Omega | BB704 | Owner Manual | Omega BB704 Owner Manual

Omega BB704 Owner Manual
TM
User’s Guide
Shop online at
omega.com
e-mail: info@omega.com
For latest product manuals:
www.omegamanual.info
BB704
Blackbody Calibrator
omega.com info@omega.com
U.S.A.
Headquarters:
Servicing North America:
Omega Engineering, Inc.
Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only)
Engineering Service: 1-800-872-9436 (USA & Canada only)
Tel: (203) 359-1660
Fax: (203) 359-7700
e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves
the right to alter specifications without notice.
BB704 Blackbody Calibrator
Table of
Contents
Table of Contents
Section ....................................................................................... Page
Section 1
Introduction .................................................................................
1.1 Precautions ..............................................................................
1.2 Safety Warnings and IEC Symbols ......................................
1.3 General Description ...............................................................
1-1
1-1
1-1
1-2
Section 2
Installation ...................................................................................
2.1 Unpacking and Inspection ....................................................
2.2 Mounting .................................................................................
2.3 Ambient Temperature ...........................................................
2.4 Power Connection ..................................................................
2-1
2-1
2-1
2-1
2-1
Section 3
Operation ......................................................................................
3.1 Front Panel Controls and Indicators ...................................
3.2 Back Panel Connections ........................................................
3.3 Changing the Temperature Setpoint ...................................
3.4 Changing the Controller Parameter Settings .....................
3.5 Heat-Up/Cool-Down Cycle Times .....................................
3.6 Thermal Reset Switch ............................................................
3-1
3-1
3-2
3-3
3-3
3-5
3-5
Section 4 Serial Communications (RS-232) ............................................. 4-1
Section 5
Maintenance .................................................................................
5.1 Calibration ..............................................................................
5.2 Cleaning ..................................................................................
5.2.1 Main Body ......................................................................
5.2.2 Target Plate ....................................................................
5.2.3 Fan ...................................................................................
5.3 Fuse Replacement ..................................................................
5-1
5-1
5-1
5-1
5-1
5-1
5-1
Section 6 Reference Probe .......................................................................... 6-1
Section 7 Specifications ............................................................................... 7-1
Section 8 Troubleshooting Guide ............................................................. 8-1
Section 9 Glossary of Terms Used in This Manual ............................... 9-1
Section 10 The OMEGA Family of Blackbody Calibrators ................. 10-1
i
Table of
Figures
BB704 Blackbody Calibrator
Table of Figures
Figure ......................................................................................... Page
1. IEC Symbols .............................................................................................. 1-2
2. Front Panel ................................................................................................. 3-1
3. Back Panel .................................................................................................. 3-2
4. Menu Hierarchy Showing Factory Default Settings ........................... 3-4
5. Programming Procedure ......................................................................... 3-4
6. Heat-Up/Cool-Down Cycle Time Table ............................................... 3-5
7. Connecting the BB704 to a Computer’s Serial Port ............................. 4-1
8. Internal Reference Probe Connections................................................... 6-1
ii
Introduction
1
Section 1 - Introduction
Your BB704 Series Blackbody Calibration Source has been designed for ease of
use and reliability whenever you have the need to test or calibrate non-contact
infrared temperature instruments. It is important that you read this manual
completely and follow all safety precautions before operating this instrument.
1.1 Precautions
• Follow all safety precautions and operating instructions outlined in this
manual.
• Never leave your calibrator unattended when in use.
• Keep out of reach of children.
• Never touch the target plate when hot.
• Never place any object within 3 inches of the target plate when hot.
• Do not operate in flammable or explosive environments.
• Never operate with a power cord other than the one provided with your unit.
• Remove and/or disconnect main power cord before attempting any
maintenance or fuse replacement.
• Do not connect and/or operate this unit to a non-grounded, non-polarized
outlet or power source.
• Do not connect the serial port or reference probe port to equipment with
exposed, hazardous, live voltages.
• Protect from moisture and rain.
• The operator of this instrument is advised that if the equipment is used
in a manner not specified in this manual, the protection provided by the
equipment may be impaired.
NOTE:
There are no user serviceable parts inside your unit. Attempting to repair or service your unit may
void your warranty.
1.2 Safety Warnings and IEC Symbols
This device is marked with international safety and hazard symbols in accordance with IEC 1010. It is important to read and follow all precautions and
instructions in this manual before operating or commissioning this device as it
contains important information relating to safety and EMC. Failure to follow all
safety precautions may result in injury and/or damage to your calibrator. Use of
this device in a manner not specified by the manufacturer may impair protection
provided within the unit.
1-1
1
Introduction
IEC symbols
Description
Caution, risk of electric shock
Caution, refer to accompanying documents
Caution, hot surface
230 VAC~ @ 50Hz (European Models)
115 VAC~ @ 60Hz (Domestic Models)
Figure 1. IEC Symbols
1.3 General Description
The Model BB704 is a portable, rugged, bench-top, hot/cold blackbody calibration source with a built-in precision PID digital controller. The calibrator is used
to test and calibrate infrared pyrometers. The large 4 inch diameter target plate
has an emissivity of 0.95 and can be set to any temperature between 100 to 398°C
(212 to 750°F).
1-2
Installation
2
Section 2 - Installation
2.1 Unpacking and Inspection
Remove the packing list and verify that you have received all your equipment.
If you have any questions about the shipment, please call our Customer Service
Department at:
1-800-622-2378 or 203-359-1660.
We can also be reached on the Internet at:
omega.com
e-mail: info@omega.com
When you receive the shipment, inspect the container and equipment for any
signs of damage. Note any evidence of rough handling in transit. Immediately
report any damage to the shipping agent.
NOTE:
The carrier will not honor any damage claims unless all shipping material is saved for inspection.
After examining and removing contents, save packing material and carton in the event reshipment
is necessary.
The following items are supplied in the box:
• BB704 Blackbody Calibration Source
• User’s Manual
• Calibration Certificate
• Power Cord
• Reference Port Cord Plug Connector
2.2 Mounting
Mount the unit on a bench, table-top, or shelf in a horizontal position and
operate at least 10 inches from any air obstructions to the fan, front panel, rear
panel, bottom, and top of the unit, in an ambient environment between the
specified 0 to 50°C (32 to 122°F).
2.3 Ambient Temperature
The target plate of the BB704 can achieve any temperature within the specified
temperature range when being operated in normal ambient temperature
environments. The maximum specified target plate temperature of 398°C
(750°F) can be achieved over the entire specified ambient temperature range.
2.4 Power Connection
Standard (115 VAC~, 50/60 Hz models)
The BB704 comes with a standard North American 3-prong AC power cord. Do
not use any other power cord other than the one provided. This cord provides
the proper grounding and has been safety tested by the proper safety agencies.
2-1
2
Installation
International (230 VAC~, 50/60 Hz models)
On 230 VAC~, 50/60Hz models, a European style power cord with the proper
color code and approvals is provided with stripped wire ends for connection to
the proper connector used in your country or local area. This connector is not
provided.
CAUTION:
• Line voltage variations are not to exceed ±10% of the rated input voltage.
• Electrical connections and wiring should be performed only by suitably trained personnel.
• To shut down the unit in an emergency, the power cord can be disconnected from the AC
outlet or rear of the unit.
2-2
Operation
3
Section 3 - Operation
3.1 Front Panel Controls and Indicators
SETPOINT TEMPERATURE
PROCESS TEMPERATURE
TARGET PLATE
U
O
TI
N:
TA
RG ET
SU RFA C E M
A
YB
E
HO
TM
BB704
T
CA
BLACK POINT
Blackbody
Calibrator
750.0 PV
750.0 SV
CA
UT
T
ION
HO
: TA
BE
RGET
AY
SURFACE M
Temperatur
e
ON/STANDBY
POWER SWITCH
PARAMETER/
ACCESS KEY
100C/212F
MODE KEY
LOWER SETPOINT KEY
399C/750F
TEMPERATURE RANGE
REFERENCE BAR
RAISE SETPOINT KEY
Figure 2. Front Panel
Process Temperature:
This field displays the current temperature of the target plate.
Setpoint Temperature:
This field displays the desired target plate temperature. Once the target plate
reaches this desired temperature, both displays will read the same value.
Target Plate:
The 4" target plate is a near ideal blackbody source. The emissivity of the plate
is 0.95. When calibrating an IR pyrometer, hold the pyrometer perpendicular to
the target plate for optimal performance. The proper distance between the IR
pyrometer and the target plate depends on the field of view of the pyrometer.
