Omega | DPF700 | Owner Manual | Omega DPF700 Owner Manual

Omega DPF700 Owner Manual
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects
in materials and workmanship for a period of 61 months from date
of purchase. OMEGA’s WARRANTY adds an additional one (1) month
grace period to the normal five (5) year product warranty to cover
handling and shipping time. This ensures that OMEGA’s customers
receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation.
OMEGA’s Customer Service Department will issue an Authorized
Return (AR) number immediately upon phone or written request. Upon
examination by OMEGA, if the unit is found to be defective, it will be
repaired or replaced at no charge. OMEGA’s WARRANTY does not
apply to defects resulting from any action of the purchaser, including
but not limited to mishandling, improper interfacing, operation outside
of design limits, improper repair, or unauthorized modification. This
WARRANTY is VOID if the unit shows evidence of having been tampered
with or shows evidence of having been damaged as a result of excessive
corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of
OMEGA’s control. Components in which wear is not warranted, include
but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various
products. However, OMEGA neither assumes responsibility for
any omissions or errors nor assumes liability for any damages
that result from the use if its products in accordance with
information provided by OMEGA, either verbal or written.
OMEGA warrants only that the parts manufactured by the
company will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT
THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING
ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION
OF LIABILITY: The remedies of purchaser set forth herein are
exclusive, and the total liability of OMEGA with respect to
this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the
purchase price of the component upon which liability is based.
In no event shall OMEGA be liable for consequential, incidental
or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used,
nor shall it be used: (1) as a “Basic Component” under 10 CFR 21
(NRC), used in or with any nuclear installation or activity; or (2) in
medical applications or used on humans. Should any Product(s) be
used in or with any nuclear installation or activity, medical application,
used on humans, or misused in any way, OMEGA assumes no
responsibility as set forth in our basic WARRANTY/DISCLAIMER
language, and, additionally, purchaser will indemnify OMEGA and hold
OMEGA harmless from any liability or damage whatsoever arising out
of the use of the Product(s) in such a manner.
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
5
6
For complete product manual:
www.omega.com/manuals/manualpdf/M1676.pdf
To Change
To Change
Scale Factor
ScaleValue
Factor Value
To Change
To the
Change
Decimal
the Decimal
Point Position:
Point Position:
1.
1.
Press1.MENU
Press
untilMENU
the
until the
display shows
display
SCALE
shows
. SCALE . sCale
RECALL
2.
/SET
Press
ADV to show
to show
Press4. ADV
current
current scale
factorscale factor
value.
value.
RECALL
5.
/SET
/ADV
Press
SET to select
Press3. SET
to select
the setting
the setting from
IN/SC from IN/SC I n
or IN .SC . or IN .SC .
RECALL
4.
/ADV
MENU
RECALL
RESET /SET
Press2. SET
Press
SET to show
to show
the current mathmatical In/SC
the current mathmatical
operation. operation.
RECALL
3.
/SET
/ADV
MENU
MENU
RECALL
RESET /SET
/ADV
RESET
2.
In/SC
MENU
RESET
3.
. C In S
S
. C
MENU
RECALL
RESET /SET
60
. 00
/ADV
MENU
RESET
/ADV
MENU
RECALL
RESET /SET
Press1.MENU
Press
untilMENU
the until the
display shows
display
dECPt
shows
. dECPt .deCpt
RECALL
/SET
/ADV
MENU
RECALL
RESET /SET
/ADV
MENU
RECALL
RESET /SET
/ADV
MENU
TM
RESET
SET to show
Press2. SET
Press
to show
. F F F FF F
. FFFF
current
the currentthe
location
of location ofF F
decimal point.
the decimalthe
point.
RECALL
/SET
/ADV
MENU
RECALL
RESET /SET
/ADV
MENU
RESET
Press
Press3. SET
to SET to
. F F FF f F
. FFF
move the decimal F f F
move the decimal
point to thepoint
right.to the right.
Continue toContinue
press to press
SET until the
meter
SET
until the meter
displays AutO
for AutO for
displays
auto ranging.
auto ranging.
RECALL
/ADV
MENU
RESET
/SET
/ADV
MENU
RECALL
RESET /SET
Press4.MENU
to store
Press
MENU to store
the decimalthe
point
into point into store
decimal
volatile memory.
volatile memory.
RECALL
6. 0 00
/SET
deCpt
/ADV
MENU
RESET
60
. 00
4.
/SET
to select
Press
ADV to select
Press5. ADV
the digit youthe
want
toyou want to
digit
change. change.
RECALL
/ADV
sCale
/SET
/ADV
DPF700
store
MENU
RECALL
RESET /SET
/ADV
MENU
RESET
6. 0 00
/ADV
MENU
Low-Cost Ratemeter/
Totalizer/Controller
RESET
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer
Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA,
PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER
FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO
AVOID PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence.
FOR WARRANTY RETURNS,
please have the following
information available BEFORE
contacting OMEGA:
1. Purchase Order number under
which the product was
PURCHASED,
2. Model and serial number of
the product under warranty, and
3. Repair instructions and/or
specific problems relative to
the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair
charges. Have the following
information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover
the COST of the repair or
calibration,
2. Model and serial number of the
product, and
3.Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes,
whenever an improvement is possible. This affords our customers the
latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC.
© Copyright 2018 OMEGA ENGINEERING, INC. All rights reserved.
This document may not be copied, photocopied, reproduced, translated, or
reduced to any electronic medium or machine-readable form, in whole or in
part, without the prior written consent of OMEGA ENGINEERING, INC.
MQS1766/0218
6.
Press6. SET
Press
SET to
to
increase theincrease
value ofthe
thevalue of the 6. 2 00
flashing digit.
flashing digit.
RECALL
7.
/SET
/ADV
MENU
RECALL
RESET /SET
6. 2 00It is the policy ofIt OMEGA
is the policy
to comply
of OMEGA
with all
to worldwide
comply with
safety
all worldwide
and EMC/EMI
safety and EMC/EMI
regulations thatregulations
apply. OMEGA
that apply.
is constantly
OMEGApursuing
is constantly
certification
pursuing
of certification
its
of its
products to the European
products toNew
the Approach
EuropeanDirectives.
New Approach Directives.
/ADV
MENU
This device is marked
This device
with the
is marked
international
with the
caution
international
symbol. caution
It is
symbol. It is
important to read important
the Setup to
Guide
read before
the Setup
installing
Guideorbefore
commissioning
installing or
this
commissioning this
device as it contains
device
important
as it contains
information
important
relating
information
to safety and
relating
EMC.
to safety and EMC.
Press
MENU to store
Press7.MENU
to store
thevolatile
value into volatile
the value into
memory. memory.
RECALL
RESET
WARNING: These
WARNING:
products are
These
not designed
products for
areuse
not in,
designed
and should
for use
not in,
beand
used
should
for, not be used for,
patient connected applications.
patient connected applications.
/SET
/ADV
MENU
RECALL
RESET /SET
/ADV
MENU
The information contained
The information
in this contained
document in
is this
believed
document
to be is
correct
believed
but to
OMEGA
be correct but OMEGA
Engineering, Inc. Engineering,
accepts no liability
Inc. accepts
for any no
errors
liability
it contains,
for any errors
and reserves
it contains,
the right
and reserves the right
toRESET
alter specifications
to alter
without
specifications
notice.
without notice.
omega.com info@omega.com
Servicing North America:
U.S.A.
Omega Engineering, Inc.
Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only)
The information contained in this document is believed to be
correct, but OMEGA accepts no liability for any errors it contains,
and reserves the right to alter specifications without notice.
Customer Service: 1-800-622-2378
(USA & Canada only)
Engineering Service: 1-800-872-9436
(USA & Canada only)
Tel: (203) 359-1660
Fax: (203) 359-7700
e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
START HERE
2
3
4
Using this Quick Reference
Use this Quick Reference with your controller for information
on wiring and to make changes to the operating mode, scale
factor, and decimal point position. For detailed instructions,
refer to the appropriate section in the Reference Manual.
Safety Consideration
Wiring Examples
TB2
EXC
12.4 V
OUTPUT
RTN
This device is marked with the international “Caution, risk of
danger” symbol.
SIG IN
Menu
Prompt
Menu
Item
Prompt is starting
point:
3
SIG RTN
4
Func
RST IN
Function
5
EXT GATE
to select rate or totalize
mode.
6
BATTERY
SCALE
Scale Factor
to select scale factor
as a divider or multiplier
and to enter scale
factor value.
Offset Value
to select offset value.
Reading on meter =
input x scale + offset
dEC Pt
Decimal Point
to select position of
decimal point in the
digital display.
SP LO
SP HI
Setpoints
Low and
High
to enter setpoints.
Shows only if
the relay option
board is installed.
COnFIG
Configuration
to program internal
software switches to
specify analog,
output, RS-232 input
or alarm conditions
An LO
Analog Low
to obtain reading for
low analog output.
An HI
Analog High
to obtain reading for
high analog output.
EXCITATION = 12.4V
POSITIVE THRESHOLD
1K PULL DOWN RESISTOR TO RTN
243K PULL UP RESISTOR TO +EXC
OFFSEt
EXC OUT
2
SIG IN
3
SIG RTN
4
RST IN
5
EXT GATE
6
BATTERY
12.4V EXC
1
2
3
4
5
6
7
8
9
10
OPEN
HIGH LEVEL SIGNAL
POSITIVE THRESHOLD
LOW PASS FILTER
1K PULL DOWN RESISTOR
243K PULL UP RESISTOR TO +EXC
OPEN
OPEN
OPEN
CLOSED
CLOSED
CLOSED
OPEN
OPEN
OPEN
OPEN
2-Wire Contact
Closure
S1
TB2
1
EXC OUT
2
SIG IN
3
SIG RTN
4
5
EXT GATE
6
BATTERY
EXCITATION= 5.0V
243K PULL-UP RESISTOR TO +EXC
1
2
3
4
5
6
7
8
9
10
OPEN
LOW LEVEL SIGNAL
ZERO THRESHOLD
NO FILTER
NoStor
No Store
to save to the volatile
(active) memory
StorE
Store
to save to the nonvolatile (permanent)
memory.
RST IN
CLOSED
Time-delay, UL 248-14 listed
125 mA (115 Vac line)
63 mA (230 Vac line)
Time-lag, IEC 127-2 recognized
100 mA (115 Vac line)
AC Powered Unit Connections
50 mA (230 Vac line)
A S
. C
A/SC
TB2
1
Mount the Unit
External Fuse Required:
OPEN
OPEN
OPEN
CLOSED
OPEN
CLOSED
OPEN
OPEN
OPEN
OPEN
3-Wire Sensor
Open Collector
(PNP)
S1
CAUTION: Do not connect AC power to your device until
you have completed all input and output connections. This
device must only be installed by a specially trained
electrician with corresponding qualifications. Failure to
follow all instructions and warnings may result in injury!
1. Cut a panel opening using the dimensions shown to the right.
2. Position the unit in the opening, making sure the front bezel is
flush with the panel.
3. Install retaining clips on both sides of the meter and tighten
against the panel.
1
2
3
4
5
6
7
8
9
10
OPEN
1. Remove the panel at the back
of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each
terminal as shown in the
following figure and tighten the lockdown screws.
4. Tug gently on the wires to verify the connections.
EXC OUT
2
CLOSED
Wiring
1
CLOSED
The instrument is a panel mount device protected in accordance
with EN61010-1 (Safety requirements for electrical equipment
for measurement, control, and laboratory standard). Remember
that the unit has no power-on switch. Building installation should
include a switch or circuit-breaker that must be compliant to IEC
947-1 and 947-3.
SAFETY:
• Do not exceed voltage rating on the label located on the top
of the instrument housing.
• Always disconnect power before changing signal and power
connections.
• Do not use this instrument on a work bench without its case
for safety reasons.
• Do not operate this instrument in flammable or explosive
atmospheres.
• Do not expose this instrument to rain or moisture.
EMC:
• Whenever EMC is an issue, always use shielded cables.
• Never run signal and power wires in the same conduit.
• Use signal wire connections with twisted-pair cables.
• Install Ferrite Bead(s) on signal wire close to the instrument if
EMC problems persist.
Using the Menus
CLOSED
OPEN
CLOSED
OPEN
OPEN
OPEN
OPEN
OPEN
OPEN
OPEN
2-Wire
Magnetic
Pickup
S1
For complete wiring information on all setup
options, please refer to the Operator’s Manual.
To Select Operating Mode
(Rate or Total):
1.
Press MENU until the
display shows Func .
Func
RECALL
2.
Press
SET to show
the current mode.
SET to
Press
select between rAtE
and totAL .
MENU
RESET
/SET
/ADV
MENU
RESET
totAL
RECALL
4.
/ADV
rate
RECALL
3.
/SET
Press MENU to store
the mode into volatile
memory.
/SET
/ADV
MENU
RESET
store
RECALL
/SET
/ADV
MENU
RESET
To store the operating mode (or any value)
into non-volatile memory:
• Press MENU until the meter displays noStor
• Press /SET once to display StorE .
• Press MENU again.
The value is stored in non-volatile memory.
.