LifeGard DS
The Most Versatile Patient Monitor in its Class
L I F E G A R D
I
F A M I L Y
D a t a
• 4-hour typical battery life; 8 hours in batterysave mode for prolonged transport.
• Remarkable simplicity of use; intuitive and easy
to learn.
• Bright color display with exceptional data
presentation and view angle.
• Spot-check and continuous monitoring modes.
• Manual or auto-set alarms (inactive in spotcheck mode).
• Manual and auto-interval NIBP modes.
• Data output for connection to LIFEGARD Vue
Central Alarm Annunciation System.
• Weighs less than 5 lbs.
Specifications
Leads: 3 leads, user selectable
Heart Rate Range: 20 to 250 bpm
Heart Rate Accuracy: ±1 bpm or ±1%
Bandwidth-Normal Monitoring: 0.5 Hz to 40 Hz
Bandwidth-Extended Low Frequency Range
(user selectable): 0.05 Hz to 40 Hz
Display Sweep Speeds: 12.5, 25, 50 mm/sec
ECG Size (sensitivity): 0.5, 1, 2, 4 mV/cm
Lead-Off Detection with Display Indicator
Pacemaker Detection Mode: Indicator on
waveform display, user selectable
Standards: ANSI/AAMI EC 13
Differential Input Impedance: > 5 MΩ
Common Mode Rejection Ratio:
> 90 dB at 50 or 60 Hz
Input Dynamic Range: ±5 mV AC, ±300 mV DC
Defibrillation Artifact Recovery Time:
< 8 sec conforming to AAMI EC 13
Blood Pressure Accuracy: Meets performance standards
of ANSI/AAMI SP10(1)
Technique: Oscillometric
Pulse Rate Range: 40 to 200 bpm
Maximum Cuff Pressure: Safety limited to 300 mmHg
Pulse Rate Accuracy: ±2 bpm or ±2%
Initial Cuff Inflation is User Adjustable; Factory Default:
160 mmHg (adult mode); 120 mmHg (pediatric mode)
Subsequent Cuff Inflation: Previous systolic ±30 mmHg
Blood Pressure Measurement Range
Systolic: 40 to 250 mmHg
Diastolic: 20 to 200 mmHg
Mean Arterial Pressure: 27 to 220 mmHg
NIBP Performance
Measurement Modes
Auto-Interval: Automatic measurements at intervals
of 1, 3, 5, 10, 15, 30, 60, and 90 minutes
Manual: Single measurement initiated by Start/Stop
button
STAT: Series of consecutive measurements for 5
minutes
S h e e t
P R E L I M I N A R Y
• 4 Non-invasive measurements –
ECG (HR), SpO2 , NIBP, and Predictive
Temperature (optional)
ECG Performance
L I F E G A R D
The Analogic LIFEGARD® I offers the best
features of both a spot-check monitor and a
continuous vital signs patient monitor. Ideal for
the general care floor, emergency department,
triage areas, recovery areas, and office-based
surgery, the LIFEGARD I sets a new standard
for versatility.
Specifications (continued)
Nellcor® Technology
% Saturation Range: 0% to 100%
Pulse Rate Range: 20 to 250 bpm
SpO2 Accuracy: 70% to 100% ±3 digits
with DS100A sensor
Pulse Rate Accuracy: ± 3 bpm
SpO2 Performance
Welch Allyn SureTemp™ Technology
Range: 84°F to 108°F (28.9°C to 42.2°C)
Accuracy: ±0.2°F or ±0.1°C
Electronic Predictive Thermometer
Response Time: oral – 4 seconds;
Monitor-continuous readings:
axillary – 10 seconds; rectal – 15 seconds
Temperature Performance
Memory Storage: 12 hours
Data Interval: 20 seconds
Display Range: 2 hours, 12 hours scrollable
Tabular Format: One table for all variables;
six fields per row (time and 4 vital signs)
Display Interval: Per NIBP measurement or
15 minutes if no NIBP, or during an alarm
condition
Stored Data: Trend data for HR, SpO2, NIBP, and
Temperature
Graphical and Tabular Trends
RS-232 digital output from the serial port for one
of the following:
nurse call
transferring data to an external computer
Interfaces
100 to 240 VAC, 50/60 Hz, 1.5A
Power
Display
TFT color-active matrix
Screen Size: 3.9" x 2.9" (101 mm x 75 mm)
5" (126 mm) diagonally
Resolution: 320 x 240 pixels
View Angles: 65° left / right, 40° up, 65° down
Backlight Life: 20,000 hours
Waveforms
Up to 3 traces (selectable in any combination ECG,
SpO2 plethysmograph, or respiration)
Indicators
Categorized alarms
(3 priority levels with different tone pitch)
Visual alarm notification
Heart rate tone
Battery status
External power LED
Battery (standard)
Internal battery: sealed lead-acid
Battery status indicator
Operating Time: 4 hours typical
(fully charged battery)
Optional External Thermal Printer:
Charge Time: <8 hours
Speeds: 12.5, 25 and 50 mm/sec
Weight: 2 lb (0.9 kg)
Paper Width: 1.9" (50 mm)
Internal Battery: sealed lead-acid
Monitor Performance
CE marking: Medical Device Directive, 93/42/EEC
Safety Standards: IEC 60601-1, IEC 60601-1-1,
UL 60601-1, CAN/CSA C22.2 No 601.1 (M90),
EN 60601-1
IEC/EN 60601-1-4
IEC/EN 60601-1-8
IEC/EN 60601-2-27
IEC/EN 60601-2-30
IEC/EN 60601-2-49
ISO 9919
EN 1060-1
EN 1060-3
ANSI/AAMI SP10
Water Resistance: IEC 60529 classification IPX1,
monitor only
Mechanical Shock: IEC 60068-2-27
Environmental Humidity: IEC 60068-2-78
Environmental Temperature: IEC 60068-2-1,
IEC 60068-2-2
Electromagnetic Compatibility: IEC 60601-1-2
Mechanical: IEC 60068-2-6
UL60601-1, CAN/CSA-22.2 No. 601.1 (M90)
Compliance and Approvals
Specifications subject to change without notice.
Weight: 5 lb (2.1 kg) excluding
accessories, options, cables
Size: 10.5" x 6.2" x 5.6"
(26.7 cm x 15.7 cm x 14.2 cm)
Physical
Operating Temperature: 32°F to 122°F (0°C to 50°C)
Storage Temperature: – 4°F to 140°F (– 20°C to 60°C)
Operating/Storage Humidity: 5% to 90% relative
humidity, non-condensing
Operating Altitude: 0 to 10,000 ft. (0 to 3,048m)
Monitor Environmental
(1) Systolic and Diastolic blood pressure measurements
determined with this device are equivalent to those
obtained by a trained observer using the cuff/stethoscope auscultation method, and mean pressure measurements determined with this device are equivalent to
those obtained by an intra-arterial blood pressure
measurement device, within the limits prescribed by the
American National Standards Institute and Association
for the Advancement of Medical Instrumentation
(Electronic or Automated Sphygmomanometers).
LIFEGARD is a registered trademark of Analogic Corporation.
Printed in U.S.A. © 2006
16-00317 Rev 00
Analogic Corporation
Life Care Systems Division
8 Centennial Drive
Peabody, MA 01960, USA
Tel: (978) 326-4000
(800) 599-7769
Fax: (978) 977-6854
www.analogic.com/lifegard