OPERATIONAL MANUAL
XJM 300
BMS Bulut Makina Sanayi Ve Ticaret Ltd. Şti.
İkitelli Organize Sanayi Bölgesi Dolapdere Sanayi Sitesi
Ada 4 No : 7-9 Başakşehir / İSTANBUL
Tel : 0 212 671 02 24 / 25 Fax : 0 212 671 02 26
web site : www.bulutmak.com e-mail : bms@bulutmak.com
Contents
Contents......................................................................................................................................................................... 1
1
Metallurgical Microscope ......................................................................................................................................... 2
2
Purpose ..................................................................................................................................................................... 3
3
Specification .............................................................................................................................................................. 3
4
Installation ................................................................................................................................................................ 5
5
Operation .................................................................................................................................................................. 6
6
Maintenance ............................................................................................................................................................. 8
7
Troubleshooting ........................................................................................................................................................ 9
8
XJM300 Metallurgical Microscope Outfits .............................................................................................................. 10
1
1 Metallurgical Microscope
1.1
Operation Instruction
Fig.1 XJM300 Inverted Metallurgical Microscope (1)
Fig.2 XJM300
Inverted
Metallurgical
Microscope
(2)
2
2 Purpose
XJM Metallurgical Microscope which with metallurgical plan objectives, polarizing attachment and 1X or 0.6X C-mount is
widely used in observation & analysis of metallurgical organization in mechanical industry, research of geological &
mineral department and viewing and measuring crystal, integrate circuit, micro-electronics etc. in electronic industry. It is
the best choice of factories, academies, scientific research organization and electronic industrial.
3 Specification
3.1
Total Magnification
Objective
2.5X
4X
10X
20X
40X
100X
12.5X
20X
50X
100X
200X
500X
10X
25X
40X
100X
200X
400X
1000X
12.5X
31X
50X
125X
250X
500X
1250X
16X
40X
64X
160X
320X
640X
1600X
Eyepiece
5X
3.2
Objectives
Magnification
Numerical
Aperture
(N.A.)
Working
Distance
(mm)
System
Colour
Ring
Obj.Mark
Obj.Type
LPL2.5/0.07
160/LPL4/0.10
160/PL10/0.25
160
PL20/0.35
160
PL40/0.65
160
PL100/1.25OIL
160
LWD
Plan
LWD
Plan
2.5×
0.07
29.7
Dry
4×
0.10
17.5
Dry
Red
Plan
10×
0.25
8.8
Dry
Yellow
Plan
20×
0.35
1.59
Dry
Green
Plan
40×
0.65
0.73
Dry
Blue
Plan
100×
1.25
0.69
Oil
White
Remark:20×、40×、100× are with spring
3
3.3
Eyepieces
Eyepiece
Mark
Eyepiece Type
Field Dia.
(mm)
Magnification
Focus
(mm)
WF10×-20
Plan Wide Field
10×
φ20
25
WF10×-18
Reticle
Eyepiece
10×
(0.1mm)
φ18
25
WF16×
Plan Eyepiece
16×
φ13
15.6
WF12.5×
Plan Eyepiece
12.5×
φ14
20
5×
Huygenian Eyepiece
5×
φ20
50
3.4 Stage
Three layer mechanical stage size:180mm×167mm。
X-Y travel range:50mm×40mm;
Division :0.1mm;
Stage plate size:φ10mm,φ20mm
Φ40mm optional
3.5
Filter set: green, blue, grey filter and white grass
3.6 Inllumination
6V/30W halogen bulb, adjustable brightness
3.7 Nosepiece
Quadruple nosepiece
3.8 Seidentopf Binocular Head
Inclined 45°,interpupillary adjustable distance is 50mm~75mm,
Diopter adjustable range ±5,
Trinocular head is fixed.
3.9 Coaxial Coarse & Fine Focusing System
Coaxial coarse & fine focusing knobs, coarse stroke 42.4mm/rotation, actual coarse stroke 12mm;
Fine stroke 0.2/rotation, fine division 0.002mm
3.10 Anti-fungal systems
Eyepiece and objective is anti-fungal.
3.11 Power Supply
Input:85V~265V,50/60Hz;
Fuse:3A 250V(φ5×20)
;
Power switch is separate with potentiometer
4
4 Installation
Please install the microscope as following:
Fig.3
4.1
4.2
4.3
4.4
4.5
4.6
Set the main body on a desk carefully.
Put out the dust cover off the eyepiece tubes, and insert a pair of eyepiece with same times into the tubes.
Turn the coarse focusing knob to low the nosepiece to bottom. Put out the dust cover off nosepiece, and drive
the objectives into the holes of nosepiece orderly.
Connect the lamp house with the connecting hole on the back of main body, and tighten it with both thumb
screws. Put the plug into the socket in the back of the main body, and tighten it with screw.
Put the power plug into power supply socket.
Put the stage plate into the big holes in stage, and insert the clamps into the small holes in stage.
5
5 Operation
Fig.4
5.1
5.2
5.3
5.4
5.5
. Adjust the potentiometer to minimize the light intensity, and put on the power switch, then adjust the light
intensity applicably.
Please insert an available stage plate into the stage, and put the sample on the plate. Please pay attention to take
the surface which would be observed down, and fix the sample with clamp. Rotate the stage moving knob to
move the sample into the light path.
Turn the nosepiece by holding its rim to bring an objective of low power into the light path, rotating the coarse
focusing knob. Observe the image with left eye through left tube to get a rough image, then put the high power
objective into the light path after get a clear image by adjusting fine focusing knob.
Interpupillary distance adjustment and diopter adjustment
4.4.1 Interpupillary distance adjustment
Interpupillar distance is different for everyone, so interpupillar distance should be adjusted before using
binocular microscope. Please take the two eyepiece tubes to turn until the bright rings observed by two eyes
are in superposition entirely.
5.6
Diopter adjustment
As focusing for binocular, user should observe right eyepiece with right eye, and make the right eyepiece
clear by focusing adjustment, then observe the left eyepiece, at the same time, adjust the diopter ring of the
left eyepiece tube to make the image of left eyepiece clear as same as the right eyepiece.
5.7
Field diaphragm adjustment
6
Stir the field diaphragm lever or turn
revolver to make the field diaphragm
view field of eyepiece, and the field
image can seen. Adjust field diaphragm
adjustment screw to make the center of
image superposed with the center of
field. Take the field diaphragm image
eyepiece view field and adjust again to
Fig.5
of field diaphragm image entirely superposed
with the center of eyepiece
then adjust field diaphragm
make field diaphragm
disappeared.
5.8
field diaphragm
smaller
than
diaphragm
center
field diaphragm
eyepiece view
smaller
than
make the center
view field,
bigger to
image
Aperture diaphragm
adjustment
Stir the condenser aperture
lever or turn
aperture ring to adjust the
size aperture
for adjusting the contrast of
the
specimen.
5.9
Using filter
Blue, green, grey and
set in the filter revolver,
choose the available filter
white filter is
and please
and put in
Fig.6
into light path.
5.10 Adjusting lamp house
5.11 Stir the aperture diaphragm to the
white filter to light path.
5.12 Turn left & right adjusting knob
and front & back adjusting knob
adjust the lamp center for making
brightest and even.
Finished the above process, please
biggest, and put
for lamp center
for lamp center to
the view field
choose the
available objective,
Fig.7
eyepiece and filter as per the sample, and adjust aperture diaphragm, lamp brightness and fine focusing knob
to get clear image.
5.13 Using polarizing attachment
Insert polarizer slide and analyzer slide into the two grooves on the right of microscope. The side with arrow
mark should be in the direction of head, and the analyzer slide with arrow mark should be up. Turn the
revolver of analyzer slide slowly to observe.
5.14 Usage of the tension adjustment ring (Shown
in Fig.4)
The tension adjustment ring can adjust the
and fine focusing unit to prevent the stage from
automatically and to improve the comfort of
clockwise makes tension decrease, and by
counterclockwise makes tension increase.
tension of the coarse
sliding down
operation. Rotation
contraries, rotation
Fig.8
7
▲ Don’t turn left and right coarse and fine focusing knobs with different direction with power at
the same time, if so, the focusing system will be damaged.(Fig.4)
5.15 Installing and using of camera and CCD attachment
5.16 Installing
Connect the C-mount with CCD camera or connect camera with camera adaptor, then connect it with
c-mount, finally put it into microscope.
5.17 Using
First get a clear image from eyepiece, then put out lever on the side of trinocular head, and collect image
with camera. Clear image should be in screen. Adjust fine focusing knobs to get it clear if image isn’t clear.
5.18 Usage of immersion oil objective
First of all, take 40X objective which has been focused well out from bright path, then take 100X objective
into bright path. At this time, nosepiece or stage moving knob should be turned slightly, and also slightly
turn fine focusing knob to clear away the air bubble of immersion oil, otherwise, the air bubble would affect
the image badly.
▲ After immersion oil used, the oil of specimen and the microscope surface should be soon cleaned by
absorbent cotton, lens paper, gauze or soft cotton cloth with moderate mixture of pure industrial alcohol
and ether (proportion 1:4).
6 Maintenance
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.8
6.9
6.10
6.11
6.12
6.13
Clean microscope
Don’t touch the lens with hand, Dust on lens should be cleaned by soft brush or absorbent cotton or cleaned by
absorbent cotton, lens paper with the mixture of alcohol and ether (proportion 1:4).
Alcohol and ether all are burnt early, please take them away from fire. Be careful for turn on and off power.
Don’t clean painted metal and galvanizing metal with organic solvent such as alcohol, ether or the mixture of
the both. Silicon cloth or soft cleaning preparation is suggested to clean it.
Plastic should be cleaned by soft cloth with clear water.
Environment of using and placing
Microscope should be used and placed in a cool, dry, non-dust, non-shake and non-corrosive gases
environment.
Microscope should be used in environment of indoor temperature 0°-40°C and maximum relative humidity
85%.
Removing equipment is suggested to be installed when microscope used in heavy humidity area to avoid fungus
and mist damage instrument.
Please pay attention to prevent microscope from violent shake and vibration in application and in carrying.
Don’t drag it on the surface of worktable to avoid damage to microscope and worktable.
Replacement of bulb
Turn off power, and pull out plug.
Wait the bulb become cool.
▲ Please be sure that the bulb is cool, then follow by the nest operations.
8
6.14 Lay aside the microscope reliably, unscrew the two knurled thumb screws in the right side of lamp housing
cover.
6.15 Pull over the lamp housing cover.
6.16 Pull out the bulb should be replaced, hold a new bulb with silk cloth to avoid fingerprint and dust affect bulb
brightness and service life, and insert fully the contact pins into the bulb socket.
6.17 Close the lamp housing cover, and screw the knurled thumb screw.
▲After working for above 10 hours continuously, better cut off the microscope about 30 minutes.
6.18
6.19
6.20
6.21
Replacement of fuse
Cut off power of microscope, and pull out the plug.
Unscrew fuse cap in the back of base, take out old fuse.
Replace a new fuse, then screw the fuse cap.
7 Troubleshooting
In the period of using XJM300 metallurgical microscope, if there is any trouble occurs, please referring to the following
sheet listed some common troubleshooting resolve them.
Trouble
Switch on but bulb Dark
Bulb is flickering or
brightness is
unsteady
Brightness of view field
isn’t enough or is
Uneven
Causation
Plug is unreliable
Plug in again
Bulb is broken
Change bulb
Fuse is broken
Change fuse
Bulb is unstable
Insert it again
Bulb is broken
Replacing bulb
Bulb specification doesn’t
meet the requirement
Replacing bulb
Brightness isn’t adjusted
correctly
Adjust rotation
potentiometer
Objective isn’t in correct
position
Make the objective in correct
position
The size of aperture diaphragm is too
small
Adjust the size of iris aperture
Lens(objective,eyepiece) has dust
Image isn’t clear
It is easy for eyes to be
tired during observing
Remedy
Clean it
Immersion oil isn’t used for 100X
objective (oil)
Use immersion oil
There is bubble in immersion oil
Clear the bubble way
Size of aperture diaphragm isn’t proper
Adjust the size of iris aperture
Diopter isn’t adjusted correctly
Readjust diopter
Interpupilary distance isn’t adjusted
correctly
Adjust interpupilary
distance according to two eyes
9
8 XJM300 Metallurgical Microscope Outfits
Items
Eyepieces
Plan Achromatic Objectives
Long Working Distance
Plan Objectives
Specifications
XJM300
WF10×/Φ18
●●
WF12.5×/Φ14
●●
WF16×/Φ13
○○
Huygenian Eyepieces 5×/Φ20
○○
WF10×/Φ18 (Reticule,0.1mm)
●
Plan 4×/0.10
○
Plan 10×/0.25
●
Plan 20×/0.35 (S)
●
Plan 40×/0.65 (S)
●
Plan 100×/1.25 (S) (Oil)
●
PL L 40×/0.60
W.D=3.64mm
○
PL L 50×/0.70
W.D=3.01mm
○
PL L 80×/0.80(S)
W.D=1.08mm
○
Trinocular Head
Inclined 45º, Light Distribution 20:80
●
Nosepiece
Quadplex
●
Mechanical Stage
Size: 180×165mm, Travel Range: 50×40mm
●
1 (Φ10)
●
2 (Φ20)
●
3 (Φ40)
○
Stage Plate
Set of Stage Plate
●
Main Body
●
Polarizing Slide
●
Analyzing Slide
●
Illuminator
6V/30W Halogen Lamp, Input voltage:100V~240V
●
Filters
Blue, Green, Grey, White
●
1X C-mount
Available for digital photo camera
○
0.6X C-mount
Available for digital camera
○
Micrometer
Division 0.01mm
○
Note:
“●”in table is standard outfits,“○”in table is optional accessories.
10