Panasonic
Oberarm-Blutdruckmessgerat
Upper Arm Blood Pressure Monitor
Bedienungsanleitung
Operating Instructions
Model No. EW3122
Lesen Sie diese Bedienungsanleitung vor Inbetriebnahme des Geräts vollständig durch und bewahren Sie sie sorgfältig auf.
Before operating this device, please read these instructions completely and save this manual for future use.
Avant d'utiliser ce dispositif, veuillez lire entièrement ces instructions et rangez soigneusement ce manuel pour le consulter par la suite.
Prima di mettere in funzione questo apparecchio leggere completamente le presenti istruzioni, e conservare questo manuale per l'utilizzo
futuro.
Antes de utilizar este dispositivo, lea completamente estas instrucciones y conserve este manual para futura referencia.
Перед использованием прибора, пожалуйста прочтите полностью эти инструкции и сохраните данное руководство для
использования на будущее.
Bu cihazi kullanmadan ónce litfen talimatlarin tamamint okuyunuz ve bu kullanim kilavuzunu daha sonra kullanmak üzere muhafaza ediniz.
Bedienungsanleitung
Operating Instructions
Mode d'emploi
Istruzioni d'uso
Instrucciones de funcionamiento
Инструкция по эксплуатации
Kullanma Talimatlari
DE1-22
EN1-22
FR1-22
[T1-22
ES1-22
RUS1-22
TR 1-22
Dear customer, congratulation to the purchase of Panasonic blood pressure
monitor. For more than a decade, Matsushita has been providing blood
pressure measuring device to our customers in Europe. The product you
have purchased has passed our strict quality control, and we hope that this
can help you manage your daily blood pressure.
We wish you all the best for your health.
PARTS IDENTIFICATION
Please check the instrument.
Pressure cuff
plug
Pressure cuff socket
Display
Set button
7 Cover
Blood pressure
indicator lamps
Connector
(for AC adapter)
Body movement
indicator lamp
Adjust button
Memory/Recall
Tube button 2
Start/Stop button
Memory/Recall
button 1
Pressure cuff
EN1
Bottom of main body
Battery cover
How to open the cover
(1) Press lightly on
the Open/Close button.
(2) Open the cover while
holding the Open/Close
button pressed.
Optional accessory
OAC Adapter
EW31225
—
I
A
y
Pressure cuff storage
example
OLarge Cuff
(35cm - 42cm)
EW3900
Open/Close button
EN2
INSTALLING BATTERIES
+ Batteries are not included. Please purchase them separately.
* Always use alkaline batteries. (Four AA-size LR6 alkaline batteries)
Turn the main body over, press
lightly on the A mark on the
battery cover, and
remove the cover by sliding
it in the direction of the arrow.
Battery cover
Check the polarities of the
batteries and install them
correctly.
3 Replace the battery cover and
slide it closed.
CAUTION:
Always follow the cautions printed on the batteries.
Check the polarities of the batteries and install them correctly.
Remove exhausted batteries promptly.
If the product will not be used for a long time, remove the batteries.
Batteries should be used before their expiration date.
Use after the expiration date may result in injury or staining of the
surrounding area due to generation of heat, bursting, or leakage.
Replace batteries when:
e The mark shown at left appears during use.
e All displays are lit for a moment after the Start/Stop
button is pressed, and then the display goes blank.
e No display appears even if the Start/Stop button is
pressed.
XI
e Approximately 500 measurements (approximately 5 months when measurements
are taken 3 times per day) can be taken with a set of fresh Panasonic alkaline
batteries (four AA-size LR6 batteries). (Measurement conditions: Room temperature
22°C; Pressurized to 170mmHg; Upper arm circumference: 30cm)
e When batteries other than Panasonic alkaline batteries are used, battery life may be
shorter. When manganese batteries or other alkaline batteries are used, battery life
becomes as short as 1/6th the life of Panasonic alkaline batteries.
e Be sure to replace batteries once a year. If battery replacement is delayed, damage
due to battery leakage may occur.
eo When replacing batteries, always use four fresh alkaline batteries of the same type
from the same manufacturer.
* When room temperature is low, battery life may become shorter.
EN3
WHEN USING THE AC ADAPTER
* Use only the exclusive AC Adapter (EW31225). (See p. 2.)
1 Insert the AC adapter plug into the connector on the side of the
main body.
2 Plug the AC adapter fully into the AC outlet.
CAUTION:
Be sure to use only the exclusive AC adapter (sold separately).
Use of other AC adapters could cause an accident.
Caution when using only the AC adapter
When no batteries are installed in the main body and the AC adapter is
unplugged from the AC outlet or main body, the set date and time will be
reset and deleted. Because of this, even when using the AC adapter, it is
recommended that batteries be installed in the main body.
When the AC adapter is used, power is supplied from the AC adapter, so
virtually no power from the batteries is used.
EN4
BEFORE STARTING MEASUREMENTS
Setting the date and time
* Please set the date and time before taking the first measurement after purchasing
the blood pressure monitor or after changing the batteries.
e Although measurements can be taken without setting the date and time, if
measurements are taken without performing these settings, the date and time will
not be stored.
e Please read the "Caution when using only the AC adapter" on p. 4.
1 Press the Set button below the display.
Bi PT 2 The date value will blink.
7 555) Set the date.
[ / | 000 Press the Change button repeatedly to set the
correct day.
Set the month.
© Press the Set button. The month value will blink.
O Press the Adjust button repeatedly to set the
correct month.
Set the hour.
© Press the Set button. The hour value will blink.
© Press the Adjust button to set the correct hour.
[ 8/1. 0:06)
[ 8/3 090)
Set the minute.
© Press the Set button. The minute value will blink.
© Press the Adjust button repeatedly to set the
correct minute.
(8/31 = 5:37) Press the Set button at the exact time the minute
E/ = Г: m =
changes tO finalize the datetime setting.
8/ 3 15:00
7 Press the Start/Stop button to switch
Off the power.
When the date or time are blinking, the clock is stopped.
During leap years, 1/3 will be displayed on February 29. For this reason, reset the date
to 29/2.
ENS
POSITIONING THE PRESSURE CUFF
<>) 1 removethe.
| pressure cuff from the
main body. (See p. 2.)
2 Insert the plug into the
socket.
e Insert the pressure cuff plug
all the way in.
3 Wrap around upper arm.
e Position the tube outlet
so that it is on the palm
side of the arm in line
with the little finger of
the hand.
e Wrap around bare arm.
NN
f The blood pressure monitor can
be used by wrapping the cuff on
either upper arm. However,
since it has been said that the
difference in blood pressure
between the left and right arms
may be around 10mmHg, be
sure to always use the same
am for measurements. В
Fasten so that there is no
gap between the cuff and the
arm, but be careful that the cuff
is not too tight.
e Be careful not to catch your
clothes when wrapping the
cuff. (Doing so may lead to
Approximately 2
to 3cm from lower accuracy.)
inside of elbow + Press the tape fastener
down well.
ENG
FOR ACCURATE MEASUREMENTS
Take measurements while sitting quietly and relaxing.
1 Rest your elbows
on a table, etc.
Position your arm so that
the pressure cuff is at the
same height as your
heart.
3 Position your hand palm
upward, and
relax your hand.
Take measurements using the upper arm on the same
side in the same position at the same time every day.
e Measurement values will vary according to the height of the pressure cuff.
e Take measurements in the morning immediately after you wake up, when
there is no urge to urinate. If this is difficult, take measurements before
breakfast and before you have moved a lot.
Relax and sit in a natural posture when taking
measurements.
e Always sit quietly for approximately 4 to 5 minutes before taking
measurements.
Relax your body, arm, and fingers and do not move their muscles.
Do not talk while taking measurements.
Do not use a cellular phone near the main body.
During measurement, do not touch the main body or the tube.
[CAUTION]
Keep away from cellular telephones and other sources of electromagnetic
radiation. Failure to do so could result in malfunctioning.
EN7
CAUTIONS DURING MEASUREMENTS
In the following situations, accurate measurements
cannot be taken.
OWithin 1 hour of eating or drinking alcohol
Olmmediately after drinking coffee or tea, or smoking
Olmmediately after exercising or taking a bath
-Wait at least 20 minutes and take a measurement while staying quiet.
Oln a moving vehicle
Oln cold places
-Take measurements at ambient temperatures of 10°C to 40°C.
OWhen you need to urinate
-Wait several minutes after urinating before taking measurements.
ENS
TAKING A READING
Measurement
y
Los
ee, CO
CO.
> e,
Las
| [
JN
| |
LL
|
JN
The entire display will
light for approximately
1 second.
e |[fthere is air left in
the pressure cuff,
the fully lit time will
be longer.
Beep
Beep
Beep
Beep
EN9
Press the Start/Stop button.
e The pressure cuff will automatically be pressurized
and a measurement will be taken.
87 3530
! ЖЖ
e If your body moves during
Jeu uy,
у
The heart mark will
blink when pulse is
detected during
pressurization.
measurement, the body movement
( “| indicator lamp will blink. (See p. 10.)
dV “<A
When measurement has been completed, a
series of beeps will sound and the blood
pressure and pulse values
will be displayed.
[EYE]
Normal-High
Normal
Hypertension
Normal-High
Normal
e Pressure will be automatically released from the pressure
cuff.
e The blood pressure indicator lamps will light (for
approximately 6 seconds).
e |fthe measured value is in the high blood pressure region,
the values will blink (for approximately 6 seconds). (For
how to read the blood pressure display, see p. 11.)
e The (AM marks will light.
e The (M) mark will blink. (For storing measurement
values, see p. 12.)
* The pulse rate value is the beats per minute calculated
based on the pulse rate during measurement.
To switch off the blood pressure monitor,
PreSS the Start/Stop button.
If you forget to switch it off, the blood pressure
monitor will switch off automatically after
approximately 5 minutes.
TAKING A READING (continuep)
About the body movement indicator function
If your arm (to which the pressure cuff is attached), etc. moves during measurement
(such as if you unconsciously bend your elbow), it may not be possible to take
measurements correctly.
e The body movement indicator function uses the body movement indicator lamp and
a warning sound to tell you when movement has occurred or when excessive
pressure has been applied to the pressure cuff during measurement.
e If the body movement indicator lamp blinks, return your arm to the correct posture.
(See p. 7.)
e When body movement has been detected, the body movement indicator lamp will
blink when the blood pressure judgment display appears. (The lamp will also blink
when measurements for which movement was detected are stored and later
recalled.)
Body movement
indicator lamp
display
O
Blinking
O
Lit
Warning sound
Beeeep beeeep
(long buzzer
sounds)
Beep beep
(short buzzer
sounds)
Explanation
Body movement was detected.
e You unconsciously bent your elbow.
e The upper arm became very tense.
e You moved your arm.
e Etc.
Return to the proper posture. (See p. 7.)
A movement which will cause large differences in
measurement values occurred.
e Body movement was large.
e Body movement occurred several times.
e Arm movement was large.
Remeasure (Remeasure mark E is shown.)
* It will also be lit in the following cases:
¢ Pressure cuff tape fastener became
unfastened.
e Tube came out.
e Etc.
(For details, see p. 21.)
* This function is intended to help ensure correct measurements. (See p. 18.)
Even if the remeasure mark does not appear (or body movement indicator lamp does
not light), it is recommended that 2 or 3 measurements be taken in a quiet condition
for best results.
EN10
Reading the blood pressure display
Whether the systolic and diastolic blood pressure values are in the high
blood pressure region or normal blood pressure region are judged based on
the WHO/ISH (World Health Organization/International Society of
Hypertension) blood pressure classifications, and the blood pressure value
blinks if required. WHO/ISH blood pressure
Lamp color classifications™
Systolic (mmHg)
(re blood pressure
mild hypertension, moderate
Orange hypertension, severe
hypertension)
140
Yellow Borderline
130 hypertension
Green Normal
— 85 90 Diastolic (mmHg)
e If the measured 873-1530)
blood pressure value LE, =‘ ,
is in the high blood { (М © Systolic blood pressure:
i ha 140mmHg or higher
pressure region, the :
values will blink for ¢ Diastolic blood pressure:
approximately 6 90mmHg or higher
seconds. (Only
immediately after
measurement)
If measurements could not be taken properly
& 3 15301 The remeasure mark E will be displayed.
When taking successive measurements, press the
Start/Stop button to switch off the blood pressure
monitor once and stay quiet for 4 to 5 minutes, and then
take the next measurement.
Nu ана
IM
Advice:
In general, blood pressure measurements are lower for the second measurement because you
are less nervous and are more tranquil.
For people who easily get tense, measuring twice each time is very effective.
When two measurements are taken, normally the smaller measurement is stored.
When taking two measurements, always stay quiet for 4 to 5 minutes after the first measurement
before taking the second measurement.
*1 1999 World Health Organization-International Society of Hypertension Guidelines for the
Management of Hypertension
EN11
STORING MEASUREMENT VALUES
This product can separately store 42 measurements
each for two persons. After measurement, the (À À will light.
Press the Memory/Recall
E. button with the number
ко Normal-High
yo corresponding to the person for which
“awd | Measurements should be stored.
Normal-High
Normal 1
Example: Person A: Press myr
Person B: Press mor
Storing is completed.
e Up to 42 measurement values can be stored.
When the memory is full and additional values
are stored, the oldest value in memory will be
deleted and the new value will be stored.
a © If measurements could not be taken properly
NE (the E mark is shown instead of the blood
a pressure value), measurement values cannot
Normal be stored.
However, even if the £ mark is shown for the
pulse value, the blood pressure value will still
be stored.
Number of stored data
To erase all measured values stored in memory:
1. Press the Memory/Recall button for the person whose data you want to delete.
The stored measurement values will be displayed.
2. Press the same Memory/Recall button again, and hold it pressed until the
measurement values disappear (approximately 3 seconds or more).
* It is not possible to erase single measurement values.
To switch off the blood pressure monitor,
press the Start/Stop button.
e |f you forget to switch it off, the blood pressure
monitor will switch off automatically after
approximately 5 minutes.
EN12
RECALLING STORED MEASUREMENT
Measurements can be recalled after measurement is
completed or after the power has been switched off.
If no measurement values are stored, no value will be shown.
(ih) 1 Press the Memory/Recall button for the
person whose data you want to recall.
Panasonic ) The average of all measured
values in memory will be Shown.
e TI mark will be shown at the bottom of the
display.
e For example, if 42 measured values are
stored, the average of the 42 measurements
Will be shown. (If only 1 measurement is
1
ry ñ stored, the KT mark will not be shown.)
a The blood pressure indicator lamps will light
mmHg
DIA. E
mmHg /
Plus/min.
sus. | 000
1 |
vv (for approximately 6 seconds).
Press the Memory/Recall button again.
A Each time you press the Memory/Recall button,
the previously stored measurement values will be
ww shown in order, starting with the newest data.
e The blood pressure indicator lamps will light
(for approximately 6 seconds).
e |f your body moves during measurement, the
body movement indicator lamp will blink (for
approximately 6 seconds).
Press the Memory/Recall button lightly for only
a short time. If you hold the button pressed for
approximately 3 seconds or longer, all data in
memory will be erased, so please be careful.
e If you forget to switch it off, the blood pressure
monitor will switch off automatically after
approximately 30 seconds.
3 To cancel measurement value recall,
press the Start/Stop button.
EN13
AFTER USE/SPECIFICATIONS
Storage
Unplug the plug from the main
body.
Roll up the pressure cuff loosely,
and Store the tube inside the
pressure cuff.
e Do not wrap the tube around the
main body, etc. Doing so may
result in excessive force on the
tube, and may result in a
malfunction.
e Be careful not to bend the tube.
Put the pressure cuff in the main
body and close the cover.
Specifications
DC 6V 4 x 1.5V (LR6/AA/Mignon/alkaline)
AC 100-240V 50/60Hz (when using optional AC adapter)
Display Digital LCD
Measurement method | Oscillometric system
Blood pressure: 0 to 280mmHg
Pulse rate: 30 to 160 beats per minute
Blood pressure: Within +3mmHg
Power source
Measurement range
Accuracy Pulse rate: Within +5%
Operation temperature / o
humidity range 10 to 40°C, 30 to 85% RH
Storage
temperature/humidity -10 to 60°C, 10 to 95% RH
range
Measurable arm Approx. 20 to 34cm
circumference pprox.
Weight Approx. 530g (not including batteries)
Dimensions 17.9 x 17.1 x 9.3cm
EN14
MEASUREMENT ACCURACY
General
This blood pressure measurement device complies with the European
regulations (EU council directive 93/42/EEC dated June 14, 1993) and
bears the CE mark "CE 0197". The quality of the device has been verified
and it confirms to the following standards:
EN 1060-1 (Dec. 1995) Non-invasive sphygmomanometers
General requirement
Non-invasive sphygmomanometers
Supplementary requirements for
electromechanical blood pressure measuring
systems
Non-invasive sphygmomanometers
Test procedures to determine the overall system
accuracy of automated non-invasive
sphygmomanometers.
EN 60601-1-2 (Nov. 2001) Electromagnetic compatibility and safety for
medical electrical equipment
EN 1060-3 (Sep. 1997)
EN 1060-4 (Sep. 2004)
The CE mark serves to facilitate the free movement of goods between the
member countries of the EU.
Panasonic ensures accurate measurement
The seal confirms that each device has been checked for
accurate measurement at the Matsushita plant. It is CALIBRATED UNTIL
recommended that the device be recalibrated every 3
years starting with the year printed on the seal. 2008
When the device has been repaired or the seal has been rendered illegible,
the accuracy of the device should be assumed to be inaccurate. In such
cases, it is recommended that the device be submitted for recalibration.
EN15
Q&A
Why are the blood pressure values measured at home different
from those measured at the doctor's office?
A Blood pressure varies 24 hours a day.
In addition, it can change greatly depending on the weather, your
mental state, exercise, etc.
In particular, there is the phenomenon of "White uniform hypertension”
at doctor's offices, where the blood pressure measured in a doctor's
office is higher because of stress or worry.
In addition, if the blood pressure measured at home is sometimes
higher and sometimes lower than the blood pressure measured at
the doctor's office, the following can be considered. Please pay
attention to these points when taking measurements.
@ ls the pressure cuff wrapped firmly?
e Make sure that the pressure cuff is not wrapped too loosely or too
tightly.
e Is the pressure cuff wrapped around the upper arm? Make sure it
is not partly covering the elbow.
For how to wrap the pressure cuff correctly, refer to p. 6.
O Do you feel worried or irritated?
Before taking measurements, take 2 or 3 deep breaths and relax to
stabilize blood pressure. In addition, it is recommended that you sit
quietly for about 4 to 5 minutes before taking measurements.
EN16
Why are measured blood pressure values different each time | take
a measurement?
@® Blood pressure changes even within a single
day, and it is different depending on the
measurement posture. Because of this,
when taking measurements, be sure to
always take measurements under the same
conditions.
For persons taking blood pressure
medication, blood pressure may change
greatly due to the effects of the medication.
When taking 2 measurements in succession,
be sure to stay quiet for 4 to 5 minutes after
the first measurement before taking the next
measurement.
Q Why are the measured values different for the left and right arm?
It is said that differences in blood pressure
values between the left and right arms may
be around 10mmHg. The blood pressure can
be measured in either the left or right arm,
but all measurements should be taken using
the same arm.
EN17
Q Even though the arm was moved, the body movement indicator
light did not come on.
A The body movement indicator function tells you that the elbow
was bent or that excess pressure was applied to the pressure cuff.
Because of this, if the movement resulted in no effect on the pressure
cuff (such as lightly twisting your arm, bending your wrist, etc.), the
indicator will not come on. If you still feel concerned about it, please
take another measurement.
Q When the body movement indicator lamp became lit, | returned my
arm to the correct measurement position, but the measured value
was higher than it always has been.
A The body movement indicator lamp is intended as only a
measurement method indicator.
If you returned to the correct posture after the body movement indicator
lamp became lit but the measured blood pressure is higher or lower
than normal, please take another measurement.
The following possible causes can also be considered:
e Are measurements being taken properly (how cuff is wrapped,
measurement posture, etc.)? Refer to pages 6 and 7 when taking
measurements.
Were measurements taken immediately after smoking?
Were measurements taken when you were feeling irritated?
Were measurements taken in a cold environment?
Are you a person who has low pulse changes?
If you feel concerned about it, please wait 4 to 5 minutes and then
take another measurement.
EN18
CARING FOR THE DEVICE
Please do not attempt to disassemble, repair or modify the device.
This may cause fire or cause the device to malfunction. It may also lead to
injury.
Do not use excessive force on the device.
Do not drop the device.
Measurements may be impaired if the device is used close to televisions,
microwave ovens, X-ray equipment or other devices with strong electrical
fields.
Please measure your blood pressure at a sufficient distance from such
devices or switch them off.
The cuff may lose some of its elasticity at temperatures below 0°C. If the
device is exposed to such extreme temperatures, it is a good idea to leave it
until it warms up to room temperature before taking the next measurement.
As a basic principle, the device should not be exposed to extreme
conditions (cold, heat, humidity, dust).
Do not allow dust or other foreign objects to enter the device as this may
cause it to malfunction.
Cleaning the device
e Do not use anything other than a cloth with warm water and a little soapy
to wipe the device.
e Do not use thinners, benzene, or alcohol on the device as this may cause
discoloration and cracks.
e Cuff can not be washed.
Storage
¢ Avoid exposing the device to high temperatures, high humidity and direct
sunlight as this can cause the device to malfunction.
Long-term storage
e Remove the batteries before storing the device for long periods of time.
Failure to do so may result in the batteries leaking, which may cause the
device to malfunction.
EN19
FOR YOUR SAFETY
Always follow your physician's instruction when taking medication.
Please consult your doctor as to the most appropriate time to measure your
blood pressure if you are receiving medication.
People with cardiac pacemakers, cardiac arrhythmia, vascular constriction
or liver disorders as well as diabetics should consult their doctor before
measuring their blood pressure themselves, since different values may be
measured in such cases. The same also applies during pregnancy.
The Diagnostec is no toy for children!
In case you measure your child's blood pressure, always consult your
doctor since the data cannot be compared with the standard adult's data.
EN20
TROUBLESHOOTING
Display Status before error Cause and solution
Remeasure mark
E is displayed.
B/ 3 15:30
IM
>
>
a Were measurements
Pressurization to more than p> | taken according to the
280mmHg was performed. proper procedure and in
the correct posture?
Pressure decreased suddenly |p See p. 6 and 7.
and the remeasure mark E
appeared. D | Was the
Measurement was taken to pressurization uf
20mmHg and then the > wrapped properly?
remeasure mark E appeared.
See p. 6.
The heart mark Y blinked only a > Tube may not be
few times or not at all.
after waiting several dozen seconds.
connected
sufficiently. Check
Pressurization doesn't occur even > the tube connection.
See p. 6.
Symptom
Possible cause
+ Position of pressure cuff is low.
dE » Pressure cuff is not wrapped correctly.
Systolic or diastolic blood pressure D |+ Poston of pressure o high, y
valu Is high or low + You moved or talked while measurements were being taken.
See p. 6 and7.
Blood pressure value is abnormally > + Posture is different for each measurement.
high or low. See p. 7.
№ : etn om that measured D | * Blood pressure varies in response to minute changes in your mental
y e doctor state, such as your reaction to a doctors presence.
Measurement values are different > o Sit quietly for 4 to 5 minutes and then take another measurement.
with each measurement. See p. 7.
No display appears even when the >
StarSiop bution pressed + Batteries are exhausted.
When the Start/Stop button was e Batteries are installed with the polarities wrong.
pressed, all displays became lit and | > Seep. 3.
then the display disappeared.
Body movement indicator lamp blinks > + Arm was moved.
during measurement. See p. 10. >
Even though the arm was not moved, e The body movement indicator lamp may respond when the arm
the body movement indicator lamp | [>| becomes tense. Relax all tension in the arm when taking
blinks. measurements. See p. 7.
Take it to the store where it was
Instrument s broken. > purchased for testing and repairs.
EN21
If problem continues
If problem continues
Ask your doctor for advice.
IMPORTANT SAFETY INSTRUCTIONS
* Before using this instrument, please read the "IMPORTANT SAFETY
INSTRUCTIONS" and be sure to use the instrument properly.
* The cautions in this section are intended to instruct you in the proper use
of this instrument and prevent danger or injury to yourself or other people.
All of the cautions contain important information
regarding safety and must be followed.
WARNING: Failure to follow the caution may result in death or serious
injury of people.
CAUTION: Failure to follow the caution may result in injury to people, or in
damage to property.
WARNING:
o Persons with serious obstructions of blood flow in the arm should consult
a doctor before using this instrument. Failure to do so may result in
deterioration of physical condition.
e When cleaning, unplug the AC adapter from the AC outlet before
cleaning. Also, do not plug in or unplug the AC adapter with wet hands.
Doing so may result in electric shock or fire.
e Do not let this product be used by children.
e Do not let this product be used by people who cannot express their
feelings, or who cannot operate the product for themselves. Such use
may result in an accident or injury.
e Do not spill water on the main body or AC adapter. Doing so may cause
electric shock, short-circuiting, or product damage.
e Do not use the AC adapter if it is damaged or if it plugs into the AC outlet
loosely. Doing so may result in electric shock or short-circuiting.
CAUTION:
e Do not make judgments yourself regarding the measurement results or self-treat yourself.
e Take measurements according to your doctor's instructions and have your doctor do the diagnosis.
e Take medicine according to your doctor's instructions.
e Do not use for purposes other than the measurement of blood pressure.
Doing so may result in an accident or malfunction.
e Never modify the product, or disassemble or attempt to repair it yourself.
Doing so may result in fire or in injury due to abnormal operation.
SAVE THESE INSTRUCTIONS