Technical Information
Liquisys M CPM223/253
pH/ORP Measurement
Transmitter for analog and digital glass and ISFET sensors
The modular design of the Liquisys M CPM223/253 allows easy
adaption of the transmitter to a variety of customer
requirements. Starting with the basic version for "measurement
and alarm generation", the transmitter can be equipped with
additional software and hardware modules for special
applications. These modules can also be retrofitted as required.
Application
• Effluent treatment
• Neutralization
• Detoxication (electroplating)
• Water treatment
• Water monitoring
Your benefits
•
•
•
•
Memosens technology
Field or panel-mounted housing
Universal application
Simple handling
– Logically arranged menu structure
– Large two-line display
– Ultrasimple two-point calibration
• Safe operation
– Overvoltage (lightning) protection
– Direct access for manual contact control
– Calibration plausibility check
– User-defined alarm configuration
The basic unit can be extended with:
• Addtional 2 or 4 contacts for use as:
– Limit contacts (also for temperature)
– P(ID) controller
– Timer for simple rinse processes
– Complete cleaning with Chemoclean
– Current input
• Plus package:
– User defined current output characteristics
– Automatic cleaning trigger on alarm or limit violation
– Sensor Check System for pH glass and reference
– Live check of sensor
– Special neutralization controller
• HART or PROFIBUS-PA/-DP
• 2nd current output for temperature, pH/ORP or continuous
controller
TI194C/07/EN/13.11
71130248
Liquisys M CPM223/253
Function and system design
Features of the basic version
pH and ORP value measurement
This is selected via the menu. During measurement, the value measured can be displayed in the other
measuring mode (e.g. pH - mV or ORP % - ORP mV). The temperature is displayed at the same time or, if
desired, not shown at all.
Calibration
pH electrodes are normally calibrated with the same pH values. Therefore the transmitter presents the settings
from the previous calibration as defaults for the next calibration. If the buffer solutions are interchanged by
accident (e.g. pH 4 buffer first, then pH 7 buffer instead of pH 7 first and then pH 4) the plausibility check
ensures that the calibration is accepted anyway.
Configuration
Different alarms are required depending on application and operator. Therefore the transmitter permits
independent configuration of the alarm contact and error current for each individual error. Unnecessary
or undesirable alarms can be suppressed in this manner. Up to four contacts can be used as limit contacts
(also for temperature) to implement a P(ID) controller or for cleaning functions.
Direct manual operation of the contacts (bypassing the menu) provides quick access to limit, control or
cleaning contacts, permitting speedy correction of deviations.
Additional functions of the
plus package
Current output
In order to output wide measuring ranges while still achieving a high resolution in specific ranges, the current
output can be configured as required via a table. This permits bilinear or quasi-logarithmic curves, etc.
Sensor-Check-System (SCS)
The sensor check system alerts to deviations of the pH glass impedance or reference impedance (analog sensors
only) from the normal range, thus indicating possible failure due to pH electrode blocking or damage.
In addition, the SCS detects glass breakage of glass electrodes and leakages of ISFET sensors.
Live-check
The live check issues an alarm when the sensor signal does not change over a defined period of time. This may
be caused by blocking, passivation, separation from the process, etc.
Neutralization controller
A special control response that cannot be handled adequately by a P(ID) controller is required to neutralize
solutions. For this reason, the transmitter provides a special neutralization controller function by combining
two P(ID) controllers.
Current input
The current input of the transmitter allows two different applications: controller shut-down in case of lower
flow rate violation or total failure in the main flow as well as feedforward control. Both functions are also
combinable.
