CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
CI-100A Indicator Operation & Service Manual
CONTENT
1. Specification ....................................................................................... 2 2. Faceplate............................................................................................. 6 3. Summary of Key function .................................................................. 7 5. Normal Weighing mode ................................................................... 21 6. Count Weighing Mode ..................................................................... 24 7. Percent Weighing Mode .................................................................. 26 8. BMI Working Mode ........................................................................... 27 9. HOLD Function ................................................................................. 28 10. Data Compare Function ................................................................. 29 11. Accumulation .................................................................................. 30 12. Calibration ...................................................................................... 31 13. Weight Fine-tune ............................................................................ 34 14. View ADC output Code .................................................................. 34 15. View or Calibrate Power Voltage ................................................... 35 16. View or Set Time............................................................................. 35 17. View or Set Date ............................................................................. 35 18. View Firmware Version .................................................................. 36 19. Display Test .................................................................................... 36 20. Keyboard and Buzzer Test ............................................................ 36 21. Serial Port1/2 (COM1/2) Receiving Test ....................................... 36 22. Serial Port1/2(COM1/2) Transmitting Test .................................... 36 23. Details about Serial Communication ............................................ 37 24. Connectors and Jumpers .............................................................. 44 25. Meaning of Some Symbols and Troubleshooting ....................... 46 26. Display Character........................................................................... 49 27. Packing List .................................................................................... 50 28. Version History ............................................................................... 51 ────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
1
CI-100A Indicator Operation & Service Manual
CI-100A Indicator Operation Manual
Thank you for purchasing the CI-100A indicator. Please read all operating instructions carefully before use and
keep the following points in mind:

Avoid lengthy exposure to extreme heat or cold, your scale works best when operated at normal room
temperature. Always allow the scale to acclimate to a normal room temperature before use

Always use a proper power supply and well connected load cell.

Allow sufficient warm up time. Turn the scale on and wait for a few minutes if possible, to give the
internal components a chance to stabilize before weighing.

These electronic scales are precision instruments. Do not operate near an in-use cell phone, radio,
computer or other electronic device. These devices emit RF and can cause unstable scale readings. If
your scale ever performs poorly, try moving the scale to a different room or location.

Avoid using in condition of heavy vibration and airflow.

Read the weight reading in short time after loading. The output signature of load cell and electronic
circuit may be little influenced after weighing for a long time.

Some functions maybe disabled by manufacture before shipping, if you need these functions
or don’t want to use some other functions, please contact with manufacture in advance,
because of this, some operations or faceplate maybe are a little bit different, and some
description sections may be deleted.
1. Specification
1.1 Outline and Bracket installation:
1.1.1 Outline:
Either a 7-pin quick disconnect cable assembly or simple load cell cable can be supplied with the
CI-100A Indicator. Both can be used to interface this indicator to a scale platform. Refer to the drawing
below illustrating the load cell cable with a 7-pin quick disconnect cable assembly.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
2
CI-100A Indicator Operation & Service Manual
1.1.2 Indicator bracket installation:
1.1.2.1 According to following figure, use locknut to fasten part1 and part2 onto the bracket, and make
sure part1 and part2 in correct direction.
1.1.2.2 From the back side of rear side, fasten the indicator to the bracket by using four M4x6 screws.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
3
CI-100A Indicator Operation & Service Manual
1.2 Power Supply:
1.2.1 4xAA size alkaline batteries: When the sign of battery is empty, this prompt you’d better to replace
batteries; When “Lo.bAt” displayed, this prompt you should replace batteries immediately, otherwise,
it will turn off automatically in 10s.
1.2.2 AC Adapter: 6-9VDC, ≥500mA, central positive:
1.2.3 USB power supply if USB interface is installed
1.2.4 Work current: ≤15mA (with backlight off, no optional Bluetooth Module and no load cells)
≤30mA (with backlight on, no optional Bluetooth Module and no load cells)
≤50mA (with backlight on and one 350 load cell, no optional Bluetooth Module)
1.3 Display:
6-digit,7-segment , 1"(25mm) LCDs with 16 annunciators and blue backlight
1.4 Keypad:
6 push buttons
1.5 Environment:
1.5.1 Working temperature: -10°C to 40°C
1.5.2 Storage temperature: -20°C to 70°C
1.5.3 Humidity: 10 to 90% RH without condensation
1.6 Load cell Excitation:
1.6.1 Voltage: 5Vdc
1.6.2 Max. Current: 60mA (can power 4-350 ohm bridge, take care the limitation when USB is power
supply)
1.6.3 Signal connection: 4 or 6 lead with sense leads
1.6.4 Max Sensitivity: -3mV/V to +3mV/V
1.6.5 Load cell wiring (7 pin socket):
Pin1:
Pin2:
Pin3:
Pin4:
Pin5:
Pin6:
Pin7:
Red Excitation +
White Excitation Green Signal +
Blue Signal Shield
Orange Sense +
Black Sense -
If other special connection is used, please refer the additional paper
1.7 Communication:
1.7.1 Serial port1: Full-duplex RS232
1.7.2 Serial port2: USB (Virtual RS232)
1.7.3 Optional Serial port3: Bluetooth
1.7.3 Baud Rate: Selectable: 1200-2400-4800-9600-19200-38400 bps
1.7.4 Data Output Format: 8N1, 7O1, 7E1, 7E2, 7O2
1.7.5 Protocol: programmable
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
4
CI-100A Indicator Operation & Service Manual
1.8 Analog Circuit characters:
1.8.1 24-bit A/D converter
1.8.2 Conversion Speed: 10Hz or 80Hz selectable
1.8.3 Input range: -15mV to +15mV
1.8.4 Output code: 1mV input between S+ and S- of load cell connector will output about 100,000 raw
Counts.
1.8.5 With Hardware low pass filter and two programmable digital low pass filters
1.9 Accuracy: ≤0.01%
1.10 Capacity and Division: Programmable
1.10.1
1.10.2
1.10.3
Max display range: -999,999 to 999,999
Division number range for primary unit: 100-100,000
Division number range for second unit: 100-125,000
(Division number will be limited by REGULA setting)
Recommended Sensitivity: >1uV/ display division
1.11 Calibration Method:
1.11.1 Software calibration with long-term storage in EEPROM
1.11.2 Provides smooth curve fit through four points.
1.11.3
1.11.4
1.11.5
Calibration can be done under kg or lb weight unit with 10% -100%FS standard weight
Optional directly weight fine adjustment (±10%)
Optional Geographical Adjustment
1.12 Real Clock: optional built-in nonvolatile real time & date
1.13 Other Main Function:
1.13.1
Programmable Initial or key Zero Range; automatic zero point tracking range;
1.13.2
Programmable pre-set tare weight
1.13.3
Programmable motion detection window
1.13.4
Programmable auto-power off time, backlight working mode
1.13.5
Programmable hold function: with peak weight holding, dynamic weighing
1.13.6
Available Check Weighing Mode; Parts Counting Mode; Percentage Working Mode
1.13.7
Available Measure Unit: kg, g, lb, oz, lb:oz, PCS, %,
1.13.8
Data Comparison and indicator of High, OK and Low is available in Weighing mode, Parts
Counting mode and Percentage Working mode.
1.13.9
Battery voltage is low and charging indicator
1.13.10
Programmable what content will be output and when they are output on serial port.
1.13.11
Optional BMI function
1.13.12
Optional Bluetooth wireless communication
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
5
CI-100A Indicator Operation & Service Manual
2. Faceplate
Meaning of symbol on faceplate:
2.1
Turn on when scale is stable
2.2
Turn on when scale is at zero point and the gross weight is 0
2.3 NET
Turn on when net weight is displaying, and the tare weight is not 0; Turn off when gross weight
is displaying
2.4 TOTAL
Turn on when display data is total times or total of weight, pieces or percentage
2.5
lb
Turn on when measure unit is lb or lb:oz
2.6
oz
Turn on when measure unit is oz or lb:oz
2.7
kg
Turn on when measure unit is kg
2.8
g
Turn on when measure unit is g
2.9
%
Turn on when in percentage weighing mode, measure unit is %
2.10
Pcs
Turn on when in counting mode, measure unit is pieces
2.11 PEAK
Turn on when in working in HOLD mode and HOLD type is PEAK-HOLD
2.12 HOLD
Turn on and flashing/not flashing: working HOLD mode and displaying number is/isn’t live
2.13
BMI
Turn on when working in BMI mode
2.14
HI
Turns on when data compare is enabled and current data (weight, pieces or percent) is over its
upper setting limitation.
2.15
OK
Turns on when data compare is enabled and current data (weight, pieces or percent) is
between its high limitation and low limitation
2.16 LO
Turn on when data compare is enabled and current data (weight, pieces or percent) is below
its lower limitation
2.17
Turn on when battery is used or charged
2.18
flash when local wireless module is installed and powered on before link, become steady on
after linked with remote wireless device
2.19
Turn on when remote wireless device is in battery low state
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
6
CI-100A Indicator Operation & Service Manual
3. Summary of Key function
Key
Condition
Function
Weighing/Counting/Percent,
Enter or exit HOLD mode
press down <3s
Weighing/Counting/Percent,
HOLD
SETUP
To enter setup mode
press down > 3s
Input data mode,
to input decimal point
press down >3s
Input data mode,
Return to last sub-menu
press down <3s
Menu selection mode
Return to last sub-menu
Weighing/Counting/Percent,
Output data to serial communication port
press down <3s
PRINT
Weighing/Counting/Percent ,
FUNC
press down > 3s
select working mode: weighing ,counting or percent
Input data mode
The digit on flashed position add 1
Menu selection mode
To last item of current menu
ACCUMULATION function is enabled in configuration
Weighing/Counting/Percent/BMI,
menu : To add up current weight/pieces/percentage to
press down <3s
memory, display times and total of accumulation; in BMI
working mode, to input height
ACC
TOTAL
Weighing/Counting/Percent,
ACCUMULATION function is enabled in configuration menu,
press down >3s
To display times and total of accumulation
Input data mode
the digit on flashed position subtract 1
Menu selection mode
To Next item of current menu
Weighing mode,
Change weighing units: kg->lb->lb:oz (not be available in
press down<3s
some conditions)->kg
Counting/Percent mode,
To enter getting piece weight or unit-percentage-weight
press down < 3s
mode (by way of sample or input directly).
UINT
Weighing /Counting/Percent mode, To input compare data (for weight , pieces or percentage )
DATA
press down> 3s
Display date or time mode,
press down >3s
of high and low limitation
To set current date or time
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
7
CI-100A Indicator Operation & Service Manual
Display voltage mode,
press down >3s
Display ADC code
Select displaying code from no-filter, filter1, filter2
Input data mode
Rotate the flashed position from left to right
Weighing/Counting/Percent ,
press down <3s
TARE
Weighing/Counting/Percent ,
PRESET press down > 3s
ON/OFF
Tare function
To input pre-set tare weight at selected weight unit
input data mode or
To confirm input data or current item selection, and go to
Menu selection mode
next item of current menu, or next operation
Display ADC code
Set or clear “tare” code
Power off mode
Power on
Weighing/Counting/Percent,
ZERO
To calibrate input voltage value
press down <3s
Weighing/Counting/Percent,
press down > 3s
Zero function
Power off
Input data mode
ignore modification
Menu selection mode
Prepare to exit from current working mode
Note:
Normally, the second function of one key need pressing it down more than 3s.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
8
CI-100A Indicator Operation & Service Manual
4. Operation Menu Structure
4.1 Main menu:
CONFIG
USER
CAL
MISC
TEST
*BT
*NOTE: BT menu only be active when Blue tooth module is installed on COM3
4.2 CONFIG Submenu:
CONFIG
SubMenu1
SubMenu2
Option
Default
CFG.ON
CFG.OFF
RESET
Final
Set
seal switch is on or off
NO
YES
REGULA
Remark
NO
reset configure parameters to default setting
NONE
USA
CANADA
NONE
select the standard that the scale will
comply with: USA,CANADA, EUROPE
EUROPE
PRIM.N
100 100,000
PRIM.D
0.0001
3000
the division number under primary unit, if
(REGULAR)≠none, the max is 10,000
1
the division value under primary unit; the
division value under second unit is
automatically determined by indicator
according to the division value under
primary unit.
0.0002
0.0005
0.001
0.002
0.005
0.01
0.02
0.05
0.1
0.2
0.5
1
2
5
10
20
50
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
9
CI-100A Indicator Operation & Service Manual
PRIM.Ut
KG
Select the primary unit from kg or lb.
The second unit is the lb if kg selected as
primary unit or kg if lb selected as primary
unit.
3000
the division number under second unit,the
max is 1.25*(PRIM.N), and
if(REGULAR)≠none, the max is 10,000
KG
LB
SECND.N
100125,000
10N.DSP
NO
YES
NO
MOTION
1-255
4
OVER.LD
0-100
0
ADC
AD.FROM
COM3
NO
AD.H.SPD
YES
KG
LB
UNITS
OZ
LBOZ
G
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
ADC
Weight Data comes from: ADC=local A/D
chip on PCB; COM3=COM3 interface;
NO
Speed of A/D convert: NO=10Hz;
YES=80Hz; if AD.FROM=COM3,this item
will not be shown
YES
YES
NO
NO
IN.IZSM
0-100
10
WEIGHT
CAL.ZRO
WEIGHT
LAST.Z.T
OV.IZSM
DSP.OVR
WEIGHT
CAL.ZRO
LAST.Z.T
Units that can be used by UNIT key select:
YES=enable this unit to be used;
NO=disable this unit to be used;
Refer to sectioon5.12 for some limitation; In
trade application, lb:oz should not be
allowed.
NO
ZRO.PNT
IZSM
Display weight at 10 times division number
under primary unit, if (REGULAR)=none, no
this item
Check motion window:
1-255=±0.25d
*(1-255), if (REGULAR)≠none, the max is
12
over load display limitation: 0=FS+9d,
1-100=101%FS -200%FS, if
(REGULAR)≠none, the max is 10
DSP.OVR
Initial zero(power on zero) point range:
0=no limitation, 1-100= (calibration zero
point) ±1%FS - (calibration zero point)
±100%FS, If (REGULAR)≠none, the max is
10
Choose which weight as current initial zero
point when current weight is in IZSM range:
WEIGHT= current weight ; CAL.ZRO=
calibration zero; LAST.Z.T=switch-off zero
and tare
If (REGULAR)≠none, the value is fixed on
WEIGHT
Choose which weight as current initial zero
point when current weight is over IZSM
range: DSP.OVR=display initial zero is
over; WEIGHT= current weight;
CAL.ZRO= calibration zero;
LAST.Z.T=switch-off zero and tare
If (REGULAR)≠none, the value is fixed on
DSP.OVR
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
10
CI-100A Indicator Operation & Service Manual
SAZSM
0-100
2
AZSM
0-100
8
FILTER
FUNC
Zero key range: 0=no limitation, 1-100=
(initial zero point) ±1%FS - (initial zero
point) ±100%FS, if (REGULAR)≠none, the
max is 2
Zero tracking window:
0=0d, no tracking;
1-100=±(0.2+0.05*(1-100))d /s, if
(REGULAR)≠none, the max is 10
FLT1.TH
0-255
40
Enter digital filter1 threshold: 0=no filter1;
1-254=filter1 be used only when vibration in
±0.5d*(1-254) ; 255= filter1 be always
used
FLT1.ST
1-64
8
Digital filter1 intensity: 1-64 ADC’s
data will be averaged
Enter digital filter2 threshold: 0=no filter2;
1-254=filter2 be used only when vibration in
±0.5d*(1-254) ; 255= filter2 be always
used
FLT2.TH
0-255
8
FLT2.ST
0-255
240
Digital filter2 intensity:
strong
HOLD
YES
YES
Yes/No=enable/disable hold function;
Operation refer to section8. In trade
application HOLD function should be
prohibited
YES
Yes/No=enable/disable counting function,
Operations refer to section6.
NO
Percent weighing function is enable or
disable: (1)NO=disable; (2)100%=enable
and display format is 100% ;
(3)100.0%=enable and display format is
100.0%; (4)100.00%=enable and display
format is 100.00%; Operations refer to
section7
NO
Yes/No=enable/disable BMI function,
Operations refer to section8.
YES
Yes/No=enable/disable data comparison
function; Operations refer to section9
NO
COUNT
YES
NO
PERCNT
NO
100%
100.0%
100.00%
BMI
YES
NO
COMPAR
YES
NO
ACCUMU
NO
MANUAL
MANUAL
AUTO
GEO.CAL
WT.ADJ
YES
NO
YES
NO
NO
NO
0-255=weak to
Accumulation Mode selection: (1)NO=no
accumulation function; (2)MANUAL=add up
current number to accumulation memory
after TATOL key is pressed;
(3)AUTO=automatically add up current
number to accumulation memory after scale
is stable and weight is over (NLD.RNG)
Yes/No=enable/disable Geographical
Adjustment Factor
Yes/No=enable/disable weight fine-tuning
using keypad in weighing mode, if
(REGULAR)≠none, this item is NO
* The setting will be limited by choice of REGUALA
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
11
CI-100A Indicator Operation & Service Manual
4.3 USER Submenu:
USER
SubMenu1
SubMenu2
RESET
NO
COM1
YES
BAUD.RT
Option
Default
NO
Remark
Set
reset user parameters to default setting
1200
2400
4800
9600
9600
selection of com1's baud rate
8N1
selection of com1's byte format:
(1)8N1=8 data bits, No parity check bit, 1 stop
bit;
(2)7O1=7 data bits, 1 Odd parity check
bit, 1 stop bit; (3)7E1=7 data bits, 1 Even
parity check bit, 1 stop bit;
(4)7O2=7 data
bits, 1 Odd parity check bit, 2 stop bit;
(5)7E2=7 data bits, 1 Even parity check bit, 2
stop bit;
PRTCMD
Selection com1 output mode: (1)NONE =No
communication; (2)CONT=continuously output;
(3)PRINT=output after PRINT key pressed;
(4)CMD=output after a request command is
received; (5)PRT.CMD= output after PRINT key
pressed or request command received;
(6)STABLE=output after scale is stable;
Note: use PRINT or CMD to output data, the
scale must be stable.
19200
38400
BYT.FMT
8N1
7O1
7E1
7O2
7E2
OUT.MOD
NONE
CONT
PRINT
CMD
PRTCMD
STABLE
LAYOUT
MULTPL
SINGLE
MULTPL
EH-SCP
SCP-12
OUT1
SCAL.ID
GROSS
YES
NO
YES
NO
TARE
YES
NO
NET
YES
NO
NO
com1 output content and format set:
(1)MULTPL= the following selected item in
OUT1 will be output using defined format;
(2)SINGLE= only displayed content and current
status will be output, it’s compatible with
NCI-SCP01; (3) EH-SCP= Command
–response mode, similar to Toledo PS60
protocol; (4)SCP-12 = only displayed content
and current status will be output, it’s compatible
with NCI-SCP12(NCI3835);
Yes/No=enable/disable output scale's ID
number, Prompt is “SCALE ID”
NO
Yes/No=enable/disable output gross weight.
Prompt is “GROSS”
NO
Yes/No=enable/disable output tare weight.
Prompt is “TARE”
YES
Yes/No=enable/disable output net weight.
Prompt is “NET”
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
12
CI-100A Indicator Operation & Service Manual
PERCNT
YES
NO
UPCTWT
YES
NO
COUNT
YES
NO
PCWT
BMI
YES
NO
ACCUMU
YES
NO
YES
NO
TIME
YES
NO
AD.CODE
YES
NO
BAT.VOL
YES
NO
STATUS
YES
NO
B.LINE
Yes/No=enable/disable output weight
percentage. Prompt is “PERCENTAGE”
NO
Yes/No=enable/disable output weight of 1%
percentage. Prompt is “1% REF WT”
NO
Yes/No=enable/disable output counts. Prompt is
“QUANTITY”
NO
Yes/No=enable/disable output piece weight.
Prompt is “PIECE WT”
YES
NO
DATE
NO
Yes/No=enable/disable output height and BMI.
Prompt is “HEIGHT” and “BMI”
Yes/No=enable/disable output accumulation
times and total. Prompt is “ACC. N” and
“TOTAL”
NO
NO
NO
Yes/No=enable/disable output date. Prompt is
“DATE”
NO
Yes/No=enable/disable output time. Prompt is
“TIME”
NO
Yes/No=enable/disable output ADC's code.
Prompt is “A/D CODE”
NO
Yes/No=enable/disable output voltage of
battery. Prompt is “VOLTAGE”
NO
Yes/No=enable/disable output scale's status.
Prompt is “STATUS”
LINE1
How many blank lines after strings output:
NONE=no blank line, LINE1/2/3/4=there're 1,
2,3 or 4 blank lines after strings, used for
paper feed forward 1/2/3/4 lines.
9600
selection of com2's baud rate
8N1
selection of com1's byte format:
(1)8N1=8 data bits, No parity check bit, 1 stop
bit;
(2)7O1=7 data bits, 1 Odd parity check
bit, 1 stop bit; (3)7E1=7 data bits, 1 Even
parity check bit, 1 stop bit;
(4)7O2=7 data
bits, 1 Odd parity check bit, 2 stop bit;
(5)7E2=7 data bits, 1 Even parity check bit, 2
stop bit;
NONE
LINE1
LINE2
LINE3
LINE4
COM2
BAUD.RT
1200
2400
4800
9600
19200
38400
BYT.FMT
8N1
7O1
7E1
7O2
7E2
OUT.MOD
NONE
PRTCMD
CONT
Selection com2 output mode: (1) NONE = No
communication; (2) CONT=continuously output;
(3)PRINT=output after PRINT key pressed;
(4)CMD=output after a request command is
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
13
CI-100A Indicator Operation & Service Manual
received; (5)PRT.CMD= output after PRINT key
pressed or request command received;
(6)STABLE=output after scale is stable;
Note: use PRINT or CMD to output data, the
scale must be stable.
PRINT
CMD
PRT.CMD
STABLE
LAYOUT
MULTPL
SINGLE
MULTPL
EH-SCP
SCP-12
OUT2
SCAL.ID
YES
NO
GROSS
YES
NO
TARE
YES
NO
NET
YES
NO
PERCNT
YES
NO
UPCTWT
YES
NO
COUNT
YES
NO
PCWT
YES
BMI
NO
YES
NO
ACCUMU
YES
NO
DATE
YES
NO
TIME
YES
NO
AD.CODE
YES
BAT.VOL
YES
NO
YES
NO
STATUS
NO
com2 output content and format set:
(1)MULTPL= the following selected item in
OUT2 will be output use defined format;
(2)SINGLE= only displayed content and current
status will be output, it’s compatible with
NCI-SCP01; (3) EH-SCP= Command
–response mode, similar to Toledo PS60
protocol; (4)SCP-12 = only displayed content
and current status will be output, it’s compatible
with NCI-SCP12(NCI3835);
NO
Yes/No=enable/disable output scale's ID
number, Prompt is “SCALE ID”
NO
Yes/No=enable/disable output gross weight.
Prompt is “GROSS”
NO
Yes/No=enable/disable output tare weight.
Prompt is “TARE”
YES
Yes/No=enable/disable output net weight.
Prompt is “NET”
NO
Yes/No=enable/disable output weight
percentage. Prompt is “PERCENTAGE”
NO
Yes/No=enable/disable output weight of 1%
percentage. Prompt is “1% REF WT”
NO
Yes/No=enable/disable output counts. Prompt is
“QUANTITY”
NO
Yes/No=enable/disable output piece weight.
Prompt is “PIECE WT”
NO
NO
Yes/No=enable/disable output height and BMI.
Prompt is “HEIGHT” and “BMI”
Yes/No=enable/disable output accumulation
times and total. Prompt is “ACC. N” and
“TOTAL”
NO
Yes/No=enable/disable output date. Prompt is
“DATE”
NO
Yes/No=enable/disable output time. Prompt is
“TIME”
NO
Yes/No=enable/disable output ADC's code.
Prompt is “A/D CODE”
NO
Yes/No=enable/disable output voltage of
battery. Prompt is “VOLTAGE”
NO
Yes/No=enable/disable output scale's status.
Prompt is “STATUS”
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
14
CI-100A Indicator Operation & Service Manual
B.LINE
NONE
LINE1
LINE2
LINE1
How many blank lines after strings output:
NONE=no blank line, LINE1/2/3/4=there're 1,
2,3 or 4 blank lines after strings, used for
paper feed forward 1/2/3/4 lines.
9600
selection of com3's baud rate
8N1
selection of com1's byte format:
(1)8N1=8 data bits, No parity check bit, 1 stop
bit;
(2)7O1=7 data bits, 1 Odd parity check
bit, 1 stop bit; (3)7E1=7 data bits, 1 Even
parity check bit, 1 stop bit;
(4)7O2=7 data
bits, 1 Odd parity check bit, 2 stop bit;
(5)7E2=7 data bits, 1 Even parity check bit, 2
stop bit;
LINE3
LINE4
COM3*
BAUD.RT
BYT.FMT
1200
2400
4800
9600
19200
8N1
7O1
7E1
7O2
7E2
OUT.MOD
NONE
PRTCMD
Select com3 output mode: (1) NONE = No
communication ; (2)CONT=continuously output;
(3)PRINT=output after PRINT key pressed;
(4)CMD=output after a request command is
received; (5)PRT.CMD= output after PRINT key
pressed or request command received;
(6)STABLE=output after scale is stable;
Note: use PRINT or CMD to output data, the
scale must be stable.
