Mode Button Operation
To activate the “MODE” button, place the tip of the
magnetic stylus over the “MODE” button and hold for
three seconds until the unit enters the “MODE” menu.
Then move away the magnetic stylus. Each time the
magnetic stylus is again placed near the “MODE” button
the meter will step through each parameter.
Set Button Operation
The “SET” button has two primary functions.
Independently the button is used to turn the
backlighting on. To activate the “SET” button, place
the tip of the magnetic stylus over the “SET” button
and hold for three seconds until the backlighting turns
on. The backlighting will remain on for the amount of
time you selected during the setup and configuration
process. Note: if you selected “Always Off” when setting
the meter up, then the “SET” button will not turn the
backlighting on from the front keypad.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects
in materials and workmanship for a period of 13 months from
date of purchase. OMEGA’s WARRANTY adds an additional one (1)
month grace period to the normal one (1) year product warranty
to cover handling and shipping time. This ensures that OMEGA’s
customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized
Return (AR) number immediately upon phone or written request.
Upon examination by OMEGA, if the unit is found to be defective, it
will be repaired or replaced at no charge. OMEGA’s WARRANTY does
not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation
outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having
been tampered with or shows evidence of having been damaged as a
result of excessive corrosion; or current, heat, moisture or vibration;
improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components in which wear
is not warranted, include but are not limited to contact points, fuses,
and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any
damages that result from the use if its products in accordance
with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by
the company will be as specified and free of defects. OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF
ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT
THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING
ANY WARRANTY OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED.
LIMITATION OF LIABILITY: The remedies of purchaser set
forth herein are exclusive, and the total liability of OMEGA
with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise,
shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for
consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used,
nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC),
used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with
any nuclear installation or activity, medical application, used on humans,
or misused in any way, OMEGA assumes no responsibility as set forth
in our basic WARRANTY/DISCLAIMER language, and, additionally,
purchaser will indemnify OMEGA and hold OMEGA harmless from any
liability or damage whatsoever arising out of the use of the Product(s)
in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA
Customer Service Department.
BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER
SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the
outside of the return package and on any correspondence.
Apply Magnet 3
Seconds Then
Remove
0%
MQS4978/1018
100%
MODE
MIN VALUE
Apply Magnet
Then Remove
Apply Magnet
Then Remove
MIN
MIN VALUE
MIN
MIN, MAX AND
AVG ALL SET TO
VALUE OF
CURRENT READING
0%
Apply Magnet
Then Remove
100%
MIN
0%
Apply Magnet
Then Remove
MAX
100%
MAX
MIN, MAX AND
AVG ALL SET TO
VALUE OF
CURRENT READING
UNITS
Apply Magnet
Then Remove
SET
UNITS
Apply Magnet
Then Remove
AVG VALUE
Apply Magnet
Then Remove
AVG
0%
MAX
100%
AVG VALUE
Apply Magnet
Then Remove
Continued from
previous column
MAX
0%
When in the “MODE” menu, the “SET” button is used to
turn the wireless transmitter “On” or “Off” and also to
select the unit of measurement.
MIN
100%
MAX VALUE
MAX VALUE
0%
100%
AVG
AVG
MIN, MAX AND
AVG ALL SET TO
VALUE OF
CURRENT READING
0%
0%
100%
UNIT
0%
100%
HAL
BLINKING STOPS
ALARM SILENCED
0%
LOW ALARM VALUE
Apply Magnet
Then Remove
0%
100%
HAL
Apply Magnet
Then Remove
0%
LAL
BLINKING STOPS
ALARM SILENCED
0%
100%
Apply Magnet
X4
Then Remove
LAL
TX
Apply Magnet
Then Remove
UNITS
MIN VALUE
MIN
100%
IN H20
WIRELESS OPTION
TX
0%
100%
LAL
WIRELESS OPTION
Apply Magnet
Then Remove
100%
LOW ALARM VALUE
LAL
High Accuracy Digital Pressure Gauge
HAL
Apply Magnet
Then Remove
100%
UNITS
HIGH ALARM VALUE
HAL
0%
DPG409 SERIES
BATT
HIGH ALARM VALUE
Apply Magnet
Then Remove
PSI
Apply Magnet
Then Remove
BAT
0%
100%
AVG
100%
BATTERY VALUE
Apply Magnet
Then Remove
FOR WARRANTY RETURNS,
please have the following
information available BEFORE
contacting OMEGA:
1. Purchase Order number
under which the product
was PURCHASED,
2. Model and serial number of the
product under warranty, and
3. Repair instructions and/or
specific problems relative
to the product.
FOR NON-WARRANTY REPAIRS,
consult OMEGA for current repair
charges. Have the following
information available BEFORE
contacting OMEGA:
1.Purchase Order number to cover
the COST of the repair or
calibration,
2. Model and serial number of the
product, and
3.Repair instructions and/or specific
problems relative to the product.
OMEGA’s policy is to make running changes, not model changes,
whenever an improvement is possible. This affords our customers
the latest in technology and engineering.
OMEGA is a trademark of OMEGA ENGINEERING, INC.
© Copyright 2018 OMEGA ENGINEERING, INC. All rights reserved.
This document may not be copied, photocopied, reproduced,
translated, or reduced to any electronic medium or machine-readable
form, in whole or in part, without the prior written consent of OMEGA
ENGINEERING, INC.
MENU MODE
For complete product manual:
www.omega.com/manuals/manualpdf/M4978A.pdf
WRLS
0%
100%
Apply Magnet
Then Remove
WRLS
Apply Magnet
Then Remove
0%
100%
MIN
0%
100%
mbar
EXIT MODE
WIRELESS OPTION
Apply Magnet
Then Remove
Apply Magnet 3
Seconds Then
Remove
omega.com info@omega.com
0%
Continued
Next Column
0%
Menu Button Operation
100%
100%
EXIT
WRLS
Menu Button Operation Cont'd
Servicing North America:
U.S.A.
Omega Engineering, Inc.
Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only)
Engineering Service: 1-800-872-9436 (USA & Canada only)
Tel: (203) 359-1660
Fax: (203) 359-7700
e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts
no liability for any errors it contains, and reserves the right to alter specifications without
notice.
The information contained in this document is believed to be correct, but
OMEGA accepts no liability for any errors it contains, and reserves the right to
alter specifications without notice.
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
ARTWORK/PRODUCT ART/
DWGS/START HERE ARROW
START HERE
2
3
4
CAUTION:
Using This Quick Start Manual
Use this Quick Start Manual with your DPG409
Series high accuracy digital pressure gauge for
quick installation and basic operation. For detailed
information, refer to the User’s Guide available at
https://www.omega.com/manuals/manualpdf/
M4978A.pdf.
General Description
Omega’s DPG409 Series Digital Pressure Gauges
incorporate rugged, 316 Stainless Steel enclosures
that are designed specifically for wash-down,
sanitary and marine applications. The large backlit
LCD display features 1" high digits that make
reading at distances up to 35 feet easy. Models
are available with integral standard or sanitary
pressure sensors that incorporate highly stable
silicon wafer technology which is micro-machined
to precision tolerances and then has strain gauges
molecularly embedded.
Failure to install these screws correctly may results in
damage to your unit when the meter is exposed to
wet conditions. When reinstalling enclosure screws
they must be tightened to 9 in-lb +/- 10%.
Switching On Your DPG409 Meter
Before you connect your USB cable to the USB
programming connector on the DPG409, you will need
to switch on the unit.
Setting Up Your Meter
Connect the USB Programming cable included with your
unit to your PC and your meter.
ON/OFF
SWITCH
USB
The DPG409 is NEMA 4X rate (water tight, dust
tight, corrosion – resistance - indoor & outdoor
use). The pressure gauge can be installed in
locations where it will be intermittently exposed to
spraying water, rain or high humidity. The meter
should never be used under water.
Software Installation
Download the Configuration Wizard software
from ftp://ftp.omega.com/public/DASGroup/
products/dpg409/. To install run the setup.exe
application. The installer will guide you through
the steps required to install the DG Configuration
Wizard on your computer.
Lid/Cover Removal
To switch on your unit and access the USB
programming connector on your meter, the
enclosure lid/cover must be removed. Four screws
must be removed. Be careful to not lose or discard
these screws. These screws play a vital part in
providing the water tight seal on your unit.
REMOVE THESE FOUR SCREWS
IN ORDER TO REMOVE
THE METER FRONT COVER
Lid/Cover Removal
Wiring (Power, Analog Output)
24Vdc POWER SUPPLY
TB2
BATTERY
4-20 mA SETTING
BATTERY
TB1
S1
0-5 Vdc/
0-10Vdc SETTING
A
A
B
B
C
C
Mounting
For mounting dimensions refer to the User’s Guide,
Page 4-1, Figure 4-1.
Now that you have connected your USB cable to your
PC and to your meter, you can now configure your
meter before placing the unit into operation. You will be
using the configuration software utility that you installed
onto your PC. If you have not installed the configuration
software utility you should do so now.
Configuring your Meter
The Omega Digital Gauge Configuration Wizard will
open once you have launched the DG Configuration
Wizard software. Follow the step-by-step instructions
given to you by the Configuration Wizard. Step 5 Choose Options: Select the type of analog output your
application requires. You can leave the default setting if
you will not be using the analog output feature.
NOTE:
You must set the wire jumpers located on the back
of your meter to match the analog output type you
selected here in the setup process.
Wiring - Power Supply
Model DPG409 requires only one battery for normal
operation (Battery 1). A second battery (Battery 2)
can be added to extend operation when on battery
power. It is also recommended that the additional
battery (Battery 2) be installed for models that include
the optional wireless transmitter.
BATTERY 1
PIN 1
TB1
4
5
Ambient Temperature
Your meter should only be installed in locations that
maintain an ambient temperature between -18 to 66°C
(0 to 150°F).
NOTE:
4
TB2
PIN 1
Battery Installation/Replacement
To install or replace the battery in your DPG409
pressure gauge you must first remove the four screws
located on the Lid of the enclosure. This will allow you
to remove the meter assembly and access the battery
holders.
USB Connector Location
5
Analog Output Options
Installation
+
–
CHART RECORDER
+
–
Wiring - Analog Output
For alarm wiring information refer to the User's
Guide, Page 4-4, Figure 4-5.
Keypad Operation
Magnetic Stylus
A magnetic stylus is attached to your meter. This
stylus is used to activate the front buttons on the
display. If you do not need to view the “MIN”,
“MAX” or “AVERAGE” readings during normal
operation, or have a need to turn the backlight on
during normal operation you can remove and store
this stylus. You will need to select “Disable Keypad”
during the setup and configuration process.
MAGNETIC TIP
+
+
BATTERY
BATTERY
A
B
C
–
–
BATTERY 2
Battery Placement
Magnetic Stylus
The “MODE”
and “SET”
buttons located
on the front of the display of your meter are
activated by waving or tapping the magnetic stylus
included with your unit above or onto the keypad
button symbol on the front meter label.