Eizo | RX650 | Owner Manual | Eizo RX650 Owner Manual

Eizo RX650 Owner Manual
Color LCD monitor
Important
Please read this “Instructions for Use”, and “Installation Manual”
(separate volume) carefully to familiarize yourself with safe and
effective usage.
Please retain this manual for future reference.
• For monitor adjustment and settings, refer to the “Installation
Manual”.
• For the latest product information including the “Instructions for
Use”, refer to our web site :
http://www.eizoglobal.com
English
Instructions for Use
SAFETY SYMBOLS
This manual and this product use the safety symbols below. They denote critical information. Please read
them carefully.
WARNING
CAUTION
Failure to abide by the information in a
WARNING may result in serious injury
and can be life threatening.
Failure to abide by the information in a
CAUTION may result in moderate injury and/or
property or product damage.
Indicates a warning or caution. For example,
indicates an “electrical shock” hazard.
Indicates a prohibited action. For example,
means “Do not disassemble”.
Indicates a mandatory action. For example,
means “Ground the unit”.
This product has been adjusted specifically for use in the region to which it was originally shipped. If
operated outside this region, the product may not perform as stated in the specifications.
No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by
any means, electronic, mechanical, or otherwise, without the prior written permission of EIZO Corporation.
EIZO Corporation is under no obligation to hold any submitted material or information confidential unless
prior arrangements are made pursuant to EIZO Corporation’s receipt of said information. Although every
effort has been made to ensure that this manual provides up-to-date information, please note that EIZO
monitor specifications are subject to change without notice.
2

PRECAUTIONS
• This product has been adjusted specifically for use in the region to which it was originally shipped.
If the product is used outside the region, it may not operate as specified in the specifications.
• For personal safety and proper maintenance, please read carefully this section and the caution
statements on the monitor.
Location of the caution statements
Monitor
AC adapter
Symbols on the unit
Symbol
This symbol indicates
AC adapter main power switch:
Press to turn the monitor’s main power off.
AC adapter main power switch:
Press to turn the monitor’s main power on.
Power switch:
Touch to turn the monitor’s power on or off.
Alternating current
Direct current
Electrical shock hazard
CAUTION:
Refer to “SAFETY SYMBOLS” (page 2).
WEEE marking:
Product must be disposed of separately; materials may be
recycled.
CE marking:
EU conformity mark in accordance with the provisions of
Council Directive 93/42/EEC and 2011/65EU.
Manufacturer
Date of manufacture
PRECAUTIONS
3
English
●●IMPORTANT
WARNING
If the unit begins to emit smoke, smells like something is burning, or makes strange noises,
disconnect all power connections immediately and contact your EIZO representative for advice.
Attempting to use a malfunctioning unit may result in fire, electric shock, or equipment damage.
Do not open the cabinet or modify the unit.
Opening the cabinet or modifying the unit may result in fire, electric shock, or burn.
Do not turn the bushing for fixing the AC adapter power cable.
Not doing so may result in fire, electric shock, or equipment damage.
Refer all servicing to qualified service personnel.
Do not attempt to service this product yourself as opening or removing covers may result in fire, electric
shock, or equipment damage.
Keep small objects or liquids away from the unit.
Small objects accidentally falling through the ventilation slots into the cabinet or spillage
into the cabinet may result in fire, electric shock, or equipment damage. If an object or
liquid falls/spills into the cabinet, unplug the unit immediately. Have the unit checked by
a qualified service engineer before using it again.
Place the unit at the strong and stable place.
A unit placed on an inadequate surface may fall and result in injury or equipment
damage. If the unit falls, disconnect the power immediately and ask your local EIZO
representative for advice. Do not continue using a damaged unit. Using a damaged unit
may result in fire or electric shock.
Use the unit in an appropriate location.
Not doing so may result in fire, electric shock, or equipment damage..
•
•
•
•
•
•
•
•
•
Do not place outdoors.
Do not place in the transportation system (ship, aircraft, trains, automobiles, etc.)
Do not place in a dusty or humid environment.
Do not place in a location where water is splashed on the screen (bathroom, kitchen, etc.).
Do not place in a location where the steam comes directly on the screen.
Do not place near heat generating devices or a humidifier.
Do not place in a location where the product is subject to direct sunlight.
Do not place in an inflammable gas environment.
Do not place in environments with corrosive gases (such as sulfur dioxide, hydrogen sulfide,
nitrogen dioxide, chlorine, ammonia, and ozone).
• Do not place in environments with dust, components that accelerate corrosion in the
atmosphere (such as sodium chloride and sulfur), conductive metals, and so on.
To avoid danger of suffocation, keep plastic packing bags away from babies and children.
Use the enclose AC adapter.
The enclosed AC adapter (PSA-073) is for use with this product only. Do not use the AC adapter with other
equipment. Connecting to power sources that do not match the power ratings of the AC adapter may result
in fire or electric shock.
Use the enclosed power cord and connect to the standard power outlet of your country.
Be sure to remain within the rated voltage of the power cord. Not doing so may result in fire or electric shock.
Power supply: 100-120/200-240Vac 50/60Hz
4
PRECAUTIONS
WARNING
English
To disconnect the power cord or adapter power cable, grasp the plug firmly and
pull.
Tugging on the cord or cable may damage and result in fire or
electric shock.
The equipment must be connected to a grounded main outlet.
Failure to do so may result in fire or electric shock.
Use the correct voltage.
• The unit is designed for use with a specific voltage only. Connection to another voltage than specified in this manual
may cause fire, electric shock, or equipment damage.
Power supply: 100-120/200-240Vac 50/60Hz
• Do not overload your power circuit, as this may result in fire or electric shock.
Handle the power cord and AC adapter with care.
Handle the power cord and AC adapter with care.
Do not place heavy objects on, pull or tie the power cord or the AC adapter. Using a
damaged cord or AC adapter may result in fire or electric shock.
For electrical safety, do not connect or disconnect the power cord in the presence of patients.
Never touch the plug, AC adapter or power cord during a thunderstorm.
Touching them may result in electric shock.
When attaching an arm stand, please refer to the user’s manual of the arm stand and install the unit
securely.
Not doing so may cause the unit to become unattached, which may result in injury or equipment damage.
Before installation, make sure that desks, walls, and others an arm stand is fixed on have adequate
mechanical strength. When the unit is dropped, please ask your local EIZO representative for advice. Do not
continue using a damaged unit. Using a damaged unit may result in fire or electric shock. When reattaching
the tilt stand, please use the same screws and tighten them securely.
Do not touch a damaged LCD panel directly with bare hands.
The liquid crystal that may leak from the panel is poisonous if it enters the eyes or
mouth. If any part of the skin or body comes in direct contact with the panel, please
wash thoroughly. If some physical symptoms result, please consult your doctor.
Fluorescent backlight lamps contain mercury (products that have LED backlight lamps contain no
mercury), dispose according to local, state or federal laws.
Exposure to elemental mercury can result in effects on the nervous system, including tremor, memory loss,
and headache.
PRECAUTIONS
5
CAUTION
Handle with care when carrying the unit.
Disconnect the power cord and cables when moving the unit. Moving the unit with the cord attached is
dangerous.
It may result in injury.
Carry or place the unit according to the correct specified methods.
• When carrying the unit, grasp and hold firmly as shown in the illustration below.
• Do not unpack or carry the unit only by a single person, since the large size unit is so
heavy.
Dropping the unit may result in injury or equipment
damage.
Do not block the ventilation slots on the cabinet.
• Do not place any objects on the ventilation slots.
• Do not install the unit in a closed space.
• Do not use the unit laid down or upside down.
Blocking the ventilation slots prevents proper airflow and may result in fire, electric
shock, or equipment damage.
Do not touch the plug or AC adapter with wet hands.
Doing so may result in electrical shock.
Use an easily accessible power outlet.
This will ensure that you can disconnect the power quickly in case of a problem.
Mind that the AC adapter becomes hot during use.
• Do not cover or place anything on top of the AC adapter. Do not place the AC adapter on top of things that trap heat
such as carpets, blankets, etc. Keep the AC adapter away from direct sunlight and heat sources such as heaters.
Not doing so may result in fire.
• Do not touch with bare hands. Doing so may result in burns.
• Before moving the monitor, be sure to turn off the power switch, disconnect the power plug from the power outlet,
and wait until it has cooled completely.
Do not suspend the AC adapter in midair.
Using it while suspended in midair may result in fire or electric shock.
Periodically clean the area around the power plug and the ventilation slot of the monitor and the AC
adapter.
Dust, water, or oil on the plug may result in fire.
Unplug the unit before cleaning it.
Cleaning the unit while it is plugged into a power outlet may result in electric shock.
If you plan to leave the unit unused for an extended period, disconnect the power cord from the wall
socket after turning off the power switch for the safety and the power conservation.
This product is only suitable for a patient environment, but not for contact with a patient.
6
PRECAUTIONS
Notice for this monitor
English
This product is intended to be used in displaying and viewing digital images for review and analysis by
trained medical practitioners. It does not support the display of mammography images for diagnosis.
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
This product should be set to horizontal view mode when used for the purposes mentioned above.
This product has been adjusted specifically for use in the region to which it was originally shipped. If the
product is used outside the region, it may not operate as specified in the specifications.
This product may not be covered by warranty for uses other than those described in this manual.
The specifications noted in this manual are only applicable when the following are used:
· Power cords provided with the product
· Signal cables specified by us
Only use optional products manufactured or specified by us with this product.
As it takes about 30 minutes for the performance of electrical parts to stabilize, adjust the monitor 30
minutes or more after the monitor power has been turned on or the monitor has recovered from the power
saving mode.
Monitors should be set to a lower brightness to reduce changes in luminosity caused by long-term use and
maintain a stable display. In addition, perform a calibration and Constancy test (including Grayscale Check)
on a periodic basis (refer to “Controlling Monitor Quality” in the Installation Manual).
When the screen image is changed after displaying the same image for extended periods of time, an
afterimage may appear. Use the screen saver or power save function to avoid displaying the same image for
extended periods of time.
If you place this product on a lacquer-coated desk, the color may adhere to the bottom of the stand due to
the composition of the rubber. Check the desk surface before use.
Periodic cleaning is recommended to keep the monitor looking new and to prolong its operation lifetime (refer
to “Cleaning” (page 8)).
The screen may have defective pixels or a small number of light dots on the screen. This is due to the
characteristics of the panel itself, and is not a malfunction of the product.
The backlight of the LCD panel has a fixed lifetime. When the screen becomes dark or begins to flicker,
please contact your local EIZO representative.
Do not press on the panel or edge of the frame strongly, as this may result in display malfunctions, such
as interference patterns, etc. If pressure is continually applied to the panel, it may deteriorate or damage
your panel. (If the pressure marks remain on the panel, leave the monitor with a black or white screen. The
symptom may disappear.)
Do not scratch or press on the panel with any sharp objects, as this may result in damage to the panel. Do
not attempt to brush with tissues as this may scratch the panel.
Notice for this monitor
7
When the monitor is cold and brought into a room or the room temperature goes up quickly, dew
condensation may occur on the interior and exterior surfaces of the monitor. In that case, do not turn the
monitor on. Instead wait until the dew condensation disappears, otherwise it may cause some damage to the
monitor.
Cleaning
Attention
• Chemicals such as alcohol and antiseptic solution may cause gloss variation, tarnishing, and fading of the cabinet
or panel, and also quality deterioration of the image.
• Never use any thinner, benzene, wax, and abrasive cleaner, which may damage the cabinet or panel.
Note
• The optional ScreenCleaner is recommended for cleaning the cabinet and panel surface.
If necessary, the stains on the cabinet and panel surface can be removed by moistening part of a soft
cloth with water.
Chemicals that may be used for cleaning
Material name
Ethanol
Isopropyl alcohol
Chlorhexidine
Benzalkonium chloride
Alkyldiaminoethylglycine
Glutaral
Glutaral
Ethanol
Isopropyl alcohol
Hibitane
Welpas
Tego 51
Sterihyde
Cidex Plus28
Product name
To use the monitor comfortably
• An excessively dark or bright screen may affect your eyes. Adjust the brightness of the monitor
according to the environmental conditions.
• Staring at the monitor for a long time tires your eyes. Take a 10-minute rest every hour.
8
Notice for this monitor
CONTENTS
English
PRECAUTIONS....................................................... 3
●●IMPORTANT.................................................... 3
Notice for this monitor.......................................... 7
Cleaning..................................................................... 8
To use the monitor comfortably.............................. 8
CONTENTS.............................................................. 9
Chapter 1
Introduction...................................... 10
1-1.
Features.........................................................10
1-2.
Package Contents.........................................11
1-3.
EIZO LCD Utility Disk...................................11
●●
●●
Disk contents and software overview.............11
Using RadiCS LE/ScreenManager Pro
for Medical.......................................................11
1-4.
Controls and Functions...............................12
Chapter 2
Setting Up......................................... 13
2-1.
Compatible Resolutions..............................13
2-2.
Connecting Cables.......................................14
2-3.
Adjusting the Screen Height and Angle.....15
2-4.
Attaching a Power Cord Holder..................16
Chapter 3
Troubleshooting............................... 17
Chapter 4
Specifications.................................. 19
Chapter 5
Glossary........................................... 22
●●Error code table..............................................18
Appendix............................................................... 24
Trademark................................................................ 24
License..................................................................... 24
Medical Standard.................................................... 25
FCC Declaration of Conformity............................. 26
EMC Information..................................................... 27
CONTENTS
9
Chapter 1 Introduction
Thank you very much for choosing an EIZO color LCD monitor.
1-1. Features
• 30.0 inches
• Supports to the resolution of 6M pixels (3280 × 2048 dots)
• Uses high-contrast panel (1000:1).
Enables the display of sharp images.
• Applicable to DisplayPort (applicable to 8 bit or 10 bit, not applicable to audio signals)
• The “PbyP (Picture by Picture)” function enables the display of two signals at the same time.
A bezel-less widescreen offers less strain on the eyes and higher work efficiency compared to having two
monitors side by side.
• The Hybrid Gamma function automatically identifies the display area for the monochrome and color
images on the same screen, and displays each set status.
*1 The identification may fail depending on the displayed image. The software to be used needs to be
verified. For validation, refer to the Installation Manual (on the CD-ROM).
• The RadiCS SelfQC function and the built-in Integrated Front Sensor allow the user to perform the
calibration and Grayscale Check for the monitor independently.
Refer to the Installation Manual (on the CD-ROM).
• The CAL Switch function allows the user to select the display mode optimum to the displayed
image.
Refer to the Installation Manual (on the CD-ROM).
• Selectable DICOM (page 22) Part 14 complied screen.
• The quality control software “RadiCS LE” used to calibrate the monitor and to manage the history is
included.
See “1-3. EIZO LCD Utility Disk” (page 11).
• The software “ScreenManager Pro for Medical” to adjust the screen using the mouse and keyboard
is included
See “1-3. EIZO LCD Utility Disk” (page 11).
• Power saving function
This product is equipped with power saving function.
-- 0 W power consumption when the main power is off
Equipped with main power switch.
When the monitor is not required, the power supply can be shut off using the main power switch
-- Presence Sensor
The sensor on the front side of the monitor detects the movement of a person. When a person moves
away from the monitor, the monitor shifts to the power saving mode automatically and does not display
the images on the screen. Therefore, the function reduces the power consumption. The sensitivity and
time until the power saving mode is activated can be set according to the monitor usage environment and
movement of the user.
Refer to the Installation Manual (on the CD-ROM).
• Space-saving thinner and lighter design
• Long service life LED backlight LCD panel
• Built-in ambient light sensor
Depending on the environment, the sensor sometimes indicates different values from those measured by a
stand-alone illuminance meter.
The usage of this sensor requires the quality control software “RadiCS/RadiCS LE” and the measurement is
always triggered by the user’s operation in the corresponding menu.
For further details about the measured value and how to execute the measurement, refer to the RadiCS/
RadiCS LE User’s Manual (on the CD-ROM).
• Displays HDCP (High-bandwidth Digital Content Protection) protected contents (DisplayPort only).
10
Chapter 1 Introduction
Attention
English
• A high temperature or high humidity environment may affect the measurement accuracy of the integrated
front sensor. We suggest storing and using the monitor under the following conditions.
-- Temperature 30˚C or less
-- Humidity 70% or less
• Avoid storing or using the sensor where it may be exposed to direct sunlight.
1-2. Package Contents
Check that all the following items are included in the packaging box. If any items are missing or damaged,
contact your local EIZO representative.
Note
• Please keep the packaging box and materials for future movement or transport of the monitor.
• Monitor
• Power cord
• USB cable : UU300
• Digital signal cable: DisplayPort - DisplayPort
(PP300) × 2 pcs.
•
•
•
•
AC adapter (PSA-073)
Power cord holder
EIZO LCD Utility Disk (CD-ROM)
Instructions for Use (this manual)
• Digital signal cable: DVI-D - DVI-D (Dual Link)
(DD300DL) × 2 pcs.
1-3. EIZO LCD Utility Disk
An “EIZO LCD Utility Disk” (CD-ROM) is supplied with this product. The following table shows the disk
contents and the overview of the software programs.
●●Disk contents and software overview
The disk includes application software programs for adjustment, and Installation Manual. Refer to
Readme.txt file on the disk for software startup procedures or file access procedures.
Contents
Overview
Readme.txt file
RadiCS LE (for Windows)
Quality control software for calibrating the monitor and managing the
calibration history.
ScreenManager Pro for Medical (for
Windows)
Software for adjusting the screen using the mouse and keyboard.
Installation Manual of this monitor (PDF file)
“Instructions for Use” of this monitor (PDF file)
●●Using RadiCS LE/ScreenManager Pro for Medical
For the installation and use of “RadiCS LE/ScreenManager Pro for Medical”, refer to the respective
User’s Manual on the disk.
When using this software, you will need to connect a PC to the monitor with the supplied USB cable.
For more information, refer to the Installation Manual (on the CD-ROM).
Chapter 1 Introduction
11
1-4. Controls and Functions
Monitor
5
4
11
Adjustment menu*1
13
12
3
67
8
9
10
14
2
1
1. Operation switches
switch
2.
3. Presence Sensor
4. Integrated Front Sensor
(slide type)
5. Ambient Light Sensor
6. USB downstream port
7. USB upstream port
8.
9.
10.
11.
12.
13.
14.
*1
Displays the menu. Follow the guide to perform the desired operations.
• Turns the power on or off.
• Indicates the monitor’s operation status.
Green: Operating, Orange: Power saving mode, Off: Main power/power off
Detects the movement of a person in front of the monitor.
Used for calibration and Grayscale Check.
Measures the environmental illumination.
Connects a peripheral USB device.
Connects the USB cable to use the software that needs USB connection, or to use
USB Hub function.
Used to connect the AC adapter power cable.
Power connector
DVI-D connector
Input signal connector
DisplayPort connector
Input signal connector
Connects a peripheral USB device.
USB downstream port
Used to adjust the height and angle of the monitor screen.
Stand
Security lock slot
Complies with Kensington’s MicroSaver security system.
Covers the monitor cables.
Cable holder
For instructions for use, refer to the Installation Manual (on the CD-ROM).
AC adapter (PSA-073)
1 2
1. Power cord holder
attachment hole
2. Power connector
3. Main power switch
4. Power cable
12
4
3
Chapter 1 Introduction
Used to attach the power cord holder.
Connects the power cord.
Turns the main power on or off.
Connects with the power connector on the monitor.
Chapter 2 Setting Up
English
2-1. Compatible Resolutions
The monitor supports the following resolutions.
√ : Supported
Resolution*1
Applicable
signal
Dot Clock
vertical
scan
frequency
640×480
VGA
60 Hz
720×400
VGA TEXT
70 Hz
800×600
VESA
60 Hz
1024×768
VESA
60 Hz
240 MHz
1280×1024
VESA
60 Hz
(Max.)
1600×1200
VESA
60 Hz
1536×2048
VESA CVT RB
46 Hz
1536×2048
VESA CVT RB
60 Hz
1640×2048
VESA CVT RB
43 Hz
1640×2048*3
VESA CVT RB
60 Hz
*1 Compatible resolutions for the left and right screens.
*2 It is also appropriate for the 10 bit.
*3 Recommended resolution.
The graphics board should comply with the VESA standard.
DisplayPort*2
DVI
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
Chapter 2 Setting Up
13
2-2. Connecting Cables
Attention
• Check that the monitor and the PC are powered off.
• When replacing the current monitor with this monitor, be sure to change the PC settings for resolution and
vertical scan frequency to those that are available for this monitor. Refer to the compatible resolution table before
connecting the PC.
Note
• When connecting multiple PCs to this product, refer to the Installation Manual (on the CD-ROM).
1. Connect the signal cables to the input signal connectors and PC.
Check the shape of the connectors, and connect the cables.
After connecting the signal cable, tighten the screws of the connectors to secure the coupling.
Connection examples :
DisplayPort 1 - DisplayPort 2 / DisplayPort 1 - DVI 2 / DVI 1 - DisplayPort 2 / DVI 1 - DVI 2
Digital connection
(DisplayPort)
Signal cable: DisplayPort - DisplayPort
(supplied)
Digital connection
(DVI)
Signal cable: DVI-D - DVI-D(Dual Link)
(supplied)
2. Connect the AC adapter power cable to the power connector on the monitor.
Attention
• Do not connect or disconnect the cable while the AC adapter main power is on.
• Do not install the AC adapter in the wrong orientation. The side with the EIZO logo should face up.
Check the shape of the monitor connector, and connect the power cable.
Tighten the screw for fixing the power cable to secure the power cable.
Connecting to the Monitor
3. Attach the power cord holder (See “2-4. Attaching a Power Cord Holder” (page
16)).
4. Plug the power cord into a power outlet and the power connector on the AC
adapter.
14
Chapter 2 Setting Up
5. Turn on the AC adapter main power.
-: On, ○: Off
English
6. Connect the USB cable when using RadiCS LE or ScreenManager Pro for
Medical.
Connecting to the Monitor
7. Touch
Connecting to a Computer
to turn on the monitor.
The monitor’s power indicator lights up green.
8. Turn on the PC.
The screen image appears.
If an image does not appear, refer to “Chapter 3 Troubleshooting” (page 17) for additional advice.
Attention
• Turn off the monitor and PC after using them.
• For the maximum power saving, it is recommended that the Power button be turned off. Turning off the AC
adapter main power switch or unplugging the power cord completely shuts off power supply to the monitor.
Note
• In order to maximize the monitor life and to minimize luminance deterioration as well as power consumption,
proceed the following:
-- Use the power saving function of the computer.
-- Turn off the monitor and PC after using them.
2-3. Adjusting the Screen Height and Angle
Hold left and right edge of the monitor with both hands, and adjust the screen height, tilt and swivel of the
screen to the best condition for working.
Attention
• Be sure that the cables are correctly connected.
Chapter 2 Setting Up
15
2-4. Attaching a Power Cord Holder
Attach the power cord holder to prevent the power cord from falling off.
Attention
• Be sure to attach the power cord holder when the main power to the AC adapter is off or before the power cable is
plugged into the power outlet.
1. Attach the supplied power cord holder to the power cord attachment hole on
the AC adapter.
2. Adjust the position to secure the power cord holder referring to the figure
shown below.
(1) Lower the power cord holder lever.
(2) Adjust the securing position.
3. Secure the power cord holder to prevent the power cord from falling off.
*Check that the power cord holder is secured.
16
Chapter 2 Setting Up
Chapter 3 Troubleshooting
Problem
1. No picture
• Power indicator does not light.
• Power indicator is lighting green.
• Power indicator is lighting orange.
• Power indicator is flashing orange and green.
2. The message below appears.
• This message appears when no signal is
input.
Example:
Possible cause and remedy
• Check whether the power cord and the adapter power
cable are connected properly.
• Turn the AC adapter main power switch on.
• Touch .
• Turn off the AC adapter main power, and then turn it on
again a few minutes later.
• Increase “Brightness”, “Contrast”, or “Gain” in the
Adjustment menu.
• Turn off the AC adapter main power, and then turn it on
again.
• Switch the input signal. For details, refer to the Installation
Manual (on the CD-ROM).
• Move the mouse or press any key on the keyboard.
• Check whether the PC is turned on.
• When Presence Sensor is set to “On”, the monitor may
have shifted to the power saving mode. Come near to the
monitor.
• Turn off the AC adapter main power, and then turn it on
again.
• The device that is connected using DisplayPort have a
problem. Solve the problem, turn off the monitor, and then
turn it on again. Refer to the User’s Manual of the output
device for further details.
This message appears when the signal is not input correctly
even when the monitor functions properly.
• The message shown left may appear, because some PCs
do not output the signal soon after power-on.
• Check whether the PC is turned on.
• Check whether the signal cable is connected properly.
• Switch the input signal. For details, refer to the Installation
Manual (on the CD-ROM).
Chapter 3 Troubleshooting
17
English
If a problem still remains after applying the suggested remedies, contact your local EIZO representative.
Problem
• The message shows that the input signal is
out of the specified frequency range.
Example:
Possible cause and remedy
• Check whether the PC is configured to meet the resolution
and vertical scan frequency requirements of the monitor
(see “2-1. Compatible Resolutions” (page 13)).
• Reboot the PC.
• Select an appropriate setting using the graphics board’s
utility. Refer to the manual of the graphics board for
details.
fD: Dot clock
fH: Horizontal scan frequency
fV: Vertical scan frequency
●●Error code table
Error Code
18
Description
0***
• Errors that occurred during SelfCalibration
1***
• Errors that occurred during Grayscale Check
*1**
• Errors that occurred during DICOM
*2**
• Errors that occurred during Custom
*3**
• Errors that occurred during CAL1
*4**
• Errors that occurred during CAL2
*5**
• Errors that occurred during CAL3
**10
• The product maximum brightness may be lower than the target brightness.
• Lower the target brightness.
