MULTI-FIELD EMF METER
TM-190
User’s Manual
B2TM1900001
1
Meun
1 Features ........................................................... 4 2 Identifying Parts ................................................ 6 3 TFT description ................................................. 7 4 Measurement Procedures ................................ 8 4.1 POWER ON/OFF Button: ................................... 8 4.2 Reading Individual and Aggregated XYZ Axial Data: .... 8 4.3 Data hold (HOLD): .............................................. 8 4.4 Menu Settings: .................................................... 9 4.4.1. Option Buttons:
10 4.4.2. Screen Brightness options:
10 4.4.3. Magnetic Unit:
10 4.4.4. RF Strength Unit:
11 4.4.5. Language:
12 4.4.6. Power Off Time:
13 4.4.7. Sound on/off:
14 4.4.8. Keys/Alarm Sound:
14 4.4.9. Keys Sound:
15 4.4.10. Alarm Sound:
15 2
English
4.4.11. 5 Information:
16 Specifications.................................................. 17 5.1 Sensor type: LF Magnetic Fields (MF) .............. 17 5.2 Sensor type: LF Electric Fields ......................... 17 5.3 Sensor type: RF Strength.................................. 18 6 Battery replacement ....................................... 19 7 External USB power supply: ........................... 20 8 Safety and maintenance standards ................ 20 9 End of life........................................................ 21 3
English
1
Features
z Data hold (HOLD)
z Low battery indication: HIGH
z Over load display “OL”.
LOW
z External USB Power Supply: ” ”
z Brightness options: low-Middle-high
z Magnetic unit:Gauss(mG) or. Tesla(uT)
z RF Strength Unit: (uW/m² ~mW/m²) (μW/cm²) (m V/m ~V/m)
(mA/m) (dBm).
z Languages: English; Traditional Chinese; Simplified Chinese;
Japanese; Español.
z Power Off Time:No; 1; 3; 5; 10; 15; 30. Factory default sets as
is displayed
“5”. Settings can be changed by the user.
on the screen after power off time is set.
; Off
z Keys/Alarm Sounds: On
z Low-Frequency EMF Readings: Individual and aggregated XYZ
.
axial readings
z RF Historical Records:
.
Up to 20 groups
z Information: Software version: V1.0
z High-frequency EMF Readings: Please perform tests according
.
to the indicated direction
4
TM-190 English
z Electric field Measurement Precautions: Please perform tests
. Please
according to the indicated direction
hold the meter at the bottom of the display, as shown in the
figure below:
Fig. 1 Electric Field measurement:
Magnetic Fields
Electric Field
RF strength
The Green zone
0~10.00mG
0~500V/m
0~0.99mW/m²;0~0.59V/m
The Yellow zone
10.01~100mG
501~1000V/m
1~9.99mW/m²;0.6~1.9V/m
The Red zone
101~2000mG
>1001V/m
>10mW/m²;>2V/m
The color zones are for reference only
z electromagnetic safety reference standard
international國際的 Council Recommendation1999/519/EC
42V/m(4.75W/m²) 59V/m(9.25W/m²)
international國際的 ICNIRP Guidelines,April1998
42V/m(4.75W/m²) 59V/m(9.25W/m²)
Austria 奧地利
ONORM S1120
Belgium比利時
Belgisch Staatsblad F.2001-1365
21V/m(1.18W/m²) 30V/m(2.31W/m²)
Germang 德國
26.Deutsche Verordnung
42V/m(4.75W/m²) 59V/m(9.25W/m²)
49V/m(6.33W/m²)
italy意大利
Decreto n.381,1998
the netherlands荷蘭 Health Council
switzerland瑞士
Verordnung 1999
united states美國
IEEE C95.1
china中國
japan日本
61V/m(10W/m²)
20V/m(1W/m²)
20V/m(1W/m²)
51V/m(6.92W/m²)
83V/m(18W/m²)
4V/m(0.04W/m²)
6V/m(0.1W/m²)
49V/m(6.33W/m²)
68V/m(12W/m²)
Draft:National Quality Technology Monitoring Bureau
49V/m(6.33W/m²)
61V/m(10W/m²)
Radio-Radiadiation Protection Guidelines,1990
49V/m(6.33W/m²)
61V/m(10W/m²)
1W/m²=0.1mW/Cm²=100uW/Cm².1mW/m²=0.1uW/Cm²
5
English
2
Identifying Parts
Fig. 2 Instrument description:
1.
2.
3.
4.
5.
6.
2.4” 240*320 resolution color TFT.
Power and Menu button
Hold and Enter button
Select and Down button
Mini USB power interface
Battery cover
6
English
3
TFT description
1
2
3
4
5
6
7
8
9
10
11
12
Data Hold indicator.
Auto power off indicator.
Buzzer indicator.
Battery indicator.
Connecting a USB power indicator.
Tesla(uT) or Gauss(mG) display.
Electric Field (V/m) display.
RF strength history histogram display.
RF strength digital (mV/m/ W/m²/μW/cm²/dB) display.
Menu key indication
Hold/ Enter key indicator
RF electronic field warning indicator:
GOOD/NORMAL/WARNING.
13 LF electric field warning indicator:
GOOD/NORMAL/WARNING.
14 Individual XYZ axial value of LF electromagnetic wave
display
15 LF electromagnetic wave warning indicator:
GOOD/NORMAL/WARNING
7
English
4
Measurement Procedures
4.1 POWER ON/OFF Button:
z Press
button to power on. LCD display measurement
screen.(see Fig. 3)
z Press
4.2
button for 3 seconds to power off LCD display.
Reading Individual and Aggregated XYZ Axial Data:
Direct the front section of the meter at the desired electromagnetic
field for measurements.
The meter simultaneously displays the electromagnetic field
readings of individual and aggregated XYZ axes, where the
aggregated calculation equation can be expressed as follows:
Because of environment-related magnetic field factors, this
electromagnetic field EMF meter may display a reading of under 0.50
mG prior to testing. This is caused by the magnetic noise in the
environment, rather than meter failure.
4.3 Data hold (HOLD):
to activate the HOLD function and
Press the left button as
hold the current readings at the upper left of the display.
To release the held data, press the left
Press the middle
button again.
button for 3 seconds to power off the meter.
8
English
4.4 Menu Settings:
On the measurement screen: Press the middle
button to enter
the main menu, where 7 options can be selected, namely,
Brightness, Magnetic unit, RF Strength Unit, Language, Power
off, Sound, and Information.
Press the right
button to make the blue brick scroll down.
button repeatedly, and the blue brick will
Press the right
cycle through the options.
Press the left
button to enter the selected option. Press the
button again to exit the selected option and return to the
left
main menu.
Press the middle
button to return to the previous menu.
9
English
4.4.1. Option Buttons:
Press the right
check box. Press the right
the check boxes.
Press the left
Press the middle
menu.
button to scroll to and check the next
button repeatedly to cycle through
button to exit to the main menu.
button to return to the previous setting
4.4.2. Screen Brightness options:
Following the operation procedures in 4.4, the Brightness setting
comprises the following three options, namely, Low, Middle, and
High. Under the sub-directory, use the option buttons to select and
confirm the desired selection.
Default: Middle
4.4.3. Magnetic Unit:
Following the operation procedures in 4.4, the Magnetic Unit setting
comprises the following two options, namely, Gauss/mG and Tesla/uT.
Under the sub-directory, use the option buttons to select and confirm the
desired selection.
Default: mG
10
English
4.4.4. RF Strength Unit:
Following the operation procedures in 4.4, the RF Strength Unit
setting comprises the following five options, namely, uW/m²~mW/m²,
uW/cm², mV/m~V/m, mA/m, and dBm (files are automatically skipped
within the selected unit). Under the sub-directory, use the option
buttons to select and confirm the desired selection.
Default: uV/m² ~ mW/m²
11
English
4.4.5. Language:
Following the operation procedures in 4.4, the Language setting
comprises the following five options, namely, English, Traditional
Chinese, Simplified Chinese, Japanese, and Español (Spanish).
Under the sub-directory, use the option buttons to select and
confirm the desired selection.
Default: English
12
English
4.4.6. Power Off Time:
Following the operation procedures in 4.4, the Power Off setting
comprises the following seven options, namely, NO, 1, 3, 5, 10, 15,
and 30 (min). Under the sub-directory, use the option buttons to
select and confirm the desired selection.
Default: 5min
13
English
4.4.7. Sound on/off:
Following the operation procedures in 4.4, the Sound setting
comprises the following two options, namely, On and Off.
Press the right
box. Press the right
button to scroll to and check the next check
button to cycle thought he options.
button to select the “Enable” setting and enter
Press the left
the Keys/Alarm Sound menu (Section 4.4.8).
button to select the “Disable” setting and
Press the left
return to the main menu(Fig. 3).
Press the middle
button to return to the measurement screen
4.4.8. Keys/Alarm Sound:
Following the selection of the Enable option in the Sound settings
menu, the Keys/Alarm Sound setting comprises the following two
options, namely, Keys and Alarm.
Press the right
button to scroll to and check the check the
button repeatedly to cycle
next check box. Press the right
through the options.
Press the left
button to enter the next options screen.
14
English
4.4.9. Keys Sound:
Following the selection of the Keys option in the Keys/Alarm Sound
settings menu, the Keys setting comprises 3 options, namely 1, 2,
and 3.
Press the right
box. Press the right
options.
Press the left
button to scroll to and check the next check
button repeatedly to cycle through the
button to exit to the main menu.
Default: 3
4.4.10. Alarm Sound:
Following the selection of the Alarm option in the Keys/Alarm Sound
settings menu, the Alarm setting comprises 3 options, namely 1, 2,
and 3.
Press the right
box. Press the right
options.
Press the left
button to scroll to and check the next check
button repeatedly to cycle through the
button to exit to the main menu.
Default: 3
15
English
4.4.11. Information:
Following the operation procedures in 4.4, select the Information
option to display the software version (V1.0).
Press the left
Press the middle
button to exit to the main menu.
button to return to the settings screen.
16
English
5
Specifications
5.1 Sensor type: LF Magnetic Fields (MF)
z The meter is equipped with three individual aerial sensors to
measure EMFs. The overload indications can be displayed
simultaneously on three axes (X, Y, Z):
z Range: 20/200/2000mG, 2/20/200µT.
z Resolution: 0.02/0.1/1 mG or 0.02/0.1/1 µT.
z Frequency response: 50/60 Hz
z Sensor: Triple Axis (X, Y, Z).
z Accuracy: ±(15%+100dgt).
5.2 Sensor type: LF Electric Fields
z Range: 50V/m to 2000V/m.
z Frequency response: 50/60Hz
z Accuracy: ± (7% + 50dgt).
17
English
5.3 Sensor type: RF Strength
z Frequency range:50MHz to 3.5GHz.
z Accuracy: ± 2dB at 2.45GHz
z Measurement units: uW/m²~mW/m²; μW/cm²; mV/m~V/m,
mA/m, and dBm
z Specified measurement range: (0.02uW/m²to 554.6mW/m²)
z (0.02uW/cm²to 55.4uW/cm²)(36.1mV/mto14.46V/m) (0.02mA/m
to 38.35mA/m)( -51dB to 16dBm)
z Display resolution: 0.02μW / m², 0.2μA / m, 0.2mV/m, m,
0.002μW/cm ², 2dB.
z Display: 4 digits Triple LCD display.
z Sample rate: 6 seconds per time.
z Battery life: Approximate 8 hours.
z Battery: 1.5V AAA Alkaline Battery*3.
z Audible Key tone alarm: Buzzer
z Operating temperature & humidity: 5°C to 40°C, below 80% RH.
z Storage temperature & humidity: -10°C to 60°C, below 70%.
z Weight: About 120g.
z Dimensions: 115(L)*60(W)*21(H) mm.
18
English
6
Battery replacement
WARNING
If the symbol “
” appears on the screen,
please replace the battery immediately
z
Turn off the instrument.
z
z
z
Open the battery covers and remove the battery.
Take three new 1.5VAAA alkaline batteries, and install the
batteries according to the polarity.
Put back the battery cover.
19
English
7
External USB power supply:
z
External USB.
Specifications: Voltage DC 4.8 ~ 5.2V current ≧ 500mA.
z
A
z
z
8
icon is displayed when an external USB power
supply is connected to the meter.
The meter determines connected USB cables to be
antennas, and thus such cables should not exceed the
top rim of the buttons during testing. If the cables
exceed this limit, the readings of subsequent tests are
considered invalid.
Safety and maintenance standards
z
z
z
z
z
z
Do not operate around combustible gases or in damp
environment.
Operating altitude: below 2,000m.
Operating environment: for indoor use, expose to pollution
level II.
This is a precision device. During use or storage, do not go
beyond its spec. to prevent any possible damage or
danger.
Do not put this device in direct sunlight or where it is hot
and/or damp.
For long storage, remove the battery to prevent the battery
from leaking and causing damage to the parts inside.
z Clean the device with a dry soft cloth. The use of wet
cloths, liquid and water is prohibited.
20
English
9
End of life
Caution
This symbol indicates that equipment and
its accessories shall be subject to a
separate collection and correct disposal
21
English
Professional Electrical and
Environment Test & Measurement
Instruments:
Battery Capacity/Impedance Tester,
TACHO Meter, LED light meter, Temperature
& Humidity meter, Infrared Thermometer,
Sound level meter, Light meter, EMF meter,
UV Light meter, RF meter, Hot wire
Anemometer, CO meter, Anemometer, Lan
cable tester, CO2 meter, Solar power meter,
Radiation meter,
Clamp meter, Multimeter, Phase Rotation
test, Digital Insulation tester.