Omega | OM-EL-1 | Owner Manual | Omega OM-EL-1 Owner Manual

Omega OM-EL-1 Owner Manual
®
An OMEGA Technologies Company
OM-EL-1
PANEL MOUNT DATALOGGER
INSTRUCTION
SHEET
M-3139
OM-EL-1 measures, displays, records and controls temperature, humidity, pH, voltage, current, etc., with a memory for 8064 readings and a battery life of up
to 3 years**. OM-EL-1 can operate as a 'stand alone' logger or be permanently connected to a system. The OM-EL-1 serial link is addressable and up to 8
loggers can be connected to one serial port. The PC software operates under Windows 3.1 and does not require specialist skill to operate. Data output is in
text format and can be easily integrated into any popular spreadsheet. Graphical output is possible under OM-EL-WIN. Consult the EasyLog software
manual for further details.
Battery Powered
Multi-function
°C
Non-volatile Data Storage
Panel Mounting
Easy to Use
Cold Junction Compensation For
Thermocouple Temperature Sensing
Windows 3.1 Control Software
Model No - OM-EL-WIN
Easy to install and use, the control software will enable the user
to control one or more EasyLogs and operate them as a complete
system. Supplied on a 3½" disk with a manual and serial link.
OM-EL-WIN will run under Windows 95.
Accessories - Cables
Model No
Data Logger
OM-EL-1
Specification
Range
Extension connector to 'daisy chain' more than one EasyLog.
One extension will be needed for each extra logger.
* Sensor dependent.
** Battery life dependent on input mode and sample
rate.
E.g. Rate Count, 5 sec. sample rate.
Battery life is approximately 30 days.
DIMENSIONS
Temperature (internal)
Voltage
Current
Rate count
Frequency
Battery
Serial link
Sensor connection
Memory capacity
Sample rate
58.0 (2.28)
1.6
(0.06)
a
Ø2.50
(0.10)
16.0
(0.63)
NOTE - BEZEL SHAPE AND
DIMENSIONS ARE LIABLE
TO CHANGE WITHOUT
PRIOR NOTICE
5.0 (0.20)
46.5(1.76)
17.8 (0.70)
30.0 (1.18)
55.0 (2.16)
25.0 (0.98)
44.0 (1.73)
All dimensions in mm (inches)
Accuracy
-25 to +200°C
1°C
±1°C
-13 to +392°F
1°F
±2°F
-10 to +50°C
1°C
±1°C
+14 to +122°F
1°F
±2°F
5 to 95% RH
1%RH
*
0 to 14pH
0.1pH*
±0.1pH*
0 to 2V D.C.
0.01V
±1%
0 to 20V D.C.
0.1V
0 to 2A
0.01A
±1%
0 to 255
1 unit
N/A
40 to 400Hz
N/A
±2Hz
3.6V ½AA lithium (up to 3 years life)**
8 Pin Mini DIN
Screw terminal
up to 8000 readings
1 sample per 5 seconds to 1 per 12 hours.
Temperature ('K' type thermocouple)
Humidity
pH
Model No - OM-EL-EASYLINK
Resolution
1.5
(0.05)
1.1
(0.044)
9.0 (0.35)
6.1
(0.24)
64.5 (2.54)
2.00
(0.08)
Panel cut-out
62 x 32 (2.44 x 1.26)
31.8 (1.25)
2.50
(0.09)
USING EXTERNAL SIGNALS
5.2 (0.20)
34.5 (1.36)
1.0
(0.04)
18.0 (0.70)
47.0 (1.84)
12.8
(0.50)
PIN FUNCTION
A1, A2, REF
Output for test diagnostics. Do not use.
HA
High Alarm output = V+ when alarmed.
SW
'Push to record event' input. Momentarily connect to V-.
RS
Reset input. Momentarily connect to V-. NOTE - logging will stop and restart at 0 time.
LA
Low Alarm output = V+ when alarmed.
A+
Output = V+ when a reading is in progress.
DO NOT APPLY AN EXTERNAL SUPPLY VOLTAGE
V+
Battery positive.
ACROSS THESE PINS WITH BATTERY IN PLACE.
VBattery negative.
Only connect outputs to high impedance inputs, otherwise performance will be affected and battery life drastically reduced.
BATTERY REPLACEMENT
Only use ½AA 3.6V lithium. The list below is not exhaustive. Check with supplier that the battery you are ordering is 'press fit' and is not fitted with solder
tags or leads. Take care to connect correctly.
MANUFACTURER
MAXELL
SAFT
SONNENSCHEIN
TADIRAN
PART NUMBER
ER 3S TC
LS3
SL-750/S
½AA/S
MANUFACTURER'S ORDER CODE
n/a
n/a
1107 501 100
1551-02-210-000
WARNING: Handle lithium batteries carefully - observe warnings on battery casing. Dispose of in accordance with local regulations.
VARIOUS OPERATING MODES
Pluggable Links (Link 1)
On-Board Link (Link 2)
OPEN
SHORTED
“K" type thermocouple
YELLOW (+VE)
A
B
Normally
OPEN
Solder to
SHORT
Normally
SHORTED
Cut to
OPEN
A
B
+
+
A
B
- 0-20V D.C.
-
pH PROBE
RED (-VE)
TEMPERATURE (with thermocouple)
A 'K' type thermocouple can be connected
directly to the OM-EL-1. The range should
be selected in the OM-EL-WIN software.
VOLTAGE*
D.C. voltage within the ranges of 0-2V or
0-20V can be connected directly to the
OM-EL-1. Ensure the appropriate range
has been selected in advance within the
OM-EL-WIN software. For higher voltage
measurement an external potential
divider network should be added.
TEMPERATURE (internal)
No sensor needed. The OM-EL-1
uses its internal temperature
A
sensor.
Ensure A to B is
B
SHORTED.
pH (with internal reference)
Connect a pH probe that has an internal
reference cell directly to the OM-EL-1.
(refer to the probe's instruction manual).
NOTES
1. There is no need to open Link 2 if the
SENTEK P14 probe is used.
2. OPEN LINK 2
+
A
B
A
B
SENSOR
1Ω
6watt
+
0-2A D.C.
A
B
0V
CURRENT*
D.C. current can be measured up to 2A
maximum by using an external 1Ω 6watt
shunt across the input. Select the voltage
'0-2V range' in OM-EL-WIN software and
'A' in the display symbols.
HUMIDITY
A capacitive humidity sensor can be
connected directly to the EasyLog.
NOTE
1. There is no RH annunciator on the LCD.
Sensor range - 200 to 300pF
0.5 to 1.0pF per %RH
0 - 5V
0
RATE COUNT*
OM-EL-1 counts the number of pulses on
the input during the sample interval (e.g.:
every 5s, 10s, 20s etc.). At the end of each
sample interval, the count number is stored
and the counter is reset to zero. A pulsed
input of 0-5V can be directly connected to
the OM-EL-1. The counter is incremented
on the falling edge.
*NOTES
1. When using the serial link, ensure that all signals are isolated from the computer ground.
2. When using several loggers together, ensure all signals are isolated or use the same ground.
omega.com
WARRANTY/DISCLAIMER
TM
®
OMEGAnet On-Line Service
http://www.omega.com
Internet e-mail
info@omega.com
SM
Servicing North America:
USA:
One Omega Drive, Box 4047
ISO 9001 Certified Stamford, CT 06907-0047
Tel: (203) 359-1660
e-mail: info@omega.com
Canada:
976 Bergar
Laval (Quebec) H7L 5A1
Tel: (514) 856-6928
e-mail: canada@omega.com
FAX: (203) 359-7700
FAX: (514) 856-6886
For immediate technical or application assistance:
USA and Canada:
Mexico and
Latin America:
SM
Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA
SM
Customer Service: 1-800-622-2378 / 1-800-622-BEST
SM
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
SM
Tel: (95) 800-TC-OMEGA
FAX: (95) 203-359-7807
En Español: (203) 359-1660 ext:2203 e-mail: espanol@omega.com
Servicing Europe:
Benelux:
Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 20 6418405
FAX: (31) 20 6434643
Toll Free in Benelux: 06 0993344
e-mail: nl@omega.com
Czech Republic:
Ostravska 767, 733 01 Karvina
Tel: 42 (69) 6311899
FAX: 42 (69) 6311114
e-mail: czech@omega.com
France:
9, rue Denis Papin, 78190 Trappes
Tel: (33) 130-621-400
FAX: (33) 130-699-120
Toll Free in France: 0800-4-06342
e-mail: france@omega.com
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: 49 (07056) 3017
FAX: 49 (07056) 8540
Toll Free in Germany: 0130 11 21 66
e-mail: germany@omega.com
United Kingdom:
25 Swannington Road,
PO Box 7, Omega Drive,
Irlam, Manchester,
ISO 9002 Certified Broughton Astley, Leicestershire,
LE9 6TU, England
M44 5EX, England
Tel: 44 (1455) 285520
Tel: 44 (161) 777-6611
FAX: 44 (1455) 283912
FAX: 44 (161) 777-6622
Toll Free in England: 0800-488-488
e-mail: uk@omega.com
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product
warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each
product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue
an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found
to be defective it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any
action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper
repair, or unauthorized modification.This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows
evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification;
misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted,
including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes
responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products
in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts
manufactured by it will be specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are
exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence,
indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability
is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under
10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any
Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way,
OMEGA assumes no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and additionally, purchaser
will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the
Product(s) in such a manner.
RETURN REQUESTS / INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS).The assigned AR number should then
be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following
FOR NON-WARRANTY REPAIRS, consult OMEGA for
information available BEFORE contacting OMEGA:
current repair charges. Have the following information
1. P.O. number under which the product was PURCHASED,
available BEFORE contacting OMEGA:
2. Model and serial number of the product under warranty,
1. P.O. number to cover the COST of repair,
and
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to
3. Repair instructions and/or specific problems relative to
the product.
the product.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our
customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written
consent of OMEGA ENGINEERING, INC.
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of
its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability for any errors it contains,
and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.