**11
• The product minimum brightness may be higher than the target brightness.
• Raise the target brightness.
**34
• The sensor may not have come out during calibration, or light may have entered the
sensor.
• Turn off the AC adapter main power, wait for a few minutes before turning the power
back on, and then execute SelfCalibration/Grayscale Check again.
**61
• The sensor may not have come out.
• Check whether there is any foreign object near the sensor.
• Execute SelfCalibration/Grayscale Check again.
**94
• SelfCalibration/Grayscale Check cannot be executed due to the following settings:
--There is no signal input to the left screen, and the vertical resolution of the right
screen is lower than 2048 dots.
--The vertical resolutions of the left and right screens are lower than 2048 dots.
**95
• The execution conditions for SelfCalibration/Grayscale Check may not be appropriate.
• Check the execution conditions and reset them if necessary.
Chapter 3 Troubleshooting
Chapter 4 Specifications
Video Signals
USB
Power
Physical
Specifications
Operating
Environmental
Requirements
Type
RX650 :
IPS (Anti-Glare)
RX650-AR :
IPS (Anti-Reflection)
Backlight
LED
Size
76 cm (30.0 inch)
Native Resolution
3280 dots × 2048 lines
Display Size (H x V)
645.5 mm × 403.0 mm
Pixel Pitch
0.197 mm
Display Colors
10-bit colors
(DisplayPort) :
1.07 billion (maximum) colors of 68 billion colors
8-bit colors
(DVI) :
16.77 million from a palette of 68 billion colors
Viewing Angles
(H / V, typical)
176˚ / 176˚
Recommended
Brightness
400 cd/m2
Response Time
(typical)
30 ms (Black-white-black)
English
LCD Panel
Input Terminals
DVI-D (Dual Link) × 2, DisplayPort × 2
Digital Scanning
Frequency (H / V)
31 kHz - 129 kHz / 29 Hz - 61 Hz (DVI), 59 Hz - 61 Hz (DisplayPort)
Frame synchronous mode: 29.5 Hz - 30.5 Hz (DVI), 59 Hz - 61 Hz
Port
Upstream port × 1, Downstream port × 2
Standard
USB Specification Revision 2.0
Input (AC Adapter)
AC 100 V - 120 V ±10 %, 50 / 60Hz 2.3 A - 1.9 A
AC 200 V - 240 V ±10 %, 50 / 60Hz 1.2 A - 1.0 A
Input (Monitor)
DC24.5V 8.5A
Maximum Power
Consumption
225 W or less
Power Save Mode
6.0 W or less
(when only the DVI connector is connected, no USB
device is connected, and “Input Selection” is set to
“Manual”, and “DC5V Output” is set to “Off”)
Standby Mode
6.0 W or less
(when only the DVI connector is connected, no USB
device is connected, and “DC5V Output” is set to “Off”)
Dimensions
692 mm × 478 mm - 612.5 mm × 331.5 mm ( W × H × D )
Dimensions
(Monitor)
692 mm × 466 mm × 109 mm ( W × H × D )
Dimensions
(AC Adapter)
148 mm × 74mm × 280mm ( W × H × D )
Net Weight
Approx. 17.6 kg
Net Weight
(Monitor)
Approx. 13.6 kg
Net Weight
(AC Adapter)
Approx. 2.6 kg
Height Adjustment
Range
165 mm (Max.), 126 mm (Tilt: 0˚)
Tilt
Up 25˚, down 0˚
Swivel
344˚
Temperature:
0 ˚C to 35 ˚C (32 ˚F to 95 ˚F)
Humidity
20 % to 80 % R.H. (no condensation)
Air pressure
540 hPa to 1060 hPa
Chapter 4 Specifications
19
Transportation/
Storage
Environmental
Requirements
Temperature
-20 ˚C to 60 ˚C (-4 ˚F to 140 ˚F)
Humidity
10 % to 90 % R.H. (no condensation)
Air pressure
200 hPa to 1060 hPa
Outside Dimensions
Unit: mm
Monitor
131
35
67.5
109
478
501.5
692
647.5
TILT RA
22.25
NGE 2
5°
296
32
184.5
296
383.5
100
405
466
612.5
643
255~381(TILT 0°)
200
146.5
181.5
29
22.25
50
331.5
LOWEST POSITION
302
344°
SWIVEL RANGE
28.5
256.5
255
281.5
HIGHEST POSITION
AC adapter (PSA-073)
74
12
280
24
71
100
77
24
148
68
3-M4
10
20
Chapter 4 Specifications
245
22
1800
100
296
Accessories
Network QC Management Software
Cleaning Kit
EIZO “RadiCS UX1” Ver. 4.3.0 or later
EIZO “RadiCS Version Up Kit” Ver. 4.3.0 or later
EIZO “RadiNET Pro” Ver. 4.3.0 or later
EIZO “ScreenCleaner”
English
Calibration Kit
For the latest information about the accessories and information about the latest compatible graphics board,
refer to our web site.
http://www.eizoglobal.com
Chapter 4 Specifications
21
Chapter 5 Glossary
DDC (Display Data Channel)
VESA provides the standardization for the interactive communication of the setting information, etc.
between a PC and the monitor.
DICOM (Digital Imaging and Communication in Medicine)
The DICOM standard was developed by the American College of Radiology and the National
Electrical Manufacturer’s Association of the USA.
The DICOM compatible device connection enables to transfer the medical image and information.
The DICOM, Part 14 document defines the digital, grayscale medical image display.
DisplayPort
This is the interface standard for image signals standardized in accordance with VESA. It was
developed with the aim of replacing the conventional DVI and analog interfaces, and it can transmit
high resolution signals and sound signals, which DVI does not support. It also supports 10-bit color,
copyright protection technology, long cables, etc. The standard size and mini size connectors have
been standardized.
DVI (Digital Visual Interface)
DVI is a digital interface standard. DVI allows direct transmission of the PC’s digital data without
loss.
This adopts the TMDS transmission system and DVI connectors. There are two types of DVI
connectors. One is a DVI-D connector for digital signal input only. The other is a DVI-I connector
for both digital and analog signal inputs.
DVI DMPM (DVI Digital Monitor Power Management)
DVI DMPM is a digital interface power saving function. The “Monitor ON (operating mode)” and
“Active Off (power saving mode)” are indispensable for DVI DMPM as the monitor’s power mode.
Gain
This is used to adjust each color parameter for red, green and blue. An LCD monitor displays the
color by the light passing through the panel color filter. Red, green and blue are the three primary
colors. All the colors on the screen are displayed by combining these three colors. The color tone
can be changed by adjusting the light intensity (volume) passing through each color’s filter.
Gamma
Generally, the monitor brightness varies nonlinearly with the input signal level, which is called
“Gamma Characteristic”. A small gamma value produces a low-contrast image, while a large
gamma value produces a highcontrast image.
22
Chapter 5 Glossary
HDCP (High-bandwidth Digital Content Protection)
Resolution
The LCD panel consists of numerous pixels of specified size, which are illuminated to form images.
This monitor consists of horizontal 3280 pixels and 2048 vertical pixels. Therefore, if the combined
resolution of the left and right screen is 3280 × 2048, all pixels are illuminated as a full screen (1:1).
sRGB (Standard RGB)
International standard for color reproduction and color space among peripheral devices (such as
monitors, printers, digital cameras, scanners). As a form of simple color matching for the Internet,
colors can be displayed using tones close to those of the transmission and reception devices.
Temperature
Color temperature is a method to measure the white color tone, generally indicated in degrees
Kelvin. The screen becomes reddish at a low temperature, and bluish at a high temperature, like
the flame temperature.
5000 K: Slightly reddish white
6500 K: White referred to as daylight-balanced color
9300 K: Slightly bluish white
Chapter 5 Glossary
23
English
Digital signal coding system developed to copy-protect the digital contents, such as video, music,
etc.
This helps to transmit the digital contents safely by coding the digital contents sent via the DVI or
HDMI connector on the output side and decoding them on the input side.
Any digital contents cannot be reproduced if both of the equipments on the output and input sides
are not applicable to HDCP system.
Appendix
Trademark
The terms HDMI and HDMI High-Definition Multimedia Interface, and the HDMI Logo are trademarks or
registered trademarks of HDMI Licensing, LLC in the United States and other countries.
The DisplayPort Compliance Logo and VESA are registered trademarks of the Video Electronics
Standards Association.
Acrobat, Adobe, Adobe AIR, and Photoshop are registered trademarks of Adobe Systems Incorporated
in the United States and other countries.
AMD Athlon and AMD Opteron are trademarks of Advanced Micro Devices, Inc.
Apple, ColorSync, eMac, iBook, iMac, iPad, Mac, MacBook, Macintosh, Mac OS, PowerBook, and
QuickTime are registered trademarks of Apple Inc.
ColorMunki, Eye-One, and X-Rite are registered trademarks or trademarks of X-Rite Incorporated in the
United States and/or other countries.
ColorVision and ColorVision Spyder2 are registered trademarks of DataColor Holding AG in the United
States.
Spyder3 and Spyder4 are trademarks of DataColor Holding AG.
ENERGY STAR is a registered trademark of the United States Environmental Protection Agency in the
United States and other countries.
GRACoL and IDEAlliance are registered trademarks of International Digital Enterprise Alliance.
NEC is a registered trademark of NEC Corporation.
PC-9801 and PC-9821 are trademarks of NEC Corporation.
NextWindow is a registered trademark of NextWindow Ltd.
Intel, Intel Core and Pentium are registered trademarks of Intel Corporation in the United States and
other countries.
PowerPC is a registered trademark of International Business Machines Corporation.
PlayStation is a registered trademark of Sony Computer Entertainment Inc.
PSP and PS3 are trademarks of Sony Computer Entertainment Inc.
RealPlayer is a registered trademark of RealNetworks, Inc.
TouchWare is a trademark of 3M Touch Systems, Inc.
Windows, Windows Media, Windows Vista, SQL Server, and Xbox 360 are registered trademarks of
Microsoft Corporation in the United States and other countries.
YouTube is a registered trademark of Google Inc.
Firefox is a registered trademark of the Mozilla Foundation.
Kensington and MicroSaver are trademarks of ACCO Brands Corporation.
EIZO, the EIZO Logo, ColorEdge, DuraVision, FlexScan, FORIS, RadiCS, RadiForce, RadiNET, Raptor,
and ScreenManager are registered trademarks of EIZO Corporation in Japan and other countries.
