HY-OPTIMATM 2700 Explosion Proof In

HY-OPTIMATM 2700 Explosion Proof In-line Hydrogen Process Analyzer
Description
H2scan’s HY-OPTIMATM 2700 real time hydrogen-specific process analyzers are designed for ease of use, interface
flexibility and true process control. The HY-OPTIMATM 2700 uses a solid-state sensor that is configured to operate in
process gas streams. The HY-OPTIMATM 2700 is ideal for hydrogen production and petrochemical applications
where real-time measurements can enhance process plant efficiencies, diagnostics and maintenance management.
Operating Conditions
Accuracy(*):
Operating Humidity: 0% to 95% RH (non-condensing)
Flow Rate: 0.1 slpm to 50 slpm
Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge
Process Gas Temperature Range: -20ºC to 110ºC
Operating Temperature Range: -20ºC to 55ºC
Storage Temperature Range: -30ºC to 80ºC
Calibration Background Gas: Nitrogen
Drift/week:
Adapter Fittings
Hydrogen Sensitivity Range:
0.5% to 100% hydrogen by volume at 1 ATM
Response Time: T90 < 30 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
± 0.3% absolute for 0.5 to 10% H2
± 1.0% absolute for 10 to 100% H2
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Repeatability:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Linearity:
± 0.2% absolute for 0.5-10% H2
± 0.4% absolute for 10-100% H2
Interface Options
Input Voltage Range: 90 VAC to 240 VAC
Input Power: 15W
Analog Outputs:
Output Currents
• 4 mA to 20 mA
• 0 mA to 20 mA
• User-specific mA range
¾” Union Tee Compression
Dimensions
Width: 13.8 cm (5.4 in)
Length: 19.2 cm (7.5)
Depth: 13.0 cm (5.7 in)
For Latest Revision go to www.h2scan.com
Performance
Serial Communication Options: RS232 or RS422
Relay Contacts: Two programmable relays and
one fault relay with both normally open (N.O.)
and normally closed (N.C.) contacts:
• 5A/240VAC
• 5A/ 30 VDC
Certifications:
UL 1203
UL 508
USC – Class I, Division 1, Groups B, C, and D
Hazardous (Classified)
∗
Sensor performance specifications are only valid for units configured for a maximum 65ºC dry process stream and at 25 ºC ambient temperature. All
figures assume pressure compensation, operating in ambient that do not contain Oxygen and are in addition to any errors in the gasses used. The
accuracy is specified for serial port output only.
TM
HY-OPTIMA is a registered Trademark of H2scan
Specifications subject to change without prior notice
90000057 R2
27215 Turnberry Lane, Unit A, Valencia CA., 91355
Tel: 661-775-9575, Fax: 661-775-9515
Email: sales@h2scan.com, Website: www.h2scan.com
HY-OPTIMATM 2720 Explosion Proof In-line Process Hydrogen Analyzer (Air)
• Calibration in an Oxygen Background
Description
H2scan’s Series 2700 real time hydrogen-specific process analyzers are designed for ease of use, interface
flexibility and true process control. The HY-OPTIMATM 2720 is a solid state sensor configured to operate in air
background gas streams having up to 95% relative humidity and temperatures up to 65ºC. The HY-OPTIMATM 2720
is ideal for the monitoring of hydrogen processes where real time measurements of hydrogen can ensure safety in
application, increase process efficiencies, improve diagnostics and enhance maintenance management translating
into higher profitability.
Operating Conditions
Hydrogen Sensitivity Range:
0.5% to 5% hydrogen by volume at 1 ATM
Response Time: T90 < 60 sec
Ingress Protection: IP64 capable
Calibration Interval: 90 days
Product Life Expectancy: 10 years
Accuracy: ± 3% of reading + 2000 ppm
Operating Humidity: 0% to 95% RH
Flow Rate: 0.1 slpm to 50 slpm
Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge
Process Gas Temperature Range: -20ºC to 65ºC
Operating Temperature Range: -20ºC to 40ºC
Storage Temperature Range: -30ºC to 80ºC
Calibration Background Gas: Air
Interface Options
Adapter Fittings
Input Voltage Range: 90 VAC to 240 VAC
Input Power: 15W
Analog Outputs:
Output Currents
• 4 mA to 20 mA
• 0 mA to 20 mA
• User-specific mA range
Serial Communication Options: RS232 or RS422
Relay Contacts: Two programmable relays and
one fault relay with both normally open (N.O.)
and normally closed (N.C.) contacts:
• 5A/240VAC
• 5A/ 30 VDC
¾” Union Tee Compression
Dimensions
Width: 13.8 cm (5.4 in)
Length: 19.2 cm (7.5)
Depth: 13.0 cm (5.7 in)
For Latest Revision go to www.h2scan.com
Performance
Certifications:
UL 1203
UL 508
USC – Class I, Division 1, Groups B, C
and D Hazardous (Classified)
∗
Sensor performance specifications are only valid for units configured for a maximum 65ºC dry process stream and at 25 ºC ambient
temperature. All figures assume pressure compensation and are in addition to any errors in the gasses used. The accuracy is specified for serial
port and digital display output only.
TM
HY-OPTIMA is a registered Trademark of H2scan
Specifications subject to change without prior notice
90000062 R1
27215 Turnberry Lane, Unit A, Valencia CA., 91355
Tel: 661-775-9575, Fax: 661-775-9515
Email: sales@h2scan.com, Website: www.h2scan.com
HY-OPTIMATM 2730 Explosion Proof Process Hydrogen Analyzer (ppm H2S & CO)
• CO and H2S Tolerance
Description
H2scan’s Series 2700 real-time hydrogen specific process analyzers are designed for ease of use, interface
flexibility and true process control. The HY-OPTIMATM 2730 is a solid-state sensor configured to operate in
background gas streams having carbon monoxide concentrations up to 100 ppm, hydrogen sulfide concentrations up
to 1000 ppm, up to 95% relative humidity, and gas temperatures up to 110ºC. The HY-OPTIMATM 2730 is ideal for
hydrogen production and petrochemical applications where real-time measurements of hydrogen can increase
process plant efficiencies, improve diagnostics and enhance maintenance management translating into higher
profitability.
Hydrogen Sensitivity Range:
0.5% to 100% hydrogen at 1 ATM on a
volume basis in a single or multicomponent gas background
Response Time: T90 < 30 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy(*):
± 0.15% absolute for 0.5 to 10% H2
± 0.5% absolute for 10 to 100% H2
Drift/week:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Repeatability:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Linearity:
± 0.2% absolute for 0.5-10% H2
± 0.4% absolute for 10-100% H2
Interface Options
Input Voltage Range: 90 VAC to 240 VAC
Input Power: 15W
Analog Outputs:
Output Currents
• 4 mA to 20 mA
• 0 mA to 20 mA
• User-specific mA range
Operating Conditions
Process Gas Tolerance:
CO tolerance: 100 ppm
H2S tolerance: 1000 ppm
Operating Humidity: 0% to 95% RH
Flow Rate: 0.1 slpm to 50 slpm
Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge
Process Gas Temperature Range: -20ºC to 110ºC
Operating Temperature Range: -20ºC to 55ºC
Storage Temperature Range: -30ºC to 80ºC
Calibration Background Gas: Nitrogen
Adapter Fittings
¾” Union Tee Compression
Dimensions
Width: 13.8 cm (5.4 in)
Length: 19.2 cm (7.5)
Depth: 13.0 cm (5.7 in)
For Latest Revision go to www.h2scan.com
Performance
Certifications:
UL 1203
UL 508
USC – Class I, Division 1, Groups B, C and D
Hazardous (Classified)
Serial Communication Options: RS232 or RS422
Relay Contacts: Two programmable relays and
one fault relay with both normally open (N.O.) and
normally closed (N.C.) contacts:
• 5A/240VAC
• 5A/ 30 VDC
∗
Sensor performance specifications are only valid for units configured for a maximum 65ºC dry process stream and at 25 ºC ambient
temperature. All figures assume pressure compensation and are in addition to any errors in the gasses used. The accuracy is specified for
serial port and digital display output only.
TM
HY-OPTIMA is a registered Trademark of H2scan
Specifications subject to change without prior notice
90000063 R1
27215 Turnberry Lane, Unit A, Valencia CA., 91355
Tel: 661-775-9575, Fax: 661-775-9515
Email: sales@h2scan.com, Website: www.h2scan.com
HY-OPTIMATM 2740 Explosion Proof In-line Area Hydrogen Analyzer (% H2S & CO)
• % Level CO and % Level H2S Tolerance
Description
H2scan’s Series 2700 real-time hydrogen specific process
analyzers are designed for ease of use, interface
flexibility and true process control. The HY-OPTIMATM 2740 is a solid-state sensor configured to operate in
background gas streams having carbon monoxide, concentrations up to 20% by volume, hydrogen sulfide
concentrations
up to 10% by volume, up to 95% relative humidity, and temperatures up to 110ºC. The HYOPTIMATM 2740 is ideal for hydrogen production and petrochemical applications where real-time measurements of
hydrogen can increase process plant efficiencies, improve diagnostics and enhance maintenance management
translating into higher profitability.
Hydrogen Sensitivity Range:
0.5% to 100% hydrogen at 1 ATM on a volume basis
in a single or multi-component gas background
Response Time: T90 < 60 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy(*):
± 0.3% absolute for 0.5 to 10% H2
± 1.0% absolute for 10 to 100% H2
Drift/week:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Repeatability:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Linearity:
± 0.2% absolute for 0.5-10% H2
± 0.4% absolute for 10-100% H2
Interface Options
Input Voltage Range: 90 VAC to 240 VAC
Input Power: 15W
Analog Outputs:
Output Currents
• 4 mA to 20 mA
• 0 mA to 20 mA
• User-specific mA range
Operating Conditions
Process Gas Tolerance:
CO tolerance: 20% by volume
H2S tolerance: 10% by volume
Operating Humidity: 0% to 95% RH (non condensing)
Flow Rate: 0.1 slpm to 50 slpm
Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge
Process Gas Temperature Range: -20ºC to 110ºC
Operating Temperature Range: -20ºC to 55ºC
Storage Temperature Range: -30ºC to 80ºC
Calibration Background Gas: Nitrogen
Adapter Fittings
¾” Union Tee Compression
Dimensions
Width: 13.8 cm (5.4 in)
Length: 19.2 cm (7.5)
Depth: 13.0 cm (5.7 in)
For Latest Revision go to www.h2scan.com
Performance
Serial Communication Options:
RS232 or RS422
Relay Contacts: Two programmable
relays and one fault relay with both
normally open (N.O.) and normally
closed (N.C.) contacts:
• 5A/240VAC
• 5A/ 30 VDC
Certifications:
UL 1203
UL 508
USC – Class I, Division 1, Groups B, C, and D Hazardous
(Classified)
∗
Sensor performance specifications are only valid for units configured for a maximum 65ºC dry process stream and at 25ºC ambient temperature.
All figures assume pressure compensation, operating in ambient that do not contain Oxygen and are in addition to any errors in the gasses used.
The accuracy is specified for serial port and digital display output only.
TM
HY-OPTIMA is a registered Trademark of H2scan
Specifications subject to change without prior notice
90000064 R1
27215 Turnberry Lane, Unit A, Valencia CA., 91355
Tel: 661-775-9575, Fax: 661-775-9515
Email: sales@h2scan.com, Website: www.h2scan.com