Hanna Instruments | HI99181 | Owner Manual | Hanna Instruments HI99181 Owner Manual

Hanna Instruments HI99181 Owner Manual
Waterproof
pH & Temperature Meter
for Skin & Scalp
INSTRUCTION MANUAL
HI99181
Thank you for choosing a Hanna Instruments product.
Dear
Customer, Please read this instruction manual carefully before
using this meter.
This manual will provide you with the necessary
information for correct use of this meter, as well as a
precise idea of its versatility.
If you need additional technical information, do not
hesitate to e‑mail us at tech@hannainst.com or view
our worldwide contact list at www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna
Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
TABLE OF CONTENTS
Preliminary examination............................................ 4
General description and Intended use......................... 5
Main features............................................................ 6
Specifications............................................................ 7
Display description.................................................... 8
Operational guide...................................................... 9
Meter setup.............................................................. 14
Battery replacement.................................................. 15
Accessories................................................................ 16
Electrode maintenance............................................... 17
Certification............................................................... 18
Recommendations for users........................................ 19
Warranty.................................................................. 19
3
PRELIMINARY EXAMINATION
4
PRELIMINARY EXAMINATION
Remove the instrument and accessories from the packaging
and examine it carefully to make sure that no damage
has occurred during shipping. Notify your nearest Hanna
Instruments Customer Service Center if damage is observed.
Each HI99181 is delivered in a rugged carrying case and
is supplied with:
• HI14143/50 pH/temperature probe with flat tip, DIN
connector and 1m (3.3’) cable
• HI70004 pH 4.01 buffer (1 sachet)
• HI70007 pH 7.01 buffer (1 sachet)
• HI700620 Cleaning and disinfection solution for skin
residuals (1 sachet)
• HI700621 Electrode cleaning solution for skin grease
and sebum (1 sachet)
• 100 mL beaker (1 pc.)
• 1.5V AAA alkaline batteries
• Instrument quality certificate
• Electrode quality certificate
• Instruction manual
Note: Save all packing material until you are sure that
the instrument works correctly. Any damaged or defective
item must be returned in its original packing material
with the supplied accessories.
The HI99181 portable pH meter together with a
HI14143/50 combination pH electrode, is designed to
measure pH on skin and scalp.
The HI99181 features a two button operation system and
is simple to use. It has a waterproof and compact casing,
large dual-line display, and automatic pH calibration at
one or two points.
The HI14143/50 has a built-in temperature sensor for
temperature compensated pH readings and an integrated
preamplifier to provide measurements impervious to noise
and electrical interferences. The flat tip of the HI14143/50
provides optimal contact between the sample and the
sensor. The HI14143/50 has an open junction design and
utilizes a non-flowing gelled reference electrolyte. Both the
pH sensing surface and reference junction come into contact
with the skin and equilibrate quickly and reproducibly. It
also has a lower profile for easier handling compared to
laboratory sensors; the glass body is just 50 mm in length.
GENERAL DESCRIPTION AND INTENDED USE
GENERAL DESCRIPTION AND INTENDED USE
5
MAIN FEATURES
6
MAIN FEATURES
• Simultaneous pH and temperature measurements on a
large dual-line LCD display
• Automatic pH calibration at one or two points within
two memorized buffer sets (standard or NIST)
• Selectable temperature unit (°C or °F)
• Electrode condition indicator
• mV of pH measurement for electrode check
• HI14143/50 dedicated pH probe with integrated
temperature sensor
• Probe quick connect system
• Battery life indication and low battery detection
• Keystroke confirmation tone
• Auto-off function
• Waterproof casing IP67
Range*
-2.00 to 16.00 pH / -2.0 to 16.0 pH
±825 mV (pH-mV)
-5.0 to 105.0 °C/23.0 to 221.0 °F
Resolution
0.01 pH / 0.1 pH
1 mV
0.1 °C/0.1 °F
SPECIFICATIONS
SPECIFICATIONS
±0.02 pH / ±0.1 pH
Accuracy
±1 mV (pH-mV)
@ 25°C / 77°F ±0.5 °C up to 60 °C; ±1.0 °C outside
±1.0 °F up to 140 °F; ±2.0 °F outside
Temperature
compensation
Automatic
-5.0 to 105.0 °C/23.0 to 221.0 °F
pH calibration
Automatic, 1 or 2 point selectable buffer
set Standard (4.01, 7.01, 10.01) or
NIST (4.01, 6.86, 9.18)
Probe (included)
HI14143/50 preamplified pH/
temperature probe with flat tip, DIN
connector and 1 m (3.3’) cable
Battery type/life
1.5V AAA (3 pcs.)
approx. 1400 hours of continuous use
Auto-off
user selectable: after 8 min, 60 min or
disabled
Environment
0 to 50 °C (32 to 122 °F)
RH max. 100%
Meter
dimensions
154 x 63 x 30 mm (6.1 x 2.5 x 1.2”)
Weight
(with batteries)
196 g (6.91 oz.)
Case ingress
protection rating
IP67
* the HI14143/50 is limited to be used from 0 to 12 pH and from
0 to 50 °C temperature (32 to 122 °F).
7
DISPLAY DESCRIPTION
8
DISPLAY DESCRIPTION
1
2
3
4
5
6
7
8
9
10
11
Automatic Temperature Compensation indicator
Stability indicator
Battery percentage
Electrode condition indicator
Low battery indicator
Measurement unit
Primary LCD
Temperature unit
Secondary LCD
Meter modes indicator
pH calibration buffer(s) used
Each meter is supplied with batteries. Before using the meter
for the first time, open the battery compartment and insert
batteries, observing the polarity (see “Battery Replacement“).
CONNECTING THE ELECTRODE
With the meter turned off, connect the HI14143/50 probe
to the DIN socket on the bottom of the meter by aligning the
pins and pushing in the plug firmly. Remove the protective
cap from the probe before taking any measurements.
TURNING THE METER ON
To turn the meter ON, press the button on
the front of the meter. If it does not turn on, make
sure that the batteries are properly installed in their place.
The meter is provided with an active acoustic signal when
a key is pressed. At start-up the meter displays all LCD
segments for a few seconds, followed by the percentage
indication of the remaining battery life, displaying “WAIT“
until electrode check is in process then the meter enters the
normal measurement mode.
LCD check
up
→
Electrode
check
→
OPERATIONAL GUIDE
OPERATIONAL GUIDE
Battery life
→
Measurement
mode
Note: The meter detects the presence and the type of the
probe at its input.
- If the probe is not connected the message “NO”
”PROBE“ appears alternatively on the secondary LCD
with “---“ blinking on the first LCD line.
- If the probe is not compatible “WRONG” ”PROBE“
message appears alternatively on the secondary LCD
with “---“ blinking on the first LCD line.
- If the readings are out of range, the nearest range
limits are displayed blinking (e.g. -2.00 pH -5.0 °C).
9
OPERATIONAL GUIDE
SELECTING MEASUREMENT RANGE
While in measurement mode, press the SET button to select
pH or pH-mV measurement on the first LCD line.
FREEZING MEASUREMENT VALUES
While in measurement mode, press and hold the SET
button until “HOLD“ appears on the secondary LCD. The
“HOLD“ remains for 1 second and reading of pH, mV
and temperature will be frozen on the LCD with “H” blinking.
→
Press any button to resume active measurements.
ENTERING CALIBRATION MODE
tag
Press and hold the button until ”POWER” and
is replaced by “STD“ and
tag. Release the button.
ENTERING SETUP MODE
tag is
Press and hold button until “STD“ and
replaced by “SETUP“ and
tag. Release the button.
TURNING THE METER OFF
While in measurement mode, press the button.”POWER”
and
tag will appear. Release the button.
pH MEASUREMENT & CALIBRATION
Make sure the meter has been calibrated before use.
If the probe is dry, soak it in HI70300 storage solution for
30 minutes to reactivate it. If soiled, clean the electrode by
soaking in cleaning solution for 20 minutes, then rinse the tip
and soak in storage solution at least 30 minutes before use.
Recalibrate before using.
Rinse the electrode off well with distilled/deionized water
or HI70960 preparation solution for solid or semi-solid
samples. Shake off excess droplets.
Place the flat tip of the electrode on
the sample to be tested. Wait until
the tag on the LCD disappears.
10
OPERATIONAL GUIDE
The LCD displays the pH value (automatically compensated
for temperature) on the primary LCD, while the secondary
LCD displays the sample temperature.
If measurements are taken in different samples successively,
rinse the probe tip thoroughly in distilled or deionized water
to eliminate cross-contamination.
For better accuracy, frequent calibration of the pH sensor
with the meter is recommended. In addition, the meter
must be recalibrated:
a) whenever the pH electrode is replaced.
b) after testing aggressive chemicals.
c) when high accuracy is required.
d) at least once a month.
e) after cleaning the sensor.
pH calibration
Enter calibration mode while in pH
measurement mode. Place the sensor
into the first calibration buffer. If
performing a two-point calibration,
use pH 7.01 (pH 6.86 for NIST) buffer first. The meter
will enter the calibration mode, displaying “pH 7.01 USE”
and tag blinking (or “pH 6.86 USE” for NIST).
Follow directions for single and two-point calibration below:
Single-point calibration
1. Place the probe in any buffer from the selected buffer
set. The meter will automatically recognize the buffer value.
2. If the buffer is not recognized or the calibration offset
is out of the accepted range “---- WRONG“ is displayed.
3. If the buffer is recognized “REC“ is displayed then
“WAIT“ until the calibration is accepted.
If using pH 7.01 (or pH 6.86 for NIST), after acceptance
of the buffer press any key to exit. The ”SAVE” message is
displayed and the meter returns to pH measurement mode.
If using pH 4.01 or 10.01 (or pH 9.18 for NIST) buffer the
”SAVE” message is displayed and the meter returns to pH
measurement mode.
11
OPERATIONAL GUIDE
12
Two-point calibration
Proceed with steps 1 through 3 under single-point
calibration using 7.01 pH (pH 6.86 for NIST) buffer first.
Then follow steps below:
The ”pH 4.01 USE” message is then displayed.
Place the probe in the second calibration buffer (pH 4.01
or 10.01, or, if using NIST, pH 4.01 or 9.18). When the
second buffer is accepted, the LCD will display “SAVE”
for 1 second and the meter will return to the normal
measurement mode.
If the buffer is not recognized or the slope is out of accepted
range ”--- WRONG” is displayed. Change the buffer, clean
the electrode or press any key to exit calibration.
For better accuracy it is always recommended to carry out
a two-point calibration.
After the calibration procedure has been completed, the
“CAL” is turned on together with the calibrated points.
Exiting calibration and resetting default values
After entering the calibration mode and before the first point
is accepted, it is possible to quit the procedure and return
to the last calibration data by pressing the button. The
LCD displays “ESC“ for 1 second and the meter returns to
normal mode.
To reset the default values and clear a previous calibration,
press the SET button after entering the calibration mode
and before the first point is accepted.
The LCD displays “CLEAR“ for 1 second, the meter resets
to the default calibration and the “CAL” with the calibrated
points on the LCD disappears.
OPERATIONAL GUIDE
ELECTRODE CONDITION
The display is provided with a probe icon (unless the
feature is disabled from setup) which indicates the electrode
status after calibration. The “condition” remains active for
12 hours (unless the batteries are removed).
The electrode condition is evaluated only if the current
calibration has two points.
5 bars: excellent condition
4 bars: very good condition
3 bars: good condition
2 bars: fair condition
1 bar: poor condition
1 bar blinking: very poor condition
With 1 bar it is recommended to clean the electrode and
recalibrate. If there is still only 1 bar or 1 bar blinking
replace the probe.
Sensor check
Setting the meter to pH-mV range the user can check the
sensor status at any time. The offset value is the reading
in pH 7.01 buffer (@ 25 °C/77 °F). If this reading is
outside the range ±30 mV, the electrode is considered
“very poor”. The slope value of the sensor is the difference
between readings in pH 7.01 and in pH 4.01 buffers. When
the slope reaches the value of about 150 mV, the electrode
is considered “very poor“. When “poor“ or “very poor“, it
is recommended to replace it with a new one.
Note: To ensure reliable readings, the electrode must
be cleaned with cleaning solution and then hydrated in
storage solution for a minimum of 30 minutes before
calibrating the probe.
13
METER SETUP
14
METER SETUP
Setup mode allows the selection of the Temperature unit,
Auto-off, Beep confirmation tone, the type of pH buffer set,
the Resolution and Information. To enter Setup mode press
and hold button until “STD“ and
tag is replaced
tag. Release the button.
by “SETUP“ and
• “TEMP“ is displayed on the secondary
LCD with the current temperature unit
(e.g. “TEMP °C“), for °C/°F selection,
use the SET button. After the
temperature unit has been selected,
press to confirm and to enter the “A-OFF“ selection.
• Use the SET button, to navigate
through the auto-off choices: 8
minutes (“8”, default value), 60
minutes (“60”) or disabled (“---”).
Press to confirm and to enter the
“BEEP“ selection.
• To switch the beep tone ON or OFF,
press the SET button; press to
confirm and to enter the calibration
buffer selection “pH 7.01 BUFF“.
• To change the buffer set, the meter
will show the current buffer set: “pH
7.01 BUFF” (for standard buffer
set: 4.01/7.01/10.01) or “pH
6.86 BUFF” (for NIST buffer set:
4.01/6.86/9.18). Change the set
with the SET button. Press to confirm and to enter pH
resolution selection “RESOL“.
• To change the pH measurement
resolution between “0.1“ and
“0.01“ use the SET button; then
to confirm and to enter
press
electrode calibration information
“INFO“ selection.
• To switch the electrode condition
indicator on the LCD ON or OFF,
press the SET button; press to exit
setup options. Change the set with the
SET button, then press to confirm
and to return to normal mode.
When the remaining battery life is less
than 10% the battery tag blinks on the
display to warn the user.
Battery Error Prevention System (BEPS)
If the battery is too weak (“0%”) the display shows “bAtt“,
“DEAD“ for a few seconds then the meter powers off.
Immediately replace the batteries with new ones.
The batteries are accessed by opening the battery cover
on the back of the instrument. Remove protective boot
if present.
BATTERY REPLACEMENT
BATTERY REPLACEMENT
Replace the three 1.5V AAA alkaline batteries located in
the battery compartment, observing the indicated polarity.
Replace the battery cover making sure that the gasket is
in place.
15
ACCESSORIES
ACCESSORIES
HI14143/50
HI7004L
HI7006L
HI7007L
HI7009L
HI7010L
HI70300L
HI7061L
HI70620L
HI70621L
HI70960
HI710029
HI76405
HI710142
HI77400P
Combination preamplified pH/temperature probe
with flat tip, DIN connector and 1 m (3.3’) cable
pH 4.01 buffer solution, 500 mL
pH 6.86 buffer solution, 500 mL
pH 7.01 buffer solution, 500 mL
pH 9.18 buffer solution, 500 mL
pH 10.01 buffer solution, 500 mL
Storage solution for pH and ORP electrodes,
500 mL
Electrode cleaning solution for general use,
500 mL
Cleaning and disinfection solution for skin
residuals, 500 mL
Electrode cleaning solution for skin grease and
sebum, 500 mL
Preparation solution for solid and semi-solid
samples, 30 mL bottle
Blue silicon rubber boot
Electrode holder with steel base
Black carrying case for HI991XX portable
instruments
Calibration kit (pH 4.01 and pH 7.01, 20 mL
sachets, 5 pcs. each)
Code and serial
number on the
other side
Temperature sensor
junction reference
pH sensor
Quick connector with Rubber
protection, and 1 meter cable
16
PREPARATION
• Remove the protective cap. Do not be alarmed if any
salt deposits are present. Rinse with water.
