Hanna Instruments HI96712 Owner Manual

Add to my manuals
2 Pages

advertisement

Hanna Instruments HI96712 Owner Manual | Manualzz
VALIDATION AND CALIBRATION PROCEDURES
Warning: do not validate or calibrate the instrument with standard
solutions other than the Hanna Instruments CAL Check™ Standards, otherwise
erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at room
temperature (18 to 25 °C; 64.5 to 77.0 °F).
Use the Hanna Instruments Office CAL Check™ cuvettes (see
“Accessories”) to validate or calibrate instruments.
Validation
1• Turn the meter on by pressing ON/OFF.
Validation
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Place the CAL Check™ Standard
HI96712-11 Cuvette A into the holder
and ensure that the notch on the cap is
positioned securely into the groove.
4• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
3
4-5
9• At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as reported on the CAL Check™
Standard Certificate.
If the value is found out of specifications,
please check that the cuvettes are free
of fingerprints, oil or dirt and repeat
validation. If results are still found out
of specifications then recalibrate the
instrument.
Calibration
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Press and hold CAL Check™ for three
seconds to enter calibration mode. The
display will show "CAL" during calibration
procedure. The blinking "ZERO" asks for
instrument zeroing.
Place the CAL Check™ Standard
HI96712-11 Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
5• Press ZERO/CFM and the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
3''
3
4
5-6
6• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and ready
for calibration. The blinking “READ” asks
for reading calibration standard.
7• Remove the cuvette.
6• Remove the cuvette.
8• Press CAL Check key and the lamp,
cuvette and detector icons together with
“CAL Check” will appear on the display,
depending on the measurement phase.
Note: It is possible to interrupt the calibration procedure at any time by pressing
CAL CHECK or ON/OFF keys.
4
5• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for validation.
7• Place the CAL Check™ Standard
HI96712-11 Cuvette B into the holder
and ensure that the notch on the cap is
positioned securely into the groove.
Calibration
7
8
9
HI96712
8• Place the CAL Check™ Standard
HI96712-11 Cuvette B into the holder
and ensure that the notch on the cap is
positioned securely into the groove.
8
9• Press READ /TIMER and the lamp,
cuvette and detector icons will appear on
the display, depending on the measurement
phase.
9-10
10•After measurment the instrument will
show for three seconds the CAL Check™
standard value.
Note: If the display shows “STD HIGH”,
the standard value was too high. If the
display shows “STD LOW”, the standard
value was too low. Verify that both
CAL Check™ Standard Cuvettes, A and B
are free of fingerprints or dirt and that they
are inserted correctly.
11• Then the date of last calibration (e.g.:
“01.08.2009”) appears on the display,
or “01.01.2008” if the factory calibration
was selected before. In both cases the year
number is blinking, ready for date input.
11-13
HI96712
12• Press GLP/ to edit the desired year
(2000-2099). If the key is kept pressed,
the year number is automatically increased.
13• When the correct year has been set,
press ZERO/CFM or READ /TIMER to
confirm. Now the display will show the
month blinking.
14• Press GLP/ to edit the desired month
(01-12). If the key is kept pressed, the
month number is automatically increased.
15• When the correct month has been set,
press ZERO/CFM or READ /TIMER to
confirm. Now the display will show the day
blinking.
16• Press GLP/ to edit the desired day
(01-31). If the key is kept pressed, the
day number is automatically increased.
Note: It is possible to change the editing
from day to year and to month by pressing
READ/TIMER.
17• Press ZERO/CFM to save the calibration
date.
18• The instrument displays “Stor” for one
second and the calibration is saved.
19• The instrument will return automatically to
measurement mode by displaying dashes
on the LCD.
or
BATTERY MANAGEMENT
14-15
or
16
17
18
In GLP mode, the last calibration date can be verified and the factory
calibration can be restored.
Last Calibration
Last Calibration Date
Date
1• Press GLP/ to enter GLP mode. The
calibration month and day will appear
on the main display and the year on the
secondary display.
1
2• If no calibration was performed, the factory
calibration message, “F.CAL” will appear on
the main display and the instrument returns
to measurement mode after three seconds.
2
To save the battery, the instrument shuts down after 10 minutes of non-use
in measurement mode and after 1 hour of non-use in calibration mode.
If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking “ZERO”
means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup and
after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows “dEAd bAtt” and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and remove the battery cover by
turning it counterclockwise.
• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.
RECOMMENDATIONS FOR USERS
Factory Calibration
It is possible to delete the calibration and restore Restore
factory calibration.
2
1• Press GLP/ to enter GLP mode.
2• Press READ /TIMER to enter in the
factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.
3-4
4• The instrument briefly indicates “donE”
upon restoration of factory calibration
prior to returning to measurement mode.
GLP
Factory Calibration Restore
3• Press ZERO/CFM to restore the factory
GLP/ again to abort factory calibration restore.
Before using these products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to
other electronic equipments, this requiring the operator to take all necessary
steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven.
For yours and the instrument safety do not use or store the instrument in
hazardous environments.
INSTRUCTION MANUAL
HI96712
Aluminum ISM
Thank You
FUNCTIONAL DESCRIPTION
MEASUREMENT PROCEDURE
Measurement
Thank you for choosing a Hanna Instruments product. Please read
this instruction manual carefully before using the instrument.
2
For more information about Hanna Instruments and our products,
visit www.hannainst.com.
For technical support, contact your local Hanna Instruments Office
or e-mail us at [email protected]
3-5
50mL
Find your local Hanna Instruments Office at www.hannainst.com
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not
damaged. If any damage occured during shipment, please contact your local
Hanna Instruments Office.
Each HI96712 Ion Selective Meter is supplied complete with:
• Sample Cuvettes and Caps (2 pcs.)
• 9V Battery
• Instruction Manual
• Quality Certificate
Note: Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and accessories see
“Accessories”.
1. GLP/ key: press to enter GLP mode. In calibration mode press to
edit the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or press
and hold for three seconds to enter calibration mode.
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore.
4. READ /TIMER key: In measurement mode, press to make a
measurement, or press and hold for three seconds to start a preprogrammed countdown prior to measurement. In GLP mode press to
view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Cristal Display (LCD)
7. Cuvette alignment indicator
8. Cuvette holder
6
7
#1
8
#1
3''
9-10
SPECIFICATIONS
Range
0.00 to 1.00 mg/L
Resolution
0.01 mg/L
Accuracy
@25 °C (77 °F)
±0.02 mg/L ±4% of reading
Light source
Tungsten lamp
Light Detector
Silicon Photocell with narrow band interference
filter @525 nm
Method
Adaptation of the aluminon method. The reaction
between aluminum and reagents causes a reddish
tint in the sample
Environment
0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type
9V (1 pc.)
Auto-Shut off
After 10’ of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder
Dimensions
192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Weight
320 g (11.3 oz.)
or
DISPLAY ELEMENTS DESCRIPTION
11
1. The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2. Error messages and warnings
3. The battery icon shows the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display
#2
12-13
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.
3• Fill a graduated beaker with 50 mL of
sample.
4• Add the content of one packet of
HI93712A-0 Aluminum Reagent A and
mix until dissolution is complete.
5• In the same beaker add the content of
one packet of HI93712B-0 Aluminum
Reagent B and mix until dissolution is
complete. This is the sample.
6• Fill two cuvettes with 10 mL of sample
each.
7• Add the content of one packet of
HI93712C-0 Aluminum Reagent C to only
one of the cuvettes. Replace the cap and
shake gently. This is the blank.
8• Place the blank into the holder and ensure
that the notch on the cap is positioned
securely into the groove.
9• Press and hold READ /TIMER for three
seconds. The display will show the countdown prior to measurement. The beeper
is playing a beep at the end of countdown
period. Alternatively, wait for 15 minutes
and press ZERO/CFM directly. In both
cases the lamp, cuvette and detector icons
will appear on the display, depending on
the measurement phase.
10• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for measurement.
11• Remove the blank and insert the second
cuvette into the instrument.
12• Press READ /TIMER. The lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
13• At the and of measurement, the instrument
directly displays concentration in mg/L of
aluminum on the Liquid Crystal Display.
Interferences
Iron above
20 mg/L
Alkalinity above
1000 mg/L
Phosphate above
50 mg/L
Flouride will interfere at all levels.
ERRORS AND WARNINGS
Battery low: The battery must be replaced soon.
On Zero Reading:
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Dead battery: This indicates that the battery is
dead and must be replaced. Once this indication
is displayed, normal operation of the instrument
will be interrupted. Change the battery and restart
the meter.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light: The instrument cannot adjust the light
level. Please check that the sample does not contain
any debris.
On Sample Reading:
Inverted cuvettes: The sample and the zero cuvette
are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for
zeroing the meter.
Under range: A blinking “0.00” indicates that the
sample absorbs less light than the zero reference.
Check the procedure and make sure you use the
same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and re-run
the test.
During Calibration Procedure:
Standard Low: The standard reading is less than
expected.
Standard High: The standard reading is higher
than expected.
Other Errors And Warnings:
Cap error: Appears when external light enters
in the analysis cell. Assure that the cuvette cap
is present.
Cooling lamp: The instrument waits for the lamp
to cool down.
ACCESSORIES
Reagent Sets
HI93712-01
Reagents for 100 tests
HI93712-03
Reagents for 300 tests
Other Accessories
HI96712-11
CAL Check™ Standard Cuvettes (1 set)
HI740029P
9V battery (10 pcs.)
HI731318
Cloth for wiping cuvettes (4 pcs.)
HI731331
Glass cuvetes (4 pcs.)
HI731335
Caps for cuvettes (4 pcs.)
HI93703-50
Cuvette cleaning solution (230 mL)
WARRANTY
HI96712 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
If service is required, contact your local Hanna Instruments Office. If under
warranty, report the model number, date of purchase, serial number and
the nature of the failure. If the repair is not covered by the warranty, you
will be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection. To
validate your warranty, fill out and return the enclosed warranty card within
14 days from the date of purchase.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc.,
Woonsocket, Rhode Island, 02895, USA
IST96712 01/18

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement