Application Report
SLOA253 – August 2017
Understanding CW Mode for Ultrasound AFE Devices
ABSTRACT
Ultrasound imaging is one of the most common medical imaging modalities in use today due to its safety
and cost-effectiveness. There are numerous ultrasound image modes like A-mode, B-mode, Doppler
ultrasound, and other variations. As opposed to A-mode and B-mode ultrasound, continuous wave (CW)
Doppler ultrasound helps to identify not just the location of tissue but also the motion of fluids and organs
within the body. Using the Doppler effect—where the frequency observed by a receiver changes based on
the motion of the transmitter—the direction and velocity of body fluids can be determined from the CW I
and Q outputs of Texas Instruments' AFE58xxxx devices.
1
2
3
Contents
Introduction ................................................................................................................... 2
Background and Theory .................................................................................................... 3
Experiment and Data ........................................................................................................ 7
List of Figures
1
CW Topology in the AFE58xxx8 ........................................................................................... 2
2
Phase Generation for 16× Clock Mode ................................................................................... 4
3
16X Mode Phase Timing Diagram
4
Phase Generation for 8X and 4X Clock Mode ........................................................................... 5
5
Sine and Cosine Phase Relationship ..................................................................................... 7
6
Filtered I and Q Outputs for Positive Delta Frequency ................................................................. 7
7
Filtered I and Q Outputs for Negative Delta Frequency ................................................................ 8
8
AFE58xxx8 Default CW Jumper Configuration .......................................................................... 8
........................................................................................
5
List of Tables
Trademarks
Texas Instruments is a trademark of Texas Instruments.
All other trademarks are the property of their respective owners.
SLOA253 – August 2017
Submit Documentation Feedback
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
1
Introduction
1
www.ti.com
Introduction
The CW signal path of Texas Instruments™ AFE58xxxx devices consist of a low-noise amplifier (LNA), a
voltage-to-current converter, two passive mixers (per channel), a pattern generator, a crosspoint switch,
and shared summing amplifiers (with low pass filters), shown in Figure 1.
I-CLK
LNA1
I-Channel
Voltage-to-Current
Converter
Q-Channel
Q-CLK
Sum Amp
with LPF
1 × fcw CLK
I-Channel
Clock Distribution
Circuits
Q-Channel
N × fcw CLK
Sum Amp
with LPF
I-CLK
LNA16
I-Channel
Voltage-to-Current
Converter
Q-Channel
Q-CLK
Figure 1. CW Topology in the AFE58xxx8
At a high level, the input signal is amplified by the LNA, converted to current by the voltage-to-current
converter, then mixed with a local oscillator (LO) clock signal to capture the frequency difference
introduced by the Doppler effect. The current outputs of all sixteen channels are then summed together
and converted into an output voltage by the summing amplifier.
The data sheet for the AFE58xxx8 provides more detail on the makeup of the various components in the
CW portion of the device, but these specifics are beyond the scope of this report. Instead, this report
focuses on detailing the CW mixers, clock phase generation, and the expected outputs of the I and Q
channels.
2
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
SLOA253 – August 2017
Submit Documentation Feedback
Background and Theory
www.ti.com
2
Background and Theory
2.1
CW Frequency Mixers
The 16 CW channels of the AFE58JD18 device each contain two frequency mixers. For a single channel,
these two mixers help extract the frequency difference information for the in-phase and quadrature signal
processing paths.
The CW frequency mixers are passive and switch based with a square wave as the local oscillator (LO)
signal. The AFE58xxx8 devices also include a proprietary harmonic suppression circuit to eliminate the
effects of the largest odd harmonics (third and fifth) from the square wave (from Fourier analysis, a square
wave consists only of odd harmonics), approximating the operation of the square wave mixer to a single
sine wave mixer.
Assuming that the received voltage signal is a cosine wave reflected off of body tissue and fluid in motion,
after amplification it is converted to current and sent to the I and Q mixers (shown in Figure 1). The output
of a frequency mixer, V0(t) is the multiplication of its two input signals: the LO signal, LO(t), and the
received cosine wave, VIN(t). See Equation 1.
VIN (t)
A u cos(Zin t
M)
(1)
The LO signal is a cosine wave at the original transmit frequency of ω0 for the in-phase or I-channel, and a
sine wave for the quadrature phase or Q-channel of the mixer.
LOI (t)
B u cos(Z LO t
LOQ (t)
V0I (t)
B u sin( Z LO t
VIN (t) u LOI (t)
J)
J)
A u B u cos( Z in t
M) u cos( Z LO t
J)
(2)
From the trigonometric identity where:
cos(x) u cos(y)
V0I
0.5 u >cos(y
B·
§
¨ A u 2 ¸ u ¬ªcos(Z LO t
©
¹
0.5 u A u B »cos((Z LO
x)
cos(y
J
Z in t
Z in )t
M
M)
J)
x)@
cos(Z LO t
cos((Z LO
Z in t
M
Z in )t
M
J )¼º
J )¼º
(3)
Assuming ωint = ω0t + ωdopplert, then ω0t = ωLOt is the frequency of the transmitted signal.
V0I
0.5 u A u B u ª¬cos( Zdoppler t
M
J)
cos(2Z0 t
Zdoppler t
M
J )º¼
(4)
With the low pass filter at the summing amplifier removing the higher frequency 2ω0 term, the frequency
difference introduced by the Doppler effect, ωdoppler can be extracted from the cosine term and
subsequently, the speed of the moving tissue/fluid. Note cos(−x) = cos(x).
While the I-channel information by itself is enough to decipher the speed of tissue, the direction of motion
cannot be obtained from just the cosine. For the Q-channel (see Equation 5):
V0Q (t)
VIN (t) u LOQ (t)
A u B u cos(Zint
M) u sin(ZLOt
J)
(5)
From the trigonometric identity, where:
sin(y) u cos(x)
0.5 u >sin(y
x)
x)@
sin(y
B·
§
¨ A u 2 ¸ u > sin(ZLO t J Zin t
©
¹
0.5 u A u B u > sin((ZLO Zin )t M
V0Q
M)
sin(ZLO t
Zin t
M
J )@
J)
sin((ZLO
Zin )t
M
J )@
(6)
Assuming ωint = ω0t + ωdopplert, ω0t = ωLOt:
V0Q
0.5 u A u B u »sin( Zdoppler t
M
J)
sin(2Z0 t
Zdoppler t
M
J )¼º
(7)
Unlike the low frequency cosine term for the I-channel, the polarity of the low-frequency sine component of
the Q-channel changes depending on whether it is a positive or negative frequency. If the Doppler-shifted
frequency of VI is higher than the frequency of LO (as opposed to lower for a positive sine frequency
term), the low frequency sine term represents the sine of a negative frequency and the output phase of
the Q-channel changes by 180° (see Equation 8).
sin(–x) = –sin(x) = sin(x + 180°)
SLOA253 – August 2017
Submit Documentation Feedback
(8)
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
3
Background and Theory
www.ti.com
In other words, the Q-channel indicates whether the Doppler frequency shift is due to fluids moving
towards or away from the receiving element (information on direction in addition to speed).
2.2
CW Clock Phase Generation
To accurately extract the frequency difference information at the output of the mixers, 16 equally-spaced
phases (and 16 possible values for γ) of the input 1X clock are generated to closely match up with
corresponding phases of the received signal. These 1X clocks with different phases are then assigned to
any of the 16 channels using the SPI-controlled cross-point switch (see Figure 2).
fIN 16X Clock
INV
D Q
fIN 1X Clock
Fin 1X Clock
16-Phase Generator
1X Clock
Phase 0º
1X Clock
Phase 22.5º
SPI
1X Clock
Phase 292.5º
1X Clock
Phase 315º
1X Clock
Phase 337.5º
16:8 Crosspoint Switch
Mixer 1
1X Clock
Mixer 2
1X Clock
Mixer 3
1X Clock
Mixer 14
1X Clock
Mixer 15
1X Clock
Mixer 16
1X Clock
Figure 2. Phase Generation for 16× Clock Mode
The AFE58xxx8 devices have up to three different ways to generate the 16 phases required for CW beam
forming:
• With a 16X clock
• With an 8X or 4X clock
• With a 1X clock
When generating 16 equal phases in 22.5° increments (360° / 16) using the 16X clock method, the 1X
clock is setup as the input to a positive-edge triggered D flip-flop with an inverted 16X clock as the clock
source for the flip-flop. Therefore, the output of the flip-flop changes on the negative edge of the 16X
clock.
4
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
SLOA253 – August 2017
Submit Documentation Feedback
Background and Theory
www.ti.com
The 1X clock with the desired phase delay (γ) is output on the next 16X positive clock edge because of
the required minimum hold time (time required to keep the flip-flop input steady for the output to be valid).
See Figure 3.
Figure 3. 16X Mode Phase Timing Diagram
When generating all 16 phases in 8X or 4X clock mode, the 4X (or 8X) clock is first used to generate 1X
clocks with four equal phases of 90° (or eight equal phases of 45°) with an inverter and positive edgetriggered flip-flop. The generated 1X clocks are then fed into an I/Q CLK generator to produce the inphase (cosine) and quadrature phase (sine) signals for the mixer at the input 1X frequency (see Figure 4).
INV
4X, 8X Clock
D Q
1X Clock
I/Q CLK
Generator
LNA2 to 16
In-Phase
CLK
Weight
Quadrature
CLK
Summed
In-Phase
I/V
Weight
LNA1
I/V
Weight
Summed
Quadrature
Weight
Figure 4. Phase Generation for 8X and 4X Clock Mode
SLOA253 – August 2017
Submit Documentation Feedback
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
5
Background and Theory
www.ti.com
Assuming a received signal frequency equal to the 1X clock frequency of ω0, in order to generate a phase
difference of x × 22.5°, or x × 2π / 16, the received cosine wave is multiplied by a weight. As an example,
in order to get a phase difference of 67.5°, weights of cos(6π / 16) and sin(6π / 16) are added to the I and
Q paths for the incoming cosine wave so that the mixer output (note from Figure 4 two mixers and a
summation block for each I, Q path) becomes:
V0I4X (t)
ª
A u B u «(cos(Z in t
¬
A u B u cos(Z in t
From cos(x
y)
§ 6S ·
M) u cos(Z0 t) u cos ¨
¸
© 16 ¹
ª
§ 6S ·
M) u «(cos(Z0 t ) u cos ¨
¸
© 16 ¹
¬
cos x u cos y sin x u sin y
§ 6S ·
(cos(Z0 t ) u cos ¨
¸
© 16 ¹
§ 6S · º
M) u sin(Z0 t) u sin ¨
¸»
© 16 ¹ ¼
§ 6S · º
sin(Z0 t) u sin ¨
¸»
© 16 ¹ ¼
§ 6S ·
sin(Z0 t ) u sin ¨
¸
© 16 ¹
A u B u cos(Z in t
Then V0I4X (t)
cos(Zin t
6S ·
§
cos ¨ Z0 t
16 ¹¸
©
6S ·
§
M) u cos ¨ Z0 t
¸
16
©
¹
(9)
Note that this is the same as the equation for the 16X mode with ωLO = ω0t, γ = (6π / 16).
The Q-channel is set up similarly, with:
V0Q4X (t)
ª
A u B u «cos( Z in t
¬
§ 6S ·
M) u cos( Z0 t) u sin ¨ ¸
© 16 ¹
ª
§ 6S ·
M) u «cos(Z0 t ) u sin ¨ ¸
© 16 ¹
¬
sin x u cos y cos x u sin y
A u B u cos(Z in t
From sin(x
y)
§ 6S ·
cos(Z0 t ) u sin ¨ ¸
© 16 ¹
Then V0Q4X (t)
(cos( Z in t
§ 6S · º
M) u sin( Z0 t) u cos ¨ ¸ »
© 16 ¹ ¼
§ 6S · º
(sin( Z0 t ) u cos ¨ ¸ »
© 16 ¹ ¼
§ 6S ·
sin(Z0 t ) u cos ¨ ¸
© 16 ¹
A u B u cos( Z in t
6S ·
§
sin ¨ Z0 t
16 ¹¸
©
6S ·
§
M) u sin ¨ Z0 t
¸
16
©
¹
(10)
For the 1X mode, synchronized clocks 90° out of phase are needed by the AFE58xxx8 devices for the
phase-delayed LO signals required by the mixers. Then the sine and cosine at the 1X clock frequency are
multiplied with the appropriate weights to produce the desired phase-delayed LO frequencies just like the
8X and 4X mode.
All 16 I-channel current outputs and 16 Q-channel current outputs are summed by two separate summing
amplifiers at the CW beam-former output in order to improve signal-to-noise ratio (SNR). When the phase
of the received signal for each of the 16 channels is perfectly matched to the phase of its corresponding
LO (φ = γ), all 16 I-channel (or Q-channel) sinusoidal signal outputs constructively sum to a higher
amplitude while their uncorrelated noise components combine by the square root of the sum of squares.
Therefore, it is important to match up the input signal of each channel to the closest of the 16 possible
phases to prevent destructive interference of the signal portion of the sinusoidal outputs that can lead to
poor SNR performance.
6
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
SLOA253 – August 2017
Submit Documentation Feedback
Experiment and Data
www.ti.com
3
Experiment and Data
Figure 5 represents (theoretically) how the cosine leads the sine for a positive angle.
1.5
1
0.5
sine
0
cosine
2
4
6
8
10
12
-0.5
-1
Figure 5. Sine and Cosine Phase Relationship
Therefore, it is expected that the I-channel leads the Q-channel, because the I and Q channels represent
the cosine and sine of the same difference frequency, respectively (from Equation 4 and Equation 7).
The AFE58JD18EVM was then tested in the laboratory in CW mode, with a LO frequency of 1.953125
MHz and an input sinusoid of 1.943125 MHz, –15 dBm (see Figure 6 for results).
Figure 6. Filtered I and Q Outputs for Positive Delta Frequency
SLOA253 – August 2017
Submit Documentation Feedback
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
7
Experiment and Data
www.ti.com
The channel 1 signal is the I-channel signal at 10-kHz and channel 2 represents the Q-channel signal at
10-kHz, but lagging the I-channel data by 90°. This is as expected for a positive frequency difference (ωLO
–ωin) as the cosine leads the sine. Figure 7 shows the oscilloscope image when the frequency of the input
signal was changed to 1.963125 MHz.
Figure 7. Filtered I and Q Outputs for Negative Delta Frequency
The Q-channel now leads the I-channel because the frequency difference turned negative and therefore,
the sine shifts by 180°.
Note that on the AFE58xxx8EVMs, CW clock generation is done using the LMK0482X devices in clock
divider mode. The onboard 125-MHz crystal is divided down to 31.25 MHz for the 16X clock and 1.953125
MHz for the 1X clock, respectively. By modifying the appropriate LMK0482X register mapped to the 16X
clock divider controls, the 8X and 4X clocks can also be generated by the EVM. Conversely, Figure 8
shows that external clocks can also be provided to the board after configuring the CW clock selection
jumpers for an external source.
Figure 8. AFE58xxx8 Default CW Jumper Configuration
8
Understanding CW Mode for Ultrasound AFE Devices
Copyright © 2017, Texas Instruments Incorporated
SLOA253 – August 2017
Submit Documentation Feedback
IMPORTANT NOTICE FOR TI DESIGN INFORMATION AND RESOURCES
Texas Instruments Incorporated (‘TI”) technical, application or other design advice, services or information, including, but not limited to,
reference designs and materials relating to evaluation modules, (collectively, “TI Resources”) are intended to assist designers who are
developing applications that incorporate TI products; by downloading, accessing or using any particular TI Resource in any way, you
(individually or, if you are acting on behalf of a company, your company) agree to use it solely for this purpose and subject to the terms of
this Notice.
TI’s provision of TI Resources does not expand or otherwise alter TI’s applicable published warranties or warranty disclaimers for TI
products, and no additional obligations or liabilities arise from TI providing such TI Resources. TI reserves the right to make corrections,
enhancements, improvements and other changes to its TI Resources.
You understand and agree that you remain responsible for using your independent analysis, evaluation and judgment in designing your
applications and that you have full and exclusive responsibility to assure the safety of your applications and compliance of your applications
(and of all TI products used in or for your applications) with all applicable regulations, laws and other applicable requirements. You
represent that, with respect to your applications, you have all the necessary expertise to create and implement safeguards that (1)
anticipate dangerous consequences of failures, (2) monitor failures and their consequences, and (3) lessen the likelihood of failures that
might cause harm and take appropriate actions. You agree that prior to using or distributing any applications that include TI products, you
will thoroughly test such applications and the functionality of such TI products as used in such applications. TI has not conducted any
testing other than that specifically described in the published documentation for a particular TI Resource.
You are authorized to use, copy and modify any individual TI Resource only in connection with the development of applications that include
the TI product(s) identified in such TI Resource. NO OTHER LICENSE, EXPRESS OR IMPLIED, BY ESTOPPEL OR OTHERWISE TO
ANY OTHER TI INTELLECTUAL PROPERTY RIGHT, AND NO LICENSE TO ANY TECHNOLOGY OR INTELLECTUAL PROPERTY
RIGHT OF TI OR ANY THIRD PARTY IS GRANTED HEREIN, including but not limited to any patent right, copyright, mask work right, or
other intellectual property right relating to any combination, machine, or process in which TI products or services are used. Information
regarding or referencing third-party products or services does not constitute a license to use such products or services, or a warranty or
endorsement thereof. Use of TI Resources may require a license from a third party under the patents or other intellectual property of the
third party, or a license from TI under the patents or other intellectual property of TI.
TI RESOURCES ARE PROVIDED “AS IS” AND WITH ALL FAULTS. TI DISCLAIMS ALL OTHER WARRANTIES OR
REPRESENTATIONS, EXPRESS OR IMPLIED, REGARDING TI RESOURCES OR USE THEREOF, INCLUDING BUT NOT LIMITED TO
ACCURACY OR COMPLETENESS, TITLE, ANY EPIDEMIC FAILURE WARRANTY AND ANY IMPLIED WARRANTIES OF
MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, AND NON-INFRINGEMENT OF ANY THIRD PARTY INTELLECTUAL
PROPERTY RIGHTS.
TI SHALL NOT BE LIABLE FOR AND SHALL NOT DEFEND OR INDEMNIFY YOU AGAINST ANY CLAIM, INCLUDING BUT NOT
LIMITED TO ANY INFRINGEMENT CLAIM THAT RELATES TO OR IS BASED ON ANY COMBINATION OF PRODUCTS EVEN IF
DESCRIBED IN TI RESOURCES OR OTHERWISE. IN NO EVENT SHALL TI BE LIABLE FOR ANY ACTUAL, DIRECT, SPECIAL,
COLLATERAL, INDIRECT, PUNITIVE, INCIDENTAL, CONSEQUENTIAL OR EXEMPLARY DAMAGES IN CONNECTION WITH OR
ARISING OUT OF TI RESOURCES OR USE THEREOF, AND REGARDLESS OF WHETHER TI HAS BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES.
You agree to fully indemnify TI and its representatives against any damages, costs, losses, and/or liabilities arising out of your noncompliance with the terms and provisions of this Notice.
This Notice applies to TI Resources. Additional terms apply to the use and purchase of certain types of materials, TI products and services.
These include; without limitation, TI’s standard terms for semiconductor products http://www.ti.com/sc/docs/stdterms.htm), evaluation
modules, and samples (http://www.ti.com/sc/docs/sampterms.htm).
Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265
Copyright © 2017, Texas Instruments Incorporated