• Shake the electrode down as you would do with a
clinical thermometer to eliminate any air bubbles inside
the glass tip.
• If the glass tip and/or junction are dry, soak the
electrode in HI70300 storage solution for a minimum
of 30 minutes.
• Rinse with water.
• Calibrate before using.
STORAGE
• To ensure a quick response, the glass tip and the junction
should be kept moist and not allowed to dry.
• Replace protective cap with a few drops of HI70300
storage solution. Follow PREPARATION above before
taking measurements.
Note: Never store the electrode in distilled water.
PERIODIC MAINTENANCE
• Inspect the electrode for any scratches or cracks. If any
present, replace the electrode.
• Rinse off any salt deposits with water.
• Follow the STORAGE procedure above.
CLEANING PROCEDURE
• Soak in Hanna HI7061 electrode cleaning solution for
general use, HI700620 cleaning and disinfection solution
for skin residuals or HI700621 electrode cleaning solution
for skin grease and sebum for 20 minutes. Rinse with
water.
• Soak the electrode in HI70300 storage solution for a
minimum of 30 minutes. Rinse with water. Calibrate
before using.
TROUBLESHOOTING
• pH Meter: Follow the meter’s operating and calibration
procedures.
• Electrode: Evaluate your electrode performance based on
the Sensor check procedure on page 13.
ELECTRODE MAINTENANCE
ELECTRODE MAINTENANCE
17
CERTIFICATION
CERTIFICATION
All Hanna Instruments conform to the CE European Directives.
RoHS
compliant
Disposal of Electrical & Electronic Equipment. The
product should not be treated as household waste. Instead
hand it over to the appropriate collection point for the
recycling of electrical and electronic equipment which will
conserve natural resources.
Disposal of waste batteries. This product contains
batteries, do not dispose of them with other household
waste. Hand them over to the appropriate collection point
for recycling.
Ensuring proper product and battery disposal prevents
potential negative consequences for the environment and
human health. For more information, contact your city,
your local household waste disposal service, the place of
purchase or go to www.hannainst.com.
MAN99181 12/18
18
Recommendations Before using this product, make sure
it is entirely suitable for your specific
for users
application and for the environment
in which it is used. Any variation
introduced by the user to the supplied
equipment may degrade the meter’s
performance. For yours and the meter’s
safety do not use or store the meter in
hazardous environments.
Warranty HI99181 is warranted for two years against defects
in workmanship and materials when used for its
intended purpose and maintained according to
instructions. Electrodes and probes are warranted
for a period of six months. This warranty is limited
to repair or replacement free of charge. Damage
due to accidents, misuse, tampering or lack of
prescribed maintenance is not covered.
If service is required, contact your local Hanna
Instruments Office. If under warranty, report the
model number, date of purchase, serial number
(see engraved on the back of the meter) and the
nature of the problem. If the repair is not covered
by the warranty, you will be notified of the
charges incurred. If the meter is to be returned to
Hanna Instruments, first obtain a Returned Goods
Authorization number from the Technical Service
department and then send it with shipping costs
prepaid. When shipping any meter, make sure it is
properly packed for complete protection.
Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance notice.
World
Headquarters
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com
MAN99181
Printed in ROMANIA
Was this manual useful for you? yes no
Thank you for your participation!