Dissolved Oxygen: Hach LDO® Probe Process - Mercado

Dissolved Oxygen: Hach LDO® Probe
Process Instruments
How does Luminescent Dissolved Oxygen work?
The HACH LDO sensor is coated
with a luminescent material. Blue
light from an LED is transmitted to
the sensor surface. The blue light
excites the luminescent material. As
the material relaxes it emits red light.
The time from when the blue light
was sent and the red light is emitted
is measured. The more oxygen that
is present the shorter the time it
takes for the red light to be emitted.
This time is measured and correlated
to the oxygen concentration.
Between the flashes of blue light a
red LED is flashed on the sensor and
used as an internal reference.
Primary Applications
• Wastewater
• Industrial Water
• Drinking Water
Specifications*
LDO Flow Cell
New and improved Air Blast Head
This sensor requires a Hach sc100 or sc1000 Digital Controller.
See pages 388-393 for details.
Prod. No.
5790000
5790001
Description
HACH LDO Dissolved Oxygen Probe
HACH LDO Probe Class I, Div 2
Price
$1,512.00
1,616.00
ACCESSORIES
5867000
Junction box for cable extensions
5796000
25 ft. extension cable
5796100
50 ft. extension cable
5796200
100 ft. extension cable
6190250
LDO Air Blast head kit
5795100
Air Blast Cleaning System, 115 V
5795200
Air Blast Cleaning System, 230 V
6860000
High Output Air Blast Cleaning System, 115 V
6860100
High Output Air Blast Cleaning System, 230 V
7300700
LDO flow cell
123.00
142.00
213.00
284.00
99.95
799.00
799.00
1,524.00
1,524.00
372.00
REPLACEMENT PARTS
5791100
Replacement Sensor Cap
142.00
MOUNTING KITS
5794400
Pole Mount Kit
5794300
Ball Float Mount Kit
379.00
521.00
For more information, call to request
Literature #2455, or visit www.hach.com
Measuring Range
0 to 20.0 ppm,
0 to 20.0 mg/L,
0 to 200% saturation
Probe Immersion Depth and
Pressure Limits
107 m (350 ft.),
1050 kPa (150 psi), maximum
Sensitivity
±0.5% of span
Transmission Distance
100 m (328 ft.) maximum
1000 m (3280 ft.) maximum when
used with a termination box
Accuracy
Measurement: ±0.2% of span
Temperature: ±0.2°C
Repeatability
±0.5% of span
Response Time at 20ºC
To 90% in less than 40 seconds
To 95% in less than 60 seconds
Resolution
Below 10 ppm:
±0.01 ppm or mg/L, ±0.1%
saturation
Above 10 ppm:
±0.1 ppm or mg/L, ±0.1% saturation
Interferences
No interferences from the following:
H2S, pH, K+1, Na+1, Mg+2, Ca+2,
NH4+1, Al+3, Pb+2, Cd+2, Zn+2,
Cr (total), Fe+2, Fe+3, Mn+2, Cu+2,
Ni+2, Co+2, CN-1, NO3-1, SO4-2, S-2,
PO4+3, Cl-1, anion active tensides,
crude oils, or Cl2-1
Operating Temperature
0 to 50°C (32 to 122°F)
Flow Rate
None required
Sensor Cable (integral)
10 m (33 ft.) terminated with quickdisconnect plug
Wetted Materials
Probe: Foamed Noryl® and 316
stainless steel
Sensor: Polybutyl methacrolate
Dimensions
60 x 292 mm (2.4 x 11.5 in.)
Weight
1.4 kg (3 lbs., 2 oz.)
Warranties
Probe: 3 Years; Sensor Cap: 1 Year
Hazardous Location Ratings
ETL listed (cETLus marked) to
Canadian and US General Safety
and Hazardous (Class I, Div. 2)
Locations
Noryl® is a registered trademark of
General Electric Co.
*Subject to change without notice.
See pages 24-47 for information on Hach laboratory and field LDO instruments.
Find it here… Buy it today on www.hach.com
U.S. customers only.
415
Process Instruments
Dissolved Oxygen: M1100 Sensor
The M1100 Luminescent
Dissolved Oxygen Sensor allows
you to monitor oxygen in beer.
• Minimal drift and annual calibration
• Highly accurate ppb oxygen measurement
• Robust optical technology without membrane or electrolyte
• Low-cost upgrade package
Prod. No.
M1100-S00
Description
ORBISPHERE M1100, Luminescent Dissolved Oxygen
Sensor (28mm)
Price
$3,250.00
M1100-S10
ORBISPHERE M1100, Luminescent Dissolved Oxygen
Sensor (12 mm)
3,250.00
Specifications*
Range
0 to 2000 ppb (dissolved O2)
Repeatability
±0.4 ppb or 1%, whichever is greater
Reproducibility
±0.8 ppb or 2%, whichever is greater
Accuracy
±0.8 ppb or 2%, whichever is greater
Limit of Detection (LOD)
0.6 ppb, minimum
Response Time (90%)
< 10 sec (gas phase)
< 30 sec (beer process)
Display Resolution
0.1 ppb
Calibration
Single point zero calibration
Calibration Sample
Standard 99.999% nitrogen (quality 50) or
equivalent oxygen free gas
Sample Temperature
-5 to 50°C (23 to 122°F)
Sample Pressure
1 to 20 bar abs (14.5 to 290 psia)
CIP
Sensor resistant to all common CIP methods
(-5 to 100°C)
Storage Temperature
-5 to 100°C (23 to 212°F)
*Subject to change without notice.
Dissolved Oxygen: G1200 Sensor
Confidence in results and analyzer
performance in a radioactive environment.
• Service requirement limited to “spot” change every 12 months—
minimal maintenance and operator intervention significantly reducing time
spent in radioactive environment
• Fully automatic, traceable calibration—results can always be trusted so time
can be focused on solving process issue not proving result
• Dry sensor with no membrane, no electrolyte and no use of chemicals—
Accuracy and reliability are not operator dependent; specialists not required to
calibrate or maintain
• Tested and approved in nuclear applications for primary circuit (2Gy cumulated)
Prod. No.
G1200-300
51G1-P400
Description
ORBISPHERE G1200, O2 Luminescent Sensor,
Nuclear Version
ORBISPHERE G1200, O2 Luminescent Sensor
+ 510 controller panel mount/cables, Nuclear Certified
Price
$2,145.00
13,719.00
Specifications*
Accuracy
(r 95) ± 1 ppb ± 2% in the 0 to 600 ppb range
Repeatability
(r 95) ±1 ppb ±2% in the 0 to 600 ppb range
Pressure
1 to 4 bar abs (14.5 to 58 psia)
Detection Range
0 to 20,000 ppb
Temperature Range
5 to 45°C (41 to 113°F) Sensor resistant to
temperature from -5 to 100°C (23 to 212°F)
Response Time (90%)
30 sec
*Subject to change without notice.
See page 157 for reagents, test kits, and accessories for measuring dissolved oxygen in the lab or field.
416
800-227-4224
Outside the United States, call 970-669-3050
Dissolved Oxygen: A1100 EC Sensor
Process Instruments
Electrochemical (EC) sensor for corrosion
control, in-line beverage, and deaerated
water applications.
• Sensor refurbishment in 3 minutes with pre-filled recharge cartridge*
• No zero drift or measurement drift (true zero sensor)—lower calibration frequency
reduces TCO
• 0.1ppb accuracy and fast response time (T90=7.2s)—most accurate/responsive
sensor available maximizes process understanding/control
• Long-life (non-consumable) sensor with 1 yr warranty + 6 monthly membrane
change—reduced cost of ownership as sensors are not disposable
*Patent-pending
Prod. No.
A1100-S00
Description
ORBISPHERE A1100, O2 Electrochemical (EC) Sensor
Price
$3,482.00
Specifications*
Accuracy
±1% of reading, or ± lower range, whichever is
greater (Assuming correct calibration)
Detection Range
1 ppb-80 ppm (Dissolved O2 measurement range)
Temperature Range (during measurement)
-5 to 95°C (23 to 203°F) –with a grille
Response Time (t90 from air)
38 sec.
Pressure (resistance)
Up to 100 bar (1450 psia)
*Subject to change without notice.
Dissolved Oxygen: 5740 sc Sensor
Replaceable membrane cartridge simplifies
the task of installing new membranes.
• Pre-installed, easy-to-replace membrane cartridge
• Rugged, foul-resistant, hydrophobic membrane
withstands harsh environments
• Unique galvanic measurement technique
Prod. No.
5740D0B
Description
5740 sc Galvanic DO Sensor with 10 m (32.8 ft.) integral cable
Price
$812.00
CONTROLLER
This sensor requires a Hach sc100 or sc1000 Digital Controller. See pages 388-393 for details.
For more information, call to
request Literature #2469,
or visit www.hach.com
Specifications*
Measuring Range
0.0 to 40.00 ppm (0 to 40 mg/L) or 200% saturation
Sensitivity
±0.5% of span
Accuracy
Measurement: ±2% of span
Temperature: ±0.2°C
Resolution
Below 10 ppm:
±0.01 ppm or mg/L, ±0.1% saturation
Above 10 ppm:
±0.1 ppm or mg/L, ±0.1% saturation
Operating Temperature
-5 to 50ºC (23 to 122ºF)
Repeatability
± 0.5% of span
Flow Rate
0.5 cm/s (0.016 ft./s), minimum
Response Time at 20ºC
To 90% in 120 seconds
Probe Immersion Depth and Pressure Limits
107 m (350 ft.),
1000 kPa (145 psi), maximum
Transmission Distance
100 m (328 ft.) maximum
1000 m (3280 ft.) maximum when used with a
termination box
Sensor Cable (integral)
10 m (33 ft.) terminated with quick-disconnect plug
Dimensions
43.7 x 203.2 mm (1.72 x 8.00 in.)
Weight
0.26 kg (0.6 lbs.)
*Subject to change without notice.
See page 157 for reagents, test kits, and accessories for measuring dissolved oxygen in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
417
Process Instruments
Dissolved Oxygen: Evita® OXY Transmitter
Intelligent, reliable, and simple DO measurement.
• Calibration just three times a year
• No need for regeneration; the sensor is simply replaced
—in less than 5 minutes
• Unique, self-cleaning transmitter design
• Cleaning is done by wiping the sensor with a cloth
—three times a year when calibrated
• Fish farming application available—call for details
Prod. No.
For more information, call to request
Literature #2494, or visit www.hach.com
Primary Applications
• Wastewater
Specifications*
OXY 4100 and 4150
TRANSMITTERS
Measuring Range
Dissolved oxygen: 0 to 10-500%,
0 to 0.1-50 mg/L or ppm,
temperature: 0 to 70°C
Measuring Uncertainty
Oxygen: ±0.5% of full scale,
temperature: ±0.5°C
Current Outputs
4-20 mA, (scalable by HART®)
galvanic isolated. Max. load 750
ohm@ 30 Vdc
Response Time
50 µm: time = 22 s
25 µm: time = 7 s
125 µm: time = 110 s
Cable
10 & 50 meters
(2 x 0.75 mm2 shielded cable)
Enclosure Rating
OXY 4100 TRANSMITTER:
IP 68 to IEC 529 (1 m)
OXY 4150 TRANSMITTER:
IP 68 to IEC 529 (10 m)
Ambient Temperature
Storage: -40 to 70°C
Operation:
Air -40 to 60°C
Medium 0 to 40°C
Power Supply
12-30 Vdc
Automatic Calibration
Compensating for temperature
(pressure, salinity, and humidity by
entering values using HART®)
Approvals
CE, C-tick
Emission: EN 50081
Immunity: EN 50081-1
Immunity: EN 61000-6-2
Description
Price
SYSTEM PACKAGES
085G4000 OXY System Package 1
$3,396.00
Includes USC 6000, OXY 4100 DO Transmitter
(Ball Float Style), OXY 1100 Membrane Cartridge,
Mounting Bracket
085G4001 OXY System Package 2
2,876.00
Includes USC 5000, OXY 4100 DO Transmitter
(Ball Float Style), OXY 1100 Membrane Cartridge,
Mounting Bracket
OXY System Package 3
Call for details
OXY 4100 DO Transmitter (Ball Float Style)
Must specify measurement range, OXY 1100
Membrane Cartridge, Mounting Bracket
REPLACEMENT TRANSMITTER
Must specify range if not used with USC, call for details.
085G4064 OXY 4100 DO Transmitter,
1,752.00
Ball-Float Style 0-20ppm, 24 Vdc, 33ft Cable,
4-20 mA Hart Transmitter, DOES NOT INCLUDE
MEMBRANE CARTRIDGE
USC (UNIVERSAL SIGNAL CONVERTOR)
For all these, other options exist, call for details. USC 7000,
up to 15 transmitter inputs, (4) 4-20 mA outputs, (5) relays.
USC 6000, (1) transmitter input, (2) 4-20 mA outputs, (3) relays.
USC 5000, (1) transmitter input, (1) 4-20 mA output
085G4146 USC 7000 Nema 6, 110-230 Vac
1,590.00
085G4130 USC 7000 w/ PROFIBUS PA Nema 6, 110-230 Vac 4,599.00
085G4131 USC 7000 w/ PROFIBUS DP Nema 6, 110-230 Vac 4,599.00
085G4140 USC 6000 Nema 6, 110-230 Vac
1,382.00
085G4145 USC 5000 Nema 6, 110-230 Vac
945.00
LP-SUN02 Sun Shield for USC 304 SS
106.00
OPTIONAL TELECOMMUNICATIONS MODULE
Call for details.
Enclosure Material
PBT/PC
Weight/Size
OXY 4100 TRANSMITTER:
2 kg/diameter: 240 mm
OXY 4150 TRANSMITTER:
1 kg/diameter: 50 mm,
length: 180 mm
OXY 1100 SENSOR
Principle
Replaceable Clark sensor measuring
dissolved oxygen. Typical life: 40,000
ppm-hours for the 50µm membrane
cartridge (temperature dependent)
Ambient Temperature
Storage: 0 to 70°C
Operation: 0 to 50°C
Material
Membrane: 50 µm (25 µm, 125 µm)
FEP Teflon®
Cathode: Gold, Anode: Silver,
Electrolyte: KCl Weight 15 g
*Subject to change without notice.
MEMBRANE CARTRIDGE
085G0022 OXY 1100 Membrane Cartridge 50 µm
Standard Single
085G0026 5 Pack
085G0027 10 Pack
085G0021 OXY 1100 Membrane Cartridge
25 µm (0-2 ppm) Single
085G0024 5 Pack
085G0025 10 Pack
085G0023 OXY 1100 Membrane Cartridge
125 µm (2-50 ppm) Single
085G0029 5 Pack
085G0030 10 Pack
192LX020 OXY 1100 Moisture Barrier Paste 5.3 Ounce Tube
MOUNTING BRACKET
085G4085 Universal Mounting to flat surface or pipe,
holds 1.5" PVC or steel pipe to attach to
transmitter, 316 SS
SPARE PARTS
081B5013 Spare Parts Kit for Mounting Bracket
(2 U-bolts, 4 washers, 4 nuts)
085G9735 OXY 4100 Collar (holds ion-permeable membrane)
215.00
982.00
1,848.00
215.00
982.00
1,848.00
311.00
1,454.00
2,699.00
29.29
197.00
68.99
37.65
See page 157 for reagents, test kits, and accessories for measuring dissolved oxygen in the lab or field.
418
800-227-4224
Outside the United States, call 970-669-3050
Distribution Monitoring: Event Monitor
• Alarm when water quality deviates from baseline
• Simplified event detection for the Water Distribution
Monitoring Panel (WDMPsc) and the Source Water
Monitoring Panel (SWMP)
• Easily upgradeable to the GuardianBlue® Early Warning
System (see pages 500-503)
• Profile and catalog events due to operational or catastrophic
excursions
• Trigger signal shows current deviation from water quality
baseline, real-time
• View trigger signal and all parameter measurements from
the main screen
• Easily communicates with your SCADA system
Process Instruments
Real-time analysis of data from Water Distribution Monitoring
Panel and astroTOC UV On-line TOC Analyzer.
• Touch screen interface for fast and easy system navigation
• Patented technology
• Hach CityGuard™ allows for remote access and control of
your systems (see pages 496-497)
The Event Monitor integrates multiple sensor outputs from the
WDMPsc and the astroTOC UV TOC Analyzer. Every 60 seconds the
system’s patented algorithm analyzes deviations in five water quality
parameters and uses the measurements to calculate a site’s water
quality baseline. The system alarms when the trigger signal exceeds
a user-set threshold, indicating a water quality deviation from the
system’s normal operating baseline parameters.
Prod. No.
6950000
Description
Event Monitor Trigger System
Includes industrial computer with
touchscreen interface mounted in a
stainless steel enclosure, 115 Vac power,
RS485 terminals for TOC analyzer and
either WDMPsc or SWMP, manual
ACCESSORIES
120161
Free-Standing Rack, wheeled
Price
$9,250.00
Specifications*
1,007.00
SERVICE WARRANTIES
69500EXW1 One year extended warranty
69500EXW2 Two year extended warranty
69500EXW3 Three year extended warranty
*
*
*
Alarms
Trigger Signal Alarm, High/Low
Parameter Alarms, Frozen Parameter
Alarm, Sensor Off-line Alarm;
Agent Alarm; Plant Alarm; Missing
Sensor; Invalid Data
Power Requirements
100-230 Vac
Operating Temperature
5 to 40ºC
Storage Temperature
-20 to 65ºC
*Call 800-227-4224, ext. 6178 for pricing.
For Water Distribution Monitoring Instrumentation, see Water
Distribution Monitoring Panel pages 420-421, Automatic Samplers
page 509, and astroTOC UV On-line TOC Analyzer page 461.
The Event Monitor Trigger System can be easily upgraded to the
GuardianBlue Early Warning System, see pages 500-503 for more
info on the GuardianBlue system.
For more information, call to request
Literature #2477, or visit www.hachhst.com
Humidity
90% at 40ºC max
Environmental
Industrial grade, meets Nema 4
and IP65 for indoor use
Weight
50 lbs. (23 kg)
Enclosure Material
316 Stainless steel
Mounting
Wall mount or rack mount
Display
15" touch screen
Certification
Listed to UL 1010 by ETS
Instrumentation
Interfaces with Hach WDM Panel or
Source Water Monitoring Panel;
astroTOC UV On-line TOC Analyzer;
Hach Sigma Portable, Refrigerated,
or All-Weather Autosamplers
Communications
RS-485 MODBUS®
Dimensions
21"(w) x 19.5"(h) x 7"(d)
(53cm x 50 cm x 18 cm )
*Subject to change without notice.
Find it here… Buy it today on www.hach.com
U.S. customers only.
419
Process Instruments
Distribution Monitoring: WDMPsc Water Panel
Know the health of your water distribution system.
• The right tool to establish your distribution system’s water
quality baseline
• Field proven instruments you can count on
• Multiparameter on-line panel includes pH, conductivity,
chlorine (free or total)
and turbidity, additional parameter optional
• Flexible system can be optimized with the Event Monitor
Trigger System, On-line TOC Analyzer, Automatic Sampler,
and ORP Probe
• Single sample inlet, outlet, and power hook-ups for
ready-to-install convenience
The Right Tool to Establish Your Distribution
System’s Baseline
The initial step in knowing the health of your water distribution
system is taking system vitals to establish a normal baseline at
critical nodes, storage reservoirs, booster stations, pump stations,
and other key monitoring points. The Water Distribution Monitoring
Panel monitors the right combination of “indicator” parameters
chosen by industry experts and recommended by the USEPA. In
combination with the Hach Event Monitor™ Trigger System, you can
now detect deviations from the baseline.
Instruments You Can Count On,
Each Ranked Top in Category
Reliability is critical for continuous, uninterrupted surveillance of your
distribution system. Each instrument in the Water Distribution
Monitoring Panel utilizes proven technology and provides readings
with little or no time lag. All data is logged to the network controller
and sent to the Event Monitor and SCADA or other remote locations
What’s on each WDMPsc and why.
Hach HST scientists chose commonly tested parameters and robust
instrumentation.
turbidity in the sample. Corrosion products and biogrowth can
elevate the turbidity level in the distribution system above that of the
plant effluent. The “E” uses USEPA approved method 180.1.
Chlorine- CL17 Chlorine Analyzer
Every 2.5 minutes the instrument obtains a sample, applies a DPD
colorimetric method based on an approved USEPA method and gives
either a free or total chlorine reading, depending on the reagent in
use at the time. You want adequate chlorine residuals to provide a
first defense against microorganism contamination, yet excess
chlorine can form DBPs in the network.
A patented differential pH measuring sensor provides information on
the acid/base nature of the water. A two-electrode conductivity
sensor measures the total ionic concentration in the water.
Temperature
Temperature is measured to ensure the probes are measuring
correctly and for other generic water quality information.
Turbidity- 1720E Turbidimeter
Continuously flowing sample enters the turbidimeter body and flows
through a bubble trap designed to vent any entrained air bubbles
from the sample stream. Turbidity is measured by directing a beam of
light from the sensor assembly into the sample in the turbidimeter
body and measuring the scatter light at 90 degrees with a photocell.
The amount of light scattered is proportional to the amount of
420
pH and Conductivity Probes from Hach/GLI
Sample Pressure
The sample pressure is measured to ensure the sample going to the
panel is within the specified range.
