Hanna Instruments | HI9124,HI9125 | Owner Manual | Hanna Instruments HI9124,HI9125 Owner Manual

Hanna Instruments HI9124,HI9125 Owner Manual
Instruction Manual
HI9124, HI9125
Portable Waterproof
pH Meters
www.hannainst.com
1
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using these
instruments.
This manual will provide you with the necessary information for correct
use of these instruments, as well as a precise idea of their versatility.
If you need additional technical information, do not hesitate to e‑mail
us at tech@hannainst.com.
The instruments are in compliance with
directives.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
2
TABLE OF CONTENTS
PRELIMINARY EXAMINATION ..........................................................4
GENERAL DESCRIPTION...................................................................4
LCD MESSAGE GUIDE......................................................................5
FUNCTIONAL DESCRIPTION..............................................................6
SPECIFICATIONS.............................................................................7
OPERATIONAL GUIDE......................................................................8
pH CALIBRATION..........................................................................11
pH BUFFER TEMPERATURE DEPENDENCE.......................................14
SETUP MENU...............................................................................15
mV CALIBRATION (HI9125 only)...................................................16
TEMPERATURE CALIBRATION.........................................................16
BATTERIES REPLACEMENT.............................................................16
TEMPERATURE CORRELATION FOR pH SENSITIVE GLASS..................18
ELECTRODE CONDITIONING & MAINTENANCE..................................19
TROUBLESHOOTING GUIDE...........................................................21
ACCESSORIES...............................................................................22
3
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it to
make sure that no damage has occurred during shipping. If there is any
damage, please contact your local Hanna Instruments Office.
Each meter is supplied with:
• HI1230B non refillable combination double-junction pH electrode
with gelled electrolyte.
• HI7662 stainless steel temperature probe with 1 m (3.3’’) cable
• pH4.01 & pH7.01 buffer solutions, 20 mL sachet
• 100 mL plastic beaker
• 1.5V AAA batteries (3 pcs.)
• Instruction manual
• Rugged carrying case
• Quality Certificate
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original
packing with the supplied accessories.
GENERAL DESCRIPTION
The HI9124 and HI9125 are state-of-the-art rugged waterproof, hand
held pH meters designed to provide laboratory results and accuracy
under harsh industrial conditions.
A large multi-level LCD, with clear indications related to the electrode
and instrument status, pH and temperature displayed simultaneously,
and user friendly symbols and text instructions during calibration.
The pH calibration procedure is automatic with 5 auto recognized
buffers (4.01, 6.86, 7.01, 9.18 and 10.01) and automatic temperature
compensation.
The HI9125 can be used with ORP (Oxidation Reduction Potential)
electrodes. mV measurements automatically change from 0.1 to 1 mV
resolution when the reading reaches 700 mV.
4
LCD MESSAGE GUIDE
TAGS & SYMBOLS
• Mode tags lights up for indicating the corresponding active mode,
and blinks for warning the user.
MEASURE on: measurement mode.
SETUP on: SETUP menu mode has been entered.
CALIBRATION on: calibration mode has been entered.
MEMORIZE on: measurement stored in the internal memory and
frozen on the display
RECALL MEMORIZED on: stored value recalled.
• Battery symbol blinking: low battery condition. Batteries should
be replaced.
• Calibration messages.
WAIT NOT READY blinking: buffer has been recognized, but
reading is not stable.
READY on: buffer has been recognized and reading is stable.
WRONG and WRONG blinking alternatively: wrong buffer,
value not recognized.
• Main active key messages light up for indicating the corresponding
active key.
CFM blinking: ask for confirmation of calibration or set value.
MR on: MR key available.
MEM on: MEM key available.
CAL on: CAL key available.
RANGE on: RANGE key available. (HI9125 only)
SETUP on: SETUP key available.
5
FUNCTIONAL DESCRIPTION
1)
2)
3)
4)
5)
6)
7)
8)
9)
10)
11)
Liquid Crystal Display (LCD).
RANGE key, to select pH or mV (HI9125 only).
ON/OFF key, to turn the meter ON and OFF.
CAL key, to enter or exit calibration mode.
SETUP/CFM key, to enter SETUP mode or to confirm calibration.
Battery compartment cap.
Temperature probe socket.
BNC electrode connector.
MR key, to recall stored value from memory.
MEM key, to store reading in memory.
 and  keys, for manual temperature setting, or selecting
pH buffer value.
12) Secondary display.
13) Primary display.
6
SPECIFICATIONS
-2.00 to 16.00 pH
±699.9 mV / ±1999 mV
RANGE
(HI9125 only)
-20.0 to 120.0 °C (-4.0 to 248.0 °F)
0.01 pH
RESOLUTION
0.1 mV / 1 mV (HI9125 only)
0.1 °C (0.1 °F)
±0.01 pH
ACCURACY
±0.2 mV / ±1 mV (HI9125 only)
@ 20 ºC / 68 ºF
±0.4 °C (±0.8 °F)
(excluding probe error)
±0.02 pH
Typical EMC Deviation ±0.2 mV / ±1 mV
±0.4 °C (±0.8 °F)
1 or 2-point,
pH Calibration
with 5 memorized buffers
(4.01, 6.86, 7.01, 9.18, 10.01)
Offset Calibration
±1 pH
Slope Calibration
From 80 to 108%
Automatic, from -20.0 to 120.0 °C
Temperature
(-4.0 to 248.0 °F) or manual,
Compensation
without temperature probe
pH Electrode
HI1230B (included)
Temperature Probe
HI7662 (included)
Input Impedance
1012 ohms
3 x 1.5V AAA batteries
Battery Type & Life
approx. 200 hours of continuous use
Auto-off
User selectable: 20 minutes or disabled
Dimensions
185x72x36 mm (7.3x2.8x1.4”)
Weight
300 g (10.6 oz.)
0 – 50 °C (32 – 122 °F)
Environment
max RH 100%
Warranty
2 years
7
OPERATIONAL GUIDE
INITIAL PREPARATION
The instrument is supplied with batteries. In order to place the batteries
inside the instrument follow the instructions from page 16.
To prepare the instrument for use, connect the pH electrode and the
temperature probe to the BNC and temperature sockets on the top of
the instrument. The temperature probe can be used independently
to take temperature measurements, or it can be used in conjunction
with the pH electrode to utilize Automatic Temperature Compensation
(ATC) mode. If the probe is disconnected, temperature can also be set
manually with the ARROW keys.
Turn the instrument ON by pressing ON/OFF.
At start-up the display will show all the LCD segments and then the
battery percentage while the instrument performs a self check (or as
long as the button is held).
The meter automatically enters measurement mode.
After measurement, switch the meter off. Clean the electrode and store
it with a few drops of HI70300 storage solution in the protective cap.
To save batteries, the auto-off feature turns the meter off after 20
minutes with no button pressed. To disable this feature, see “Setup
Menu” section on page 15.
pH MEASUREMENTS
Calibrate the meter and pH electrode before taking
measurements. See page 10 for details. To take a
pH measurement remove the electrode’s protec‑
tive cap, rinse off pH sensing tip. Submerse the
electrode and the temperature probe 3 cm (1¼”)
into the sample and stir gently.
If necessary, press the RANGE key until the
display shows pH mode (HI9125 only).
Allow time for the reading to stabilize.
8
The LCD will show the pH measurement and the temperature of the
sample.
If several measurements are taken successively in different samples,
rinse the electrode thoroughly with deionized or tap water and a small
amount of the sample to be measured.
The pH value of the sample is directly affected by temperature. In
order for the meter to measure the pH accurately, the value must be
compensated for temperature. If the sample temperature is different
from the temperature at which the pH electrode was kept, allow a few
minutes for thermal equilibrium.
To use the meter’s Automatic Temperature Compensation feature, sub‑
merse the temperature probe into the sample as close to the electrode
as possible and wait for a few minutes.
If manual temperature compensation is desired the temperature
probe must be disconnected from the instrument.
The display will show the default temperature of “25 °C”, or the last
temperature set with the “°C” (or “°F”) indicator blinking.
The temperature can now be adjusted with the ARROW keys. Set the
value at the sample temperature.
Notes: When the reading is out of
range, the display will flash the
closest full-scale value.
If using pH electrode while in mV
mode, the meter will measure the
mV generated by the pH electrode.
9
ORP MEASUREMENTS (HI9125 only)
To perform ORP measurements, connect an optional ORP electrode (see
“Accessories” section) to the meter and turn it ON.
If necessary, enter the mV mode by pressing RANGE until the display
changes to mV.
Submerse the ORP electrode 3 cm (1¼”) into the sample to be tested
and wait a few minutes for the reading to stabilize.
Measurements within the ±699.9 mV range are displayed with
0.1 mV resolution, while outside this range the resolution automatically
switches to 1 mV.
For accurate ORP measurements, the surface of the electrode must be
clean and smooth. Pretreatment solutions are available to condition
the electrode and improve its response time (see “Accessories” section).
MEM & MR FUNCTIONS
The instrument allows the user to store the current measurement (pH
and temperature, or mV and temperature) into internal memory by
pressing the MEM key. The “MEMORIZE” tag lights up on display.
Stored values can be recalled by pressing MR: the display will show
the values and the “RECALL MEMORIZED” tag as long as the button
is pressed.
10
pH CALIBRATION
It is recommended to calibrate the instrument frequently, especially if
high accuracy is required.
The pH range should be recalibrated:
• Whenever the pH electrode or temperature probe is replaced.
• At least once a week.
• After testing aggressive chemicals.
• When extreme accuracy is required.
PREPARATION
Pour a small quantity of buffer solution into clean beakers. For accurate
calibration use two beakers for each buffer solution, the first one for
rinsing the electrode and the second one for calibration.
PROCEDURE
In order to perform pH calibration:
• Make sure that the meter is in the pH mode (HI9125 only).
• Remove the protective cap and rinse the electrode with some of the
buffer solution to be used for the first calibration point.
There is a choice of 5 memorized buffers: 4.01, 6.86, 7.01, 9.18
and 10.01 pH.
TWO-POINT CALIBRATION
• Press the CAL key. The “CAL” and “ ” indicators will be dis‑
played. The secondary LCD will display buffer “7.01”. If a different
calibration buffer is desired (e.g. “6.86”), use the ARROW keys to
change the displayed value.
or
• Submerse the electrode approx. 3 cm (1¼”) into the solution,
place the temperature probe as close as possible to the electrode
and stir gently.
11
• The LCD will flash the “WAIT NOT READY” message.
• Once the reading is stable, if it is not close to the selected buffer,
“WRONG “ and “WRONG “ will blink alternatively; if it is
close to the selected buffer the display will change to “READY”
and blinking “Press: CFM to confirm calibration point”.
• Press the CFM key to confirm the calibration: the meter stores the
offset calibration point. The calibrated reading is then displayed
on the primary LCD while the secondary LCD will show the second
buffer to be used for calibration (pH4.01).
• After the first calibration point is confirmed, submerse the electrode
into the second buffer (pH4.01, 10.01 or 9.18) and stir gently.
Choose pH4.01 for acidic samples, and pH10.01 or 9.18 for alkaline
solutions.
• Submerse the electrode approx. 3 cm (1¼”) into the solution,
place the temperature probe as close as possible to the electrode
and stir gently.
• Select the second buffer value on the secondary display by pressing
the ARROW keys.
12
• If the reading is not close to the selected buffer, “WRONG ”
and “WRONG ” will blink alternatively;
• If the reading is close to the selected buffer and the reading is
stable, the “READY” symbol is displayed and the “CFM” symbol
starts blinking on the LCD, asking for confirmation.
• Press the CFM key: the value is stored into memory and the meter
returns to normal mode.
Notes: • The meter automatically skips the buffer used for the first
calibration point to avoid erroneous procedure. A difference
of at least 1.5 pH unit is required between the two buffers
used for the offset and slope calibration: once calibrated at
either pH7.01 or 6.86, the instrument automatically ignores
the other value for the second point (same for pH10.01 and
9.18).
• During calibration, the secondary LCD shows the selected
buffer value. For the HI9125 model, it is possible to dis‑
play the buffer temperature during calibration by pressing
RANGE.
• To clear a previous calibration and return to the default
values, in calibration mode press CFM, then CAL before
the first buffer is accepted. The LCD will show “CLr CAL”
for one second, and then will return to normal mode.
ONE-POINT CALIBRATION
For optimum accuracy it is always recommended to perform a two-point
calibration, but for a faster operation a single-point calibration can be
used. pH7.01 or pH6.86 (NIST) are normally used for this purpose,
even though the meters can be calibrated with any of the 5 memorized
calibration values.
After calibrating the first point (see above), press the CAL key to end
the calibration procedure.
13
pH BUFFER TEMPERATURE
DEPENDENCE
Temperature has an effect on pH. The calibration buffer solutions are
affected by temperature changes to a lesser degree than normal solutions.
During calibration the instrument will automatically calibrate to the pH
value corresponding to the measured or set temperature.
TEMP
pH VALUES
°C
°F
4.01
6.86
7.01
9.18
10.01
0
32
4.01
6.98
7.13
9.46
10.32
5
41
4.00
6.95
7.10
9.39
10.24
10
15
50
59
4.00
4.00
6.92
6.90
7.07
7.05
9.33
9.27
10.18
10.12
20
68
4.00
6.88
7.03
9.22
10.06
25
77
4.01
6.86
7.01
9.18
10.01
30
86
4.02
6.85
7.00
9.14
9.96
35
95
4.03
6.84
6.99
9.11
9.92
40
104
4.04
6.84
6.98
9.07
9.88
45
113
4.05
6.83
6.98
9.04
9.85
50
122
4.06
6.83
6.98
9.01
9.82
55
131
4.08
6.84
6.98
8.99
9.79
60
140
4.09
6.84
6.98
8.97
9.77
65
149
4.11
6.84
6.99
8.95
9.76
70
158
4.12
6.85
6.99
8.93
9.75
75
167
4.14
6.86
7.00
8.91
9.74
80
176
4.16
6.87
7.01
8.89
9.74
85
185
4.17
6.87
7.02
8.87
9.74
90
194
4.19
6.88
7.03
8.85
9.75
95
203
4.20
6.89
7.04
8.83
9.76
During calibration the instrument will display the pH buffer value at
25 °C.
14
SETUP MENU
Instrument allows the user to configure several parameters through
the Setup Menu.
To enter the SETUP mode, while in measurement mode, press and hold
the SETUP key for about 5 seconds. When the key is released the first
parameter will be displayed.
Once the menu is entered, each parameter can be changed by using
the ARROW keys; then pressing the CFM key will confirm the value
and scroll to the next parameter.
1. Audible signal: On (default) or Off
2. Auto-off feature: 20 minutes (default) or disabled
3. Temperature measure unit: ºC (default) or ºF
After setting the last parameter, pressing the CFM key will confirm the
value and return to measurement mode.
15
mV CALIBRATION (HI9125 only)
HI9125 has been precalibrated for mV range at the factory.
For optimum accuracy, it is recommended to recalibrate the meter for
mV readings at least once a year.
Contact your local Hanna Instruments Office.
TEMPERATURE CALIBRATION
HI9124 & HI9125 have been precalibrated for temperature at the
factory.
For optimum accuracy, it is recommended to recalibrate the meter for
temperature at least once a year.
Contact your local Hanna Instruments Office.
BATTERIES REPLACEMENT
The instrument is supplied with batteries. First time you start working
with the instrument, insert the supplied batteries in the battery
compartment observing the correct polarity (see page 17).
At start-up the battery percentage is displayed.
If the batteries become weak, the display will flash the battery symbol
to advise the user that approximately 1 hour of working time remains.
It is recommended to change the batteries as soon as the battery
symbol appears blinking.
16
The instrument is also provided with the BEPS (Battery Error Prevention
System) feature which automatically turns the instrument off when
the battery level is too low to ensure reliable readings. At start-up
the display will show “0 batt” for few seconds, then the instrument
automatically turns off.
To replace the batteries, follow the next steps:
• Turn the instrument OFF.
• Open the battery compartment cap (located on the top of the
instrument).
• Remove old batteries.
• Insert three new 1.5V AAA batteries in the battery compartment,
observing the polarity on the rear of the instrument.
• Reattach the battery compartment cap.
17
TEMPERATURE CORRELATION
FOR pH SENSITIVE GLASS
The resistance of glass electrodes partially depends on temperature. The
lower the temperature, the higher the resistance. It takes more time for
the reading to stabilize if the resistance is higher. In addition, the re‑
sponse time will suffer to a greater degree at temperatures below 25 °C.
Since the resistance of the pH electrode is in the range of
50 – 200 Mohms, the current across the membrane is in the pico
Ampere range. Large currents can disturb the calibration of the electrode
for many hours. For these reasons high humidity environments, short
circuits and static discharges can be detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.
Typical Electrode Life
Ambient Temperature
1 – 3 years
90 °C
Less than 4 months
120 °C
Less than 1 month
Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline
solutions. The pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is called
alkaline error and causes the pH to be underestimated.
Hanna Instruments’ glass formulations have the indicated characteristics.
Sodium Ion Correction for Glass at 20-25 °C
Concentration
pH
13.00
0.1 Mol L-1 Na+
13.50
14.00
12.50
13.00
1.0 Mol L-1 Na+
13.50
14.00
18
Error
0.10
0.14
0.20
0.10
0.18
0.29
0.40
ELECTRODE CONDITIONING & MAINTENANCE
Not present in gel electrodes.
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is
normal with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may have formed inside the glass
bulb. The electrode cannot function properly under these conditions.
These bubbles can be removed by “shaking down” the electrode as
you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in HI70300
Storage Solution for at least one hour or longer.
19
For refillable electrodes, if the refill solution (electrolyte) is more than
2½ cm (1”) below the fill hole, add the appropriate Electrolyte Solution.
MEASUREMENT
Rinse the electrode tip with distilled water, submerse it 3 cm (1¼”) in
the sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with the solution to be tested, before taking any
measurements.
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time, the glass bulb
and the junction should always be kept moist.
When not in use, store it with a few drops of HI70300 storage solution
in the protective cap.
NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED
WATER.
PERIODIC MAINTENANCE
Inspect electrode and cable. The cable must be intact. No cracks should
be seen on the electrode stem or bulb. If any scratches or cracks are
present, replace the electrode. Rinse off any salt deposits with water.
Connectors must be perfectly clean and dry.
For refillable electrodes:
Refill the electrode with fresh electrolyte (see the electrode’s specifications
to select the correct refilling solution). Allow the electrode to stand upright
for 1 hour. Follow the Storage Procedure above.
For nonrefillable electrodes:
Do not be concerned if crystals form in the gel. This will not effect pH
electrode response.
CLEANING PROCEDURE
• General
Soak in Hanna Instruments HI7061 General
Cleaning Solution for approximately ½ hour.
• Protein
Soak in Hanna Instruments HI7073 Protein
Cleaning Solution for 15 min.
• Inorganic
Soak in Hanna Instruments HI7074 Inorganic
CleaningSolution for 15 min.
• Oil/grease
Rinse with Hanna Instruments HI7077 Oil &
Fat Cleaning Solution for 1 min.
IMPORTANT: After performing any of the cleaning procedures, rinse
the electrode thoroughly with distilled water and soak it in HI70300
Storage Solution for at least 1 hour before taking measurements.
20
TROUBLESHOOTING GUIDE
SYMPTOMS
Slow reponse/excessive
drift.
PROBLEM
Dirty pH electrode.
SOLUTION
The electrode needs to be
cleaned. Follow the clean‑
ing procedure on page 20.
Reading fluctuates up Clogged/dirty junc‑
Clean the electrode.
and down (noise).
tion. Low electrolyte Refill with fresh electro‑
level (refillable
lyte (refillable electrodes
electrodes only).
only). Check cable and
connector.
Display shows blinking Reading out of range. Verify the electrode is
full scale value.
connected. Check that
the sample is within
measurable range.
Verify that the electrode
is in contact with the
solution.
Display shows blinking Broken temperature Replace temperature
“°C” or “°F”.
probe.
probe.
Meter does not work
Broken temperature Replace temperature
with temperature probe. probe.
probe.
Wrong temperature
probe.
Display shows blinking Low battery level.
Replace the batteries.
battery symbol.
Meter fails to calibrate Broken pH electrode. Replace electrode.
or gives faulty readings.
“WRONG CALIBRAWrong or contaminat‑ Check that buffer
TION” is displayed
ed buffer.
solution is correct and
during pH calibration
fresh.
procedure.
Meter shuts off.
Dead batteries or
Replace the batteries;
Auto-off feature is
Press ON/OFF.
enabled: in this case,
meter shuts off after
20 min of non-use.
“Er0“, “Er1“, “Er2”
EEPROM error.
Contact your local Hanna
message at start up.
Instruments.
“CLr” message at
Loaded default pH
Perform pH calibration.
start up.
calibration values.
21
ACCESSORIES
pH CALIBRATION SOLUTIONS
HI70004P
pH4.01 Buffer Solution, 20 mL sachet, 25 pcs
HI70007P
pH7.01 Buffer Solution, 20 mL sachet, 25 pcs
HI70010P
pH10.01 Buffer Solution, 20 mL sachet, 25 pcs
HI7004L
pH4.01 Buffer Solution, 500 mL bottle
HI7004M
pH4.01 Buffer Solution, 230 mL bottle
HI7006L
pH6.86 Buffer Solution, 500 mL bottle
HI7006M
pH6.86 Buffer Solution, 230 mL bottle
HI7007L
pH7.01 Buffer Solution, 500 mL bottle
HI7007M
pH7.01 Buffer Solution, 230 mL bottle
HI7009L
pH9.18 Buffer Solution, 500 mL bottle
HI7009M
pH9.18 Buffer Solution, 230 mL bottle
HI7010L
pH10.01 Buffer Solution, 500 mL bottle
HI7010M
pH10.01 Buffer Solution, 230 mL bottle
ELECTRODE STORAGE SOLUTION
HI70300L
Storage Solution, 500 mL bottle
HI70300M
Storage Solution, 230 mL bottle
ELECTRODE CLEANING SOLUTIONS
HI70000P
Electrode Rinse Solution, 20 mL sachet, 25 pcs
HI7061L
General Cleaning Solution, 500 mL bottle
HI7061M
General Cleaning Solution, 230 mL bottle
HI7073L
Protein Cleaning Solution, 500 mL bottle
HI7073M
Protein Cleaning Solution, 230 mL bottle
HI7074L
Inorganic Cleaning Solution, 500 mL bottle
HI7074M
Inorganic Cleaning Solution, 230 mL bottle
HI7077L
Oil & Fat Cleaning Solution, 500 mL bottle
HI7077M
Oil & Fat Cleaning Solution, 230 mL bottle
REFILLING ELECTROLYTE SOLUTIONS (50 mL, 4 pcs)
HI7071
3.5M KCl+AgCl Electrolyte for single junction electrodes
HI7072
1M KNO3 Electrolyte
HI7082
3.5M KCl Electrolyte for double junction electrodes
HI8093
1M KCl+AgCl Electrolyte
ORP PRETREATMENT SOLUTIONS
HI7091L
Reducing Pretreatment Solution, 500 mL bottle
HI7091M
Reducing Pretreatment Solution, 230 mL bottle
HI7092L
Oxidizing Pretreatment Solution, 500 mL bottle
HI7092M
Oxidizing Pretreatment Solution, 230 mL bottle
ORP SOLUTIONS
HI7021L
Test Solution 240 mV, 500 mL bottle
HI7021M
Test Solution 240 mV, 230 mL bottle
HI7022L
Test Solution 470 mV, 500 mL bottle
HI7022M
Test Solution 470 mV, 230 mL bottle
22
pH ELECTRODES
All electrodes part numbers ending in B are supplied with a BNC connector
and 1 m (3.3’) cable, as shown below:
HI1043B
Glass-body, double junction, refillable, combination pH electrode.
Use: strong acid/alkali.
HI1053B
Glass-body, triple ceramic, conic shape, refillable, combination pH
electrode.
Use: emulsions.
HI1083B
Glass-body, micro, Viscolene, non-refillable, combination pH electrode.
Use: biotechnology, micro titration.
HI1131B
Glass-body, single junction, refillable, combination pH electrode.
Use: general purpose.
23
HI1330B
Glass-body, semimicro, single junction, refillable, combination pH
electrode.
Use: laboratory, vials.
HI1331B
Glass-body, semimicro, single junction, refillable, combination pH
electrode.
HI1230B
Plastic-body (PES), double junction, gel-filled, combination pH
electrode.
Use: general, field.
HI2031B
Glass-body, semimicro, conic, refillable, combination pH electrode.
Use: semisolid products.
HI1332B
Plastic-body (PES), double junction, refillable, combination pH
electrode.
Use: general purpose.
24
FC100B
Plastic-body (PVDF), double junction, refillable, combination pH
electrode.
Use: general purpose for food industry.
FC200B
Plastic-body (PVDF), open junction, conic, Viscolene, non-refillable,
combination pH electrode.
Use: meat & cheese.
FC210B
Glass-body, double junction, conic, Viscolene, non-refillable, combination
pH electrode.
Use: milk, yogurt.
FC220B
Glass-body, triple-ceramic, single junction, refillable, combination
pH electrode.
Use: food processing.
FC911B
Plastic-body (PVDF), double junction, refillable with built-in amplifier,
combination pH electrode. Use: very high humidity.
25
HI1413B
Glass-body, single junction, flat tip, Viscolene, non-refillable, combi‑
nation pH electrode. Use: surface measurement.
ORP ELECTRODES
HI3131B
Glass-body, refillable, combination platinum ORP electrode.
Use: titration.
HI3230B
Plastic-body (PES), gel-filled, combination platinum ORP electrode.
Use: general purpose.
HI4430B
Plastic-body (PES), gel-filled, combination gold ORP electrode.
Use: general purpose.
Consult the Hanna Instruments General Catalog for a complete and
wide selection of electrodes.
OTHER ACCESSORIES
HI721317 Rugged carrying case
HI740157 Plastic electrode refilling pipet (20 pcs)
HI76405 Electrode holder
HI7662
Temperature probe with 1 m (3.3’) screened cable
pH and ORP electrode simulator with 1 m (3.3’) coaxial
HI8427
cable ending in female BNC connectors
pH and ORP electrode simulator with LCD and 1 m
HI931001
(3.3’) coaxial cable ending in female BNC connectors
26
RECOMMENDATIONS FOR USERS
Before using this product, make sure it is entirely suitable for your
specific application and for the environment in which it is used. Any
variation introduced by the user to the supplied equipment may degrade
the meters' performance. For yours and the meter's safety do not use
or store the meter in hazardous environments.
WARRANTY
HI9124 & HI9125 are guaranteed for two years against defects in
workmanship and materials when used for their intended purpose
and maintained according to instructions. Electrodes and probes
are guaranteed for six months. This warranty is limited to repair or
replacement free of charge.
Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered.
If service is required, contact your local Hanna Instruments Office. If
under warranty, report the model number, date of purchase, serial
number and the nature of the problem. If the repair is not covered
by the warranty, you will be notified of the charges incurred. If these
instruments are to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization number from the Technical Service de‑
partment and then send it with shipping costs prepaid. When shipping
any instrument, make sure it is properly packed for complete protection.
Hanna Instruments reserves the right to modify the design,
construction and appearance of its products without advance notice.
27
World Headquarters
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com
Local Office
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
Phone: 800.426.6287
Fax: 401.765.7575
e-mail: tech@hannainst.com
MAN9125 10/18
Printed in ROMANIA
28
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising