OPERATION 6
IFC 100
6.2 Menu structure
Measuring mode
^
Select
menu
Select menu and/or sub-menu
Select
function
and set
data
>
Press
> 2.5 s
A quick setup
> A1 language
^
A2 Tag
>
^
A3 reset
A3.1 reset errors
A3.2 counter 1
A3.3 counter 2
A4 analogue outputs
A4.1 measurement
A4.2 unit
A4.3 range
A4.4 low flow cutoff
A4.5 time constant
A5 digital outputs
A5.1 measurement
5.2 pulse value unit
A5.3 value p. pulse
A5.4 low flow cutoff
A7 process input
A7.1 device serial no.
A7.2 zero calibration
A7.3 size
A7.5 GKL
A7.6 coil resistance Rsp
A7.7 calib. coil temp.
A7.8 target conduct.
A7.9 EF electr. factor
A7.10 field frequency
A7.11 flow direction
^
B test
> B1 simulation
^
B2 actual values
>
^
B3 information
>
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
49
6 OPERATION
Measuring mode
^
IFC 100
Select
menu
Select menu and/or sub-menu
Select
function
and set
data
>
Press
> 2.5 s
C setup
> C1 process input
^
> 1.1 calibration
^
1.2 filter
>
^
1.3 self test
1.4 information
1.5 simulation
^
> C2 I/O (Input/Output)
^
> 2.1 hardware
^
2. current output X
2.
frequency output X
2.
pulse output X
2.
status output X
2.
limit switch X
>
^
^
> C3 I/O counter
^
> 3.1 counter 1
^
3.2 counter 2
>
^
^
> C4 I/O HART
^
> 4.1 PV is
^
4.2 SV is
>
^
4.3 TV is
4.4 4V is
4.5 HART units
^
> C5 device
^
> 5.1 device info
^
5.2 display
>
^
5.3 1. meas. page
5.4 2. meas. page
5.5 graphic page
5.6 special function
5.7 units
5.8 HART
5.9 quick setup
>
50
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
6.3 Function tables
6.3.1 Menu A, quick setup
No.
Function
Settings / descriptions
language
Language selection depends on the device version.
Tag
Measuring point identifier (Tag no.) (also for HART® operation), appears in
the LC display header (up to 8 digits).
A1 language
A1
A2 Tag
A2
A3 reset?
A3
reset?
A3.1
reset errors
reset errors? Select: no/yes
A3.2
reset counter 1
reset counter? Select: no / yes (available if activated in C5.9.1)
A3.3
reset counter 2
reset counter? Select: no / yes (available if activated in C5.9.2)
A4 analogue outputs (only for HART®)
A4
analogue outputs
Applicable to current output (terminal A), frequency output (terminal D),
limit switch (terminal C and/or D) and the first display page / line 1
A4.1
measurement
1) Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
2) Use for all outputs? (also use this setting for Fct. A4.2...A4.5!)
Setting: no (applies only to the main current output ) / yes (applies to all
analogue outputs)
A4.2
unit
Selection of the unit from a list, depending on the measurement.
A4.3
range
1) Setting for main current output (range: 0...100%)
Setting: 0…x.xx (format and unit, depending on measurement, see A4.1 and
A4.2 above)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
A4.4
low flow cutoff
1) Setting for main current output (sets output value to "0")
Setting: x.xxx ± x.xxx% (range: 0.0…20%)
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
2) Use for all outputs? Make setting, see Fct. A4.1 above!
A4.5
time constant
1) Setting for main current output (applicable to all flow measurements)
Setting: xxx.x s (range: 000.1…100 s)
2) Use for all outputs? Make setting, see Fct. A4.1 above!
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
51
6 OPERATION
IFC 100
A5 digital outputs
A5
digital outputs
Valid for pulse output (terminal D) and counter 1.
A5.1
measurement
1) Select measurement: volume flow / mass flow
2) Use for all outputs? (also use this setting for Fct. A5.2...A5.4!)
Setting: no (only for pulse output D) / yes (for all digital outputs)
A5.2
pulse value unit
Selection of the unit from a list, depending on the measurement.
A5.3
value p. pulse
1) Setting for pulse output D (volume or mass value per pulse)
Setting: xxx.xxx in l/s or kg/s
2) Use for all outputs? Make setting, see Fct. A5.1 above!
A5.4
low flow cutoff
1) Setting for pulse output D (sets output value to "0")
Setting: x.xxx ± x.xxx% (range: 0.0…20%)
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
2) Use for all outputs? Make setting, see Fct. A5.1 above!
process input
A7.1
device serial no.
Serial no. of the system.
The following process input parameters are only available, if the quick access has been activated in the menu "setup /
device / quick setup.
A7.2
zero calibration
Display of current zero calibration value.
Query: calibrate zero?
Settings see Fct. C1.1.1.
A7.3
size
Select from size table.
A.7.5
GKL
Set value acc. to nameplate; range: 0.5…12 (20)
A7.6
coil resistance Rsp
Field coil resistance at 20°C; range: 10.00...220
A7.7
calib. coil temp.
The coil temperature is derived from the coil resistance at the reference
temperature.
Settings see Fct. C1.1.8.
A7.8
target conduct.
Reference value for on-site calibration; range: 1.000...50000 S/cm
A7.9
EF electr. factor
For calculation of the conductivity based on the electrode impedance.
A7.10
field frequency
Settings see Fct. C1.1.11.
Setting as on measuring sensor nameplate =
line frequency x value (from the following list):
2; 4/3; 2/3; 1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50
A7.11
flow direction
Define polarity of flow direction.
forwards (according to the arrow on the measuring sensor) or backwards (in
the opposite direction to the arrow)
52
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
6.3.2 Menu B, test
No.
Function
Settings / descriptions
B1 simulation
B1
simulation
Displayed values are simulated.
B1.1
flow speed
Simulation of the flow speed
Select: break (exit function without simulation) /
set value (range: -12…+12 m/s; unit selection in Fct. C5.7.7)
Query: start simulation?
Settings: no (exit function without simulation) / yes (start simulation)
B1.2
volume flow
Volume flow simulation, sequence and settings similar to B1.1, see above!
X stands for one of the connection terminals A, C or D
stands for Fct. no. B1.3...1.6
B1.
current output X
B1.
pulse output X
B1.
frequency out X
B1.
limit switch X
B1.
status output X
simulation X
X stands for one of the connection terminals A, C or D
Sequence and settings similar to B1.1, see above!
For pulse output a set number of pulses are output in 1 s!
B2 actual values
B2
actual values
Display the current values; exit the displayed function with key ^.
B2.1
operating hours
Display the actual operating hours; exit the displayed function with key ^.
B2.2
act. flow speed
Display the actual flow speed; exit the displayed function with key ^.
B2.3
act. coil temp.
See also Fct. C1.1.7...C1.1.8
B2.4
electr. temperature
Display the actual electronics temperature; exit the displayed function with
key ^.
B2.5
act. conductivity
See also Fct. C1.3.1...C1.3.2
B2.6
act. electr. noise
See also Fct. C1.3.13...C1.3. 15
B2.8
act. coil resistance
Display the actual resistance of the field coils depending on the current coil
temperature.
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
53
6 OPERATION
IFC 100
B3 information
B3
information
LC display
1st line: ID No. of the circuit board
2nd line: software version
3rd line: production date
B3.1
C number
CG number, cannot be changed (input/output version)
B3.2
process input
Process input part of the electronics
B3.3
SW.REV.MS
Information about the device part of the electronics (for HART ® devices,
including the HART® part) and main software
B3.4
SW.REV.UIS
Information about the user interface of the device
B3.5
"bus interface"
Only appears with Profibus, Modbus and FF.
B3.6
device serial no.
Serial no. of the system.
B3.7
electronic serial no.
Serial number of the electronics assembly
B3.8
Electronic Revision ER
Reference identification number, electronic revision and production date of
the device; includes all hardware and software changes
54
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
6.3.3 Menu C, setup
No.
Function
Settings / descriptions
C1 process input
C1.1 Calibration
C1.1
calibration
Grouping of all functions related to measuring sensor calibration.
C1.1.1
zero calibration
Display of actual zero calibration value.
calibrate zero?
Select: break (return with ^ key) /
standard (with factory setting) /
manual (display last value, set new value, range: -1.00…+1 m/s) /
range: -1.00…+1 m/s) /
automatic (shows the current value as the new zero calibration value)
C1.1.2
size
Select from size table.
C1.1.5
GKL
Set value acc. to nameplate; range: 0.5…12 (20)
C1.1.7
coil resistance Rsp
Field coil resistance at 20°C; range: 10.00...220
C1.1.8
calib. coil temp.
The coil temperature is derived from the coil resistance at the reference
temperature.
Set coil temperature:
break (return with ^ key)
standard (= 20°C)
automatic (set current temperature); range: -40.0…+200°C
Set coil resistance:
break (return with ^ key)
standard (= setting from Fct. C1.1.7)
automatic (= calibration with the current resistance)
C1.1.9
density
Calculation of mass flow with constant density of product;
range: 0.1…5 kg/l
C1.1.10
target conduct.
Reference value for on-site calibration; range: 1.000...50000 S/cm
C1.1.11
EF electr. factor
For calculation of the conductivity based on the electrode impedance.
Query: calibrate EF?
Select: break (return with ^ key) /
standard (with factory setting) /
manual (set desired value) /
automatic (determines EF according to the setting in Fct. C1.1.10)
C1.1.13
field frequency
Setting as on measuring sensor nameplate =
line frequency x value (from the following list):
1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50
C1.1.14
select settling
Select settling (special function)
Select: standard (fixed allocation) /
manual (manual time setting for the settling time for the field current)
C1.1.15
settling time
Only when "manual" selected under Fct. C1.1.14; range: 1.0…250 ms
C1.1.16
line frequency
Set line frequency.
automatic (measuring & setting; for DC systems fixed setting 50 Hz)
Select: 50 Hz or 60 Hz (fixed setting)
C1.1.17
act. coil resistance
07/2010 - 4000041004 - MA IFC 100 R04 en
Display of the actual resistance of the field coil for calculation of the
temperature.
www.krohne.com
55
6 OPERATION
IFC 100
C1.2 filter
C1.2
filter
Grouping of all functions related to filter of measuring sensor electronics.
C1.2.1
limitation
Limitation of all flow values, before smoothing by time constant, affects all
outputs.
Settings: -xxx.x / +xxx.x m/s; condition: 1st value < 2nd value
Range 1st value: -100.0 m/s
value
Range 2nd value: +0.001 m/s
C1.2.2
flow direction
value
-0.001 m/s
+100 m/s
Define polarity of flow direction.
forwards (according to the arrow on the measuring sensor) or backwards (in
the opposite direction to the arrow)
C1.2.3
time constant
C1.2.4
pulse filter
For all flow measurements and outputs.
xxx.x s; range: 0.0…100 s
Suppresses noise due to solids, air/gas bubbles and sudden changes in pH.
Select: off (without pulse filter) / on (with old pulse filter) / automatic (with
new pulse filter)
Pulse filter "on": The change from one measurement value to the next one is
limited to the value "pulse limitation" for the total time "pulse width".
This filter allows a faster signal tracking for slowly changing flow values.
Pulse filter "automatic": The raw flow values are collected into a buffer,
covering two times the "pulse width" values. This filter is called "median"
filter.
This filter allows a better suppression of pulse shaped disturbances
(particles or air bubbles in very noisy environment).
C1.2.5
pulse width
Length of interference and delays to be suppressed on sudden changes in
flow.
Only available, if pulse filter (Fct. C1.2.4) is "on" or "automatic"
xx.x s; range: 0.01...10 s
C1.2.6
pulse limitation
C1.2.7
noise filter
Dynamic limitation from one measured value to the next; only effective, if
pulse filter (Fct. C1.2.4) is "on".
xx.x s; range: 0.01...100 m/s
Suppresses noise at low conductivity, high solids content, air and gas
bubbles, and chemically inhomogeneous media.
Select: off (without noise filter) / on (with noise filter)
C1.2.8
noise level
Range within which changes are evaluated as noise, and outside of which
changes are evaluated as flow (only with noise filter switched on,
Fct. C1.2.7).
xx.xx m/s; range: 0.01...10 m/s
C1.2.9
noise suppression
Set noise suppression (only when noise filter switched on, Fct. C1.2.7).
Range: 1…10, noise suppression factor [min = 1...max = 10]
C1.2.10
low flow cutoff
Sets output value of all outputs to "0":
x.xxx ± x.xxx m/s (ft/s); range: 0.0…20 m/s (0.0...32.8 ft/s)
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C1.2.11
56
conductivity
Setting the time constant for the conductivity measurement; only available, if
the conductivity measurement is activated
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
C1.3 self test
C1.3
self test
Grouping of all functions related to self test of measuring sensor
electronics.
C1.3.1
empty pipe
Switch conductivity measurement off and on (measurement of the electrode
resistance with or without empty pipe detection).
Select: off /
conductivity (only conductivity measurement) /
cond. + empty pipe [F] (conductivity measurement and empty pipe indication,
error category [F] application);
Flow indication "= 0" when pipe empty /
cond. + empty pipe [S] (conductivity measurement and empty pipe indication,
error category [S] measurement out of specification);
Flow indication " 0" when pipe empty
C1.3.2
limit empty pipe
Only available when empty pipe activated [..] in Fct. C1.3.1.
Range: 0.0…9999 S (set max 50% of the lowest occurring conductivity in
operation. Conductivity below this value = signal as empty pipe)
C1.3.3
act. conductivity
Only available when empty pipe activated [..] in Fct. C1.3.1.
Actual conductivity is indicated. Activation takes place only after setting
mode is exited!
C1.3.13
electrode noise
Automatic test switched off / on.
Select: off /on
C1.3.14
limit electr. noise
Only with electrode noise activated, see Fct. C1.3.13.
Range: 0.000…12 m/s (noise above this threshold generates an error of
category [S])
C1.3.15
act. electr. noise
Only available when electrode noise "on" activated in Fct. C1.3.13. Activation
takes place only after the setting mode is exited!
C1.3.16
settling of field
Automatic test switched off / on.
Select: off /on
C1.3.17
diagnosis value
Select diagnosis value for testing the various analogue outputs.
Select:
off (no diagnosis) / electrode noise (activate Fct. C1.3.13) /
terminal 2 DC (electrode DC voltage at electrode terminal 2) /
terminal 3 DC (electrode DC voltage at electrode terminal 3)
C1.4 information
C1.4
information
Grouping of all functions related to information with respect to measuring
sensor and sensor electronics.
C1.4.1
liner
Shows material of the liner.
C1.4.2
electr. material
Shows material of the electrodes.
C1.4.3
calibration date
Not available at this time.
C1.4.4
serial no. sensor
Shows serial no. of the measuring sensor.
C1.4.5
V no. sensor
Shows the order number of the measuring sensor.
C1.4.6
sensor electr. info
Sows serial number of circuit board, software version number and
calibration date of circuit board
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
57
6 OPERATION
IFC 100
C1.5 simulation
C1.5
simulation
Grouping of all functions for simulating measuring sensor values. These
simulations have effect on all outputs, including counters and display.
C1.5.1
flow speed
Sequence see Fct. B1.1
C1.5.2C1.5.
2
volume flow
Sequence see Fct. B1.2
No.
Function
Settings / descriptions
C2 I/O (Inputs/Outputs)
C2.1 hardware
C2.1
hardware
Assignment of connection terminals dependent on signal converter version:
active / passive / NAMUR
C2.1.1
terminal A
Select: off (switched off) / current output
C2.1.3
terminal C
Select: off (switched off) / status output / limit switch
C2.1.4
terminal D
Select: off (switched off) / frequency output / pulse output / status output /
limit switch
C2.
current out X
C2.
current out X
X stands for connection terminals A
stands for Fct. no. C2.2 (A)
C2. .1
range 0%…100%
Current range for the selected measurement, e.g. 4...20 mA,
corresponds to 0…100%
xx.x … xx.x mA;
range: 0.00…20 mA (condition: 0 mA
C2. .2
extended range
1st value
2nd value
Defines the min. and max. limits.
xx.x…xx.x mA; range: 03.5…21.5 mA (condition: 0 mA
21.5 mA)
C2. .3
error current
20 mA)
1st value 2nd value
Specify error current.
xx.x mA; range: 3…22 mA (condition: outside of extended range)
C2. .4
Error Condition
The following error conditions can be selected.
Select: error in device (error category [F]) / application error (error category
[F]) / out of specification (error category [S])
C2. .5
measurement
Measurements for activating the output.
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C2. .6
range
0…100% of the measurement set in fct. C2. .5
0…xx.xx _ _ _ (format and unit depend on the measurement, see above)
C2. .7
polarity
Set polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2. .8
limitation
Limitation before applying the time constant.
±xxx … ±xxx%; range: -150…+150%
58
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
No.
Function
Settings / descriptions
C2. .9
low flow cutoff
Sets output value to "0"
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C2. .10
time constant
Range: 000.1…100 s
C2. .11
special function
Select: off (switched off) /
automatic range (range is changed automatically, extended lower range,
only makes sense together with a status output)
C2. .12
threshold
Appears only when Fct. C2. .11 threshold is activated between extended
and normal range. The automatic range function always changes from the
extended to the normal range when the 100% current is reached.
The upper 100% value of the hysteresis is then = 0. The threshold is then the
hysteresis value, instead of "threshold ± hysteresis" as shown in the display.
Range: 5.0…80%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C2. .13
information
Serial number of the I/O circuit board, software version number and
production date of the circuit board
C2. .14
simulation
Sequence see B1.
C2. .15
4mA trimming
Trimming of the current at 4 mA
current out X
Reset to 4 mA restores the factory calibration.
Used for HART® setting.
C2. .16
20mA trimming
Trimming of the current at 20 mA
Reset to 20 mA restores the factory calibration.
Used for HART® setting.
C2.
frequency out X
C2.
frequency out X
X stands for connection terminals D
stands for Fct. no. C2.5 (D)
C2. .1
Pulse Shape
Specify the pulse shape.
Select: symmetric (about 50% on and 50% off) / automatic (constant pulse
with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate,
setting see below Fct. C2. .3 100% pulse rate)
C2. .2
pulse width
Only available if set to "fixed" in Fct. C2. .1.