800-604-3493
Outside the United States, call 970-669-3050
Distribution Monitoring: WDMPsc Water Panel
Process Instruments
Specifications*
Hach Distribution
Monitoring Panel
Hach CL17 Chlorine Analyzer
Hach On-line pH Monitor
sc1000 CONTROLLER
Range
0 to 5 mg/L free or total residual chlorine
Range
0-14 pH
Dimensions
22 x 51.5"
Accuracy
±5% or 0.035 mg/L as Cl2, whichever is greater
Sensitivity
Less than 0.005 pH
SAMPLE REQUIREMENTS
Precision
±5% or 0.005 mg/L as Cl2, whichever is greater
Sample Inlet
1/2" OD Tube Connection
Stability
0.03 pH per 24 hours, non-cumulative
Minimum Detection Limit
0.035 mg/L
Sample Flow
400-600 mL/min
Hach On-line
Conductivity Monitor
Hach 1720E Turbidimeter
Sample Pressure
20 - 100 psig
Sample Operating Temperature
5 to 40ºC
Waste/Drain
3/4" NPT Hose Barb Connection
Waste/Drain Pressure
ambient, free flowing
Range
0-2000 µS/cm
Range
0.01-100 Nephelometric Turbidity Units (NTU)
Accuracy
±0.01 % of reading
Accuracy
±2% of reading or ±0.015 NTU (whichever is greater) from
0 to 40 NTU; ±5% of reading from 40 to 100 NTU
Stability
0.05 % of span per 24 hours, non-cumulative
Displayed Resolution
0.0001 NTU up to 9.9999 NTU; 0.001 NTU from
10.000 to 99.999 NTU
Sampler Connection
1/4" NPT on inlet manifold
ELECTRICAL REQUIREMENTS
Line Voltage
115 Vac / 60 Hz
Power Consumption
90 VA maximum for Cl17; 30 VA for others
Repeatability
0.1% of span or better
Temperature
20 to 200ºC
Repeatability
Better than ±1.0% of reading or ±0.002 NTU,
whichever is greater
Pressure Sensor (Gems)
Sample Flow Required
200 to 750 mL/minute (3.1 to 11.9 gal/hour)
Range
0 - 150 psi
Digital Output
RS 485 MODBUS
*Subject to change without notice.
Hach’s Water Distribution Monitoring Panels are shipped fully tested
on a panel for wall mounting with a start-up kit and manual. The
start-up kit includes reagents, and calibration standards for the CL17,
1720E, pH, and conductivity instruments. All panels include three
flow meters, one Y-strainer, one pressure sensor, and one sample
pressure regulator with gauge.
Prod. No.
Prod. No.
6846000
ACCESSORIES
6846100
6846200
6846300
Description
Price,
WDMP sc
$13,750.00
Includes: sc1000 with Modbus Probe
Module; sc1000 Display module;
pH calibration buffers; Conductivity
Reference Solution 1000 µs/cm, 100 mL,
quantity 3; StablCAL 20 NTU, 1L for
turbidity; Free Chlorine reagent set;
Total Chlorine reagent set; Sulfuric Acid,
19.2 N, 100mL MDB (for CL17 cleaning);
CL17 start-up kit and 10 feet each of inlet
and drain tubing, hose clamp, beaker,
and manuals.
WDMP sc
Same as Prod. No. 6846000 above,
within a NEMA 4X enclosure.
WDMP sc Start-up Kit
Included with Prod. No. 6846000.
Can be ordered separately if you require
additional kits. Includes: CL17 reagents,
calibrations, standards, and accessories
for first month of operation.
17,950.00
Description
Price
MONTHLY REAGENTS REQUIRED FOR WDMPsc
2556900
Free Chlorine Reagent Set
$44.39
2557000
Total Chlorine Reagent Set
46.59
6846400
ORP Accessory Kit
Includes: ORP sensor; 200mV buffer, 500 mL
705.00
6846600
Pressure Regulator
Brings sample pressure down to 125 psi
where the panel can safely use it,
upward limit of 300 psi.
175.00
6846700
sc1000 Controller Attachment Plate
Allows sc1000 controller to be attached
to the WDMP sc wheeled rack.
140.00
6844600
WDMP sc Mounting Rack, Wheeled
945.00
6840000
Probe Plug (extra)
*
SERVICE WARRANTIES
397.00
WDMP sc Maintenance Kit
285.00
Includes: all the necessary items to perform
routine maintenance for the first six months.
NOTE: You must order your chlorine reagent
separately and specify Total Chlorine Reagent
Set Prod. No. 2557000 or Free Chlorine Reagent
Set Prod. No. 2556900. Chlorine reagents
sets last one month.
FSPWDMPscbasic Field Service Partnership
for Water Distribution
Monitoring Panel
*
FSPWDMPscfull
Field Service Partnership
for full water quality system
including the Event Monitor
*
59800EXW1
One-Year Extended Warranty
*
59800EXW2
Two-Year Extended Warranty
*
59800EXW3
Three-Year Extended Warranty
*
*Call 800-227-4224, ext. 6178 for pricing.
For Hach Event Monitor Trigger System see page 419,
for astroTOC UV On-line TOC Analyzer see page 461.
For more information, call to request
Literature #2566, or visit www.hachhst.com
Find it here… Buy it today on www.hach.com
U.S. customers only.
421
Process Instruments
Distribution Monitoring: astroTOC™ Analyzer
Increases system sensitivity to organic chemicals
in the distribution system.
• Greatly enhances the detection and classification
capabilities of GuardianBlue Early Warning System
• When combined with the Water Panel, the TOC Analyzer
exponentially increases the system’s sensitivity to organic
chemicals, creating one of the industry’s most unique and
innovative early warning systems. Total organic carbon is a
crucial part of the fingerprint structure.
• Combines chemical and ultraviolet oxidation techniques
in a low-temperature reactor to deliver direct TOC
measurements
• Uses a multi-staged UV oxidation reactor and a chemically
impervious non-dispersive infrared (NDIR) CO2 detector
system, assuring full compliance with Standard Methods
5310 C and EPA method 415.1
One of Hach’s most sophisticated
water quality sensors
In the first analysis step, the sample is mixed with acid, converting
the total inorganic carbon (TIC) into CO2. The TIC sparger removes all
the CO2 from the sample solution. Subsequently, the TIC-free sample
is mixed with sodium persulfate and routed through the UV reactor,
oxidizing the TOC into CO2.
The gas/liquid mixture is transported by the carrier gas into the
gas-liquid separator (GLS), where the sample gas is separated and
diverted into the NDIR detector for the direct, interference-free CO2
measurement. The resulting CO2 concentration measurement is
directly proportional to the original TOC concentration found in the
sample. The front panel displays the TOC concentration in mg/L.
TOC Analyzers with communication hardware to interface with Event
Monitor Trigger System, WDM Panel, and PipeSonde In-pipe Probe.
Includes one-month’s supply of reagents, 2 UV lamps, view window,
drain, and reagent level indicators.
Prod. No.
H-6195-1030DS
8814000K
Specifications*
8814100K
Range
0-25 mg/L
8814300
Accuracy
±2% of full scale at 25ºC
Repeatability
±2% of reading at 25ºC
ACCESSORIES
4300-0008
Minimum Detection Limit
≤0.015 mg/L for 0-5 mg/L
4300-0009
Response Time
T90 ≤8 min.
120161
Serial Communication
Multi-function RS-232 or RS-485 serial port (MODBUS®, CSV)
SERVICE WARRANTIES
FSP1950+
Field Service Partnership for 1 year
Mounting
Wall mount or optional rack mount
*Call 800-227-4224, ext. 6178 for pricing.
*Subject to change without notice.
422
Description
WDM astroTOC UV CRS enclosure,
0-25 mg/L (Recommended for use
with the Event Monitor.)
TOC IR Bench Calibration Gas,
Nitrogen Zero
TOC IR Bench Calibration Gas,
Carbon Dioxide, 1000 PPM
Adjustable Regulator and Flowmeter
Combo for 103L gases
Purge Gas Generator
with compressor, 110V
Purge Gas Generator
with compressor, 230V
Free Standing Rack, wheeled
Price
$16,000.00
115.00
145.00
250.00
7,350.00
7,350.00
1,007.00
For more information, call to request
Literature #2573, or visit www.hachhst.com
800-604-3493
Outside the United States, call 970-669-3050
*
Flow: Tee-Mount Sensors
Process Instruments
Accurate flow measurement at an economical price.
• For pipe sizes from 1/2 to 4 inches
• Non-magnetic sensing technique
• Variable frequency square wave signal
• Wide rangeability
• Superior low-velocity performance
• FDA-approved wetted materials
• For use with F53, F33 or PRO-F3 controllers/ transmitters
Principal of Operation
Hach's Impeller Flow sensors unique non-magnetic sensing technique is impervious to fouling
from metal particles. The non-magnetic sensing technique reduces drag for superior low-flow
and low-volume measurement accuracy.
The forward-swept, six-bladed impeller has significantly better low-flow, low-velocity
characteristics than conventional four-bladed impellers to provide higher measuring accuracy.
Economical and Practical
An F53 or F33 controller can independently monitor four sensors, making this system both
economical and convenient to use.
Prod. No.
F1A11A1T
F1A11B2T
Description
Polypropylene Sensor
PVDF Sensor
Price
$399.00
499.00
For more information, call to request
Literature #G501 or visit www.hach.com
Mounting Tee Size
Mounting Tees
1/2 inch
3/4 inch
1 inch
1 inch
(high flow)
PVC tees
1 1/4 inch
1 1/2 inch
2 inch
3 inch
4 inch
MHF15H2
MHF15J2
MHF15K2
MHF15B2
MHF15L2
—
MHF15D2
MHF15E2
MHF15G2
Cast bronze tees
—
—
MHF13B1
—
MHF13C1
MHF13D1
—
—
—
PVDF tees
—
—
—
—
—
MHF16D2
MHF16E2
MHF16G2
MHF16H2
Specifications*
MODEL F1A11A1T POLYPROPYLENE SENSOR
MODEL F1A11B2T PVDF SENSOR
Maximum Temperature
In PVC Tee
In Cast Bronze Tee
In PVDF Tee
140°F at 40 psi (60°C at 2.75 bar)
176°F at 400 psi (80°C at 27.5 bar)
140°F at 40 psi (60°C at 2.75 bar)
Maximum Temperature
In PVC Tee
In Cast Bronze Tee
In PVDF Tee
140°F at 40 psi (60°C at 2.75 bar)
176°F at 400 psi (80°C at 27.5 bar)
176°F at 40 psi (60°C at 2.75 bar)
Maximum Pressure
In PVC Tee
In Cast Bronze Tee
In PVDF Tee
Repeatability (in any tee)
Linearity (in any tee
100 psi at 77°F (6.9 bar at 25°C)
400 psi at 176°F (27.5 bar at 80°C)
100 psi at 77°F (6.9 bar at 25°C)
±0.5% of full scale
±1% of full scale
Maximum Pressure
In PVC Tee
In Cast Bronze Tee
In PVDF Tee
Repeatability (in any tee)
Linearity (in any tee
100 psi at 77°F (6.9 bar at 25°C)
400 psi at 176°F (27.5 bar at 80°C)
230 psi at 77°F (6.9 bar at 25°C)
±0.5% of full scale
±1% of full scale
Accuracy**
In PVC Tee
In Cast Bronze Tee
In PVDF Tee
Accuracy**
In PVC Tee
±1% of full scale from 1 to 30 ft./sec.
(0.3 to 9.0 m/sec.)
±1% of full scale from 1 to 20 ft./sec.
(0.3 to 6.0 m/sec.)
±1% of full scale from 1 to 30 ft./sec.
(0.3 to 9.0 m/sec.)
In Cast Bronze Tee
In PVDF Tee
±1% of full scale from 1 to 30 ft./sec.
(0.3 to 9.0 m/sec.)
±1% of full scale from 1 to 20 ft./sec.
(0.3 to 6.0 m/sec.)
±1% of full scale from 1 to 30 ft./sec.
(0.3 to 9.0 m/sec.)
Sensor Cable
2 conductor (plus shield), 20 ft. (6 m)
Sensor Cable
2 conductor (plus shield), 20 ft. (6 m)
Wetted Materials
Polypropylene body, Polypropylene impeller, TTZ
(transformation toughened zirconia), ceramic shaft and EPR O-rings
Wetted Materials
PVDF body, PVDF impeller, TTZ (transformation toughened zirconia), ceramic
shaft and Viton O-rings
*Subject to change without notice.
**Accuracy is attained with at least 10 diameters of straight pipe upstream of sensor and
at least 5 diameters of straight pipe downstream from sensor, and with full pipe flow.
Find it here… Buy it today on www.hach.com
U.S. customers only.
423
Process Instruments
Fluoride: CA610™ Analyzer
Accurate, cost-effective fluoride analysis.
• Ion-selective electrode with patented,† replaceable tip
• Automatic calibration
• Temperature-controlled flow cell
• Rugged, lightweight enclosure
• Ultra-low reagent consumption
• Virtually immune to interferences
†U.S.
Patent #5,393,402
Full-Time Assurance. Long-Term Affordability.
Using proven ion-selective electrode technology, Hach’s CA610
Fluoride Analyzer provides continuous assurance that the fluoride
concentration in your drinking water is correct. The CA610 also offers
unmatched cost-efficiency and versatility. Reagent consumption is
ultra-low. With a compact, self-contained design, the analyzer is ideal
for monitoring in your plant or in remote locations.
Dependable Technology. Practical Design.
The CA610 delivers accurate fluoride readings regardless of
changes in a sample’s ionic strength, pH, or temperature. It is
virtually immune to interference. Hach’s unique electrode—with a
patented, replaceable tip—makes maintenance simple, infrequent,
and inexpensive. There’s no need to replace the entire electrode—
only the fluoride crystal tip.
A patented electrode—with a molded
replaceable lanthanum fluoride crystal tip—
delivers an exceptionally long working life.
Hach CA610 Fluoride Analyzers are shipped with a one-month supply
of reagents, installation kit, annual maintenance kit, and manual. The
power cord is ordered separately.
Prod. No.
5740001
5740002
Primary Applications
• Drinking Water
Price
$5,448.00
5,711.00
ACCESSORIES
5448800
Power Cord with Strain Relief,
115 Vac, 10 A, 1.83 m (6')
5448900
Power Cord with Strain Relief,
230 Vac, 10 A, 1.83 m (6')
• Industrial Water
Specifications*
Range
0.1 - 10 mg/L Fluoride
REPLACEMENT ITEMS
2816900
Reagent Set, CA610 Fluoride Analyzer
Includes Reagent 1 TISAB, Standard 1,
and Standard 2
5742100
Maintenance Kit**
5744400
Pump Tubing Set
5742700
Instrument Tubing Set
5744800
Electrode Kit***
Accuracy
±10% or ±0.10 mg/L, whichever is greater
Detection Limit
0.10 ppm
Cycle Time
4.2 minutes
Recorder Outputs
One 4-20 mA, with output span programmable over any portion
of the 0.1 to 10 mg/L range. 130 V isolation from earth ground
Alarms
Two alarms, selectable for sample concentration, system warning or system
shutdown. Each alarm equipped with an SPDT relay with contacts rated
for 5 A resistive load at 230 Vac.
Optional External Outputs
AquaTrend® Network interface
Description
CA610 Fluoride Analyzer
CA610 Fluoride Analyzer
with AquaTrend® Network Capability
34.15
36.35
63.09
120.00
31.50
57.10
735.00
**Kit includes instrument and pump tubing sets (1-year supply), replacement
filter screen, and spare fittings.
***Kit includes ISE and pH electrodes, electrode tips, syringe, and inner
fill solution.
For more information, call to request
Literature #4540 or visit www.hach.com
*Subject to change without notice.
See pages 158-159 for reagents, test kits, and accessories for measuring fluoride in the lab or field.
424
800-227-4224
Outside the United States, call 970-669-3050
Hardness: APA 6000™ Analyzer
Process Instruments
Accurate, continuous hardness measurement.
• Measures low-range hardness using EPA-approved
calmagite chemistry
• Accurately and continuously measures up to
two sample streams (requires sample sequencing kit)
• Operates unattended for one month
• Self-calibrating, self diagnostics
• Makes your water softening system more efficient
and less costly
Industrial and Ultrapure Water Applications
The APA 6000 Hardness Analyzer can be used to monitor both
influent and effluent in industrial applications and water treatment
processes including demineralizer effluent, boiler feed-water, boiler
water, and process water. It is also appropriate for monitoring
hardness in ultrapure processes used by pharmaceutical, electronic
chip, and cosmetics manufacturers.
Primary Applications
• Drinking Water
• Pure Water/Power
• Industrial Water
Specifications*
LOW-RANGE ANALYZER
APPLY TO BOTH ANALYZERS
Range
Sample Temperature Range
5 to 50ºC (41 to 122ºF)
0.05 to 10 mg/L for hardness as
calcium carbonate
Accuracy
±5 % of reading or ± 0.05 mg/L
whichever is greater
Each analyzer includes an installation kit, one month’s supply of
reagents, a maintenance kit, a sample conditioning kit, an illustrated
manual, and a quick reference card. Power cords must be ordered
separately.
Prod. No.
5100210
6200010
Description
APA 6000
Low-Range Hardness Analyzer
50 to 10,000 µg/L, with AquaTrend®
APA 6000
High-Range Hardness Analyzer
10 to 1,000 mg/L, with AquaTrend®
Price
$9,590.00
10,678.00
ACCESSORIES
Low-Range Hardness NOTE: both sets are required for operation.
6001900
Low-Range Hardness Reagent Set
91.55
6002000
Low-Range Hardness Standards Set
46.89
High-Range Hardness NOTE: both sets are required for operation.
6002100
High-Range Hardness Reagent Set
104.00
6002200
High-Range Hardness Standards Set
46.89
5133900
APA 6000 Micro Filter System, 115 V
2,698.00
5133901
APA 6000 Micro Filter System, 230 V
3,022.00
4630600
Power cord kit, 120 Vac
39.60
4630800
Power cord kit, 240 Vac
45.00
6200900
Sample Sequencing Kit
343.00
Repeatability
±3% of reading or ± 0.03mg/L
whichever is greater
Response Time
Less than 5 minutes for 90%
response
Cycle Time
4 minutes
HIGH-RANGE ANALYZER
Range
10 to 1000 mg/l as CaCO3
Accuracy
±5 % of reading or ± 2 mg/L CaCO3
whichever is greater
Sample Flow
100 to 2000 mL/min. max.
Inlet Pressure
0.5 to 30.0 psig (0.03 to 2.04 bar)
Drain Fitting
3/4" NPT barbed hose fitting
Outputs
Two 4-20 mA outputs suitable for
recorders or PID control. Output
span programmable over any portion
of the measuring range (130 Vac
isolation from earth ground).
Alarms
Two SPDT relays with contacts rated
for 5A resistive load at 230 Vac.
Additional relays available with
optional Signal Output Modules.
Network Connectivity
AquaTrend™ network, using the
Lonworks® protocol
Repeatability
±5% of reading or ±2 mg/L as
CaCO3, whichever is greater
Certification
NRTL certified to UL and CSA
standards and CE approved
Response Time
Less than 17 minutes for 90%
response
Power Requirements
95-240 Vac, 50/60 Hz ± 2 Hz
Cycle Time
8.2 minutes
Enclosure
NEMA-4X(indoor)/IEC 529 (IP66) with
provision for air purge. Reagent
enclosure is drip-proof.
Dimensions
21 x 25 x 21" (522 x 627 x 526 mm)
Weight
56 lbs (25.5 kg)
For more information, call to request
Literature #1584 or visit www.hach.com
*Subject to change without notice.
See pages 161-163 for reagents, test kits, and accessories for measuring hardness in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
425
Process Instruments
Hardness: SP 510™ Monitor
Maximize your softener cycle time and minimize
your regeneration cost.
• Low Maintenance—operates unattended for two months
• Low reagent consumption
• Rugged, lightweight, and self-contained
• Operates unattended for two months
• Immediately signals hardness breakthrough to
activate regeneration
• Makes your water softening system more efficient
and less costly
• Reliable, simple, and accurate—with automated calibration
• Continuous monitoring in “real time”
Continuous Hardness Detection
The SP 510 detects hardness breakthrough when the capacity of the
water softener is exhausted, immediately signaling the need for
regeneration. Alarm points 0.3, 1, 2, 5, 10, 20, 50 and 100 ppm
(expressed as mg/L of CaCO3) are selected by choosing the
corresponding Hach model. Easy-to-read LED indicators show a
simple “HARD” or “SOFT” sample status. You can also use the SP
510’s built-in alarm relay to actuate an external annunciator.
Low Maintenance Requirements
The SP 510 samples water every two minutes, operating
automatically for up to 60 days. Spend only about 15 minutes of your
time every two months to replenish and standardize the reagents.
Replace tubing in the pump system every six months.
Convenient, Trouble-Free Operation
The SP 510 eliminates the guesswork, so your softener is
regenerated only when needed. Regeneration based on calculation
or set times can be replaced with continuous monitoring and
automatic control.
Primary Applications
• Drinking Water
• Pure Water/Power
• Industrial Water
Specifications*
Repeatability
± 10% of set point value in 0.3-2 mg/L ranges
± 4% of set point value in 5-100 mg/L ranges
Cycle Time
2.0 minutes, average
Output
One SPDT relay
Power
115/230 Vac, 50/60 Hz, selectable
Certification
NRTL certified to UL and CSA standards, and CE approved
Reagent Requirements
1 each of indicator and buffer. Replenish every two months for
continuous operation.
Alarm Delay
2 consecutive cycles above set point activate alarm.
1 cycle below set point cancels alarm.
Includes a two-month supply of reagents, an installation kit, a
maintenance kit, and an instruction manual. (Recorder output
capabilities not available.) Note: Select the model with the alarm trip
point 40 to 50 percent higher than your normal effluent hardness.
Prod. No.
5410003
5410001
5410002
5410005
5410010
5410020
5410050
5410099
5448800
5448900
Alarm Trip Point
0.3 mg/L
1 mg/L
2 mg/L
5 mg/L
10 mg/L
20 mg/L
50 mg/L
100 mg/L
Power cord Kit with strain relief, 120 Vac
Power cord Kit with strain relief, 240 Vac
Price
$1,813.00
1,813.00
1,813.00
1,813.00
1,813.00
1,813.00
1,813.00
1,813.00
34.15
36.35
A customized SP 510 is available that activates the alarm after one
cycle instead of two. Contact your Hach representative for details.
For more information, call to request
Literature #1457 or visit www.hach.com
*Subject to change without notice.
See pages 161-163 for reagents, test kits, and accessories for measuring hardness in the lab or field.
426
800-227-4224
Outside the United States, call 970-669-3050
Molybdate: MO42 Analyzer
• Uses fast, reliable, and economical catechol chemistry
for up to 30 days of unattended operation
• No extensive maintenance and minimal use of reagents
means low operating costs
• Manual or automatic feed pump control
• The instrument performs self-diagnostics on every cycle
and will issue warnings or alarms when an issue is detected
—making troubleshooting simple
• Cost-effective alternative to grab samples and lab testing
Ideal for Industrial Water Treatment Applications
The MO42 Molybdate Analyzer is ideal for industrial water treatment
applications where it is necessary to continuously monitor and
control the concentration of molybdate within specified levels—
cooling towers and closed recirculating systems, such as hot water
heating (boilers) and chilled water systems (coolers).
Process Instruments
Reliable and economical molybdate analysis.
Molybdate compounds are used extensively in these areas as
corrosion inhibitors or as tracer agents for the determination of
concentrations of other treatment chemicals. It is important that the
concentration of molybdate be maintained at the specified levels in
order to balance the benefits of corrosion protection without the
additional expenses of chemical overfeed.
The Hach MO42 Molybdate Analyzer is an effective alternative to
laboratory methods for molybdate monitoring and allows for real time
feed control.
Hach MO42 Molybdate Analyzers include a one-month supply of
reagents, maintenance kit, installation kit, and user manual.
(Power cord sold separately.)
Prod. No.
6180004
Description
Hach MO42 Molybdate Analyzer
• Pure Water/Power
• Industrial Water
Price
$5,339.00
ACCESSORIES
5448800
Power Cord; 125V, 10A, 1.83 m (6 ft.)
34.15
5448900
Power Cord; 230V, 10A, 1.83 m (6 ft.),
36.35
continental European plug
6181100
Maintenance Kit; 1 year, includes tubing,
119.00
caps, funnel, and fittings
6181101
Maintenance Kit with Preassembled Tubing;
144.00
1 year, includes tubing, caps, funnel, and fittings
4643600
Flow Meter; with 1/4-inch OD tubing
139.00
REAGENTS
The Hach MO42 Molybdate Analyzer requires two (2) 500 mL bottles
of reagent for a 30-day operating period.
2890549
MO42 Reagent, 500 mL
48.95
Service Plans available, call 800-227-4224, ext. 6178 for pricing.
For more information, call to request
Literature #2589 or visit www.hach.com
Primary Applications
Specifications*
Range
0 to 5 mg/L as Mo6+
Accuracy
±5% or ±0.05 mg/L as Mo6+, whichever is greater
Precision
±3% or 0.03 mg/L as Mo6+, whichever is greater
Minimum Detection Limit
0.03 mg/L
Cycle Time
2.5 minutes
Operating Temperature
5 to 40°C (41 to 104°F)
Recorder Outputs
One 4-20 mA with an output span programmable over any portion of the 0 to
5 mg/L range, 130 V isolation from earth ground, 500 ohm maximum
Alarm Relay Outputs
Two alarms selectable for sample concentration alarm, analyzer system
warning, or analyzer system shut-down alarm.
Shipping Weight
11.3 kg (25 lbs.)
*Subject to change without notice.
See page 171 for reagents, test kits, and accessories for measuring molybdate in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
427
Monochloramine/Ammonia: APA 6000™ Analyzer
Process Instruments
Total chloramination control.
• Automatic operation—the system auto-calibrates, cleans,
primes, and displays alarms on the front panel.
• Two-stream operation—optionally monitor up to two
separate sample streams. The system completes one
measurement per stream every 5 minutes.
• Flexible alarms and outputs—user-configured outputs
include two 4-20mA analog outputs and two alarm relays.
The analyzer can be expanded to 14 analog and/or relay
outputs by adding additional Signal Output Modules.
Complete Confidence
The APA 6000 Ammonia/ Monochloramine Analyzer was developed
to meet the strict monitoring requirements of the Metropolitan Water
District of Southern California—one of the largest water districts in
the USA. It also meets your requirements for easy installation and
continuous operation.
Each analyzer includes an installation kit, one month’s supply of reagents,
a maintenance kit, a sample conditioning kit, an illustrated manual, and a
quick reference card. Power cords must be ordered separately.
Primary Applications
• Drinking Water
Prod. No.
5500610
Specifications*
Speciation
Free and Total Ammonia (NH3) and
Monochloramine (NH2Cl) as N.
Also menu selectable to display
as Cl2
Outputs
Two 4-20 mA outputs included;
total of up to 14 programmable
4-20 mA outputs (with optional
Signal Output Modules)
Range
0.02 to 2 mg/L as N
(0.1 to 10.0 mg/L as Cl2)
Network Connectivity
AquaTrend network using the
LonWorks protocol
Accuracy
± 5% of reading or ± 0.02 mg/L as N
(0.1 mg/L as Cl ), whichever is
greater
Compliance/Certification
UL, CSA and IEC safety standards,
FCC & European RFI standards and
European EMI standards
Repeatability
± 3% of reading or ± 0.01 mg/L as N
(0.05 mg/L as Cl ), whichever is
greater
Dimensions
Approximately 522 x 627 x 527 mm
(21" tall, 25" wide, 21" deep)
2
2
Cycle Time
5.0 minutes per analysis for
1 sample stream (Optionally capable
of sampling 2 streams—requires
sample sequencing kit)
Alarms
Two SPDT alarm relays included;
total of up to 14 programmable
alarm relays (with optional Signal
Output Modules)
Description
APA 6000 Ammonia and
Monochloramine Analyzer with AquaTrend®
Price
$12,733.00
ACCESSORIES
6001400
Ammonia/Monochloramine Reagent Set
6001500
Ammonia/Monochloramine Standards Set
NOTE: both sets are required for operation.
5133900
APA 6000 Micro Filter System, 115 V
5133901
APA 6000 Micro Filter System, 230 V
4630600
Power cord kit, 120 Vac
4630800
Power cord kit, 240 Vac
6200900
Sample Sequencing Kit
105.00
34.95
2,698.00
3,022.00
39.60
45.00
343.00
Enclosure
NEMA 4X (indoor)/IEC529 (IP66) with
provision for air purge Reagent
enclosure is drip-proof
Weight
25.5 kg (56 lbs.)
Power
95-240 Vac, 50/60 Hz
*Subject to change without notice.
For more information, call to request
Literature #1607, or visit www.hach.com
See page 134 for reagents, test kits, and accessories for measuring ammonia/monchloramine in the lab or field.
428
800-227-4224
Outside the United States, call 970-669-3050
Nitrate: NITRATAX™ sc UV Sensors
Process Instruments
UV analysis—the reagent-free alternative.
• UV absorption method—proven, continuous, and precise
• Eliminates reagents, sampling, and sample conditioning
• Self-cleaning sensor using built-in wiper
• Life-long factory calibration
• Full-featured plug-and-play with sc100 or sc1000 Digital Controllers
UV Analysis—Eliminates Reagents, Sampling, and
Sample Conditioning
Using advanced ultraviolet (UV) absorption technology, Hach NITRATAX UV Nitrate
Sensors offer unprecedented simplicity, accuracy, and economy in nitrate analysis.
UV Absorption Method—Proven, Continuous, Precise
NITRATAX sensors rely on the principle that molecular bonds can absorb ultraviolet
(UV) light—in this case, nitrate (NO3) and nitrite (NO2) absorb UV light. As the
concentration of nitrate or nitrite increases, UV absorption also increases. A built-in
photometer measures the absorbance, while a second beam of UV light provides
a reference standard and corrects for interference caused by turbidity and
organic matter.
Primary Applications
Specifications*
• Drinking Water
• Wastewater
• Industrial Water
• Environmental
NITRATAX plus sc
NITRATAX eco sc
NITRATAX clear sc
Measuring Principal
Reagent-free UV absorption with patented 2-beam technique (applies to all)
Measuring Gap/Path Length 1, 2, and 5 mm
1 mm
5 mm
Measuring Range NO2+3-N 0.1 to 100.0 mg/L (1 mm) 1.0 to 20.0 mg/L (1 mm)
0.5 to 20.0 mg/L (5 mm)
0.1 to 50.0 mg/L (2 mm)
0.1 to 25.0 mg/L (5 mm)
Detection Limits NO3-N
0.1 to 100 mg/L
1.0 to 20 mg/L
0.5 to 20 mg/L
Accuracy
± 3% of the mean ± 0.5
± 5% of the mean ± 1.0
± 5% of the mean ± 0.5
Resolution
0.1 mg/L
0.5 mg/L
0.1 mg/L
*Subject to change without notice.
Prod. No.
Description
Price
NITRATAX sc UV NITRATE ANALYZER SYSTEMS
6139200
NITRATAX plus sc System
includes sensor with 2 mm path length,
mounting hardware, and sc100 controller
6139500
NITRATAX plus sc System includes sensor
with 2 mm path length, flow-through cell,
and sc100 controller.
6139400
NITRATAX eco sc System includes
sensor with 1 mm path length, mounting
hardware, and sc100 controller
6139300
NITRATAX clear sc System includes
sensor with 5 mm path length, mounting
hardware, and sc100 controller
6139600
NITRATAX clear sc System includes sensor
with 5 mm path length, flow through cell,
and sc100 controller
INDIVIDUAL NITRATAX sc UV NITRATE SENSORS
All sensors are equipped with 10 m (32.8 ft.) cable.
LXV417.99.10002 NITRATAX plus sc Sensor
1 mm path length
LXV417.99.20002 NITRATAX plus sc Sensor
2 mm path length
LXV417.99.50002 NITRATAX plus sc Sensor
5 mm path length
LXV420.99.50002 NITRATAX clear sc Sensor
5 mm path length
LXV415.99.10002 NITRATAX eco sc Sensor
1 mm path length
$15,372.00
16,876.00
11,624.00
12,870.00
14,549.00
13,747.00
13,747.00
13,747.00
CONTROLLER
This sensor requires a Hach sc100 or sc1000 Digital Controller.
See pages 388-393 for details.
Prod. No. Description
Price
ACCESSORIES
LZX414.00.10000 Mounting hardware for sensor
LZX869
Flow-through cell for NITRATAX
plus sc-sensors, 2 mm path length
LZX867
Flow-through cell for NITRATAX
plus sc-sensors, 5 mm path length
LZX866
Flow-through cell for NITRATAX
clear sc-sensors, 5 mm path length
LCW825
Calibration standard 50 mg/L
(11.3 mg/L NO3-N)
LZX148
Spare wiper blades for 1 mm Nitratax, pk/5
LZX012
Spare wiper blades for 2 mm Nitratax, pk/5
LZX117
Spare wiper blades for 5 mm Nitratax, pk/5
CABLE ACCESSORIES
5867000
Junction box for extension cables
5796000
Extension cable, 7.6 m (25 ft.)
5796100
Extension cable, 15.2 m (50 ft.)
5796200
Extension cable, 30.5 m (100 ft.)
$378.00
1,934.00
1,934.00
1,934.00
34.75
402.00
188.00
117.00
123.00
142.00
213.00
284.00
11,381.00
10,000.00
For more information, call to request
Literature #2464 or visit www.hach.com
See page 173 for reagents, test kits, and accessories for measuring nitrate in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
429
Process Instruments
Nitrate: NO3D Sensor
The easiest to maintain nitrate ISE sensor for continuous trending.
• Improved Accuracy—with the NEW CARTRICAL™ technology
• Simplified Maintenance—cartridge replacement in minutes
• No Sample Preparation—sensor immersion directly
in aeration basin
• Economical to Operate—with continuous measurement
and trending
Principal of Operation:
Hach’s innovative technology,
CARTRICAL makes the NO3D sc
the easiest to maintain nitrate ISE
sensor for continuous trending.
The Hach NO3D sc Nitrate Sensor uses an ion-selective electrode
(ISE) to detect nitrate ions (NO3-) directly in the aeration basin as
nitrate nitrogen (NO3-N). A differential electrode is used as the
reference electrode for superior stability. The most probable
interference is from chloride ions (Cl-) which is compensated
through the use of an integrated chloride ISE to correct the nitrate
value. Other interferences are further reduced using CARTRICAL
technology, which calibrates each electrode individually and
calibrates all three electrodes to each other. A temperature sensor
is also included to improve accuracy.
Primary Applications
• Municipal Wastewater
Specifications*
Measurement Method
Ion-selective electrodes for nitrate
and chloride with differential
reference electrode and temperature
sensor
Range
0 to 1000 mg/L NO3-N
Accuracy
5% of measured value ±0.2 mg/L
(with standard solution)
Detection Limit
0.2 mg/L
Response Time
Less than 3 minutes (T90)
Operating Temperature
-20 to 45°C (-4 to 113°F)
Sample Temperature
2 to 40°C (35 to 104°F)
Sample pH
5 to 9
Sensor Immersion Depth
0.3 to 3.0 m (1 to 10 ft.) maximum
Sample Pressure
0.3 bar (4.4 psi) maximum
Storage Temperature
Sensor: -20 to 60°C (-4 to 140°F)
Sensor Cartridge: 5 to 40°C
(41 to 104°F)
Calibration
Sensor cartridge, calibrated
(sensor code contains factory
calibration in code form)
Entry of sensor code calibrates
the sensor
No standard solutions needed
1- or 2-point inline matrix correction
adapts sensor to the wastewater
matrix
Certifications
CE approved
Sensor Construction
316 stainless steel with Ryton® ends
Dimensions
48 x 361 mm (1.9 x 14.2 in.)
Cable Length
Standard: 10 m (33 ft.)
Optional Extensions: 7.6 m (25 ft.),
15.2 m (50 ft), or 30.5 m (100 ft.)
Maximum Total Length: 100 m
(328 ft.)
Prod. No.
Description
Price
LXV442.99.00002 NO3D sc Nitrate Sensor,
$5,851.00
includes 10 m (32.8 ft.) integral
cable and a calibrated sensor cartridge
CONTROLLER
This sensor requires a Hach sc100 or sc1000 Digital Controller.
See pages 388-393 for details.
MOUNTING KITS
6184900
Rail Mount Kit
LZX914.99.12400 Chain Mount Kit
ACCESSORIES
LZY331
6860000
6860100
Cleaning Unit
High Output Air Blast Cleaning
Compressor, 115 Vac
High Output Air Blast Cleaning
Compressor, 230 Vac
REPLACEMENT PARTS
6188401
Calibrated Sensor Cartridge
6188300
Test-Cartridge
356.00
336.00
731.00
1,524.00
1,524.00
664.00
139.00
Weight
1.3 kg (2.9 lbs.)
Warranty
1 year
Ryton® is a trademark of
Phillips Petroleum Company.
*Subject to change without notice.
For more information, call to request
Literature #2649, or visit www.hach.com
See page 173 for reagents, test kits, and accessories for measuring Nitrate in the lab or field.
430
800-227-4224
Outside the United States, call 970-669-3050
Oil-In-Water: FP 360 sc Sensor
Process Instruments
Continuous oil–in-water
monitoring for the right price.
• Probe design makes it the low price solution among
competitive UV fluorescence instruments
• No tubes, pumps, or valves that can foul or require frequent
maintenance
• Accurate online measurement reduces time-consuming
laboratory testing
• Can detect polycyclic aromatic hydrocarbons (PAHs)
from 1.2 ppb to up to 5000 ppb (µg/L) or mineral oils
from 0.1 to 150 ppm (mg/L)
• UV fluorescence method makes it impervious to interferences
by turbid water or natural organic matter (NOM)
• Available in stainless steel or titanium housing for
measurement in the harshest of conditions
• Plug-and-play with Hach Digital Controllers
Principle of Operation
The FP 360 sc is the only online oil-in-water instrument
that delivers the highest sensitivity and selectivity with
the lowest total cost of ownership.
The FP 360 sc measures intensity of fluorescence light at a
wavelength of 360 nm emitted by polycyclic aromatic hydrocarbons
(PAH) after UV irradiation of the sample at 254 nm. Since PAHs are
components of most mineral oils, the FP 360 sc can detect the
presence of oil contamination in surface, process, or industrial
waters. In addition, since the intensity of the emitted light is
proportional to the PAHs concentration, the FP 360 sc can be
calibrated to measure oil concentration in stable matrices.
For more information, call to request
Literature #2668 or visit www.hach.com
Primary Applications
Prod. No.
STAINLESS STEEL
LXV441.99.11102
LXV441.99.11302
LXV441.99.21102
LXV441.99.21302
Description
WITHOUT CLEANING UNIT
0–500 µg/L, 32.8 ft (10 m) cable
0–500 µg/L, 5 ft (1.5 m) cable
0–5,000 µg/L, 32.8 ft (10 m) cable
0–5,000 µg/L, 5 ft (1.5 m) cable
Price
$14,990.00
14,990.00
14,990.00
14,990.00
STAINLESS STEEL WITH CLEANING UNIT
LXV441.99.11202 0–500 µg/L, 32.8 ft (10 m) cable
LXV441.99.21202 0–5,000 µg/L, 32.8 ft (10 m) cable
15,290.00
15,290.00
TITANIUM WITHOUT CLEANING UNIT
LXV441.99.12102 0–500 µg/L, 32.8 ft (10 m) cable
LXV441.99.12302 0–500 µg/L, 5 ft (1.5 m) cable
LXV441.99.22102 0–5,000 µg/L, 32.8 ft (10 m) cable
LXV441.99.22302 0–5,000 µg/L, 5 ft (1.5 m) cable
16,990.00
16,990.00
16,990.00
16,990.00
TITANIUM WITH CLEANING UNIT
LXV441.99.12202 0–500 µg/L, 32.8 ft (10 m) cable
17,290.00
LXV441.99.22202 0–5,000 µg/L, 32.8 ft (10 m) cable
17,290.00
Note: Probes with cleaning unit cannot be operated in combination
with the flow cell, Prod. No. LZY669.
CONTROLLER
This sensor requires a Hach sc100 or sc1000 Digital Controller.
See pages 388-393 for details.
RECOMMENDED ACCESSORIES
Mounting Hardware:
LZX914.99.11110 SS chain mounting set
LZY669
Flow cell with mounting panel
293.00
1,859.00
• Drinking Water
• Wastewater
• Pure Water/Power
• Industrial Water
Specifications*
Measurement Method
UV fluorescence method for
polycyclic aromatic
hydrocarbons (PAH)
Light Source
Miniature xenon flashlamp with
interference filter
Detector
UV photodiode with interference
filter; Compensation of daylight and
flashlamp intensity fluctuations
Excitation Wavelength
254 nm
Measurement Wavelength
360 nm
Measuring Range
Low Measuring Range:
0–50 ppb (µg/L) and
0–500 ppb (µg/L) (PAH)**
0.1–1.5 ppm (mg/L) and
0.1–15 ppm (mg/L) (oil)**
High Measuring Range:
0–500 ppb (µg/L) and
0–5,000 ppb (µg/L) (PAH)**
0.1–15 ppm (mg/L) and
0.1–150 ppm (mg/L) (oil)**
Resolution
0.1 ppb (µg/L) (PAH) in the
lowest measuring range
Limit of Detection (LOD)
is 1.2 ppb (PAH)
Reproducibility
2.5% of measured value at constant
temperature
Response Time
10 s (T90)
Calibration
Factory calibrated with UV
fluorescence standard or process
calibration with results of a grab
sample analysis.
Sample Temperature
33.8 to 104°F or 1 to 40°C
Pressure Range
Max. 30 bar or 435 psia
(measurement probe)
Housing
Stainless steel 316Ti (1.4571)
or titanium
Dimensions
2.68” x 12.05” or 68 x 306 mm
(D x H; without connector and
suspension pin)
Weight
Stainless Steel: 6.2 lbs or 2.8 kg
Titanium: 4 lbs or 1.8 kg
**With Calibration Standard.
*Subject to change without notice.
Find it here… Buy it today on www.hach.com
U.S. customers only.
431
Process Instruments
Organics: UVAS sc Sensor
Continuously protect plant treatment processes
from high influent organic loads.
• Continuous, automatic early warning systems
• Control activated sludge processes
• Integrated self-cleaning wiper system
• Monitor efficiency of UV disinfection process
• Self diagnostics and easy maintenance
Principal of Operation
The Hach UVAS sc UV Absorbance / %Transmittance Sensor
determines the Spectral Absorption Coefficient (SAC) at a wavelength
of 254 nm. Measurements can be expressed in absorption units
(m–1), mE, AU, %T, %T/cm, mg/L, or ppm.
Primary Applications
• Drinking Water
• Wastewater
•Industrial Water
Specifications*
Measurement Technique
UV absorption measurement
(2-beam technique), reagent-free
Measurement Method
SAC 254 in accordance with
DIN 38404 C3
Measurement Path Length
UVAS sc Tank Sensors
1, 2, 5 and 50 mm
UVAS sc Bypass Sensors
2, 5, and 50 mm
Measurement Range
UVAS sc Tank Sensors
Choice of:
0.01 to 60 m–1 at 50 mm
0.1 to 600 m–1 at 5 mm
0 to 1500 m–1 at 2 mm
2 to 3000 m–1 at 1 mm
UVAS sc Bypass Sensors
Choice of:
0.01 to 60 m–1 at 50 mm
0.1 to 600 m–1 at 5 mm
0 to 1500 m–1 at 2 mm
Compensation
550 nm
Measurement Interval
≥ 1 minute
Sample Temperature
2 to 40°C (35.6 to 104°F)
Sample pH
4.5 to 9 pH
Probe Pressure Limit at Inlet
UVAS sc Tank Sensors
0.5 bar (7.25 psi) maximum
UVAS sc Bypass Sensors
n/a
Sample Flow Rate
UVAS sc Tank Sensors
n/a
UVAS sc Bypass Sensors
0.5 L/hour minimum
Prod. No.
Price
$15,954.00
15,988.00
15,954.00
15,954.00
UVAS sc SENSOR ONLY
LXV418.99.10002
1 mm UVAS sc sensor only
LXV418.99.20002
2 mm UVAS sc sensor only
LXV418.99.50002
5 mm UVAS sc sensor only
LXV418.99.90002
50 mm UVAS sc sensor only
14,684.00
14,684.00
14,684.00
14,684.00
CONTROLLER
This sensor requires a Hach sc100 or sc1000 Digital Controller.
See pages 388-393 for details.
Sample Connection
UVAS sc Tank Sensors
n/a
UVAS sc Bypass Sensors
4 mm ID/6 mm OD hose
Cable Length
10 to 120 m (32.8 to 393.7 ft.)
Control Function
PID, time control, 2-point controller
(with sc100 or sc1000)
Inspection Interval
6 months
User Maintenance
1 h / month, typical
Dimensions
70 x 333 mm
(2.75 x 13.11 in.) approximate
BYPASS PANEL
LZX868
Bypass Panel for 50 mm sensor
LZX867
Bypass Panel for 5 mm sensor
LZX869
Bypass Panel for 2 mm sensor
MOUNTING HARDWARE
LZX414.00.10000
Mounting Hardware
with 90 degree adapter
ACCESSORIES
LZX148
Spare wiper blades for 1 mm UVAS sc, pk/5
LZX012
Spare wiper blades for 2 mm UVAS sc, pk/5
LZX119
Spare wiper blades for 50 mm UVAS sc, pk/5
LZX396
Calibration Verification filter
Weight
3.6 kg (7.9 lb.) approximate
*Subject to change without notice.
432
Description
UVAS sc UV ABSORBANCE/TRANSMITTANCE SYSTEM
The following systems include the Hach sc100 controller.
6945000
1 mm UVAS sc sensor
6945100
2 mm UVAS sc sensor
6945200
5 mm UVAS sc sensor
6945300
50 mm UVAS sc sensor
For more information, call to request
Literature #2485, or visit www.hach.com
800-227-4224
Outside the United States, call 970-669-3050
1,814.00
1,934.00
1,934.00
378.00
402.00
188.00
99.55
97.60
Oxygen Scavenger/Hydrazine: 9186 Analyzer
Process Instruments
Greater sensitivity in your
feedwater.
• 3 electrode-controlled potential amperometry
to ensure low signal drift
• High sensitivity
• Quick response time
• No moving parts or pumps
• Low reagent consumption
• Self-cleaning electrode
Low Drift
The Hach 9186 Oxygen Scavenger analyzer uses a three electrode
amperometric cell which offers an excellent zero stability. The
working electrode, counter electrode, and reference electrode allow
this stability. The working electrode is self-cleaning by turning Teflon®
beads. Calibrations are done via comparison method.
Low Maintenance
No moving parts or pumps and insignificant signal drift make
maintenance commonly less than 15 minutes per month.
Self-Cleaning Electrodes
Teflon® beads, driven by sample flow, circulate on the surface of the
platinum electrode to prevent deposits. This reduces maintenance
costs and analyzer downtime.
The Hach 9186 Oxygen Scavenger Analyzer is shipped on a panel
complete with controller, probe, cable, flow cell and all hardware
necessary for installation.
Prod. No.
Description
Price
COMPLETE ANALYZER
19186=A=3011
Hach 9186
$11,321.00
Oxygen Scavenger Analyzer
2834453
Diisopropylamine, 99% 1L
64.49
ACCESSORIES
09186=C=0360
4630600
4630800
Bottle cap adapter
Power cord 125V
Power cord 230V
143.00
39.60
45.00
REPLACEMENT PARTS
09186=A=8000
Maintenance kit for 2 years**
1,319.00
**Maintenance kit includes 6 filters, 1 reference electrode, 1 Venturi injection
nozzle, 7 plastic beads, 2 meters of 4x 6mm PE tubing.
Primary Applications
• Pure Water/Power
• Industrial Water
Specifications*
Range
5 - 500 µg/L Hydrazine
2 - 100 µg/L Carbohydrazide
Repeatability
< ± 2% of reading or < ± 1 µg/L N2H4 whichever is greater
Response Time
T90 approximately 60 seconds
Service Intervals
Every 4-5 weeks
Analog Output
Two 0/4-20 mA outputs, max. impedance 500 Ohms
Operating Temperature Range
5 to 45ºC (41 to 113ºF)
Weight
9.2 kg (20 lbs.)
For more information, call to request
Literature #2452, or visit www.hach.com
*Subject to change without notice.
See page 178 for reagents, test kits, and accessories for measuring oxygen scavenger/hydrazine in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
433
Process Instruments
Ozone, Dissolved: 9185 sc Analyzer
Repeatable dissolved ozone measurements.
• Great for readings in low conductivity water
with no interferences from oxidants or pH
• Reagentless analysis of ozone—ion-selective
membrane separates electrolyte from sample water
• All-inclusive, pre-assembled panel reduces
installation expenses
• Plug-and-play with digital controllers simplifies setup
• Integral temperature sensor provides more
accurate readings
• Includes 2 years of typical maintenance parts
Optional acidification (for
cleaning) and intermittent flow
accessories link in series so
they only take up one sensor
port in the controller.
Prod. No.
Note: sc100 shown with optional mounting panel.
Primary Applications
• Drinking Water
• Wastewater
• Pure Water/Power
• Industrial Water
Specifications*
Range
0.005 ppm to 2 ppm (0.005 mg/L to 2 mg/L)
Accuracy
±3% or ±10 ppb O3 whichever is greater
Cycle Time
90% in T<90sec
Description
Price
COMPLETE ANALYZERS
6043300
9185 sc Ozone Sensor
$3,604.00
Preassembled panel including Ozone
probe with integral temperature and
flow control, sc100 controller, and
mounting panel for sc100
6043301
Same as 6043300 but with
3,758.00
MODBUS® RS485 output
6043302
Same as 6043300 but with
3,823.00
MODBUS®RS232 output
SENSOR ONLY
LXV433.99.00001 9185 sc Ozone Sensor
2,224.00
CONTROLLERS ONLY
This sensor requires a Hach sc100 or sc1000 Digital Controller.
See pages 388-393 for details.
ACCESSORIES
LZY051
LZY052
LXY060
9180 sc Acidification Unit
Also used for cleaning
9180 sc Intermittent Flow
Mounting panel for sc100
1,384.00
1,024.00
236.00
For more information, call to request
Literature #2407, or visit www.hach.com
*Subject to change without notice.
See page 178 for reagents, test kits, and accessories for measuring ozone in the lab or field.
434
800-227-4224
Outside the United States, call 970-669-3050
Ozone: C1100 Sensor
Process Instruments
The NEW C1100 Ozone Sensor for bottled water,
UPW pharmaceutical, and drinking water applications.
• Installs in-line to eliminate continual product loss
• Sensor refurbishment in under 5 minutes with
pre-filled recharge cartridge
• Quick and easy air calibration
• Stainless steel body extends sensor life
Principle of Operation
The sensor is constructed of two metal electrodes, the noble
working electrode, immersed in an electrolytic solution, and
separated by a gas permeable membrane from the sample
if interest. An auxiliary guard ring electrode surrounds the
working electrode to shield against the influence of other
gases and improve stability.
Price
Prod. No. Description
C1100-T00 Electrochemical ozone sensor,
$4,725.00
Titanium version, Maximum pressure 100 bar,
with Smart capability
C1100-S00 Electrochemical ozone sensor,
3,360.00
Stainless Steel version, Maximum pressure
40 bar, with Smart capability
2956A-C
Recharge kit of 4 pre-filled cartridges
299.00
with premounted 2956A membranes for
C1100 ozone sensors. Includes o-rings,
cotton filters, cleaning tools and dacrons
Primary Applications
• Bottled Water
• Ultrapure Water Pharmaceutical
Specifications*
Measuring Range
0 ppb-50 ppm O3
Response Time
30 s
Accuracy
±0.4 ppb or ±5% whichever is the
greater
Working Operating Range
-5 to 45°C
Repeatability
±0.4 ppb or ±5% whichever is
the greater
Pressure Rating
(Stainless Steel/Titanium)
40 bars / 100 bars
580 psi / 1450 psi
For more information, call to request
Literature #2680 or visit www.hach.com
• Drinking Water
Limit of Detection
0.6 ppb
Maximum Operating Range
-5 to 100°C
Membrane
2956A-C
Typical Flow Rate
350 mL/min
*Subject to change without notice.
See page 178 for reagents, test kits, and accessories for measuring ozone in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
435
Particle Counting: 2200 PCX
Process Instruments
Optimize your filter analysis.
• Supports up to 32 size channels and 8 analog inputs
(turbidity, pH, etc.)
• Operates stand-alone or as part of a networked system
• Volumetric—all particles pass through the sensing area
For maximum flexibility in an on-line particle counter, the Model 2200
PCX is the most versatile counter available. The 2200 PCX offers
serial connection to powerful software, simultaneous 4-20 mA
SCADA serial connection, and serial connection to SCADA.
Vista Software
For medium to large systems, up to to 32 sensors, Vista offers
“intelligent monitoring.” Intelligent monitoring minimizes total data
stored, but maximizes useful information. Intelligent monitoring
ensures that you capture critical information about filter-to-waste,
impending filter breakthrough, or unexpected excursions.
Additionally, Vista Data Management software is a high-performance
package for Windows 2000/XP/Vista computers. Each provides
clear tabular and graphical displays to assess plant performance
with a glance.
OPC Explorer Software
If you want a direct RS-485 connection between our particle counters
and your SCADA system, use the 2200 PCX Explorer. It is an OLE for
Process Control (OPC) driver for Microsoft® Windows® designed to
easily connect particle counters to OPC clients (such as SCADA
software or data loggers).
Prod. No.
Primary Applications
• Drinking Water
Specifications*
Counting Range
2-750 microns
MODEL 2200 PCX PARTICLE COUNTER - 230V VERSION
5705001
2200 PCX Particle Counter
5,360.00
with ANALOG with WATER WEIR
5704001
2200 Particle Counter
4,563.00
with WATER WEIR
Flow Rate
100 mL/minute
Maximum Pressure
65 psig, not more than 1 minute; 55 psig continuous
Sample Time
1 second to 24 hours
Fluid Connections Inlet
Barbed fitting with self-sealing quick disconnect for 1/4" O.D. tubing
Power
100- 230 Vac; 50-60 Hz
Dimensions
13.8"H x 8.3"W x 7"D
MODEL WGS 267 GRAB SAMPLING INSTRUMENT
20844445-01 WGS 267 Grab Sampling Instrument
10,975.00
DATA ACQUISITION SOFTWARE - 2200 PCX
5702500
OPC Software - 2200 PCX
5701525
Vista Software V1.3.1
919.00
1,594.00
DATA ACQUISITION SOFTWARE - WGS 267
5700510
PortAll Software
797.00
For more information, call to request
Literature #4599, or visit www.hach.com
*Subject to change without notice.
436
Price
MODEL 2200 PCX PARTICLE COUNTER - 115V VERSION
5705000
2200 PCX Particle Counter
$5,360.00
with ANALOG with WATER WEIR
5704000
2200 PCX Particle Counter
4,563.00
with WATER WEIR
• Pure Water/Power
Fluid Connections Outlet
Quick disconnect for 1/4" O.D. tubing
Description
800-227-4224
Outside the United States, call 970-669-3050
pH/ORP: Differential Sensors
Process Instruments
The smart choice for accurate, reliable, and
dependable pH/ORP measurement.
Differential Electrode Measurement Technique
This field-proven technique uses three
electrodes instead of the two used in
conventional pH sensors. Process and
reference electrodes measure the pH
differentially with respect to a third
ground electrode.
The result is unsurpassed
measurement accuracy, reduced
reference junction potential, and
elimination of sensor ground loops.
These sensors provide greater
reliability, resulting in less downtime
and maintenance.
Patented Technology
The former GLI, now a Hach Company brand, invented the
Differential Electrode Technique for pH measurement in 1970. The
pHD™ sensor series (U.S. Patent Number 6395158B1, dated May
28, 2002) takes this field-proven technology to a new level.
Hach Digital pHD sc sensors are available in convertible (PEEK® or Ryton®),
insertion, and sanitary body styles. Three electrodes are used in these sensors
to increase measurement accuracy and eliminate sensor ground loops.
Replaceable Salt Bridge/Protector
The unique, replaceable salt bridge holds an extraordinary volume of
buffer to extend the working life of the sensor by protecting the
reference electrode from harsh process conditions.
Built-in Encapsulated Preamp
Encapsulated construction protects the sensor’s built-in preamp from
moisture and humidity, ensuring reliable sensor operation. The
preamp in the pHD analog sensor produces a strong signal, enabling
the sensor to be located up to 1000 m (3280 ft.) from the analyzer.
Wide range pH glass
(HF resistant glass also available)
Durable Body Materials
Both the digital and analog pH and ORP differential sensors feature a
durable PEEK® body for chemical compatibility with most process
solutions. For less aggressive solutions, Hach offers a Ryton® sensor
in a convertible style for pH and ORP measurement. A sensor with a
stainless steel body is available for immersion applications.
Primary Applications
Versatile Mounting Styles
Sensors are available in four mounting styles—convertible, insertion,
immersion, and sanitary.
• Drinking Water
• Wastewater
• Pure Water/Power
• Industrial Water
• Environmental
• Food and Beverage
General Specifications*
Differential Sensor Warranty
Hach Company offers the best sensor warranty in the industry on
its Differential Sensors. We will replace any Differential Sensor that
fails due to defects in materials or workmanship within one year
from the date of shipment—and up to 30 months on a prorated
basis for any failure
Available in analog or digital versions.
pH Range
-2 to 14 pH
Operating Temperature Range
Analog Sensor:-5 to 105°C (23 to 221°F)
Digital Sensor: -5 to 70°C (23 to 158°F)
Pressure Range
0 to 6.9 bar (100 psi)
ORP Range
-1500 to +1500 mV
See next page for ordering information.
For more information, visit
www.hach.com/ProcesspHSensors
By replacing the salt bridge and standard cell solution,
Hach Differential Sensors can be regenerated for repeated use.
Pressure Range
0 to 6.9 bar (100 psi)
*Dependent on specific sensor and mounting.
See pages 179-181 for reagents, test kits, and accessories for measuring pH in the lab or field.
Find it here… Buy it today on www.hach.com
U.S. customers only.
437
Process Instruments
pH/ORP: Differential Sensors
Prod. No.
Price
CONTROLLER REQUIRED
For information about Hach digital and analog controllers,
see pages 388-398.
ORP Sensors
Body Material
PEEK1
PEEK1
PEEK1
Ryton2
Ryton2
Stainless Steel
PHD sc ANALOG SENSORS
pH Sensors
Body Material
PD1P1
PEEK1
3
PD1P3
PEEK1
PD2P1
PEEK1
PD3P1
PEEK1
PD1R1
Ryton2
PD1R33
Ryton2
Body Style
Convertible
Convertible
Insertion
Convertible
Convertible
Immersion
845.00
895.00
870.00
795.00
845.00
843.00
Body Style
Convertible
Convertible
Insertion
Sanitary
Convertible
Convertible
642.00
792.00
667.00
692.00
592.00
742.00
Body Style
Convertible
Convertible
Insertion
Convertible
Convertible
694.00
744.00
719.00
644.00
694.00
ORP Sensors
RD1P5
RD1P6
RD2P5
RD1R5
RD1R6
Body Material
PEEK1
PEEK1
PEEK1
Ryton2
Ryton2
1Polyetheretherketone; 2Polyphenelene
Sulfide; 3HF resistant glass
DIGITAL GATEWAY
6120500
Use the Digital Gateway to connect pHD
analog sensors to a Hach
digital controller.
Price
pHD sc DIGITAL AND PHD ANALOG SENSOR ACCESSORIES
pHD sc DIGITAL DIFFERENTIAL pH/ORP SENSORS
All digital sensors include built-in digital electronics and
integral 10 m (33 ft.) cable terminated with connector for a
digital controller. Body styles:
• Convertible – 1-inch NPT threads at both ends, designed for teemounting or other flow through mountings, and pipe mounting
for immersion
• Insertion – no threads on the electrode end, designed for use
with insertion valve assembly or flow-through cell
• Sanitary – 2-inch flange for a tri-clover style fitting
• Immersion – used with chain mounting or pipe mounting
pH Sensors
Body Material
Body Style
DPD1P1
PEEK1
Convertible
$793.00
DPD1P33
PEEK1
Convertible
943.00
DPD2P1
PEEK1
Insertion
818.00
DPD3P1
PEEK1
Sanitary
843.00
DPD1R1
Ryton2
Convertible
743.00
DPD1R33
Ryton2
Convertible
893.00
DPS1
Stainless Steel
Immersion
844.00
DRD1P5
DRD1P6
DRD2P5
DRD1R5
DRD1R6
DRS5
Prod. No.
Cables
Extension cables are used only with digital sensors or digital
gateways when connecting to a digital controller.
6122400
Digital Extension Cable, 1 m (3.2 ft.)
$112.00
5796000
Digital Extension Cable, 7.7 m (25 ft.)
142.00
5796100
Digital Extension Cable, 15 m (50 ft.)
213.00
5796200
Digital Extension Cable, 31 m (100 ft.)
284.00
Interconnect cables are used only with analog sensors, junction box,
and controller.
1W1100
Analog Interconnect Cable, order per foot
2.19
Digital Termination Box
Required when the length of cable between the digital sensor/digital
gateway and a digital controller is between 100 m (328 ft.) and
1000 m (3280 ft.)
5867000
Digital Termination Box
123.00
Analog Junction Box
Required when the length of cable between the analog sensor and
analog controller is greater than standard length of sensor cable.
Each junction box includes terminal strip and gasket.
60A2053
Junction Box, Surface-mount, aluminum
37.49
(includes mounting hardware)
60A9944
Junction Box, Pipe-mount, PVC
69.45
(for 1/2-inch diameter pipe, includes
mounting hardware)
60G2052
Junction Box, Pipe-mount, PVC
68.39
(for 1-inch diameter pipe, includes
mounting hardware)
76A4010-001
Junction Box, NEMA 4X
156.00
(no mounting hardware included)
pHD sc Digital and pHD Analog Sensor Reagents and Standards
25M1A1025-115 Standard Cell Solution,
53.15
to replenish standard cell chamber
in Hach pHD sensors while
replacing salt bridge, 500 mL
25M8A1002-101 Gel Powder,
46.85
for >95°C applications, 2 g
pH Buffers
2283549
2283449
2283649
Description
pH 7
pH 4
pH 10
ORP Reference Solutions (in resealable
Description
25M2A1001-115
200 mV
25M2A1002-115
600 mV
Volume
500 mL (1 pint)
500 mL (1 pint)
500 mL (1 pint)
9.69
9.69
9.69
plastic bottles)
Volume
500 mL (1 pint)
500 mL (1 pint)
46.85
46.85
pH/ORP systems with Class I Division II safety classification are
available— please contact your Hach representative.
111.00
For additional information and available mounting
hardware options, call Hach or download product data
sheet (Lit. #2467) from www.hach.com/ProcesspHSensors.
See pages 179-181 for reagents, test kits, and accessories for measuring pH in the lab or field.
438
800-227-4224
Outside the United States, call 970-669-3050