2
Endress+Hauser
Liquisys M CPM223/253
Measuring system
A complete measuring systems comprises:
• The transmitter Liquisys M CPM223 or CPM253
• A pH/ORP electrode with or without an integrated temperature sensor
• An immersible, flow or retractable assembly
• A measuring cable(e.g. CPK9)
Options: extension cable, junction box VBA or VBM
1
2
3
REL 1
ALARM
CAL
REL
REL 2
REL 1
E
REL 2
4
8
9
4
7
REL 1
ALARM
CAL
REL
REL 2
REL 1
E
REL 2
5
6
a0007781
Complete measuring system Liquisys M CPM223/253
Flow assembly CPA250
1
Junction box VBA
2
Liquisys M CPM253
3
Measuring cable e.g. CPK9
4
Liquisys M CPM223
5
Endress+Hauser
6
7
8
9
Retractable assembly Cleanfit W CPA450
Electrode, e.g Orbisint CPS11
Immersion assembly CPA111
Extension cable
3
Liquisys M CPM223/253
Input
Measured variables
pH (analog or digital sensors)
ORP
Temperature
Measuring range
pH:
ORP:
Temperature:
Pt 100
Pt 1000 (versions IS / PS)
NTC 30K (versions IS / PS)
–2 to 16
–1500 to +1500 mV / 0 to 100 %
-50 to +150 °C (-58 to +302 °F)
-50 to +150 °C (-58 to +302 °F)
-20 to +100 °C (-4 to +212 °F)
Input resistance
> 1012 Ω (for nominal operating conditions) for standard sensors
Cable specification
Length of cable (analog):
Length of cable (digital):
max. 50 m (164 ft)
max. 100 m (328 ft)
Binary inputs
Voltage:
Power consumption:
10 to 50 V
max. 10 mA
Current input
4 to 20 mA, galvanically separated
Load: 260 Ω at 20 mA (voltage drop 5.2 V)
Output
Output signal
0/4 to 20 mA, galvanically separated, active
HART
Signal coding
Frequency Shift Keying (FSK) + 0.5 mA via current output signal
Data transfer rate
1200 Baud
Galvanic isolation
yes
PROFIBUS PA
Signal coding
Manchester Bus Powered (MBP)
Data transfer rate
31.25 kBit/s, voltage mode
Galvanic isolation
yes (IO-Module)
PROFIBUS DP
Signal coding
RS485
Data transfer rate
9.6 kBd, 19.2 kBd, 93.75 kBd, 187.5 kBd, 500 kBd, 1.5 MBd
Galvanic isolation
yes (IO-Module)
Signal on alarm
2.4 or 22 mA
Load
maximum 500 Ω
4
Endress+Hauser
Liquisys M CPM223/253
Output range
pH:
ORP:
absolute:
relative:
Temperature:
adjustable, min. Δ 1 pH
adjustable, min. Δ 50 mV
fixed, 0 to 100 %
adjustable, Δ 10 to Δ 100 % of upper range value
Resolution
max. 700 digits/mA
Min. distance for
0 / 4 to 20 mA signal
10% of measuring range
Isolation voltage
max. 350 VRMS/500 V DC
Overvoltage protection
according to EN 61000-4-5
Auxiliary voltage output
Output voltage:
Output current:
15 V ± 0.6
max. 10 mA
Contact outputs
Switching current with ohmic load (cos ϕ = 1):
Switching current with inductive load (cos ϕ = 0.4):
Switching voltage:
Switching power with ohmic load (cos ϕ = 1):
Switching power with inductive load (cos ϕ = 0.4):
max. 2 A
max. 2 A
max. 250 V AC, 30 V DC
max. 500 VA AC, 60 W DC
max. 500 VA AC, 60 W DC
Limit contactor
Pickup/dropout delay:
0 to 2000 s
Controller
Function (adjustable):
Controller response:
Control gain Kp:
Integral action time Tn:
Derivative action time Tv:
Period for pulse length controller:
Frequency for pulse frequency controller:
Basic load:
pulse length/pulse frequency controller
PID
0.01 to 20.00
0.0 to 999.9 min
0.0 to 999.9 min
0.5 to 999.9 s
60 to 180 min-1
0 to 40% of max. set value
Alarm
Function (selectable):
Alarm threshold adjustment range:
Alarm delay:
latching / momentary contact
pH / temperature: complete measuring range
0 to 2000 s
0 to 2000 min
Endress+Hauser
5
Liquisys M CPM223/253
Protocol specific data
HART
Manufacturer ID
11h
Device type code
0091h
Transmitter specific revision
0001h
HART specification
5.0
DD files
www.products.endress.com/profibus
Load HART
250 Ω
Device variables
None (dynamic variables PV, SV, only)
Features supported
-
PROFIBUS PA
Manufacturer ID
11h
Ident number
1516h
Device revision
11h
Profile version
2.0
GSD files
www.products.endress.com/profibus
GSD file version
Output values
Main value, temperature value
Input values
Display value of PLC
Features supported
Device locking: The device can be locked by hardware or software.
PROFIBUS DP
Manufacturer ID
11h
Ident number
1520h
Profile version
2.0
GSD files
www.products.endress.com/profibus
GSD file version
6
Output values
Main value, temperature value
Input values
Display value of PLC
Features supported
Device locking: The device can be locked by hardware or software.
Endress+Hauser
Liquisys M CPM223/253
Power supply
Electrical connection of analog
sensors
41
pH
42
A
43
Ref.
SRC
B
47
48
DRN.
C
49
S
58
59
11
12
E
F
52
G
53
mA
32
33
55
56
mA
34
10-50 V
24 mA
82
J
P
~ =
L1 L+
N L-
10-50 V
O
23 +
93
I
N
54
pH
81
H
M
51
13
31
L
57
PA/
PM
D
K
Q
94
85
86
15 V
a0007753
Electrical connection Liquisysy M
J
Standard sensor
A
K
ISFET sensor
B
L
Outer screen connection for glass electrodes
C
M
Potential matching
D
N
Temperature sensor
E
O
Signal output 1 pH/ORP
F
Signal output 2 temperature, pH/ORP or controller P
G
Q
Binary input 1 (Hold)
H
Binary input 2 (Chemoclean)
I
Endress+Hauser
Auxiliary voltage output
Alarm (contact position currentless)
Relay 1 (contact position currentless)
Relay 2 (contact position currentless)
Relay 3 (contact position currentless)
Relay 4 (contact position currentless)
Current input 4 to 20 mA
Power supply
7
Liquisys M CPM223/253
Electrical connection of
Memosens sensors
A
B
41
S
42
BN
87
43
WH
88
47
H
48
GN
A
96
49
YE
B
97
57
I
58
31
C
mA
32
33
D
J
59
51
pH
52
K
53
mA
34
54
55
81
E
10-50 V
82
23 +
24 -
93
F
10-50 V
L
56
mA
94
M
~ =
L1 L+
N L-
85
G
86
N
15 V
a0007754
Transmitter electrical connection with Memosens technology
H
Screen
A
I
Sensor
B
J
C Signal output 1 pH/redox
K
D Signal output 2 temperature, pH/redox or controller
L
Binary input 1 (Hold)
E
M
Binary input 2 (Chemoclean)
F
N
G Auxiliary voltage output
8
Alarm (contact position currentless)
Relay 1 (contact position currentless)
Relay 2 (contact position currentless)
Relay 3 (contact position currentless)
Relay 4 (contact position currentless)
Current input 4 to 20 mA
Power supply
Endress+Hauser
Liquisys M CPM223/253
Sensor cable
The pH and ORP electrodes are connected using special terminated and shielded multicore cables. The
measuring cable can be extended with a junction box and an extension cable.
Termination instructions are supplied with the measuring cable.
J
PA/
PM
Schirm /
Screen
S
11 12 13
GN WH YE
BN
S
pH/
Redox
BNC
PA/
PM
11 12 13 DRN
GN WH YE RD
BN
S
Ref
innen / BK
internal
PE
SRC
SRC Ref
innen / BK
internal
PE
Ref
BK
)
CPK12
CPK9
A
B
A
BN
B
BN
C
C
CPS471
CPS11
a0007765
Connection CPS11 with CPK9 and CPS471 with CPK12 to Liquisys M
A
B
C
Panel-mounted instrument
Field instrument
Potential matching PM for symmetrical connection
PA/
Ref
BN
PA/
PM
S NC
BNC
Ref
Schirm /
screen
Schirm /
Screen
S NC PM
BK
BN BNC
BK
CPK1
CPK1
A
B
A
BN
B
BN
C
CPS12
CPS12
a0007766
Unsymmetrical and symmetrical connection of ORP electrodes to Liquisys M
A
B
C
Endress+Hauser
Panel-mounted instrument
Field instrument
Potential matching PM for symmetrical connection
9
Liquisys M CPM223/253
S
87 88 96 97
BN WH GN YE
CYK10
CPS11D
a0007773
Connection of digital sensor CPS11D with CYK10
Supply voltage
Fieldbus connection
Depending on ordered version:
100/115/230 V AC +10/-15 %, 48 to 62 Hz
24 V AC/DC +20/-15 %
HART
Supply voltage
n/a, active current outputs
Integrated reverse voltage protection
n/a, active current outputs
PROFIBUS PA
Supply voltage
9 V to 32 V, max. 35 V
Polarity sensitive
no
FISCO/FNICO compliant acc. to IEC 60079-27
no
PROFIBUS DP
Supply voltage
9 V to 32 V, max. 35 V
Polarity sensitive
n/a
FISCO/FNICO compliant acc. to IEC 60079-27
no
Power consumption
max. 7.5 VA
Mains protection
Fine-wire fuse, medium-slow blow 250 V/3.15 A
10
Endress+Hauser
Liquisys M CPM223/253
Performance characteristics
Reference temperature
25 °C (77 °F)
Resolution
pH:
ORP:
Temperature:
Maximum measured error1)
Display
pH:
ORP:
Temperature:
Signal output
pH:
ORP:
Temperature:
max. 0.75 % of measuring range
max. 0.75 % of measuring range
max. 1.25 % of measuring range
Repeatability1)
pH:
ORP:
max. 0.2 % of measuring range
max. 0.2 % of measuring range
Zero point
Glass:
Antimon:
ISFET:
pH 5.0 to 9.0 (nominal pH 7.00)
pH –1.0 to 3.0 (nominal pH 1.00)
-500 to +500 mV
Slope
Glass:
Antimon:
ISFET:
38.00 to 65.00 mV/pH (nominal 59.16 mV/pH)
25.00 to 65.00 mV/pH (nominal 59.16 mV/pH)
38.00 to 65.00 mV/pH (nominal 59.16 mV/pH)
Offset
pH:
ORP:
Temperature:
±2 pH
±120 mV/±50 %
±5 °C
1)
0.01 pH
1 mV/0.1 %
0.1 °C
max. 0.5 % of measuring range
max. 0.5 % of measuring range
max. 1.0 % of measuring range
acc. to IEC 746-1, for nominal operating conditions
Endress+Hauser
11
Liquisys M CPM223/253
Installation
Installation instructions
Pg 13.5
Pg 7
Pg 16
ALARM
REL
REL2
REL1
REL2
Ø6/
Ø 0.24
E
247 / 9.72
REL1
CAL
70 / 2.76
157 / 6.18
M5
11/
0.43
70 / 2.76
115 / 4.53
154 / 6.06
170 / 6.69
mm / inch
a0005733
Field instrument
1
1
2
3
Ø max. 60 /
2.36
mm / inch
1
3
a0005737
Mounting on pipes
1 - 3 Mounting screws and mounting plate
12
Endress+Hauser
Liquisys M CPM223/253
1
2
3
4
Removable electronics box
Partition plate
Terminal blocks
Fuse
4
1
2
3
a0005734
View into the field instrument
2
Ø6/
Ø 0.24
1
154 / 6.06
mm / inch
a0005736
Wall mounting of the field instrument
1
2
Mounting holes
Protecting cap
3
2
1
3
a0005738
Mounting of the field instrument with mounting post and weather protection cover
1 - 3 Mounting holes
Endress+Hauser
13
Liquisys M CPM223/253
96 / 3.78
6/
0.24
149.5 / 5.89
139 / 5.47
92 / 3.62
90 / 3.54
mm / inch
a0005735
Dimensions panel-mounted instrument
92+0.5 / 3.62+0.02
92+0.5 / 3.62+0.02
1
3
96 / 3.78
92 /
3.62
90 / 3.54
57 / 2.24
33 /
1.30
2
3
approx. 25 /
0.98
139 / 5.47
max. 45 /
1.77
approx. 165 / 6.50*
6/
0.24
3
mm / inch
a0005739
Installation of the panel-mounted instrument
1
2
3
*
14
Wall of control cabinet
Gasket
Tensioning screws
Required installation depth
Endress+Hauser
Liquisys M CPM223/253
Environment
Ambient temperature
-10 to +55 °C (+14 to +131 °F)
Storage temperature
–25 to +65 °C (-13 to +149 °F)
Electromagnetic compatibility
Interference emission and interference immunity as per EN 61326-1:2006, EN 61326-2-3:2006
Ingress protection
Panel mounted instrument:
Field instrument:
Electrical safety
according EN/IEC 61010-1:2001, Installation Category II, for use up to 2000 m above sea level
CSA
Apparatus with CSA General Purpose Approval are certified for indoor use.
Relative humidity
10 to 95%, non-condensing
Pollution degree
The product is suitable for pollution degree 2.
IP 54 (front), IP 30 (housing)
IP 65 / tightness acc. to NEMA 4X
Mechanical construction
Dimensions
Field instrument:
96 x 96 x 145 mm (3.78 x 3.78 x 5.71 inches)
Mounting depth: approx. 165 mm (6.50")
247 x 170 x 115 mm (9.72 x 6.69 x 4.53 inches)
Weight
Panel mounted instrument:
Field instrument:
max. 0.7 kg (1.5 lb)
max. 2.3 kg (5.1 lb)
Materials
Housing of panel mounted instrument:
Field housing:
Front membrane:
Polycarbonate
ABS PC FR
Polyester, UV-resistant
Terminals
Cross section
2.5 mm2 (14 AWG)
Endress+Hauser
Panel mounted instrument:
15
Liquisys M CPM223/253
Operability
Operating concept
All instrument control functions are arranged in a logical menu structure. Following access code entry, the
individual parameters can be easily selected and modified as needed.
Display elements
7
1
pH
7.05 R262
Setpoint
6
2
5
3
4
a0007771-en
Operating elements
1
2
3
4
5
6
7
LC display for display of measured values, configuration data and current menu field
Field for user labeling
4 main control keys for calibration and instrument configuration
Key for switching between automatic/manual operation
LED indicators for switched limit outputs
LED indicator for alarm function
Display of active contact and key for relay switching in manual mode
The display simultaneously shows the current measured value and the temperature - the essential process data.
Brief information texts in the configuration menu provide assistance with parameter configuration.
Certificates and approvals
4 symbol
Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus complies with the legal
requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.
Ex approval for zone 2
Application of transmitter as related electrical equipment in non-hazardous area or in simple pressurized
apparatus; application of sensor in hazardous area zone 2
C.M2.3-..4...
C.M223-..6...
C.M253-..A...
CSA general purpose
16
ATEX II (3)G (Ex nL) IIC
C.M2.3-..2...
C.M2.3-..3...
C.M2.3-..7...
Endress+Hauser
Liquisys M CPM223/253
Ordering information
Product structure
Sensor input, software
IS
MR
MS
PR
PS
pH (glass/ISFET) / ORP; Plus package
pH (digital sensor); pH (glass/ISFET)/ORP; basic version
pH (digital sensor); pH (glass/ISFET)/ORP; Plus package
pH (glass)/ORP; basic version
pH (glass)/ORP; Plus package
Power supply, approval
A
0
1
2
3
4
5
6
7
8
24 V AC/DC; ATEX II (3)G (Ex nL) IIC (CPM253 only)
230 V AC
115 V AC
230 V AC; CSA Gen. Purp.
115 V AC; CSA Gen. Purp.
230 V AC; ATEX II (3)G [Ex nL] IIC
100 V AC
24 V AC/DC; ATEX II (3)G [Ex nL] IIC (CPM223 only)
24 V AC; CSA Gen. Purp.
24 V AC/DC
Output
0
1
3
4
5
6
1 x 20 mA, primary value
2 x 20 mA, primary value + secondary value
PROFIBUS PA
PROFIBUS DP
1 x 20 mA, primary value, HART
2 x 20 mA, primary value, HART + secondary value
Additional contacts
05
10
15
16
20
25
26
not selected
2 relays (limit/P(ID)/timer)
4 relays (limit/P(ID)/Chemoclean) (not with PROFIBUS DP)
4 relays (limit/P(ID)/timer) (not with PROFIBUS DP)
1 x 4 ... 20 mA input + 2 relays (limit/P(ID)/timer)
1 x 4 ... 20 mA input + 4 relays (limit/P(ID)/Chemoclean) (not with PROFIBUS DP)
1 x 4 ... 20 mA input + 4 relays (limit/P(ID)/timer) (not with PROFIBUS DP)
Additional features (CPM223 only)
PRL
Protective layer
Marking
1
Tagging (Tag), see additional spec.
CPM253complete order code
CPM223-
Additional functions of the
Plus package
2)
•
•
•
•
Current output table to cover large areas with varying resolution
Monitoring of sensor and process for safe operation
Neutralization controller to keep pH value constant by dosing acid and alkali
Automatic cleaning function start2)
In combination with "Additional contacts" only, see product structure
Endress+Hauser
17
Liquisys M CPM223/253
Scope of delivery
The delivery of the field instrument includes:
• 1 transmitter CPM253
• 1 plug-in screw terminal
• 1 cable gland Pg 7
• 1 cable gland Pg 16 reduced
• 2 cable glands Pg 13.5
• 1 Operating Instructions BA194C/07/EN
• versions with HART communication:
1 Operating Instructions Field Communication with HART, BA208C/07/EN
• versions with PROFIBUS communication:
1 Operating Instructions Field Communication with PROFIBUS PA/DP, BA209C/07/EN
• versions with explosion protection for hazardous area zone II (ATEX II 3G):
Safety Instructions for use in explosion-hazardous areas, XA194C/07/A3
The delivery of the panel mounted instrument includes:
• 1 transmitter CPM223
• 1 set of plug-in screw terminals
• 2 tensioning screws
• 1 BNC-plug (solder-free)
• 1 Operating Instructions BA194C/07/EN
• versions with HART communication:
1 Operating Instructions Field Communication with HART, BA208C/07/EN
• versions with PROFIBUS communication:
1 Operating Instructions Field Communication with PROFIBUS PA/DP, BA209C/07/EN
• versions with explosion protection for hazardous area zone II (ATEX II 3G):
Safety Instructions for use in explosion-hazardous areas, XA194C/07/A3
Accessories
Sensors
Glass sensors
Orbisint CPS11/CPS11D
• pH sensor for process applications
• Optional SIL version for connection to SIL approved transmitters
• With dirt-repellent PTFE diaphragm
• Ordering acc. to product structure, see Technical Information (TI028C/07/en)
Orbisint CPS12/CPS12D
• ORP electrode for process applications
• With dirt-repellent PTFE diaphragm
• Ordering acc. to product structure, see Technical Information (TI367C/07/en)
Ceraliquid CPS41/CPS41D
• pH sensor
• With ceramics diaphragm and liquid KCl electrolyte
• Ordering acc. to product structure, see Technical Information (TI079C/07/en)
Ceraliquid CPS42/CPS42D
• ORP electrode
• With ceramics diaphragm and liquid KCl electrolyte
• Ordering acc. to product structure, see Technical Information (TI373C/07/en)
Ceragel CPS71/CPS71D
• pH sensor
• With double chamber reference system and integrated bridge electrolyte
• Ordering acc. to product structure, see Technical Information (TI245C/07/en)
Ceragel CPS72/CPS72D
• ORP electrode
• With double chamber reference system and integrated bridge electrolyte
• Ordering acc. to product structure, see Technical Information (TI374C/07/en)
Orbipore CPS91/CPS91D
• pH sensor
• With open aperture for media with high dirt load
• Ordering acc. to product structure, see Technical Information (TI375C/07/en)
18
Endress+Hauser
Liquisys M CPM223/253
Orbipore CPS92/CPS92D
• ORP sensor
• With open aperture for media with high dirt load
• Ordering acc. to product structure, see Technical Information (TI435C/07/en)
ISFET sensors
Tophit CPS471
• Sterilizable and autoclavable ISFET sensor
• For food and pharmaceuticals, process technology, water treatment and biotechnology;
• Ordering acc. to product structure, see Technical Information (TI283C/07/en)
Tophit CPS441
• Sterilizable ISFET sensor for media with low conductivity
• With liquid KCl electrolyte
• Ordering acc. to product structure, see Technical Information (TI352C/07/en)
Tophit CPS491
• ISFET sensor with open aperture for media with high dirt load
• Ordering acc. to product structure, see Technical Information (TI377C/07/en)
Connection accessories
CPK9 special measuring cable
• For sensors with TOP68 plug-in head, for high-temperature and high-pressure applications, IP 68
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CPK1 special measuring cable
• For pH/ORP electrodes with GSA plug-in head
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CPK2 special measuring cable
• For pH/ORP electrodes with GSA plug-in head, with three sensor plugs
• Ordering acc. to product structure, see Technical Information (TI 118C/07/en)
CPK12 special measuring cable
• For pH/ORP glass electrodes and ISFET sensors with TOP68 plug-in head
• Ordering acc. to product structure, see Technical Information (TI118C/07/en)
CYK10 Data cable for digital sensors
• For digital pH sensors with Memosens technology
• Ordering according to product structure, see below
Certificates
A
G
O
S
Standard, non Ex
ATEX II 1G EEx ia IIC T6/T4
FM CI.I Div. 1 AEx ia IIC T6/T4
CSA IS CI.I Ex ia IIC T6/T4
Cable length
03
05
10
15
20
25
88
89
Cable length: 3 m / 9.84 ft
Cable length: 5 m / 16.41 ft
Cable length: 10 m / 32.81 ft
Cable length: 15 m / 49.22 ft
Cable length: 20 m / 65.62 ft
Cable length: 25 m / 82.03 ft
... m length
... ft length
Ready-made
1
CYK10-
Endress+Hauser
Wire terminals
complete order code
19
Liquisys M CPM223/253
CYK81 measuring cable
• Non-terminated measuring cable for extension of sensor cables of e.g. Memosens sensors, CUS31/CUS41
• 2 wires, twisted pair with shield and PVC-sheath (2 x 2 x 0.5 mm2 + shield)
• Sold by the meter, order no.: 51502543
Junction box VBM
• For cable extension
• 10 terminals
• Cable entries: 2 x Pg 13.5 or 2 x NPT ½"
• Material: aluminum
• Ingress protection: IP 65 (i NEMA 4X)
• Order numbers:
– cable entries Pg 13.5: 50003987
– cable entries NPT ½": 51500177
Junction box VBA
• For cable extension of pH/ORP sensors
• 10 terminals, protection class: IP 65 (i NEMA 4X)
• Cable entries: 2 x Pg 13.5, 2 x Pg 16
• Material: polycarbonate
• Order no.: 50005276
Junction box RM
• For cable extension (e.g. for Memosens sensors)
• 5 terminals
• Cable entries: 2 x Pg 13.5
• Material: PC
• Ingress protection: IP 65
• Order no.: 51500832
Mounting accessories
CYY101 weather protection cover for field devices, absolutely essential if operating the unit outdoors
• Material: stainless steel 1.4031 (AISI 304)
• Order No. CYY101-A
a0012427
Weather protection cover for field devices
*
**
20
Wall and post mounting
Rail mounting
Endress+Hauser
Liquisys M CPM223/253
Post mounting kit
• For mounting of field housing on horizontal or vertical pipes (Ø max. 60 mm (2.36"))
• Material: stainless steel 1.4301
• order no. 50086842
90 / 3.54
70 / 2.76
M6
60 / 2.36
90 / 3.54
Ø8/
0.31
70 / 2.76
Ø8/
0.31
30 /
1.18
Ø max. 60 / 2.36
Ø 6 / 0.24
70 / 2.76
90 / 3.54
mm / inch
a0005676
Post mounting kit
150 /
5.91
1495 / 58.86
CYY102 universal post
• Square pipe for mounting transmitters
• Material: stainless steel 1.4301 (AISI 304)
• Order No. CYY102-A
120 /
4.72
60 x 60 /
2.36 x 2.36
mm/inch
a0005742
Universal post
Endress+Hauser
21
Liquisys M CPM223/253
Buffer solutions
High-quality buffer solutions of Endress+Hauser - CPY20
The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal
Physico-technical Institute) and to standard reference material of NIST (National Institute of Standards and
Technology) according to DIN 19266 by a DKD (German Calibration Service) accredited laboratory.
pH value
A
C
E
G
I
K
M
pH 2.00 (accuracy ± 0.02 pH)
pH 4.00 (accuracy ± 0.02 pH)
pH 7.00 (accuracy ± 0.02 pH)
pH 9.00 (accuracy ± 0.02 pH)
pH 9.20 (accuracy ± 0.02 pH)
pH 10.00 (accuracy ± 0.05 pH)
pH 12.00 (accuracy ± 0.05 pH)
Quantity
01
02
10
50
20 x 18 ml (0.68 fl.oz) only buffer solutions pH 4.00 and 7.00
250 ml (8.45 fl.oz)
1000 ml (0.26 US gal)
5000 ml (1.32 US gal) canister for Topcal S
Certificates
A
Buffer analysis certificate
Version
1
CPY20-
Standard
complete order code
Technical buffer solutions for ORP electrodes
• +220 mV, pH 7.0, 100 ml (3.4 fl.oz.); order no. CPY3-0
• +468 mV, pH 0.1, 100 ml (3.4 fl.oz.); order no. CPY3-1
KCl-electrolyte solutions for liquid filled electrodes
• 3.0 mol, T = -10 ... 100 °C (14 ... 212 °F), 100 ml (3.4 fl.oz.), order no. CPY4-1
• 3.0 mol, T = -10 ... 100 °C (14 ... 212 °F), 1000 ml (34 fl.oz.), order no. CPY4-2
• 1.5 mol, T = -30 ... 100 °C (-22 ... 266 °F), 100 ml (3.4 fl.oz.), order no. CPY4-3
• 1.5 mol, T = -30 ... 100 °C (-22 ... 266 °F), 1000 ml (34 fl.oz.), order no. CPY4-4
Optoscope
22
Optoscope
• Interface between transmitter and PC / laptop for service purposes.
• The Windows software "Scopeware" required for the PC or laptop is supplied with the Optoscope. The
Optoscope is supplied in a sturdy plastic case with all the accessories required.
• Order no. 51500650
Endress+Hauser
Liquisys M CPM223/253
Endress+Hauser
23
TI194C/07/EN/13.11
71130248