Range: 0.05…2000 ms
Note: max. setting value Tp [ms] 500 / max. pulse rate [1/s], gives the pulse
width = time where the output is activated
C2. .3
100% pulse rate
Pulse rate for 100% of the measuring range.
Range: 0.0…10000 1/s
Limitation 100% pulse rate 100/s: Imax
Limitation 100% pulse rate > 100/s: Imax
C2. .4
measurement
100 mA
20 mA
Measurements for activating the output.
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C2. .5
rangemeasurement range 0…100% of the measurement set in fct. C2. .4
0…xx.xx _ _ _ (format and unit depend on the measurement, see above)
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
59
6 OPERATION
IFC 100
No.
Function
Settings / descriptions
C2. .6
polarity
Set polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2. .7
limitation
C2. .8
low flow cutoff
Limitation before applying the time constant.
±xxx … ±xxx%; range: -150…+150%
Sets output value to "0":
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C2. .9
time constant
Range: 000.1…100 s
C2. .10
invert signal
Select:
off (activated output generates a high current at the output, switch closed) /
on (activated output generates a low current at the output, switch open)
C2. .12
information
Serial number of the I/O circuit board, software version number and
production date of the circuit board
C2. .13
simulation
Sequence see B1.
C2.
frequency out X
pulse output X
C2.
pulse output X
X stands for connection terminals D
stands for Fct. no. C2.5 (D)
C2. .1
pulse shape
Specify the pulse shape.
Select: symmetric (about 50% on and 50% off) / automatic (constant pulse
with about 50% on and 50% off at 100% pulse rate) / fixed (fixed pulse rate,
setting see below Fct. C2. .3 100% pulse rate)
C2. .2
pulse width
Only available if set to "fixed" in Fct. C2. .1
Range: 0.05…2000 ms
Note: max. setting value Tp [ms] 500 / max. pulse rate [1/s], gives the pulse
width = time where the output is activated
C2. .3
max. pulse rate
Pulse rate for 100% of the measuring range.
Range: 0.0…10000 1/s
Limitation 100% pulse rate 100/s: Imax
Limitation 100% pulse rate > 100/s: Imax
C2. .4
measurement
100 mA
20 mA
Measurements for activating the output.
Select: volume flow / mass flow
C2. .5
pulse value unit
Selection of the unit from a list, depending on the measurement.
C2. .6
value p. pulse
Set value for volume or mass per pulse.
xxx.xxx, range in [l] or [kg] (volume or mass for current output C2. .6)
At max. pulse rate see above 2. .3 pulse output.
C2. .7
polarity
Set polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2. .8
low flow cutoff
Sets output value to "0"
x.xxx ± x.xxx%; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C2. .9
60
time constant
Range: 000.1…100 s
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
No.
Function
Settings / descriptions
C2. .10
invert signal
Select:
off (activated output generates a high current at the output, switch closed) /
on (activated output generates a low current at the output, switch open)
C2. .12
information
Serial number of the I/O circuit board, software version number and
production date of the circuit board
C2. .13
simulation
Sequence see B1.
C2.
pulse output X
status output X
C2.
status output X
X (Y) stands for one of the connection terminals C or D
stands for Fct. no. C2.4 (C) / C2.5 (D)
C2. .1
Mode
The output shows the following measuring conditions:
out of specification (output activated, signals application error or error in
device refer to
on page 70 /
application error (output activated, signals application error or error in
device refer to
on page 70 /
polarity flow (polarity of the current flow) /
over range flow (over range of the flow) /
counter 1 preset (activates when counter X preset value is reached) /
counter 2 preset (activates when counter X preset value is reached) /
output A (activated by the status of output Y, additional output data see
below) /
output C (activated by the status of output Y, additional output data see
below) /
output D (activated by the status of output Y, additional output data see
below) /
off (switched off) /
empty pipe (when pipe empty, output activated) /
error in device (when error, output activated)
C2. .2
current out Y
Only appears if output A is set under "mode (see above)", and this output is a
"current output".
Select:
polarity (is signalled) /
over range (is signalled) /
automatic range signals lower range
C2. .2
frequency out Y and pulse
output Y
Only appears if output D is set under "mode (see above)", and this output is a
"frequency/pulse output".
Select:
polarity (is signalled) /
over range (is signalled)
C2. .2
status output Y
Only appears if output C or D is set under "mode (see above)", and this
output is a "status output".
Same signal (like other connected status output, signal can be inverted, see
below)
C2. .2
limit switch Y
Only appears if output C or D is set under "mode (see above)", and this
output is a "limit switch".
Status off (is always selected here if status output X is connected with a limit
switch.
C2. .2
off
Only appears if output A, C or D is set under "mode (see above)" and this
output is switched off.
C2. .3
invert signal
Select:
off (activated output supplies a high current, switch closed) /
on (activated output supplies a low current, switch open)
C2. .4
information
Serial number of the I/O circuit board, software version number and
production date of the circuit board
C2. .5
simulation
Sequence see B1.
07/2010 - 4000041004 - MA IFC 100 R04 en
status output X
www.krohne.com
61
6 OPERATION
C2.
IFC 100
limit switch X
C2.
limit switch X
X stands for one of the connection terminals C or D
stands for Fct. no. C2.4 (C) / C2.5 (D)
C2. .1
measurement
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C2. .2
threshold
Switching level, set threshold with hysteresis
xxx.x ±x.xxx (format and unit depend on the measurement, see above)
(1st value = threshold / 2nd value = hysteresis),
condition: 2nd value 1st value
C2. .3
polarity
Set polarity, please note flow direction in C1.2.2!
Select: both polarities (plus and minus values are displayed) / positive
polarity (display for negative values = 0) / negative polarity (display for
positive values = 0) / absolute value (use for the output)
C2. .4
time constant
Range: 000.1…100 s
C2. .5
invert signal
Select:
off (activated output generates a high current, switch closed) /
on (activated output generates a low current, switch open)
C2. .6
information
Serial number of the I/O circuit board, software version number and
production date of the circuit board
C2. .7
simulation
Sequence see B1.
62
limit switch X
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
No.
Function
Settings / descriptions
C3 I/O counter
C3.1
counter 1
Set function of counter
stands for 1, 2 (= counter 1, 2)
The basic version (standard) has only 2 counters!
These functions are only available for HART® devices.
C3.2
counter 2
C3. .1
function
Select: sum counter (counts positive and negative values) / +counter (counts
only the positive values) / -counter (counts only the negative values) / off
(counter is switched off)
C3. .2
measurement
Selection of the measurement for counter
Select: volume flow / mass flow
C3. .3
low flow cutoff
Sets output value to "0"
Range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C3. .4
time constant
Range: 000.1…100 s
C3. .5
preset value
If this value is reached, positive or negative, a signal is generated that can be
used for a status output at which "preset counter X" has to be set.
Preset value (max. 8 digits) x.xxxxx in selected unit, see C5.7.10 + 13
C3. .6
reset counter
Sequence see Fct. A3.2 and A3.3
C3. .7
set counter
Set counter
to the desired value.
Select: break (exit function) / set value (opens the editor to make the entry)
Query: set counter?
Select: no (exit function without setting the value) / yes (sets the counter and
exits the function)
C3. .8
stop counter
Counter
stops and holds the current value.
Select: no (exits the function without stopping the counter) / yes (stops the
counter and exits the function)
C3. .9
start counter
Start counter
after that counter is stopped.
Select: no (exits the function without starting the counter) / yes (starts the
counter and exits the function)
C3. .10
information
07/2010 - 4000041004 - MA IFC 100 R04 en
Serial number of the I/O circuit board, software version number and
production date of the circuit board
www.krohne.com
63
6 OPERATION
No.
IFC 100
Function
Settings / descriptions
I/O HART
Selection / display of the 4 dynamic variables (DV) for HART®.
C4 I/O HART
C4
The HART® current output (terminal A basic I/Os) always has a fixed link to
the primary variables (PV). Fixed links of the other DVs (1-3) are only
possible if one additional analogue output (frequency) is available; if not, the
measurement can be freely selected from the following list: see Fct. A4.1
"measurement".
stands for 1, 3 or 4
X stands for connection terminals A, C or D
C4.1
PV is
Current output (primary variable)
C4.2
SV is
(secondary variable)
C4.3
TV is
(tertiary variable)
C4.4
4V is
(4th variable)
C4.5
HART units
Changes units of DVs (dynamic variables) in the display
Break: return with ^ key
HART display®: copies the settings for the display units to the settings for
DVs
Standard: sets factory defaults for DVs
C4. .1
current out X
Shows the current analogue measured value of the linked current output.
The measurement cannot be changed!
C4. .1
frequency out X
Shows the current analogue measured value of the linked frequency output,
if present. The measurement cannot be changed!
C4. .1
HART dynamic var.
Measurements of the dynamic variables for HART®.
Linear measurements: volume flow / mass flow / diagnosis value / flow
speed / coil temperature / conductivity
Digital measurements: counter 1 / counter 2 / operating hours
No.
Function
Settings / descriptions
C5 device
C5.1 device info
C5.1
device info
Grouping of all functions that have no direct effect on the measurement or
any output.
C5.1.1
Tag
Settable characters (max. 8 digits):
A…Z; a…z; 0…9; / - , .
C5.1.2
C number
CG number, non-alterable (input/output versions)
C5.1.3
device serial no.
Serial no. of the system.
C5.1.4
electronic serial no.
Serial no. of the electronic assembly, cannot be changed.
C5.1.5
SW.REV.MS
Serial no. of the circuit board, version no. of the main software, production
date of the circuit board
C5.1.6
Electronic Revision ER
Reference identification number, electronic revision and production date of
the device; includes all hardware and software changes
64
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
C5.2 display
C5.2
display
-
C5.2.1
language
Language selection depends on the device version.
C5.2.2
contrast
Adjust display contrast for extreme temperatures.
Setting: -9…0…+9
C5.2.3
default display
This change takes place immediately, not just when setting mode is exited!
Specification of the default display page that is returned to after a short
delay period.
Select:
none (the current page is always active) / 1. meas. page (show this page) /
2. meas. page (show this page) / status page (show only status messages) /
graphic page (trend display of the 1st measurement)
C5.2.5
SW.REV.UIS
Serial no. of the circuit board, version no. of user software, production date
of the circuit board
C5.3 & C5.4 1. meas. page & 2. meas. page
C5.3
1. meas. page
C5.4
2. meas. page
C5. .1
function
stands for 3 = meas. page 1 and 4 = meas. page 2
Specify number of measured value lines (font size)
Select: one line / two lines / three lines
C5. .2
measurement 1.line
Specify measurement for 1st line.
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity
C5. .3
range
C5. .4
limitation
0…100% of the measurement set in Fct. C5. .2
0…xx.xx _ _ _ (format and unit depend on the measurement)
Limitation before applying the time constant.
xxx%; range: -120…+120%
C5. .5
low flow cutoff
Sets output to "0": x.xxx ± x.xxx %; range: 0.0…20%
(1st value = switching point / 2nd value = hysteresis),
condition: 2nd value 1st value
C5. .6
time constant
Range: 000.1…100 s
C5. .7
format 1.line
Specify decimal places.
Select: automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 digits)
C5. .8
measurement 2.line
Specify measurement 2.line (only available if this 2.line is activated)
Select: bar graph (for the measurement selected in 1.line) / volume flow /
mass flow / diagnosis value / flow speed / counter 1 / counter 2 / conductivity
/ coil temperature / operating hours
C5. .9
format 2.line
Specify position after decimal point (only available if this 2.line is activated)
Select: automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 digits)
C5. .10
measurement 3.line
Specify measurement 3.line (only available if this 3.line is activated)
Select: volume flow / mass flow / diagnosis value / flow speed / coil
temperature / conductivity / counter 1 / counter 2 / operating hours
C5. .11
format 3.line
Specify position after decimal point (only available if this 3.line is activated)
Select: automatic (adaptation carried out automatically) /
X (= none) …X.XXXXXXXX (max. 8 digits)
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
65
6 OPERATION
IFC 100
C5.5 graphic page
C5.5
graphic page
Graphic page always shows trend curve of the measurement of the 1. meas.
page / 1.line, see Fct. C5.3.2
C5.5.1
select range
Select: manual (set range in Fct. C5.5.2) /
automatic (automatic depiction based on the measured values)
Reset only after parameter change or after switching off and on.
C5.5.2
range
Set the scaling for the Y axis. Only available if "manual" is set in C5.5.1.
+xxx ±xxx%; range: -100…+100%
(1st value = lower limit / 2nd value = upper limit),
condition: 1st value 2nd value
C5.5.3
time scale
Set the time scaling for the X axis, trend curve
xxx min; range: 0…100 min
C5.6 special function
C5.6
special function
-
C5.6.1
reset errors
reset errors?
Select: no/yes
C5.6.2
save settings
Save current settings.
Select: break (exit function without saving) / backup 1 (save in storage
location 1) / backup 2 (save in storage location 2)
Query: continue copy? (cannot be done afterwards)
Select: no (exit function without saving) / yes (copy current settings to
storage backup 1 or backup 2)
C5.6.3
load settings
Load saved settings.
Select: break (exit function without loading) / factory settings (load in state
as delivered) / backup 1 (load data from storage location 1) / backup 2 (load
data from storage location 2) / load sensor data (factory settings of
calibration data)
Query: continue copy? (cannot be done afterwards)
Select: no (exit the function without saving) / yes (load data from the selected
storage location)
C5.6.4
password quick set
Password required to change data in the quick setup menu.
0000 (= to Quick Setup menu without password)
xxxx (password required); range 4digit: 0001…9999
C5.6.5
password setup
Password required to change data in the setup menu.
0000 (= to Quick Setup menu without password)
xxxx (password required); range 4digit: 0001…9999
66
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
C5.7 units
C5.7
units
-
C5.7.1
volume flow
m³/h; m³/min; m³/s; l/h; l/min; l/s (l = litres);
ft³/h; ft³/min; ft³/s; gal/h; gal/min; gal/s;
IG/h; IG/min; IG/s;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.2
Text free unit
For text to be specified refer to
C5.7.3
[m³/s]*factor
Specification of the conversion factor, based on m³/s:
xxx.xxx refer to
on page 68:
on page 68
C5.7.4
mass flow
kg/s; kg/min; kg/h; t/min; t/h; g/s; g/min; g/h;
lb/s; lb/min; lb/h; ST/min; ST/h (ST = Short Ton); LT/h (LT = Long Ton);
free unit (set factor and text in the next two functions, sequence see below)
C5.7.5
Text free unit
For text to be specified refer to
C5.7.6
[kg/s]*factor
Specification of the conversion factor, based on kg/s:
C5.7.7
flow speed
m/s; ft/s
C5.7.8
conductivity
µS/cm; S/m
C5.7.9
temperature
°C; °F; K
C5.7.10
volume
m³; l (litre); hl; ml; gal; IG; in³; ft³; yd³;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.11
Text free unit
For text to be specified refer to
C5.7.12
[m³]*factor
Specification of the conversion factor, based on m³:
xxx.xxx refer to
xxx.xxx refer to
on page 68:
on page 68
on page 68:
on page 68
C5.7.13
mass
kg; t; mg; g; lb; ST; LT; oz;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.14
Text free unit
For text to be specified refer to
C5.7.15
[kg]*factor
Specification of the conversion factor, based on kg:
xxx.xxx refer to
on page 68:
on page 68
C5.7.16
density
kg/l; kg/m³; lb/ft³; lb/gal;
free unit (set factor and text in the next two functions, sequence see below)
C5.7.17
Text free unit
For text to be specified refer to
C5.7.18
[kg/m³]*factor
Specification of the conversion factor, based on kg/m³:
xxx.xxx refer to
07/2010 - 4000041004 - MA IFC 100 R04 en
www.krohne.com
on page 68:
on page 68
67
6 OPERATION
IFC 100
C5.8 HART
C5.8
HART
This function is only available for devices with a HART® interface!
C5.8.1
HART
Switch HART® communication on/off
Select:
on (HART® active) current = 4…20 mA /
off (HART® not active) current = 0…20 mA
C5.8.2
address
Set address for HART® operation.
Select: 00 (point-to-point operation, current output has normal function,
current = 4…20 mA) / 01...15 (Multi-Drop operation, current output has a
constant setting of 4 mA)
C5.8.3
message
C5.8.4
description
Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
Set required text:
A…Z ; a…z ; 0…9 ; / - + , . *
C5.9 quick setup
C5.9
quick setup
Activate quick access in quick setup menu; default setting: quick setup is
active (yes)
Select: yes (activated) / no (not activated)
C5.9.1
reset counter 1
Reset counter 1 in quick setup menu
Select: yes (activated) / no (not activated)
C5.9.2
reset counter 2
C5.9.4
process input
Reset counter 2 in quick setup menu
Select: yes (activated) / no (not activated)
Activate quick access to the important process input parameters
Select: yes (activated) / no (not activated)
6.3.4 Set free units
Free units
Sequences to set texts and factors
Texts
Volume flow, mass flow and density:
3 digits before and after the slash
xxx/xxx (max. 3 digits before / after the slash)
Volume, mass:
xxx (max. 3 digits)
Permissible characters:
A…Z ; a…z ; 0…9 ; / - + , .*; @ $ % ~ () [] _
Conversion factors
Desired unit
= [unit see above] * conversion factor
Conversion factor
Max. 9 digits
Shift decimal point:
68
to the left and
to the right
www.krohne.com
07/2010 - 4000041004 - MA IFC 100 R04 en
OPERATION 6
IFC 100
6.4 Description of functions
6.4.1 Reset counter in quick setup menu
Button
Display
Description and setting
>
quick setup
Press and hold for 2.5 s, then release the key.
>
language
2x
reset
>
reset errors
counter 1
Select desired counter.
counter 2
counter 3
>
reset counter
no
or
reset counter
yes
^
counter 1,2
3x^
Measuring mode
Counter has been reset.
6.4.2 Deleting error messages in the quick setup menu
.
Key
Display
Description and setting
>
quick setup
Press and hold for 2.5 s, then
release the key.
>
language
2x
reset
>
reset errors
>
reset?
no
or
reset?
yes
^
reset errors
3x^
Measuring mode
07/2010 - 4000041004 - MA IFC 100 R04 en
Error has been reset.
www.krohne.com
69
Download PDF