MULTPL
Com3 output content and format set:
(1)MULTPL= the following selected item in
OUT3 will be output use defined format;
(2)SINGLE= only displayed content and current
status will be output, it’s compatible with
NCI-SCP01; (3) EH-SCP= Command
–response mode, similar to Toledo PS60
protocol; (4)SCP-12 = only displayed content
and current status will be output, it’s compatible
with NCI-SCP12(NCI3835);
CONT
PRINT
CMD
PRTCMD
STABLE
LAYOUT
MULTPL
SINGLE
EH-SCP
SCP-12
OUT3*
SCAL.ID
YES
NO
GROSS
TARE
NET
PERCNT
YES
NO
YES
NO
YES
NO
YES
NO
NO
Yes/No=enable/disable output scale's ID
number, Prompt is “SCALE ID”
NO
Yes/No=enable/disable output gross weight.
Prompt is “GROSS”
NO
YES
NO
Yes/No=enable/disable output tare weight.
Prompt is “TARE”
Yes/No=enable/disable output net weight.
Prompt is “NET”
Yes/No=enable/disable output weight
percentage. Prompt is “PERCENTAGE”
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
15
CI-100A Indicator Operation & Service Manual
UPCTWT
COUNT
PCWT
BMI
ACCUMU
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
DATE
TIME
YES
NO
YES
NO
AD.CODE
YES
BAT.VOL
STATUS
B.LINE
BEEP
KEY
NO
YES
NO
YES
NO
NONE
LINE1
LINE2
LINE3
LINE4
YES
NO
COMPAR
Yes/No=enable/disable output weight of 1%
percentage. Prompt is “1% REF WT”
Yes/No=enable/disable output counts. Prompt is
“QUANTITY”
Yes/No=enable/disable output piece weight.
Prompt is “PIECE WT”
NO
NO
NO
Yes/No=enable/disable output height and BMI.
Prompt is “HEIGHT” and “BMI”
Yes/No=enable/disable output accumulation
times and total. Prompt is “ACC. N” and
“TOTAL”
Yes/No=enable/disable output date. Prompt is
“DATE”
Yes/No=enable/disable output time. Prompt is
“TIME”
NO
NO
NO
NO
Yes/No=enable/disable output ADC's code.
Prompt is “A/D CODE”
NO
Yes/No=enable/disable output voltage of
battery. Prompt is “VOLTAGE”
Yes/No=enable/disable output scale's status.
Prompt is “STATUS”
NO
NO
LINE1
Yes/No=enable/disable beep after a key
pressed down
YES
NONE
IN.LMT
(1)NONE=not beep; (2)L.Low=beep when lower
than low limitation; (3)IN.LMT=beep when in
range of low and high limitation;
(4)O.HIGH=beep when over high limitation;
(5)OUT.LMT=beep when lower than low
limitation or higher than high limitation
AUTO
HOLD Mode: (1)NONE=no hold function
(2)PS.PEAK=Positive Peak number Hold mode:
scale will display and refresh the positive peak
value from last zero setting
(3)NG.PEAK=Negative PEAK number Hold
mode. it's Similar with PS.PEAK, but negative
number is used. (4)TOGGLE=Press HOLD key
to enter HOLD mode, if weight is over
(NLD.RNG) and stable, the data will be frozen
until press HOLD key again to exit. (5)
AVERAG= Average HOLD mode: in this
mode, if weight is over (NLD.RNG), and its
variation is less than (HLD.RNG), the average
data in (AVG.TIM) will be frozen. Press HOLD
key or (HLD.TIM) time elapsed to exit this
mode (6)AUTO=Auto hold mode: it's similar with
L.LOW
IN.LMT
O.HIGH
OUT.LMT
HOLD
HLD.MOD
NONE
PS.PEAK
NG.PEAK
TOGGLE
AVERAG
How many blank lines after strings output:
NONE=no blank line, LINE1/2/3/4=there're 1,
2,3 or 4 blank lines after strings, used for
paper feed forward 1/2/3/4 lines.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
16
CI-100A Indicator Operation & Service Manual
AVERAG mode, but if the one held load is
removed, and a new load that is over
(NLD.RNG) put on scale, the new load will be
automatically frozen.
AUTO
AVG.TIM
1-60
3
average data time for HOLD mode: 1-60s
STB.TIM
3*AVG.TIM
- 255
9
Waiting time for scale stable in HOLD mode:
3*(AVG.TIM) - 255S
0
Data HOLD time: 0=data will be frozen until
HOLD key pressed; 1-65535=data frozen time is
1-65535s, after the time elapses, scale will exit
HOLD mode
HLD.TIM
HLD.RNG
0-65535
0
- 255
5
OTHER
NLD.RNG
1-255
CMD.SRC
NONE
10
COM.1
COM.2
COM.1.2
COM.1
COM.3
COM.1.3
COM.2.3
COM.1.2.3
Vibration range of data that can be averaged
and held in HOLD mode: 0=any data can be
averaged;
1-255= only the data which
vibration is in 1-255d can be averaged and
held;
1-255=the range of weight is 1-255d; when
current weight is less than this value, the scale
can be regarded as empty , or the load on scale
is removed. It must be bigger than
(CONFI.MOTION).
Source of the executed command selection:
(1)NONE=no any command will be executed;
(2)COM.1/.2/.3= command from COM1/2/3 will
be executed;
(3)COM.1.2/.1.3/.2.3/= command from
COM1,COM2/COM1,COM3 or
COM2,COM3 will be executed;
(4)COM.1.2.3= command from COM1,COM2
or COM3 will be executed;
NOTE: if AD.FROM=COM3, then COM.x.x.3 will
not be active
5
Auto off time: 0=not auto power off; 1-255=auto
power off after 1-255 minutes, in this period, no
operation or no weight changing
OFF
Auto off mode:
(1)OFF=turn off instrument;
(2)DSP.TIM= display time; (3)AC.TIME=turn
off when only battery is used, display time when
AC adaptor is used。
If time is displayed and enabled continuously
output to COMx, the time will be sent out
0-255
30
LCD backlight set: (1)0=always off
(2)1=always on (2)2=press down ZERO+UNIT
together more than 3s to turn on or turn off
(4)3-255=auto on when key operation or weight
changing, auto off after 3-255s elapsed
LCD.CST
CST1---8
CST8
SCAL.ID
000000999999
123456
A.OFF.T
OFF.MD
0-255
OFF
DSP.TIM
AC.TIME
LCD.BLT
LCD contraction level selection
scale's ID number: 000000-999999
NOTE: if AD.FROM=COM3, then COM3 and OUT3 will not be active !
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
17
CI-100A Indicator Operation & Service Manual
4.4 CAL Submenu:
CAL
SUBMENU1 SUBMENU2
CAL.ON
CAL.OFF
seal switch is on or off
ZERO
LINE
only do zero point calibration, then go to CAL.END to
end
CAL.P0
Linear calibration point0: do zero point
calibration, this point can't be omitted.
CAL.P1
END.Y
Linear calibration point1: do first weight point
calibration, this point can't be omitted and standard
weight must be over 10%FS.
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P2
END.Y
Linear calibration point2: do second weight point
calibration, standard weight must be over 10%FS
and be larger than it in CAL.P1, this point can be
omitted.
YES
End calibration? YES=go to CAL.END to end; NO=go
to do next point calibration
NO
CAL.P3
CODE
GEO
REMARK
OPTION
GRAVT
Linear calibration point3: do third weight point
calibration, standard weight must be over 10%FS
and be larger than it in CAL.P2, this point can be
omitted.
Selection of Geographical Position Code 00-70,refer
TABLE12-1
00-70
9.76183
-9.99999
Input Gravity of User Location by keyboard
INPUT
Input or view calibration parameters value
CAL.END
calibration end and restart
NOTE:
The details can be referred in section “12.CALIBRATION”
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
18
CI-100A Indicator Operation & Service Manual
4.5 MISC Submenu:
MISC
SUBMENU1
REMARK
CODE
display ADC's code, this code can be after no-filter, filter1 or filter2; details refer to
section14
VOL
display voltage; calibrate voltage; details refer to section15
DATE
display date and set date; details refer to section17
TIME
display time and set time; details refer to section16
VER
display firmware version; details refer to section18
4.6 TEST Submenu:
TEST
SUBMENU1
REMARK
DISP.TST
test LCD or LED; details refer to section19
COM1.RD
test COM1 receiving; details refer to section21
COM1.TD
test COM1 transmitting; details refer to section22
COM2.RD
test COM2 receiving; details refer to section21
COM2.TD
test COM2 transmitting; details refer to section22
KEY.TST
test keys and buzzer; details refer to section20
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
19
CI-100A Indicator Operation & Service Manual
4.7 BT Submenu:
BT
SUBMENU1
REMARK
BT…?
Go to Blue Tooth module setup mode. Refer section “24.Bluetooth Configuration”
INIT
To initialize blue tooth module some parameters: In initialization, INIT.i will be
shown, i=1/2/3…, this means processing initialization step1/2/3… These
parameters including:
(1) Set no flow control in serial communication
(2) Set local blue tooth name to EHBTxxxxxx (xxxxxx is last six address of local
blue tooth )
(3) Set always need authentication to access the Blue tooth in the indicator
(4) Set blue tooth device type to 000000
(5) Set blue tooth module function is only serial communication
(6) Set to throw away all received data before the module connection
(7) Set BAUD=9600bps, UARTMD=8N1, PASSWD=1234,TYPE=0=slave
(8) Set ROLE=0=salve, MAXCON=1,BIND=0=not bind address
(9) Clear all memorized address
LADD
Go to display Local blue tooth address mode, The address will shown in Li.xxxx
format, i=1/2/3, xxxx is four hexadecimal digits
RADD
Go to display or input Remote blue tooth address mode, The address will shown in
Ri.xxxx format, i=1/2/3, xxxx is four hexadecimal digits
BAUD
Set or view serial communication baud rate :1200-2400-4800-9600-19200-38400,
default is 9600
UARTMD
Set or view serial communication mode: 8N1-7O1-7E1-7O2-7E2
(1) stop bit=1/2 bit stop bits;
(2)parity bit=N(none)/O(odd)/E(even); default is 8N1
PASSWD
Set or view access password
TYPE
Set or view type of device: 0=slave; 1=master; 2=repeater; 3=relay node
ROLE
Set or view device role: 0=slave,1-7=master, and can connect 1-7 devices; default
is slave
MAXCON
Set maxim number of connected devices: 0-7, default=1
BIND
Set or view whether address binding is need: 0=no binding, 1=bind address;
default=0
CLRADR
Clear all memorized address
BTVER
View the version of device
Note:
Only proper blue tooth module is installed (e.g BTM0304C1H, BTM0804C2H…) on COM3,
corresponding items can be active! Details refer to section24.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
20
CI-100A Indicator Operation & Service Manual
5. Normal Weighing mode
5.1 During key operation, please note to use the second function of one key need pressing the key over 3 seconds;
To input data or select menu, use
to process.
5.2 Power on scale: when scale is off, short press ON/OFF key to turn on;
Power off scale: when scale is on, long press ON/OFF key to turn off the scale.
5.3 Change working mode: long press FUNC key, then use
to choose and confirm to enter into
weighing mode, counting mode or percentage working mode
5.4 Enter to or exit from HOLD mode: press HOLD key
5.5 ZERO: When the weights is stable and within the zero range, press ZERO key to set new zero point. Please
refer the ZERO and TARE limitations in 5.9 section.
5.6 TARE: When the gross weight is big than zero, and the scale is stable, press TARE key, the indicator will show
net weight of zero, the NET annunciator will be lighted. Please refer the ZERO and TARE limitations in 5.9
section.
5.7 Preset tare weight: long press PRESET key, Pr.Tare will show, and the TARE annunciator flashes, it means it is
in preset TARE weight mode, use
to input tare weight, and its unit is same as
the unit that it used before, there’s no limitation to preset tare weight, but it should bigger than zero. After input
a tare weight, “NET” annunciator will be lighted. Note: this indicator can only save one tare weight, the new
tare weight will automatically replace the old one. Also, please refer the ZERO and TARE limitations in 5.9
section
5.8 Clear tare weight: remove any weight on platform, wait till the scale is stable, short press TARE key, please
refer the ZERO and TARE limitations in 5.9section.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
21
CI-100A Indicator Operation & Service Manual
5.9 Limitation to ZERO and TARE operation under different conditions:
Table5-1
Standard
Weight
on
platform
≤0
USA
>0
≤0
Canada
>0
≤0
Europe
Tare key
no
No action
yes
Clear the tared weight
No
≤0
no
No action
Yes
Clear the tared weight
No
Tare
Yes
No action
No
No action
Yes
Clear the tared weight
Tare
Yes
No
No action
yes
Clear the tared weight
No
>0
Zero key
Zero
Tare
Yes
No
>0
None
(same with
Europe)
key function
Data in TARE
memory unit
Tare
yes
Zero
Zero
zero and clear the
tared weight
zero
zero and clear the
tared weight
Zero
zero and clear the
tared weight
zero
zero and clear the
tared weight
NOTE: (1) ZERO only be active when scale is stable and weight is in SAZSM setting range.
(2) TARE only be active when scale is stable
(3) Clear TARE weight or ZERO scale, make indicator to enter displaying GROSS mode
(4) TARE weight, make indicator to enter displaying NET mode
5.10 Output data: When scale is stable, press PRINT key.
5.11 Accumulation:
Press ACC key to add displayed number to accumulation memories, and accumulation times will also
add up 1. and then to display accumulation result
5.12 Change Weight UNIT:
Short press UNIT key to select kg, lb, lb:oz, g, or oz unit, note: under some condition, g and lb:oz is not
available. And , In trade application, lb:oz should be prohibited. Please refer the following tables:
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
22
CI-100A Indicator Operation & Service Manual
Table5-2: use Kg as primary unit:
Calibration
Display division value in different weight unit that can be used
division value
kg
g
lb
oz
lb:oz
0.0001kg
0.001kg
0.01kg
0.1kg
1kg
10kg
0.0002kg
0.002kg
0.02kg
0.2kg
2kg
20kg
0.0005kg
0.005kg
0.05kg
0.5kg
5kg
50kg
0.0001kg
0.001kg
0.01kg
0.1kg
1kg
10kg
0.0002kg
0.002kg
0.02kg
0.2kg
2kg
20kg
0.0005kg
0.005kg
0.05kg
0.5kg
5kg
50kg
0.1g
1g
10g
100g
Not available
Not available
0.2g
2g
20g
200g
Not available
Not available
0.5g
5g
50g
500g
Not available
Not available
0.0002lb
0.002lb
0.02lb
0.2lb
2lb
20 lb
0.0005 lb
0.005 lb
0.05 lb
0.5 lb
5 lb
50 lb
0.001 lb
0.01 lb
0.1 lb
1 lb
10 lb
Not available
0.005oz
0.05oz
0.5oz
5oz
50oz
Not available
0.01oz
0.1oz
1oz
10oz
Not available
Not available
0.02oz
0.2oz
2oz
20oz
Not available
Not available
Not available
Not available
0.5oz
Not available
Not available
Not available
Not available
0.1 oz
1 oz
Not available
Not available
Not available
Not available
0.2 oz
2oz
Not available
Not available
Not available
Table5-3: use LB as primary unit:
Calibration
Display division value in different weight unit that can be used
division value
kg
g
lb
oz
lb:oz
0.0001lb
0.001 lb
0.01 lb
0.1 lb
1 lb
10 lb
0.0002 lb
0.002 lb
0.02 lb
0.2 lb
2 lb
20 lb
0.0005 lb
0.005 lb
0.05 lb
0.5 lb
5 lb
50 lb
Not available
0.0005 kg
0.005 kg
0.05 kg
0.5 kg
5 kg
0.0001 kg
0.001 kg
0.01 kg
0.1 kg
1 kg
10 kg
0.0002 kg
0.002 kg
0.02 kg
0.2 kg
2 kg
20 kg
Not available
0.5g
5g
50g
500g
Not available
0.1g
1g
10g
100g
Not available
Not available
0.2g
2g
20g
200g
Not available
Not available
0.0001lb
0.001 lb
0.01 lb
0.1 lb
1 lb
10 lb
0.0002 lb
0.002 lb
0.02 lb
0.2 lb
2 lb
20 lb
0.0005 lb
0.005 lb
0.05 lb
0.5 lb
5 lb
50 lb
0.002oz
0.02oz
0.2oz
2oz
20oz
Not available
0.005 oz
0.05 oz
0.5 oz
5 oz
50 oz
Not available
0.01 oz
0.1 oz
1 oz
10 oz
Not available
Not available
Not available
Not available
0.2 oz
2 oz
Not available
Not available
Not available
Not available
0.5 oz
Not available
Not available
Not available
Not available
0.1 oz
1 oz
Not available
Not available
Not available
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
23
CI-100A Indicator Operation & Service Manual
5.13
Check Weight in weighing mode:
5.13.1
To make weight compare function be available, CONFIG-FUNC-COMPAR item should set to YES,
and high and low limitation of weight should be set correctly according to following steps:
5.13.2
In weighing mode, Press down DATA key more than 3s to input compare data of high and low.
5.13.3
After HIGH being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input high weight
number and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press ZERO
key to exit and back to weighing mode.
5.13.4
After Low being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input low weight
number and press TARE key to confirm. Annunciator of Lo will be shown in this step. Press ZERO
key to exit and back to weighing mode.
NOTE:
If High number is 0 or is equal or less than low number, the comparison will be disabled, and the
input data has no limitation.
5.13.5
After a reasonable limitation is set and compare is be active, one of annunciator HI, OK, LO will be
lighted, and the beeper will sound according to its setting in USER-BEEP.
5.14
Enter Setup Mode:
5.14.1
If need to set configuration parameters, set user parameters, calibrate the scale, set current date or
time, test some hardware…
It’s need to long press SETUP key to enter setup mode
5.14.2
After Entering Setup Mode, the main menu item CONFIG will be shown first.
5.14.3
In Setup mode, use
to select wanted submenu and
wanted menu item, select wanted choice, set wanted number, confirm and save data, or exit this
mode. Please refer the “4.Operation Menu Structure”.
6. Count Weighing Mode
6.1
In this mode, scale will weigh goods weight on scale, calculate and display its counts after the piece
weight of goods is obtained
6.2
To make counting function be available, CONFIG-FUNC-COUNT item should be set to YES in
CONFIG menu.
6.3
To enter counting working mode, in normal weighing or percent-weighing mode, long press FUNC
key, WEIGH/PERCEN will be shown, use ACC or PRINT key to select COUNT, then press TARE to
confirm go to parts counting mode. Before new piece weight is got, the last piece weight will be
used.
6.4
In counting working mode, the function of ZERO, TARE, PRINT, HOLD, PRESET TARE, ACC,
SETUP, ON/OFF are available.
6.5
There’re two ways to obtain the piece weight: (1) input piece weight directly, refer operations of
step6.5.1 (2)weigh samples weight which quantity is known, refer operation of step 6.5.2
6.5.1
Input piece weight from keypad: in counting mode, press UNIT key, When InP.PWt is shown, press
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
24
CI-100A Indicator Operation & Service Manual
TARE key to enter input piece weight mode.
6.5.1.1
When UNIT.KG is shown, use PRINT or ACC key to select the unit of piece weight, use TARE key to
confirm and go to next. Press ZERO key to exit getting piece weight mode and back to counting
mode.
6.5.1.
When last stored piece weight is shown, use PRINT, ACC, UNIT key to input new piece weight,
press SETUP key more than 3s to input decimal point. Press TARE key to confirm and save it, then
go back to counting mode. If the input piece weight is less than 0.5d, the indicator will display
PWt.Er and go back to counting mode.
6.5.2
Obtain piece weight by weighing samples weight which quantity: in counting mode, press UNIT key,
When InP.PWt is shown, use PRINT or ACC key to select SPL.PWT, press TARE key to weigh
samples (which quantity is known) weight, calculate piece weight. Press ZERO key to exit getting
piece weight mode and back to counting mode.
6.5.2.1
When SPL.Lo is shown, move away any sample on scale and press TARE key to confirm, before
scale is stable, SPL.Lo will be flashed. After it is stable, it will go to next step. Press ZERO key to
exit getting piece weight mode and back to counting mode.
6.5.2.2
When SPL.Hi is shown, put samples (its quantity is known) onto the scale, Press TARE key to
confirm reading weight. Before scale is stable, SPL.Hi will be flashed. After it is stable, it will go to
next step. Press ZERO key to exit getting piece weight mode and back to counting mode.
6.5.2.3
After INP.PCS being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input the
quantity of samples and Press TARE key to confirm. If the calculated piece weight is less than 0.5d,
the indicator will display PWt.Er and go back to counting mode, otherwise, after the reasonable
piece weight being got, the scale will go back to counting mode. The got piece weight can be saved
after the power off and can be used next time.
6.6
Check Counts (counts compare) in Counting mode:
6.6.1
To make counts compare function be available, CONFIG-FUNC-COMPAR item should set to YES,
and high and low limitation of pieces should be set correctly according to following steps:
6.6.2
In counting working mode, Press down DATA key more than 3s to input compare data of high and
low.
6.6.3
After HIGH being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input high
quantity number and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press
ZERO key to exit getting piece weight mode and back to counting mode.
6.6.4
After Low being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input low quantity
number and press TARE key to confirm. Annunciator of Lo will be shown in this step. Press ZERO
key to exit getting piece weight mode and back to counting mode.
NOTE:
If High number is 0 or is equal or less than low number, the comparison will be disabled.
6.6.5
After a reasonable limitation is set and compare is be active, one of annunciator HI, OK, LO will be
lighted, and the beeper will sound according to its setting in USER-BEEP.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
25
CI-100A Indicator Operation & Service Manual
7. Percent Weighing Mode
7.1
In this mode, scale will weigh goods weight on it, calculate and display its percentage after the
unit-percentage-weight of goods is obtained. (NOTE: If 100% display format is set to 100%,
100.0% or 100.00% in CONFIG-FUNC-PERCEN menu item, then, the
unit-percentage-weight is the weight of 1%, 0.1% or 0.01%)
7.2
To make percent weighing function be available, CONFIG-FUNC-PERCEN menu item shouldn’t be
set to NONE.
7.3
To enter percent weighing mode, in normal weighing or counting mode, long press FUNC key,
WEIGH/COUNT will be shown, use ACC or PRINT key to select PERCEN, then press TARE to
confirm go to percent weighing mode. Before new unit-percentage-weight is got, the last
unit-percentage-weight will be used.
7.4
In percent weighing mode, the function of ZERO, TARE, PRINT, HOLD, PRESET TARE, ACC,
SETUP, ON/OFF are available.
7.5
To obtain the unit-percentage-weight, there’re two ways : (1) input weight and its percentage, then
scale calculates the unit-percentage-weight, refer operations of step 7.5.1 (2) weigh samples weight
which percentage is known, refer operation of step 7.5.2
7.5.1
Input weight and its percentage from keypad, and calculate unit-percentage-weight: in percent
weighing mode, press UNIT key, When InP.Pct is shown, press TARE key to enter this mode:
7.5.1.1
Before input weight, use PRINT or ACC key to select the percentage from 1%, 2%, 5%, 10%, 20%,
50% and 100%, this percentage is corresponding to the weight you will input in following steps
7.5.1.2
When UNIT.KG is shown, use PRINT or ACC key to select the unit of input weight, use TARE key to
confirm and go to next. Press ZERO key to exit and back to percent weighing mode.
7.5.1.3
When last stored unit-percentage-weight data is shown, use PRINT, ACC, UNIT key to input new
unit-percentage-weight, press HOLD key more than 3s to input decimal point. Press TARE key to
confirm and save it, then go back to percent weighing mode. If the calculated unit-percentage-weight
is less than 0.5d, the indicator will display Pct.Er and go back to percent weighing mode.
7.5.2
Obtain unit-percentage-weight by weighing samples weight which percentage is known: in percent
weighing mode, press UNIT key, When InP.Pct is shown, use PRINT or ACC key to select SPL.Pct,
press TARE key to weigh samples (which percentage is known) weight, calculate piece weight.
Press ZERO key to exit and back to percent weighing mode.
7.5.2.1
When SPL.Lo is shown, move away any sample on scale and press TARE key to confirm, before
scale is stable, SPL.Lo will be flashed. After it is stable, it will go to next step. Press ZERO key to
exit and back to percent weighing mode.
7.5. 2.2 When SPL.Hi is shown, put samples (its percentage is known) onto the scale, Press TARE key to
confirm reading weight. Before scale is stable, SPL.Hi will be flashed. After it is stable, it will go to
next step. Press ZERO key to exit and back to percent weighing mode.
7.5.2.3
After
INP.PCT
being
shown,
000000
(position
of
decimal
point
is
determined
by
CONFIG-FUNC-PERCEN setting) will be displayed, use PRINT, ACC, UNIT key to input the
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
26
CI-100A Indicator Operation & Service Manual
percentage of samples and Press TARE key to confirm. If the calculated unit-percentage-weight is
less than 0.5d, the indicator will display Pct.Er and go back to percent weighing mode, otherwise,
after the reasonable unit-percentage-weight being got, the scale will go back to percent weighing
mode. The got unit-percentage-weight can be saved after the power off and can be used next time.
7.6
Check Percent (percentage compare) in Percent weighing mode:
7.6.1
To make percentage compare function available, CONFIG-FUNC-COMPAR menu item should set to
YES, and high and low limitation of percentage should be set correctly according to following steps:
7.6.2
In percent weighing mode, Press DATA key more than 3s to input compare data of high and low.
7.6.3
After HIGH being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input high
percentage number and press TARE key to confirm. Annunciator of Hi will be shown in this step.
Press ZERO key to exit and back to percent weighing mode.
7.6.4
After Low being shown, 000000 will be displayed, use PRINT, ACC, UNIT key to input low
percentage number and press TARE key to confirm. Annunciator of Lo will be shown in this step.
Press ZERO key to exit and back to counting mode.
NOTE: If High number is 0 or is equal or less than low number, the comparison will be disabled.
7.6.5
After a reasonable limitation is set and compare is be active, one of the annunciator, HI, OK, LO will
be lighted, and the beeper will sound according to its setting in USER-BEEP.
8. BMI Working Mode
8.1
To make BMI working Mode be available, CONFIG-FUNC-BMI menu item should be set to YES and
factory setting should be also enable this function
8.2
To enter BMI Working mode:
8.2.1
When CONFIG-FUNC-ACCUMU= Yes: If In normal weighing mode, percent weighing mode, or
counting mode, long press FUNC key, one of WEIGH/COUNT/PERCEN will be shown, use ACC or
PRINT key to select BMI, then press TARE to confirm go to BMI mode.
8.2.2
When CONFIG-FUNC-ACCUMU= NO: press BMI (ACC) key, go to BMI mode.
8.3
After scale go to this mode, “CM.xxx” (means: last input height is xxx cm) or “IN.xx.x” (means: last
input height is xx.x inch) will be displayed, and to wait for input height: (1)to change height unit to cm
or inch by pressing DATA key; (2) to change height number by using PRINT, ACC(BMI) keys
(3)Press and hold down PRINT or ACC(BMI) key will increase or decrease number fast, (4) Press
TARE key to confirm the input. Press ON/OFF key to exit input data mode and back to BMI working
mode. The range of height is 50-250cm(19.7-98.4inch) and default is 170cm(66.9inch)
8.4
In this mode, when BMI number is shown (BMI annunciator is on also), or weight number is shown
(BMI and kg or lb annunciator are on), Press ACC key to select weight or BMI number to be
displayed, when weight is displayed, the weight unit can be selected by pressing UNIT key, and BMI
and weight unit will be displayed at same time.
8.5
In this mode, when current net weight is less than NLD.RNG, the indicator will go to display weight
number if CONFIG-FUNC-ACCUMU= No; or the indicator will back to original working mode if
CONFIG-FUNC-ACCUMU= Yes.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
27
CI-100A Indicator Operation & Service Manual
9. HOLD Function
NOTE:
In trade application, HOLD function may not be allowed, please check with your local
legal organization of measurement!
9.1
HOLD function can be used to freeze display number. In this mode, scale can catch a dynamic
number, hold a stable number, or average a unstable number, then HOLD (freeze) this number
temporary for user to watch or record. This function can be used in normal weighing mode, counting
mode and percent weighing mode. After entering HOLD mode, the speed of A/D converter can be
increased to 80Hz (if CONFIG-AD.H.SPD is set to YES) from original 10Hz for some dynamic
weighing applications. With the hold function, it is possible to weigh restless weighing samples such
as live animals, moving objects. The indicator provides special mode settings to accommodate
sample’s movements.
9.2
To make HOLD function be active, the CONFIG-FUNC-HOLD menu item must be set to YES; menu
items
of
USER-HOLD-HLD.MOD
/-AVG.TIM
/-HLD.TIM
/-HLD.RNG
/-STB.TIM,
USER-OTHER-NLD.RNG need be set to reasonable value.
To speedup sampling of weight, set CONFIG-AD.H.SPD menu item to YES.
To enter HOLD working mode, press down HOLD key when scale works in normal weighing mode,
counting mode or percent weighing mode.
9.3
There’re several HOLD mode to freeze display data:
(1) Positive Peak Number HOLD mode
(2) Negative Peak Number HOLD mode
(3) Toggle HOLD mode
(4) Average HOLD mode
(5) Auto HOLD mode
The following are details of these HOLD modes:
9.3.1
Positive Peak HOLD:
When USER-HOLD-HLD.MOD is set to PS.PEAK, the hold mode is positive peak hold mode. When
scale first enters this working mode, it will display the largest positive number that is from the time of
zero-pint set. After entering this working mode, scale will always catches and refresh positive larger
number and display it. To exit HOLD mode, press HOLD key again.
9.3.2
Negative Peak HOLD:
When USER-HOLD-HLD.MOD is set to NG.PEAK, the hold mode is negative peak hold mode. When
scale first enters this working mode, it will display the largest negative number that is from the time of
zero-point set. After entering this working mode, scale will always catches negative larger number and
display it. To exit HOLD mode, press HOLD key again.
9.3.3 Toggle HOLD:
When USER-HOLD-HLD.MOD is set to TOGGLE, the hold mode is toggle hold mode ---a manual Hold
function. After entering this working mode, scale will freeze and display number if scale is stable. Only
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
28
CI-100A Indicator Operation & Service Manual
the weight that is over USER-OTHER-NLD.RNG (zero 'dead' band ) can be held. To exit HOLD mode,
press HOLD key again. If the time of scale being unstable is more than USER-HOLD-STB.TIM, STB.ER
will be shown, press TARE key to start averaging again, or press HOLD key to exit.
9.3.4 Average HOLD:
When USER-HOLD-HLD.MOD is set to AVERAG, the hold mode is average hold mode. After entering
this working mode, scale will freeze and display number if scale is stable. If scale is not stable, but the
variation is less than USER-HOLD-HLD.RNG, scale will average data in USER-HOLD-AVG.TIM, then
freeze and display the number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen.
Scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If the time of scale
variation being over USER-OTHER-NLD.RNG is more than USER-HOLD-STB.TIM, STB.ER will be
shown, press TARE to start averaging again, or press HOLD key to exit.
9.3.5 Auto HOLD:
When USER-HOLD-HLD.MOD is set to AUTO, the hold mode is auto hold mode--- different subjects
can be weighed one after another without pressing any buttons. After entering this working mode, scale
will freeze and display number if scale is stable. If scale is not stable, but the variation is less than
USER-HOLD-HLD.RNG, scale will average data in USER-HOLD-AVG.TIM, then freeze and display the
number. Only the weight that is over USER-OTHER-NLD.RNG can be frozen. If the held weight is
moved away, and a new load put on the scale, scale will automatically hold new number of load. Scale
will exit HOLD mode according to the setting of USER-HOLD-HLD.TIM. If the time of scale variation
being over USER-OTHER-NLD.RNG is more than USER-HOLD-STB.TIM, STB.ER will be shown,
press TARE to start averaging again, or press HOLD key to exit.
9.4 In Positive or Negative Peak HOLD mode, the red HOLD (for LED version) or PEAK and HOLD (for LCD
version) announciator will be lighted, in other HOLD mode, green HOLD (for LED version) or HOLD (for
LCD version) announciator will be lighted. When HOLD announciator flash, the displayed number is live,
When HOLD announciator become steady, the displayed number is frozen.
10. Data Compare Function
10.1 Data compare function can be used in normal weighing mode, counting mode and percent weighing
mode, and call it as Check Weight, Check Counts and Check Percentage. When this function is enabled,
you can set a higher and a lower limitation of weight, counts or percentage independently, and these
limitation can be saved permanently. Then, the current data of weight , counts or percentage will be
compared with the setting limitation, and corresponding annunciator will be lighted.
10.2 To make data compare function be available, CONFIG-FUNC-COMPAR menu item should set to YES,
and high and low limitation should be set correctly according to following steps:
10.3 In normal weighing mode, counting mode or percent weighing mode, Press down DATA key more than
3s to enter input compare data of high and low mode.
10.4 After HIGH being shown, last setting data of high will be displayed, use PRINT, ACC, UNIT key to input
new number of high and press TARE key to confirm. Annunciator of Hi will be shown in this step. Press
ZERO key to exit and back to original working mode.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
29
CI-100A Indicator Operation & Service Manual
10.5 After Low being shown, last setting of low will be displayed, use PRINT, ACC, UNIT key to input new
number of low and press TARE key to confirm. Annunciator of Lo will be shown in this step. Press ZERO
key to exit and back to original working mode.
NOTE:
If High number is 0 or is equal or less than low number, the comparison will be disabled.
10.6 After a reasonable limitation is set and compare function is be active, one of announciators HI, OK, LO
will be lighted, and the beeper will sound according to its setting in USER-BEEP.
10.7 For details, please refer to section of section5.13, 6.6 and 7.6
11. Accumulation
11.1 Accumulation function can be used in normal weighing mode, counting mode and percent weighing mode,
When this function is enabled, you can accumulate current net weight, piece, and percentage. Note, only
the load on scale is larger than USER-OTHER-NLD.RNG, the displayed positive number can be added
up. The accumulation times and total can be displayed or printed.
11.2 To make data accumulation function be available, CONFIG-FUNC-ACCUMU menu item should set to
MANUAL or AUTO, Following are details.
11.3 When CONFIG-FUNC-ACCUMU is set to MANUAL, the stable and positive displayed net weight (must
be larger than USER-OTHER-NLD.RNG), piece or percentage can be accumulated by long pressing
TOTAL key, and indicator will display accumulation times first, and then display total of number. To avoid
repeating accumulation for same load, one load only can be accumulated once. So, before a new load put
onto the scale, the original load should be removed and let load on scale be smaller than
USER-OTHER-NLD.RNG.
11.4 When CONFIG-FUNC-ACCUMU is set to AUTO, the stable and positive displayed net weight (must be
larger than USER-OTHER-NLD.RNG), piece or percentage can be accumulated automatically, and
indicator will display accumulation times first, and then display total of number. To avoid repeating
accumulation for same load, one load only can be accumulated once. So, before a new load put onto the
scale,
the
original
load
should
be
removed
and
let
load
on
scale
be
smaller
than
USER-OTHER-NLD.RNG.
11.5 To view total, when display number is zero, long pressing TOTAL key, and indicator will display
accumulation times first, and then display total of number. It will alternative display accumulation times
and weight (or quantity, or percentages) until the TOTAL key being pressed again.
NOTE:
When HOLD function is enable, and scale is working in PEAK HOLD mode ( CONFIG-HOLD=YES,
USER-HOLD-HLD.MOD=PS.PEAK/NG.PEAK),
Accumulation
function
will
be
automatically
disabled!!!
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
30
CI-100A Indicator Operation & Service Manual
12. Calibration
Note:
(1) Before calibrate the scale, you should prepare a standard weight (more than 10% of FS weight) for
calibration.
(2) In following steps, to press ON/OFF will show “EXIT?”, and press ON/OFF again or press TARE will
exit calibration
12.1 Go to setup mode, select “CAL”, then press TARE to confirm to enter calibration mode.
12.2 After entering this mode, the number of this indicator has been calibrated will be shown first, this number
will be increased one after every calibration and calibration data saved, and this counter can’t be modified
or erased by any other ways, it counts from 0000 to 9999, when it reaches 9999, it starts over at 0000.
After the counter number being displayed, it will show “CAL.OFF” or “CAL-ON” according to the status of
the sealed calibration switch is OFF or ON. If the switch is OFF, the following steps can be done, but the
result will not be saved. Press TARE key to go to next step.
12.3 When “ZERO” is shown, use PRINT or ACC key to select do zero point calibration (refer step12.4), do
linearity calibration (refer step12.5), do Geographical calibration (refer step 12.6) or Input/view calibration
parameters value (refer step 12.7).
12.4 When ZERO is selected, remove all weight on scale and then press TARE key to confirm, the ZERO will
flash when in catching zero point state. After getting reasonable data, it automatically goes to step12.8
12.5 When LINE is selected, press TARE key to confirm to enter linearity calibration.
12.5.1 0% weight will be displayed after CAL.P0 being shown, remove all weight on scale and then press
TARE to confirm to calibrate the zero point; the zero weight will flash in catching zero point state. After
getting the reasonable zero-point data, the zero weight will become steady and then go to next step.
12.5.2 When first default standard weight is displayed after CAL.P1 being shown. It will be calibrated on
standard weight for first point. Put corresponding weight (more than 10%FS weight) onto scale. The
default standard weight is 100%FS.
Use PRINT, ACC, UNIT keys to input the value of the loaded weight. Before input this value, you
can long press DATA key to change weight’s unit to kg or lb. Press TARE key to confirm, then, the
indicator will flash the input standard weight.
When this weight number becomes steady, it means the stable and reasonable data
corresponding to the standard weight has been gotten.
After this, the indicator will automatically go to next step. If this point can’t be calibrated correctly
(maybe the weight load onto scale is too small, maybe the input data is incorrect…), it will display
“CAL.Er” and return back to step12.5.1 for re-calibration.
12.5.3 When End.y is shown and y is flashing, it’s waiting command to exit calibration or go on next
calibration. Use PRINT or ACC key to select yes or no, use TARE to confirm. If yes is selected, it will
go to step12.8 to end calibration; if no is selected, it will go to next step.
12.5.4 When 100%FS weight is displayed after CAL.P2 being shown. It will be calibrated on standard weight
for second point. Put corresponding weight (more than 10%FS weight, and larger than the weight
used on CAL.P1) onto scale. Next operation is same as what in step12.5.2
12.5.5 When End.y is shown and y is flashing, Use PRINT or ACC key to select yes or no, use TARE to
confirm. Similar with doing in step12.5.3
12.5.6 When third standard weight displayed after CAL.P3 being shown. It will be calibrated on standard
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
31
CI-100A Indicator Operation & Service Manual
weight for third point. Put corresponding weight (more than 10%FS weight, and larger than the weight
used on CAL.P2) onto scale. Next operation is same as what in step12.5.2.
12.5.7 When the stable and reasonable data corresponding to the standard weight has been gotten. The
indicator will automatically go to Step12.8. Otherwise, it will display “CAL.Er” and return back.
12.6
When GEO is selected, press TARE key to confirm to enter Geographical Adjustment
12.6.1 When “CODE” is shown, use PRINT or ACC key to select geographical position code (refer step12.6.2)
or input user local gravity value directly (refer step12.6.3).
12.6.2 When CODE is selected, select the position code of scale being used (00-70) according to the
elevation and latitude from Table12-1 by using PRINT, ACC, UNIT keys. Press TARE key to confirm.
12.6.3 When GRAVT is selected, Use PRINT, ACC, UNIT keys to input the gravity value of the position that
scale is used (9.76183-9.99999). Press TARE key to confirm.
NOTE: Only an authorized manufacturer’s representative or certified verification personnel
may make these changes. Changing the geographical setting alters the calibration
values !!!
12.7
12.7.1
12.7.2
12.7.3
12.7.4
When INPUT is selected, press TARE key to confirm to enter Input calibration parameters value that
were got before, or view current calibration parameters value.
All parameters about calibration are divided to 18 pages to be displayed on LCD by “nn:xxxx” format
(“nn” is a decimal number of page, “xxxx” is an hexadecimal value of parameter, e.g. 02:85E2).
01-02 pages: zero code;
03-04 pages: standard weight of CAL.P1;
05-06 pages: codes of CAL.P1;
07-08 pages: standard weight of CAL.P2;
09-10 pages: codes of CAL.P2;
11-12 pages: full capacity net code;
13-14 pages: the coefficient of weight fine-tune;
15-16 pages: gravity value of calibration location;
17-18 pages: gravity value of the location that the scale is used at.
When no digit blink on LCD, that means calibration parameters value are being viewed, and use UNIT
key to view next page, use ZERO key to exit.
When parameters value are being viewed, use DATA key to prepare to modify, When first digit is
blinked, that means the value is being modified, and you can use UNIT key to make next digit flash (if
current flashing position is the last one, next page value will be shown), use PRINT, ACC, UNIT key to
input number, use TARE key to confirm.
In this mode, press down PRINT key more than three seconds, these parameters will be sent out from
COM1, the print out format is <LF>nn:xxxx<CR>; there’re totally eighteen lines.
12.8 After the indicator gets all need data, it will calculate and store all calibration parameters into EEPROM,
or after finishing calibration works, it will display CAL.End. At last, it will re-start and go back to original
mode.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
32
TABLE12-1: Location Code for different elevation and latitude
CI-100A Indicator Operation & Service Manual
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
33
CI-100A Indicator Operation & Service Manual
13. Weight Fine-tune
With this function, the user can adjust displayed weight a little, and no need standard weight. But please
Note:
(1)The scale must have been calibrated before this adjustment
(2)The range of adjustment is “(current displayed weight) x (0.9-1.1)”. it means the range is about ±10%
(3)The “CONFIG-REGULA =NONE” and “CONFIG-FUNC-WT.ADJ=YES” must be set.
(4) Only an authorized manufacturer’s representative or certified verification personnel may make
these changes. Changing this value alters the calibration values !!!
13.1 To enter this mode, turn on indicator, after indicator displaying 0 weight, put a load (suppose: its correct
weight is 1230.0lb) onto scale, then indicator will display the load’s weight, say “1234.5 lb”. Press down
TARE and ZERO at same time until first digit flashes, this means indicator has entered into “weight
fine-tune” mode.
13.2 Use
to input correct weight (1230.0). After confirmed by TARE, the active correct
weight will be displayed and no any digit will be flashed. After this, displayed weight will be adjusted by
this ratio (1230.0/1234.5) and this ratio will keep active until next modification on it.
13.3 To remove effect of this ratio, there’re two ways: “12.3.1 way” and “12.3.2 way”
13.3.1 Do standard calibration, like in “12.CALIBRATION”
13.3.2 Move away weight on scale, Press ZERO to make 0 displayed, Put a load onto scale, a number
will be displayed, suppose it’s 1230.0lb (but original number is 1234.5) ; Press down TARE and
ZERO at same time until first digit flashes, this means indicator has entered into “weight fine-tune”
mode.
Press
key, the displayed weight will be restored to 1234.5, and then press
to confirm
and exit to normal weighing mode.
14. View ADC output Code
14.1 In this mode, you can examine the stability of weighing system, the increment value of ADC output code
corresponding to the loaded weight.
Note:
① The increment of ADC code for FS weight must be larger or equal to 10 times of selected display
division; otherwise, the calibration cannot be properly completed. e.g. The display division is 0.1kg.
Load 100kg standard weight on the platform, the increment of ADC code should be at least more than
10x100kg/0.1kg= 10x1000=10000. In this case, the scale can be calibrated. Otherwise, smaller
division needs to be Chosen.
② The variation of ADC code should be small; otherwise, the calibration cannot properly complete also.
14.2 To go to this working mode, press down SETUP until CONFIG is shown, using PRINT , ACC and TARE
key to to go to MISC - Code item, press TARE to enter this mode and display ADC output raw code.
14.3 In this mode, first press TARE key can set current code as a reference zero, and then to display net code,
press TARE again to clear this reference and display gross code.
14.4 In this mode, press UNIT key to select displaying code that has been filtered by no-filter, filter1 or filter1
and filter2, and corresponding announciator Lo, OK, HI will be lighted.
14.5 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
34
CI-100A Indicator Operation & Service Manual
15. View or Calibrate Power Voltage
15.1 In this mode, you can examine the voltage of battery, or you can examine the voltage that regulated out
from AC adaptor when no battery is used, you also can calibrate the displayed voltage and set the
voltage value of low battery point.
Note:
(1)The end customer normally no need to calibrate the displayed voltage, these have been done in
factory.
(2) The normal displaying voltage is 4.0V-6.0V
15.2 To go to this working mode, press down SETUP until CONFIG is shown, using PRINT , ACC and TARE
key to to go to MISC - VoL item, press TARE to enter this mode and display battery voltage.
15.3 If the voltage is sure not correct, to calibrate the voltage according to following steps:
15.3.1 Prepare a DC power supply which output voltage can be adjusted from 5V to 8V, output current must
be larger than 0.5A. Power off the indicator, move away AC adaptor, Connect this DC power to battery
connector on main board, adjust voltage to about 5V, power on the indicator, enter battery voltage
display mode by the way of step15.2.
15.3.2 Press down UNIT until CAL.5V is shown, adjust voltage to 5V, press TARE key to confirm 5V
calibration.
15.3.3 When CAL.6V is shown, adjust voltage to 6V, press TARE key to confirm 6V calibration. When CV.End
is shown, that means the voltage calibration is completed and then exit to display voltage.
15.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
16. View or Set Time
16.1 After entering SETUP mode (by press down SETUP key more than 3s), using PRINT or ACC key to
select MISC-TIME item, press TARE to display current time.
16.2 Time display Format is: xx:xx:xx(hh-mm-ss) , 24h format
16.3 Press down UNIT more than 3s to enter modification time mode. Using PRINT, ACC, UNIT, TARE keys
to modify current time. If time of no operation is more than 5s, it will automatically exit modification mode.
16.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
17. View or Set Date
17.1 After entering SETUP mode (by press down SETUP key more than 3s), using PRINT or ACC key to
select MISC-DATE item, press TARE to display current time.
17.2 Date display Format is: xx.xx.xx(yy-mm-dd)
17.3 Press down UNIT more than 3s to enter modification date mode. Using PRINT, ACC, UNIT, TARE keys
to modify current date. If time of no operation is more than 5s, it will automatically exit modification mode.
17.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
35
CI-100A Indicator Operation & Service Manual
18. View Firmware Version
18.1 Press down SETUP until CONFIG is shown, using PRINT or ACC key to select MISC-VER item, press
TARE to display current Version.
18.2 Firmware Version display Format is:
Vxx.yy,
xx is hardware version, yy is software version
18.3 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
19. Display Test
19.1 Press down SETUP more than 3s to enter SETUP mode, using PRINT or ACC key to select
TEST-DISP.TST item, press TARE to enter test display mode and all segments will be lighted first.
19.2 In this mode, every pressing of ACC key will light next segment, every pressing of UNIT will light next
digit, press PRINT will automatically light all segments and all digits.
19.3 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
20. Keyboard and Buzzer Test
20.1 Press down SETUP more than 3s to enter SETUP mode, using PRINT or ACC key to select
TEST-key.tst item, press TARE to enter test keypad mode, and key. -- will be displayed.
20.2 In this mode, press a key, the value of this key will be displayed on -- position and buzzer will beep
whatever USER-BEEP-KEY item is set.
20.3 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
21. Serial Port1/2 (COM1/2) Receiving Test
21.1 Before test the receiving function of COM1 or COM2, a cable is need to connect a PC and this indicator,
and a software be similar with Super Terminal of Windows is also need to run on PC to send bytes to this
indicator. Please note: baud rate is selected by USER-COM1/2-BAUDRT, 8N1 byte format is fixed,
Hex data (0x00 – 0xff) are used.
21.2 Press down SETUP more than 3s to enter SETUP mode, using PRINT or ACC key to select
TEST-COM1.RD or TEST-COM2.RD item, press TARE to enter test COM1/2 receiving function, and
rd1.-- or rd2.--will be displayed first.
21.3 In this mode, received hex data (0x00 – 0xff) will be displayed on -- position.
21.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
22. Serial Port1/2(COM1/2) Transmitting Test
22.1 Before test the transmitting function of COM1 or COM2, a cable is need to connect a PC and this
instrument , and a software be similar with Super Terminal of Windows is also need to run on PC to
receive bytes from this instrument. Please note: baud rate is selected by USER-COM1/2-BAUDRT, 8N1
byte format is fixed, Hex data (0x00 – 0xff) are used..
22.2 Press down SETUP more than 3s to enter SETUP mode, using PRINT or ACC key to select
TEST-COM1.TD or TEST-COM2.TD item, press TARE to enter test COM1/2 transmitting function, and
td1.-- or td2.-- will be displayed first.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
36
CI-100A Indicator Operation & Service Manual
22.3 In this mode, transmitted hex data (0x00 – 0xff) will be displayed on -- position, and PRINT, ACC, UNIT,
TARE keys can be used to modify transmitted data.
22.4 Press SETUP key to return to last menu item, press ON/OFF key to prepare to exit this mode
23. Details about Serial Communication
23.1 COM1 is RS232, communication wires come from RS232 connector, and TXD0, RXD0 and GND are
used. Please refer to section 24.5 connector details
23.2 COM2 is USB used as a virtual RS232, communication wires come from USB connector, and TXD1,
RXD1 and GND are used, Please refer to section 24.6 for connector details.
23.3 The baud rate and byte format is set by USER-COM1/2-BAUD.RT and USER-COM1/2-BYT.FMT.
Responses to serial commands will be immediate, or within one weight measure cycle of the scale. One
second should be adequate for use as a time-out value by remote (controlling) device.
23.4 The length of each item in a transition string:
23.4.1
Reading data --- 6bytes
Data polarity ----1byte: “-” for negative, and followed the first digit; “ ” for positive.
Decimal point ---1byte: “.”
Measure unit ----1-5bytes:“ lb”,“ kg”, “lb:oz”, “pcs”,“%”, Units are always lower case, left aligned
Current status-- 4bytes
23.4.2 If the weight is overcapacity, the scale will return eight “^” characters (the field of polarity, decimal
point, weight data is filled by “^”).
23.4.3 If the weight is under capacity, it will return eight “_” characters (the field of polarity, decimal point,
and weight data is filled by “_”).
23.4.4 If the zero point is error, it will return eight “-“ characters (the field of polarity, decimal point, and
weight data is filled by “-”).
23.4.5 Useless leading 0 before digits is suppressed. Reading weight is right aligned.
23.5 Key to symbols used
<LF>
<CR>
<ETX>
<SP>
H1H2H3 H4
Line Feed character (hex 0AH)
Carriage Return character (hex 0DH)
End of Text character (hex 03H)
Space (hex 20H)
Four current status bytes
<P>
Polarity character: “-” or “ ”
W 1---W6
<DP>
U1U2 U3U4U5
<Add>
<Prompt>
Reading data, 1-6 bytes (six digits)
Decimal point
Measure units, kg, lb, lb:oz , % or pcs; 2-5 bytes
Address of scale; 2 bytes (00-99)
Prompt characters of output content; max. 11bytes
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
37
CI-100A Indicator Operation & Service Manual
The bit definition of H1H2H3 H4:
Bit
0
1
2
3
4
5
6
7
Byte 1 (H1)
Byte 2 (H2)
0=stable
1= not stable
0= not at zero point
1= at zero point
0=RAM ok
1= RAM error
0= eeprom OK
1= eeprom error
always 1
always 1
always 0
parity
Byte 3 (H3)
Byte 4 (H4)
0= not under capacity
1= under capacity
0= not over capacity
1= over capacity
0=ROM ok
1=ROM error
0=calibration ok
1=calibration error
00=compare disable
01=lower limit
10=ok
11=upper limit
00=normal weighing
01=count weighing
10=percent weighing
11=other mode
0= gross weight
1= net weight
0=initial zero ok
1=initial zero error
0=not in HOLD
1=in HOLD
0=battery ok
1=low battery
always 1
always 1
always 1
Parity
always 1
always 1
always 1
parity
always 1
always 1
always 0
Parity
23.6 Communication Details when USER-COM1/2-LAYOUT is set to SINGLE:
23.6.1 Commands and response
23.6.1.1 Command: W<CR> (57h 0dh), request current reading
Response:
<LF>^^^^^^^^U1U2 U3U4U5<CR><LF> H1H2H3 H4<CR><ETX>---over capacity
<LF>_ _ _ _ _ _ _ _ U1U2 U3 U4U5<CR><LF> H1H2H3 H4<CR><ETX>---under capacity
<LF> - - - - - - - - U1U2 U3 U4U5<CR><LF> H1H2H3 H4<CR><ETX>---zero-point error
Note: U1U2 U3 U4U5 is 1,2,3 or 5 bytes according to current unit: %, kg, lb, pcs, lb:oz
<LF><P>W1W2W3W4W5<DP>W6 U1U2 U3 U4U5<CR><LF> H1H2H3 H4 <CR><ETX>---normal data
Note: (1) The decimal point position is determined by CONFIG-PRIM.D
(2) If current unit is “lb:oz”, the format will be similar with following:
<LF><P>W1W2W3lb<SP>W4W5<DP>W6oz<CR><LF> H1H2H3 H4 <CR><ETX>
23.6.1.2 Command: S<CR>
(53h 0dh) , request current status
Response: <LF> H1H2H3 H4<CR><ETX>
23.6.1.3 Command: Z<CR> (5ah 0dh)
Response: Zero function is activated (simulate ZERO key) and it returns to current scale status.
<LF> H1H2H3H4<CR><ETX>
If ZERO function cannot be activated, it will return to current scale status.
23.6.1.4 Command: T<CR>
(54h 0dh)
Response: TARE function is activated (simulate TARE key), and then returns scale status.
<LF> H1H2H3 H4 <CR><ETX>
If TARE function cannot be activated, it will return to current scale status.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
38
CI-100A Indicator Operation & Service Manual
23.6.1.5 Command: U<CR>
(55h 0dh)
Response: Changes units of measure (simulate UNIT key) and return scale status with new units,
The new measure unit should be allowed to use
<LF> U1U2 U3 U4U5<CR><LF> H1H2H3 H4<CR><ETX>
23.6.1.6 Command: L<CR>
(4ch 0dh)
Response: If Hold function can be activated, it will enable/disable hold function (simulate HOLD key),
and returns scale status.
<LF> H1H2H3H4<CR><ETX>
23.6.1.7 Command: X<CR> (58h 0dh)
Response: power off the scale, just like press down the ON/OFF key to turn off the scale.
23.6.1.8 Command: all others
Response: Unrecognized command
<LF>? <CR><ETX>
23.6.2 Summary of Command and Response:
Command
ASCII
Response
HEX
Read scale weight:
W<CR>
57 0d
<LF>^^^^^^^^U1U2 U3U4U5<CR><LF> H1H2H3 H4<CR><ETX>---over capacity
<LF>________U1U2U3U4U5<CR><LF> H1H2H3 H4<CR><ETX>---under capacity
<LF>-------- U1U2 U3 U4U5<CR><LF> H1H2H3 H4<CR><ETX>---zero-point error
<LF><p>W1W2W3W4W5<dp>W6U1U2U3U4U5<CR><LF>H1H2H3H4<CR><ETX>
---normal data
S<CR>
53 0d <LF> H1H2H3H4<CR><ETX>; read scale status
Z<CR>
5a 0d <LF> H1H2H3H4<CR><ETX> ; simulate ZERO key
T<CR>
54 0d <LF> H1H2H3H4<CR><ETX> ; simulate TARE key
U<CR>
55 0d <LF> U1U2 U3 U4U5<CR><LF> H1H2H3H4<CR><ETX>; simulate UNIT key
L<CR>
4c 0d <LF> H1H2H3H4<CR><ETX>; simulate HOLD key
X<CR>
58 0d power off the scale, simulate OFF key
others
<LF>? <CR><ETX>
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
39
CI-100A Indicator Operation & Service Manual
23.7 Communication Details when USER-COM1/2-LAYOUT is set to MULTIPLE:
23.7.1 Output string frame:
<LF><Prompt><p>W1W2W3W4W5<Dp>W6 U1U2 U3 U4U5<CR>
……
---- Line number and content are determined by setting of USER-OUT1/2-xxxx
<LF><Prompt>H1H2H3 H4<CR>
---- USER-OUT1/2-STATUS is set to YES
……
<LF>
(1)The decimal point position is determined by CONFIG-PRIM.D
(2)The unit position and bytes is determined by which current unit is used.
(3)The details of <Prompt> refer to the content in 4.3USER Submenu.
(4)In hold mode, if ADC conversion speed is set high speed (80Hz), and USER-COM1/2-LAYOUT is
set to MULTIPLE, and many contents are selected to output, the output contents from COM1 or COM2 may
not catch up with the data processed in indicator, So, if you want to watch “real time” data, you need to select
fewer output contents and set higher baud rate for C<CR>
--- USER-OUT1/2-LINE is set to LINE1/2/3/4
……
---The number of blank lines is determined by USER-OUT1/2-LINE setting
<ETX>
--- Last byte of string frame
23.7.2 Examples of some layout when USER-OUT1/2-xxxx is set to YES (USER-COM2-EN.ADDR=No):
23.7.2.1 In weighing mode:
SCALE ID:
123456
GROSS:
123lb 4.56oz
TARE:
11lb 2.22oz
NET:
112lb 2.34oz
ACC. N:
8
TOTAL:
789lb 15.2oz
DATE:
2011-06-12
TIME:
12:34:56
A/D CODE:
1234567
VOLTAGE:
6.7V
STATUS:
bpq2
23.7.2.2 In counting mode:
SCALE ID:
123456
GROSS:
1234.55kg
TARE:
12.15kg
NET:
1222.40kg
QUANTITY:
24448pcs
PIECE WT:
0.05kg
ACC. N:
10
TOTAL:
23456pcs
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
40
CI-100A Indicator Operation & Service Manual
DATE:
2011-06-12
TIME:
12:34:56
A/D CODE: 1234345
VOLTAGE:
6.7V
STATUS:
bpq2
23.7.2.3 In percent weighing mode:
SCALE ID:
123456
GROSS:
12345lb
TARE:
10lb
NET:
12335lb
PERCENTAGE:
91.4%
1% REF. WT:
ACC. N:
135lb
3
TOTAL:
271.6%
DATE:
2011-06-12
TIME:
12:34:56
A/D CODE:
1231234
VOLTAGE:
6.7V
STATUS:
bpq2
23.7.2.4 In BMI mode:
SCALE ID:
GROSS:
123456
110.0kg
TARE:
10.0kg
NET:
100.0kg
HEIGHT:
170cm
BMI :
34.6
DATE:
2011-06-12
TIME:
12:34:56
A/D CODE:
1231234
VOLTAGE:
6.7V
STATUS:
bpq2
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
41
CI-100A Indicator Operation & Service Manual
23.8 Communication Details when USER-COM1/2-LAYOUT is set to EH-SCP:
23.8.1 This protocol of serial communication is similar to TOLEDO PS60 protocol. The baud rate and data
format is set by User menu.
23.8.2 Output status bit meaning:
Bit
0
1
2
3
4
5
6
7
Status Byte
0=Stable weight data
1=Scale in motion
0= Within weighing range
1= Over capacity
0=Within weighing range
1= Under zero
0= Within range
1= Outside zero capture range
0= Not at center of zero
1= Center of zero
always 1
always 1
parity
23.8.3 Summary of Command and Response:
Command
ASCII
Response
HEX
Read scale weight:
①normal data
<STX> W1 W2<dp>W3W4W5<CR>
②if current weight is invalid
<STX>?<Status Byte><CR>
W
57
Z
5a
L
4c
Switch to and send standard weight. Same as W above
K
4b
Switch to and send metric weight. Same as W above
others
Simulate ZERO key:
<STX>?<Status Byte><CR>
;
Un-known commands:
<STX>?<Status Byte><CR>
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
42
CI-100A Indicator Operation & Service Manual
23.9 Communication Details when USER-COM1/2-LAYOUT is set to SCP-12:
23.9.1 This protocol of serial communication is similar to with NCI3835 protocol. The baud rate and data
format is set by User menu.
23.9.2 Output status bit meaning:
Bit
Status Byte1
Status Byte2
0=Scale in motion
1=Stable
0= Scale at zero
1= Not at zero
0=RAM error
1= RAM okay
1 = Under capacity
0 = Not under capacity
1 = Over capacity
0 = Not over capacity
1 = ROM error
0 = ROM okay
4
0= EEPROM error
1= EEPROM okay
Always 1
1 = Faulty calibration
0 = Calibration okay
Always 1
5
6
7
always 1
always 0
parity
always 1
always 0
parity
0
1
2
3
23.9.3 Key to symbols used:
<ETX> End of TeXt character (03 hexadecimal).
<LF>
Line Feed character (0A hex).
<CR>
Carriage Return character (0D hex).
Xxxxxx Weight characters from display including minus sign and out-of-range characters.
p
Polarity character (ie ‘-’ for negative, space for positive)
hh
Two status bytes. (see 23.9.2)
UU
Units of measure (LB, KG or OZ all upper case).
23.9.4 Summary of Command and Response:
Command
ASCII
Response
HEX
Returns decimal lb, kg or oz weight, units and status.
<LF>pxxx.xxUU<CR>hh<ETX>
W<CR>
Returns ounces weight with units plus scale status.
57 0D <LF>p00xxxxxOZ<CR>hh<ETX>
Scale status only if initial zero error.
<LF>hh<CR><ETX>
S<CR>
53 0D Read scale status :
Z<CR>
5A 0D Simulate ZERO key:
others
<LF>hh<CR><ETX>
no response from scale.
Un-known commands: <LF>?<CR>
23.9.5 If your indicator need work with UPS worldship, you can try following settings:
(1) USER-COM1(or 2)-BAUD.RT=4800
(2) USER-COM1(or 2)-BYT.FMT=7E1
(3) USER-COM1(or 2)-LAYOUT=SCP-12
(4) Set scale port to NCI3835 in UPS Worldship.
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
43
CI-100A Indicator Operation & Service Manual
24. Connectors and Jumpers
24.1 Overview of Connectors or jumpers on PCB
0
0
1
2
3
7
13
12
14
11
15
10
16
9
17
8
18
7
19
6
20
5
21
4
22
3
9
8
6
7
5
6
4
5
3
4
2
3
1
2
1
1
2
4
6
8
10
1
3
5
7
9
23
2
24
1
2
3
4
8
7
6
5
4
3
2
1
5
6
1
2
7
3
2
2
1
1
24.2 Load Cell Connector
PIN # DEFINITION
IN/OUT/POWER
ELECTRICAL LEVEL
1
Excitation +
Power output
5±0.3 Vdc (≤0.12A)
2
Sense +
Power input
5±0.3 Vdc
3
Excitation-
Power ground
0Vdc
4
Sense -
Power input
≤0.5 Vdc
5
Signal +
Signal Input
2.5±0.3 Vdc
6
Signal -
Signal Input
2.5±0.3 Vdc
7
Shield
-
-
24.3 ADP---adapter power input connector
PIN #
DEFINITION
IN/OUT/POWER
ELECTRICAL LEVEL
1
Adapter input voltage +
Power input
7-9Vdc (I≥0.5A)
2
Adapter input voltage – (GND)
Power ground
0Vdc
24.4 BAT---Battery power input Connector
PIN #
DEFINITION
IN/OUT/POWER
ELECTRICAL LEVEL
1
Battery input voltage +
Power input
4-6.8Vdc
2
Battery input voltage – (GND)
Power ground
0Vdc
3
Temperature sensor on Battery input
Power ground
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
44
CI-100A Indicator Operation & Service Manual
24.5 SIO---Serial Input Output Connector
PIN #
DEFINITION
IN/OUT/POWER
ELECTRICAL LEVEL
1
RS485 signal A (if RS485 installed)
Input/output
0-5Vdc
2
RS232 Transmit on COM1
Output
-12 to +12Vdc
3
RS232 Receive on COM1
Input
-12 to +12Vdc
GND
Power ground
0Vdc
RS485 signal B (if RS485 installed)
Input/output
0-5Vdc
4
5
6
7
8
9
24.6 J1---USB Connector for virtual RS232 #1 and USB power supply
PIN #
DEFINITION
IN/OUT/POWER
ELECTRICAL LEVEL
1
VDD
Power output
5±0.3 Vdc
2
RXD Receive on COM2
Input
0-5Vdc
3
TXD Transmit on COM2
Output
0-5Vdc
4
GND of VDD
Power ground
0Vdc
5
GND1 of VUSBH
Power ground
0Vdc
6
USB Power DC/DC select
Output
0-5Vdc
7
USB Power DC/DC output
output
6±0.3 Vdc
24.7 KCAK Jumper set:
CONNECTED
1-2
PINS
FUNCTION
Calibration Disabled
2-3
Calibration Enabled
24.8 JP1 Jumper:
CONNECTED
PINS
FUNCTION
1-2
Two shorter on pin1-2: 4 wires Load cell is used
2-3
Two shorter on pin2-3: 6 wires load cell is used
24.8 ICM1 Socket: used for optional Bluetooth, RS485 or USB interface
24.9 JTAG and IOSEL used by Manufacture
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
45
CI-100A Indicator Operation & Service Manual
25. Meaning of Some Symbols and Troubleshooting
25.1 Meaning of Symbols:
25.1.1 0﹉﹉﹉﹉ ------ Zero is over the setting range
25.1.2 0﹍﹍﹍﹍ ------ Zero point is below the setting range
25.1.3 Ad﹉﹉﹉ ------ Signal to ADC is over max. range)
25.1.4 Ad﹍﹍﹍ ------ Signal to ADC is below min. range
25.1.5 ﹉﹉﹉﹉
------ Weight is over upper limitation, or display data is over limitation
25.1.6 ﹍﹍﹍﹍
------ Weight is below lower limitation
25.1.7 EEP.E1
------ CONFIG or CAL parameters are not correctly set
25.1.8 EEP.E2
------ USER parameter is not correctly set
25.1.9 Lo.bAt
------ Battery voltage is lower than setting.
25.1.10 CAP.- - -
---- Next displaying content is Capacity
25.1.11 CAP.ER
-----
Parameters about Capacity is not correct
25.1.12 CAL.Px
-----
Calibration on point(x)
25.1.13 CAL.OFF -----
Calibration Seal Switch is on OFF position
25.1.14 CAL.ON
Calibration Seal Switch is on ON position
25.1.15 CAL.Er
---------
Calibration error, maybe input data or loaded weight is incorrect, unstable, un-linear
25.1.16 CAL.End -----
Calibration is end
25.1.17 OFF
-----
Power OFF the indicator
25.1.18 STB.ER
------ Unstable time is lager than setting of USER-HOLD-STB.TIM
25.1.19 AC.xxxx
----- Accumulation times is xxxx
25.1.20 PR.TARE
----- To Preset TARE weight
25.1.21 COMP
---- To go to input COMPARE data mode
25.1.22 HIGH
----- To input HIGH limitation data of Comparison
25.1.23 LOW
----- To input LOW limitation data of Comparison
25.1.24 SPL.Lo ---------- Sample load weight of low point.
25.1.25 SPL.HI ----------- Sample load weight of high point.
25.1.26 SPL.PWT ------- Sample goods weight to calculate piece weight
25.1.27 INP.PCS --------- input pieces number of weighted goods
25.1.28 UNIT.KG --------- Unit kg is selected
25.1.29 UNIT.LB --------- Unit lb is selected
25.1.30 PWT.ER --------- Piece weight is error, it’s too small (<0.5d).
25.1.31 SPL.PCT -------- Sample goods weight to calculate
25.1.32 INP.PCT --------- input percentage of weighted goods
25.1.33 PCT.ER ---------- Unit-Percentage -Weight is too small (<0.5d).
25.1.34 CACU.ER---------Internal calculation overflows
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
46
CI-100A Indicator Operation & Service Manual
25.2 Troubleshooting
SYMPTOM
Ad﹉﹉﹉
Ad﹍﹍﹍
0﹉﹉﹉﹉
0﹍﹍﹍﹍
PROBABLE
Load
cell
wires
CAUSE
to
indicator
REMEDY
are
Make sure wires are ok and correctly
incorrectly connected, or shorted, or
connected. Replace load cell or ADC
opened; or ADC, load cell are damaged
chip, Service required.
Weight reading exceeds Power On Zero
limit.
Perform zero calibration.
Weight reading below Power On Zero
zero calibration.
or The weight value cannot be displayed
in the current unit of measure because it
exceeds 6 digits..
﹍﹍﹍﹍
EEP.E1
EEP.E2
Install platform on scale. Perform
limit.
Weight reading exceeds Overload limit,
﹉﹉﹉﹉
Make sure scale platform is empty.
Weight reading below Under load limit.
CONFIG or CAL parameters are not
correctly set
Reduce load on scale until weight
value can be displayed. Use a more
appropriate unit of measure. Re-set
some parameters of CONFIG or
UAER.
Install platform on scale.
Perform zero calibration
Re-set items in CONFIG, do
calibration
USER parameter is not correctly set
Re-set items in USER
Set PRIM.N/PRIM.d/SECND.n to
CAP.ER
Capacity parameters are not correct
correct number, make sure capacity
not more than 6 digit
CAL.Er
Calibration error, maybe input data or
Input correct data, load correct
loaded weight is too small, too big,
weight onto platform, Service
unstable, un-linear
required
Piece weight is error, it’s too small
PWT.ER
(<0.5d), The weight on the platform is
too small to define a valid reference
Use a greater weight for the sample.
weight.
Unit-Percentage -Weight is error, it’s too
small (the weight of 1%, 0.1%, or 0.01%
PCT.ER
determined by
Use a greater weight for the sample.
CONFIG-FUNC-PERCNT
is less than 0.5d)
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
47
CI-100A Indicator Operation & Service Manual
CACU.ER
STB.ER
Internal calculation overflow
Adjust the value of the PWT or PCT
USER-HOLD-STB.TIM is too short,
Set USER-HOLD-STB.TIM longer,
USER-HOLD-HLD.RNG is too small,
or set USER-HOLD-HLD.RNG
other failure
bigger. Service required
Power cord not plugged in or properly
connected. Power outlet not supplying
Not turn on.
electricity. Battery discharged. Other
failure.
Check power cord connections.
Make sure power cord is plugged
into the power outlet. Check power
source. Replace batteries. Service
required.
Cannot zero the
Load on scale exceeds allowable limits.
Remove load on scale.
display or will not zero
Load on scale is not stable.
Wait for load to become stable.
when turned on.
Load cell damage.
Service required.
Cannot display weight
in desired weighing
unit.
Unit not set to enable, or d≥5oz,when
unit is lb:oz
Enable unit in CONFIG-UNITS
Battery symbol is
empty or Lo.bAt
is
Batteries are discharged.
Charge batteries
shown
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
48
CI-100A Indicator Operation & Service Manual
26. Display Character
ASCII
LCD/LED Show
ASCII
LCD/LED Show
ASCII
0
A
N
1
B
O
2
C
P
3
D
Q
4
E
R
5
F
S
6
G
T
7
H
U
8
I
V
9
J
W
K
X
L
Y
M
Z
LCD/LED Show
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
49
CI-100A Indicator Operation & Service Manual
27. Packing List
No.
CONTENT
QTY
1
Indicator
1
2
User manual
1
3
Swivel bracket
1
4
1x CR2032 battery for RTC
Optional
5
AC Adapter
1
6
RS232 cable
Optional
7
USB cable
Optional
8
Wireless (Bluetooth) Module
Optional
9
4xAA Alkaline batteries
Optional
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
50
CI-100A Indicator Operation & Service Manual
28. Version History
VERSION
V1.0
V1.1
V1.2
V1.3
V1.4
DESCRIPTION
Initial version
DATE
2011-09-21
(1)Add Geographical Adjustment in Calibration
(2)Add Description for Connectors and Jumpers on PCB in Section24
(1)Add Fig. of load cell wire go to indicator from bottom
(2)Modify 1.6.5.1 section, add 1.11.5 and 1.13.11
(3)Modify “3. Summary of Key function”, add BMI key function
(4)Modify 4.5 MISC-VOL; 4.6TEST
(5)Add “ 8. BMI Working Mode”
(6)Modify the content in TABLE12-1
(7)Modify “15. Display and Calibrate Power Voltage”
(8)Modify 23.1 and 23.2 in “23. Details about Serial Communication”
(9)Add 23.8 section about EH-SCP
(1)Modify some default setting in configuration and user parameters
(2)Modify power supply to 7-9V,500mA;
(3)Modify the max number of load cell that can be powered to 4(x350Ω)
(1) Delete BMI faceplate in “2.Faceplate”
(2) Delete BMI key in “3.Summary of key function”
(3) Delete “1.6.5.3 4 holes socket used”
2011-10-07
2011-12-6
2012-1-14
2012-3-14
V1.5
(1) Modify the range of zero tracking and center of zero point
(2) Add prompt for the prohibition of lb:oz and HOLD for trade application
2012-05-15
V2.0
(1) Add unit “g” and “oz” and modify CONFIG-UNITS item and modify
Table5-1/2 in section5.12
(2) Add “change weight unit” in calibration in section12.5.2
(3) Add contents about Bluetooth module, in section24
2012-6-15
V04.11
(1)
(2)
(3)
(4)
(5)
2013-08-02
V04.12
(1) Add calibration parameters print out function in calibration section
(2) Modify the operation of accumulation
2013-08-06
(1) Delete some ways of load cell connect to indicator
(2) Modify the description in CONFIG-PRIM.Ut
(3) Here, V04.12.01 means firmware version is V04.12, this user manual
modification version is 01 on V04.12
2013-08-15
V04.12.01
Add more controlling of Bluetooth for end user
Add 7O2 & 7E2 selection in USER-COMx-BYT.FMT menu
Add serial communication protocol SCP-12 of NCI3835
Modify EH-SCP serial communication protocol
Delete displaying indicator version, voltage, when power on
────────────────────────────────────────────────────────
CAS-USA Corporation
www.cas-usa.com
51