Dwyer Series 485 Operating instructions

Add to my manuals
2 Pages

advertisement

Dwyer Series 485 Operating instructions | Manualzz
485 IOM
8/8/02
8:31 AM
Page 1
Bulletin E-85-A
Model 485-2 Digital Hygrometer
Specifications and Operating Instructions
Specifications
Relative Humidity Measurement
Range: 0-100%
Accuracy: ±2%
Resolution: 0.1%
Sensor Operating Temperature:
-22° to 185°F ( -30° to 85°C)
Temperature Measurement
Range: -22° to 185°F (-30° to 85°C)
Accuracy: ±1°F (0.5°C)
Resolution: 0.1°
Ambient Temperature Operating Range
32° to 104°F (0° to 40°C)
Storage Temperature Limits
-40° to 176°F (-40° to 80°C)
Power Source
9 Volt Alkaline Battery
Operating Instructions & Specifications
Introduction
The Model 485 Digital Hygrometer is a versatile, hand-held,
battery-operated humidity and air temperature measurement
system. The instrument will directly measure temperature, in
either °C or °F, and percent relative humidity. The 485 will also
calculate the dew point temperature and the wet bulb temperature based on the relative humidity and temperature readings.
MIN and MAX functions are provided to save minimum and
maximum temperature and relative humidity readings. The
reading may also be frozen using the HOLD function. In addition, a memory function is provided which will hold up to 25
humidity and temperature samples.
Battery Installation
The unit is shipped with a separate 9 Volt alkaline battery
which must be installed before operation. Remove the two
screws holding the bottom endcap in place and remove it.
Connect the battery to the enclosed battery clip, observing correct polarity. Be careful not to trap the wires between the case
or foam pad which retains the battery. This could make it difficult to install the battery or remove it later for replacement. Be
sure the rubber gasket is properly seated in the gasket channel
and replace the endcap. Note that the endcap will only fit one
way because the holes are slightly off-center. Place the “Z”
shaped wrist strap clip in one of the screw recesses and replace
the screws. Do not overtighten. Attach the wrist strap to the
clip.
When battery replacement becomes necessary, use only 9
Volt alkaline type batteries such as a Duracell® MN1604, or
Eveready® 522 or equivalent. Zinc-carbon types, often labeled
Heavy-Duty are not recommended because of their shorter life
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.
and an increased potential for leakage. Alkaline batteries are a
better value because they last typically up to three times longer
in this device.
On-Off Operation
The on-off control is a toggle function. Press the ON/OFF key
once to turn the unit on; again to turn it off. If the hygrometer is
left on with no activity for approximately 22 minutes, the unit
will turn itself off to conserve the battery. Each time a key is
pressed, the timeout will be reset to 22 minutes.
An extended timeout function is provided for situations
where the 22 minute timeout is not long enough. With the unit
off, press the ON/OFF key and the DP/WB/STORE key simultaneously during power on and release them together. The
“ALARM 2” annunciator will be illuminated. The power timeout
will now be approximately 23 minutes from the time the unit is
turned on. This function is also useful for min/max sampling.
Turn the unit on in the extended timeout mode then clear the
MIN/MAX memory and place the unit in the desired location. It
will record the minimum and maximum temperature and humidity measured during the sample time.
Display Backlight
The Model 485 includes a display backlight to allow use in
the dark or in poor lighting conditions. The unit must be off
before this feature can be activated. Next, press and hold the
ON/OFF key down. After about 1 second the backlight will
come on and remain lighted for approximately 2 minutes after
which it will turn itself off to conserve the battery.
Selecting Temperature Units
The temperature readings may be displayed in either °F or
°C. The currently selected units will be indicated on the display.
Phone: 219/879-8000
Fax: 219/872-9057
Lit-By Fax: 888/891-4963
www.dwyer-inst.com
e-mail: [email protected]
485 IOM
8/8/02
8:31 AM
Page 2
To change the units, press the UNITS/LOC key. The selected
units will remain in memory even when the power is shut off.
This way, your preference will always be displayed after the initial selection.
Display Hold
There may be situations where you want to temporarily retain
a reading. The Model 485 provides a display hold function which
freezes the current reading and holds the reading until cleared.
To activate this function, momentarily press the HOLD/MEMORY key when the reading you want is displayed. A HOLD indicator will appear in the display to indicate that the reading
shown is frozen. To return to normal operation, press the
HOLD/MEMORY key again. The HOLD indicator will disappear
and the current temperature and humidity readings will again be
shown.
Dew Point and Wet Bulb Temperatures
The Model 485 provides calculated dew point and wet bulb
temperatures. To display the dew point temperature, press the
DP/WB/STORE key. The large display will show the dew point
temperature in the selected units. The %RH indicator will be
extinguished and the smaller lower left display will show “dP” to
indicate dew point measurement. To display the wet bulb temperature, press the DP/WB/STORE key again. The lower left display will now display “bt” to indicate wet bulb temperature. To
restore the display to %RH, press the DP/WB/STORE key again.
The %RH indicator will be illuminated and the relative humidity
and ambient temperature will again be displayed.
Min-Max Readings
A minimum and maximum reading function is provided by the
model 485. This will save the minimum and maximum temperature and relative humidity readings. To display the minimum
temperature and relative humidity readings press and hold the
MIN key. While the key is depressed the minimum recorded
readings will be displayed. To return to normal operation,
release the key. Likewise, to display the maximum readings,
press and hold the MAX key; release when done. To clear the
MIN/MAX memory, press both the MIN and MAX keys simultaneously. The display will display “----” while the memory is
being cleared. When the switches are released the MIN/MAX
memory begins storing the current minimum and maximum values. The minimum and maximum values are stored in nonvolatile memory which retains the minimum and maximum values even when the power is off.
Memory Function
A memory function is included in the Model 485 that allows
you to store up to 25 humidity and temperature readings. The
samples are stored in nonvolatile memory so the samples are
retained even when the power is off.
then be released. The active memory location will be briefly displayed, starting initially with “01”, then humidity will be displayed.
Storing Readings
To store a reading, press the DP/WB/STORE key. The readings will be stored in the previously indicated memory location
and a beep will sound to indicate the reading has been saved.
As each reading is saved, the memory location is advanced to
the next location and briefly displayed in the lower left display.
Viewing Stored Readings
To view the contents of memory, the unit must be in the memory mode. To enter the view mode, press the UNITS/LOC button. The HOLD indicator will then be displayed with the MEM
indicator to indicate that the memory values are being displayed
rather than the current readings, and the location number will be
briefly displayed. Each time the UNITS/LOC button is pressed
the memory location is advanced, and the new location is briefly
displayed. If the button is held down, the location will automatically increment until the button is released. This operation may
be used to select a particular memory location to store the next
reading. To resume temperature and humidity measurements,
press the HOLD/MEMORY button. The HOLD indicator will disappear and the display will resume displaying the current measurements. The last viewed memory location will remain as the
location in which the next sample will be stored.
Clearing Memory
The memory may be cleared of all previously stored values by
pressing and holding the DP/WB/STORE key, then simultaneously pressing the ON/OFF key. During the clearing operation “---” will be displayed. When the memory is cleared, the current
readings will be displayed and the memory location will be set
to “01”.
Exiting Memory Mode
To exit memory mode, press the HOLD/MEMORY key. The
memory indicator will disappear. All values stored in memory will
be retained for later viewing.
Low Battery Indicator
A weak battery may cause improper operation or inaccurate
measurements. A low battery indicator is provided on the display to show when the battery needs to be replaced. Although
the unit may appear to function and read properly, the accuracy
of the readings cannot be guaranteed when the “LOW BAT”
indicator is illuminated. Replace the battery with a fresh one. Do
not leave an exhausted battery in the unit due to potential battery leakage.
Entering Memory Mode
To enter memory mode, press and hold the HOLD/MEMORY
key until the MEM indicator appears in the display. The key can
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
Lit-By Fax: 888/891-4963
www.dwyer-inst.com
e-mail: [email protected]

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement