Application Report
SNAA270 – July 2015
Using Ultrasonic Sensing to Monitor Level in Tanks
Matthew Minasi
ABSTRACT
Ultrasonic sensing Time Of Flight (TOF) measurement techniques can be utilized to measure fluid levels
in tanks. Those measurements can either be done from inside or outside the wall of the tank. In
automotive, industrial and even medical applications the ability to perform non-invasive measurements is
driven from the target fluid's corrosiveness and/or sterile requirements. This application note describes
how to utilize the TDC1000 and piezoelectric ultrasonic transducer to perform highly accurate fluid level
measurements on a tank externally.
Topic
1
2
3
4
5
...........................................................................................................................
Page
System Requirements .......................................................................................... 2
Approach ............................................................................................................ 3
Implementation .................................................................................................... 4
Conclusion ........................................................................................................ 27
Tools and Resources .......................................................................................... 27
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
1
System Requirements
1
www.ti.com
System Requirements
The requirements are:
1. Measure water level non-invasively (for corrosive or sterile fluids) in a container
2. Tank material: plastic or metalized plastic
3. Maximum height: 20cm
4. Minimum height: 2cm
5. Accuracy: 1mm
6. Frequency of measurement: 10 samples per sec
7. No moving parts
8. High reliability to environmental factors
2
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Approach
www.ti.com
2
Approach
There are several methods for noninvasive level measurement. However only ultrasonic TOF techniques
can measure tank levels from outside conductive tank walls, which is the case for metalized plastic.
Ultrasound has no moving parts and using a 1 MHz transducer will yield sub-mm accuracy. Ultrasound
also isn’t affected by external electric field changes which can be an issue with other sensing
technologies.
Ultrasonic TOF level measurement works by using a single piezoelectric transducer to create a pulse from
the bottom of a tank. That pulse travels through the tank wall, through the fluid in the tank until it reaches
the fluid surface. At the fluid surface (fluid to air interface) an echo is created. Measuring how long it takes
for the echo to return is referred to as TOF (Time Of Flight) measurement.
At the highest level an ultrasonic level measuring system consists of a signal transmitter, signal receiver
and signal transmission path. The signal that needs to be detected reliably is the ultrasonic echo that is
created at the acoustic boundary between the material level we are measuring (liquid) and the lack of it
(air etc.). The ultrasonic TOF measurement system will have 3 parts (as shown in Figure 1 ), the piezo
electric transducer, the analog front end (AFE) which interfaces between the transducer and the
microprocessor. The AFE drives the transducer and converts the analog echo signals into digital signals
that represent the beginning (START) and end (STOP) of the TOF measurement. The microprocessor
controls the analog interface, measures the time delta between the Start and Stop signals and processes
the TOF information created by the AFE into a liquid level value.
Time-of-flight =
(2 * Fluid Level)
Fluid Speed of Sound
Fluid
Fluid
Fluid
Level
Transducer
START
TX
STOP
RX
TDC1000-Q1:
Ultrasonic
AFE
TRIGGER
8MHz CLK
MSP430/
C2000
Figure 1. Ultrasonic Level Block Diagram
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
3
Implementation
3
www.ti.com
Implementation
System Design Challenges
Section 3 describes the challenges regarding architecture choices in designing an ultrasonic level system.
Those tradeoffs are: transducer selection (what size) , TDC1000 configuration (transmit and receive
configuration), high or low voltage excitation, single or resonant excitation and finally physical tank
features that can help performance.
3.1
Transducer Selection
When selecting a transducer the parameters that determine which is right for the application are: resonant
frequency, transducer diameter, and packaging.
Transducer diameter has the largest impact as it is both the transmitter and the receiver for our signal
chain. In general use the largest transducer that fits your application.
For this fluid measurement application, external to the tank, the lowest cost solution is to choose an
unsealed transducer, with a 1Mhz resonant frequency to yield easy sub-mm level resolution. Three
transducer diameters tested were 7mm, 10mm and 15mm.
They are the following:
3.2
Dia (mm)
Model Number
Website
7
BPU-P7-1000B-W200
www.bestartech.com
10
SMD10T2R111WL
https://www.steminc.com/PZT/en/piezo-ceramic-disc-10x2mm-r-215-khz-wire-leadssmd10t2r111wl
15
SMD15T21R111WL
https://www.steminc.com/PZT/en/piezo-ceramic-disc-1-mhz
Transducer Mounting
The transducers were mounted as described in App note How to Select and Mount Transducers in
Ultrasonic Sensing for Level Sensing and Fluid ID on the bottom of the tank.
4
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
3.3
Making Measurements
Measurements were taken with approximately 3cm of water in the tank. Using our test container and
TDC1000-C2000EVM the default configuration and resulting signals are described below.
3.3.1
TDC1000 Configuration
The TDC1000 platform used here is the TDC1000-C2000 EVM. To properly configure the TDC1000, an
oscilloscope is required to measure echo amplitudes (COMPIN_BUF), START and STOP pulse creation.
For our testing the setup is as follows:
TDC1000 signals connected as follows: START(TP23):(Ch1), STOP(TP22):(Ch2),
COMPIN_BUF(TP21):Ch3.
Transducer is connected to TX1/RX2 J5-pins 10 and 9.
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
5
Implementation
3.3.2
www.ti.com
Default Settings
The default settings for the TDC1000-C2000 EVM are set for the 10mm transducer in the list. A 3cm level
is a good place to start to access the performance of your transducer. The TDC1000 needs to start
measurements before the waveforms can be seen on the oscilloscope. To do this, toggle the switch on the
GUI as shown below to turn it green.
6
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
Below is an oscilloscope picture showing the TDC1000 signals once the EVM is in continuous
measurement mode.
TDC1000 signals are connected as follows: START(TP23):(Ch1), STOP(TP22):(Ch2),
COMPIN_BUF(TP21):Ch3
Note the timing relationship from Start to Stop with respect to the amplified echo on CH3. In the default
mode the TDC1000 will create a Stop pulse that lasts for the duration of the echo such that the amplitude
of the echo is greater than the programmed Voltage threshold. For more detail, see the TDC1000 data
sheet ( TDC1000: Ultrasonic Analog-Front-End ) section 8.4.4.2.
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
7
Implementation
3.3.3
www.ti.com
Transducer Ringdown
Now that a working echo is seen the next step is to understand how much the ringdown our transducer
has as it can be a limiting factor on signal detection on the low side and high side of the level
measurement. Ringdown is the mechanical “ringing” of the transducer even after the TDC1000 has
stopped exciting it. Piezoelectric transducers can be viewed as very specific RLC resonant filters and as
such once they are “excited” or stimulated via an external voltage pulse it will take a specific amount of
time to stop. With the single transducer topology (as used in level applications) the transducer is the
speaker as well as the microphone and the ringdown needs to stop or reduce in order for the transducer
to “hear” an echo. The transducers ringdown is the largest factor to defining minimum level detectable.
Setting the following configuration registers as shown will enable this to be measured.
TDC1000 tab with changed registers highlighted here.
3.3.4
Timeout Register
The specific fields changed are “Enable Short TOF and ShortTOF Blanking = 1us. These changes allow
the receive channel to observe and respond to the echo 1us after the “excitation phase” is completed. In
terms of distance, this means echos that are less than 1mm in front of the transducer. For this example
this is unusable but can be used to illustrate the behavior of the transducer.
8
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
3.3.5
Config3 Register
The specific bit field changed was “Echo_Qual_Thld” which determines the minimum voltage level for the
echo that will trigger the STOP pulse creation. Here it has been increased to -1500mv which disables Stop
pulse creation. It is useful to do this as depending on other register settings the receive circuit can be
turned off after the echo has been detected and the Stop pulse created. This can also be achieved by
setting the Config1 register “Num_rx” bit field to “No RX Event count”.
Below is a scope picture showing the TDC1000 signals.
Note the ringdown section of the waveform occurs right after excitation completes.
Now that we’ve seen “ringdown” is seen you can reduce it for any given transducer is by using mechanical
dampening techniques and/or reducing the excitation time. The result of additional dampening will be a
less efficient, less sensitive transducer that will stop “ringing” sooner. The “excitation time” for the above
example was 4us or 4 pulses from the TDC1000. See below:
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
9
Implementation
www.ti.com
We can reduce the number of pulses to shorten the ringdown but that can also reduce our transmit signal
amplitude (see section TX below). To get the shortest ringdown reducing the number of TX pulses to 1
which yields the following waveform:
To illustrate ringdown versus echo reducing the water level down to 1cm shows the relative size of the
water surface echo versus ringdown.
Another approach with the TDC1000 is to use “autozero” to electronically “blank out” a programmable time
period after the START pulse when the TDC1000 will “ignore” the echo. This method limits how close to
“tank empty scenario” that can be measured. To maximize low level sensitivity “blanking time” should be
minimized. However, the downside to this approach is as the level increases in taller tanks the echo
amplitude at max level approaches the “ringdown amplitude” and therefore makes it difficult to distinguish
between the two. This is what is referred to as a “false echo”. So it will be a tradeoff between low level
sensing and maximum level sensing.
10
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
3.3.6
Autozero in Practice
This section illustrates how to use “Autozero” to set the lowest detection level for this system. Without
changing Gain settings the following Figures 2-4 demonstrate the effect of Autozero on the echo. Note
how as the Autozero time is increased the ringdown amplitude is reduced.
In Figure 5 Vthreshold was reduced to re-enable STOP pulse creation. In Figure 6 the water level was
then reduced to illustrate the lowest level detectable. Register settings for this are shown in Figure 7.
Figure 2. Autozero @ 2us
Figure 3. Autozero @ 4us
Figure 4. Autozero @ 8us
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
11
Implementation
www.ti.com
Figure 5. VThreshold Reduced to Enable STOP Pulse
Figure 6. Water Level Reduced to Show Echo and Ringdown Mixing (Lowest Measurable Level)
12
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
Figure 7. GUI Settings for Above
NOTE: By using ShortTOF, 8us blanking and -775mV Echo_qual_thld sub-1cm level detection is
possible.
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
13
Implementation
www.ti.com
The following scope picture is from the same setup as above with only the Vthreshold lower (-410). This
configuration would yield Stop pulses that do not show the correct level. Note that the STOP pulse is
occurring during ringdown.
Figure 8. False Echos
14
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
Figure 9. False Echo GUI Settings
In the “false echo” case above, the STOP pulse circuitry triggered on the ringdown and then shutoff as the
NUM_RX was set to one. This shows the importance of choosing the correct threshold and blanking to
filter out false echoes from transducer ringdown.
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
15
Implementation
3.3.7
www.ti.com
Ringdown for Different Size Transducers for the Same Settings
The following three measurements were acquired using the exact same TDC1000 register settings using
three different sized transducers.
Figure 10. With the 15cm Transducer
Figure 11. With the 10mm Transducer
Figure 12. With a 7mm Transducer
Looking at the three scope pictures in Figs 10-12 it can be observed that as the transducer size increases
so too does the desired signal (the echo) versus the ringdown. Thus the signal to noise ratio increases
with transducer size.
16
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
3.3.8
Looking Beyond the Low End Measurement and Ringdown
ShortTOF mode was utilized in the previous section to observe transducer ringdown and how it can be
controlled. However ShortTOF mode will limit the maximum range the TOF measurement can take. A
TDC1000 with an 8Mhz input clock, in ShortTOF mode can only measure up 128us for a TOF
measurement. Level measurements NOT requiring ShortTOF mode enables the TDC1000 to measure
TOFs up to approximately 2ms.
As most level applications have these requirements this section describes the register setting changes to
enable those longer TOF measurements.
3.3.9
Setting the GUI for Non-ShortTOF Measurements
The GUI settings below are for typical level applications where a minimum level of 1.5cm up to 20cm are
required.
Note the changes are to the following registers:
• Config3: Blanking disabled: as blanking doesn’t apply to this application since the echo will need to be
detected anywhere from 1.5 to 20cm.
• TOF-1: TimingREG[9:8] = 3
• TOF-1: TimingREG[7:0] = 255 These 10 bits determine the maximum amount of time the TDC1000 will
allow for the echo to be detected before an error is set in the ERROR_FLAGs register. In this case I’ve
maximized the time to 993us as the GUI calculates.
• Timeout: ShortTOF is disabled
• Clockrate: Autozero period has been set to 16us to blank out ringdown allowing me to increase Gain
as needed to detect the echo from the maximum tank level. At this setting the earliest echo should be
at approximately 1.1cm in front of the transducer. (1480 m/s)*(16e-6s)/2 = .0118m
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
17
Implementation
3.3.10
www.ti.com
Level Detection on the High End
Thus far the focus has been on the low end of level detection. This section focuses on how to maximize
the signal to noise ratio for the TDC1000 to detect echoes on the maximum height end. The challenge so
far has been how to deal with transducer ringdown and how it affects the ability to distinguish between
real echoes and false ones. There are two ways to increase the output signal for a transducer. The
transducer can be excited “harder” by using greater excitation voltage or the transducer can be excited
resonantly. The following slides show the effects of both options for all three of the transducer sizes we’ve
looked at so far.
3.3.11
The Effect of Resonant Excitation
The slides that follow show the resultant echo with the 10mm transducer while increasing the number of
excitation pulses sequentially. The water level was increased to 7cm and all transducers' performance
were measured at the same water level.
18
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
Figure 13. 10mm Transducer with 1 Pulse
Figure 14. 10mm Transducer with 2 Pulses
Figure 15. 10mm Transducer with 3 Pulses
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
19
Implementation
www.ti.com
Figure 16. 10mm Transducer with 4 Pulses
Figure 17. 10mm Transducer with 5 Pulses
As the number of pulses increases, note the amplitude reaches a maximum at 4 pulses. Additional pulses
just extend the echo length without significantly increasing the echo amplitude.
Table 1. Excitation Pulses vs. Echo Amplitude for 10mm Transducer
20
Number of Pulses
Echo (V)
1
.97
2
1.59
3
2.0
4
2.06
5
2.09
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
Performing the same experiment with the 7mm and 15mm transducers yielded the following results.
Figure 18. 7mm Transducer: Max Signal at 5 Pulses
Table 2. Excitation Pulses vs. Echo Amplitude for 7mm Transducer
Number of Pulses
Echo (V)
1
.25
5
.5
Figure 19. Excitation Pulses vs. Echo Amplitude for 15mm Transducer
Table 3. Excitation Pulses vs. Echo Amplitude for 15mm Transducer
Number of Pulses
Echo (V)
1
2.0
2
2.28
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
21
Implementation
3.3.12
www.ti.com
The Effect of Voltage Excitation vs. Echo Amplitude
The slides that follow show the resultant echo with the 10mm transducer while increasing the excitation
voltage to 30V using the TIDA00322 TI design PCB. The TIDA00322 is functionally equivalent to
TDC1000-C2000 EVM used up to this point excepting it has a 30V boost supply and circuitry to voltage
level shift the TDC1000 5V excitation up to 30V. The transducer performance was measured at 20cm
(maximum) water level.
22
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
Figure 20. 5V Drive at 20cm Water Height
Figure 21. 30V Drive 20cm Water Height
Table 4. Excitation Pulses Voltage vs. Echo Amplitude for 10mm Transducer at 20cm
Excitation Voltage (V)
Echo (V)
5
.910
30
2.34
Why use higher voltage excitation? This question is best illustrated in the 7mm transducer example below.
The question becomes why add the cost and complexity of a boost DC-DC and level shift circuit if a user
could just increase the gain of the receive chain.
Figure 22. 7mm Transducer at 20cm Water Height at 9db Gain with 30V Excitation: Echo Amplitude:
0.81V
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
23
Implementation
www.ti.com
Figure 23. Ringdown Amplitude: ~.25Vat 9db Gain with 30V Excitation
Figure 24. 7mm Transducer at 20cm Water Height at 21db Gain with 5V Excitation: Echo Amplitude
Figure 25. Ringdown Amplitude: ~1.0V at 21db Gain with 5V Excitation
Results: If low fluid level (<35us) is required then high voltage excitation will be required. Otherwise the
user can maximize the TDC1000 gain and use blanking to filter out the ringdown echoes.
24
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Implementation
www.ti.com
3.3.13
Final System Configuration
After all of the analysis of the different size transducers and different excitation voltages, what are the best
choices of transducers and register configuration for the original system requirement?
1. ShortTOF or not? Due to the 2cm low level requirement this isn’t required. So non-ShortTOF is best.
2. Transducer size? 10mm will be chosen for cost considerations.
3. High voltage or not? At 20cm only the 10mm and 15mm transducers were capable of creating a large
enough echo with a minimum of ringdown so the low level requirement could be met. In either case the
high voltage wasn’t required. However, in some circumstances where the container is in motion
(automotive etc.) the high voltage can guarantee a return echo in the presence of surface perturbations
as the echoes may be reduced in amplitude and how frequently they can be observed. The 10mm is
cheaper than the 15mm so it was selected.
4. Register Configuration: See Figure 26:
The following are the results for the 10mm transducer at 20cm (15db, Vthreshld 410, autozero 8us, 5V
excitation)
And at 2cm Level (15db, Vthreshld 410, autozero 8us, 5V excitation)
SNAA270 – July 2015
Submit Documentation Feedback
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
25
Implementation
www.ti.com
Figure 26. Final Register Settings
3.3.14
Tank Features
There are several factors in tank construction that can affect level measurements. The single best physical
advantage to increase distance (level) measurement and reduce measurement filtering is using a level
waveguide. A level waveguide is a vertical tube within the tank with multiple inlets that allow the fluid level
in it to be the same as in the main tank. Using a waveguide aids level measurements by increasing level
height measurement capability for a given electronic system as it increases the signal to noise ratio for the
transducer. It does this by reducing the amount of signal loss due the transducer beam spreading. In
essence it confines the acoustic beam so less signal is lost due to transducer beam spreading and aids
measurements by acting as physical low pass filter for surface perturbations. Lastly the waveguide helps
reduce echo signal loss due to container tilting. If a container is not kept level the surface is no longer
parallel to the tank bottom which translates into less efficient signal transmission back to the transducer.
26
Using Ultrasonic Sensing to Monitor Fluid Level in Tanks
Copyright © 2015, Texas Instruments Incorporated
SNAA270 – July 2015
Submit Documentation Feedback
Conclusion
www.ti.com
4
Conclusion
Given a specific set of system requirements the TDC1000 can enable a variety of solutions for any given
level measurement problem due to its high level of programmability. We have demonstrated in this
application note how to measure level of liquid in a specific container. The TDC1000's high degree of
programmability enables the systems designer to build a sub-mm accurate, non-invasive, level sensing
solution that is both flexible and low cost. Ultrasonic level measurement has the added advantage to
operating through conductive (metallic) tanks.
5
Tools and Resources
Further information:
• TDC1000-TDC7200 EVM (http://www.ti.com/tool/tdc1000-tdc7200evm)
• TDC1000-C2000 EVM (http://www.ti.com/tool/tdc1000-c2000evm)
• Automotive Ultrasonic Fluid Level/Quality Measurement Reference Design TIDA-00322
(http://www.ti.com/tool/TIDA-00322)
• Application Note: How to Select and Mount Transducers in Ultrasonic Sensing for Level Sensing and
Fluid ID (http://www.ti.com/lit/an/snaa266/snaa266.pdf)
• Application Note: Ultrasonic Sensing for Fluid Identification and Contamination
(http://www.ti.com/lit/pdf/snaa265.pdf)
• Videos and other information (http://www.ti.com/ultrasonic)
Revision History
DATE
REVISION
NOTES
July 2015
*
Initial release.
SNAA270 – July 2015
Submit Documentation Feedback
Revision History
Copyright © 2015, Texas Instruments Incorporated
27
IMPORTANT NOTICE
Texas Instruments Incorporated and its subsidiaries (TI) reserve the right to make corrections, enhancements, improvements and other
changes to its semiconductor products and services per JESD46, latest issue, and to discontinue any product or service per JESD48, latest
issue. Buyers should obtain the latest relevant information before placing orders and should verify that such information is current and
complete. All semiconductor products (also referred to herein as “components”) are sold subject to TI’s terms and conditions of sale
supplied at the time of order acknowledgment.
TI warrants performance of its components to the specifications applicable at the time of sale, in accordance with the warranty in TI’s terms
and conditions of sale of semiconductor products. Testing and other quality control techniques are used to the extent TI deems necessary
to support this warranty. Except where mandated by applicable law, testing of all parameters of each component is not necessarily
performed.
TI assumes no liability for applications assistance or the design of Buyers’ products. Buyers are responsible for their products and
applications using TI components. To minimize the risks associated with Buyers’ products and applications, Buyers should provide
adequate design and operating safeguards.
TI does not warrant or represent that any license, either express or implied, is granted under any patent right, copyright, mask work right, or
other intellectual property right relating to any combination, machine, or process in which TI components or services are used. Information
published by TI regarding third-party products or services does not constitute a license to use such products or services or a warranty or
endorsement thereof. Use of such information may require a license from a third party under the patents or other intellectual property of the
third party, or a license from TI under the patents or other intellectual property of TI.
Reproduction of significant portions of TI information in TI data books or data sheets is permissible only if reproduction is without alteration
and is accompanied by all associated warranties, conditions, limitations, and notices. TI is not responsible or liable for such altered
documentation. Information of third parties may be subject to additional restrictions.
Resale of TI components or services with statements different from or beyond the parameters stated by TI for that component or service
voids all express and any implied warranties for the associated TI component or service and is an unfair and deceptive business practice.
TI is not responsible or liable for any such statements.
Buyer acknowledges and agrees that it is solely responsible for compliance with all legal, regulatory and safety-related requirements
concerning its products, and any use of TI components in its applications, notwithstanding any applications-related information or support
that may be provided by TI. Buyer represents and agrees that it has all the necessary expertise to create and implement safeguards which
anticipate dangerous consequences of failures, monitor failures and their consequences, lessen the likelihood of failures that might cause
harm and take appropriate remedial actions. Buyer will fully indemnify TI and its representatives against any damages arising out of the use
of any TI components in safety-critical applications.
In some cases, TI components may be promoted specifically to facilitate safety-related applications. With such components, TI’s goal is to
help enable customers to design and create their own end-product solutions that meet applicable functional safety standards and
requirements. Nonetheless, such components are subject to these terms.
No TI components are authorized for use in FDA Class III (or similar life-critical medical equipment) unless authorized officers of the parties
have executed a special agreement specifically governing such use.
Only those TI components which TI has specifically designated as military grade or “enhanced plastic” are designed and intended for use in
military/aerospace applications or environments. Buyer acknowledges and agrees that any military or aerospace use of TI components
which have not been so designated is solely at the Buyer's risk, and that Buyer is solely responsible for compliance with all legal and
regulatory requirements in connection with such use.
TI has specifically designated certain components as meeting ISO/TS16949 requirements, mainly for automotive use. In any case of use of
non-designated products, TI will not be responsible for any failure to meet ISO/TS16949.
Products
Applications
Audio
www.ti.com/audio
Automotive and Transportation
www.ti.com/automotive
Amplifiers
amplifier.ti.com
Communications and Telecom
www.ti.com/communications
Data Converters
dataconverter.ti.com
Computers and Peripherals
www.ti.com/computers
DLP® Products
www.dlp.com
Consumer Electronics
www.ti.com/consumer-apps
DSP
dsp.ti.com
Energy and Lighting
www.ti.com/energy
Clocks and Timers
www.ti.com/clocks
Industrial
www.ti.com/industrial
Interface
interface.ti.com
Medical
www.ti.com/medical
Logic
logic.ti.com
Security
www.ti.com/security
Power Mgmt
power.ti.com
Space, Avionics and Defense
www.ti.com/space-avionics-defense
Microcontrollers
microcontroller.ti.com
Video and Imaging
www.ti.com/video
RFID
www.ti-rfid.com
OMAP Applications Processors
www.ti.com/omap
TI E2E Community
e2e.ti.com
Wireless Connectivity
www.ti.com/wirelessconnectivity
Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265
Copyright © 2015, Texas Instruments Incorporated