APPLICATIONs sTANdARd FEATUREs sPECIFICATIONs

CALIBRATION STANDARDS
Condec 3030 Hand-Held Pressure/Temperature Calibrator
Standard Features
calibration standards
• Many models and external modules available
(contact factory for additional items)
• 4 AA cells included, > 35-hours, typical usage
• Ranges from vacuum to 10,000 psi, -15 to 15 psi,
-15 to 30 psi
• 0.025% of full scale for models with 15 psi thru
3000 psi internal pressure ports
• 0.015% measurement of 0-24 mA or 0-10V
• RS-232/External pressure module port standard
• High-contrast back lit Liquid Crystal Display (LCD),
up to 3 process sub-regions
• Multi-engineering units standard: psi, bar, mbar,
kPa, mPa, kg/cm2, mmH2O @ 4° C, mmH2O @
20° C, cmH2O @ 4° C, cmH2O @ 20° C, inH2O @
4° C, inH2O @ 20° C, inH2O @ 60° F, mmHg @ 0° C, inHg @ 0° C
• Built-in stainless steel sensor for gas and liquid
compatibility, 1/8-27 NPT female pressure port
• Operating temperature -10° C to 50° C (14° F to 122° F)
• Switch test on pressure 1, pressure 2 or external
pressure module
• Built-in damping function
• Minimum/Maximum pressure readings
• Percent error displayed
• Built-in 24 VDC power supply to power unit under test
• Instrument setup recall, 5 last used on power-up
• Calibration Certificate included - traceable to the
National Institute of Standards and Technology
(NIST)
Applications
• Boiler feed pressure
• Condenser vacuum
• Pneumatic system checks
• Transmitter, transducer and switch testing
• Biomedical quality assurance
• Pressure transfer standard
• Temperature measurement
• Gas custody transfer
specifications
Pressure Ranges:
“120462 - NO TOOLS REQUIRED”
Gauge: 0.3* psi, 1* psi, 5 psi, 15 psi, 30 psi, 100 psi,
300 psi, 500 psi, 1000 psi, 3000 psi,5000 psi,
10000 psi
Absolute: 15* psi, 30* psi, 100 psi, 300 psi
Compound: -15 to 15* psi, -15 to 30* psi
Accuracy:
0.3 psi .................................. ±0.10% F.S.
1 psi, 5 psi............................. ±0.05% F.S.
15 psi through 3000 psi.........±0.025% F.S.
5000 psi, 10000 psi...............±0.035% F.S.
Isolated stainless steel pressure ports except those
noted with an (*) above.
Stated accuracies above are for internal pressure
ports only.
Consult factory for external pressure modules.
Electrical Ranges:
“106594 - NO TOOLS REQUIRED”
0 to 24.000 mA or 0 to 30.000 VDC
Accuracy:
mA ±0.015% of reading ±0.002 mA
Volts ±0.015% of reading ±0.002 VDC
Temperature:
RTD (ohms) -40.0°F to 220.0°F (-40.0°C to 105.0°C)
Accuracy: ±0.015% of reading ±0.02 ohms; or ±0.1°C @ 0°C for Pt100
Media:
All gases and liquids compatible with 316
stainless steel and nickel-plated brass
Optional LPT1000A Probe:
Meets PT-100 ALPHA 385 Class “A” specifications
Connectors/Ports:
Pressure - two 1/8 in. NPT
External pressure module/RS-232
RTD probe
EMI/RFI Conformance:
EN50082-1: 1992 and EN55022: 1994 Class B
Power Requirements:
Battery: Four (4) standard alkaline AA cells
Battery Life: > 35 hours, typical usage
Dimensions:
8.3 in. H x 3.9 in. W x 1.8 in. D (21.082 x 9.906 x
4.572 cm)
Accessories Included:
Test leads, carrying case, 4-AA alkaline batteries,
NIST Certificate, and instruction manual
Weights:
1 lb. 4 oz. (0.567 kg)
Above specifications 18°C to +28°C, unless otherwise noted.
Environmental Temperature:
Operating +14°F to +122°F (-10°C to +50°C)
Storage -4°F to +140°F (-20°C to +60°C)
Temperature Stability:
No effect on accuracy on all functions from
59°F to 95°F (15°C to 35°C)
Add ±0.002% F.S./°C for temps outside of
59°F to 95°F (15°C to 35°C )
12
© 2013 Rice Lake Weighing Systems February 1, 2013 Prices and specifications subject to change without notice. Visit www.4condec.com for further information.
CALIBRATION STANDARDS
Condec 3030 Hand-Held Pressure/Temperature Calibrator
PART #
DESCRIPTION
105951
CONDEC 3030 pressure/temperature
calibrator; 15/1500 PSIG
CONDEC 3030 pressure/temperature
calibrator; 30/3000 PSIG
CONDEC 3030 pressure/temperature
calibrator; 100/300 PSIG
CONDEC 3030 pressure/temperature
calibrator; 30/1000 PSIG
CONDEC 3030 pressure/temperature
calibrator; +/- 15/1 PSIG
CONDEC 3030 pressure/temperature
calibrator; 100/500 PSIG
CONDEC 3030 pressure/temperature
calibrator; 3000/10,000 PSIG
CONDEC 3030 pressure/temperature
calibrator; 3000 PSIG
CONDEC 3030 pressure/temperature
calibrator; 1500/3000 PSIG
CONDEC 3030 pressure/temperature
calibrator; 300 PSIA
CONDEC 3030 pressure/temperature
calibrator; 5/-12-30 PSIG
105952
112942
114698
120683
120684
120686
114822
129379
126772
131995
Options/Accessories
EST. SHIP
WEIGHT
PRICE
PART #
DESCRIPTION
PRICE
112958
External pressure module adaptor cable
External pressure module; 0 - 1000 psig (0.025% F.S.)
External pressure module; 0 - 3000 psig (0.05% F.S.)
External pressure module; 0 - 5000 psig (0.05% F.S.)
Additional CONDEC 3030 manual
Electrical signal test lead kit
RS-232 adapter cable
RTD Probe PR100-385
Input range -40°C to 105°C (-40°F to 220°F)
Additional soft carrying case
Output (test) port quick-disconnect adjustable
adapter assembly, male coupling body to 1/8 NPT
swivel other end, 5000 psig working pressure (for
hands-free display reading when used between
Source 3000 and CONDEC 3000 or CONDEC 3030)
Output (test) port quick-disconnect adjustable
adapter assembly, 1/8 NPT swivel to 1/4-37 Degree
AN-4 swivel other end, 5000 psig working pressure
(for hands-free display reading when used between
Orion 2C Cased unit test port and CONDEC 3030,
requires one Cheat Seal 54854)
Consult
6.5 lb
Consult
112953
6.5 lb
Consult
6.5 lb
Consult
6.5 lb
Consult
112956
112957
106375
55092
106376
6.5 lb
Consult
106410
6.5 lb
Consult
106378
6.5 lb
Consult
6.0 lb
Consult
6.5 lb
Consult
6.5 lb
Consult
6.5 lb
Consult
106594
120462
calibration standards
MODEL/PRICE
Consult
Consult
Consult
Consult
Consult
Consult
Consult
Consult
Consult
Consult
NOTE: Consult factory for additional external pressure modules
NOTE: Consult factory for additional models
Control Functions
1. Power On/Off Key -This key is used to turn the calibrator on and off.
2. Display Backlight On/Off Key -This key is used to turn the display backlight on and off.
3. Zero Key -This key is used to zero pressure measurements.
4. Soft Function Keys -These keys are used primarily to configure the calibrator.
5. UPPER Display - Indicates what is being measured and the unit of measure
(in one of 17 available pressure units). For mA and mA Loop functions, also
indicates the percent of the 4 to 20 mA span.
6. MIDDLE Display - Indicates what is being measured and the unit of measure
(in one of 17 available pressure units). For rnA and rnA Loop functions, also
indicates the percent of the 4 to 20 mA span.
7. LOWER Display - Indicates what is being measured and the unit of measure
(in one of 17 available pressure units). For rnA and rnA Loop functions, also
indicates the percent of the 4 to 20 mA span.
8. MAIN MENU Display - Functions assigned to the three Function Keys (4) appear
above each key. The functions displayed change as the operating mode and
options are selected.
9. Internal Pressure Ports - Internal Sensors PI (left) and P2 (right).
10.Input Terminals -These terminals are used to measure current, voltage, and a contact closure for switch test.
11.External Module lnput -This connection is used to interface with optional
CONDEC-P Digital Pressure Modules (Module Adapter required).
12.RTD lnput -This connection accepts a Model LTP-100A PT-100 RTD for temperature measurements.
© 2013 Rice Lake Weighing Systems February 1, 2013 Prices and specifications subject to change without notice. Visit www.4condec.com for further information.
13