EMI-Hardened Op Amps Reduce Errors in EKGs
Soufiane Bendaoud
Precision Amplifiers
Medical devices such as infusion pumps, EKG’s and
pulse oximeters are designed with great care for
electromagnetic compatibility (EMC). In other words,
they must be compatible with their environment and
not cause any interference.
Before the creation of electromagnetic interference
(EMI)-hardened amplifiers, board and system-level
engineers struggled to come up with the best possible
filtering schemes without compromising signal chain
performance, especially that of the analog front end.
Operational amplifiers needed special care and
attention due to common mishaps, such as placing a
capacitor across the inputs of the amplifier.
While the goal was to eliminate undesirable voltage
spikes at the output of the amplifier, it usually resulted
in unstable circuits.
316 mV is present at the output of the op amp [ (100
mV / 31.6) × 100]. Now assume the output is fed to a
12-bit ADC with a 5-V FSR.
The loss of counts caused the injected signal (EMI)
can be computed as follows:
5 V / (212) =1.22 mV. Dividing the output of the op amp
by 1.22 mV (316 mV / 1.22 mV) to determine a loss of
nearly 260 counts.
Using an op amp like the OPA192 reduces the count
loss to roughly eight.
When making a design more immune to EMI without
compromising the circuit performance, look up TI’s
precision amplifiers portfolio.
See this clip for additional information on avoiding
electromagnetic interference (EMI).
Electrodes
MUX
Similarly, filters on the output, passive, low-pass filters
(RC) particularly, can be a headache. Lowering the
cutoff frequency often implies using a large capacitor,
which not every op amp can drive. Conversely,
increasing the resistor value causes gain errors.
ADC
MCU
RS-232
REF
Put The Op Amp To Work
Impedance
Match
The easiest way to deal with rejecting unwanted RF
signals and electromagnetic interference is selecting
active components with integrated filters. Every TI op
amp designed in the last seven years provides such
filters.
Figure 1. Typical Block Diagram for EKG
160.0
140.0
The choice of amplifiers ranges from the INA828 highvoltage instrumentation amplifier to the ubiquitous
OPA192, which comes in single, dual, and quad
options.
To better appreciate the benefits of EMI filters in the
amplifier, consider an example:
A first-order filter at 10 MHz rejects 40 dB at 1 GHz.
However, a 10 MHz device with a cut-off at 100 MHz
has a rejection of only 20 dB at 1 GHz.
PRF = -10 dBm
VSUPPLY = ±18 V
VCM = 0 V
120.0
EMIRR IN+ (dB)
To determine whether the EMI rejection is sufficient for
the application, a plot like the CMRR and PSRR,
called the EMIRR (EMI rejection ratio), is provided in
the datasheet.
Output
Buffer
100.0
80.0
60.0
40.0
20.0
0.0
10M
100M
1G
Frequency (Hz)
10G
C017
Figure 2. OPA192 EMI Rejection Ratio Plot Versus
Frequency
EMI errors can have serious consequences on the
system. Suppose 100 mV is injected into an amplifier
with gain of 100. This example uses an op amp with
no EMI filters, but still provides 30 dB of rejection (1
GHz).
SBOA329A – January 2019 – Revised August 2019
Submit Documentation Feedback
EMI-Hardened Op Amps Reduce Errors in EKGs Soufiane Bendaoud
Copyright © 2019, Texas Instruments Incorporated
1
www.ti.com
Table 1. Alternative Device Recommendations
Device
Unity Gain Bandwidth
Description
OPA189
14 MHz
14 MHz, MUX-Friendly, Low-noise, Zero-Drift, RRO, CMOS Precision
Operational Amplifier
OPA188
2 MHz
Precision, Low-Noise, Rail-to-Rail Output, 36 V Zero-Drift Operational Amplifier
OPA388
10 MHz
10 MHz, CMOS, Zero-Drift, Zero-Crossover, True RRIO Precision Operational
Amplifier
OPA333
350 kHz
1.8 V, 17 µA, microPower, Precision, Zero-Drift CMOS Op Amp
OPA187
550 kHz
Zero Drift, Low-Power, Rail-to-Rail Output Amplifier
2
EMI-Hardened Op Amps Reduce Errors in EKGs Soufiane Bendaoud
SBOA329A – January 2019 – Revised August 2019
Submit Documentation Feedback
Copyright © 2019, Texas Instruments Incorporated
IMPORTANT NOTICE AND DISCLAIMER
TI PROVIDES TECHNICAL AND RELIABILITY DATA (INCLUDING DATASHEETS), DESIGN RESOURCES (INCLUDING REFERENCE
DESIGNS), APPLICATION OR OTHER DESIGN ADVICE, WEB TOOLS, SAFETY INFORMATION, AND OTHER RESOURCES “AS IS”
AND WITH ALL FAULTS, AND DISCLAIMS ALL WARRANTIES, EXPRESS AND IMPLIED, INCLUDING WITHOUT LIMITATION ANY
IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR NON-INFRINGEMENT OF THIRD
PARTY INTELLECTUAL PROPERTY RIGHTS.
These resources are intended for skilled developers designing with TI products. You are solely responsible for (1) selecting the appropriate
TI products for your application, (2) designing, validating and testing your application, and (3) ensuring your application meets applicable
standards, and any other safety, security, or other requirements. These resources are subject to change without notice. TI grants you
permission to use these resources only for development of an application that uses the TI products described in the resource. Other
reproduction and display of these resources is prohibited. No license is granted to any other TI intellectual property right or to any third
party intellectual property right. TI disclaims responsibility for, and you will fully indemnify TI and its representatives against, any claims,
damages, costs, losses, and liabilities arising out of your use of these resources.
TI’s products are provided subject to TI’s Terms of Sale (www.ti.com/legal/termsofsale.html) or other applicable terms available either on
ti.com or provided in conjunction with such TI products. TI’s provision of these resources does not expand or otherwise alter TI’s applicable
warranties or warranty disclaimers for TI products.
Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265
Copyright © 2019, Texas Instruments Incorporated