Dwyer | Series DPG-200 | Instruction manual | Dwyer Series DPG-200 Instruction Manual

Dwyer Series DPG-200 Instruction Manual
Bulletin A-34A
Series DPG-200 Digital Pressure Gage
®
Specifications - Installation and Operating Instructions
2-3/8 [60.45]
1-5/8
[40.39]
3 [76.20]
3-5/8
[91.39]
1/4 NPT
The Series DPG-200 Digital Pressure Gage has a precise ±0.25% full scale
accuracy. The 4 digit digital display will reduce the potential for errors in readings by
eliminating parallax error commonly produced with analog gages. The DPG-200 is
packaged in a durable extruded aluminum case designed to meet NEMA 4X (IP66).
The unit is powered by 12 to 24 VDC/VAC and contains two alarm set points along
with a 4 to 20 mA process output. A four-button keypad allows easy access to features.
These features include backlight, peak and valley, auto zero and conversion of the
pressure units.
INSTALLATION
When installing gage always use 1˝ hex at the base of the housing to tighten the gage
to a mating fitting. Do not apply wrench to housing.
WIRING
WHITE
RED
BLACK
BLUE
ORANGE
GREEN
RED
BLACK
ORANGE(+)
BROWN (-)
N.O.
COM
N.C.
HIGH ALARM
N.O.
COM
N.C.
LOW ALARM
POWER SUPPLY
10 to 28 VDC
OR VAC
4 to 20 mA
RECEIVER
SHEILD
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
SPECIFICATIONS
DIGITAL GAGE SPECIFICATIONS
Service: Liquids and non-combustible compatible gases.
Wetted Materials: Type 316L SS.
Housing: Black polycarbonate front & back cover, anodized aluminum extruded
housing with recessed grooves, polycarbonate overlay, Buna-N O-rings, 316L SS
sensor construction.
Accuracy: 0.25% FS ±1 least significant digit. (Includes linearity, hysteresis,
repeatability).
Pressure Limit: 2x pressure range for models 1000 psi; 5000 psi for 3000 psi
range; 7500 psi for 5000 psi range.
Temperature Limits: 0 to 158°F (0 to 70°C).
Process Connection: 1/4˝ male NPT.
Display: 4 digit (.425 H x .234 W digits).
Size: 3.00˝ OD x 1.90 deep (not including cables).
Weight: 8.84 oz (275 g).
SWITCH SPECIFICATIONS
Switch Type: 2 SPDT Form C contacts.
Electrical Rating: 0.5A @ 125 VAC resistive, 1A @ 24 VDC.
Relay Differential: 1 least significant digit.
Electrical Connections: Two 3 ft (.91 m) cables.
Mounting Orientation: Mount in any position.
Set Point Adjustment: Via menu.
TRANSMITTER SPECIFICATIONS
Temperature Limits: 0 to 158°F (0 to 70°C).
Thermal Effect: Between 70 to 158°F = 0.016%/°F. Between 0 to 70°F =
0.026%/°F.
Power Requirements: 12 to 24 VAC ±20% 50 to 400 HZ, 12 to 24 VDC ±20%.
Output Signal: 4 to 20 mA.
Loop Resistance: 600 Ω maximum.
Power Consumption: 0.8 W max.
Electrical Connections: 2 three foot cables.
Enclosure Rating: Designed to meet NEMA 4X (IP66).
Phone: 219/879-8000
www.dwyer-inst.com
Fax: 219/872-9057
e-mail: info@dwyermail.com
POWER UP
HOME
• Top line shows units.
• Bottom line shows pressure.
• If pressure is 1% FS below range of instrument the pressure value will alternate UFL on a 1 second interval.
• If pressure is 1% FS above range of instrument the pressure value will alternate with OFL on a 1 second interval.
• Factory set to PSI.
BUTTON
OPERATION
See menu page.
Zeros display if within ±4% F.S. of zero.
See units chart below.
See peak/valley.
PEAK VALLEY
PEAK
• Highest pressure since peak and valley was reset.
• Zero button resets both peak and valley to the current pressure.
VALLEY
• Lowest pressure since peak and valley was reset.
• Zero button resets both peak and valley to the current pressure.
UNITS CHART
Range Pressure Ranges
Model
psi
kg/cm2 bar
in Hg ft wc ft sw* kPa
oz/in2 in wc mbar cm wc mm Hg
DPG-200 -14.70-0 -1.033 -1.013 -29.93 -33.94 -33.06 -101.4 -235.2 -407.3 -1013 -1034 -760.7
DPG-202 15.00
1.055
1.034 30.54 34.61 33.73 103.4 240
415.2 1034 1055 775.7
DPG-203 30.00
2.109
2.069 61.08 69.21 67.45 206.9 480
830.4 2069 2109 1551
DPG-204 50.00
3.515
3.448 101.8 115.4 112.4 344.8 800
1384 3448 3515 2586
DPG-205 100.0
7.03
6.895 203.6 230.7 224.8 689.5 1600 2768 6895 7031 5172
DPG-206 200.0
14.06
13.79 407.2 461.4 449.7 1379 3200 5536
DPG-207 300.0
21.09
20.69 610.8 692.1 674.5 2069 4800 8304
DPG-208 500.0
35.15
34.48 1018 1154 1124
3448 8000
DPG-209 1000
70.3
68.98 2036 2307 2248 6895
DPG-210 3000
210.9
206.9 6108 6921 6745
DPG-211 5000
351.5
344.8
Compound Ranges Available: DPG-220 Range: 30˝ Hg-0-15 psi; DPG-221 Range: 30˝ Hg-0-30 psi; DPG222* Range: 30˝ Hg-0-45 psi; DPG-223*: Range 30˝ Hg-0-60 psi; DPG-224: 30˝ Hg-0-100 psi.
*feet of seawater @ 4°C
This symbol indicates waste electrical products should not be disposed
of with household waste. Please recycle where facilities exist. Check with
your Local Authority or retailer for recycling advice.
MENU
SECURITY
• 1=full access, password is 1110.
• 2=protected, password is 1101.
• Factory set to full access.
• All menus except backlight are protected.
ALARM #1 SETPOINT
• Activates Alarm #1 relay on increase.
• AL1 indicator and alarm #1 relay turned on when pressure is
greater than or equal to setpoint.
ALARM #1 DIFFERENTIAL
• AL1 indicator and alarm #1 relay turned off when pressure is less
than the setpoint minus the differential.
ALARM #2 SETPOINT
• Activates Alarm #2 relay on decrease.
• AL2 indicator and alarm #2 relay turned on when pressure is less
than or equal to setpoint.
ALARM #2 DIFFERENTIAL
• AL2 indicator and alarm #2 relay turned off when pressure is
greater than the setpoint plus the differential.
ALARM #2 INHIBIT
• Prevents low alarm on power up.
• When set to “On” alarm #2 is disabled until the pressure is greater
than the setpoint plus the differential.
• Factory set to ON.
MENU BUTTON OPERATION
Advances menu.
Hold for 1 second to go back to home page.
Decreases value - top of display beings to blink.
Hold for 1 second to decrease fast.
Increases value - top line of display begins to blink.
Hold for 1 second to increase fast.
Stores value - top line of display stops blinking.
If menu pressed before enter button pressed the value is lost.
MENU CONTINUED
BACKLIGHT
• Turns backlight on or off.
• Factory set to off.
PROCESS OUTPUT HIGH
• Pressure at which current output is set to 20 mA.
• Factory set to maximum scale.
PROCESS OUTPUT LOW
• Pressure at which current output is set to 4 mA.
• Factory set to minimum scale.
SPAN ADJUSTMENT
• Display shows pressure using existing span.
• Apply full scale pressure to instrument and hold enter for one
second to acquire new span.
MENU BUTTON OPERATION
Advances menu.
Hold for 1 second to go back to home page.
Decreases value - top of display beings to blink.
Hold for 1 second to decrease fast.
Increases value - top line of display begins to blink.
Hold for 1 second to increase fast.
Stores value - top line of display stops blinking.
If menu pressed before enter button pressed the value is lost.
MENU BUTTON OPERATION
50
40
PRESSURE
(psi)
50
AL1S
AL1D
30
PRESSURE
(psi)
20
10
AL2S + AL2D
40
30
AL2D
20
10
AL1S - AL1D
AL2S
0
0
0
0
TIME
AL1S sets the relay on point, and AL1S - AL1D sets the relay off point. The relay’s output
functions in the direct acting mode, which means the relay activates with an increase
in pressure. In the above graph, an instrument with a 50 psi range has the AL1S set at
40. The relay will turn on at 40 psi increase. The AL1D is set at 30. The relay will turn off
at 10 psi decrease (40 - 30 = 10).
©Copyright 2019 Dwyer Instruments, Inc.
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
TIME
AL2S sets the relay on point, and AL2S + AL2D sets the relay off point. The relay’s output
functions in the reverse acting mode, which means the relay activates with a decrease
in pressure. In the above graph, an instrument with a 50 psi range has the AL2S
relay turn on at 10 psi decrease. The AL2D is set at 30. The relay will turn off at 40 psi
increase (10 + 30 = 40).
Printed in U.S.A. 2/19
FR# 443202-30 Rev. 2
Phone: 219/879-8000
www.dwyer-inst.com
Fax: 219/872-9057
e-mail: info@dwyermail.com