HI 93501 - HI 93502 HI 93503 - HI 93510

Instruction Manual
HI 93501 - HI 93502
HI 93503 - HI 93510
Portable
Microprocessor
Thermistor
Thermometers
ins ru
m n
s
°C
CLEA
R
HI
LO
ON
OFF
HOLD
PRINTED IN ITALY
°C
°F
MANL93R2
03/97
h t t p : / / w w w . h a n n a c a n . c o m
http://www.hannacan.com
These Instruments are in
Compliance with the CE Directives
Dear Customer,
Thank you for choosing a Hanna product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with the necessary information for a
correct use of the instrument, as well as a more precise idea of its
versatility. If you need additional technical information, do not
hesitate to e-mail us at techserv@hannacan.com.
These instruments are in compliance with the
directives
EN 50081-1 and EN 50082-1.
CE DECLARATION OF CONFORMITY
DECLARATION OF CONFORMITY
We
TABLE OF CONTENTS
Hanna Instruments Srl
V.le delle industrie 12
35010 Ronchi di Villafranca (PD)
ITALY
herewith certify that the thermometers
PRELIMINARY EXAMINATION ........................................................ 3
GENERAL DESCRIPTION ................................................................ 3
SPECIFICATIONS OF HI 93501 .................................................... 4
SPECIFICATIONS OF HI 93502 .................................................... 5
SPECIFICATIONS OF HI 93503 .................................................... 6
SPECIFICATIONS OF HI 93510 .................................................... 7
OPERATIONAL GUIDE ................................................................... 8
DISPLAY CODE GUIDE ................................................................ 10
CALIBRATION ............................................................................. 10
TEMPERATURE PROBES .............................................................. 11
SHOCKPROOF RUBBER BOOTS ................................................... 14
OTHER ACCESSORIES .................................................................. 14
WARRANTY ................................................................................ 14
CE DECLARATION OF CONFORMITY .............................................. 15
ISO 9000 Certified
Company since 1992
2
HI 93501
HI 93502
HI 93503
HI 93510
have been tested and found to be in compliance with the following regulations:
IEC 801-2
IEC 801-3
EN 55022
Electrostatic Discharge
RF Radiated
Radiated, Class B
Date of Issue: 02-10-1995
D.Volpato - Engineering Manager
On behalf of
Hanna Instruments S.r.l.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the
environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60 VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.
15
SHOCKPROOF RUBBER BOOTS
These rubber boots are specially made to
prolong the life of your instruments and to
prevent damages due to bangs and falls.
They measure 155 x 90 x 45 mm
(6.1x3.5x1.8") and are available in two
colors for your convenience:
HI 710007
= blue colored
HI 710008
= orange colored.
°C
CLEAR
ON
OFF
HI
HOLD
LO
°C
°F
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is
any damage, notify your Dealer or the nearest Hanna Customer
Service Center.
Note: Save all packing material until you are sure that the
instrument functions correctly. All defective items must be
returned in the original packing with the supplied accessories.
GENERAL DESCRIPTION
OTHER ACCESSORIES
HI 710004
MANL93R2
Soft carrying case (for meter only)
Instruction manual
WARRANTY
All Hanna Instruments meters are warranted for two years
against defects in workmanship and materials when used for their
intended purpose and maintained according to instructions. The
probes are warranted for a period of six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of
purchase, serial number and the nature of the failure.
First obtain a Returned Goods Authorization number from the
Customer Service department, then return the instrument with the
Authorization # included along with shipment costs prepaid.
If the repair is not covered by the warranty, you will be notified of the
charges.
When shipping any instrument, make sure it is properly packaged for
complete protection.
To validate your warranty, fill out and return the enclosed warranty
card within 14 days from the date of purchase.
14
HI 93501, HI 93502, HI 93503 and HI 93510 are powerful and
flexible instruments which have been designed using the latest
microprocessor technology to provide you with reliable and accurate
measurements.
Standard features include interchangeable probes, low battery detection, auto-off capability, long battery life and two year warranty.
HI 93501, HI 93502, HI 93503 have been designed specifically by
Hanna Instruments for use in the food industry.
HI 93501, HI 93503, HI 93510 come supplied complete with a
temperature probe and a 9V battery.
HI 93502 is supplied with a 9V battery.
All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright owner,
Hanna Instruments Inc.
Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.
3
SPECIFICATIONS OF HI 93501
Main features:
• Interchangeable HI 762 probes
• Interchangeable HI 762 calibration test plugs
• Range -20 to +120°C
• Display of temperature in Celsius
• Low battery detection
• Auto-off capability
General purpose liquid probe, with 10 m (33')
cable and white handle
HI 765PBL
Penetration probe with 1 m (3.3') cable and
blue handle
HI 765PBL/10 Penetration probe with 10 m (33') cable and
blue handle
HI 765PG
Penetration probe with 1 m (3.3') cable and
green handle
HI 765PG/10 Penetration probe with 10 m (33') cable and
green handle
HI 765PR
Penetration probe with 1 m (3.3') cable and red
handle
HI 765PR/10 Penetration probe with 10 m (33') cable and
red handle
HI 765PW
Penetration probe with 1 m (3.3') cable and
white handle
HI 765PW/10 Penetration probe with 10 m (33') cable and
white handle
HI 765W
Wire probe without handle for hard-to-reach
places with 1 m (3.3') cable
HI 765W/10
Wire probe without handle for hard-to-reach
places with 10 m (33') cable
HI 765L/10
HANNA INSTRUMENTS TEMPERATURE PROBES
205 mm
8.1"
100 mm
3.9"
3 mm
0.12"
HI 762L
HI 762L/LUM
HI 765L
Range
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Battery
Environment
Dimensions
Weight
SPECIFICATIONS
-20.0 to 120.0°C
0.1°C
±0.4°C (-20 to 80°C)
±0.4% of Full Scale (outside)
for one year, excluding probe error
±0.4°C
with HI 762PW probe
9 Volt battery, 1000 hours of continuous use
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
143 x 80 x 38mm (5.6 x 3.2 x 1.5")
320 g (11.3 oz.)
4
205 mm
8.1"
100 mm
3.9"
3 mm
0.12"
HI 762P
HI 762PW/LUM
HI 765P
205 mm
8.1"
100 mm
3.9"
HI 762A
HI 762A/LUM
HI 765A
3mm
0.12"
HI 762W
HI 765W
13
3 mm
0.12"
SPECIFICATIONS OF HI 93502
Main features:
• LUMBERG socket for temperature probes
• Interchangeable LUMBERG probes
• Range -20 to +120°C
• Display of temperature in Celsius
• Low battery detection
• Auto-off capability
ins rum n s
e
General purpose liquid probe, with 10 m (33')
cable and white handle
HI 762PBL
Penetration probe with 1 m (3.3') cable and
blue handle
HI 762PBL/10 Penetration probe with 10 m (33') cable and
blue handle
HI 762PG
Penetration probe with 1 m (3.3') cable and
green handle
HI 762PG/10 Penetration probe with 10 m (33') cable and
green handle
HI 762PR
Penetration probe with 1 m (3.3') cable and red
handle
HI 762PR/10 Penetration probe with 10 m (33') cable and
red handle
HI 762PW
Penetration probe with 1 m (3.3') cable and
white handle
HI 762PW/10 Penetration probe with 10 m (33') cable and
white handle
HI 762W
Wire probe without handle for hard-to-reach
places with 1 m (3.3') cable
HI 762W/10
Wire probe, without handle for hard-to-reach
places with 10 m (33') cable
HI 762L/10
Foodcare
ON
OFF
RANGE:
-20°C to 120°C
ACCURACY:
±0.4°C
FOR HI 93502 ONLY
Air probe, with 1 m (3.3') cable and white
handle, Lumberg connector
HI 762BP/T/LUM Frozen Food between pack probe, with 1 m (3.3')
cable and white handle, Lumberg connector
HI 762L/LUM General purpose liquid probe, with 1 m (3.3')
cable and white handle, Lumberg connector
HI 762PW/LUM Penetration probe, with 1 m (3.3') cable and
white handle, Lumberg connector
HI 762A/LUM
FOR HI 93503 ONLY
HI 765A
HI 765A/10
HI 765BL
HI 765BL/10
HI 765L
Air probe, with 1 m (3.3') cable and white
handle
Air probe, with 10 m (33') cable and white
handle
General purpose liquid probe, with 1 m (3.3')
cable and black handle
General purpose liquid probe, with 10 m (33')
cable and black handle
General purpose liquid probe, with 1 m (3.3')
cable and white handle
12
HI 93502
Range
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Battery
Environment
Dimensions
Weight
SPECIFICATIONS
-20.0 to 120.0°C
0.1°C
±0.4°C (-20 to 80°C)
±0.4% of Full Scale (outside)
for one year, excluding probe error
±0.4°C
with HI 762PW/LUM probe
9 Volt battery, 1000 hours of continuous use
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
143 x 80 x 38mm (5.6 x 3.2 x 1.5")
320 g (11.3 oz.)
5
SPECIFICATIONS OF HI 93503
Main features:
• Interchangeable HI 765 probes
• Interchangeable HI 765 calibration test plugs
• Range -50 to +150°C
• Display of temperature in Celsius
• Low battery detection
• Auto-off capability
FOR HI 93503 ONLY
Calibration key, -18.0°C ±0.4°C
Calibration key, 0.0°C ±0.4°C
Calibration key, +70.0°C ±0.4°C
HI 765-18C
HI 765000C
HI 765070C
TEMPERATURE PROBES
Range
Resolution
Accuracy
(@20°C/68°F)
Typical EMC
Deviation
Battery
Environment
Dimensions
Weight
SPECIFICATIONS
-50.0 to 150.0°C
0.1°C
±0.4°C (-20 to 80°C)
±0.3% of Full Scale (outside)
for one year, excluding probe error
±0.4°C
with HI 765PW probe
9 Volt battery, 1000 hours of continuous use
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
143 x 80 x 38mm (5.6 x 3.2 x 1.5")
320 g (11.3 oz.)
6
Hanna offers a wide range of probes to meet your needs with air,
liquid, penetration and surface measurements.
These probes use highly sensitive thermistor sensors which provide
greater accuracy, faster response and wider temperature range than
conventional probes.
All Hanna temperature probes are supplied pre-calibrated from the
factory and are ready to be used with your meter. The probes are
easily connected to your meter with a standard connector.
Completely interchangeable, these probes make it possible for you to
switch from one probe to another without wasting time and money by
going through tedious calibration procedures.
The Hanna probes are also available in different handle colors to
avoid cross contamination during testing:
FOR HI 93501 AND HI 93510 ONLY
HI 762A
HI 762A/10
HI 762BL
HI 762BL/10
HI 762L
HI 762L/2
Air probe, with 1 m (3.3') cable and white
handle
Air probe, with 10 m (33') cable and white
handle
General purpose liquid probe, with 1 m (3.3')
cable and black handle
General purpose liquid probe, with 10 m (33')
cable and black handle
General purpose liquid probe, with 1 m (3.3')
cable and white handle
General purpose liquid probe, with 2 m (6.6')
cable and white handle
11
DISPLAY CODE GUIDE
The following error messages, low battery detection warning and selfexplanatory symbols are displayed to help the user:
DISPLAY
DESCRIPTION
Er
The probe is not connected to the instrument;
the probe cable might be damaged: change the probe;
the reading is out of range.
Eb Weak battery: replace the battery.
HI 93510 only:
°C
The readings are in Centigrade.
°F
The readings are in Fahrenheit.
M
Hold function is activated. press HOLD to return to current
measurement
HI
Highest temperature is being displayed
LO Lowest temperature is being displayed
SPECIFICATIONS OF HI 93510
Main features:
• Interchangeable HI 762 probes
• Interchangeable HI 762 calibration test plugs
• Range -50 to +150°C or -58 to 302°F
• Display of temperature in Celsius and Fahrenheit
• Hold function
• HI and LO (Max and Min)
memory recall function
ins rum n s
• Low battery detection
• Auto-off capability
CLEAR
°C
ON
OFF
HI
HOLD
LO
°C
°F
CALIBRATION
All Hanna thermometers have been accurately pre-calibrated at the
factory. It is generally recommended to have all thermometers
recalibrated at least once a year. For an accurate recalibration,
contact your nearest Hanna Customer Service Center.
FOR HI 93501, HI 93503 AND HI 93510 ONLY
Range
Hanna Test Plugs provide a quick and easy way to check the meter's
accuracy by simply connecting these Test Plugs to the probe connector
of the meter. If the reading differs by more than ±0.4°C (±0.8°F)
from the Test Plug value, the unit is due for recalibration.
Choose the right Test Plug to suit your needs:
Resolution
Accuracy
(@20°C/68°F)
FOR HI 93501 AND HI 93510 ONLY
HI 762-18C
HI 762000C
HI 762070C
HI 762-004F
HI 762032F
HI 762158F
Calibration key, -18.0°C ±0.4°C
Calibration key, 0.0°C ±0.4°C
Calibration key, +70.0°C ±0.4°C
Calibration key, -0.4°F ±0.8°F
Calibration key, +32.0°F ±0.8°F
Calibration key, +158.0°F ±0.8°F
10
Typical EMC
Deviation
Battery
Environment
Dimensions
Weight
SPECIFICATIONS
-50.0 to 150.0°C
-58.0 to 199°F/200 to 302°F
0.1°C / 0.1°F / 1°F
±0.4°C (-20 to 80°C)
±0.8°F (-4 to 176°F)
±0.7% of Full Scale (outside)
for one year, excluding probe error
±0.5°C / ±1°F
with HI 762BL probe
9 Volt battery, 500 hours of continuous use
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
143 x 80 x 38mm (5.6 x 3.2 x 1.5")
320 g (11.3 oz.)
7
OPERATIONAL GUIDE
Battery
Compartment
INITIAL PREPARATION
Remove the battery cover located on the rear of
your thermometer as shown here.
Remove the 9V battery. Unwrap the battery
and connect it to the battery clips. Replace the
battery and the cover.
Release the key and the meter will return to normal operational
mode.
Press and hold the LO key and the instrument
LO
will display the lowest temperature reached
together with the "LO" symbol.
9 VOLT BATTERY
CLEAR HI AND LO MEMORY (HI 93510 ONLY)
SLIDE OFF
To switch on, press the ON/OFF key.
Your thermometer will carry out a self diagnostic test routine and the LCD will light up all
segments.
The thermometer will then revert to the measurement mode.
If a temperature probe is plugged in, the unit
will display the temperature. If a probe is not
plugged in, the display will show "Er".
To switch your thermometer off, press the ON/OFF key again.
°C / °F SWITCH (HI 93510 ONLY)
Measurements can be performed in either Centigrade or Fahrenheit. To change the scale,
press °C/°F. When the meter is turned on, it
automatically defaults to the °C scale.
HOLD FUNCTION (HI 93510 ONLY)
The hold function is activated by the HOLD
key. The measured temperature is frozen on
the display until as HOLD is pressed again.
The blinking symbol "M" will appear on the
display when the HOLD is in function to alert
the user.
HI and LO stored values can be erased by
pressing the CLEAR key. The meter will restart
memorizing the maximum and minimum values automatically.
CLEAR
LOW BATTERY DETECTION AND BATTERY REPLACEMENT
When the battery is low the user is
warned by the symbol "Eb" on the display.
Battery replacement must only take place in a non-hazardous area
using an alkaline 9V battery.
Remove the battery compartment cover and replace the 9V battery
with a new one. Make sure the battery is properly connected before
replacing the cover.
AUTO SHUT-OFF
To save battery life, you can have your thermometer switch itself off
after 7 minutes. A switch is located in the battery compartment for
this feature. Remove the battery cover and slide the switch across to
the right to disable the thermometer's auto-off feature. If the switch
is moved to the left, the meter will shut off automatically after 7
minutes of non-use.
HI AND LO MEMORY (HI 93510 ONLY)
HI and LO memory recall keys make it possible to check and recall the
maximum and minimum temperature measured since the meter was
last switched on or CLEAR was pressed.
Once the HI key is pressed, the instrument will
HI
automatically display the highest temperature
reached during the measurement together with
the "HI" symbol.
8
AUTO-OFF SWITCH
9 VOLT BATTERY
BATTERY
BATTERY COVER
9