Digital Phosphor Oscilloscopes and Digital Serial

Digital Phosphor Oscilloscopes
and Digital Serial Analyzers
DPO/DSA70000 Series
Features & Benefits
On all four channels simultaneously
– 20, 16, 12.5, 8, 6, and 4 GHz
bandwidth models
– Up to 50 GS/s real-time
sample rate
– Up to 200 megasamples record
length with MultiView Zoom™
feature for quick navigation
– Fastest waveform capture
rate with >300,000 wfms/s
maximum per channel
Digital serial analyzer models
with dedicated configuration for
high speed serial design and
compliance testing
Enhanced bandwidth to the
probe tip extended to support
multiple bandwidth steps for
advanced signal integrity
Unmatched Performance for Greater Insight Into Your Design to
Get Your Work Done Faster
The DPO70000 and the DSA70000
Series are the new generation of realtime digital phosphor oscilloscopes and
are the industry’s best solution to the
challenging signal integrity issues faced
by designers verifying, characterizing,
debugging and testing sophisticated
electronic designs. The specialized
DSA70000 Series provides a complete
and dedicated solution to address the
challenges of high speed serial designs.
The family features exceptional performance in signal acquisition and analysis,
operational simplicity and unmatched
debugging tools to accelerate your
day-to-day tasks. The largest screen
in the industry and the intuitive user
interface provide easy access to the
maximum amount of information.
also be switched off to take advantage of true
analog bandwidth for applications needed the
highest available raw data capture
Bandwidth Enhance to the probe tip, extended
to support bandwidth steps, gives you an
oscilloscope with bandwidth adjustable to
capture transitions accurately without excess
frequencies and noise
High sample rate on all models, on all channels, to capture more signal details (transients,
imperfections, fast edges)
50 GS/s on all four channels for the 12.5, 16
and 20 GHz models
25 GS/s on all four channels for the 4, 6 and
8 GHz models
Lowest jitter noise floor and vertical accuracy
for very accurate measurements
Unique serial pattern triggering
up to 3.125 Gb/s and 8 b/10 b
standard protocol triggering for
isolation of pattern-dependent
effects and NRZ serial test
pattern triggering up to 6.25 Gb/s
Serial data analysis and compliance for PCI express, serial ATA,
FB-DIMM, SAS, fibre channel,
IEEE1394b, RapidIO, XAUI,
HDMI, DVI, Ethernet, USB 2.0
Most popular jitter timing and eye
diagram analysis package
DDR memory bus analysis
12.1" largest XGA touch screen
display in the industry
Event search and mark to facilitate the comprehension of event
relationships
MyScope® custom windows
and right mouse click menus
for exceptional efficiency
OpenChoice® software with
Microsoft Windows XP OS
enables built-in networking
and extended analysis
Applications
Unmatched Acquisition
Performance
Signal integrity, jitter and
timing analysis
Verification, debug and characterization of sophisticated designs
Signal Fidelity of Tektronix
Oscilloscopes Ensures Confidence
in Your Measurement Results
High bandwidth up to 20 GHz matched across
two, three or four channels and enabled by
Tektronix proprietary DSP enhancement. The
user-selectable DSP filter on each channel
provides magnitude and phase correction
plus bandwidth extension to 20 GHz for more
accurate representation of extremely fast
signals. The DSP filter on each channel can
Pinpoint® triggering, with over
1400 combinations to address
virtually any triggering situation
Debugging and compliance testing
of serial data streams for telecom
and datacom industry standards
Investigation of transient
phenomena
Spectral analysis
User selectable bandwidth limiting choices.
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Zoom in on four areas of interest
simultaneously to compare them.
Maximize the probability of capturing
elusive glitches and other infrequent events
with FastAcq acquisition mode.
Isolate glitches down to 100 ps wide.
Accelerate the Debug of
Complex Electrical Designs
FastAcq Acquisition Mode
Expedites Debugging by Clearly
Showing Imperfections
Highest performance P7500 TriMode™ probes.
Longest acquisition of the industry to provide
more resolution and longer time sequence
– Standard 10 M samples per channel on
the DPO70000 Series and 20 M on the
DSA70000 Series
– Optional up to 100 M samples on all four
channels for the 4, 6 and 8 GHz models
– Optional up to 200 M samples on all four
channels for the 12.5, 16 and 20 GHz
models
– Easily manage this deep record length,
provide detailed comparison and analysis
of multiple waveform segments with the
MultiView Zoom™ feature. Automatically
scroll through deep records visually, or
create a math expression to instantly highlight differences
Highest performance probing solutions with
bandwidth enhanced to the probe tip for
differential and single-ended voltage signals,
because accurate design verification
depends on high bandwidth access to critical
signals and high-fidelity signal capture
2
More than just color-grading, FastAcq’s
proprietary DPX® acquisition technology
captures signals at more than 300,000
waveforms per second on all four
channels simultaneously, dramatically
increasing the probability of discovering
infrequent fault events. And with a simple
turn of the intensity knob you can clearly
“see a world others don’t see,” displaying
the complete picture of your circuit’s
operation. Some oscilloscope vendors
claim high waveform capture rates for
short bursts of time, but only Tektronix
oscilloscopes, enabled by DPX technology, can deliver these fast waveform
capture rates on a sustained basis —
saving minutes, hours or even days by
quickly revealing the nature of faults so
sophisticated trigger modes can be
applied to isolate them.
The Ability to Trigger an
Oscilloscope on Events of Interest
is Paramount in Complex Signal
Debug and Validation
Whether you’re trying to find a system
error or need to isolate a section of a
complex signal for further analysis, like
a DDR read or write burst, Tektronix’
Pinpoint® triggering provides the solution. The Pinpoint trigger system uses
Silicon Germanium (SiGe) technology to
provide very high trigger sensitivity with
DPO/DSA70000 • www.tektronix.com/oscilloscopes
Isolate only the valid glitches.
very low trigger jitter and ability to capture
very narrow glitches. Pinpoint triggering
allows selection of virtually all trigger
types on both A and B trigger events.
Other trigger systems offer multiple
trigger types only on a single event (A
event), with delayed trigger (B event)
selection limited to edge type triggering
and often do not provide a way to reset
the trigger sequence if the B event
doesn’t occur. But Pinpoint triggering
provides the full suite of advanced
trigger types on both A and B triggers,
logic qualification to control when to
look for these events, and reset triggering to begin the trigger sequence
again after a specified time, state or
transition so that even events in the
most complex signals can be captured.
Other oscilloscopes typically offer less
than 20 trigger combinations; Pinpoint
triggering offers over 1400 combinations, all at full performance.
With Enhanced Triggering, you can
choose to compensate for the difference in time there is between the
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
NRZ Pattern Lock Triggering of a 640-bit
long serial test pattern SATA 6 Gb/s.
trigger path and the display path and
eliminate virtually any trigger jitter at the
trigger point. In this mode, the trigger
point can be used as a measurement
reference.
Protocol and Serial Pattern
Triggering
To debug serial architectures, use the
serial pattern triggering for NRZ serial
data stream with built-in clock recovery
and correlate events across physical
and link layer. This feature comes standard on the DSA70000 series and is
available on DPO70000 models as
Opt. PTH. The instrument can recover
the clock signal, identify transitions,
and. allow you to set the desired
encoded words for the serial pattern
trigger to capture. Opt. PTH and the
DSA70000 Series cover serial standards up to 3.125 Gb/s.
Pattern Lock Triggering adds a new
dimension to NRZ serial pattern triggering by enabling the oscilloscope to
take synchronized acquisitions of a long
serial test pattern with outstanding time
base accuracy. Pattern lock triggering
can be used to remove random jitter
from long serial data patterns. Effects of
specific bit transitions can be investigated, and averaging can be used with
mask testing. This feature supports up
to 6.25 Gb/s NRZ serial data stream
and is standard on the DSA70000
instruments, or included as part of
Option PTH on the DPO70000 models.
How does a 12.1" display compare to the
display size of other oscilloscopes?
Three modes of operation of the horizontal
time base.
Large 12.1" XGA Display Screen
Unmatched Versatility
The DPO/DSA70000 Series have the
largest display in the industry with a
12.1" XGA touch screen that gives up
to 15% more waveform display than
other oscilloscopes of their classes.
Get the Most of Your Oscilloscope
by Fully Controlling its Waveform
Acquisition and Display
Parameters
10 vertical divisions give you 25% more
vertical measurement resolution than
other oscilloscopes.
Unmatched Usability
The TekConnect™ probe interface
provides versatility and ease of use
enabled by intelligent bi-directional
oscilloscope-to-probe communication
and maintain signal fidelity.
The DPO/DSA70000 Series instruments contain a comprehensive suite
of features, such as a touch-screen,
shallow menu structures, intuitive
graphical icons, knob-per-channel
vertical controls, support for right
mouse clicks, mouse wheel operation
and intuitive Export/Save/Recall menus.
You have the choice of three horizontal
time base modes of operations. If you
are simply doing signal exploration and
want to interact with a lively signal, you
will use the Automatic or interactive
default mode that provides you with the
liveliest display update rate. If you want
a precise measurement and the highest
real-time sample rate that will give you
the most measurement accuracy, then
the Constant Sample Rate mode is for
you. It will maintain the highest sample
rate and provide the best real-time
resolution. The last mode is called the
Manual mode because it ensures direct
and independent control of the sample
rate and record length.
Interoperability with Logic Analyzers
for Digital Design and Debug
Tektronix’ Integrated View (iView™) data
display enables digital designers to solve
signal integrity challenges and effectively
debug and verify their systems more
quickly and easily. This integration allows
designers to view time-correlated digital
and analog data in the same display
window, and isolate the analog characteristics of the digital signals that are
causing systems failures. No user calibration is required. And, once set up,
the iView feature is completely automated.
DPO/DSA70000 • www.tektronix.com/oscilloscopes
3
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Drag and drop menu items of interest to
create the MyScope control window.
Capture data into Microsoft Excel using the
unique Excel toolbar, and create custom
reports using the Word toolbar.
With the MyScope® Feature, Create
Your Own Control Windows With
Only the Controls, Features, and
Capabilities that You Care About
With OpenChoice® Software,
Customize Your Test and
Measurement System with
Familiar Analysis Tools
Easily create your own personalized
“toolbox” of oscilloscope features in a
matter of minutes using a simple,
visual, drag and drop process. Once
created, these custom control windows
are easily accessed through a dedicated MyScope button and menu
selection on the oscilloscope
button/menu bar, just like any other
control window. You can make an
unlimited number of custom control
windows, enabling each person who
uses the oscilloscope in a shared environment to have their own unique
control window. MyScope control
windows will benefit all oscilloscope
users, eliminating the ramp-up time that
many face when returning to the lab
after not using an oscilloscope for a
while, and enables the power user to
be far more efficient. Everything you
need is found in one control window
rather than having to constantly navigate through menu after menu to
repeat similar tasks.
The analysis and networking features of
OpenChoice software add more flexibility to Tektronix’ Windows XP oscilloscopes: Using the fast embedded bus,
waveform data can be moved directly
from acquisition to analysis applications
on the Windows desktop at much
faster speeds than conventional GPIB
transfers. Tektronix’ implementation of
industry-standard protocols, such as
TekVISA™ interface and ActiveX
controls, are included for using and
enhancing Windows applications for
data analysis and documentation. IVI
instrument drivers are included to enable
easy communication with the oscilloscope using GPIB, serial data, and LAN
connections from programs running on
the instrument or an external PC. Or,
use the Software Developer’s Kit (SDK)
to help create custom software to automate multi-step processes in waveform
collection and analysis with Visual
BASIC, C, C++, MATLAB, LabVIEW,
4
DPO/DSA70000 • www.tektronix.com/oscilloscopes
LabWindows/CVI and other common
Application Development Environments
(ADE). Integration of the oscilloscope
with external PCs and non-Windows
hosts is also supported. In addition, the
OpenChoice architecture provides a
comprehensive software infrastructure
for faster, more versatile operations.
Data transfer programs, such as the
Excel or Word toolbar are used to simplify
analysis and documentation on the
Windows desktop or on an external PC.
More Insight into Your
Complex Electrical Design
for Characterization and
Compliance Testing
Such as a simple math expression,
waveform mask testing, a pass/fail
compliance test, event searching, event
marking or a custom application that you
develop yourself, the DPO/DSA70000
Series offers the industry’s most
comprehensive set of analysis and
compliance tools.
A Wide Range of Built-in Advanced
Waveform Analysis Tools
Waveform cursors make it easy to
measure trace-to-trace timing characteristics, while cursors that link between
YT and XY display modes make it easy
to investigate phase relationships and
Safe Operating Area violations. Select
from 53 automatic measurements
using a graphical palette that logically
organizes measurements into
Amplitude, Time, Combination,
Histogram, and Communications categories. Gather further insight into your
measurement results with statistical
data such as mean, min, max, standard deviation, and population.
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Basic spectral UI control window.
Define and apply math expressions to
waveform data for on-screen results in
terms that you can use. Access
common waveform math functions with
the touch of a button. Or, for advanced
applications, create algebraic expressions consisting of live waveforms,
reference waveforms, math functions,
measurement values, scalars and user
adjustable variables with an easy-touse calculator-style editor.
FFT – To analyze your signal in the
spectral domain, use the basic spectral
(provides you with the best parameter),
or use advanced spectral with the
manual time base horizontal mode (to
directly control the frequency span,
center frequency and resolution bandwidth).
Filtering – Enhance your ability to isolate
or remove some important component
of your signal (noise or specific harmonics
of the signal) by creating your own filters,
or using the filters provided as standard
with the instrument. These customizable
FIR filters can be used to implement
today’s preferred signal-filtering techniques, including to remove the preemphasis or to minimize the effects of
fixtures and cables connected to the
device under test.
Jitter, Timing and Eye Diagram Analysis.
A Breadth of Tools to Extend
Waveform Analysis Even Further
Jitter, Timing and Eye Diagram
Analysis (Opt. DJA) – Tight timing
margins associated with today’s serial
buses demand stable, low-jitter
designs. DPOJET extends the oscilloscope capability by making jitter, timing
and eye diagram measurements over
contiguous clock and data cycles in a
single-shot real-time acquisition. With
multiple measurements and a variety of
analysis tools including spectral and
trend plots, DPOJET quickly shows
system timing under variable conditions. It also provides Rj/Dj on signals
without a repeating pattern and without
requiring a fixed pattern or length. You
can get insight into the signal characteristics like SSC profile using the
analysis features and perform pass-fail
testing using eye diagram masks and
limit files for testing against statistical
limits using the compliance features.
Accelerating the research of specific events
in an acquired waveform.
support for more advanced event types
like transition, setup and hold or logic
pattern, is provided with the ASM
option on the DPO70000 Series, standard on the DSA70000.
Waveform Limit Testing (Opt. LT) –
This feature consists of comparing an
acquired waveform to boundaries.
These boundaries are typically defined
by the user to specify a tolerance band
around a reference waveform. If any
part of the acquired waveform falls
outside the limit, the software returns a
failure message and the location of the
failure on the waveform.
This tool is available for the DPO70000
and DSA70000 Series as Opt. DJA.
Advanced Event Search and Mark
(Opt. ASM) – Event Search and Mark
will relieve the user from the tedious
task of examining data by highlighting
important events, skipping the unimportant ones, and enhancing the
comprehension of event relationships.
You can navigate between the events
of interest effortlessly. A basic event
(edge-only) search and mark is
provided as a standard feature; and
DPO/DSA70000 • www.tektronix.com/oscilloscopes
5
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Test eye diagram in equivalent time against
the standard mask.
Communications Mask Testing – This
feature provides a complete portfolio of
masks for verifying compliance to serial
communications standards. It supports
156 Standards Masks.
ITU-T (1.544 Mb/s to 155 Mb/s)
ANSI T1.102 (1.544 Mb/s to 155 Mb/s)
Ethernet IEEE 802.3, ANSI X3.263
(1.544 Mb/s to 3.125 Gb/s XAUI)
Sonet/SDH (51.84 Mb/s to 2.4883 Gb/s)
Fibre Channel (133 Mb/s to 4.25 Gb/s*1)
InfiniBand (2.5 Gb/s)
USB (12 Mb/s to 480 Mb/s)
Serial ATA (1.5 Gb/s, 3 Gb/s)
Serial Attached SCSI (1.5 Gb/s, 3 Gb/s)
IEEE 1394b (491.5 Mb/s to 1.966 Gb/s)
Rapid I/O (1.25 Gb/s to 3.125 Gb/s)
OIF Standards (2.488 Gb/s to 3.11 Gb/s)
PCI Express (2.5 Gb/s)
RT-Eye® version 2.0 - PCI Express Rev2
compliance test.
Serial Data Compliance and Analysis –
Patented Real-time Eye (RT-Eye® clock
recovery and eye-rendering) provides
high-speed serial data domain expertise
to enable analysis and compliance
measurements for testing high-speed
serial standards like PCI Express, Serial
ATA, SAS, InfiniBand, FB-DIMM, as well
as Front Side Bus (FSB), XAUI, Fibre
Channel, IEEE 1394b, and RapidIO. It
recovers the clock of the serial stream
to ≥10 Gb/s and generates very high
precision eye diagrams with an accumulated waveform database. Serial
data compliance and analysis comes
standard on the DSA70000 Series, and
optional on the DPO70404, DPO70604
and DPO70804 as Opt. RTE. The
compliance modules for PCI Express,
Serial ATA, SAS, InfiniBand, and FBDIMM are options on both DSA70000
Series and DPO70000 Series (Opt.
PCE, SST, IBA, or FBD).
*1 A 4.25 Gb/s mask supported using Glitch Trigger. It is standard on the DSA70000 Series, and optional as Opt. MTH on
DPO70404, DPO70604 and DPO70804.
6
DPO/DSA70000 • www.tektronix.com/oscilloscopes
HDMI compliance testing.
Optional HDMI Compliance Testing
(Opt. HT3) – Compliance testing: This
is your complete solution for HDMI
compliance testing, enabling unprecedented efficiency by offering a complete
solution of unmatched reliable automation to support the widest range of
tests in the industry.
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
UWB WiMedia analysis and measurements.
Power measurements and analysis.
USB compliance testing.
Ethernet compliance testing.
Identify and separate all DDR read from
write bursts.
Optional Ultra-Wideband Spectral
Analysis and Ultra-Wideband
Spectral Analysis Essentials
UWBE – Ultra-Wideband microwave,
optical and electrical signals require
more real-time bandwidth than is
possible with spectrum analyzer based
solutions. Spectral Analysis and Digital
Down Conversion of RF data is fast
and easy and the down converted
frequency span of interest may be
exported for further analysis in tools
such as RSAVu and MATLAB.
UWB in addition adds – With automatic packet, TFC and data rate
detection, support for all band groups,
Time Frequency Codes and data rates,
WiMedia PHY 1.2 analysis provides a
complete solution. Rapid visualization,
debug and report generation of the
Spectrograms, Power Spectral Density,
QPSK/DCM Constellations, EVM-vsSymbol, EVM-vs-Subcarrier, CommonPhase-Error-vs-Symbol, and Voltagevs-Time plots and complete measurements are captured and documented
for each test condition.
Optional Power Measurement and
Analysis (Opt. PWR) – Analyze power
dissipation in power supply switching
devices and magnetic components,
and generate detailed reports in
customizable formats. The HiRes
acquisition mode delivers greater than
8 bits of vertical resolution on singleshot or repetitive signals at bandwidth
up to 125 MHz. The powerful and
flexible measurements, math, and
math-on-math capabilities make it an
ideal solution for performing power
measurements, such as voltage,
current, instantaneous power and
energy, for power device designers.
Optional Ethernet Compliance
Testing (Opt. ET3) – Provides compliance testing for 10/100/1000Base-T
signals.
Optional DVI Compliance Testing
(Opt. DVI) – Provides Digital Visual
Interface physical layer validation and
compliance testing with automated eye
diagram generation and parametric
testing.
Optional USB Compliance Testing
(Opt. USB) – Provides compliance
testing for USB2.0 signals.
Optional DDR Memory System
Analysis (Opt. DDRA) – Accelerate the
validation of a memory system based
on DDR2, DDR3 or DDR variants technology, like GDDR3. This new DDR
search algorithm automatically detects
the rates and the voltage levels of the
data and strobe signals and marks every
occurrence of read or write bursts. You
can then generate an eye-diagram of
the data or perform JEDEC standard
measurements qualified on read or
write bursts with DPOJET.
DPO/DSA70000 • www.tektronix.com/oscilloscopes
7
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
DPO/DSA70000 Series
9
5
9
11
4
7
6
Protocol Trigger.
Spread Spectrum Clock profiling.
10
1
5
Protocol Trigger and Decode.
User-selectable bandwidth limit for optimal
performance.
3
8
User-definable FIR filters for advanced
analysis.
Compliance testing for many serial standards like HDMI, XAUI, PCI-Express and
FB-DIMM.
1 Large 12.1-inch XGA
Touch Screen Display
The DPO70000 and the
DSA70000 Series touch
screen gives up to 15%
more waveform display
than any other oscilloscopes of its class.
8
2 TekConnect®
Probe Interface
It provides performance,
versatility and ease of use.
It enables bi-directional
communication between
the oscilloscope and the
probe and bandwidth
enhancement down to
the probe tip.
10
3 Exceptional
Performance
A very lively instrument
and the performance of
the highest oscilloscope
with 50 GS/s real-time
sample rate and 200 M
record length on all four
channels simultaneously.
DPO/DSA70000 • www.tektronix.com/oscilloscopes
4 With MultiView
Zoom™
Easily deep into very long
record of acquired data,
analyze multiple waveform segments simultaneously and scroll automatically through the
deepest records visually.
2
5 Unmatched Usability
With MyScope,
create your own control
window with only the
controls you care about.
The versatile graphical
user interface allows you
to use the touch screen
or the mouse.
®
6 Accelerate the Debug
of Complex Designs
with Pinpoint™
Triggering
7 FastAcq Acquisition
Expedites Debugging
by Clearly Showing
Faults
With the industry’s
largest trigger combinations, you can address
virtually any triggering
situation. Take synchronized snapshot of test
patterns up to 6.25 Gb/s.
More than 300,000
waveforms per second, and
with a simple turn of the
intensity knob, you can
see the frequency of
occurrence.
8 Easy Connectivity
Built-in USB port at the
front for easy access to
easily save your work on
a memory stick. Front
access to recovered
clock and data.
9 A Wide Range of
Built-in Advanced
Analysis Tools
Four types of cursors.
53 automatic
measurements.
Waveform boundary
tolerance testing.
Many advanced math
functions like Filtering
and Spectral.
10 For Insight into
Your High-speed
Serial Designs
40 bit 8 b/10 b Serial
Pattern triggering with
clock recovery plus
industry-unique protocol
decode and triggering
to capture problems
at the link or at the
physical layers.
11 A Breadth of Optional
Software Packages
for Expanded Waveform Analysis and
Compliance Testing
Accelerate the research
of specific events in an
acquired waveform.
Advanced jitter and
timing measurements.
Application-specific
measurements and
compliance testing.
DPO/DSA70000 • www.tektronix.com/oscilloscopes
9
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
DSA70000 Series
For Developing with Today’s
High-Speed Serial Standards,
the DSA70000 Digital Serial
Analyzer is Your Uncompromised
High-Performance, Dedicated
Solution to Efficiently Address
Your Design Challenges
The DSA70000 Series is a new generation of real-time digital serial analyzers
based on the same advanced technology
as the DPO70000 real-time digital
phosphor oscilloscopes. As high-speed
serial technology becomes more pervasive, more designers are looking for
easy to use, complete, and dedicated
solutions for verifying, characterizing,
debugging and testing sophisticated
high-speed serial designs. The DSA70000
Series is specifically targeted to address
the challenging high-speed serial design
issues faced by designers, by encapsulating extended high-speed serial data
domain expertise. It inherits exceptional
signal acquisition performance, operational simplicity, and unmatched debugging tools from the DPO70000 Series,
to accelerate your day-to-day tasks. It
also features the extended analysis tools
that enable high-speed serial signal
analysis and compliance measurements in a specialized instrument.
The DSA70000 Series Analyzers
provides the signal fidelity of
Tektronix oscilloscopes to ensure
confidence in your measurement
results: high sample rate on all models,
on all channels, to capture more signal
details (transients, imperfections, fast
edges), 25 GS/s on all four channels for
the 4, 6 and 8 GHz models, 50 GS/s
on all four channels for the 12.5, 16
and 20 GHz models, bandwidth
enhancement as well as best low jitter
noise floor and vertical accuracy for
very accurate measurements.
The DSA70000 Series provides the
longest acquisition of the industry to
provide more resolution and longer time
10
sequence—a standard 20 M on the DSA
Series, or an optional up to 100 M
samples on all four channels for the 4,
6 and 8 GHz models, 200 M samples
on all four channels for the 12.5, 16
and 20 GHz models. Easily manage
this deep record length and provide
detailed comparison and analysis of
multiple waveform segments with the
MultiView Zoom™ feature.
The DSA70000 analyzers share the
DPX technology of the DPO70000 and
can deliver high waveform capture rate
at more than 300,000 waveforms per
second. The DSA70000 Series capture
these intermittent fault events that can
break a design with the FastAcq
acquisition mode. With Pinpoint®
triggering, the DSA70000 series is also
equipped to isolate a section of a
complex signal for further analysis.
To debug serial architectures, the
DSA70000 Series features the NRZ
serial pattern triggering and protocol
decode with built-in clock recovery. It
recovers the clock signal, identifies the
transitions and decodes characters and
other protocol data. You can see the
captured bit sequences decoded into
their words for convenient analysis (for
8 b/10 b and other encoded serial data
streams), or you can set the desired
encoded words for the serial pattern
trigger to capture. Lastly, you can
synchronize long serial test pattern
acquisitions up to 6.25 Gb/s to remove
random jitter. The DSA70000 Series
covers serial standards up to 3.125 Gb/s.
The DSA70000 Series features the
highest accuracy jitter and timing
measurements as well as comprehensive analysis algorithms. Tight
timing margins demand stable, lowjitter designs. You can make jitter
measurements over contiguous clock
cycles from every valid pulse in a
single-shot acquisition. Multiple measurements and trend plots quickly show
system timing under variable condi-
DPO/DSA70000 • www.tektronix.com/oscilloscopes
tions. It also includes Random Jitter
and Deterministic Jitter separation as
well as Total Jitter measurement at Bit
Error Ratio to 10-18.
Communications Mask Testing
provides a complete portfolio of masks
for verifying compliance to serial
communications standards. It supports
156 Standards Masks - ITU-T (1.544 Mb/s
to 155 Mb/s)/ANSI T1.102 (1.544 Mb/s
to 155 Mb/s); Ethernet IEEE 802.3;
ANSI X3.263 (1.544 Mb/s to 3.125 Gb/s
XAUI); Sonet/SDH (51.84 Mb/s to
2.4883 Gb/s); Fibre Channel (133 Mb/s
to 4.25 Gb/s*1). InfiniBand (2.5 Gb/s);
USB (12 Mb/s to 480 Mb/s); Serial ATA
(1.5 Gb/s, 3 Gb/s); Serial Attached
SCSI (1.5 Gb/s, 3.0 Gb/s); IEEE 1394b
(491.5 Mb/s to 1.966 Gb/s); Rapid I/O
(1.25 Gb/s to 3.125 Gb/s); OIF
Standards (2.488 Gb/s to 3.11 Gb/s);
PCI Express (2.5 Gb/s).
Accurate, Simple and Customizable
Physical Layer Testing on High
Speed Serial Standards. When
designing to industry standards, analog
validation and compliance testing (Front
Side Bus, PCI Express, FB-DIMM,
Serial ATA, Serial Attached SCSI, Fibre
Channel, XAUI, IEEE1394b, RapidIO) is
critical to ensure device interoperability.
Patented Real-Time (RT-Eye®) clock
recovery and Eye Rendering provides
standard specific clock recovery, high
precision eye diagrams for transition
and non-transition bits and accurate
jitter measurements, and de-emphasis
measurements. Standard-specific
compliance and analysis modules that
configure the pass/fail waveform mask
and measurement limit testing are also
available as an option for PCI Express
(Option PCE), for Serial ATA and SAS
(Option SST), for FB-DIMM (Fully
Buffered - Dual Inline Memory Module)
(Option FBD), or InfiniBand (Option IBA).
*1 A 4.25 Gb/s mask supported using Glitch Trigger. It is standard on the DSA70000 Series, and optional as Opt. MTH on
DPO70404, DPO70604 and DPO70804.
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Characteristics
Vertical System
DPO/DSA Models
Input Channels
Bandwidth
(user selectable DSP enhance)
Rise Time 10% to 90% (typical)
Rise Time 20% to 80% (typical)
Hardware Analog Bandwidth (–3 dB)
DC Gain Accuracy
Bandwidth Limits
70404
70604
70804
71254
71604
72004
4
4 GHz
4
6 GHz
4
8 GHz
4
12.5 GHz
4
16 GHz
93 ps
65 ps
4 GHz
62 ps
43 ps
6 GHz
47 ps
33 ps
8 GHz
4
2 settings: 20 GHz
and 18 GHz
22.5 ps
17 ps
16 GHz (typical)
34.3 ps
27.5 ps
23 ps
21 ps
12.5 GHz
16 GHz (typical)
±2% (of reading)
Depending on instrument model: 19 GHz, 18 GHz, 17 GHz, 16 GHz, 15 GHz, 14 GHz, 13 GHz, 12 GHz, 11 GHz, 10 GHz,
9 GHz, 8 GHz, 7 GHz, 6 GHz, 5 GHz, 4 GHz, 3 GHz, 2 GHz, 1 GHz or 500 MHz
DC (50 Ω), GND
50 Ω ±1.5%, 1 MΩ with TCA–1MEG adapter
Input Coupling
Input Impedance
Input Sensitivity
18 GHz and below
20 GHz and 19 GHz
Vertical Resolution
Max Input Voltage, 50 Ω
Position Range
Offset Range
10 mV/div to 1 V/div (100 mV to 10 V full scale)
20 to 99.5 mV/div and 200 mV/div to 1 V/div
8 bit (11 bit with averaging)
< 5.5 VRMS for ≥1 V full scale; also determined by TekConnect® accessory
±5 div
10 mV/div: ±450 mV
20 mV/div: ±400 mV
50 mV/div: ±250 mV
100 mV/div: ±4.5 V
200 mV/div: ±4.0 V
500 mV/div: ±2.5 V
1.0 V/div: 0
10 mV/div – 99.5 mV/div: ± (0.35% (offset value-position) + 1.5 mV + 1% of full scale)
100 mV/div – 1 V/div : ± (0.35% (offset value-position) + 15 mV + 1% of full scale)
≤100 ps for any two channels with equal V/div and coupling settings
≤50 ps with BW enhance enabled (BW+)
≥120:1 (for input frequency 0 to 10 GHz)
≥80:1 (for input frequency >10 GHz to 12 GHz)
≥50:1 (for input frequency >12 GHz to 15 GHz)
≥25:1 (for input frequency >15 GHz)
Offset Accuracy
Delay between any two channels (typical)
Channel-to-channel Isolation
(Any Two Channels at Equal Vertical Scale Settings)
Time Base System
DPO/DSA Models
Time Base Range
Time Resolution (in ET/IT mode)
Time Base Delay Time Range
Channel-to-channel Deskew
Delta Time Measurement Accuracy
(typical) Over <100 ns duration;
single shot; with signal rise time
= 1.2X scope rise time
Trigger Jitter (RMS)
Jitter Noise Floor(typical)
(With BW+ bandwidth enhance enabled)
Time Base Accuracy
70404
70604
70804
71254
20 ps/div to 1000 s/div
200 fs
888 fs
450 fs
71604
72004
10 ps/div to 1000 s/div
100 fs
695 fs
–5.0 ks to 1.0 ks
Range ±75 ns
611 fs
504 fs
482 fs
525 fs
450 fs
1 psRMS (typical) with enhanced triggering OFF
< 100 fsRMS with enhanced triggering ON
450 fs
300 fs
300 fs
400 fs
±1.5 ppm initial accuracy, aging <1 ppm per year
DPO/DSA70000 • www.tektronix.com/oscilloscopes
11
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Acquisition System
DPO/DSA Models
70404/70604/70804
71254/71604/72004
25 GS/s
5 TS/s
50 GS/s
10 TS/s
Sample Rates
Real time mode 1, 2, 3 or 4 channel (max)
ET/IT Mode (max)
Maximum Record Length per Channel
With Standard Configuration
With Record Length Opt. 2XL
With Record Length Opt. 5XL
With Record Length Opt. 10XL
With Record Length Opt. 20XL
N/A
10 M on all four channels (DPO70000 Series only)
20 M on all four channels (DSA70000 Series only)
20 M on all four channels (DPO70000 Series only)
50 M on all four channels
100 M on all four channels
200 M on all four channels
Maximum Duration at Highest Real-time Resolution
DPO/DSA Models
Resolution
Max Duration with Standard Memory
Max Duration with Opt. 2XL
Max Duration with Opt. 5XL
Max Duration with Opt. 10XL
Max Duration with Opt. 20XL
70404/70604/70804
71254/71604/72004
40 ps (25 GS/s)
0.4 ms DPO70000 Series; 0.8 ms for DSA70000 Series
0.8 ms (DPO70000 Series only)
2.0 ms
4.0 ms
N/A
20 ps (50 GS/s)
0.2 ms DPO70000 Series; 0.4 ms for DSA70000 Series
0.4 ms (DPO70000 Series only)
1.0 ms
2.0 ms
4.0 ms
Acquisition Modes
FastAcq Acquisition Mode
Maximum FastAcq Waveform Capture Rate
Waveform Database
Sample
Peak Detect
Averaging
Envelope
Hi-Res
FastFrame™ Acquisition
Roll Mode
FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events
>300,000 wfms/s on all four channels simultaneously
Accumulate waveform database providing three-dimensional array of amplitude, time, and counts
Acquire sampled values
Captures narrow glitches at all real-time sampling rates: 1 ns at ≤125 MS/s; 1/sample rate at ≥250 MS/s
From 2 to 10,000 waveforms included in average
From 1 to 2x109 waveforms included in min-max envelope
Real-time boxcar averaging reduces random noise and increases resolution
Acquisition memory divided into segments; maximum trigger rate >310,000 waveforms per second
Time of arrival recorded with each event. Frame finder tool helps to visually identify transients
Up to 10 MS/s with a maximum record length of 40 M
Pinpoint ® Trigger System
DPO Models
70404/70604/70804/71254/71604/72004
DSA Models
70404/70604/70804/71254/71604/72004
Sensitivity
Internal DC Coupled
External (Auxiliary Input) 50 Ω
4% of full scale from DC to 50 MHz
10% of full scale at 4 GHz
20% of full scale at 8 GHz
50% of full scale at 11 GHz
250 mV from DC to 50 MHz, increasing to 350 mV at 1.0 GHz
Trigger Characteristics
A Event and Delayed B Event Trigger Types
Main Trigger Modes
Enhanced Triggering
Trigger Sequences
12
Edge, Glitch, Runt, Width, Transition Time, Timeout, Pattern, State, Setup/Hold, Window—all except
Edge, Pattern and State can be Logic State qualified by up to two channels
Auto, Normal and Single
User-selectable; it corrects the difference in timing between the trigger path and the acquired data path
(it supports all Pinpoint trigger types on both A- and B-Events except pattern trigger and not available in FastAcq)
Main, Delayed by Time, Delayed by Events, Reset by Time, Reset by State, Reset by Transition
All sequences can include separate horizontal delay after the trigger event to position the acquisition window in time
DPO/DSA70000 • www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Communications-related Triggers
Requires Opt. MTH
Standard
Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications signals
Select among isolated positive or negative one, zero pulse form or
eye patterns as applicable to the standard
Serial Pattern Trigger
Requires Opt. PTH
Standard
Up to 64 bit serial word recognizer, bits specified in binary (high, low, don’t care) or hex format.
Trigger on NRZ-encoded data up to 1.25 GBaud.
Trigger on 8 b/10 b-encoded data from 1.25 to 3.125 GBaud (40 bits)
Clock Recovery System
Clock Recovery Phase Locked Loop Bandwidth
Frequency Range
Clock Recovery Jitter (RMS)
Tracking/Acquisition Range
Minimum Signal Amplitude needed for Clock Recovery
Trigger Level Range Internal
AUX Trigger
Line
Trigger Coupling
Trigger Holdoff Range
Requires Opt. PTH or MTH
Standard
Fixed at FBaud/1600
1.5 MBaud to 3.125 GBaud
<0.25% bit period + 2 psRMS for PRBS data patterns
<0.25% bit period + 1.5 psRMS for repeating “0011” data pattern
±2% of requested baud
1 divp-p up to 1.25 Gbaud
1.5 divp-p above 1.25 Gbaud
±120% of full scale from center of screen
TekConnect® interface: ±5 V
Fixed at 0 V
DC, AC (attenuates <100 Hz),
HF Rej (attenuates >20 kHz),
LF Rej (attenuates <200 kHz),
Noise Reject (reduces sensitivity)
250 ns min to 12 s max
Trigger Modes
Edge – Positive or negative slope on any channel or
front panel auxiliary input. Coupling includes DC, AC,
noise reject, HF reject and LF reject.
Glitch – Trigger on or reject glitches of positive,
negative or either polarity. Minimum glitch width is
down to 150 ps (typical) with re-arm time of 300 ps.
Width – Trigger on width of positive or negative
pulse either within or out of selectable time limits
(down to 150 ps).
Runt – Trigger on a pulse that crosses one
threshold but fails to cross a second threshold
before crossing the first again. Event can be time- or
logic-qualified.
Timeout – Trigger on an event which remains high,
low or either, for a specified time period. Selectable
from 300 ps.
Transition – Trigger on pulse edge rates that are
faster or slower than specified. Slope may be positive, negative or either.
Setup/Hold – Trigger on violations of both setup
time and hold time between clock and data present
on any two input channels.
Pattern – Trigger when pattern goes false or stays
true for specified period of time. Pattern (AND, OR,
NAND, NOR) specified for four input channels
defined as high, low, or don’t care.
State – Any logical pattern of channels (one, two,
three) clocked by edge on channel four. Trigger on
rising or falling clock edge.
Window – Trigger on an event that enters or exits a
window defined by two user-adjustable thresholds.
Event can be time- or logic-qualified.
Trigger Delay by Time – 3.2 ns to 3 Ms.
Trigger Delay by Events – 1 to 2 G events.
Comm – Standard feature on the DSA70000,
provided as part of Opt. MTH on the DPO70000
Series. Support for AMI, HDB3, BnZS, CMI, MLT3
and NRZ encoded signals.
Serial Pattern – Trigger on NRZ-encoded data up
to 3.125 Gbaud; above 1.25 Gbaud requires 8 b/10 b
encoded data. Extended with pattern lock triggering
to capture repeated acquisitions of long serial test
patterns up to 6.25 Gb/s.
Histogram Related – Waveform Count, Hits in Box,
Peak Hits, Median, Maximum, Minimum, Peak to
Peak, Mean (µ), Standard Deviation (σ), µ+1σ,
µ+2σ, µ+3σ
Eye Pattern Related – Extinction Ratio (absolute,
%, dB), Eye Height, Eye Width, Eye Top, Eye Base,
Crossing %, Jitter (peak to peak, RMS, 6σ), Noise
(peak to peak, RMS), Signal/Noise Ratio, Cycle
Distortion, Q-Factor.
Search and Mark Events
Waveform Processing/Math
Basic – Mark any events and document waveforms.
Search positive, negative slopes or both on any
channels. Event table summarizes all found events.
All events are time stamped in reference to trigger
position. Users can choose to stop acquisitions when
an event is found.
Advanced – Search glitches or runts, as well as
transition rate, pulse width, setup and hold, timeout,
window violations or find any logic or state pattern
on any number of channels. Search DDR read or
write bursts with Opt. DDRA.
Arithmetic – Add, Subtract, Multiply, Divide
Waveforms and Scalars.
Algebraic Expressions – Define extensive algebraic
expressions including Waveforms, Scalars, Useradjustable Variables and Results of Parametric
Measurements e.g. (Integral (CH.1-Mean
(CH.1)) x1.414xVAR1).
Math Functions – Average, Invert, Integrate,
Differentiate, Square Root, Exponential, Log 10, Log e,
Abs, Ceiling, Floor, Min, Max, Sin, Cos, Tan, ASin,
ACos, ATan, Sinh, Cosh, Tanh.
Relational – Boolean result of comparison >, <, ≥,
≤, ==, !=.
Frequency Domain Functions – Spectral
Magnitude and Phase, Real and Imaginary Spectra.
Vertical Units – Magnitude: Linear, dB, dBm Phase:
Degrees, radians, group delay IRE and mV units.
Window Functions – Rectangular, Hamming,
Hanning, Kaiser-Bessel, Blackman-Harris, Gaussian,
Flattop2, Tek Exponential.
Waveform Definition – As an arbitrary math expression.
Filtering Functions – User-definable filters. Users
specify a file containing the coefficients of the filter.
Several filter files provided.
Mask Function – A function that generates a
Waveform Database pixmap from a sample waveform. Sample count can be defined.
Waveform Measurements
Automatic Measurements – 53, of which eight
can be displayed on screen at any one time; measurement statistics, user-definable reference levels,
measurement within gates isolating the specific
occurrence within an acquisition to take measurements on.
Amplitude Related – Amplitude, High, Low,
Maximum, Minimum, Peak to Peak, Mean, Cycle
Mean, RMS, Cycle RMS, Positive Overshoot,
Negative Overshoot.
Time Related – Rise Time, Fall Time, Positive Width,
Negative Width, Positive Duty Cycle, Negative Duty
Cycle, Period, Frequency, Delay.
Combination – Area, Cycle Area, Phase, Burst Width.
DPO/DSA70000 • www.tektronix.com/oscilloscopes
13
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Display Characteristics
Rear Panel
Display Type – Liquid crystal active-matrix color display.
Display Size – Diagonal: 307.3 mm (12.1 in.).
Display Resolution – XGA 1024 horizontal x 768
vertical pixels.
Waveform Styles – Vectors, Dots, Variable
Persistence, Infinite Persistence.
Color Palettes – Normal, Green, Gray, Temperature,
Spectral and User-defined.
Display Format – YT, XY.
External Time Base Reference In – BNC connector;
allows time base system to phase lock to external
10/100 MHz reference .Optimized (via a software
switch) for either a highly stable clock or tracking mode.
Time Base Reference Out – BNC connector;
provides TTL-compatible output of internal 10 MHz
reference oscillator.
AUX Trigger Output – BNC connector, 0 to 3 V;
default output is A-Event Trigger low true.
Parallel Port – IEEE 1284, DB-25 connector.
Audio Ports – Miniature phone jacks for stereo
microphone input and stereo line output.
USB2.0 Ports – Four in back. Allow connection or
disconnection of USB keyboard, mouse, or storage
device while oscilloscope power is on.
Keyboard Port – PS/2 compatible.
Mouse Port – PS/2 compatible.
LAN Port – RJ-45 connector, supports 10Base-T,
100Base-T and 1000Base-T.
Serial Port – DB-9 COM1 port.
Windows Video Port – 15 pin D-sub connector on
the rear panel; connects a second monitor to use
dual-monitor display mode allowing analysis results
and plots to be viewed along with the oscilloscope
display. Video is DDC2B compliant.
GPIB Port – IEEE 488.2 standard.
Scope XGA Video Port – 15 pin D-sub connector
on the rear panel, video is IBM XGA compatible.
Connects to show the oscilloscope display, including
live waveforms on an external monitor or projector.
The primary Windows desktop can also be displayed
on an external monitor using this port.
TekLink™ – Proprietary interface for connecting
multiple Tektronix instruments.
Power – 100 to 240 VRMS, ±10%, 50/60 Hz; 115
VRMS ±10%, <870 Watts, 400 Hz; CAT II, <1100 VA
typical.
Computer System and
Peripherals
Operating System – Windows XP.
CPU – Intel Pentium 4, 3.4 GHz processor.
PC System Memory – 2 GB.
Hard Disk Drive – Rear-panel, removable hard disk
drive, 80 GB capacity.
CD-R/W Drive – Front-panel CD-R/W drive with CD
creation software application.
DVD Drive – Read only.
Mouse – Optical wheel mouse, USB interface.
Keyboard – USB interface.
Input/Output Ports
Front Panel
Aux Trigger Input – See trigger specifications.
Recovered Clock – SMA connector, ≤1.25 Gb/s,
Output swing ≥130 mVp-p into 50 Ω at 1.25 Gb/s.
Requires Opt. PTH or Opt. MTH to enable on
DPO70000, standard on DSA70000.
Recovered Data – SMA connector, ≤1.25 Gb/s,
Output swing of 1010 repeating pattern 200 mV
into 50 Ω at 1.25 Gb/s. Requires Opt. PTH or Opt.
MTH to enable on DPO70000, standard on
DSA70000.
DC Probe Calibration Output – BNC connector,
±10 V DC for DC probe calibration. (Signal available
only during probe calibration.)
Fast Edge Output – SMA connector provides fast
edge signal. 1 kHz ±20%; 810 mV (base to top)
±20% into ≥10 kΩ load; 440 mV ±20% into a 50
Ω load.
AUX Trigger Output – BNC connector, provides a
TTL-compatible, polarity switchable pulse when the
oscilloscope triggers.
USB2.0 Port – One in front, four on back. Allows
connection or disconnection of USB keyboard,
mouse or storage device while oscilloscope is on.
Physical Characteristics
Benchtop Configuration
Dimensions
mm
in.
Height
Width
Depth
298
451
489.97
11.74
17.75
19.29
Weight
kg
lbs.
Net
Shipping
20
34
44
75
Rackmount Configuration
Dimensions
mm
Height
311
Width
480.1
Depth (from rackmounting
ear to back of instrument) 546.1
in.
12.25
18.9
21.5
Weight
kg
lbs.
Net
Kit
20
2.7
44
6
Mechanical
Cooling — Required Clearance
Top
Bottom
Left side
Right side
Front
Rear
mm
in.
0
0
76
76
0
0
0
0
3
3
0
0
Environmental
Temperature
Operating – 5 ºC to +45ºC.
Non-Operating – –20ºC to +60ºC.
Humidity
Operating – 8% to 80% relative humidity (RH) at up
to 32ºC. 5% to 45% RH above +32ºC up to +45ºC.
Non-Operating – 5% to 95% relative humidity (RH).
Upper limit derated to 45% RH above +30ºC up to
+60ºC.
Altitude
Operating – 10,000 ft. (3,048 m).
Non-Operating – 40,000 ft. (12,190 m).
Regulatory
Electromagnetic Compatibility – 93/68/EEC;
EN61326:1997 +A1 1998+A2:2000.
Certifications – UL 3111-1, CSA1010.1,
ISO11469,EN61010-1, IEC 61010-1.
14
DPO/DSA70000 • www.tektronix.com/oscilloscopes
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
DPO/DSA70000 Series
Ordering Information
DPO70404
4 GHz Digital Phosphor Oscilloscope.
DPO70604
6 GHz Digital Phosphor Oscilloscope.
DPO70804
8 GHz Digital Phosphor Oscilloscope.
DPO71254
12.5 GHz Digital Phosphor Oscilloscope.
DPO71604
16 GHz Digital Phosphor Oscilloscope.
DPO72004
20 GHz Digital Phosphor Oscilloscope.
DSA70404
4 GHz Digital Serial Analyzer.
DSA70604
6 GHz Digital Serial Analyzer.
DSA70804
8 GHz Digital Serial Analyzer.
DSA71254
12.5 GHz Digital Serial Analyzer.
Software Options for
DPO70000 Series and
DSA70000 Series
Recommended Accessories
Opt. LT – Waveform Limit Testing.
Opt DDRA*10 – DDR Memory Bus Analysis.
Opt. DJA – DPOJET Jitter and Eye Diagram Analysis.
Opt. ET3*2 – Ethernet Compliance Test Software.
Opt. USB*3 – USB2.0 Compliance Test Software only.
Opt. PWR*4 – Power Measurement and
Analysis Software.
Opt. HT3 – HDMI Compliance Test Software.
Opt. DVI – DVI Compliance Test Solution.
Opt. PCE*5 – PCI Express™ Compliance Module for
RT-Eye Serial Data Compliance and Analysis Software.
Opt. SST*5 – SATA and SAS Analysis Software
Module for RT-Eye Serial Data Compliance and
Analysis Software.
Opt. FBD*5 – FB-DIMM Compliance Module for RTEye Serial Data Compliance and Analysis Software.
Opt. IBA*5 – InfiniBand Compliance Module for RTEye Serial Data Compliance and Analysis Software.
Opt. UWBE – Ultra-Wideband Spectral
Analysis Essentials.
Opt UWB – Ultra-Wideband Spectral Analysis.
P7520 – 20 GHz TriMode™ probe.
P7516 – 16 GHz TriMode™ probe.
P7513 – 13 GHz TriMode™ probe.
P7313 – 13 GHz Z-Active™ differential probe.
P7313SMA – 13 GHz TekConnect® differential
SMA probe.
P7380A – 8 GHz Z-Active™ differential probe.
P7380SMA – 8 GHz TekConnect differential
SMA probe.
P7360A – 6 GHz Z-Active differential probe.
P7340A – 4 GHz Z-Active differential probe.
P6251 – DC to 1 GHz, 42V, Differential Probe
(requires TCA-BNC adapter).
P6250 – DC to 500 MHz, 42V, Differential Probe
(requires TCA-BNC adapter).
TCPA300/TCPA400 – Series current
measurement systems.
P5200/P5205/P5210 – High voltage
differential probes.
DSA71604
User Manual Options
16 GHz Digital Serial Analyzer.
Opt. L0 – English manual.
Opt. L1 – French manual.
Opt. L3 – German manual.
Opt. L5 – Japanese manual.
Opt. L7 – Simple Chinese manual.
Opt. L8 – Standard Chinese manual.
Opt. L9 – Korean manual.
Opt. L10 – Russian manual.
Opt. L99 – No manual.
DSA72004
20 GHz Digital Serial Analyzer.
All Models Include: Accessory pouch, front cover,
mouse, keyboard, quick start user manual
(071-173x-xx), DPO70000 Series product software
CD/DVD-ROM, DPO70000 Series operating system
restoration CD/DVD-ROM, Optional applications
software CD/DVD-ROM, performance verification
procedure PDF file, GPIB programmer’s reference
(on product software CD/DVD-ROM), calibration
certificate documenting NIST traceability, Z 540-1
compliance and ISO9001, power cord, one year
warranty.User to specify quick start user manual
language and power plug when ordering.(4)
TekConnect® to 2.92 mm adapters (TCA-292MM)
and (1) Tekconnect to BNC adapter (TCA-BNC).
Options
Instrument Options
Record Length Options for DPO70000 Series.
Opt. 2XL – 20 MSamples/ch.
Opt. 5XL – 50 MSamples/ch.
Opt. 10XL – 100 MSamples/ch.
Opt. 20XL*9 – 200 MSamples/ch.
Record Length Options for DSA70000 Series.
Opt. 5XL – 50 MSamples/ch.
Opt. 10XL – 100 MSamples/ch.
Opt. 20X*9 – 200 MSamples/ch.
Software Options for
DPO70000 Series
Opt. PTH – Protocol Triggering and Decoding for
8b/10b-encoded Serial Signals up to 3.125 Gb/s.
Includes hardware clock recovery and pattern lock
triggering.
Opt. MTH – Mask testing for Serial Standards up to
4.25 Gb/s. Includes hardware clock recovery.
Opt. ASM – Advanced Event Search and Mark.
Power Plug Options
Opt. A0 – North America.
Opt. A1 – Universal European Union.
Opt. A2 – UK.
Opt. A3 – Australia.
Opt. A5 – Switzerland.
Opt. A6 – Japan.
Opt. A10 – China.
Opt. A11 – India.
Opt. A99 – No power cord.
Service Options
Opt. CA1 – Provides a single calibration event or
coverage for the designated calibration interval,
whichever comes first.
Opt. C3 – Calibration Service 3 Years.
Opt. C5 – Calibration Service 5 Years.
Opt. D1 – Calibration Data Report.
Opt. D3 – Calibration Data Report 3 Years
(with Opt. C3).
Opt. D5 – Calibration Data Report 5 Years
(with Opt. C5).
Opt. R3 – Repair Service 3 Years.
Opt. R5 – Repair Service 5 Years.
Probes
Adapters
TCA-292MM – TekConnect to 2.92 mm connectors.
TCA-SMA – TekConnect-to-SMA adapter.
TCA-N – TekConnect-to-N adapter.
TCA-BNC – TekConnect-to-BNC adapter.
TCA75 – 4 GHz precision TekConnect 75 Ω to 50
Ω adapter with 75 Ω BNC input connector.
TCA-1MEG – TekConnect high-impedance buffer
amplifier. Includes P6139A passive probe.
Cables
GPIB Cable (1 m) – 012-0991-01.
GPIB Cable (2 m) – 012-0991-00.
RS-232 Cable – 012-1298-00.
Centronics Cable – 012-1214-00.
Accessories
Service Manual – 071-1740-xx.
Instrumented DIMM for DDR3 – Order Scope
NEXVu card for UDIMM Raw Card E. (Contact
www.nexustechnology.com).
Transit Case – 016-1977-00.
Rackmount Kit – 016-1985-00.
Oscilloscope Cart – K4000.
TDSUSBF – Test fixture for use with Opt. USB.
Probe Calibration and Deskew Fixture (4 GHz) –
067-0484-xx.
Probe Deskew Fixture (>4 GHz) – 067-1586-xx.
Power Deskew Fixture – 067-1686-xx.
Ethernet Test Fixture – Order through Crescent
Heart Software (http://www.c-h-s.com).
DPO/DSA70000 • www.tektronix.com/oscilloscopes
15
Digital Phosphor Oscilloscopes and Digital Serial Analyzers
Contact Tektronix:
ASEAN / Australasia (65) 6356 3900
DPO/DSA70000 Series
Austria +41 52 675 3777
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 07 81 60166
Instrument Upgrades
To upgrade your DPO70000 Series Oscilloscope
or your DSA70000 Series Serial Analyzer, order
option as noted:
XL02*6 – To upgrade record length on DPO70000
Series from standard configuration to Opt. 2XL
configuration.
XL05 – To upgrade record length from standard
configuration to Opt. 5XL configuration.
XL010 – To upgrade record length from standard
configuration to Opt. 10XL configuration.
XL020*9 – To upgrade record length from standard
configuration to Opt. 20XL configuration (only available on instruments of bandwidth >= 12.5 GHz).
XL25*6 – To upgrade record length on DPO70000
Series from Opt. 2XL configuration to Opt. 5XL
configuration.
XL210*6 – To upgrade record length on
DPO70000 Series from Opt. 2XL configuration to
Opt. 10XL configuration.
XL220*9*6 – To upgrade record length on
DPO70000 Series from Opt. 2XL configuration to
Opt. 20XL configuration (only available on instruments of bandwidth >= 12.5 GHz).
XL510 – To upgrade record length on DPO70000
Series or DSA70000 Series from Opt. 5XL configuration to Opt. 10XL configuration.
XL520*9 – To upgrade record length on DPO70000
Series or DSA70000 Series from Opt. 5XL configuration to Opt. 20XL configuration (only available on
instruments of bandwidth >= 12.5 GHz).
XL1020*9 – To upgrade record length on
DPO70000 Series or DSA70000 Series from Opt.
10XL configuration to Opt. 20XL configuration
(only available on instruments of bandwidth >=
12.5 GHz).
DDRA10 – To upgrade DPO70000 or DSA70000
Series with Opt. DDRA.
DJAH – To upgrade DPO70404, 70604, 70804
with Opt. DJA.
DJAU – To upgrade DPO71254, 71604, 72004
with Opt. DJA.
DJUP – To upgrade DSA70000 with Opt. DJA.
DVI – To upgrade DPO70000 Series or
DSA70000 Series with Opt. DVI.
SST*5 – To upgrade DPO70000 Series or
DSA70000 Series with Opt. SST.
Brazil & South America (11) 40669400
ET3 – To upgrade DPO70000 Series or
Canada 1 (800) 661-5625
DSA70000 Series with Opt. ET3.
LT – To upgrade DPO70000 Series or DSA70000 Central East Europe, Ukraine and the Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Series with Opt. LT.
Denmark +45 80 88 1401
ASM – To upgrade DPO70000 Series with Opt. ASM.
Finland +41 52 675 3777
USB – To upgrade DPO70000 Series or
France +33 (0) 1 69 86 81 81
DSA70000 Series with Opt. USB.
Germany +49 (221) 94 77 400
PWR – To upgrade DPO70000 Series or
Hong Kong (852) 2585-6688
DSA70000 Series with Opt. PWR.
India (91) 80-22275577
5
PCE* – To upgrade DPO70000 Series or
Italy +39 (02) 25086 1
DSA70000 Series with Opt. PCE.
Japan 81 (3) 6714-3010
IBA*5 – To upgrade DPO70000 Series or
Luxembourg +44 (0) 1344 392400
DSA70000 Series with Opt. IBA.
Mexico,
Central
America
& Caribbean 52 (55) 5424700
5
FBD* – To upgrade DPO70000 Series or
Middle East, Asia and North Africa +41 52 675 3777
DSA70000 Series with Opt. FBD.
The Netherlands 090 02 021797
HT3 – To upgrade DPO70000 Series or
Norway 800 16098
DSA70000 Series with Opt. HT3.
People’s Republic of China 86 (10) 6235 1230
MTH*6 – To upgrade DPO70000 Series with Opt. MTH.
Poland +41 52 675 3777
PTH*6 – To upgrade DPO70000 Series with Opt. PTH.
Portugal 80 08 12370
7
CP2* – TDSCPM2 ANSI/ITU Telecom pulse
Republic
of
Korea
82 (2) 6917-5000
compliance testing software (requires Opt. MTH
Russia & CIS +7 (495) 7484900
on DPO70000 Series).
South Africa +27 11 206 8360
UWBE – To upgrade DPO70000 Series or
Spain (+34) 901 988 054
DSA70000 Series with Opt. UWBE.
Sweden 020 08 80371
UWB – To upgrade DPO70000 Series or
Switzerland +41 52 675 3777
DSA70000 Series with Opt. UWB.
Taiwan 886 (2) 2722-9622
J2 – TDSDDM2 disk drive analysis software.
United Kingdom & Eire +44 (0) 1344 392400
VNM*8 – TDSVNM CAN and LIN Timing and
USA 1 (800) 426-2200
Protocol Decode (no CAN triggering included).
For other areas contact Tektronix, Inc. at: 1 (503) 627-7111
*2 Requires Ethernet Test Fixture.
Updated 12 November 2007
*3 Requires TDSUSBF (USB Test Fixture).
*4 At least Opt. 2XL and a TCA-1MEG TekConnect 1 MΩ buffer
amplifier are recommended for use.
*5 Requires Opt. RTE on DPO70000 Series.
*6 DPO70000 only.
*7 Requires Opt. MTH on DPO70000 Series.
*8 Requires ATM1 CAN/LIN trigger module - Order through
Crescent Heart Software.
*9 For models of bandwidth >= 12.5 GHz only.
*10 Requires DJA and ASM.
For Further Information
Tektronix maintains a comprehensive, constantly expanding collection of
application notes, technical briefs and other resources to help engineers
working on the cutting edge of technology. Please visit www.tektronix.com
Product(s) are manufactured
in ISO registered facilities.
Copyright © 2008, Tektronix. All rights reserved. Tektronix products are
covered by U.S. and foreign patents, issued and pending. Information
in this publication supersedes that in all previously published material.
Specification and price change privileges reserved. TEKTRONIX and
TEK are registered trademarks of Tektronix, Inc. All other trade names
referenced are the service marks, trademarks or registered trademarks
of their respective companies.
05/08 JS/WOW
4HW-19377-11