OM-CP-QUADPROCESS and OM-CP-OCTPROCESS
Product Overview
Engineering Units
Engineering units are used to convert one measurement reading to another. The OM-CP Data Logger software
allows for software level Engineering Units (conversion applied to data after download). Certain devices have
device level Engineering Units, which upon download automatically appear in the chosen unit of measure.
Wiring the Data Logger
The OM-CP-QUADPROCESS and OM-CP-OCTPROCESS both have two-position removable screw terminal
connections. The OM-CP-QUADPROCESS has 4 connections, the OM-CP-OCTPROCESS has 8 connections.
They accept 3-wire configurations.
Warning: Note the polarity instructions. Do not attach wires to the wrong terminals.
OM-CP-QUADPROCESS and OM-CP-OCTPROCESS
Installation Guide
Installing the Interface Cable
- OM-CP-IFC200
Insert the device into a USB port. The drivers will install automatically.
- OM-CP-IFC110
Plug the serial cable into the port and verify it is secure.
Installing the software
Insert the Omega Software Flash Drive in an open USB port. If the autorun does not appear, locate the drive
on the computer and double click on Autorun.exe. Follow the instructions provided in the Wizard.
Device Operation
Connecting and Starting the data logger
- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer.
- The device will appear in the Connected Devices list, highlight the desired data logger.
- For most applications, select “Custom Start” from the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the data logging application and click “Start”. (“Quick
Start” applies the most recent custom start options, “Batch Start” is used for managing multiple loggers at once,
“Real Time Start” stores the dataset as it records while connected to the logger.)
- The status of the device will change to “Running”, “Waiting to Start” or “Waiting to Manual Start”,
depending upon your start method.
- Disconnect the data logger from the interface cable and place it in the environment to measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point
the device cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Highlight the data logger in the Connected Devices list. Click “Stop” on the menu bar.
- Once the data logger is stopped, with the logger highlighted, click “Download”. You will be prompted to
name your report.
- Downloading will offload and save all the recorded data to the PC.
OM-CP-QUADPROCESS and OM-CP-OCTPROCESS
Device Maintenance
Battery Replacement
Materials: 3/32” HEX Driver (Allen Key) and Replacement Battery (OM-CP-BAT103)
- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
The OM-CP-QUADPROCESS or OM-CP-OCTPROCESS standard calibration is at two points. The points are
dependent on the range of the data logger.
Range
1mA
25mA
100mA
Calibration Point
0mA and 0.9-1mA
0mA and 22.5-25mA
0mA and 90-100mA
Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in
the software when the device is due. Specifications subject to change. See Omega’s terms and conditions at
www.omega.com
OM-CP-QUADPROCESS and OM-CP-OCTPROCESS
OM-CP-QUADPROCESS General Specifications
Description OM-CP-QUADPROCESS
Current Range
Current Resolution
Current Accuracy
*See Table Below
Input Impedance
Channels 4
Memory 32,767/channel
Reading Rate 1 reading every second up to 1 reading every 12 hours
LED Indicator None
Required Interface Package IFC110 or IFC200
Baud Rate 2,400
Typical Battery Life 1 year
Operating Environment -20°C to +60°C, 0%RH to 95%RH (non–condensing)
Material Anodized aluminum
Dimensions
3.5” x 4.4” x 1.0”
(89 mm x 112 mm x 26 mm)
Weight 13 oz (370 g)
*OM-CP-QUADPROCESS Range, Resolution and Calibrated Accuracy
Nominal Range
+-1mA
+-25mA
+-100mA
Measurement Range
+-1.5mA
+-30mA
+-120mA
Common Mode Input Range
0 to 2.5V
0 to 2.5V
0 to 2.5V
0.05µA
1µA
5µA
+-0.5 %FSR
+-0.1%FSR
+-0.1 %FSR
Resolution
Calibrated Accuracy @25 oC
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH,
PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 60°C (140°F).
Specifications subject to change.
See Omega’s terms and conditions at www.omega.com
Still need help?
For more troubleshooting tips and information, refer to the built in help section visit us online at www.omega.com
or contact us for support at 1 (800) 872-9436.
OM-CP-QUADPROCESS and OM-CP-OCTPROCESS
OM-CP-OCTPROCESS General Specifications
Description OM-CP-OCTPROCESS
Current Range
Current Resolution
Current Accuracy
*See Table Below
Input Impedance
Channels 8
Memory 16,383/channel
Reading Rate 1 reading every second up to 1 reading every 12 hours
LED Indicator None
Required Interface Package IFC110 or IFC200
Baud Rate 2,400
Typical Battery Life 1 year
Operating Environment -20°C to +60°C, 0%RH to 95%RH (non–condensing)
Material Anodized aluminum
Dimensions
3.5” x 4.4” x 1.5”
(89 mm x 112 mm x 39 mm)
Weight 17 oz (480 g)
*OM-CP-OCTPROCESS Range, Resolution and Calibrated Accuracy
Nominal Range
+-1mA
+-25mA
+-100mA
Measurement Range
+-1.5mA
+-30mA
+-120mA
Common Mode Input Range
0 to 2.5V
0 to 2.5V
0 to 2.5V
0.05µA
1µA
5µA
+-0.5 %FSR
+-0.1 %FSR
+-0.1 %FSR
Resolution
Calibrated Accuracy @25 oC
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH,
PENETRATE OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 60°C (140°F).
Specifications subject to change.
See Omega’s terms and conditions at www.omega.com
Still need help?
For more troubleshooting tips and information, refer to the built in help section visit us online at www.omega.com
or contact us for support at 1 (800) 872-9436.
omega.com info@omega.com
U.S.A.
Headquarters:
Servicing North America:
Omega Engineering, Inc.
Toll-Free: 1-800-826-6342 (USA & Canada only)
Customer Service: 1-800-622-2378 (USA & Canada only)
Engineering Service: 1-800-872-9436 (USA & Canada only)
Tel: (203) 359-1660
Fax: (203) 359-7700
e-mail: info@omega.com
For Other Locations Visit omega.com/worldwide
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any errors it contains, and reserves the
right to alter specifications without notice.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product
warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an
Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found
to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from
any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits,
improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with
or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper
specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components in which wear is not
warranted, include but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes
responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its
products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only
that the parts manufactured by the company will be as specified and free of defects. OMEGA MAKES NO OTHER
WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth
herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10
CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will
indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in
such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER
SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the
outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1. Purchase Order number under which the product was
PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the
product.
FOR NON-WARRANTY REPAIRS, consult OMEGA for current
repair charges. Have the following information available BEFORE
contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the
product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and
engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2016 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.