ColorNavigator, EcoView NET, EIZO EasyPIX, EIZO ScreenSlicer, i • Sound, Screen Administrator, and
UniColor Pro are trademarks of EIZO Corporation.
All other company and product names are trademarks or registered trademarks of their respective
owners.
License
A round gothic bold bit map font designed by Ricoh is used for the characters displayed on this product.
24
Appendix
Medical Standard
Classification of Equipment
------
Type of protection against electric shock : Class I
EMC class: EN60601-1-2:2007 Group 1 Class B
Classification of medical device (MDD 93/42/EEC) : Class I
Mode of operation : Continuous
IP Class : IPX0
Appendix
25
English
• It shall be assured that the final system is in compliance to IEC60601-1-1 requirement.
• Power supplied equipment can emit electromagnetic waves, that could influence, limit or result in
malfunction of the monitor. Install the equipment in a controlled environment, where such effects
are avoided.
FCC Declaration of Conformity
For U.S.A. , Canada, etc. (rated 100-120 Vac) Only
FCC Declaration of Conformity
We, the Responsible Party
EIZO Inc.
5710 Warland Drive, Cypress, CA 90630
Phone: (562) 431-5011
declare that the product
Trade name: EIZO
Model: RadiForce RX650
is in conformity with Part 15 of the FCC Rules. Operation of this product is subject to the
following two conditions: (1) this device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may cause undesired
operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant
to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in
a particular installation. If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures.
*
*
*
*
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
Note
Use the attached specified cable below or EIZO signal cable with this monitor so as to keep interference
within the limits of a Class B digital device.
- AC Cord
- Shielded Signal Cable (enclosed)
Canadian Notice
This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de le classe B est comforme à la norme NMB-003 du Canada.
26
Appendix
EMC Information
CAUTION
The RadiForce series requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or specified by us.
Using other cables may cause the increase of emission or decrease of immunity.
Cable Length: Max. 3m
Do not put any portable and mobile RF communications equipment close to the RadiForce series.
Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts,
configuring a medical system, is responsible that the system complies with the requirements of
IEC/EN60601-1-2.
Guidance and manufacturer’s declaration - electromagnetic emissions
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Emission test
RF emissions
CISPR11/EN55011
Compliance
Group 1
RF emissions
Class B
CISPR11/EN55011
Harmonic emissions Class D
IEC/EN61000-3-2
Voltage fluctuations / Complies
flicker emissions
IEC/EN61000-3-3
Electromagnetic environment - guidance
The RadiForce series uses RF energy only for its internal function.
Therefore, its RF emission are very low and are not likely to cause any
interference in nearby electronic equipment.
The RadiForce series is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration - electromagnetic immunity
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC/EN61000-4-2
Electrical fast
transient / burst
IEC/EN61000-4-4
Surge
IEC/EN61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC/EN61000-4-11
IEC/EN60601 test
level
±6kV contact
±8kV air
Compliance level Electromagnetic environment - guidance
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
±2kV
for power supply lines commercial or hospital environment.
±1kV
for input/output lines
±1kV line(s) to line(s) Mains power quality should be that of a typical
±2kV line(s) to earth commercial or hospital environment.
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in
UT) for 5sec
3A/m
<5% UT (>95% dip in
UT) for 0.5 cycle
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5sec
3A/m
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the RadiForce series requires continued
operation during power mains interruptions, it
is recommended that the RadiForce series be
powered from an uninterruptible power supply or
a battery.
Power frequency magnetic fields should be at
Power frequency
levels characteristic of a typical location in a
(50/60Hz)
typical commercial or hospital environment.
magnetic field
IEC/EN61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Appendix
27
English
Essential performance of RadiForce series is to display images and operate functions normally.
Guidance and manufacturer’s declaration - electromagnetic immunity
The RadiForce series is intended for use in the electromagnetic environment specified below. The customer or the user
of the RadiForce series should assure that it is used in such an environment.
Immunity test
IEC/EN60601 test
level
Compliance level
Conducted RF
IEC/EN61000-4-6
3Vrms
150kHz to 80MHz
3Vrms
Radiated RF
IEC/EN61000-4-3
3V/m
80MHz to 2.5GHz
3V/m
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the RadiForce series, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended Separation distance
d = 1.2 √P
d = 1.2 √P, 80MHz to 800MHz
d = 2.3 √P, 800MHz to 2.5GHz
Where “P” is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and “d” is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site surveya,
should be less than the compliance level in each
frequency rangeb.
Interference may occur in the vicinity of equipment
marked with the following symbol.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the RadiForce series is used exceeds
the applicable RF compliance level above, the RadiForce series should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
RadiForce series.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
28
Appendix
Recommended separation distances between portable and mobile RF communications equipment
and the RadiForce Series
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Appendix
29
English
The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the RadiForce series can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
RadiForce series as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output
m
power of transmitter
150kHz to 80MHz
80MHz to 800MHz
800MHz to 2.5GHz
d = 1.2 √P
d = 1.2√P
d = 2.3√P
W
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance “d” in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where “P” is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan
Siemensallee 84, 76187 Karlsruhe, Germany
中国苏州市苏州工业园区展业路 8 号中新科技工业坊 5B
http://www.eizoglobal.com
Copyright © 2013-2017 EIZO Corporation. All rights reserved.
00N0L842E1
U.M-RX650-4
5th Edition-June, 2017 Printed in Japan.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising