Hanna Instruments HI122-01 Owner Manual
PDF
Download
Document
Advertisement
Advertisement
Instruction Manual HI120 & HI122 pH/mV/Temperature Bench Meters with Calibration Check ON OFF HI122 CAL pH / mV Meter CFM CLR LOG Calibration www.hannainst.com 1 Check Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using these instruments. This manual will provide you with the necessary information for correct use of these instruments, as well as a precise idea of their versatility. If you need additional technical information, do not hesitate to e‑mail us at [email protected] or view our worldwide contact list at www.hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION.................................................................. 4 GENERAL DESCRIPTION.......................................................................... 4 FUNCTIONAL DESCRIPTION HI120 ......................................................... 5 FUNCTIONAL DESCRIPTION HI122 ......................................................... 6 HI120 AND HI122 SPECIFICATIONS..................................................... 11 OPERATIONAL GUIDE........................................................................... 12 pH CALIBRATION................................................................................. 15 ENHANCED CALIBRATION MESSAGES.................................................... 20 ELECTRODE CONDITION & ELECTRODE RESPONSE TIME.......................... 22 RELATIVE mV CALIBRATION.................................................................. 22 GOOD LABORATORY PRACTICE (GLP)...................................................... 23 SETUP................................................................................................ 27 LOGGING............................................................................................. 32 AutoHOLD........................................................................................... 36 PRINTING (HI122 only)....................................................................... 36 TEMPERATURE CALIBRATION (for technical personnel only)..................... 40 mV CALIBRATION (for technical personnel only)...................................... 41 PC INTERFACE..................................................................................... 42 pH BUFFER TEMPERATURE DEPENDENCE.............................................. 44 ELECTRODE CONDITIONING & MAINTENANCE......................................... 45 PRINTER MAINTENANCE (HI122 only)................................................... 47 TROUBLESHOOTING GUIDE.................................................................. 48 TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS......................... 50 ACCESSORIES...................................................................................... 51 3 PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. If there is any damage, contact your local Hanna Instruments Office. Each instrument is supplied with: • HI1131P Glass-body Combination pH Electrode • HI7662-T Temperature probe • pH4.01 & 7.01 Buffer solutions, 20 mL each • HI7082S Electrolyte solution • 5 paper rolls (HI122 only) • 12VDC Power Adapter • Instruction Manual Note: Save all packing material until you are sure that the instrument functions correctly. All defective items must be returned in the original packing with the supplied accessories. GENERAL DESCRIPTION The Hanna HI120 and HI122 are professional bench meters for pH, ORP (Oxidation Reduction Potential) and Temperature measurements with Calibration Check. Relative mV feature is also provided. Calibration Check performs a set of diagnostic tests during calibration using the history of electrode slope and offset to detect problems that can cause loss of accuracy. Calibration Check Features are: • Enhanced Calibration Messages • Electrode Condition on LCD Display • Electrode Response Time on LCD Display • Calibration Alarm Time-Out • Out of Calibration Range Other features include: • Up to five-point calibration with seven memorized buffers (1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45 pH) and two custom buffers. • pH calibration using pH buffers with 0.001 resolution. • pH reading with manual or automatic temperature compensation. • Up to 50 samples for Log on demand mode and up to 1000 samples for AutoLOG mode. • Two selectable alarm limits. • User selectable AutoLOG modes. • Printing feature in four selectable languages (HI122 only). • Large easy-to-read LCD which shows the pH or mV and Temperature simultaneously, together with graphic symbols. 4 • AutoHOLD feature to freeze first stable reading on the LCD. • GLP feature to view last calibration data for pH and Relative mV. • PC interface. FUNCTIONAL DESCRIPTION HI120 Front Panel Electrode Condition Primary LCD Secondary LCD Electrode Response ON OFF HI120 AutoHOLD GLP pH / mV Meter SET CAL CFM CLR AutoLOG MODE pH mV RelmV Reso NUM LOG 2 4 1 3 Rear Panel PIN pH/ORP 5 1) 2) 3) 4) 5) 6) 7) 8) 9) TEMP POWER RS232 7 6 RCL Calibration Check 9 8 Left Keyboard ON switch OFF switch Right Keyboard Pin input socket BNC electrode connector Temperature probe socket Power adapter socket RS232 serial communication connector 5 FUNCTIONAL DESCRIPTION HI122 Front Panel Paper Roll Electrode Condition Primary LCD Secondary LCD Electrode Response ON OFF HI122 AutoHOLD Print pH / mV Meter Paper CAL CFM CLR AutoLOG pH mV RelmV Reso NUM LOG Calibration Check 2 4 1 3 Rear Panel PIN pH/ORP 5 POWER RS232 7 6 1) 2) 3) 4) 5) 6) 7) 8) 9) 10) TEMP 9 8 10 Left Keyboard ON switch OFF switch Right Keyboard Pin input socket BNC electrode connector Temperature probe socket Power adapter socket RS232 serial communication connector Printer 6 HI120 KEYBOARD ON THE LEFT Shortcuts to alternate functions 1) AutoHOLD key, to freeze the first stable reading on the LCD. 2) GLP key, to display Good Laboratory Practice Information. 3) SET key, to enter/exit SETUP mode. 4) key, to toggle between parameters while in RECALL or SETUP modifying mode (backwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode. 5) AutoLOG key, to start/stop AutoLOG mode. 6) key, to toggle between parameters while in RECALL or SETUP modi‑ fying mode (forwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode. 7) pH key, to select pH range. 8) mV key, to select mV range. 9) RelmV key, to select Relative mV range. 11) Resolution key, to select pH resolution. Numerical keys 10) key, to change sign. 12) NUM key, to activate the numerical keys (0 to 9). 7 HI120 MAIN KEYBOARD ON THE RIGHT HI120 pH / mV Meter 3 1 2 CAL CFM CLR 5 6 4 MODE 7 9 8 LOG RCL Calibration Check 1) CAL key, to enter and exit/escape calibration mode. SET key (second function), to enter/exit SETUP mode. 2) CFM key, to confirm different values. GLP key (second function), to display Good Laboratory Practice Infor‑ mation. 3) CLR key, to clear calibration or logged data. 4) key, to manually increase temperature value or other parameters. AutoHOLD key (second function), to freeze the first stable reading on the LCD. 5) key, to manually decrease temperature value or other parameters. AutoLOG key (second function), to start/stop AutoLOG mode. 6) MODE key, to select the measurement unit or to switch focused data. 7) 2nd key, to select second key function. 8) LOG key, to store measured data. 9) RCL key, to enter/exit view logged data mode. Note: To select second key function, press 2nd and then the desired key. The “2nd” tag will appear on the LCD until the desired key is pressed. To leave second key function selection, press 2nd again. 8 HI122 KEYBOARD ON THE LEFT Shortcuts to alternate functions 1) AutoHOLD key, to freeze the first stable reading on the LCD. 2) Print key, to obtain a printout or to cancel printing. 3) Paper key, to pull out the paper. 4) key, to toggle between parameters while in RECALL or SETUP modifying mode (backwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode. 5) AutoLOG key, to start/stop AutoLOG mode. 6) key, to toggle between parameters while in RECALL or SETUP modi‑ fying mode (forwards), to toggle between absolute mV and temperature while in Relative mV mode and between pH buffer and temperature while in pH calibration mode. 7) pH key, to select pH range. 8) mV key, to select mV range. 9) RelmV key, to select Relative mV range. 11) Resolution key, to select pH resolution. Numerical keys 10) key, to change sign. 12) NUM key, to activate the numerical keys (0 to 9). 9 HI122 MAIN KEYBOARD ON THE RIGHT HI122 pH / mV Meter 3 1 2 CAL CFM CLR 5 6 4 7 9 8 LOG Calibration Check 1) CAL key, to enter and exit/escape calibration mode. SET key (second function), to enter/exit SETUP mode. 2) CFM key, to confirm different values. GLP key (second function), to display Good Laboratory Practice Infor‑ mation. 3) CLR key, to clear calibration or logged data. 4) key, to manually increase temperature value or other parameters. AutoHOLD key (second function), to freeze the first stable reading on the LCD. 5) key, to manually decrease temperature value or other parameters. AutoLOG key (second function), to start/stop AutoLOG mode. 6) MODE key, to select the measurement unit or to switch focused data. Paper key (second function), to pull out the paper. 7) 2nd key, to select second key function. 8) LOG key, to store measured data. 9) RCL key, to enter/exit view logged data mode. Print key (second function), to obtain a printout or to cancel printing. Note: To select second key function, press 2nd and then the desired key. The “2nd” tag will appear on the LCD until the desired key is pressed. To leave second key function selection, press 2nd again. 10 HI120 AND HI122 SPECIFICATIONS –2.00 to 16.00 pH –2.000 to 16.000 pH RANGE ±999.9 mV ±2000 mV –20.0 to 120.0 ºC (–4.0 to 248.0 ºF) 0.01 pH 0.001 pH RESOLUTION 0.1 mV 1 mV 0.1 ºC (0.1 ºF) ACCURACY @ 25 °C / 68 °F ±0.2 mV (±699.9 mV) ±0.5 mV (±999.9 mV) ±1 mV (± 2000 mV) ±0.4 ºC (±0.7 ºF) (excluding probe error) Relative mV offset range ±2000 mV pH Calibration Up to five-point calibration, 7 standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and 2 custom buffers Temperature compensation Manual or Automatic from: –20.0 to 120.0 ºC (–4.0 to 248.0 ºF) pH Electrode HI1131P Temperature probe HI7662 Logging Interval 5 seconds to 180 minutes Printer (HI122 only) Dot matrix, 44 mm wide paper PC interface opto-isolated RS232 Input impedance 1012 ohm Power supply 12 VDC adapter Dimensions 280 x 203 x 84 mm (11.0 x 8.0 x 3.3”) Weight 1.8 Kg (4.1 lb); kit with printer: 1.9 Kg (4.2 lb) Environment 0 – 50 ºC (32 – 122 ºF) max. 95% RH non-condensing Warranty 2 years 11 OPERATIONAL GUIDE POWER CONNECTION Plug the 12 VDC adapter into the power supply socket. Notes: • These instruments use non volatile memory to retain the pH, mV, temperature calibrations and all other settings, even when unplugged. • Make sure a fuse protects the main line. ELECTRODE AND PROBE CONNECTIONS For Hanna Instruments P type pH or ORP electrodes (with internal reference) connect the electrode’s BNC to the socket on the back of the instrument (#6 on page 4 and 5) and the pin to the appropiate socket (#5 on page 4 and 5). Notes: • Electrode condition and response information is displayed on the bar graph gauges during the day the calibration is performed only if Hanna Instruments P type (PIN) electrodes are used. • If the electrode is not recognized as a Hanna Instruments P type electrode, the bar graph gauges will blink (25 seconds OFF, 4 seconds ON, full bar graph). For temperature measurements and automatic temperature compensation, connect the temperature probe to the appropriate socket. INSTRUMENT START-UP • Turn the instrument on by pressing the ON switch. • All LCD tags are displayed and a beep is heard while the instrument performs a self test. • The instrument displays then the date on the primary LCD and the time on the secondary LCD, along with the “Remove protective cap” and “Unscrew electrode refilling cap” messages alternatively blinking. These messages alert the user to follow displayed instructions in order to take proper measurements and to improve electrode response. • The instrument automatically defaults to pH or mV measurement mode, if a Hanna Instruments P type pH or ORP electrode is detected. • If no Hanna Instruments P type electrode is detected, the instrument starts in the same range as it was at power off. 12 pH MEASUREMENTS Make sure the instrument has been calibrated before taking pH measurements. • Press MODE to enter pH mode. Note: To change pH resolution, press MODE again or simply Resolution from the left keyboard. • Submerge the electrode tip and the temperature probe approximately 4 cm (1½”) into the sample to be tested. Allow time for the electrode to stabilize. • The pH is displayed on the primary LCD and the temperature on the secondary LCD. OUT CAL RANGE feature warns the user if the current reading is out of the calibrated area. The calibrated area is that part of the pH range in which the calibration point assures an accurate reading. If the reading is taken out of the calibration area, the “OUT CAL RANGE” message will blink. The calibrated area is calculated according to the pH resolution used during the reading. To avoid having this message, the calibration points have to be well distributed in the desired measurement range. If measurements are taken successively in different samples, it is recommended to rinse the electrode thoroughly with deionized water or tap water and then with some of the next sample to prevent cross-contamination and to condition the electrode before immersing it into the sample solution. The pH reading is affected by temperature. In order to measure the pH accurately, the temperature effect must be compensated for. To use the Automatic Temperature Compensation feature, connect and submerge the HI7662-T temperature probe into the sample as close as possible to the electrode and wait for a few seconds. If the temperature of the sample is known, manual temperature compensation can be performed by disconnecting the temperature probe. The display will then show the default temperature of 25 ºC (77 ºF) or the last temperature reading with the “ºC” (or “ºF”) tag blinking. The temperature can be adjusted with the ARROW keys or the numerical keypad (from –20.0 ºC to 120.0 ºC or from –4.0 ºF to 248.0 ºF). Press NUM to change the temperature value with the numerical keys. The “2nd” tag will blink. 13 Press CLR if you want to delete digits of the displayed value. The remaining digits will shift to right. Introduce the desired value. If the value is out of temperature range, a long beep will be heard. Press NUM to confirm the new value or CAL to escape without changing the temperature. mV/ORP MEASUREMENTS Oxidation-reduction potential (REDOX) measurements provide the quantification of the oxidizing or reducing power of the tested sample. To correctly perform a redox measurement, the surface of the ORP electrode must be clean and smooth. • Press MODE or simply mV from the left keyboard to enter mV range. • Submerge the tip of the ORP electrode (4 cm/1½”) into the sample to be tested and allow a few seconds for the reading to stabilize. • The instruments display the mV reading on the primary LCD and the temperature on the secondary LCD. • If the reading is out of range, the closest full-scale value will be displayed blinking on the primary LCD. RELATIVE mV MEASUREMENTS To enter Relative mV mode, press MODE or simply RelmV from the left keyboard. The relative mV reading will be displayed on the primary LCD and the current temperature value on the secondary LCD. Note: Press or from the left keypad to toggle between temperature and absolute mV reading on the secondary LCD. The Relative mV reading is equal to the difference between the absolute mV input value and relative mV offset established in the relative mV calibration. TEMPERATURE MEASUREMENTS Connect the HI7662-T temperature probe to the appropriate socket. Immerse the temperature probe into the sample and allow the reading on the secondary LCD to stabilize. 14 Note: The temperature can be displayed in Celsius degrees (º C) or in Fahrenheit degrees (º F) (see SETUP for details, page 27). pH CALIBRATION Calibrate the instrument often, especially if high accuracy is required. The instrument should be recalibrated: • Whenever the pH electrode is replaced. • At least once a week. • After testing aggressive chemicals. • If “CAL DUE” tags are blinking during measurement. • If “OUT CAL RANGE” message blinks during pH measurement (the measurement range is not covered by current calibration). PREPARATION Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic or glass beakers to minimize any EMC interferences. For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution. One for rinsing the electrode and one for calibration. If you are measuring in the acidic range, use pH7.01 or 6.86 as first buffer and pH4.01 or 1.68 as second buffer. If you are measuring in the alkaline range, use pH7.01 or 6.86 as first buffer and pH10.01/9.18 or 12.45 as second buffer. For extended range measurements (acidic and alkaline), perform a five-point calibration by selecting five of the available buffers. PROCEDURE If 0.001 pH resolution is selected, each selected standard buffer value can be updated according to the value on the production lot certificate at 25 ºC (77 ºF). Press 2nd then SET key when a standard pH buffer with 0.001 resolution is selected. The buffer value will start blinking and it can be changed with the ARROW keys in a ±0.020 pH window. Calibration has a choice of 7 memorized buffers: pH1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 and 2 custom buffers. The custom buffers are a special option that allows the user to calibrate in a buffer solution different from a standard one. Up to two custom buffers can be set in SETUP menu (see page 27). When selected during calibration, the “CUSTOM C1” or “CUSTOM C2” tags are displayed on the LCD and the custom buffer value can be changed in a ±1.0 pH window, around the set value. For accurate measurements it is recommended to perform a five-point calibration. However, at least a two-point calibration is suggested. 15 The instruments will automatically skip the buffers used during calibration and the buffers which are in a ±0.2 pH window, around one of the calibrated buffers. All new calibrations will override existing stored calibration data, in a ±0.2 pH window, at these calibration points. The slopes adjacent to the calibration points will be reevaluated. If the new calibration point has no correspondence in the existing stored calibration data, it is added to it if this is not full, or the instrument will ask which buffer will be replaced by the current buffer. If at least a two-point calibration has been performed and an offset correction of the electrode is desired, keeping unchanged the existing slopes, perform a one-point calibration with “OFFS” option selected in SETUP menu. If “Pnt” option is selected, the slopes adjacent to the calibration points will be reevaluated. FIVE-POINT CALIBRATION • Immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into a buffer solution of your choice (pH1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45, custom buffer 1 or 2, if these were set) and stir gently. The temperature probe should be close to the pH electrode. • Press CAL. The “CLEAR CAL IF NEW ELECTRODE” message will be displayed blinking on the LCD for a few seconds if the instrument was calibrated before CAL and calibration was not cleared. Press CLR if you are using a new electrode or want to clear calibration history, or wait a few seconds to continue. Press CAL again or the ARROW keys to skip this message. Note: It is very important to clear calibration history when a new electrode is used because most errors and warning messages that appear during calibration depend on calibration history. • The instruments will display the measured pH on the primary LCD and the “7.01” buffer on the secondary LCD, together with “CAL” and “Cal Point 1” tags and “7.01” tag blinking. 16 • If necessary, press the ARROW keys to select a different buffer value. • The “ ” tag will blink on the LCD until the reading is stable. • When the reading is stable and close to the selected buffer, “CFM” tag blinks. • Press CFM to confirm calibration. • The calibrated value is then displayed on the primary CFM LCD and the secondary LCD will display the second expected buffer value, together with “CAL”, “Cal Point 2” and “7.01” tags and the corresponding buffer tag blinking. • After the first calibration point is confirmed, immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into the second buffer solution and stir gently. The temperature probe should be close to the pH electrode. • If necessary, press the ARROW keys to select a different buffer value. • The “ ” tag will blink on the LCD until the reading is stable. • When the reading is stable and close to the selected buffer, “CFM” tag blinks. • Press CFM to confirm calibration. • The calibrated value is then displayed on the primary CFM LCD and the secondary LCD will display the third expected buffer value. • After the second calibration point is confirmed, immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into the third buffer solution and stir gently. The temperature probe should be close to the pH electrode. • If necessary, press the ARROW keys to select a different buffer value. • The “ ” tag will blink on the LCD until the reading is stable. 17 • When the reading is stable and close to the selected buffer, “CFM” tag blinks. • Press CFM to confirm calibration. • The calibrated value is then displayed on the primary LCD and the secondary LCD will display the fourth CFM expected buffer value. • After the third calibration point is confirmed, immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into the fourth buffer solution and stir gently. The temperature probe should be close to the pH electrode. • If necessary, press the ARROW keys to select a different buffer value. • The “ ” tag will blink on the LCD until the reading is stable. • When the reading is stable and close to the selected buffer, “CFM” tag blinks. • Press CFM to confirm calibration. • The calibrated value is then displayed on the primary CFM LCD and the secondary LCD will display the fifth expected buffer value. • After the fourth calibration point is confirmed, immerse the pH electrode and the temperature probe approximately 4 cm (1½”) into the fifth buffer solution and stir gently. The temperature probe should be close to the pH electrode. • If necessary, press the ARROW keys to select a different buffer value. • The “ ” tag will blink on the LCD until the reading is stable. • When the reading is stable and close to the selected buffer, “CFM” tag blinks. • Press CFM to confirm calibration. • The instruments store the calibration values and CFM return to normal measurement mode. FOUR, THREE or TWO-POINT CALIBRATION • Proceed as described in “FIVE-POINT CALIBRATION” section. • Press CAL after the appropriate accepted calibration point. The instru‑ ments will return to measurement mode and will memorize the calibration data. ONE-POINT CALIBRATION Two SETUP selectable options are available: “Pnt” and “OFFS”. If the “Pnt” option is selected, the adjacent slopes will be reevaluated. If the “OFFS” option is selected, an electrode offset correction is performed keeping unchanged the existing slopes. • Proceed as described in “FIVE-POINT CALIBRATION” section. • Press CAL after the first calibration point was confirmed. The instru‑ ments will memorize the one-point calibration data and will return to measurement mode. 18 Notes: • Press MODE or / from the left keyboard to toggle between pH buffer and temperature reading during calibration. • Each time a buffer is confirmed, the new calibration parameters replace the old calibration parameters of the corresponding buffer. If current confirmed buffer has no correspondence in the existing stored calibration and this is not full, the current buffer is added to the existing stored calibration. If the existing stored calibration is full (five calibration points), the instrument ask which buffer will be replaced by current buffer. Press the ARROW keys to select another buffer to be replaced. Press CFM to confirm the buffer that will be replaced. Press CAL to leave calibration without replacing. Note: The replaced buffer is not removed from calibration list and it can be selected for the next calibration points. WORKING WITH CUSTOM BUFFERS If at least one custom buffer was set in SETUP menu, it can be selected for calibration by pressing the ARROW keys. Press 2nd then SET key if you want to adjust the buffer value. The buffer value, displayed on the secondary LCD, will start blinking. Use the ARROW keys to change the buffer value. After about 5 seconds you performed the last change, the buffer value is updated. Press 2nd then SET key if you want to change it again. Note: Custom buffer value can be adjusted in a ±1.00 pH window, around the set value. 19 CLEAR CALIBRATION Press CLR in any moment during calibration. The “CLEAR CAL” tag will appear and “donE” message will be displayed on the secondary LCD. All old calibrations, starting with current selected buffer are cleared and the instrument continues calibration. If CLR is pressed when “CLEAR CAL IF NEW ELECTRODE” message is displayed blinking, the calibration history is deleted and the instrument will display “hiSt” message on the primary LCD and “donE” message on the secondary LCD, along with “CLEAR CAL” tags. The calibration process will continue starting with 7.01 pH buffer as first buffer. Note: If CLR is pressed during the first calibration point, the instrument returns to measurement mode. CLEAR CALIBRATION Press CLR in any moment during calibration. The “CLEAR CAL” tag will appear and “donE” message will be displayed on the secondary LCD. All old calibrations, starting with current selected buffer are cleared and the instrument continues calibration. If CLR is pressed when “CLEAR CAL IF NEW ELECTRODE” message is displayed blinking, the calibration history is deleted and the instrument will display “hiSt” message on the primary LCD and “donE” message on the secondary LCD, along with “CLEAR CAL” tags. The calibration process will continue starting with 7.01pH buffer as first buffer. Note: If CLR is pressed during the first calibration point, the instrument returns to measurement mode. ENHANCED CALIBRATION MESSAGES The stored calibration history to used issue error and warning messages during calibration to help ensure the highest accuracy. As electrode aging is normally a slow process, substantial changes from previous calibrations are likely due to a temporary problem with the electrode or buffers. Calibrating under these conditions will result measurement errors. ERROR MESSAGES Error messages appear if one or all of the calibration parameters are out of accepted windows. When these messages are displayed, calibration cannot be confirmed. WRONG BUFFER This message appears when the difference between the pH reading and the value of the selected buffer is too big. If this error message is displayed, check if you have selected the proper calibration buffer. 20 WRONG BUFFER TEMPERATURE This message appears if the temperature of the buffer is outside the defined buffer temperature range. WRONG & CONTAMINATED BUFFER / CHECK ELECTRODE This message appears if the buffer used is contaminated or the electrode is supposed to be either broken or very dirty. WRONG & CHECK ELECTRODE / CLEAN ELECTRODE This message appears if the electrode is supposed to be broken or very dirty. WRONG & OLD / CLEAR CAL & OLD This message appears as a result of an erroneous slope condition. If the slope between current calibration point and one of the previous calibration, that was not overriden in current calibration, exceeds slope window (80% to 110%), this message will appear. Press CLR to clear old parameters and continue calibration process or CAL to leave calibration. WARNING MESSAGES During calibration, the Calibration Check feature analyzes the electrode calibration history and warns the user when problems have been detected. It is possible to override the warning messages and confirm the calibration but it is not recommended. CLEAR CAL IF NEW ELECTRODE This warning is displayed any time you enter calibration and calibration is not cleared or the new calibration parameters are better than the previous ones. You can clear calibration history by pressing CLR, or continue by pressing CAL. CLEAN ELECTRODE This warning message appears in order to alert the user that some dirt or deposits could be on the electrode. Refer to the electrode Cleaning Procedure. This ensures the removal of film, dirt or deposits on the glass bulb and reference junction. CONTAMINATED BUFFER This warning message appears in order to alert that the buffer could be contaminated. Refresh your buffer and continue the calibration procedure. CHECK ELECTRODE / CHECK BUFFER This warning message appears when the electrode slope exceeds the highest accepted slope limit (110%). Check your electrode and use fresh buffers. CLEAN ELECTRODE / CHECK BUFFER This warning appears in order to alert the user that the electrode can be dirty or the buffer contaminated. Refer to the electrode Cleaning Procedure and use fresh buffers. 21 ELECTRODE CONDITION & ELECTRODE RESPONSE TIME When using an appropriate Hanna Instruments BNC electrode with pin, the instrument will assess electrode condition and response time during each calibration and will display the calibration status for the rest of the day. The digital gauge for electrode condition is a representation of the offset and slope performance of the electrode. The response gauge is a function of the stabilization time between the first and second calibration buffers. These gauges reflect electrode performance and should be expected to slowly decrease over the life of the electrode. The condition and response gauges show the electrode’s condition at the time of calibration only and are displayed for the rest of the day the calibration is performed. For a continuous display of electrode condition at the time of calibration, daily calibration is necessary. The electrode condition and response time are also visible when viewing GLP data. If the instrument is not calibrated, it has been calibrated only in one point, or if calibration history was deleted, the electrode condition and response gauges will be empty. When using an appropriate Hanna Instruments BNC electrode with pin, the instrument will assess electrode condition and response time during each calibration and will display the calibration status for the rest of the day. The electrode response is evaluated only when calibration has been performed using pH 7.01 or pH 6.86, pH 4.01 and pH 10.01 or pH 9.18 buffers. When the instrument cannot evaluate the electrode response or pH 1.68/12.45 buffer was used as calibration buffer, the response gauge will be empty. If the electrode is in a very poor condition, the first condition segment will blink. If the electrode response is very slow, the first response segment will blink. RELATIVE mV CALIBRATION • Press CAL when the instrument is in RELATIVE mV measurement mode. The relative mV value is displayed on the primary LCD and the absolute mV value on the secondary LCD. • Use the ARROW keys or the numerical keys if you want to change the displayed relative mV value. 22 Notes: • Press MODE to select another resolution if the displayed value allows it (e.g. if 199.9 is displayed, by pressing MODE the value will change to 1999; if 19.9 is displayed, nothing will happen if pressing MODE). • Pressing CLR the displayed value is set to 0.0 mV. • Press NUM from the left keyboard if you want to change the value using the numerical keys. The “2nd” tag will blink. • Press CLR to delete the last digit. • Press MODE to select another resolution (see above). • Press NUM again to leave the numerical keyboard. • When the reading is stable, in mV range and the Relative mV offset is inside the offset window (±2000 mV), “CFM” tag blinks. • Press CFM to confirm relative mV calibration. The instrument will return to measurement mode. • If the absolute mV reading is out of range or the Relative mV offset is out of the offset window, “WRONG” tag will blink. Change the input value or the Relative mV offset to complete the calibration process. GOOD LABORATORY PRACTICE (GLP) GLP is a set of functions that allows storage and retrival of data regarding the maintenance and status of the electrode. All data regarding pH and Rel mV calibration is stored for the user to review when necessary. CALIBRATION ALARM TIME OUT For pH calibration, all the instruments allow the user to set the number of days before the next required pH calibration. This value can be set from 1 to 7 days. The default setting is OFF (disabled). The instrument checks if the time-out has expired. If the time elapsed, “CAL DUE” tags will blink as a reminder. Note: If the instrument was not calibrated or all calibration parameters were cleared, “CAL” “DUE” tags will be displayed even if the feature is disabled in SETUP menu. LAST pH CALIBRATION DATA The last pH calibration data is stored automatically after a successful calibration. To view the pH calibration data, press 2nd then GLP key or simply GLP from the left keyboard (HI120) when the instrument is in pH (mV) measurement mode. 23 The instrument will display the date (yyyy.mm.dd) of the last calibration. Press the ARROW keys to view the next calibration parameter (pressing the key): • The time (hh:mm) of the last calibration. • The pH calibration offset. • The pH calibration slope (the GLP slope is the average of the calibration slopes; the percentage is referred to the ideal value of 59.16 mV/pH). 24 • The calibration buffers in calibrating order, with the corresponding warnings. The first pH calibration buffer: The second pH calibration buffer: The third pH calibration buffer: The fourth pH calibration buffer: The fifth pH calibration buffer: 25 Notes: • The “OLd” message displayed beside the pH value means that this buffer was not used during last calibration. Press 2nd then SET key if you want to see calibration date (or time, if old calibration was performed in the same day with current calibration). • Each calibration buffer is displayed with the resolution from calibration moment. • If “no bUF” message appears on the LCD, the instrument informs you that calibration was performed in less than five points. • The Calibration Alarm Time Out status. If disabled, or the number of days until the calibration alarm will be displayed (e.g. 5 days), or from the time calibration expired (e.g. –3 days). • The instrument ID. LAST RELATIVE mV CALIBRATION DATA Last Relative mV calibration data is stored automatically after a successful calibration. To view the Relative mV calibration data, press 2nd then GLP key or simply GLP from the left keyboard (HI120) while in Relative mV measurement mode. The instrument will display the Relative mV GLP information. • The date (yyyy.mm.dd) of the last calibration as in pH GLP mode. Press the ARROW keys to view the next logged calibration parameter (pressing the key): • The Relative mV calibration offset and time (hh:mm) as in pH GLP mode. • The instrument ID as in pH GLP mode. 26 Notes: • Press 2nd then GLP key at any moment and the instrument will return to measurement mode. • If calibration has not been performed, the instrument displays “no CAL” message blinking. SETUP Setup mode allows viewing and modifying the following parameters: • Calibration Alarm Time Out • One-point calibration behaviour • Custom Buffer 1 • Custom Buffer 2 • Alarm High Limit • Alarm Low Limit • AutoLOG Start Condition • AutoLOG End Condition • AutoLOG Interval • Temperature Unit • Current Date (yyyy.mm.dd) • Current Time (hh:mm) • Printing Language (HI122 only) • Printer Status (HI122 only) • Beep Status • Baud Rate (serial communication) • Command prefix (serial communication) • Instrument ID 27 To enter SETUP mode, press 2nd then SET key, or simply SET from the left keyboard (HI120 only), while the instrument is in measurement mode. Select the desired setup parameter using the ARROW keys. Press CAL if you want to change the item value. The selected item (e.g. hour, in setting up the correct time) and “CFM” tag will start blinking. Press the ARROW keys to change the displayed value or simply use the numerical keys for all numerical parameters. Note: To use the numerical keys, press NUM from the left keyboard. The “2nd” tag will blink. Set the desired number digit by digit. The new introduced digit is always the last one. All the previous digits will shift to left. To delete digits press CLR. Last introduced digit will be deleted and the number will shift to right. If there is another item to be set (e.g. minutes), press MODE or / from the left keyboard. The other item will start blinking. Press the ARROW keys to change the displayed value or simply use the numerical keys for all numerical parameters. Press CFM to confirm or CAL to escape. Press the ARROW keys to select the next/previous parameter. Press 2nd then SET key, or simply SET from the left keyboard (HI120 only) to exit SETUP menu at any time. The instrument asks for printing a Setup Report (HI122 only). Press CFM to print the Setup Report or CAL to return to measurement mode. The following table lists the SETUP parameters, their valid values range and the factory settings (default): 28 Item Description Valid values Default CAL DUE Alarm Time Out OFF or 1 to 7 days OFF 1Pnt Pnt or OFFS Pnt One-point cal. behaviour Custom C1 Custom Buffer 1-2.00 to 16.00 pH no Custom C2 Custom Buffer 2-2.00 to 16.00 pH no AL.HI Alarm High Limit pH/mV/RelmV ranges no AL.LO Alarm Low Limit pH/mV/RelmV ranges no Strt AutoLOG Start Condition See Time/Date or “btn” “btn” End AutoLOG End Condition “dur”, “SAMP”, “rdG” “dur” Interval AutoLOG Interval 5 s to 180 min 5s tEMP Temperature Unit ºC or ºF ºC Date Date (yyyy.mm.dd) 2000.01.01 to 2099.12.31 Time Time (hh:mm) 00:00 to 23:59 LAnG Printing Language EnG, ItA, ESP, FrA Printer Status ON/OFF OFF Beep Status ON/OFF OFF bEEP 2004.01.01 English USA bAUd Baud Rate 600; 1200; 2400; 4800; 9600 PrEF Command Prefix 0 to 47 In Id Instrument ID 0000 to 9999 29 00:00 2400 16 0000 Notes: • The custom buffers can be set only with 0.001 pH resolution. If 0.01 pH resolution is selected during calibration, the displayed custom buffer value is a rounded one. • For calibration alarm time out, custom buffers, alarm limits, start and stop condition items, if pressing CLR while in changing mode, the selected item will be set to default. • Printer status and printing language items appear only for HI122. The available printing languages are: EnG (English), ItA (Italiano), ESP (Español) and FrA (Français). ALARMS SETUP • Select one of the alarm items. The displayed alarm value will be the previous set one. • Press MODE to select the range for alarm. The corresponding range tag will blink. • Press CAL to enter in changing mode. Set the new value using the ARROW or the numerical keys. Notes: • Press MODE while in changing mode to select another resolution if the displayed value allows it (e.g. if 199.9 is displayed, by pressing MODE the value will change to 1999; if 19.9 is displayed, nothing will happen if pressing MODE). • Pressing CLR the displayed value will be set to default (“no”). The first displayed value after “no”, if one of the ARROW keys is pressed, will be 7.00 pH or 0.0 mV, according to the selected range. • Press NUM from the left keyboard if you want to change the value using the numerical keys. The “2nd” tag will blink. • Press CLR to delete the last digit. • Press MODE to select another resolution (see above). • Press NUM or CFM to confirm the introduced value from the numerical keyboard. When in normal measurement mode: • If only “AL.LO” item is set, the instrument will beep when the reading is below alarm low value. • If only “AL.HI” item is set, the instrument will beep when the reading is above alarm high value. • If both alarms are set, the instrument will beep when the reading is above alarm high value or below alarm low value. Note: If “AL.HI” item is less than or equal to “AL.LO” item, “WRONG” tag will blink. 30 AutoLOG SETUP AutoLOG Start Condition • Select the “Strt” item. The displayed AutoLOG start condition will be the previous set one. The default value is “btn” – button (start with key). • Press and keep hold down MODE while in view mode, when time is displayed, to view the set date. • If pressing CAL while in view mode, the start time will be displayed. Notes: • Pressing CLR while in changing mode, the start condition will be set to default (“btn”). • The first displayed value after “btn” will be the current time/date. AutoLOG End Condition • Select the “End” item. The displayed AutoLOG end condition will be the previous set one. One of the following options will be available: “dur” (duration), “SAMP” (samples number), “rdG” (reading limits). • Pressing MODE while in view mode, the instrument will display one of the options: “dur”, “SAMP”, “rdG” & “pH”, “rdG” & “mV”, “rdG” & “Rel mV”. • Pressing CAL, the instrument will enter in changing mode in according with the selected option and will display one of the following messages: • “dur” on the secondary LCD and duration (hh:mm) on the primary LCD, along with “TIME” tag. The default value is 3 hours. The duration can be set from 1 minute to 199 hours and 59 minutes. Note: Pressing CLR while in changing mode, the end condition value will be set to default (3 hours). • “SAMP” on the secondary LCD and sample number on the primary LCD. The default value is 1000 samples. Note: Pressing CLR while in changing mode, the end condition value will be set to default (1000 samples). • “rdG” on the secondary LCD and last set limit value on the primary LCD, with the corresponding range tag blinking. The default value is 7.00 pH or 0.0 mV, in according with the selected range. Notes: • Pressing MODE while in view mode, another resolution is selected. • Pressing CLR while in changing mode, the end condition will be set to default (7.00 pH or 0.0 mV). The AutoLOG will be started in according with “Strt” item, from button (“btn”) by pressing AutoLOG or when the starting time condition is reached. The AutoLOG will stop in according with the selected option for “AutoLOG End Condition” item or when AutoLOG is pressed. 31 LOGGING This feature allows the user to log pH and Rel mV, together with temperature au‑ tomatically. All logged data can be transferred to a PC through the RS232 port. The maximum logging space is 1000 record locations (samples). This can be divided in 1 up to 50 lots (one lot can use all the free space). The lot ID (number) is between 1 and 99 (after 99 it restarts from 1). Only up to 50 lots can be memorized at one time, even if there is free space available. Note: When the logged lots are more than one, the total number of samples can be less than 1000, even with full memory, due to the fact that logging memory is divided in pages of 20 samples each. The appropriate logging interval can be set from 5 seconds to 180 minutes (see SETUP for details, page 27). Up to 50 record locations are also provided. LOGGING THE CURRENT DATA (LOG ON DEMAND) To store the current reading into memory, press LOG while in measurement mode. LOG The instrument will display the current date (mm.dd) on the primary LCD, the record number on the secondary LCD, the “LOG” tag blinking for a few seconds and then the free locations number. If there are less than 6 memory locations remaining, the record number and “Lo” message will blink for a few seconds to alert the user and then the free locations number is displayed on the LCD. If the LOG space is full, “FULL LOC” message will be displayed on the LCD for a few seconds with “LOG” tag blinking, and then ”FrEE 0” message. 32 The instrument returns to normal measurement mode. AutoLOGGING The setable AutoLOG modes make these instruments very useful in a wide range of applications: • Start at button or set time/date; stop after a set duration. • Start at button or set time/date; stop when a set value is reached. • Start at button or set time/date; stop when a set sample number is reached. To start autologging using the keyboard, press 2nd then AutoLOG key or simply AutoLOG from the left keyboard while the instrument is in measurement mode. The instrument will display for a few seconds the lot number on the primary LCD and the AutoLOG interval on the secondary LCD, together with “AutoLOG” tag. Notes: • For the other AutoLOG modes, the autologging starts/stops automatically after the set start/stop condition in SETUP menu is reached and the stop condition is correct. • The “AutoLOG” tag will blink if one of these modes is selected and the start condition is not reached. After data logging is started, the current value is displayed on the primary LCD and the temperature on the secondary LCD, along with “AutoLOG” tag. 33 To stop autologging, press 2nd then AutoLOG key again, or simply AutoLOG from the left keyboard. The instrument will display for a few seconds the lot number on the primary LCD and the sample number on the secondary LCD, together with “AutoLOG” tag, and will return to measurement mode. Notes: • If printer is ON, each logged sample is printed only if the AutoLOG interval is at least 30 s (HI122 only). • If the AutoLOG pages are full, the “FULL LOC” message will be displayed on the LCD, as in Log on demand mode. VIEW LOGGED DATA Press RCL to retrieve the information stored while in measurement mode. If no data were logged, the instrument displays: Otherwise, the instrument will display the last pH or Rel mV memorized reading on the primary LCD and the lot number on the secondary LCD. Note: If LOG mode was invoked while the instrument was in mV/Relative mV measurement range, the corresponding memorized reading will be displayed on the primary LCD, along with “RelmV” tag. 34 Press 2nd then SET key while in RECALL mode and the instrument will toggle between the lot or record number on the secondary LCD. Use the ARROW keys to select another lot or record. Press MODE or from the left keyboard and the instrument will display the next logged parameter as shown in the table below: Parameter Primary LCD Secondary LCD mV mV reading Temperature TIME Hour & minutes Seconds DATE Year Month & day OFFSET Offset value Lot (or record) number SLOPE Slope Value Lot (or record) number from the left keyboard, the previous logged Notes: • If pressing parameter will be displayed. • If in Rel mV RECALL mode regarding the slope, the instrument will display “----” message on the primary LCD. • The record number refers to an identification number inside a lot. Last displayed parameter is “dEL” message on the primary LCD and the lot number on the secondary LCD. Notes: • The “LdM” message on the secondary LCD shows that data has been stored in Log on demand mode. •Pressing 2nd then SET key, the instrument toggles between lot number, record number (Log on demand only) or all lots. • Press CLR to delete the selected lot/record or all lots. Note: Positions remain free in the Log on demand lot by deleting the last logged samples or all the lot. • If “dEL ALL” option was selected, all logged data is deleted and the instrument returns to measurement mode. Press RCL at any time to return to measurement mode. 35 AutoHOLD To freeze the first stable reading on the LCD, press 2nd then AutoHOLD key, or simply AutoHOLD from the left keyboard while the instrument is in measurement mode. The “Auto” ”HOLD” tags will be displayed blinking on the LCD until the reading will stabilize. When the reading is stable, the “Auto” “HOLD” tags stop blinking and the reading is frozen. Press 2nd then AutoHOLD key again, or simply AutoHOLD from the left keyboard to return to normal measurement mode. Note: Pressing MODE or the specific range keys from the left keyboard, the instrument will skip to the displayed range, without leaving AutoHOLD mode. The LOG key also holds AutoHOLD mode. Pressing 2nd then SET, GLP or RCL key, the instrument leaves AutoHOLD mode and performs the selected function. PRINTING (HI122 only) A complete set of information based on the measured, set or recorded data can be printed. Data can be printed on demand (for current reading in measurement mode, GLP and SETUP modes) by pressing 2nd then Print key, or simply Print from the left keyboard, or automatic (for AutoLOG and Log on demand modes). Automatic data printing is possible in AutoLOG mode only if the printer is ON and the AutoLOG interval is greater than 30 seconds. Note: To cancel printing, press 2nd then Print key again or simply Print from the left keyboard. When in pH measurement mode, the printout provides the following information: 36 Note: For Relative mV measurement mode, the Rel. mV value will be printed. When in GLP mode, the printout provides the following information: • For pH range: • For Rel mV range: When in SETUP mode, the following information can be printed: • If 2nd then Print key are the first keys pressed after entering SETUP mode, a SETUP table of contents will be printed: 37 • When exiting SETUP mode, the instrument asks if a SETUP REPORT shall be printed. The “Prn” message will be displayed on the LCD, together with “ ” tag and “CFM” tag blinking. • Press CFM to print the SETUP REPORT or CAL to escape without printing. When in SETUP mode, if pressing CAL, then Print for a chosen parameter, a help printout will come out, providing the following information (e.g. AutoLOG Interval): 38 When in LOG mode, a printout will automatically come out, providing the following information: • For pH/mV Log on demand: • For AutoLOG mode: Note: If selecting a different printing language, all data will be printed in the selected language. 39 TEMPERATURE CALIBRATION (for technical personnel only) All the instruments are factory calibrated for temperature. Hanna Instruments’ temperature probes are interchangeable and no tem‑ perature calibration is needed when they are replaced. If the temperature measurements are inaccurate, temperature recalibration should be performed. For an accurate recalibration, contact your local Hanna Instruments Office, or follow the instructions below. • Prepare a vessel containing ice and water and another one containing hot water (at approximately 50 ºC or 122 ºF). Place insulation material around the vessels to minimize temperature changes. • Use a calibrated thermometer with a resolution of 0.1 ºC or 0.1 ºF as a reference thermometer. Connect the HI7662-T temperature probe to the appropriate socket. • With the instrument off, press and hold down the CAL & keys, then power on the instrument. The “CAL” tag will appear and the secondary LCD will show 0.0 ºC or 32.0 ºF. The primary LCD will display the mea‑ sured temperature or the ”----” message, if the measured temperature is out of range. • Immerse the temperature probe into the vessel with ice and water as close as possible to the reference thermometer. Allow a few seconds for the probe to stabilize. • Use the ARROW keys to set the reading on the secondary LCD to that of ice and water, measured by the reference thermometer. When the reading is stable and close to the selected calibration point, “CFM” tag will blink. • Press CFM to confirm. The secondary LCD will display CFM 50.0 ºC or 122.0 ºF. • Immerse the temperature probe into the second vessel as close as possible to the reference thermometer. Allow a few seconds for the probe to stabilize. • Use the ARROW keys to set the reading on the secondary LCD to that of the hot water. • When the reading is stable and close to the selected calibration point, “CFM” tag will blink. 40 • Press CFM to confirm. The instrument returns to measurement mode. Note: If the reading is not close to the selected calibration point, “WRONG” tag will blink. Change the temperature probe and restart calibration. CFM mV CALIBRATION (for technical personnel only) All the instruments are factory calibrated for mV. Hanna Instruments’ ORP electrodes are interchangeable and no mV calibra‑ tion is needed when they are replaced. If the measurements are inaccurate, mV recalibration should be performed. For an accurate recalibration, contact your local Hanna Instruments Office, or follow the instructions below. A two or three-point calibration can be performed at 0.0 mV, 600.0 mV and 1800.0 mV. • Attach to the BNC connector (#6 on page 4 and 5) a mV simulator with an accuracy of ±0.1 mV. • With the instrument off, press and hold down the CFM & LOG keys, then power on the instrument. The “CAL” tag will appear and the secondary LCD will show 0.0 mV. • Set 0.0 mV on the simulator. When the reading is stable and close to the selected calibration point, “CFM” tag will blink. • Press CFM to confirm. The secondary LCD will display 600 mV. • Set 600.0 mV on the simulator. When the reading is stable and close to the selected calibration point, “CFM” tag will blink. • Press CFM to confirm. The secondary LCD will display 1800 mV. • Set 1800.0 mV on the simulator. When the reading is stable and close to the selected calibration point, “CFM” tag will blink. • Press CFM to confirm. The instrument returns to measurement mode. Notes: • If the reading is not close to the selected calibration point, “WRONG” tag will blink. Verify calibration condition or contact your vendor if you can not calibrate. • Press CAL in any moment of the calibration process. The instru‑ ment will return to measurement mode. If calibration process is stopped after 600 mV is confirmed, the 600 mV range is calibrated and calibration parameters are memorized. 41 PC INTERFACE Data transmission from the instrument to the PC can be done with the HI92000 Windows® compatible software (optional). HI92000 also offers graphing and on-line help feature. Data can be exported to the most popular spreadsheet programs for further analysis. To connect your instrument to a PC, use the optional Hanna Instruments HI920010 cable connector. Make sure that your instrument is switched off and plug one connector to the instrument’s RS232C socket and the other to the serial port of your PC. To allow our users access to the latest version of Hanna Instruments PC compatible software, we made the products available for download at http:// software.hannainst.com. Select the product code and click Download Now. After download is complete, use the setup.exe file to install the software. Notes: • Other cables than HI920010 may use a different configuration. In this case, communication between instrument and PC may not be possible. • If you are not using Hanna Instruments HI92000 software, please see the following instructions. SENDING COMMANDS FROM PC It is also possible to remotely control the instrument with any terminal program. Use HI920010 cable to connect the instrument to a PC, start the terminal program and set the communication options as follows: 8, N, 1, no flow control. COMMAND TYPES To send a command to the instrument the scheme is: <command prefix> <command> <CR> where: <command prefix> is a selectable ASCII character between 0 and 47 (default 16). <command> is the command code. Note: Either small or capital letters can be used. SIMPLE COMMANDS MOD Is equivalent to pressing MODE CAL Is equivalent to pressing CAL CFM Is equivalent to pressing CFM UPC Is equivalent to pressing the UP arrow key DWC Is equivalent to pressing the DOWN arrow key LOG Is equivalent to pressing LOG RCL Is equivalent to pressing RCL ALG Is equivalent to pressing AutoLOG SET Is equivalent to pressing SET PRT Is equivalent to pressing PRINT (HI122 only) CLR Is equivalent to pressing CLR OFF Is equivalent to pressing OFF AHD Is equivalent to pressing AutoHOLD 42 CHR xx Change the instrument’s range according to the parameter ‘s value (xx): • xx=00 pH range/0.001 resolution • xx=01 pH range/0.01 resolution • xx=03 mV range • xx=04 Relative mV range The meter will answer for these commands with: <STX> <answer> <ETX> where: <STX> is 02 ASCII code character (start of text) <ETX> is 03 ASCII code character (end of text) <answer>: <ACK> is 06 ASCII code character (recognized command) <NAK> is 21 ASCII code character (unrecognized command) <CAN> is 24 ASCII code character (corrupted command) COMMANDS REQUIRING AN ANSWER Causes the instrument to send a complete set of readings according to the current range: • pH, mV and temperature reading on pH range. • mV and temperature reading on mV range. • Rel mV, absolute mV and temperature reading on Rel mV range. Note: The temperature is sent with two decimals and in Celsius degrees, even if the set temperature unit was Fahrenheit degrees. DA? Requests the date. TI? Requests the time. MDR Requests the instrument’s model name and firmware code. GLP Requests the calibration data record. PAR Requests the setup parameters setting. NSL Requests the number of logged samples. ?ML Requests the information about AutoLOG. /MLPxxx Selects the lot. ?DM Downloads the selected AutoLOG. LODPxxx Requests the xxxth record logged data. LODPALL Requests all Log on demand. Notes: • “Err8” is sent if the instrument is not in measurement mode. • “Err7” is sent if a different range is requested during AutoLOGGING mode (for CHR xx command only). • “Err6” is sent if the requested range is not available. • “Err4” is sent if the requested set parameter is not available. • “Err3” is sent if the Log on demand is empty. • Invalid commands will be ignored. RAS 43 pH BUFFER TEMPERATURE DEPENDENCE Temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. During calibration the instrument will automatically calibrate to the pH value corresponding to the measured or set temperature. TEMP pH BUFFERS ºC ºF 1.679 4.010 6.862 7.010 9.177 10.010 12.454 0 32 1.670 4.007 6.982 7.130 9.459 10.316 13.379 5 41 1.670 4.002 6.949 7.098 9.391 10.245 13.178 10 50 1.671 4.000 6.921 7.070 9.328 10.180 12.985 15 59 1.673 4.001 6.897 7.046 9.273 10.118 12.799 20 68 1.675 4.004 6.878 7.027 9.222 10.062 12.621 25 77 1.679 4.010 6.862 7.010 9.177 10.010 12.450 30 86 1.683 4.017 6.851 6.998 9.137 9.962 12.286 35 95 1.688 4.026 6.842 6.989 9.108 9.919 12.128 40 104 1.693 4.037 6.837 6.983 9.069 9.881 11.978 45 113 1.700 4.049 6.834 6.979 9.040 9.847 11.834 50 122 1.707 4.062 6.834 6.978 9.014 9.817 11.697 55 131 1.715 4.076 6.836 6.979 8.990 9.793 11.566 60 140 1.724 4.091 6.839 6.982 8.969 9.773 11.442 65 149 1.734 4.107 6.844 6.987 8.948 9.757 11.323 70 158 1.744 4.123 6.850 6.993 8.929 9.746 11.211 75 167 1.755 4.139 6.857 7.001 8.910 9.740 11.104 80 176 1.767 4.156 6.865 7.010 8.891 9.738 11.003 85 185 1.780 4.172 6.873 7.019 8.871 9.740 10.908 90 194 1.793 4.187 6.880 7.029 8.851 9.748 10.819 95 203 1.807 4.202 6.888 7.040 8.829 9.759 10.734 During calibration the instrument will display the pH buffer value at 25 ºC (77 ºF). 44 ELECTRODE CONDITIONING & MAINTENANCE PREPARATION PROCEDURE Remove the protective cap of the pH electrode. DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal with electrodes. They will disappear when rinsed with water. During transport, tiny bubbles of air may form inside the glass bulb affect‑ ing proper functioning of the electrode. These bubbles can be removed by “shaking down” the electrode as you would do with a glass thermometer. If the bulb and/or junction is dry, soak the electrode in HI70300 or HI80300 Storage Solution for at least one hour. For refillable electrodes: If the filling solution (electrolyte) is more than 2½ cm (1”) below the fill hole, add HI7082 or HI8082 3.5M KCl Electrolyte Solution for double junction or HI7071 or HI8071 3.5M KCl+AgCl Electrolyte Solution for single junction electrodes. For faster response, unscrew the fill hole screw during measurements. For AmpHel® electrodes: If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced. 45 MEASUREMENT Rinse the pH electrode tip with distilled water. Immerse the tip (bottom 4 cm /1½”) in the sample and stir gently for a few seconds. For a faster response and to avoid cross-contamination of the samples, rinse the electrode tip with a few drops of the solution to be tested, before taking measurements. Make sure the sleeve holes of the ORP probe are completly submerged. Tap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve. STORAGE PROCEDURE To minimize clogging and assure a quick response time, the glass bulb and the junction of pH electrode should be kept moist and not allowed to dry out. Replace the solution in the protective cap with a few drops of HI70300 or HI80300 Storage Solution or, in its absence, Filling Solution (HI7071 or HI8071 for single junction and HI7082 or HI8082 for double junction electrodes). Follow the Preparation Procedure on page 45 before taking measurements. Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER. PERIODIC MAINTENANCE Inspect the electrode and the cable. The cable used for connection to the instrument must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present, replace the electrode. Rinse off any salt deposits with water. pH Probe Maintenance For refillable electrodes: Refill the reference chamber with fresh electrolyte (HI7071 or HI8071 for single junction or HI7082 or HI8082 for double junction electrodes). Allow the electrode to stand upright for 1 hour. Follow the Storage Procedure above. pH CLEANING PROCEDURE • General Soak in Hanna Instruments HI7061 or HI8061 General Cleaning Solution for approximately ½ hour. • Protein Soak in Hanna Instruments HI7073 or HI8073 Protein Cleaning Solution for 15 minutes. Soak in Hanna Instruments HI7074 Inorganic Cleaning Solution for 15 minutes. Rinse with Hanna Instruments HI7077 or HI8077 Oil and Fat Cleaning Solution • Inorganic • Oil/grease 46 IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled water, refill the reference chamber with fresh electrolyte (not necessary for gel-filled electrodes) and soak the elec‑ trode in HI70300 or HI80300 Storage Solution for at least 1 hour before taking measurements PRINTER MAINTENANCE (HI122 only) HI122 instrument uses plain paper rolls 44 mm width. To insert a new paper roll, follow the procedure below: • For an easier insertion, cut the paper edge as shown in the next figure: • Put the paper roll on the paper holder attached to the printer. CUT • Insert the paper edge into the printer slot as shown below: INSERT • Simply press Paper from the left keyboard to feed the printer. Allow about 5 cm (2”) to exit from the printer. • Tear the paper out for its edge to be straight, as shown below: TEAR The paper is now properly inserted and the printer is ready to print. 47 TROUBLESHOOTING GUIDE SYMPTOM PROBLEM SOLUTION Slow response/ excessive drift. Dirty pH electrode. Soak the electrode tip in HI7061 solution for 30 minutes and then clean the electrode. Readings fluctuate up and down (noise). Clogged/dirty junction. Low electrolyte level (refillable electrodes only). Clean the electrode. Refill with fresh solution (for refillable electrodes only). The display shows Out of range in the pH blinking dashes scale. during pH measurements. Make sure the pH sam‑ ple is in the specified range. Recalibrate. Check the electrolyte lev‑ el and the general state of the pH electrode. The display shows Reading out of range. blinking reading during measurement. Electrode not connected. Out of range in the mV scale. Dry membrane/ junction. Soak in HI70300 Storage solution for at least one hour. The meter does not work with the temperature probe. Out of order tempera‑ ture probe. Replace the probe. The meter fails to calibrate or gives faulty readings. Broken or out of order pH electrode. Replace the electrode. Explicit warnings are displayed during pH cali‑ bration. Dirty/broken pH elec‑ trode, contaminated reference or buffers. Follow displayed instructions. The electrode condition and response gauges are not displayed after calibration. Electrode without pin or pin is not con‑ nected. Use Hanna Instruments P type pH electrode and check the pin connection. SYMPTOMS 48 TROUBLESHOOTING GUIDE (cont.) SYMPTOM PROBLEM SOLUTION The electrode con‑ dition and response gauges are empty The meter has not been calibrated on the current day. Only one-point calibration has been performed. Perform a two-point calibration procedure. Calibrate using pH7.01 and pH4.01 buffers. The meter displays empty Electrode Response scale. Not enough data available to eval‑ uate the response time. Repeat calibration more carefully. At startup the meter displays all LCD permanently. One of the keys is blocked. Check the keyboard or contact the vendor. Long beep heard when pressing a key. Key without function in current mode. End of range reached with the ARROW keys. Don’t press keys without function in current mode. Digits can not be introduced with the numeric keys. Maximum value reached for corre‑ sponding range. Delete digits. “Prn Eror” message displayed. Printer error. Turn off the meter and then turn it on. If the error persists, contact the vendor. “Err xx” error mes‑ sage displayed. Internal error. Turn off the meter and then turn it on. If the error persist, contact the vendor. SYMPTOM 49 TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes more time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at temperatures below 25 ºC (77 ºF). Since the resistance of the pH electrode is in the range of 50 – 200 Mohm, the current across the membrane is in the pico Ampere range. Large currents can disturb the calibration of the electrode for many hours. For these reasons high humidity environments, short circuits and static discharges are detrimental to a stable pH reading. The pH electrode’s life also depends on the temperature. If constantly used at high temperatures, the electrode life is drastically reduced. Typical Electrode Life Ambient Temperature 90 ºC (194 ºF) 120 ºC (248 ºF) 1 – 3 years Less than 4 months Less than 1 month Alkaline Error High concentrations of sodium ions interfere with readings in alkaline solutions. The pH at which the interference starts to be significant depends upon the composition of the glass. This interference is called alkaline error and causes the pH to be underestimated. Hanna Instruments’ glass formulations have the indicated characteristics. Sodium Ion Correction for the Glass at 20-25 ºC (68-77 ºF) Concentration pH Error 0.1 Mol L-1 Na+ 13.00 0.10 13.50 0.14 14.00 0.20 12.50 0.10 13.00 0.18 1.0 Mol L-1 Na+ 13.50 0.29 14.00 0.40 50 ACCESSORIES pH BUFFER SOLUTIONS HI70004P pH4.01 Buffer Sachets, 20 mL, 25 pcs. HI70007P pH7.01 Buffer Sachets, 20 mL, 25 pcs. HI70010P pH10.01 Buffer Sachets, 20 mL, 25 pcs. HI7001L pH1.68 Buffer Solution, 500 mL bottle HI7004L pH4.01 Buffer Solution, 500 mL bottle HI7006L pH6.86 Buffer Solution, 500 mL bottle HI7007L pH7.01 Buffer Solution, 500 mL bottle HI7009L pH9.18 Buffer Solution, 500 mL bottle HI7010L pH10.01 Buffer Solution, 500 mL bottle HI7001L pH1.68 Buffer Solution in FDA approved bottle, 500 mL HI8004L pH4.01 Buffer Solution in FDA approved bottle, 500 mL HI8006L pH6.86 Buffer Solution in FDA approved bottle, 500 mL HI8007L pH7.01 Buffer Solution in FDA approved bottle, 500 mL HI8009L pH9.18 Buffer Solution in FDA approved bottle, 500 mL HI8010L pH10.01 Buffer Solution in FDA approved bottle, 500 mL ELECTRODE STORAGE SOLUTIONS HI70300L Storage Solution, 460 mL bottle HI80300L Storage Solution in FDA approved bottle, 460 mL ELECTRODE CLEANING SOLUTIONS HI70000P Electrode Rinse Sachets, 20 mL, 25 pcs. HI7073L Protein Cleaning Solution, 460 mL bottle HI7074L Inorganic Cleaning Solution, 460 mL bottle HI7077L Oil & Fat Cleaning Solution, 460 mL bottle HI8073L Protein Cleaning Solution in FDA approved bottle, 460 mL HI8077L Oil & Fat Cleaning Solution in FDA approved bottle, 460 mL 51 ELECTRODE REFILL ELECTROLYTE SOLUTIONS HI7071 3.5M KCl + AgCl Electrolyte, 4x30 mL, for single junction electrodes HI7072 1M KNO3 Electrolyte, 4x30 mL HI7082 3.5M KCl Electrolyte, 4x30 mL, for double junction electrodes HI8071 3.5M KCl + AgCl Electrolyte in FDA approved bottle, 4x30 mL, for single junction electrodes HI8072 1M KNO3 Electrolyte in FDA approved bottle, 4x30 mL HI8082 3.5M KCl Electrolyte in FDA approved bottle, 4x30 mL, for double junction electrodes ORP PRETREATMENT SOLUTIONS HI7091L Reducing Pretreatment Solution, 460 mL bottle HI7092L Oxidizing Pretreatment Solution, 460 mL bottle pH ELECTRODES All electrodes with code ending with P are supplied with BNC & pin connectors and 1 m (3.3’) cable, as shown below: HI1043P Glass-body, double junction, refillable, combination pH electrode. Use: strong acid/alkali. HI1043 52 HI1053P Glass-body, triple ceramic, conic shape, refillable, combination pH electrode. Use: emulsions. HI1053 HI1131P Glass-body, double junction, refillable, combination pH electrode. Use: general purpose. HI1131 HI1083P Glass-body, micro, Viscolene, non-refillable, combination pH electrode. Use: biotechnology, micro titration. HI1083 HI1330P Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory, vials. HI1330 HI1332P Plastic-body (Ultem®), double junction, refillable, combination pH electrode. Use: general purpose. HI1332 53 ORP ELECTRODES HI3131P Glass-body, refillable, combination platinum ORP electrode. Use: titration. HI3131 Consult the Hanna General Catalog for more electrodes with BNC and pin connectors. OTHER ACCESSORIES HI710005 Voltage adapter from 115 VAC to 12 VDC (USA plug) HI710006 Voltage adapter from 230 VAC to 12 VDC (European plug) HI710012 Voltage adapter from 240 VAC to 12 VDC (UK plug) HI710013 Voltage adapter from 230 VAC to 12 VDC (South African plug) HI710014 Voltage adapter from 230 VAC to 12 VDC (Australian plug) HI8427 pH and ORP electrode simulator with 1 m (3.3’) coaxial cable ending in female BNC connectors HI931001 pH and ORP electrode simulator with LCD and 1m (3.3’) coaxial cable ending in female BNC connectors HI7662-T Temperature probe with 1 m (3.3’) cable HI710032 Pack of 10 plain paper spare rolls (HI122) HI710033 Replacement ink cartridge (HI122) HI740157 Plastic electrode refilling pipette (20 pcs.) HI92000 Windows® compatible software HI920010 9 to 9-pin RS232 cable 54 RECOMMENDATIONS FOR USERS Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the meters' performance. For yours and the meter's safety do not use or store the meter in hazardous environments. CERTIFICATION All Hanna Instruments conform to the CE European Directives. Disposal of Electrical & Electronic Equipment. The product should not be treated as household waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic equipment which will conserve natural resources. Disposal of waste batteries. This product contains batteries, do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling. Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, the place of purchase or go to www.hannainst.com. WARRANTY HI120 and HI122 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are guaranteed for six months. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packed for complete protection. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner. Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice. 55 World Headquarters Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA www.hannainst.com Printed in ROMANIA MAN122 06/19 56 ">

Public link updated
The public link to your chat has been updated.
Advertisement