If the pyrometer is too far away, it will scan unwanted surfaces outside of the
perimeter of the target plate. Holding the pyrometer too close could introduce
undesirable heat to the IR detector of the pyrometer.
CAUTION:
• The BB704’s target plate can be set to very high temperatures. Exercise extreme caution when operating the unit. Keep hands and fingers away from the target plate area.
Keep flammable products such as paper, plastics, and clothing far from the BB704.
• The BB704 is a Class II instrument. It is intended to be operated in a laboratory
environment only. Never operate the unit outdoors or around children.
• Nothing should come in contact with the target plate, even when the unit is off.
• Never unplug the unit while it is on or during “cool down.”
3-1
3
Operation
NOTE:
PID Control:
Proportional, integral, derivative control (PID) is a temperature control algorithm used in high-end
temperature controllers. The controller causes the process to attain the desired temperature by turning the process on or off. The process may be a heater or refrigerator. As the process temperature
approaches the setpoint temperature, the hot or cold process will be pulsed to reduce the corrective measures and minimize overshooting. The controller provides a visual representation of the
process status through LED indicators. An indicator may light continuously, blink, or shut off entirely
to indicate that the process is on, being pulsed, or off, respectively.
Parameter/Access Key:
Press to scroll through menu parameters.
Raise Key:
Press to increase the selected parameter or scroll upward in the list of possible
settings.
Lower Key:
Press to decrease the selected parameter or scroll downward in the list of possible settings.
Mode Key:
Press to save settings and exit a menu level.
3.2 Back Panel Connections
RTD REFERENCE
PROBE CONNECTION
FUSE
FAN
RS-232 COMMUNICATIONS PORT
Figure 3. Back Panel
3-2
AC MAINS POWER INPUT
AND FUSE COMPARTMENT
Operation
3
AC Power Mains Input and Fuse Compartment:
The customer connects the power cord to the AC Power Input. As a safety precaution, the power cord cannot be connected if the fuse compartment is open.
Refer to Section 5.3 for information on fuse replacement.
Reference Probe Connection:
The reference probe enables the user to monitor the target plate temperature
with an external instrument. The wires are connected to a Class A, 0.00385 RTD
Sensor.
Refer to Section 6 for pinout details.
RS-232 Communications Port:
The female DB-9 port allows the customer to make a 3-wire RS-232 interface
with the BB704. A detailed description of this port is in Section 4.
3.3 Changing the Temperature Setpoint
The layout of the front panel is shown in Figure 2. The BB704 incorporates a PID
digital setpoint controller. The upper display indicates the blackbody target plate
temperature known as (PV) Process Variable, while the lower display indicates
the programmed setpoint known as (SV) Setpoint Variable. Changes to the
setpoint, units of measure and communication settings are made via the
and
keys. Holding a key in continuously will cause the setpoint temperature to advance more quickly to a desired value. Three scanning speeds are
provided: slow, medium, and fast. The minimum and maximum setpoints are
locked.
3.4 Changing the Controller Parameter Settings
The BB704 operates at optimum performance when left with its factory
parameter settings. The only internal parameters that the operator should need
to change are the engineering units (°F or °C) or the serial communications
parameters. See next page for menu hierarchy with factory default settings
(Fig. 4) and programming procedure (Fig. 5).
3-3
3
Operation
Menu 00
Key Lock
Menu 01
SETPOINT
Menu 02
Ac.Cd = 02
Menu 03
Ac.Cd = 03
Menu 04
Ac.Cd = 04
Menu 05
Ac.Cd = 05
Ac.Cd
Gn.o1
64
ALr1
775
id.no
01
SnSr
775
05
BAUd
CAL.L
CAL.H
12.o.7
29
146
ALr2
Cy.t1
Cy.t2
Sn.00
dEC.P
FILt
H.Hys
HyS.1
SP.tt
L.SP.L
OFF
212
OUt.1
OUt.2
Ht.P
Alr
C.HyS
HyS.2
C.SPr
L.SCL
U.SP.L
H.SCL
750
CoL.t
A1.HL
A1.Pd
nor
HI
Pr
A1.OP
A2.HL
LAt
Hl
A2.Pd
A2.OP
Unit
Pr
LAt
F
Gr.o2
rAtE
rSEt
SPr.2
dPnG
nl
P
Figure 4. Menu Hierarchy Showing Factory Default Settings
NOTE:
Only the boldface parameters are active for the default mode of operation.
Changing the Controller’s Parameter Settings:
key to enter the programming mode. The lower display will
1. Press the
alternately display the menu level and “Ac.Cd.”
2. Use the
and
keys to change to the desired menu level.
key to scroll through the
3. Once you have chosen the desired menu use the
parameters. To change the setting of a given parameter, use the
and
keys.
4. To save settings press the
key. The controller now exits the programming
menu and returns to the normal operating mode.
5. To change settings on other menu levels, you must re-enter the programming
menu (from step #1).
Putting the Controller In or Out of Standby Mode:
1. To enter the Standby mode hold the
flashes “StbY.”
key for 4 seconds until the window
2. To exit the Standby mode hold the
key for 4 seconds until the window
flashes “tUne.” Then press the
key for 4 seconds again until “tUne” stops flashing.
Figure 5. Programming Procedure
3-4
Operation
3
NOTE:
In the event that the unit has been operated outside of operating range, the controller alarm will
disable the heater and the target plate temperature will not rise. Let the unit cool down completely.
Turn the power off and then back on. If the problem is not corrected, contact the factory.
3.5 Heat-Up/Cool-Down Cycle Times
Approximate cycle times for heat up and cool down are given in the table above.
To find a given transition time from an initial temperature to a second target
temperature, follow this procedure: Look for an initial temperature in the left
column. Next, look for the target temperature along the top row. The intersection of the row and column provides the approximate transition time.
FROM:
TO:
121°C
260°C
398°C
121°C
260°C
398°C
15 MIN. 25 MIN. 250°F
25 MIN.
18 MIN. 500°F
32 MIN. 10 MIN.
750°F
250°F
500°F
750°F FROM:TO:
Figure 6. Heat-Up/Cool-Down Cycle Time Table
3.6 Thermal Reset Switch
In the event that the target plate temperature and/or ambient temperature
exceeds the acceptable limit, the thermal reset switch will pop open. The controller will appear to be demanding a temperature increase but the target plate
temperature will not rise. Let the unit cool down completely. Press the switch
located on the right side of the unit near the handle hinge until a click is heard. If
this has no effect, refer to the troubleshooting section.
If the reset switch keeps resetting, call Omega’s Customer Service Department at
(203) 359-1660.
3-5
3
Operation
NOTES:
3-6
Serial Communications (RS-232)
4
Section 4 - Serial Communications (RS-232)
This method allows bi-directional data transfer via a three-conductor cable consisting of signal ground, receive input, and transmit output. It is recommended
that less than 50 feet of shielded cable be used between the computer and this
instrument. This will assure performance of the BB704 to EN61326 standards,
under the EMC. Note that multiple instruments cannot be tied to the same port
in this configuration. The RS-232 port is optically isolated to eliminate ground
loop problems.
Below is a pinout diagram for the serial port of the BB704 as well as the pinout
for a 9-pin PC serial port. Use a straight DB9 (female) to DB9 (male) connector
cable to connect your computer to the BB704. The cable should be attached only
when the computer and BB704 are off.
Only parameters in the parameter list should be modified or queried via the
serial port. Other parameters should be viewed or queried from the controller
directly. It is highly recommended that baud rate for the controller be modified
on the controller directly. Note that both the BB704 and the computer must be
communicating with the same serial communications parameters to establish a
working communication link.
The serial communications feature can be tested using a terminal emulation
package. Note that this controller does not time-out waiting for the next character to be transmitted. Be sure not to use the XON/XOFF or hardware handshaking. Lastly, it should be noted that following a complete transmission to the
BB704, a response is sent back. If the message was valid, the changed or queried
parameter is echoed back (following the same format). If the message was not
according to acceptable format or was attempting to force a parameter out of
range, an “ERROR” message is echoed.
Figure 7. Connecting the BB704 to a Computer’s Serial Port
BB704
BLACKBODY
CALIBRATOR
COMPUTER
SERIAL
PORT
(RS-232)
4-1
4
Serial Communication (RS-232)
Parameter List (only relevant parameters shown):
PAR#:
Parameter:
Range/Units:
00
Process Temp.
Input determined
01
Setpoint
Input determined
19
Baud Selection
bAUd
Baud Selections:
Code:
Baud:
Parity:
Data Bits:
Stop Bits:
3.o.7
300
odd
7
2
6.o.7
600
odd
7
2
12.o.7
1200
odd
7
2
(factory default settings)
24.o.7
1200
odd
7
2
3.n.8
300
no
8
1
6.n.8
600
no
8
1
12.n.8
1200
no
8
1
24.n.8
2400
no
8
1
General Message Format:
#[controller id][command][parameter number]<new value><units>[CR/LF]
Definitions:
# – This character initiates an “escape sequence” that the controller will recognize.
[controller id] – Up to 2 numeric characters, 00 to 99 (factory default = 01)
[command] – 1 character, upper case or lower case
R – To read a parameter from the controller
M – To temporarily modify a controller parameter (lost upon shutdown)
E – To modify a controller parameter in non-volatile memory (saved even
after shutdown)
[parameter #] – Up to 2 numeric characters, 00 to 99
<new value> – This control word is used only when entering or modifying a
parameter. Up to 6 characters may be entered. The first character
can be a space, a “+”, or a “-”. The next 4 characters are for entering the new value parameter value. Be sure to use the exact same
field format as is currently being used (i.e., if the XXX.X format
is used to express temperature, be sure to enter a new value that
conforms to the same format).
<units> – This optional control word is used to specify units, F for °F, C for °C.
[CR/LF] – Every transmission must be terminated with a carriage return [CR]
character. The line feed [LF] character is optional.
4-2
Maintenance
5
Section 5 - Maintenance
5.1 Calibration
This unit has been fine-tuned at the factory and calibrated to give optimum
performance at its full temperature range. It is recommended that the unit be
returned annually for re-calibration.
5.2 Cleaning
CAUTION:
Remove all electrical connections and power before attempting any maintenance or cleaning.
5.2.1 Main Body
Only a damp, soft rag with a mild cleaning solution should be used when cleaning the main body of this unit.
5.2.2 Target Plate
Do not attempt to clean the target plate. The target plate has a special coating applied and cleaning may change the emissivity and performance of your unit.
5.2.3 Fan
The fan filter should be cleaned at least monthly by washing the filter with warm
water and then blowing dry with air. It can be removed by firmly pulling the
black plastic frame outward. The internal protective grill that is seated against
the fan can be cleaned with a soft bristle brush.
5.3 Fuse Replacement
WARNING:
Disconnect all power from source before attempting fuse replacement.
CAUTION:
For continued protection against the risk of fire, replace with only the same size, type, and fuse rating indicated here and on the rear panel of your unit.
For model BB704 use:1 ea. 250 VAC~, T5A (Time-Lag, 5 Amp) UL/CSA APPROVED (0.25” dia. x 1.25” long).
For model BB704-230VAC use:2 ea. 250 VAC~, F3.5A (Fast-Acting, 3.15 Amp)
VDE APPROVED (5 mm dia. x 20 mm long).
5-1
6
Reference Probe
Section 6 - Reference Probe
An RTD probe has been embedded into the target plate heater assembly to be
used as a reference to check the calibrator’s calibration and accuracy. A connector has been provided with your calibrator for connection to this reference probe.
Listed below are a few calibration points and the resistance readings you can
expect from the reference probe.
Target Plate Temperature
Reference Probe Resistance (Ohms)
100°C (212°F)
138.51
149°C (300°F)
156.96
204°C (400°F)
177.33
260°C (500°F)
197.71
316°C (600°F)
217.73
371°C (700°F)
237.05
399°C (750°F)
246.71
TARGET PLATE ASSEMBLY
BLACK WIRE
RED WIRE
1
2
3
BLACK WIRE
RTD REFERENCE PROBE
Figure 8. Internal Reference Probe Connections
6-1
REFERENCE
PROBE
CONNECTOR
Specifications
7
Section 7 - Specifications
Target Plate Temperature Range:
100 to 398°C (212 to 750°F)
Overall Accuracy:
±0.8°C (±1.4°F)
Display Accuracy:
±0.5°C (±1.0°F)
Stability:
±0.1°C or less (±0.2°F or less)
Ambient Environmental Conditions
Temperature:
0 to 50°C (32 to 122°F )
Humidity:
0 to 90% RH, Non-condensing
Target Plate Size:
101.6 mm (4")
Cavity Emissivity:
0.95*
Internal Control Sensor:
RTD, Class A, Alpha = 0.00385
Reference Sensor:
Accuracy:
RTD, Class A, Alpha = 0.00385
±1.8°F (±1.0°C)
Warm-up Time:
See Table, Page 3-5
Cool-down Time:
See Table, Page 3-5
Power
BB704:
BB704-230 VAC:
115 VAC~, 50/60 Hz, 425W
230 VAC~, 50/60 Hz, 425W
Dimensions:
190 mm H x 410 mm W x 264 mm L
(7.5 H x 16.1 W x 10.4" L)
Weight:
7.2kg (16 lbs)
Installation Category II
* Reference to 8-14 microns wavelength bandwidth.
7-1
8
Troubleshooting Guide
Section 8 - Troubleshooting Guide
Problem
Solution
1. Unit will not turn on.
a. Check all electrical connections.
b. Check rear panel fuses.
c. If previous steps don’t solve the problem,
contact our customer service department.
2. Unit turns on but the target
a. Check that you have entered a setpoint
plate will not get hot. above the ambient temperature.
b. Verify that the controller is set to its factory default settings.
c. Unit may have been operated outside of
operating range. The controller alarm
may have disabled the heater. Let the
unit cool down completely. Turn the
power off and back on. If the problem is
not corrected, contact the factory.
d. Unit may have been operated outside
of operating range. Safety cutoff switch
may have disabled heater. Let the unit
cool down completely. Press the reset
switch located on the left side of the
unit, behind the carry handle, until a
click is heard. If the problem is not corrected, contact the factory.
e. If previous steps don’t solve the problem,
contact our customer service department.
3. Controller display shows “Error” a. Unit requires service; contact our
and the target plate will not get customer service department.
hot or cold.
4. Target plate temperature will not a. Verify that the controller is set to its
stabilize to within ±0.5°F of the factory default settings.
setpoint temperature.
b. If previous steps don’t solve the problem,
contact our customer service department.
5. Controller has no output.
a. Controller may be in STANDBY mode.
Pressing buttons has no effect. Hold down the
key continuously
TUNE blinks on the display. until “TUNE" stops blinking.
6. Controller has no output.
a. Controller may be in LOCK mode.
Controller does not respond to Hold down the
key continuously
any key presses.until “Ac.Cd” is displayed. Change to a
menu level other than level 1 and exit the
menu selection mode by pressing the
key once.
8-1
Glossary of Terms Used in This Manual
9
Section 9 - Glossary of Terms Used in This Manual
Blackbody
A theoretical object that radiates the maximum amount of energy at a given temperature, and absorbs all the energy incident upon it.
Calibration
The process of adjusting an instrument or compiling a deviation chart so that its
reading can be correlated to the actual value being measured.
Emissivity
The ratio of energy emitted by a surface to the energy emitted by a blackbody at
the same temperature.
IEC
International Electrotechnical Commission.
Infrared (IR)
A range of the electromagnetic spectrum extending beyond red visible light from
760 nanometers to 1000 microns.
PID
Proportional, Integral, Derivative. A three-mode control action where the controller has time proportioning, integral (auto reset), and derivative rate action.
RTD
Resistance temperature detector.
9-1
10
The OMEGA Family of Blackbody Calibrators
The OMEGA Family of Blackbody Calibrators
Listed below is a selection guide to OMEGA’s current line of blackbody calibration sources including the one you have selected. This family of rugged, portable, and accurate
calibrators covers a wide range of temperatures, target plate sizes, and features, making
them perfect for infrared pyrometer field service testing and laboratory calibrations.
BB701 Hot/Cold Blackbody Calibration Source
Calibration Range: -18 to 149°C (0 to 300°F)
Emissivity: 0.95
Cavity Size: 63.5 mm (2.5 in.)
Accuracy: ±0.8°C (±1.4 °F)
Ambient Temp.: 4 to 43°C (40 to 110°F)
Power: 115/230V, 50/60 Hz, 175W
BB702 Blackbody Calibration Source
Calibration Range: 32 to 215°C (amb. 90 to 420°F)
Emissivity: 0.95
Cavity Size: 63.5 mm (2.5 in.)
Accuracy: ± 0.5°C (±0.9°F), ±0.25% rdg
Ambient Temp.: 5 to 45°C (41 to 113°F)
Power: 115/230V, 50/60 Hz, 175W
BB703 Mini Blackbody Calibration Source
Calibration Range: 32 to 400°C (90 to 752°F)
Emissivity: 0.95
Cavity Size: 28.6 mm (1.125 in.)
Accuracy: ±1.4°C (±2.5°F)
Ambient Temp.: 0 to 40°C (32 to 104°F)
Power: 115/230V, 50/60 Hz, 175W
BB704 4” Target Plate Blackbody Calibration Source
Calibration Range: 100 to 398°C (212 to 750°F)
Emissivity: 0.95
Cavity Size: 101.6 mm (4 in.)
Accuracy: ±0.8°C (±1.4°F)
Ambient Temp.: 0 to 50°C (32 to 122°F)
Power: 115/230V, 50/60 Hz, 425W
BB705 Laboratory Grade Blackbody Calibration Source
Calibration Range: 100 to 1046°C (212 to 1915°F)
Accuracy: ±1.0°C (±1.8°F), ±0.25% rdg
Emissivity: 0.99
Ambient Temp.: 0 to 35°C (32 to 95°F)
Cavity Size: 44 mm (1.75 in.)
Power: 115/230V, 50/60 Hz, 1100W
BB-4A High Temperature Blackbody Calibration Source
Calibration Range: 100 to 982°C (212 to 1800°F)
Accuracy: ±1.0°C (±1.8°F), ±0.25% rdg
Emissivity: 0.99
Ambient Temp.: 0 to 50°C (32 to 122°F)
Cavity Size: 22.2 mm (0.88 in.)
Power: 115/230V, 50/60 Hz, 400W
For a complete, updated specification sheet and price on any of the calibrators listed here, visit our
web site at omega.com. Please call our sales or customer service department for information and
pricing on any new models available.
9-2
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a
period of 25 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace
period to the normal two (2) year product warranty to cover handling and shipping time. This ensures
that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser,
including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result of excessive corrosion;
or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However,
OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for
any damages that result from the use of its products in accordance with information provided
by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of
OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical
applications or used on humans. Should any Product(s) be used in or with any nuclear installation or
activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility
as set forth in our basic WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)
NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the outside of the return package and on any
correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR NON-WARRANTY REPAIRS, consult
FOR WARRANTY RETURNS, please have the
OMEGA for current repair charges. Have
following information available BEFORE contacting
the following information available BEFORE
OMEGA:
contacting OMEGA:
1.Purchase Order number under which the product
1. Purchase Order number to cover the COST
was PURCHASED,
of the repair,
2.Model and serial number of the product under
2. Model and serial number of the product, and
warranty, and
3. Repair instructions and/or specific problems
3. Repair instructions and/or specific problems
relative to the product.
relative to the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC.
© Copyright 2017 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior
written consent of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for
Process Measurement and Control?
OMEGA…Of Course!
Shop online at omega.com
TEMPERATURE
M
U Thermocouple, RTD & Thermistor Probes, Connectors,
Panels & Assemblies
M
U Wire: Thermocouple, RTD & Thermistor
M
U Calibrators & Ice Point References
M
U Recorders, Controllers & Process Monitors
M
U Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
M
U Transducers & Strain Gages
M
U Load Cells & Pressure Gages
M
U Displacement Transducers
M
U Instrumentation & Accessories
FLOW/LEVEL
M
U Rotameters, Gas Mass Flowmeters & Flow Computers
M
U Air Velocity Indicators
M
U Turbine/Paddlewheel Systems
M
U Totalizers & Batch Controllers
pH/CONDUCTIVITY
M
U pH Electrodes, Testers & Accessories
M
U Benchtop/Laboratory Meters
M
U Controllers, Calibrators, Simulators & Pumps
M
U Industrial pH & Conductivity Equipment
DATA ACQUISITION
M
U Communications-Based Acquisition Systems
M
U Data Logging Systems
M
U Wireless Sensors, Transmitters, & Receivers
M
U Signal Conditioners
M
U Data Acquisition Software
HEATERS
M
U Heating Cable
M
U Cartridge & Strip Heaters
M
U Immersion & Band Heaters
M
U Flexible Heaters
M
U Laboratory Heaters
ENVIRONMENTAL
MONITORING AND CONTROL
M
U Metering & Control Instrumentation
M
U Refractometers
M
U Pumps & Tubing
M
U Air, Soil & Water Monitors
M
U Industrial Water & Wastewater Treatment
M
U pH, Conductivity & Dissolved Oxygen Instruments
M3016/1217
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising