steindorff - New York Microscope Co.

STEINDORFF
®
LCD Digital Microscope
NYMCS-1262
Instruction Manual
This manual is for LCD digital microscope NYMCS-1262. To ensure the safety, obtain optimum performance and
to familiarize yourself fully with the use of this microscope, it is strongly recommended that you study this manual
thoroughly before operating the microscope.
STEINDORFF
®
Content
Safety Notice ………………………………………………………………………………….....
3
Optical Parts
1. Components ……………………………………………………………………………….... 7
2. Technical parameter and specification ………………………………………………........
9
3. Configuration ……………………………………………………………………………….
11
4. Assembly …………………………………………………………………………………….
12
5. Operation …………………………………………………………………………………… 18
6. Troubleshooting ………………………………………………………………………….....
25
7. Maintenance …………………………………………………………………………….......
27
Digital Parts
8. Introduction ………………………………………………………………………………...
28
9. Structure …………………………………………………………………………………….
33
10. Usage and Operation ……………………………………………………………………….
35
11. Common Failure and Solution …………………………………………………………….
53
STEINDORFF
®
Safety notice
1. Transportation
As microscope is a precision instrument, handle with care, avoiding impact or abrupt movement during
transportation. Do not push or pull the microscope during using, otherwise the precision for image will be
reduced.
(1). Hold the curve and keep the microscope in balance
(2). Do not hold the focusing knob, eyepiece tubes and stage as these parts are movable.
Troubles maybe caused by such handles.
★Do not make specimen or filters fall off.
2. Working Environment
As microscope is a precision instrument, improper using will make it unworkable or reduce its precision.
(1). Do not expose the microscope in the sun directly.
(2). Temperature range is 0-40 and the max. humidity is 85%.
(3). Avoid high temperature and humidification otherwise there will be fog or mold on the lens.
(4). Avoid violent vibration as the vibration will reduce the images quality.
(5). Place microscope on a stable plane and keep it in balance. Keep the environment breezy and cover the
microscope with the dust-cover.
(6). Do not place microscope in moist room to avoid short circuit. Please turn off the power supply as soon
as water drop in the microscope. If there are other things unsafe come into the microscope may cause short
circuit too, please stop using and contact with manufacturer.
3. Focusing knob
Never turn the left and right focusing knob in the adverse direction at the same
time. Do not turn the coarse focusing knob when the stage reaches max position.
These improper operations will make focusing structure unworkable.
3
Curve
STEINDORFF
®
4. Clamp
The clamps fix the microscope firmly during
transportation, please remove them before using.
Clamp
5. Do not take the microscope apart
As it will be damaged. Such operations might have bad effect on the performance and make user get an
electric shock or injured. Please contact with manufacturer if there is any problem.
6. Make sure the voltage
There is correct input voltage in the label on microscope. Make sure
the voltage in your position is right. Improper input voltage may
cause short circuit and fire. The microscope will be damaged.
Input voltage in the label: 100-240V, 50/60Hz
7. Use the spare parts supplied by the manufacturer
The bulb, fuse and electronic cord have been assembled already in the factory and please make sure to use the
spare parts of them supplied by the manufacturer. Improper bulb, fuse and electronic cord will destroy
microscope and cause fire. Please make sure to use PE electronic cord when using extra-long electronic cord.
Spare bulbs: halogen bulb 12V20W (or 12V30W)
Spare fuse: 250V 1A delayed model, mini type fuse
4
STEINDORFF
®
8. Mounting the bulb
Never touch the surface of bulb with your hand directly. Please use gloves or cloth material when you
mount the bulb to avoid leaving fingerprints. Fingerprints or stains should be wiped off with a tissue
moistened with a small amount of alcohol. The fingerprints will etch on the bulb
s surface and make
brightness lower and life cycle shorter.
Please make sure the bulb’s touching points are Ok as it will not light or short circuit if the bulb’s touching
points are damaged. Insert the bulb’s legs into the bulb holder firmly. The bulb will go out if its legs have
not been inserted into the holder firmly. Such Operation may cause short circuit or smoking. At last, please
make sure if the collector has been mounted properly.
9. Temperature for illumination
There will be high temperature when the bulb is lightened, never move the collector when the bulb is
lightened. Do not touch the bulb within 30 minutes after it has been turned off. Make sure the bulb has been
cooling for enough time (at least 30 minutes) to avoid scald. Do not put fiber, papers and incendive things (eg.
Gasoline, aether, methanol and ethanol) close to the bulb.
10. Make sure to turn off the power supply before assembling microscope,
replacing bulb or fuse.
11. Observing with oil immersion
Using small amount of oil immersion is enough. The redundant oil will adhere to stage or
condenser and these will reduce microscope’s performance.
Get rid of the redundant oil or clean the lens by using aether or pure alcohol according to this
instruction. Pay attention during process as these things are incentive
5
STEINDORFF
®
12. Operating the microscope only indoor and under the altitude 2000m,
IP52 grads.
If the equipment is used in manner not specified by the manufacturer, the equipment may be
impaired.
6
STEINDORFF
®
Optical Part
1. Components
10X eyepiece
Binocular head
Nosepiece
Head fix screw
Objective
Microscope body
Specimen holder
Mechanical stage
Switch
Condenser
Collector
Brightness
adjust knob
Electronic cord
Fine focus knob
Stage movement knob (X-axis)
Stage movement knob (Y-axis)
7
STEINDORFF
®
Condenser movement knob
Coarse focus knob
Fine focus
Condenser diaphragm
knob
8
STEINDORFF
2. Technical parameter and specification
2.1 Specification
1. Optical system:
Infinite optical system
2. Illumination:
Inserted transmitted illumination, 12V20W halogen bulb(standard outfit)
Input voltage: 100V~240V
50/60Hz
Voltage undulation
Rating
Fuse
10%
100V~240V 50/60Hz 0.8A
250V 1A delayed model, mini type fuse 2pcs
3. Focusing adjustment:
Division of fine focusing adjustment: 0.002mm
Fine focusing knob control range: stage will be up or down 0.2mm per circle
Coarse focusing knob control range: stage will be up or down 20mm
4. Mechanical stage
X-movement range: 54mm, Y-movement range: 78mm
5. Nosepiece:
Roller bear quintuple nosepiece
6. Condenser:
Abbe NA=1.25 with iris diaphragm
7. Eye tubes:
Interpupillary range 55~75mm
9
®
STEINDORFF
®
8. Working environment:
Temperature: 0 ~40 , Humidity: max. 85%, no dew
2.2 Optical parameter
1. Optical parameter
Objectives: infinite plan achromatic objectives
Objectives
4X
N.A.
0.1
Cover glass thickness(mm)
0.17
W.D.
mm
25
Dry/Oil
Dry
10X
0.25
0.17
6.7
Dry
40X
0.65
0.17
0.6
Dry
2. Eyepiece: WF10X/22mm
2.3 Nomenclature
Total magnification
Total magnification = Eyepiece’s magnification * objective’s magnification
Numerical aperture (N.A.)
N.A. value will affect resolution and image’s brightness, it is the leading parameter for
objectives. N.A.= n*sin
n stands for refractive index of the medium between objective and specimen or condenser (air or oil
immersion)
a stands for half angle of max. aperture angle in the axis
The image will be sharper and brighter when the N.A. value is bigger
Resolution
Resolution can be measured by the distance value from one point to another one which could
be distinguished on the object surface
10
STEINDORFF
®
Resolution=  2N.A.
stands for wavelength (= 0.55 m)
Working distance
The distance between the object surface to the first surface of optical system when you get
the sharpest image. The W.D. will become more short when the magnification for the
objective is bigger.
Eyepiece view field
The max diameter measured when the diaphragm is open. 10X/20 means the magnification is
10X and the max diameter of view field is 20mm
Effective view field
Liner view field which is observed on the object surface
Effective view field = eyepiece view field / objective’s magnification
Depth of field
Depth of field stands for depth of the space in which there is sharp image on the object surface.
Depth of field will be longer when the diaphragm reduces. Depth of field will be shorter when
the N.A. becomes bigger.
3. Configuration
Microscope’s body ( including stage, nosepiece, coaxial coarse and fine
1
focusing adjustment, bracket for condenser, adjusting power supply, halogen
i
1pc
bulb and normal collector)
2
Siedentopf binocular inclined at 30° (360° rotatable)
1pc
3
Condenser ( with iris diaphragm, without auxiliary lens)
1pc
4
Objectives( Infinite achromatic objectives)4X10X40X100X
1 set
5
WF10X/22 eyepiece
2 pcs
11
STEINDORFF
®
6
Blue filter
1 set
7
Electronic cord
1 pc
8
Immersion oil
1 bottle
9
Dust cover
1 pc
10
Instruction and quality certificate
1 pc
11
Halogen bulb ( 12V20W)
2 pcs
12
Wrench 2.5, 1.5
1pc
Optional
Item
Content
1
Dark field condenser
2
WF10X eyepiece with reticle
3
WF15X eyepiece (16mm)
4
phase contrast condenser
5
LED transmitted illumination
6
110V, 240V transmitted illumination
7
Disc phase contrast condersor
8
Polarization attachment
9
WF20X eyepiece (12mm)
Piece
4. Assemblage
Please read the safety notice carefully before assembly and make the assembly according to the following
steps ever turn the left and right focusing knob in the adverse direction at the same time.
Tools: wrench (2 pcs)
12
STEINDORFF
®
4.1. Checking the input voltage
There is correct input voltage in the label on microscope. Make sure the voltage in your position is right.
Improper input voltage may cause short circuit and fire. The microscope will be damaged. Turn off the
power supply (turn the switch to “O” ) and insert one end of the electronic cord (connector) into the
entrance for AC. Insert the other end of the electronic cord (pin) into the grounded AC connector and
make sure that the electronic cord has been connected safely.
★Please use the supplied electronic cord by manufacturer
★Please make sure to use PE electronic cord when using extra-long
electronic cord.
★Pay attention that the microscope should be placed near AC
connector and the AC connector is touchable for users
4.2. Removing the clamps for transportation
AC input socket
The clamps fix the stage and focusing adjustment firmly during transportation.
Handle the grooves beside collector and pull it out. Remove the clamps by wrenches.
Mechanical stage
There is one clamp fixing the mechanical stage in Y axis direction. Remove the bolts and clamps.
13
STEINDORFF
®
Clamps of Y axis
Focusing adjustment
Pull out the collector. The bracket is fixed by one clamp.
Please remove the bolts and clamps.
Clamps of stage
Fixing screw for binocular head
4.3. Mount the binocular head
Loosen the fixing screw in the eye tubes and insert the
binocular head into the eye tubes. Then tighten the screw.
4.4. Mount the blue filter
Take out the filter from the bottom of condenser and mount it
in the bracket then move the bracket into the condenser.
14
STEINDORFF
®
4.5. Adjustment and replacement
(1). Condenser
The condenser is mounted in the microscope before shipment. Remove or
replace the condenser according to the following steps.
a.
Handle the grooves of the condenser and pull it out.
b.
Turn the lifting knob for condenser and make the bracket in a suitable
position.
c.
Loosen the screw in the left side of condenser and take out the condenser. Mount the blue filter in the
bracket then move it into the bottom of the condenser. Make the label in the condenser forward and move the
condenser into the bracket. At last tighten the screw.
d.
Handle the grooves and push the collector in.
e.
Turn the lifting knob of condenser and make the condenser to the highest position.
f.
Turn the lifting knob of condenser and make the condenser a lit bit lower. The image will be focused in
the right place of object (center of light path) when the light transit condenser. Move the condenser up and
down a little to make the dispersion image disappeared.
(2). Objectives
The objectives have been mounted in the microscope before shipment.
Take off the specimen from the stage and make the stage lower when you replace the objectives.
Handle the objective with your two hands and remove it. Be carefully do not make it fall off.
Turn the nosepiece to get the objective you need. The magnification for the objectives becomes bigger in
clockwise direction (observing from the top of the microscope).
(3). Specimen clip
15
STEINDORFF
®
The clip has been mounted in the stage before shipment.
Loosen two screws by wrenches supplied by manufacturer to remove the clip.
(4). Eyepieces
10X eyepieces have been in the foam box before shipment. Insert the eyepiece into the eyepiece tubes and
tighten the fixing screws by wrench.
Handle the eyepieces and replace them according to following picture.
Left and right eyepiece should be replaced together when replacing to use 15X eyepiece to keep the same
magnification between left and right eye tubes.
16
STEINDORFF
®
★Notice: The 10X eyepiece should touch the end of eye tube.
The “o” scale should be consistent with standard line when the eyepieces are inserted in or pulled out of
the eye tubes.
(5). Other attachments
Replace other attachments (eg, camera) according to the instruction.
(6). Replacement
a. Replacement for bulb
Turn off the power supply (turn the switch to “O”) and pull out the connector.
Wait for 30 minutes till the bulb is cooling Handle the grooves in condenser and
pull it out Pull out the original bulb. Replace the bulb by gloves or cloth material
then insert the bulb into the socket entirely.
Pull the condenser back to its original position. Connect the electronic cord to the
power supply
b. Replacement for fuse
Turn off the power supply (turn the switch to “O”) and pull out the connector.
Open the cover for fuse by screwdriver.
Mount the new fuse
★Make sure that the voltage shown in the cover for fuse is same to the
working voltage.
17
STEINDORFF
®
5. Operation
5. 1. Illumination
Turn on the power supply (turn the switch to “-”) then the bulb
will be on. The lightness can be adjusted by turning the
adjusting knob.
5.2. Interpupillary range
Adjust the siendentopf binocular to make sure that view field in the right and left eyepiece is consistent.
The point stands for the interpupillary distance.
5.3. Placing the specimen
Put the specimen in the stage and make the cover glass on it. Make the clips
clamp them entirely. Two slides can be put in the stage in the same time.
5.4. Focus by 10X objective
Put the 10X objective into the optic path and focus by turning coarse and fine focusing knob.
18
STEINDORFF
®
Turn the stage and focusing knob according to the direction shown in the picture.
There is no coarse focusing knob in one side of stage’s moving knob and in the other side there are both
coarse and fine focusing knob.
Never turn the left and right focusing knob in the adverse direction at the same time.
Never turn the coarse focusing knob when the stage is in the end of removing range. Otherwise it will
make damage to the instrument.
Detailed operation
It is difficult to do focusing if turning the focusing knob optionally.
The specimen may be crushed when 100X objective is used. Please
read the instruction carefully to avoid making damage to the cover
glass or objective
a.
Put the 10X or 4X objective into optic path.
b.
Remove the stage to the top by turning the coarse focusing knob.
c.
Observe through the eyepiece and turn the coarse focusing knob
slowly to let the stage down and down till there is sharp image.
d.
Turn the fine focusing knob to focus finely. Use the10X objective or
4X objective firstly before 40X and 100X objective. Turn the fine
focusing knob to focus finely.
5.5. Adjusting the diopter for eyepieces
Adjust the diopter ring according to user’s diopter for left and right eyes. This function could make use of
advantages of objectives fully. Meanwhile it also can be react as focusing.
a.
Put the 40X objective in the optic path and turn the coarse
19
STEINDORFF
®
and fine focusing knob to do focusing.
b.
Put the 10X objective (or 4X objective) in the optic path.
c.
Let your left eye observe through the left eyepiece and do
focusing by the diopter adjusting ring in left eyepiece.
d.
Repeat the above steps till sharp image can be observed
by left and right eyepiece at the same time.
5. 6. Adjusting the upright position for condenser
Turn the lifting knob for condenser till it reaches the top position then fall it a little bit down. If there is
dispersion image in the view field, please remove the condenser a little bit of up and down to make the
dispersion image disappeared.
5.7. Objectives
Turn the nosepiece to choose the objective and adjust the view field diaphragm and aperture accordingly.
5.8. View field diaphragm
Adjust the view field diaphragm according to the objective till its image touches the edge of view field
as this could eliminate the outside light and increase the contrast of the image.
★Notice: There will be no image for view field diaphragm when 100X objective is used. So adjust the
diaphragm to its min diameter
Centering view field diaphragm
a. Use the 10X objective and focus the image. Turn the view field diaphragm adjusting ring and reduce
the diaphragm’s diameter.
b. Turn the lifting knob for condenser to focus the image of view field diaphragm.
c. Turn the two centering screws for condenser and make the image in the center of view field.
d. Check the centering and open the view field diaphragm till the image of the diaphragm touches the
20
STEINDORFF
®
edge of view field. Please do the centering again if the image of the diaphragm does not touch the
edge of view field.
e. Open the view field diaphragm and make its image a little bigger than the view field in practice.
5.9.
Adjusting aperture diaphragm
Adjust the aperture diaphragm adjusting shaft for condenser till the position which stands for the magnification
of objective adjusting the aperture diaphragm
a. Adjust the aperture diaphragm by the adjusting shaft. The brightness and resolution will be decrease
while the contrast and depth of view will increase if the diameter of aperture diaphragm reduces.
b. Adjust the value of aperture diaphragm to 70%-80% of the value of objectives N.A..
c. The aperture diaphragm could control the N.A. of condenser. Do not adjusting the brightness by it
while use the brightness adjusting knob to make the adjustment.
d. N.A. is marked in each objective e.g. 40X/0.65 stands for that magnification is 40X and N.A. is 0.65.
e. Remove the eyepiece and observe through the eye tubes by eyes directly N.A. of condenser could tell
users the position of aperture diaphragm adjusting shaft in corresponding magnification. (That is to say,
the value of aperture diaphragm should be 70% to 80% of the N.A. of objective when the aperture
diaphragm adjusting shaft reaches some position) If the objective has been changed, please move the shaft to
the position with the same value which is in the objective in the optic path and this will make ideal contrast.
21
STEINDORFF
100X
60X
40X
20X
10X
®
4X
same value to the objective’s magnification
5. 10. Observing with oil immersion
There is “oil” marked in the oil immersion objective. Please use the oil immersion, which supplied by
the manufacturer, between the objective and cover glass.
a. Operation
Condenser
Remove the slide back and fall the condenser down a little. One drop of oil could be placed in the
top of condenser by through the long aperture of stage. Then move the slide forward and lift the
condenser up.
Objectives
Turn the nosepiece and take the objective out of the optic path. Place one drop of oil on the slide
then turn the condenser slowly and mount the objective.
b.
Air bubble in oil immersion
★Never make the oil immersion into your eyes. Please take following steps as soon as the oil
immersion touches your skin or eyes.
Use soap or clean water to wash your skin carefully.
Please wash your eyes with clean water (at least wash 15 minutes) and go to the hospital at once.
22
STEINDORFF
®
★ Do not expose the oil immersion in the sun or ultraviolet radiation directly.
As the air bubble in the oil immersion will make bad affect to the quality of image, please make sure
that there is no air bubble in the oil immersion before use. Please check the air bubble as following,
remove the eyepieces and open the view field diaphragm and aperture diaphragm entirely then observe
the exit pupil which is light and in nummular shape.
★Remove the air bubble as following:
Turn the nosepiece slowly and turn the oil immersion objective once or twice.
Turn the lifting knob for condenser slowly and make the condenser a little bit of up and down.
Increase the amount of oil immersion or change the old oil immersion with new one.
c. Operation for oil immersion
Use the oil immersion as little as possible. The oil immersion will be conglutinated to the stage and
condenser and this will have bad effect the performance if there is too much oil immersion. Please
clean the redundant oil on the objective and condenser after finishing the observation otherwise the
image will be affected. Use the aether to clean the oil then use the pure alcohol (ethanol or carbinol) to
make entirely cleaning. Please repeat the clean three or four times.
★Notice:
Please follow the instruction from manufacture when the aether or pure alcohol is used. Keep them
away from the fire or electronic spark.
d. Attention
★Keep the bottle for oil immersion sealed as much as possible and check it periodically.
★Never press the bottle overly as it may caused oil gushed out of the bottle.
★Clean the out surface of the bottle from remanent oil.
23
STEINDORFF
5.11.Adjusting the coarse focusing knob
The tension of coarse focusing knob is adjustable. Please turn the tension controlling
ring of the coarse focusing knob clockwise then the tension will be increased. This
ring is near the coarse focusing knob. If you want decrease the tension, please turn the
ring withershins. The stage will fall down automatically if the tension of the ring
decreased too much.
24
®
STEINDORFF
®
6. Troubleshooting
The performance of the microscope can be made fully by unfamiliar using and this table will give some
advice. Please see the following table and please contact with manufacturer if the troubles could not be
solved.
6.1. Optical Part
Trouble
Cause
Remedy
1.optical
Revolve to click position
Darkness at the periphery or
Revolving nosepiece not in click stop
position(objective not centered in
optical path)
(switch the objective correctly into the
optical path)
uneven view field brightness
Filament image not in center
Centering
Dirt or dust on the lens (condenser,
objective, eyepiece, collector)
Cleaning
Dirt or dust on the lens (condenser,
objective, eyepiece, collector)
Cleaning
Dirt or dust on the slide
Cleaning
Condenser position too low
No cover glass on the specimen
Correct position
Attach cover glass
Cover glass too thick or thin
Use cover glass of
specified thickness (0.17mm)
Slide upside-down
Immersion oil on dry objective
Turn over the slide
Cleaning
(especially 40X)
Dirt or dust on the lens (condenser,
Cleaning
objective eyepiece collector)
No immersion oil used on
Use immersion oil
i bubbles
i
bj immersion
i
Air
in
oil
Remove air bubbles
Unspecified immersion oil used
Use specified immersion oil
Condenser aperture and field
Close properly
Dirt or dust on the entrance lens
Cleaning
Condenser aperture closed too far
Close properly
Condenser position too low
Raise to the position where
the field diaphragm image is
in focus
Condenser not in the center of the
Reinstall condenser and carefully
Dirt or dust in the view field
Poor image quality
(low resolution poor contrast)
25
STEINDORFF
®
Revolving nosepiece not in click
Revolve to click-stop position
t
iti
Floating
specimen
Fasten securely
Insufficient illumination
Specimen rise from stage surface
Place it stable
brightness
Revolving nosepiece not in
Revolve to click-stop position
Image tinged yellow
Blue filter not used
Use blue filter
Condenser aperture too small
Readjust aperture
Condenser position too low
Correct position
Image shift on one side
Insufficient illumination
brightness
Dirt or dust on the lens (condenser
Cleaning
objective, eyepiece, collector)
6.2. Mechanical Part
Trouble
Cause
Remedy
Image not focusable with
Slide upside-down
Turn slide cover
high power objective
Cover glass too thick
High power objective
Side upside down
Use cover glass of specified
thickness (0.17mm)
Turn slide over
contacts slide when changed
Cover glass to thick
Use cover glass of specified
thickness (0.17mm)
Mechanical stage not securely fastened
Tighten all fastener
Interpupillary distance not correctly
adjusted
Incorrect diopter adjustment
Adjustment
over from low power
Specimen movement
unsmooth
Binocular images not
integrated
Excessive eye fatigue
Correct adjustment
Inadequate brightness or illumination
Adjust brightness with control
dial
Cause
Remedy
6.3. Electrical
Trouble
No electrical power
Lamp dose not light when Lamp bulb not inserted
switched ON
Lamp failure
26
Check power cord connection
Insert correctly
Replacement
STEINDORFF
Sudden lamp failure
Insufficient
brightness
illumination
®
Replace with specified lamp
Unspecified lamp bulb used or input bulb. If the same symptom
voltage too high
occurs after replacing the
lamp, contact your dealer
Unspecified lamp bulb used
Voltage too low
Flickering or unstable lamp
Lamp bulb about to fail
brightness
Lamp bulb not correctly inserted into
socket
Replace with specified lamp
bulb
Increase brightness with control
dial
Replacement
Check for positive connection
7. Maintenance
7.1. Cleaning Lenses
Dust is best removed with a soft brush or gauze.
More persistent dirt, such as fingerprints, grease and oil, may be removed with soft cotton, lens tissue,
or gauze lightly moistened with absolute alcohol (ethyl or methyl alcohol).
To clean immersion oil off the oil-immersion type objective, use lens tissue, soft cotton or gauze lighty
moistened with petroleum benzine only.
Do not use petroleum benzine to clean the entrance lens at the bottom of the eyepiece tube or prism
surfaces inside the eyepiece tube.
Absolute alcohol and petroleum benzine are quite inflammable. Take great care when handling them
and when setting the power switch on and off. Be very careful with fire.
7.2. Cleaning painted or plastic surfaces
Avoid use of and organic solvents (such as alcohol, ether, thinner,ect.) to clean the painted or plastic
surfaces of the instrument. We recommend the use of silicon cloth.
More persistent dirt may be cleaned with mild detergent solution.
Printed plastic surfaces should be cleaned only by soft cloth moistened with water.
27
STEINDORFF
®
7.3. When not in use
When the microscope is not in use, cover it up with dust cover, and store in a dry place not subject to
mold.
We especially recommend that the objectives and eyepieces be kept in a container (such as a
plastic bottle).
7.4. Periodical inspection
To maintain the performance of the microscope, periodical inspection is recommended.
Digital Part
. Introduction
1.1 About Digital Display
Digital display is a new generation display that embedded Android system. It is
with 8 inch IPS LCD screen, A7 dual-core processor, friendly man-machine
interface and powerful image processing functions (supporting image browsing,
photos, video, storage, dynamic calibration, measurement etc.). This display
supports 5 points touch, a variety of peripheral equipment interfaces, and
comfortable ergonomic design. It is very easy to operate and is your good
assistant on the teaching, scientific, electronic detection fields.
1.2 Main Features
28
STEINDORFF
®
1. Embedded OS Android 4.2.2, can link mouse and keyboard, just like a
micro computer.
2. 8 inch IPS LCD screen with clear and bright vivid vision.
3. Friendly human-machine interface, can use the mouse, keyboard or touch
screen to operate.
4. Built-in a 5.0M camera, providing the high quality images.
5. Provides a powerful software of international advanced technology. It
supports real-time preview, photos, video, storage, network, Bluetooth,
dynamic calibration and measurement.
6. LCD screen supports 5 points touch. Very convenience to use.
7. Supports more interfaces, such as USB, SD card, Mini USB, HDMI
interface and so on.
8. Supports WI-FI wireless communication, can transmit real-time
image
and video.
9. Images and videos can be captured and saved to the SD card and
USB driver for subsequent analysis.
10. Automatic measurement, real-time displays measurement results.
1.3 System Specification
Cortex-A7 CPU
29
STEINDORFF
®
Interface

Built-in 2GB memory.

Supports external extension SD card.
Camera video interface

Support external USB camera.

5 million pixels camera to image when using LCD. Frame rate: 10-15
fps.
Man-machine interface

8.0 inch (Diagonal)IPS screen, resolution:1024*768.

Supports the keyboard and mouse input.

Capacitive touch screen.

Hardware key: reset, power switch.
Expansion card

Supports large capacity SD card. The maximum capacity is 32G.
External interface

USB 2.0 HOST interface in flank.
30
STEINDORFF

®
USB 2.0 OTG interface at the back

HDMI interface

Mini USB interface
Audio module

Stereo speaker with double channels
Network module

Supports WI-FI access

Supports bluetooth
Power management

Power mode: ordinary, sleep (lock screen) and power off. System default
sleep time is 20 days when no operating.

Supports some module shut to minimize power consumption.
Operation environment

5V 2A DC power adapter

LED power lamp

OS support all kinds of languages.
31
STEINDORFF

Video player software, picture viewer software.
1.4 Camera Specification
Sensor size (diagonal)
1/2.5-inch (4:3)
Effective image size
5.70mm(H)x4.28mm(V)
7.13mm(diagonal)
Effective pixels
2592H x 1944V
Pixel size
2.2um x 2.2um
Dynamic range
66.5dB
S/N (maximum)
40.5dB
A/D conversion resolution
12-bit, on-chip
Sensitivity
0.53V/lux-sec(550nm)
Video format and frame rate
1280x720, 15fps
640x480, 30fps
Exposure range and mode
Electronic shutter (ERS), auto
White balance
Manu/Auto
32
®
STEINDORFF
Software
TouchScope
Software function
Capture, Video and dynamic measure
®
. Structure
Structure is shown as below:
Power
lock key
USB interface
of connecting
camera
Supply
hub
HDMI
interface
33
STEINDORFF
®
Mini USB interface
SD card slot
USB interface
When booting, Please long press the power key until the screen appear image.
When shutting down, please long press the power key, popping up a box as
below:
Then click “OK”.
34
STEINDORFF
®
. Usage and Operation
3.1 Software-TouchScope
TouchScope is a special software of microscope system that apply to android
platform. Using this software, you can finish the image capture, video and
dynamic image calibration and measurement.
3.1.1 Main Function
The main function of the software as follows:
1) Real-time image preview
2) Capture image
3) View the captured images
4) Real-time measurement.
5)
Support bluetooth, WI-FI and Gmail.
3.2 Operate TouchScope
1. Boot to run android OS. Please long press the power key until the screen
appear image.
2. Click the desktop icon
, the main interface of software shown as below:
35
STEINDORFF
®
Photo
files
Tool
bar
Capture
Preview video
Video
3.2.1 Capture
Icon
Instruction
Click”capture”icon, start to capture and display image on
the photo viewing thumbnails
36
STEINDORFF
®
3.2.2 Video
Icon
Instruction
Click ”video” icon, the software is in video mode. Then
click”
”, software starts to video.
If want to stop video, please click”
”and will end the
function of video. Then click”
”,software interface
will restore to the original state.
3.2.3 Color Adjustment
Click button “
Icon
”, pop-up function box:
Instruction
Click icon, software performs automatic white balance
mode..
37
STEINDORFF
®
RGB : manual white balance adjustment
Click icon, software performs incandescent lamp mode.
Click icon, software performs fluorescent lamp mode
Click icon, software performs sunny mode
Click icon, software performs cloudy mode
3.2.4 Dynamic Calibration and Dynamic Measurement
TouchScope provides a variety of measurement tools and supports calibration
and dynamic measurement. Before measuring, please do calibration.
Click”
”, unfold a control panel of measuring tool:
38
STEINDORFF
Icon
®
Instruction
Spot measurement
Circle measurement
Line measurement
Rectangle measurement
Cross measurement
Angular measurement
Tool for modifying the color and thickness of
lines and tag text.
Export graphics to images
Delete the specified measurement graphic
(Click the button
Delete all measurement graphics
again and hide measuring tools)
When measuring, it is recommended to use the mouse to stretch and drag the
lines. Please do the specific setting as follows:
39
STEINDORFF
Click“
®
”,enter the setting interface of software:
As shown in the red box, check the mode of using mouse, which is a convenient
way to measure. The measurement data will be more accurate.
A. Calibration
Before measuring, please do calibration.
Here the objective is 10X and the micro ruler is 0.1mm.
40
STEINDORFF
®
1. Put a micro ruler on the stage. select the objective (here is 10X), focus the
microscope, and make the micro ruler display clearly on the LCD screen.
2.Set calibration value
Click button“
”, pop up a box:
click “edit calibration value”, pop up a box:
41
STEINDORFF
Click “
”, pop up a box::
42
®
STEINDORFF
®
Enter the name of the calibration item and choose a unit. Then click “Yes”.
There will come out an adjustable calibrating line.
3. Move the line by mouse and set the starting point.
43
STEINDORFF
Click “
®
”, enter the physical length of this line.
Click “
”, save the result of calibration.
Then click return key, back to the video.
B
Line measurement
Click the icon “
”on the control panel.
Select the line on the video, move line, click one side of line and adjust the start
point and end point of line. Then click “
”, loading the calibration,
now the measurement value will display on the screen.
44
STEINDORFF
®
Here the length is 0.413mm. It is matched with the physical length.
C.
Rectangle measurement
Click the icon “
”and display the Rectangle tool on the screen. Click the
center of Rectangle can move Rectangle. Click four corners of rectangle can
adjust the size of rectangle. Then the length, width and area of rectangle will
display on the screen. As shown in the below, area is 0.04078mm2,, which is
matched with physical area.
45
STEINDORFF
D.
®
Reticle ruler
Click the reticle button. There will be a reticle tool on the image, click the center
of the reticle with the mouse, there will be a little white reticle in the center of
the reticle, drag the reticle, the reticle will move.
E.
Property of ruler
Click the button “
”, there are many tools to modify the color, line width and
text size of ruler. As shown in the below:
46
STEINDORFF
®
Click the color selection bar of line, pop up a box:
Click and choose one color you wanted on the annulus. Then click inner circle
and you can get the choosed color successfully.
Click the button “
”, add the width of line.
Click the button “
”, minus the width of line.
The way to change the color and size of text is the same as the way to change
line.
F
Delete graph
Click on the selected graph, then click button “
Continuously pressing the button “
” can delete all graph.
3.2.5 Software Setting
Click button “
” and delete it.
”, pop up a box to set software .
47
STEINDORFF
®
1. Camera
Control
panel
◆Preview size: choose one preview resolution of video. Two
resolutions: 640*480;1024*768.
◆Capture
size:
choose
one
capture resolution. Three
resolutions: 2592*1944;1600*1200;2048*1536.
Function
◆Camera state: check the option to show the state of camera.
There will display the resolution and frame rate of video on the
video.
Initial Set
◆If getting video preview resolution higher, the video frame
rate will drop.
2. Storage
48
STEINDORFF
®
Control
panel
◆Storage location:Choose the storage path of image and video.
Three paths: internal storage, SD card, USB flash disk.
Function
◆JPEG compression quality: inferior quality (50%), medium
quality (70%), high quality (80%), ultra quality (90%)
◆Choose different storage path to save files.
Initial set
◆The compression quality of software is
medium quality (70%)
3. Measurement
Control
panel
Function
◆Mouse mode: check the mouse mode, which is convenient to
measure. The measurement value will be more accurate.
49
STEINDORFF
Initial set
®
When measuring, it is more advantageous to improve the
accuracy of measurement after choosing the mouse mode.
4. Interface
Control
panel
◆Autohide: check the autohide mode to set delay time to hide
Function
the widget on the interface of software.
◆The tolerant delay time is 5S. In 5 seconds, there is no people
operate software, the widget will be hide. If don’t check this
Initial set
mode, software will not hide widget.
5. Other
Control
panel
◆Clear data: permanently remove all data on the software.
50
STEINDORFF
®
◆Feedback: feedback suggestion to developer of software.
Function
◆About: software version
◆Please don’t easily remove the data of software.
Initial set
3.2.6 Image Processing Module
Click the button “
” and process the captured image
1. Optimize image
Item
Icon
Instruction
Image
There are many different forms set for the color of
color
image: punch, vintage, black & white, instant, coffee
latte, blue, plate mark, negative.
2. Image border
Item
Icon
Instruction
Adding different borders for image.
Border
3. Crop image
Item
Icon
Instruction
51
STEINDORFF
Straighte
®
Drag the image by finger to change image size.
n
You can select the proportion to crop the image, also can
Crop
drag the white border by finger to cut the size of image.
Drag the image by finger to rotate
Rotate
Drag the image by finger to mirror
Mirror
4. Image color
Item
Icon
Instruction
Set the color of image: automatically adjust the color,
Image
exposure, halation, contrast, cloudy, vibrance, acutance,
color
curve, saturation, black and white filter.
5. Send image
Click the button “
”, there are three ways to send image: bluetooth, gmail,
picasa
Item
Icon
Instruction
52
STEINDORFF
®
Bluetooth
Open bluetooth function in the setting of system and
send image by bluetooth.
Gmail
Through the gmail to send image
Picasa
Share image to the Google web photo album.
Common Failure And Solution
4.1 Software Crash
Phenomenon: Software crash, point out “Camera error: 100”.
Cause: The interface of camera is unsteadiness or wrong operation.
Solution: 1. Exit software, check the connection of USB line and plug
line again to ensure the connection is tightness. then restart
software.
2. After restarting software, if the software can not run
normally,
please shut down and restart display.
4.2 System Halted
Phenomenon: The system halt, and no response of any operation.
Causes: Wrong operation or the unstable voltage.
53
STEINDORFF
®
Solutions: 1) Press the “Power Button” for more than 2 seconds,
2) Click the small “Rst” hole with a needle (or other things
that have a small, sharp-point) to reset your system, as
the picture below.
4.3 Can Not Switch On The System
Phenomenon: Can not switch on the system or the system can not work.
Causes: Power isn’t connected properly.
Solutions: Check the power supply of the LED head, make sure
the 5V DC is connect properly. Or Plug the power and
link again.
54