined
impedance.
VA:
Calculates and displays Volts x Amps.
Data Logger:
Manual or automatic storage of 100 measurements. Storage interval
1s to 9999s, manually from keyboard, or by remote contact closure.
INTERFACES
Full remote control facilities are available through the RS232 (all models) or GPIB (alternative
mains-only version) interfaces.
RS232:
Baud rates 2400, 9600 or 19200. Complies fully with the ARC
(Addressable RS232 Chain) interface standard. Address selectable
from the front panel. Operational only when the meter is powered from
the AC input.
GPIB (IEEE-488):
The meter can be specified with an IEEE-488 interface. This is an
alternative version which operates only from AC mains; IEEE-488 is
not a retrofittable option. Address selectable from the front panel.
POWER REQUIREMENTS
AC Input:
220V-240V or 110-120V AC ±10%, 50/60Hz, by internal adjustment;
5VA max. Installation Category II.
Batteries:
6 x C cells, disposable or rechargeable.
Battery Life:
>150 hours from alkaline cells; typically 70 hours from rechargeable
cells.
GENERAL
6
Operating Range:
+5°C to + 40°C, 20% to 80% RH
Storage Range:
"20°C to + 60°C
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 1.
Safety:
Complies with EN61010-1.
EMC:
Complies with EN61326.
Size:
260(W) x 88(H) x 235(D)mm, excl. handle and feet.
Weight:
2.0kg.