Hanna Instruments | HI955501,HI955502,HI93501,HI935001,HI935007,HI935004,HI935008,HI9350011,HI9350041 | Owner Manual | Hanna Instruments HI955501,HI955502,HI93501,HI935001,HI935007,HI935004,HI935008,HI9350011,HI9350041 Owner Manual

Hanna Instruments HI955501,HI955502,HI93501,HI935001,HI935007,HI935004,HI935008,HI9350011,HI9350041 Owner Manual
Portable Thermometers
INSTRUCTION MANUAL
HI93501, HI935001,
HI935004, HI935007,
HI935008, HI9350011,
HI9350041
Thank you for choosing a Hanna Instruments product.
Dear
Customer, Please read this instruction manual carefully before
using this meter.
This manual will provide you with the necessary
information for correct use of this meter, as well as
a precise idea of its versatility.
If you need additional technical information, do not
hesitate to e‑mail us at tech@hannainst.com or view
our worldwide contact list at www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is
prohibited without the written consent of the copyright owner, Hanna
Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
TABLE OF CONTENTS
Table of Contents.........................................................3
Preliminary Examination..............................................4
Intended Use..............................................................4
General Description......................................................5
HI93501 Specifications...............................................6
HI935001 Specifications.............................................8
HI935004 Specifications.............................................10
HI935007 Specifications.............................................12
HI935008 Specifications.............................................14
HI9350011 Specifications...........................................16
HI9350041 Specifications........................................ 18
Operational Guide.......................................................20
Meter Setup................................................................23
Battery Replacement....................................................24
Temperature Probes.....................................................25
Accessories..................................................................25
Certification.................................................................26
3
PRELIMINARY EXAMINATION
Remove the meter from the packing material and exam‑
ine it to make sure that no damage has occurred during
shipping. If there is any damage, contact your local Hanna
Instruments Office.
Each meter is supplied complete with:
• Temperature probe
• Alkaline batteries: 1.5V AAA (3 pcs.)
• Quick reference guide
• Instruction manual
Note: Save all packing material until you are sure that
the meter functions correctly. All defective items must
be returned in the original packing with the supplied
accessories.
INTENDED USE
The thermometers are waterproof and are rated
IP65. The meters are specifically designed for
measuring the temperature in all areas of the
food sector.
Both air temperature and product internal temperature can be
measured with these thermometers.
The probe shaft is designed for continuous contact with foodstuffs
in accordance with Regulation (EC) 1935/2004.
Refer to the probe section of this manual for more information
on the available probes.
4
GENERAL DESCRIPTION
The process variable temperature plays a central
function in food commerce. Many aspects of safe food handling
require the accurate measurement of temperature. Food
service, food preparation, packaging, storage and transport of food
require temperature to be monitored or controlled. Spot checking
temperatures with Hanna food thermometers ensures daily work
routines are carried out at the correct temperature.
Hanna Instruments manufactures food thermometer
models to meet the tough environment encountered in
the food sector. Standard features of all models include a
ruggedized professionally designed waterproof case and
stainless steel probe for measurements in semi solid foods,
liquids, pastes or air. They provide good accuracy and
include features such as CAL Check, low battery detection, auto-off
capability, and long battery life.
The model HI93501 is a thermistor style thermometer. It
measures temperatures from -50.0 to 150.0 °C (-58.0 to
302.0 °F). It comes with a stainless steel replaceable style
penetration probe (FC762PW).
The models HI935001 and HI935007 thermometers utilize a
K-type thermocouple style probe and provide the large range
of temperature measurement; from -50.0 to 300 °C (-58.0 to
572 °F). The model HI935001 uses a replaceable style probe
(FC766PW) and the model HI935007 has a fixed thermocouple
probe and is not replaceable.
The models HI935004 and HI935008 thermometers utilize a
T-type thermocouple style probe and provide the large range
of temperature measurement; from -50.0 to 300 °C (-58.0 to
572 °F). The model HI935004 uses a replaceable style probe
(FC767PW) and the model HI935008 has a fixed thermocouple
probe and is not replaceable.
Models HI93501, HI935004 and HI935008 are compliant to
standard EN 13485.
The model HI9350011 thermometer utilize a K-type thermo‑
couple style probe and provide the large range of temperature
measurement; from -50.0 to 300 °C (-58.0 to 572 °F).
HI9350011 uses a replaceable style probe (FC766C1).
The model HI9350041 thermometer utilize a T-type
thermocouple style probe and provide the large range of
temperature measurement; from -50.0 to 300 °C (-58.0
to 572 °F). The model HI9350041 uses a replaceable style
probe (FC767C1).
5
HI93501 SPECIFICATIONS
Main features:
• EN 13485 compliant
• FC762PW thermistor probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI93501
Interchangeable FC762 series
Thermistor probes
6
-50.0 to 150.0 °C
-58.0 to 302.0 °F
Resolution
0.1 °C
0.1 °F
Meter Accuracy
@ 23.0 °C ±5 °C
Probe Accuracy
FC762PW
±0.1 °C (-50.0 to 150.0 °C)
±0.2 °F (-58.0 to 302.0 °F)
±0.3 °C (-10.0 to 80.0 °C)
±0.5 °F (14 to 176 °F)
±0.7 °C / ±1.3 °F remaining range
Response time for
10 seconds
90% of final value
Battery type / life
1.5V AAA (3 pcs.)
approx. 4500 hours of continuous use
Auto-Off
user selectable: after 8 min, 60 min
or disabled
HI93501 SPECIFICATIONS
Range*
For air measurement: Type E
Environment
For product internal measurement:
Rated operating condition:
-20 to 50 °C (-4 to 122 °F)
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions
140 x 57 x 28 mm (5.5 x 2.2 x 1.1”)
Mass
175 g (6.17 oz.)
Certification
EN 13485:2001
Suitability: Storage and transport;
Climatic Environment: E;
Accuracy Class: 1;
*The measurement range may be limited by probe type, and
applies to the probe shaft.
7
HI935001 SPECIFICATIONS
Main features:
• FC766PW K-type thermocouple probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI935001
Interchangeable
K-type Thermocouple FC766
series probes
8
-50.0 to 199.9 °C ; 200 to 300 °C
-58.0 to 399.9 °F ; 400 to 572 °F
Resolution
0.1 °C (-50.0 to 199.9 °C)
1 °C (200 to 300 °C)
0.1 °F (-58.0 to 399.9 °F)
1 °F (400 to 572 °F)
@ 23.0 °C ±5 °C
±0.4 °C (-50.0 to 300 °C)
±0.7 °F (-58.0 to 572 °F)
Probe Accuracy
FC766PW
±1.6 °C (-50.0 to 300 °C)
±2.9 °F (-58.0 to 572 °F)
Meter Accuracy
Response time for
20 seconds
90% of final value
Battery type/life
1.5V AAA (3 pcs.)
approx. 3500 hours of continuous use
Auto-Off
user selectable: 8 min, 60 min or
disabled
HI935001 SPECIFICATIONS
Range*
Rated operating condition:
-20 to 50 °C (-4 to 122 °F)
Environment
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions
140 x 57 x 28 mm (5.5 x 2.2 x 1.1”)
Mass
178 g (6.27 oz.)
*The measurement range may be limited by probe type, and
applies to the probe shaft.
9
HI935004 SPECIFICATIONS
Main features:
• EN 13485 compliant
• FC767PW T-type thermocouple probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI935004
Interchangeable
T-type Thermocouple
FC767 series probes
10
-50.0 to 199.9 °C ; 200 to 300 °C
-58.0 to 399.9 °F ; 400 to 572 °F
Resolution
0.1 °C (-50.0 to 199.9 °C)
1 °C (200 to 300 °C)
0.1 °F (-58.0 to 399.9 °F)
1 °F (400 to 572 °F)
Meter Accuracy
±0.4 °C (-50.0 to 300 °C)
±0.7 °F (-58.0 to 572 °F)
Probe Accuracy
FC767PW
±0.6 °C (-50 to 100.0 °C)
±1.6 °C (100.0 to 300 °C)
±1.1 °F (-58 to 212 °F)
±2.9 °F (212 to 572 °F)
@ 23.0 °C ±5 °C
HI935004 SPECIFICATIONS
Range*
Response time for
20 seconds
90% of final value
Battery type/life
1.5V AAA (3 pcs.)
approx. 3500 hours of continuous use
Auto-Off
user selectable: 8 min, 60 min or
disabled
For air measurement: Type E
Environment
For product internal measurement:
Rated operating condition:
-20 to 50 °C (-4 to 122 °F)
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions
140 x 57 x 28 mm (5.5 x 2.2 x 1.1”)
Mass
178 g (6.27 oz.)
Certification
EN 13485:2001
Suitability: Storage and transport;
Climatic Environment: E;
Accuracy Class: 1;
*The measurement range may be limited by probe type, and
applies to the probe shaft.
11
HI935007 SPECIFICATIONS
Main features:
• Fixed K-type thermocouple probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI935007
Fixed
K-type Thermocouple probe
12
-50.0 to 199.9 °C ; 200 to 300 °C
-58.0 to 399.9 °F ; 400 to 572 °F
Resolution
0.1 °C (-50.0 to 199.9 °C)
1 °C (200 to 300 °C)
0.1 °F (-58.0 to 399.9 °F)
1 °F (400 to 572 °F)
System Accuracy
(Meter @ 23.0 °C ±5 °C)
±1 °C (-50.0 to 100.0 °C)
±2 °C (100.0 to 300 °C)
±1.8 °F (-58.0 to 212°F)
±3.6 °F (212 to 572 °F)
Response time for
90% of final value
20 seconds
Battery type/life
1.5V AAA (3 pcs.)
approx. 3500 hours of continuous use
Auto-Off
user selectable: 8 min, 60 min or
disabled
HI935007 SPECIFICATIONS
Range*
Rated operating condition:
-20 to 50 °C (-4 to 122 °F
Environment
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions (meter)
140 x 57 x 28 mm (5.5 x 2.2 x 1.1”)
Mass (meter)
178 g (6.27 oz.)
*The measurement range applies to the probe shaft.
13
HI935008 SPECIFICATIONS
Main features:
• EN 13485 compliant
• Fixed T-type thermocouple probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI935008
Fixed
T-type Thermocouple probe
14
-50.0 to 199.9 °C ; 200 to 300 °C
-58.0 to 399.9 °F ; 400 to 572 °F
Resolution
0.1 °C (-50.0 to 199.9 °C)
1 °C (200 to 300 °C)
0.1 °F (-58.0 to 399.9 °F)
1 °F (400 to 572 °F)
System Accuracy
(Meter @ 23.0 °C ±5 °C)
±0.5 °C (-50.0 to 100.0 °C)
±1 °C (100.0 to 300 °C)
±0.9 °F (-58.0 to 212 °F)
±1.8 °F (212 to 572 °F)
Response time for
90% of final value
20 seconds
Battery type/life
1.5V AAA (3 pcs.)
approx. 3500 hours of continuous use
Auto-Off
user selectable: 8 min, 60 min or
disabled
HI935008 SPECIFICATIONS
Range*
For air measurement: Type E
Environment
For product internal measurement:
Rated operating condition:
-20 to 50 °C (-4 to 122 °F)
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation
condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions (meter)
140 x 57 x 28 mm (5.5 x 2.2 x 1.1”)
Mass (meter)
178 g (6.27 oz.)
Certification
EN 13485:2001
Suitability: Storage and transport;
Climatic Environment: E;
Accuracy Class: 0.5;
*The measurement range applies to the probe shaft.
15
HI9350011 SPECIFICATIONS
Main features:
• FC766C1 K-type thermocouple probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI9350011
Interchangeable
K-type Thermocouple FC766C1
series probes
16
-50.0 to 199.9 °C ; 200 to 300 °C
-58.0 to 399.9 °F ; 400 to 572 °F
Resolution
0.1 °C (-50.0 to 199.9 °C)
1 °C (200 to 300 °C)
0.1 °F (-58.0 to 399.9 °F)
1 °F (400 to 572 °F)
@ 23.0 °C ±5 °C
±0.4 °C (-50.0 to 300 °C)
±0.7 °F (-58.0 to 572 °F)
Probe Accuracy
FC766C1
±1.6 °C (-50.0 to 300 °C)
±2.9 °F (-58.0 to 572 °F)
Meter Accuracy
Response time
4 seconds
for 90% of final value
Battery type/life
1.5V AAA (3 pcs.)
approx. 3500 hours of continuous
use
Auto-Off
user selectable: 8 min, 60 min or
disabled
HI9350011 SPECIFICATIONS
Range*
Rated operating condition:
-20 to 50 °C (-4 to 122 °F)
Environment
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation
condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions
140 x 57 x 28 mm (5.5 x 2.2 x
1.1”)
Mass
178 g (6.27 oz.)
*The measurement range may be limited by probe type, and
applies to the probe shaft.
17
HI9350041 SPECIFICATIONS
Main features:
• FC767C1 T-type thermocouple probe
• Calibration Check feature
• Remaining battery life indication/low battery detection
• Auto-off capability
• Waterproof casing IP65
HI9350041
Interchangeable
T-type Thermocouple
FC767C1 series probes
18
-50.0 to 199.9 °C ; 200 to 300 °C
-58.0 to 399.9 °F ; 400 to 572 °F
Resolution
0.1 °C (-50.0 to 199.9 °C)
1 °C (200 to 300 °C)
0.1 °F (-58.0 to 399.9 °F)
1 °F (400 to 572 °F)
Meter Accuracy
±0.4 °C (-50.0 to 300 °C)
±0.7 °F (-58.0 to 572 °F)
Probe Accuracy
FC767C1
±0.6 °C (-50 to 100.0 °C)
±1.6 °C (100.0 to 300 °C)
±1.1 °F (-58 to 212 °F)
±2.9 °F (212 to 572 °F)
Response time for
90% of final value
4 seconds
Battery type/life
1.5V AAA (3 pcs.)
approx. 3500 hours of continuous use
Auto-Off
user selectable: 8 min, 60 min or
disabled
@ 23.0 °C ±5 °C
HI9350041 SPECIFICATIONS
Range*
For air measurement: Type E
Environment
For product internal measurement:
Rated operating condition:
-20 to 50 °C (-4 to 122 °F)
Limiting condition:
-30 to 50 °C (-22 to 122 °F)
Storage and transportation
condition:
-40 to 70 °C (-40 to 158 °F)
Relative Humidity 100 %
Storage/transport
temperature
-40 to 70 °C (-40 to 158 °F)
Dimensions
140 x 57 x 28 mm (5.5 x 2.2 x 1.1”)
Mass
178 g (6.27 oz.)
*The measurement range may be limited by probe type, and
applies to the probe shaft.
19
OPERATIONAL GUIDE
Each meter is supplied with batteries. Remove the battery
cover, unwrap the supplied batteries and install them while
paying attention to their polarity (see the “Battery replacement”
section for more details).
TAKING MEASUREMENTS
Connect a suitable temperature probe to the meter (see page
25 for a complete list of Hanna Instruments probes).
Note: HI935007 and HI935008 have a fixed probe.
To switch the meter ON, press the ON/OFF button on the front
of the meter. If it does not turn on, make sure that the batteries
are properly installed in their place.
At start-up the thermometer displays all LCD segments for a few
seconds, followed by the percentage indication of the remaining
battery life. It then enters CAL Check mode followed by measurement
mode.
HI93501
HI935001 and HI935007
HI935004 and HI935008
HI9350011
HI9350041
20
Battery Life
CALIBRATION CHECK
All the models are provided with an automatic CAL Check
feature at start-up, which allows the user to check the meter
calibration status.
If the meter is calibrated, the display will stabilize and show
“0.0 °C” (±0.3 °C) for a few seconds.
OPERATIONAL GUIDE
Note: The full display can be frozen by keeping the
ON/OFF button pressed.
MEASUREMENT
If a temperature probe is plugged in, the meter displays the
measured temperature.
HI93501
HI935001 and HI935007
HI935004 and HI935008
HI9350011
HI9350041
21
OPERATIONAL GUIDE
If the measurement is unstable, the meter will display ⌛ until
the measurement becomes stable.
If the reading is slightly beyond the measurement range, the
meter flashes the closest full-scale value.
E.g.HI93501
To switch the meter OFF, press the ON/OFF key.
If no temperature probe is plugged in, or the signal is interrupted
the display shows blinking dashes and the “NO PROBE” message
is displayed.
WARNING MESSAGES
NO PROBE message
If no temperature probe is plugged in, or the signal is interrupted
the display shows blinking dashes and the “NO PROBE” message
is displayed.
DAMAGED PROBE message
When the readings are very far from the measurement range,
the probe may be damaged. The display shows blinking dashes
and “NO PROBE” message is displayed.
CALIBRATION CHECK ERROR MESSAGE
If during CALIBRATION CHECK the meter displays more than
±0.3 °C, then an “Err” message is displayed.
Contact your local Hanna Instruments Office for recalibration.
22
Enter Setup Mode to change the temperature unit (°C or °F) or
to adjust the auto-off time. To enter the Setup Mode, press the
SET button located inside the battery compartment.
METER SETUP
METER SETUP
SET button
• Use the ON/OFF button to switch between “°C” or “°F”.
Long press the ON/OFF button once to save and continue
to auto-off settings.
• Use the ON/OFF button to cycle through the following
Auto-Off choices: 8 minutes (“d08”, default value),
60 minutes (“d60”) or OFF (“d---”). Long press the
ON/OFF button once to save and continue to normal
operation.
8 minutes
60 minutes
Disabled
23
BATTERY REPLACEMENT
Each model is provided with an auto-off feature, which switches
the meter off after a certain period of non-use, to save battery life.
BATTERY REPLACEMENT
When the remaining battery life is less than
10% the battery tag blinks on the display to
warn the user.
If the display shows battery level “0%” it means the battery
charge is not enough to guarantee good measurements and
the Battery Error Prevention System (BEPS) turns the meter off.
Immediately replace the batteries with new ones.
The batteries are accessed by opening the battery cover on the
back of the instrument.
Replace the three 1.5V AAA alkaline batteries located in the
battery compartment, observing the indicated polarity.
Replace the battery cover making sure that the gasket is in place.
24
TEMPERATURE PROBES
All Hanna Instruments temperature probes are pre-calibrated at
the factory and are ready-to-use. These interchangeable probes
make it possible to switch from one probe to another within the
family without requiring recalibration.
Note: To clean the meters and the probes, do not use
aggressive detergents. It is recommended to use water.
General purpose penetration probe with
1 m (3.3’) white cable & white handle
for HI93501
FC766PW
Penetration, K-type thermocouple probe
with 1 m (3.3’) white cable & white
handle for HI935001
FC767PW
Penetration, T-type thermocouple probe
with 1 m (3.3’) white cable & white
handle for HI935004
FC766C1
Penetration, K-type thermocouple probe
with 1 m (3.3’) white cable & white
handle for HI9350011
FC767C1
Penetration, T-type thermocouple probe
with 1 m (3.3’) white cable & white
handle for HI9350041
HI710026
Protective Rubber Boot with circular
hole used for HI93501, HI935007 and
HI935008
HI710027
Protective Rubber Boot with rectangular
hole used for HI935001, HI935004,
HI9350011 and HI9350041
ACCESSORIES
FC762PW
25
CERTIFICATION
All Hanna Instruments conform to the CE European
Directives.
RoHS
compliant
Disposal of Electrical & Electronic Equipment. The
product should not be treated as household waste. Instead
hand it over to the appropriate collection point for the
recycling of electrical and electronic equipment which will
conserve natural resources.
Disposal of waste batteries. This product contains
batteries, do not dispose of them with other household
waste. Hand them over to the appropriate collection point
for recycling.
Ensuring proper product and battery disposal prevents
potential negative consequences for the environment and
human health, which may be caused by inappropriate
handling. For more information, contact your city, your local
household waste disposal service, the place of purchase or
go to www.hannainst.com.
MAN93500X 07/18
26
Recommendations Before using these products, make
sure they are entirely suitable for
for Users
your specific application and for the
environment in which they are used.
Any variation introduced by the user
to the supplied equipment may
degrade the meters’ performance.
For yours and the meter’s safety
do not use or store the meter in
hazardous environments.
Warranty HI93501, HI935001, HI935004, HI935007,
HI935008, HI9350011 and HI9350041
are warranted for two years against defects
in workmanship and materials when used
for their intended purpose and maintained
according to instructions. This warranty is
limited to repair or replacement free of
charge. Damage due to accidents, misuse,
tampering or lack of prescribed maintenance
is not covered.
If service is required, contact your local Hanna
Instruments Office. If under warranty, report
the model number, date of purchase, serial
number and the nature of the problem. If the
repair is not covered by the warranty, you will
be notified of the charges incurred. If the meter
is to be returned to Hanna Instruments, first
obtain a Returned Goods Authorization number
from the Technical Service department and
then send it with shipping costs prepaid. When
shipping any meter, make sure it is properly
packed for complete protection.
Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance notice.
27
World
Headquarters
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com
Local Office
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
Phone: 800.426.6287
Fax: 401.765.7575
e-mail: tech@hannainst.com
MAN93500X
Printed in ROMANIA
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF