Omega | TCL-3M165E2 | Owner Manual | Omega TCL-3M165E2 Owner Manual

Omega TCL-3M165E2 Owner Manual
Series TCL-3M165E2
.
Temperature Calibrator
Series TCL-3M165E2
-2-
Series TCL-3M165E2
Table of contents
page
0 About this operating manual ...................................................................................... 5
1 Device description ..................................................................................................... 6
1.1 Delivery, unpacking and accessories ...................................................................... 7
1.2 Intended use............................................................................................................ 8
1.3 Exclusion of liability ................................................................................................. 8
2 Safety Instructions ..................................................................................................... 9
2.1 Safety instructions for the application of calibration liquids ................................... 11
3 Construction, Function and Test Task ..................................................................... 12
3.1 Construction .......................................................................................................... 12
3.2 Function ................................................................................................................ 13
3.3 Test tasks .............................................................................................................. 14
4 Commissioning and operation ................................................................................. 15
4.1 Operating conditions ............................................................................................. 15
4.2 Electrical connection ............................................................................................. 16
4.3 Preparing of the calibrator ..................................................................................... 17
4.3.1 Dry block calibrator ........................................................................................... 18
4.3.2 Calibrator for infrared thermometer ................................................................... 19
4.3.3 Calibrator for surface temperature sensors ....................................................... 20
4.3.4 Micro bath calibrator.......................................................................................... 21
4.4 Switching on, cool down and switching off ............................................................ 25
5 Operation................................................................................................................. 27
5.1 Main window.......................................................................................................... 28
5.2 Configuration range ............................................................................................... 29
5.3 Window structure, operating elements and icons .................................................. 29
5.3.1 Operating elements ........................................................................................... 30
5.3.2 Toolbar and icons.............................................................................................. 31
5.4 Functionality of the calibrator................................................................................. 32
5.4.1 Select Function ................................................................................................. 32
5.4.2 Manage Function .............................................................................................. 32
5.4.3 Configure Function ............................................................................................ 33
5.5 Test specimen (DUT - Device under test) ............................................................. 34
5.5.1 Select DUT ........................................................................................................ 34
5.5.2 Manage DUT ..................................................................................................... 35
5.5.3 Configure DUT .................................................................................................. 35
Technical changes reserved
-3-
Series TCL-3M165E2
5.6 Test Task .............................................................................................................. 36
5.6.1
Name ........................................................................................................... 37
5.6.2
Data acquisition ........................................................................................... 37
5.6.3
Functionality .................................................................................................. 37
5.6.4
Test specimen (DUT) .................................................................................. 38
5.6.5
Test points ..................................................................................................... 38
5.6.6
Alarm settings ............................................................................................. 40
5.6.7
Scan / Barcode .......................................................................................... 40
5.7 Calibrator settings ................................................................................................. 41
5.7.1 Calibrator Set Up .............................................................................................. 41
5.7.2 Measurement reports ........................................................................................ 45
6 Testing process / Calibration ................................................................................... 46
7 Problems ................................................................................................................. 48
7.1 Return shipment to the manufacturer .................................................................... 48
8 Maintenance and cleaning ...................................................................................... 50
8.1 Maintenance.......................................................................................................... 50
8.2 Cleaning ................................................................................................................ 51
8.3 Recalibration ......................................................................................................... 52
8.4 Adjustment ............................................................................................................ 53
9 Decommissioning and disposal ............................................................................... 54
10 Technical data ......................................................................................................... 55
10.1 Characteristics TCL-3M165E2 .............................................................................. 55
10.2 Heating up and cooling times ................................................................................ 57
10.2.1 TCL-3M165E2................................................................................................... 57
Copyright notice:
The reproduction, distribution and utilization of this operating manual as well as the communication of
its contents to others without express authorization is prohibited. Offenders will be held liable for the
payment of damages. All rights reserved in the event of the grant of a patent, utility model or design.
-4-
Series TCL-3M165E2
0
About this operating manual
About this operating manual
•
The operating manual is aimed at specialists and semi-skilled personnel.
•
Before each step, read through the relevant advice carefully and keep to the specified
order.
•
Thoroughly read and understand the information in the section “Safety Instructions”.
•
For calibrators with a cooling function, the term "Cooling" is also used for temperatures
below room temperature, in the meaning of "Heating".
If you have any problems or questions, please contact your supplier or contact us directly
at:
Omega.com
Tel: (203) 359-1660
e-mail: info@omega.com
Hazard signs and other symbols used:
DANGER! Risk of death due to electric current!
This sign indicates dangers which could lead to serious health defects or to death.
WARNING! / CAUTION! Risk of injury!
This sign indicates dangers that cause personal injuries that can lead to health defects or
cause considerable damage to property.
CAUTION! High temperature!
This sign indicates dangers resulting from high temperature that can lead to health defects or
considerable damage to property.
CAUTION! Material damage!
This sign indicates actions which could lead to possible damage to material or environmental
damage.
NOTICE!
ADHERE TO OPERATING MANUAL!
This symbol indicates important notices,
tips or information.
NO DOMESTIC WASTE!
The device must not be disposed of
together with domestic waste.
Pay attention to and comply with
information that is marked with this symbol.
 Follow the specified instructions and steps.
Adhere to the given order.
Technical changes reserved
 Check the specified points or notices.
 Reference to another section, document or
source.
• Item.
-5-
Device description
1
Series TCL-3M165E2
Device description
The TCL-3M165E2 series of calibrators test and calibrate different temperature measuring
instruments and temperature sensors, and measure temperatures.
The calibrator can be very easily operated using the touch-sensitive screen or touch screen.
Pre-defined or self-defined testing tasks, by which the testing process is governed, are
employed for this purpose.
The portable instruments are of compact and robust construction and thus allow use directly
on-site or in a laboratory. The calibrators meet the requirements of EN 61326-1, class A
(industrial sector).
This TCL-3M165E2 series is used for service purposes and for different industrial and
laboratory tasks.
Thus, for example, thermometers, temperature switches/thermostats, resistance
thermometers and thermo-elements can be directly connected and checked.
Versions:
The TCL-3M165E2 is available as a micro bath, dry block, infrared, surface or multi-function
calibrator.
It can be supplied both with a heating function as well
as a cooling and heating function.
The series and the various types are different in their
construction (tank or metal block, the temperature
range, the accuracy, the adapter sleeves and
calibration inserts, the fittings and the available
accessories.
Further information on the series, the models of the
individual types that can be supplied, and their fittings
can be found in our catalogues.
Type plate:
You find the type plate on the rear of the device.
It includes the type designation, the serial number and the key electric
specifications ( example).
Type plate example
-6-
Series TCL-3M165E2
Device description
1.1 Delivery, unpacking and accessories
All units have been carefully checked for their operational reliability before shipment.
 Immediately after receipt, please check the outer packaging for damages or any signs of
improper handling.
 Report any possible damages to the forwarder and your responsible sales
representative. In such a case, state a description of the defect, the type and the serial
number of the device.
Report any in-transit damage immediately. Damage reported at a later date shall not be
recognized.
Unpacking:
 Carefully unpack the unit to prevent any damage.
 Check the completeness of the delivery based on the delivery note.
Save the packaging!
Temperature calibrators are delivered in special protective packaging.
 Save the packaging for returning the instrument safely to the manufacturer for
recalibration or repair.
Scope of delivery and accessories:
Scope of delivery












TCL-3M165E2 according to the order data.
Test certificate.
Power cable.
Sleeve exchange tool.
PC- and network cable.
Operating manual.
Protective packaging / transport
protection.
Sensor cage.
Drain syringe.
Transport cover.
Magnetic stirrer with magnet lifter.
Work cover with 5 silicone plugs.
Accessories (optional)
Transport case.
Adapter sleeve(s).
Infrared insert.
Surface insert and exchange tool.
Tub insert.
External reference sensor TF 255-3-300.
PC software.
Network switch, barcode reader, WLAN
router.
 Works certificate.








IMPORTANT!
 Use the type plate to check if the delivered unit corresponds to your order.
 In particular, for devices with electrical components, check to see if the correct power
supply voltage is specified.
Technical changes reserved
-7-
Device description
Series TCL-3M165E2
1.2 Intended use
The TCL-3M165E2 series of calibrators may only be used for testing and calibration of
suitable temperature measuring instruments, temperature sensors and for measuring
temperatures.
The calibrators may not be used for warming up or heating other parts or gases.
The calibrators have been designed for indoor use only.
The micro baths may only be used with suitable media. Permitted liquids are silicone oils,
mineral oils and water ( § 2.1 "Safety instructions for the application of calibration liquids").
Hazardous media (flammable or explosive liquids or gases) may not be used.
The operational safety of the device supplied is only guaranteed by intended use. The
specified limits ( § 10 "Technical data") may under no circumstances be exceeded.
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
 Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C (95 °F or below 50 °F).
 Allow the calibrator to cool before you remove the test specimen, clean the tank,
change the adapter sleeve or the calibration insert or switch off the calibrator.
CAUTION! Material damage!
The opening in the metal block of the calibrator is only intended to be used with adapter
sleeves or calibration inserts.
Using heat transfer media (oil, thermal paste or other media) can lead to incorrect
measurements and damage to the calibrator.
 Never fill the metal block opening with a heat transfer medium.
 Only micro baths are suitable for use with heat transfer medium.
It is your responsibility to select the instrument which is suitable for your specific application,
to connect it correctly, to carry out tests and to maintain all the components.
1.3 Exclusion of liability
Omega accepts no liability for any damage or malfunctions resulting from incorrect
installation, in-appropriate use of the device or failure to follow the instructions in this
operating manual.
-8-
Series TCL-3M165E2
2
Safety Instructions
Safety Instructions
Before you install the TCL-3M165E2, read through this operating manual carefully. If the
instructions contained within are not followed, in particular the safety guidelines, this
could result in danger for people, the environment, and the device and the system it is
connected to.
The TCL-3M165E2 corresponds to the state-of-the-art technology. This concerns the
accuracy, the operating mode and the safe operation of the device.
In order to guarantee that the device operates safely, the operator must act competently and
be conscious of safety issues.
OMEGA ENGINEERING provides support for the use of its products either personally or via
relevant literature. The customer verifies that our product is fit for purpose based on our
technical information. The customer performs customer-specific and application-specific tests
to ensure that the product is suitable for the intended use. With this verification all hazards and
risks are transferred to our customers; our warranty is not valid.
Qualified personnel:
The personnel in charge of the installation, operation and maintenance of the TCL3M165E2 must hold a relevant qualification. This can be based on training or relevant
instructions.
Personnel must be aware of this operating manual and have access to it at all times.
The electrical connection should only be carried out by a fully qualified electrician.
General safety instructions:
In all work, the existing National regulations for accident prevention and safety in the
workplace must be complied with. Any internal regulations of the operator must also be
complied with, even if these are not mentioned in this manual.
Ensure that the complete operating instructions are always available in excellent condition
at the calibrator installation site.
Degree of protection according to EN 60529:
Ensure that the ambient conditions at the site of use does not exceed the requirements
for the stated protection rating ( § 10.1 "Characteristics TCL-3M165E2").
Structural safety in accordance with EN 61010-1:
The calibrator must be installed in such a way that the requirements for structural safety
are met.
Only use the TCL-3M165E2 if it is in perfect condition. Damaged or faulty devices must
be checked without delay and, if necessary, replaced.
If problems cannot be cleared, immediately shut down the calibrator and ensure that it
cannot be started up accidentally.
Never leave the calibrator unattended when it is in operation or in the cooling phase.
Do not remove or destroy type plates or other markings on the device, or the warranty is
rendered null and void.
Technical changes reserved
-9-
Safety Instructions
Series TCL-3M165E2
Special safety instructions:
Thermal fuse!
The calibrator is equipped with a temperature fuse that works independently. If there is an
over-temperature in the inside of the housing, the power supply to the heating system is
cut off. The calibrator cannot then be started any more.
 After it has cooled, send the calibrator for inspection to OMEGA ENGINEERING.
Risk of injury from hazardous gases!
When liquids are heated, the evaporation can result in dangerous gases being formed.
The calibrator may not be used in an explosion-endangered atmosphere (ignitable or
explosive atmosphere).
 Remove all the easily flammable media from the vicinity of the calibrator.
 Ensure that the calibrator cannot come in contact with easily flammable or explosive
media.
Operate the calibrator only in the temperature range permissible for the test sample.
Ensure that the test sample is securely fixed in the calibrator.
 Use only suitable adapter sleeves or calibration inserts.
When doing so, also ensure that the structural safety of the calibrator is retained.
Expert mode!
When selecting the function, test sample and testing tasks, you can switch into the expert
mode for administration and configuration. The settings that are made in this mode
require detailed knowledge of the calibration and method of working of calibrators.
If the settings are incorrect, the calibrator can get damaged!
IMPORTANT NOTICE!
The transport cover is equipped with a safety valve, which is activated once the pressure
reaches ~1.5 bar (21.75 psi). This can result in hot steam being released.
 Always unscrew the transport cover before putting the micro bath into service, in
order to avoid excessive pressure.
 Wait until the micro calibration bath has cooled down before screwing on the transport
cover.
Further warnings that are specifically relevant to individual operating procedures or activities
can be found at the beginning of the relevant sections of this operating manual.
- 10 -
Series TCL-3M165E2
Safety Instructions
2.1 Safety instructions for the application of calibration liquids
Before using calibration liquids, read the entire safety data sheet thoroughly. Pay
particular attention to the information on the physical and chemical properties.
Only use calibration liquids that are suitable for the required temperature range and which
are not flammable.
Always wear safety goggles for the eyes when handling calibration liquids.
We recommend the following calibration liquids for the various temperature ranges:
Calibration liquid
Distilled water
Calibration Range
Flashpoint
2…95 °C (36…203 °F)
none
-35 °C (-31 °F) 155 °C (311 °F)
165 °C (329 °F)
Silicone oil from XIAMETER®:
PMX-200 SILICONE FLUID 10 CS
Water:
Only use distilled water, otherwise excessive limescale and soiling will build up in the
calibrator tank.
Silicone oil:
Use only the silicone oil recommended here.
Always read the safety data sheet supplied with the silicone oil before using it.
Always ensure adequate ventilation when working with silicone oil, since hazardous
substances can be released.
Prevent silicone oil from coming into contact with your eyes.
Since silicone oil is hygroscopic, always use the transport cover to close the calibration
bath after use.
Mineral oil:
OMEGA ENGINEERING supplies the calibrators only with silicone oil.
Using mineral oil is possible, but must be done on your own responsibility. The danger
and the risk must be borne by the customer and our warranty will be rendered null and
void.
Please follow the safety data sheet of the mineral oil used.
The safety instructions for silicone oil apply analogously for mineral oils as well. The same
also applies to the corresponding sections for silicone oil in this operating manual.
Technical changes reserved
- 11 -
Construction, Function and Test Task
3
Series TCL-3M165E2
Construction, Function and Test Task
3.1 Construction
The calibrator consists of a robust, black-and-red painted steel housing  and has a carrying
handle at the top .
The front part of the housing holds the electronics of the TCL-3M165E2 for controlling the
reference temperature and for the touch screen .
On the front side are the main switch with fuse and power plug connection , the
connection for the external reference , as well as the ports for the PC, the network and
service .
The grille
for the exhaust air is located, in the TCL-3M calibrators, in both the side walls of
the housing.
The rear part of the housing holds a heat-insulated calibration block with heating or cooling
elements and an integrated sensor for the reference temperature. In the case of the dry block
and multifunction calibrator, this is a metal block  and in the micro bath calibrator, it is a
tank .
In the base, there are grilles for cooling the calibrator. The built-in fans control the inlet air of
the housing
and of the calibration block .
An overview of the most important components of the TCL-3M165E2:
 Carrying handle.
 Steel housing.
 Touch screen for operation and





measurement value display.
Main switch with fuse and power plug
socket.
Tank (Micro bath).
Calibrator ventilation:
Exhaust through lateral and upper
housing grilles.
Metal block (Dry block).
Connection for external reference.
Ports for PC, network and service.
Calibrator, bottom view:
 Calibrator ventilation
Inlet air for housing cooling.
Inlet air for tank / metal block cooling.
- 12 -
Series TCL-3M165E2
Construction, Function and Test Task
3.2 Function
IMPORTANT NOTICE!
For your safety, manual operation of the calibrator, i.e. directly starting it up and its
remaining at a certain temperature, is not possible.
The testing procedure is always started with a testing task ( § 3.3).
This ensures that the calibrator is always started with a defined "Behavior at test end"
( p. 25).
The calibrators can be used with dry block, infrared, surface or micro bath function.
The function is determined by the calibrator type and the used measuring insert.
Type
Function
Measuring insert:
Dry block
Infrared
Surface
Micro bath
Adapter sleeve
Infrared insert
Surface insert
Calibration liquid




TCL-3M165E2
Method of working:
Adapter sleeves or calibration inserts that are suitable for the temperature sensors or
temperature measuring instruments to be tested are inserted in the calibrator
( § 4.3 "Preparing of the calibrator").
They give the test specimen a secure fit and ensure optimum heat transfer.
Once all the preparations have been carried out, the calibrator can be switched on ( § 4.4).
The calibrator heats or cools the metal block or tank to the set temperature. As soon as this
temperature has been reached and is stable, the calibrator signals accordingly.
The calibration of the test specimen can now be carried out. Then, the next test point is
approached or the testing procedure is ended.
Functionality:
The measuring inserts ( p. 17) of the calibrator have different characteristic values, which
were determined in our laboratory through calibration. These characteristic values are
defined at the factory and stored in the calibrator as a protected function.
To be able to use a measuring insert, the relevant "Functionality of the calibrator" ( § 5.4)
must be selected.
IMPORTANT! Protected list entries!
The functions, test specimens and test tasks marked with the
symbol are factorydefined. They can be neither deleted nor configured.
They serve as templates for your own functions, test specimens and test tasks, which
are marked with the
symbol.
Technical changes reserved
- 13 -
Construction, Function and Test Task
Series TCL-3M165E2
3.3 Test tasks
Test tasks are containers for defined test conditions. They are helpful for recurring testing
processes, for standardizing test sequences and for generating measurement logs.
All the required settings and configurations for the calibration of a test specimen are compiled
in a test task. The parameters of the testing task are saved and linked to the selected
function and test specimen ( § 5.6).
The operation concept of the calibrator is based on pre-installed and self-defined test tasks.
They are a central component of the function and operation of the TCL-3M165E2.
The emphasis of the operating manual is on the use of test tasks during operation of the
calibrator.
The calibrator is shipped with functions, test samples and testing tasks defined in the factory,
in which the basic settings of the calibrator are stored. If desired, customer-specific test tasks
can also be pre-set.
These protected test tasks can be neither deleted nor edited. They serve as templates for
your own self-defined test tasks. They can be copied and can then be configured.
You can define your own test tasks for different test specimens or test sequences. These test
tasks are directly saved in the calibrator and can then be easily activated. This makes quick
access to recurring test tasks possible.
Upon switching on the calibrator, the first test task of the selection list is loaded automatically,
with the relevant parameters.
IMPORTANT! The function and test specimen are independent!
During the configuration of test tasks ( § 5.6) when selecting the function or the test
sample, you can switch to the respective expert mode for administration and
configuration.
 Note here that changes to existing functions and test specimens always apply to all
the test tasks linked to them.
- 14 -
Series TCL-3M165E2
4
Commissioning and operation
Commissioning and operation
CAUTION! Risk of injury or material damage!
The calibrator can become very hot during operation. If the calibrator is operated without
supervision, third-party persons in the vicinity could get injured. Moreover, flammable
material could get into the calibrator and cause significant damage to property.
 Never leave the calibrator unattended when it is in operation or in the cooling phase.
For safe operation of the calibrators of the TCL-3M165E2 series, a proper commissioning
procedure is necessary.
Commissioning includes the installation, the electrical connections, the preparation for the
test task as well as correct switching on and off of the calibrator.
Further, a visual inspection for damage is required before use.
The required steps are described in the following sections.
IMPORTANT!
Please also follow the instructions for intended use ( § 1.2), the safety instructions
( § 2) and the information on the ambient conditions ( p. 56).
4.1 Operating conditions
Select a safe installation site for commissioning the calibrator.
Installation site and operating position:
 Only suitable for indoor use, do not use outdoors.
 Operate only in the vertical position on an even surface. The surface must be stable,
clean, and dry.
If the operation positions do not conform to the above, the structural safety and the
specified properties of the calibrator are not guaranteed.
 At higher testing temperatures, please use a sufficiently big, fire-resistant supporting
surface.
 Sufficient clearance around the calibrator
On the front side > 1 m, behind and to the sides > 0.5 m. Head clearance and sufficient
clear space above the calibrator.
 Sufficient ventilation must be ensured.
 Do not operate in the vicinity of flammable materials.
 Do not install in a cupboard or other similar location.
 The ventilation openings must not be blocked or covered.
 The calibrator must be installed so that it can be switched off at any time.
IMPORTANT! "KILL" switch!
The plug of the power connecting cable serves as a "KILL" switch.
 Ensure that the plug is always easily accessible and easy to reach.
 In an emergency, pull the plug, so that the calibrator is isolated from the power.
Technical changes reserved
- 15 -
Commissioning and operation
Series TCL-3M165E2
4.2 Electrical connection
The electrical connections are made with the supplied power connecting cable.
DANGER! Risk of death due to electric current!
The TCL-3M165E2 is operated at power voltages up to 240 VAC. Contact with the power
voltage can result in serious or fatal injuries.
 Switch off the TCL-3M165E2 and remove the power connecting cable before you start
any work on live parts.
Check the following points before you connect the calibrator:
 Operate the calibrator only with the approved supply voltage ( § 10.1). Ensure that the
power voltage is the same as that specified on the type plate.
 Please follow the local regulations of the energy supplier.
 Connect the calibrator only to a properly installed and earthed 3-pole outlet.
 Do not use any extension cables or adapter plugs.
IMPORTANT! Power connecting cable!
The power connecting cable may only be replaced by an equivalent cable.
 Use only original cables from OMEGA ENGINEERING or approved cables of the
same type with the correct design as replacements ( "Electrical characteristics").
Connect TCL-3M165E2:
 Connect the power connecting cable with the connector plug of the TCL-3M165E2.
 Insert the plug of the power connecting cable in a suitable outlet.
- 16 -
Series TCL-3M165E2
Commissioning and operation
4.3 Preparing of the calibrator
The preparations for the test task must be carried out with the calibrator switched off and
cooled to ambient temperature.
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious burn injuries.
 Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C (95 °F or below 50 °F).
 Allow the calibrator to cool before you remove the test specimen, clean the tank,
change the adapter sleeve or the calibration insert or switch off the calibrator.
IMPORTANT! Empty the tank and clean it!
The tank must be emptied and cleaned to prevent the adapter sleeve or the calibration
insert from getting stuck.
 Empty and clean the tank before you insert another adapter sleeve or calibration
insert.
Measuring insert:
The function of the calibrator is determined by the measuring insert. The required measuring
insert is inserted in the opening of the metal block or the tank.
It is thus easily possible to switch between the dry block, infrared, surface and micro bath
functions.
Each measuring insert has its own characteristic values, which are defined at the factory.
These characteristic values are influenced by different factors, such as: Diameter, type and
material of the measuring insert, number and diameter of the bore holes or the physical
properties of the calibration liquid.
These characteristic values are stored as protected functions for the accompanying
measuring inserts of the calibrator ( § 5.4).
Technical changes reserved
- 17 -
Commissioning and operation
Series TCL-3M165E2
4.3.1 Dry block calibrator
Type
TCL-3M165E2
Dry block

Adapter sleeves with single or multiple holes are used for the calibration of straight
temperature sensors.
To achieve the specified accuracy of the calibrators ( § 10.1) the temperature
sensor (test specimen) and the adapter sleeve must be matched to one another:
•
•
The borehole of the adapter sleeve may be a maximum of 0.5 mm (0.12 in)
larger than the diameter of the test specimen.
The measurement element of the test specimen must be located in the
homogeneous temperature zone of the adapter sleeve.
Inserting:
 The appropriate adapter sleeve is inserted
into the metal block with the aid of the
sleeve exchange tool.
External reference sensor (optional):
 Align the adapter sleeve so that the hole
for the external reference sensor is located
at 12 o’clock.
Removing and cleaning:
 Let the calibrator cool before you remove
the adapter sleeve.
Adapter sleeves and sleeve exchange tool
 Pull the adapter sleeve out of the metal
block with the help of the sleeve exchange tool.
 Clean the adapter sleeve and the metal block.
This prevents the adapter sleeve from getting stuck in the metal block.
- 18 -
Series TCL-3M165E2
Commissioning and operation
4.3.2 Calibrator for infrared thermometer
Type
TCL-3M165E2
Infrared

A special infrared insert is used for the contact-less measurement infrared thermometers.
This makes calibration fast and easy.
The infrared insert has a special design and surface coating on the
inside. As a result, an emissivity of 0.9994 (black body) is reached.
The additional holes in the border (2 x 3.5 mm • 1 x 4.5 mm /
2 x 0.14 in • 1 x 0.18 in) are for external reference sensors. In this
manner, the exact temperature on the inside of the infrared insert can
be measured.
IMPORTANT! Size of the measuring spot!
The measuring spot of the infrared thermometer must project itself to the floor of the
infrared insert during the calibration. The measuring spot must be smaller than the inner
diameter and must not touch the wall of the infrared insert.
Inserting:
 The infrared insert is inserted into the metal block with
the aid of the sleeve exchange tool.
External reference sensor (optional):
 Align the infrared insert so that the hole for the external
reference sensor is located at 12 o’clock.
Removing and cleaning:
 Let the calibrator cool before you remove the infrared
insert.
 Pull the infrared insert out of the metal block with the
help of the sleeve exchange tool.
Infrared insert
 Clean the infrared insert and the metal block.
This prevents the infrared insert from getting stuck in the metal block.
FORMATION OF ICE AND DEW!
At temperatures < 0 °C (32 °F) and higher humidity levels ice or condensation can form
in the infrared insert. This can result in the calibration of the infrared thermometer being
distorted.
The forming of ice or condensation can be significantly reduced by covering the measuring
opening of the infrared insert.
 Keep the measuring opening closed for as long as possible.
 Only open the measuring opening briefly for measuring.
 Existing ice or condensation can be removed by heating the infrared insert.
Technical changes reserved
- 19 -
Commissioning and operation
Series TCL-3M165E2
4.3.3 Calibrator for surface temperature sensors
Type
TCL-3M165E2
Surface

A special surface insert is used for calibrating surface temperature sensors. This makes
calibration fast and easy.
The surface insert is hollow from the bottom and longer than
the adapter sleeve. For reference sensors, three boreholes
are located directly under the surface (1 x 3 mm •
1 x 3.1 mm • 1 x 4.5 mm / 1 x 0.12 in • 1 x 0.12 in •
1 x 0.18 in). Thus, the exact surface temperature of the
insert can be measured.
The two threaded boreholes in the border are for the
accompanying exchange tool.
Inserting:
 The surface insert is inserted into the metal block with
the aid of the exchange tool.
External reference sensor (optional):
 Align the surface insert so that the hole for the external
reference sensor is located at 12 o’clock.
Surface insert and
exchange tool
Removing and cleaning:
 Let the calibrator cool before you remove the surface insert.
 Pull the surface insert out of the metal block with the help of the exchange tool.
 Clean the surface insert and the metal block.
This prevents the surface insert from getting stuck in the metal block.
IMPORTANT NOTICE!
The calibration of surface temperature sensors is very difficult and not uncontroversial.
The temperature sensor, which are placed on the surface, conduct heat away from the
surface and thus produce a cold spot on the surface to be measured.
Using a multifunction calibrator, the calibration temperature is generated in a specially
constructed surface insert and measured directly under the surface with an external
reference thermometer.
The reference thermometer also determines the temperature of the cold spots by integrating
the temperature over the sensitive length of the reference thermometer and can therefore
offer a true temperature calibration of surface temperature test specimens.
The surface insert is constructed in such a way that the included external reference delivers
the best possible results because the depth of the boreholes is adapted to the sensitive
length. If you need to make your own external reference for a comparison calibration, ensure
that the sensitive length is known and that it is situated in the middle of the calibration surface
- 20 -
Series TCL-3M165E2
Commissioning and operation
4.3.4 Micro bath calibrator
Type
TCL-3M165E2
Micro bath

CAUTION! Risk of injury - Wear safety goggles!
Calibration liquid can be ejected when working with the micro bath calibrator.
 Always wear safety goggles for the eyes when handling calibration liquids.
The micro bath is used for calibrating sensors with special shapes or dimensions. Direct
contact between the sensor and the calibration liquid ensures excellent heat transfer.
The calibration liquid is poured directly into the tank or into a tub insert ( p. 22).
The micro bath includes the transport cover, the work cover, the sensor cage, the magnetic
stirrer, the drain syringe, the magnetic lifter and as an accessory, the tub insert. The
individual components are described below.
Transport cover:
The transport cover serves for secure closing of the micro bath. It prevents spillage of the
calibration liquid during transport.
CAUTION! Safety valve!
The transport cover is equipped with a safety valve, which is activated
once the pressure reaches ~1.5 bar (~21.75 psi). This can result in hot
steam being released.
 Always unscrew the transport cover before putting the micro bath into service, in
order to avoid excessive pressure.
Work cover:
The work cover fulfils various tasks during the operation.
• It reduces the evaporation of the calibration liquid to a minimum.
• It reduces the cooling on the surface of the calibration liquid.
• It ensures stable positioning of the test specimens in the micro bath.
The work cover is screwed on to the micro bath and has five openings for
test specimens. The unused openings can be closed with suitable silicon
plugs.
Sensor cage and magnetic stirrer:
The sensor cage protects the magnetic stirrer. It prevents the sensors from
blocking the magnetic stirrer. The function of stirring is ensured by the
sensor cage.
The magnetic stirrer ensures a uniform temperature distribution in the
calibration liquid. The speed of the magnetic stirrer is set in Configuration
range ( § 5.2) or in the window "Configure Function" ( § 5.4.3 "Stirring
speed").
IMPORTANT! Limited Life Part!
The magnetic stirrer is a limited life part.
 Replace worn-out magnetic stirrers. ( § 8.1 "Maintenance").
Technical changes reserved
- 21 -
Commissioning and operation
Series TCL-3M165E2
Drain syringe and magnetic lifter:
The drain syringe is used for pumping out the calibration liquid from the tank of the micro
bath. The magnetic stirrer is removed with the help of the magnetic lifter.
Both the activities are required before another adapter sleeve or calibration insert is inserted
in the calibrator.
Tub insert (optional / accessories):
We recommend using a tub insert if you
•
•
frequently change between dry block, infrared, surface and micro
bath function.
frequently work with different calibration liquids.
The tub insert is placed into the tank with the help of the sleeve exchange
tool.
Just like the tank, the tub insert can be closed with the associated cover.
Both threaded covers are leakproof, so the calibration liquid can be left in
the tank or the tub insert during transport.
4.3.4.1 Notes on calibration liquid
Different calibration liquids supply varying calibration results due to their specific
characteristics. Adjustment to the respective calibration liquid has to be carried out by the
manufacturer.
In order to achieve the best possible accuracy of a micro bath, it has to be filled with a
suitable calibration liquid.
The calibration liquid is poured directly into the tank or a tub insert.
When using water as the calibration liquid:

Only use distilled water, otherwise excessive limescale and soiling will build up in the
tank.
When using silicone oil as the calibration liquid:

Wear safety goggles for the eyes without fail when working with silicone oil!

Use only the silicone oil recommended here.

Always read the safety data sheet supplied with the silicone oil before using it.

Always ensure adequate ventilation when working with silicone oil, since hazardous
substances can be released.

Spilled or leaked silicone oil results in an extreme danger of slipping. Clean the
affected ranges by suitable means.

Since silicone oil is hygroscopic, always use the associated transport cover to
close the tank or the tub insert after use.
- 22 -
Series TCL-3M165E2
Commissioning and operation
IMPORTANT NOTICE!
Only use clean calibration liquid. Checking temperature sensors and other temperature
detection means can lead to a contamination of the calibration liquid. This contamination
can lead to smeary gel effect on the bottom of the tank due to the rotation of the
magnetic stirrers.




Clean the tank.
Clean before calibration the sensors.
Exchange the worn magnet stirrer.
Exchange dirty, smeary calibration fluid.
4.3.4.2 Information on filling amounts
CAUTION! Risk of incorrect measurement or material damage!
Do not exceed the maximum fill level during operation!
• Above the maximum fill level the heat dissipation is too great, preventing compliance
with the specified tolerances.
• Overflow of the calibration liquid causes contamination and can damage the calibrator.
 Ensure that the maximum fill level is not exceeded during operation.
The fill level in the tank or tub insert rises as a result of
•
Thermal expansion
Calibration liquids expand to varying degrees as a
result of heating. The increase in fill level depends
on the calibration liquid that is used and the
reference temperature setting.
•
Displacement by sensors
The volume displaced by the sensors being
calibrated must be taken into account in the filling
amount.
•
Rise due to stirring
The rotation of the magnetic stirrer forms a
whirlpool in the liquid. This raises the fill level at
the wall.
Tank:
The maximum fill level in the tank is displayed by
the upper edge of the aluminum lining.
max. fill level of the micro bath
The maximum fill level is ~0.45 liters.
Tub insert:
The maximum fill level with the tub insert is below the fixture for the sleeve exchange tool.
The maximum fill level is ~0.32 liters.
Technical changes reserved
- 23 -
Commissioning and operation
4.3.4.3 Filling the micro bath *
Series TCL-3M165E2
1
IMPORTANT! Pay attention to § 4.3.4.2 "Information on filling amounts"!
 When filling, leave enough room for thermal expansion, displacement by sensors and
level rise due to stirring.
 Unscrew the cover of the tank/ tub insert.
 Insert the tub insert in the tank using the sleeve exchange tool
(only for tub insert).
 Place the magnetic stirrer in the tank / tub insert.
 Insert the sensor cage.
 Insert the test specimens in the sensor cage.
This accounts for the volume of the sensors to be
tested.
 Fill the calibration liquid in the tank / tub insert.
Leave sufficient reserve space for an additional rise in
the level.
 If necessary, remove the test specimens again.
 Screw the work cover onto the tank and insert the sensors through the work cover into
the tank or tub insert.
*1 For tubs already filled, some step are not required.
- 24 -
Series TCL-3M165E2
Commissioning and operation
4.4 Switching on, cool down and switching off
IMPORTANT!
Keep in mind the instructions and notes in the previous sections ( § 4.1 +  § 3.3).
For reasons of safety, upon switching on the calibrator, the fan runs at the fastest speed. As
soon as the internal reference has measured a safe block temperature, the fan speed is
adjusted.
Switching on:
 Put on the main switch.
 The fan of the device starts and the OMEGA logo appears for ~ 40 s on the screen.
 The type designation and the current software version are displayed for ~ 4 s.
 The main window is displayed and the device is ready for operation.
 The first entry in the selection list, with its parameters, is displayed as the test task.
IMPORTANT! Keep in mind after transport or storage!
After transport, storage or long periods of non-use, moisture can seep into the heating
elements (magnesium oxide).
For drying, the calibrator must be slowly heated up. During this process, the calibrator
has not yet reached the required insulation voltage for protection class I.
 For drying the heating elements, heat the calibrator for at least 15 min to 120 °C
(248 °F)
Cool down the calibrator:
To avoid injuries or material damage, it is necessary to bring the calibrator into a defined
operational state.
To that end, a "Safety temperature " ( p. 40) is saved in the calibrator. Upon switching off,
the calibrator will go to it via the main window.
IMPORTANT! Power failure or separation from the power!
If there is a power failure, or if the main switch is turned off, or upon removal of the power
connection ("EMERGENCY STOP"), the built-in fan does not blow any more cooling air.
Sufficient thermal de-coupling between the metal block, the tank and the housing is
nonetheless guaranteed
Technical changes reserved
- 25 -
Commissioning and operation
Series TCL-3M165E2
Switching off:
CAUTION! Before switching off, pay attention to the block temperature!
The calibrator should first reach a safe block temperature before switching off. Upon
switching off outside the safe temperature range, the calibrator can get damaged.
 Switch off the calibrator only when the safe temperature range has been reached.
 Press the
key until the main window is displayed.
 Press the
key to switch off the calibrator.
 The calibrator regulates the temperature in the safe range:
The message "Please wait – device is being brought to a safe temperature" is
displayed.
 The safe temperature range was reached:
The message "You can now switch off the device" is displayed.
 Switch off the calibrator via the main switch.
 Clean the calibrator after use ( § 8.2).
- 26 -
Series TCL-3M165E2
5
Operation
Operation
The TCL-3M165E2 is operated by means of the touch screen.
You can use the icons and operating elements of the screen to
navigate through the window. There, you can edit parameters
and input values.
The main window, which is displayed after starting up, is the
starting point for the operation of the TCL-3M165E2.
From there, you can get to the various windows or start the
testing process.
Upon tapping the corresponding symbols or parameter fields, the
respective window of the parameter is opened or displayed.
The windows "Calibrator Set Up" and "Select measurement report" are opened with the
icons
and
. The window "Select Test Task" is opened from the parameter field with the
name of the test task.
Operating structure:
The following diagram gives an overview of the operation structure of the TCL-3M165E2. The
zones are schematically depicted with their windows and parameters. Furthermore, the
parameters of the other operating elements of the main window can be seen.
The operating concept of the TCL-3M165E2 is based on pre-installed and self-defined test
tasks for recurring testing processes.
One-off testing processes can be directly set via the configuration range and then started.
Technical changes reserved
- 27 -
Operation
Series TCL-3M165E2
5.1 Main window
The individual icons and the most important ranges of the main window are presented in brief
in the following.
 Calibrator Set Up:








By using the
icon, you can reach the "Calibrator Set Up" window. There, you can
change the settings of the calibrator ( p. 41).
Measurement reports:
By using the
icon, you can reach the "Select log data" window. There, you can select
saved measurement logs and view them ( p. 45).
Remote Control:
Enables Internet communication.
Test Task:
Using this parameter field, you can reach the "Select Test Task" window. There, you can
select the saved test tasks and create or configure your own test tasks (p. 36).
Switching off:
By using the
icon, you can switch the calibrator off ( § 4.4 "Switching on, cool down
and switching off").
Display range:
The reference temperature, the set temperature as well as additional information related
to the selected test task (p. 47).
Alarm signal:
You can switch the alarm signal on or off using the icon
/
.
In the window "
Alarm settings" ( p. 40) you can define under what criteria the alarm
signal is to be activated.
Start / Stop slider:
The start / stop slider is used to start
or end
the testing process.
Configuration range ( § 5.2):
You can use the parameter fields of the configuration range to get to the parameter
windows of the calibrator.
By using the
icon, you can get to the parameter fields of the other parameters.
- 28 -
Series TCL-3M165E2
Operation
5.2 Configuration range
The calibrator settings can be changed directly from the configuration range. You can get to
the windows of the parameters from the corresponding parameter fields.
The operation of the calibrator via the configuration range is particularly well suited for one-off
testing processes. You can directly set the required test conditions and then start the testing
process.
You can save the new test conditions under the current test task or a new test task. Without
saving, the test conditions are discarded.
IMPORTANT! In expert mode, keep the following in mind!
From the windows "Select Function" ( § 5.4.1) and "Select DUT" ( § 5.5.1) you can
switch to the expert mode for administering and configuring.
 Follow the instructions in the corresponding sections.
 Changes to functions and test specimens always affect all the test tasks linked to
them.
5.3 Window structure, operating elements and icons
The windows are divided into three ranges: the information line, the toolbar and the
parameter range.
Information line:
The information line is located at the top edge of
the window. It shows the name of the window and
the reference temperature.
Toolbar:
The toolbar is located along the right edge of the
window. It contains the icons available in this
window.
Parameter range:
The parameter range is the biggest part of the
window. Information about the window and other operating elements are displayed in it.
Technical changes reserved
- 29 -
Operation
Series TCL-3M165E2
5.3.1 Operating elements
The following operating elements can be found in the parameter range of the window.
Information box:
Different kinds of information is displayed in the
information box. This information can be
explanations for the window, information on the
selected element of the selection list,
measurement values or measurement value
curves.
If the information is extensive, such as
measurement value curves, the section of the
information box can be scrolled (adjusted).
Selection list:
The selection list displays the elements available in the window. This can be available
test tasks, recorded measurement logs, saved test specimens or functions of the
calibrator.
The desired element is selected by tapping
and its information is displayed in the
information range.
The scrollbar on the right can be used to easily view long lists.
Parameter fields:
Parameter fields are active elements of the window. From the parameter fields, you can
get to the next window or carry out defined actions.
Numeric keypad:
The numeric keypad is for inputting numeric values.
Virtual keyboard and input line:
The alphanumeric virtual keyboard can be used for inputting text and numbers in the
relevant input line.
Dialog box:
Special messages are displayed in a dialog box.
- 30 -
Series TCL-3M165E2
Operation
5.3.2 Toolbar and icons
Some icons occur in different windows, selection fields / lists and input ranges and have the
same meaning.
Cancel / back:
Return to the previous window. Changes are discarded without saving.
Input confirmation / Save:
The selected value or setting is confirmed and saved.
Manage / configure:
The selected entry in the selection list is processed and the relevant window is
displayed.
Copy:
The selected entry in the selection list is copied and the relevant window is displayed.
Create new:
A new entry is generated for the selection list and the relevant window is displayed.
Delete:
The selected entry in the selection list is deleted.
Search:
The selection list is searched for an entry ( for example p. 36 "search for test
tasks") and the result is displayed in the selection list. Tapping the icon once again
shows the entire list.
Sorting A-Z:
The selection list is sorted alphabetically in ascending order.
Sorting Z-A:
The selection list is sorted alphabetically in descending order.
Export:
The selected entry in the selection list is copied and the relevant window is displayed.
For better differentiation and simpler operation, most of the parameter fields are marked with
additional symbols before the text in the windows. These icons are also used in the operating
manual.
Technical changes reserved
- 31 -
Operation
Series TCL-3M165E2
5.4 Functionality of the calibrator
Here, you can select and manage the function of the calibrator for the relevant measuring
insert. You can create new functions or tailor unprotected functions to your requirements.
IMPORTANT! When changing functions, please consider!
The individual functions of the calibrator are independent. Changes to existing functions
always affect all the test tasks that are linked to the function.
You can get to the "Select Function" window via the parameter field in the configuration
range of the main window or from the window "Test Task configuration".
5.4.1 Select Function
Information on the selected function is displayed in
the information box.
All the functions of the calibrator are displayed in the
selection list.
 Select the desired function by tapping on
in the
selection list.
 The information on the function is immediately
displayed in the information box.
From the toolbar, you can accept the selected function, or switch to expert mode for
managing and configuring the functions.
Accept function:
 Confirm your selection with the
icon.
 The selected function is adopted with all its settings, and you will be returned to the
main window or to the previous window.
Managing functions:
 Tap on the
icon to manage the functions in the expert mode.
 The "Manage Function" window is displayed. This window offers special functions for
managing and configuring the functions. Details can be found in the following section.
5.4.2 Manage Function
From the "Select Function" window, you can use the
Function".
This window is different from the "Select Function"
window only in the other elements in the toolbar.
The information box and selection list are the same.
The functions can be sorted, copied, deleted or
configured.
- 32 -
icon to go to the window "Manage
Series TCL-3M165E2
Operation
Copy Function:
 Tap on the
icon to copy the selected function.
 A copy is created and the "Configure Function" window is opened.
From the parameter fields of this window, you can set the parameters of the new
function. Details can be found in the following section.
Delete Function:
 On tapping the
icon, the selected function is immediately deleted.
Configure Function:
 Tap on the
icon to edit the parameters of the selected function.
 The "Configure Function" window is opened. Details can be found in the following
section.
5.4.3 Configure Function
From the "Manage Function" window, you can use the
"Configure Function".
icon to go to the window
Here, you can change the parameters of the selected
function.
 Tap on the parameter field of the desired
parameter.
 The parameter’s window is opened. You can
configure the parameter there.
Name
:
You can change the name of the function here. After tapping, the virtual keyboard with the
input line is displayed.
 Input the name of the function and confirm the input.
 The new name is adopted and you will be returned to the previous window.
Serial number
:
The parameter field of the serial number is disabled and cannot be edited.
Minimum temperature :
 Input the new value in the input window and confirm the input.
 The new value is adopted and you will be returned to the previous window.
Maximum temperature :
 Input the new value in the input window and confirm the input.
 The new value is adopted and you will be returned to the previous window.
Own adjustment values :
If required, you can save the corresponding adjustment values for your own measuring
inserts here. Further information on this can be found under "Adjustment" ( § 8.4).
Technical changes reserved
- 33 -
Operation
Series TCL-3M165E2
Setting time :
 Input the new value in the input window and confirm the input.
 The new value is adopted and you will be returned to the previous window.
Stirring speed :
You can change the stirring speed here. The parameter field is only displayed with "Select
Function" as micro bath.
The parameter field of the stirring speed is only
displayed if the calibrator is operated as a micro bath
and the corresponding functionality has been selected.
 Tap on the parameter field "Stirring speed"
 Input the desired speed and confirm the input.
 The new stirring speed is adopted and you will be
returned to the previous window.
Stability range :
 Input the new value in the input window and confirm the input.
 The new value is adopted and you will be returned to the previous window.
5.5 Test specimen (DUT - Device under test)
Here, you can select and manage existing test specimens. You can create new test
specimens or tailor unprotected test specimens to match your requirements.
IMPORTANT! When changing test specimens, please consider!
The individual test specimens of the calibrator are independent. Changes to existing test
specimens always affect all the test tasks that are linked to the test specimen.
You can get to the "Select DUT" window via the parameter field in the configuration range of
the main window or from the window "Test Task configuration".
5.5.1 Select DUT
Information on the selected test specimen is
displayed in the information box.
All the test specimens of the calibrator are displayed
in the selection list.
From the toolbar, you can accept the selected test
specimen, or switch to expert mode for managing and
configuring the test specimens.
Operation is analogous to the section "Select Function" ( § 5.4.1).
- 34 -
Series TCL-3M165E2
Operation
5.5.2 Manage DUT
From the "Select DUT" window, you can use the
icon to go to the window "Manage DUT".
This window is different from the "Select DUT"
window only in the other elements in the toolbar.
The information box and selection list are the same.
The test specimens can be sorted, copied, deleted or
configured, or new ones created.
Operation of the window is analogous to the section
"Manage Function" ( § 5.4.2).
In addition, you can also create a new test specimen.
Create new DUT:
 Tap on the
icon to create a new test specimen.

A new test specimen is created and the window "Configure DUT" is opened. Details
pertaining to the window can be found in the following section.
5.5.3 Configure DUT
From the "Manage DUT" window, you can use the
DUT".
icon to go to the window "Configure
Here, you can change the parameters of the selected
test specimen.
You can define the name, the serial number, the
minimal and maximal temperature, the adjustment
value and the type.
 Tap on the parameter field of the desired
parameter.
 The window of the parameter is opened. You can
configure the parameter there.
Operation is analogous to the section "Configure Function" ( § 5.4.3), with the difference
being that the parameter field of the serial number is activated.
Serial number
:
 Input the new value of the serial number and confirm the input.
 The new value is adopted and you will be returned to the previous window.
Technical changes reserved
- 35 -
Operation
Series TCL-3M165E2
5.6 Test Task
Here, you can select and manage existing test tasks. You can create new test tasks or tailor
unprotected test tasks to your requirements.
From the parameter field "Test Task" of the main window, you get to the window "Select Test
Task".
Select Test Task:
Information on the selected test task is displayed in the
information box.
All the test tasks of the calibrator are displayed in the
selection list.
From the toolbar, you can accept the selected test
task, or search for test tasks or switch to expert mode
for managing and configuring the test tasks.
Operation is analogous to the section "Select Function" ( § 5.4.1).
In addition, you can also search for test tasks:
 Tap on the
icon to search for a test task.
 A virtual keyboard with an input line is displayed.
 Input the desired search term and confirm your input.
 The result of the search is shown in the selection list.
Manage Test Task:
From the "Select Test Task" window, you can use the
Test Task".
icon to go to the window "Manage
This window is different from the "Select Test Task"
window only in the other elements in the toolbar.
The information box and selection list are the same.
The test tasks can be sorted, copied, deleted or
configured, or new ones created.
Operation of the window is analogous to the sections
( § 5.4.2) and ( § 5.5.2).
Configure Test Task:
From the "Manage Test Task" window, you can use the
"Configure Test Task".
Here, you can change the parameters of the selected
test task.
 Tap on the parameter field of the desired
parameter.
 The window of the parameter is opened. You can
configure the parameter there.
- 36 -
icon to go to the window
Series TCL-3M165E2
5.6.1
Operation
Name
Here, you can change the name of the test task. After tapping, the virtual keyboard with the
input line is displayed.
 Input the new name of the test task and confirm the input.
 The new name is adopted and you will be returned to the previous window.
5.6.2
Data acquisition
Here, you can change the name of the measurement log under which the measurement
values of the test task are saved. Moreover, you can define the type of data acquisition.
 Tap on the parameter field to change the desired
parameter.
 The window of the parameter is opened.
Record name:
 Input the new name of the measurement log and confirm the input.
 The new name is adopted and you will be returned to the previous window.
Data acquisition:
Information on the selected type of data acquisition is
displayed in the information box.
 Select the desired type and confirm the input.
 The new type of data acquisition is adopted and
you will be returned to the previous window.
When the data acquisition is activated, the test
specimen temperature is additionally saved to the
measurement log. To that end, a window is opened, in
which you have to input the test specimen temperature.
5.6.3
Functionality
Here, you can link the desired function of the calibrator with the test task.
IMPORTANT!
The selected function must be suitable for the calibrator type ( § 3.2) and the
measuring insert used ( § 4.3).
Details regarding the selected function are displayed
in the information box in the window.
 Select the desired function by tapping on
in the
selection list.
 Confirm the selection by using the
icon.
 The new function is adopted and you will be
returned to the previous window.
You can use the
functions.
icon to switch into expert mode for administering and configuring the
Caution! Changes to existing functions always affect all the test tasks that are linked to
the function ( § 5.4).
Technical changes reserved
- 37 -
Operation
5.6.4
Series TCL-3M165E2
Test specimen (DUT)
Here, you can link the desired test specimen with the test task.
Details of the selected test specimen are displayed in
the information box in the window.
 Select the desired test specimen by tapping on
in the selection list.
 Confirm your selection with the
icon.
 The new test specimen is adopted and you will
be returned to the previous window.
You can use the
icon to switch into expert mode
for administering and configuring the test specimens.
Caution! Changes to existing test specimens always affect all the test tasks that are linked
to the test specimen ( § 5.5).
5.6.5
Test points
The parameters of the test points are set here.
IMPORTANT! Dynamic designation and depiction of the parameter fields.
Depending on the selection of the "Behavior at test end" ( p. 40), individual parameter
fields have another designation or are suppressed.
The information box has several parameter fields for
setting the test points.
All the test points are displayed in the selection list.
The test points can be moved, copied, created,
deleted or changed from the toolbar.
Move test point:
 Select the desired test point by tapping on
in the selection list.
 Move the test point down or up by tapping the
/
icons.
Copy test point:
 Tap on the
icon to copy the selected test point.
 The test point is copied and appended at the end of the selection list.
Create new test point:
 Tap on the
icon to create a new test point.
 The window for the test point setting is opened.
 Input the temperature, the dwell time and the gradient and confirm the relevant input.
 The input values are adopted and you will be returned to the previous window.
Delete test point:
 On tapping the
- 38 -
icon, the selected test task is immediately deleted.
Series TCL-3M165E2
Operation
Change test point:
 Tap on the
icon to modify the selected test point.
 The window for the test point setting is opened.
 Change the temperature and the dwell time and confirm the new values.
 The new values are adopted and you will be returned to the previous window.
Test point calculation
Here, you can easily have the steps between two temperatures calculated. If a change is
made to a parameter, the other parameters dependent on it are calculated afresh
automatically.
 Tap the parameter field to open the window "Calculate test point".
Step, one-sided
:
 Activate the parameter field by tapping
if the test points are to be approached in
succession from the first to the last test point.
Step, two-sided
:
 Activate this parameter field by tapping
if the test points are to be approached in
succession from the first to the last back to the first test point.
Define first and last test point
 To do so, tap on the parameter fields, input the new values and confirm the input.
 The new values are adopted, the interval is calculated afresh and you will be returned to
the previous window.
Cycles:
You can set the number of cycles here. For every cycle, the test points in the selection list
are approached successively.
 Tap on the parameter field to change the number of cycles.
 Input the desired number in the input window and confirm the input.
 The new values are adopted and you will be returned to the previous window.
Set interval:
 For fixed intervals, tap on the parameter field "Interval", input the desired step width and
confirm the input.
 The new interval is adopted and the number of steps is calculated. If necessary, the last
test point is also adjusted to match, and you will be returned to the previous window.
Defining the number of steps:
Note in this context that the number of test points is greater than the number of steps by 1.
 Tap on the parameter field "Number of Steps" to change the number of steps.
Input the desired number and confirm the input.
 The number of steps is adopted and the interval is calculated afresh.
Set dwell time:
 Tap on the parameter field "Dwell time", input the new values and confirm the input.
 The new value is adopted and you will be returned to the previous window.
Technical changes reserved
- 39 -
Operation
Series TCL-3M165E2
Set gradient:
 Tap on the parameter field "gradient", input the new values and confirm the input.
 The new value is adopted and you will be returned to the previous window.
Behavior at test end / Safety temperature value
Using the parameter fields of the new window, you can set the safety temperature value and
define the behavior at test end.
The parameter field "Safety range value" is only visible if "Safety temperature value" was
selected as the behavior at test end.
Behavior at test end:
 Tap on the parameter field "Behavior at test end".
 The "Select behavior at test end" window is displayed.
 Select the desired behavior and confirm the input.
 The new behavior is adopted and you will be
returned to the previous window.
Safety temperature value:
 Tap on the parameter field "Safety temperature
value".
 Input the respective safety temperature value and
confirm the input.
 The new value is adopted and you will be returned to the previous window.
5.6.6
Alarm settings
Here, you can define at what times the calibrator activates the alarm signal.
Details on the selectable times are displayed in the
information box.
 Select the desired times by tapping on
in the
selection list.
 Confirm your selection with the
icon.
 The new times are adopted and you will be
returned to the previous window.
5.6.7
Scan / Barcode
Via the parameter field of the configuration range, an input window is opened with a virtual
keyboard. A connected USB barcode reader can replace the keyboard input here.
When scanning a barcode, all the test tasks and test specimens are searched for the code.
All the entries linked to the barcode are displayed in a selection list.
By using the parameter field of the test tasks, a scanned barcode can be linked to the test
task.
- 40 -
Series TCL-3M165E2
Operation
5.7 Calibrator settings
The most important windows for the setting of the calibrator are described in the following
sections.
5.7.1 Calibrator Set Up
In the "Calibrator Set Up" window, the settings of the calibrator are displayed and the
parameters are changed.
The window has two pages. You can switch between the two pages with the parameter fields
" next" and " back".
By using the parameter fields of the "Calibrator Set Up" window, you can get to the windows
of the parameters. There, the relevant information is displayed or the parameters are
configured.
Parameter fields of the "Calibrator Set Up" window
System information:
The most important data of your calibrator is
displayed. These data are the calibrator type, the production
date, the version of the software, the serial number, the
operating hours (total / since the last calibration) and the
calibration date.
Language selection:
The current language is indicated in the information
box. All the installed languages of the calibrator can be seen
in the selection list.
 Select the desired language by tapping
in the
selection list.
 Confirm your selection with the
icon.
 The selected language is adopted and you will be returned to the "Calibrator Set Up"
window.
Technical changes reserved
- 41 -
Operation
Series TCL-3M165E2
Network settings:
Using the parameter fields of the window, you can
integrate the calibrator in your network.
First connect the calibrator with your network:
 Insert one plug of a suitable network cable in the socket of the calibrator.
 Insert the other plug of the network cable in a free socket of your network.
Now configure the calibrator for the network.
 First set the network mode on.
"DHCP" = automatic assignment of the IP address.
"Manual" = Manual input of the IP address and subnet mask.
 Select the desired mode by tapping in the selection list.
 Confirm your selection with the
icon.
 "DHCP": The IP address and subnet mask are requested from the network. The two
parameter fields are disabled.
In addition, for "Manual":
 Input the IP address from the numeric keypad of the window and confirm the input with
the
icon.
 Input the subnet mast from the numeric keypad of the window and confirm the input with
the
icon.
Date and Time:
Using the parameter fields of the window, you can
set the date and time of the calibrator, as well as change the
format of the date and time.
By tapping the respective parameter field, a window is opened for the input/change.
Upon confirming with the
and Time" window.
icon, the input/change is adopted and you return to the "Date
Display layout:
The depiction on the screen is changed using the
parameter fields in this window.
Upon tapping on the corresponding parameter field, the display on the screen is immediately
adjusted.
- 42 -
Series TCL-3M165E2
Operation
Temperature units:
In the selection list, you can select between Celsius,
Fahrenheit and Kelvin as the temperature unit.
 Select the desired unit by tapping on
in the selection
list.
 Confirm your selection with the
icon.
 The selected unit is adopted, the temperature values
are converted and you will be returned to the "Calibrator Set Up" window.
Software update:
For a software update, you need a new software version. The newest software
version for your calibrator can be found on our website.
 Compare the version level of your calibrator ( "System information") with the version
level of the latest software.
 Download the latest software to a USB flash drive.
 Insert the USB flash drive in the calibrator and wait for 20-30 s so that the USB flash
drive is recognized. Now tap on the parameter field "Software update".
 Follow the instructions on the screen.
Decimal places:
Here, you can define the format of the temperature values
in the windows.
 Select the desired format by tapping on
in the selection list.
 Confirm your selection with the
icon.
 The selected format is adopted, the temperature values are
displayed accordingly and you will be returned to the "Calibrator Set Up" window.
Display settings:
You can use the parameter field of this window to set the
screen brightness.
 Input the desired values from the numeric keypad of the window and confirm the input
with the
icon.
Input values: 1 … 100%.
Presentation format:
You can use the parameter fields of this window to
change the behavior of dialog boxes, configure the toolbar
and set the access level.
Technical changes reserved
- 43 -
Operation
Series TCL-3M165E2
System backup:
When taking a system backup, all the settings of the calibrator, including the software,
are saved to the connected medium.
 Insert the USB flash drive in the calibrator and wait for 20-30 s so that the USB flash
drive is recognized. Now tap on the parameter field "System backup".
 Follow the instructions on the screen.
System recovery:
You can use the system restore function to restore a previously saved system backup
to the calibrator. Next, the calibrator is restarted to activate the restoration.
 Insert the USB flash drive in the calibrator and wait for 20-30 s so that the USB flash
drive is recognized. Now tap on the parameter field "System recovery".
 Follow the instructions on the screen.
Import user files:
You can use this parameter field to load previously saved user settings to the
calibrator. Moreover, you can copy additional OMEGA ENGINEERING factory settings of
measurement tasks, test specimens and functions.
 Insert the USB flash drive in the calibrator and wait for 20-30 s so that the USB flash
drive is recognized. Now tap the parameter field "Import user files".
 Follow the instructions on the screen.
Export user files:
You can use this parameter field to save all the user settings of test tasks, test
specimens and functions to the connected medium.
 Insert the USB flash drive in the calibrator and wait for 20-30 s so that the USB flash
drive is recognized. Now tap the parameter field "Export user files".
 Follow the instructions on the screen.
File format test certificate:
You can use this parameter field to select the file format of the test certificate.
Internal factory access:
You can use this parameter field in the window to set the system parameters of the
calibrator. These settings may only be done by authorized OMEGA ENGINEERING technical
staff.
This range is password-protected!
CAUTION! Material damage!
Service operations are protected by a password. Unauthorized access to this range can
result in damage to the calibrator.
 Do not try to hack the password.
 Quit this range by tapping the
icon.
Start screen:
You can use this parameter field to select remote or standard control upon startup.
- 44 -
Series TCL-3M165E2
Operation
5.7.2 Measurement reports
Here, you can select saved data of testing processes and have them displayed. The
measurement log is saved under the name of the testing process. Several measurement logs
can have the same name, if a testing process was carried out at different times.
From the main window, you get to the window "Select measurement report".
Information on the selected measurement log is
displayed in the information box.
All the measurement logs of the calibrator are
displayed in the selection list.
The measurement logs can be sorted, deleted,
exported or displayed from the toolbar.
 Select the desired measurement log by tapping on
in the selection list.
 The information on the measurement log is immediately displayed in the information
box.
Delete measurement report:
 On tapping the
icon, the selected measurement log is immediately deleted.
Export measurement report:
 Tap on the
icon to export the selected measurement log.
 Follow the instructions on the screen.
Show measurement report:
 Confirm your selection with the
icon.
 The window for displaying the measurement logs
is opened and the data of the selected
measurement log are loaded.
 Move the scrollbar below the measurement value
range to move the time axis of the recording.
 The relevant measurement data are displayed in
the measurement value range. The temperature
axis is scaled automatically.
 Tap on the (+/-) symbols next to the measurement
value range to change the scaling of the time axis.
Technical changes reserved
- 45 -
Testing process / Calibration
6
Series TCL-3M165E2
Testing process / Calibration
CAUTION! Risk of burns!
The calibrator can become very hot when in operation. Touching hot parts can result in
serious injuries.
 Never touch the metal block, the tank, the adapter sleeve or the test specimen at
temperatures above 35 °C or below 10 °C (95 °F or below 50 °F).
 Allow the calibrator to cool before you remove the test specimen, clean the tank,
change the adapter sleeve or the calibration insert or switch off the calibrator.
IMPORTANT! Allow the calibrator to warm up before first calibration!
Allow the calibrator to warm up for at least an hour before the first calibration. This can
be achieved during the calibrator approaches the first testing point, for example.
Before the testing process:
Before the start of the testing process, check whether
 the instructions regarding the installation site and the operational position have been
complied with ( § 4.1).
 the electrical connections have been made correctly ( § 4.2).
 the inside of the block and the surface of the adapter sleeve or the calibration insert are
dry. You can safely remove existing ice or condensation water by heating to above
100 °C (212 °F).
 the correct measuring insert has been selected for the test task ( § 3.3).
 the test specimen is securely fixed in the calibrator.
 the calibrator has sufficient structural stability.
Starting the testing process:
IMPORTANT! Parameter fields disabled!
During the testing process, all the parameter fields, with the exception of the Start / Stop
slider and the alarm signal, are disabled.
• Push the button of the start / stop slider completely to the right.
 The testing process is started. Information on the
testing process and the temperature progression
is displayed in the window.
Information on the testing process is displayed in the
display range ( p. 47).
Testing process finished:
 A dialog box with the message that the test task was successfully
completed is displayed.
• Tap "OK" to close the dialog box.
• Push the button of the start / stop slider completely to the left.
 The main window is displayed.
- 46 -
Series TCL-3M165E2
Testing process / Calibration
Cancel testing process:
• Push the button of the start / stop slider completely to the left.
 The testing process is aborted and the calibrator gets programmed to
reach the temperature of the "Behavior at test end" ( p. 40) of the test task.
Display range:
The following information is shown in the display range:
 Current temperature values of the reference used.
 Current test point.
 Probable remaining test duration for the current







test point.
Time and date.
Operating mode.
Type of data acquisition.
Functionality.
Reference used.
Control.
Status of reference and DUT.
After the testing process:
 Allow the calibrator to cool to room temperature.
 Note the instructions in the section "Test tasks" ( § 3.3).
 Clean the calibrator ( § 8.2).
Technical changes reserved
- 47 -
Problems
7
Series TCL-3M165E2
Problems
CAUTION! Material damage!
The TCL-3M165E2 cannot be repaired by the user! In case of a defect, the device must
be returned to the manufacturer for repair.
 Never open the TCL-3M165E2 and perform any repair yourself.
The following table details what problems you can solve yourself and how to solve them.
Problem
Possible cause
Remedy
Calibrator and touch screen Unknown.
no longer react.
Turn off calibrator, wait a few
minutes, and then restart.
Sensor break.
External reference sensor not
properly connected.
Recheck connection and connect
properly.
Cable break or short circuit.
Service call required.
Fan not running.
The fan is defective or blocked.
The temperature switch is possible
triggered, switching off the power
supply to the heat cartridge
(Service call required).
End temperature is not
achieved.
Solid state relay is defective or
heating / cooling element has short
circuited or aged.
Service call required.
No display.
Controller defective.
Service call required.
No function.
Network connection not established Check the network connection and
correctly or fuse defective.
fuse.
If you are unable to remedy any particular problem, then immediately disconnect the
calibrator in order to protect it from unintended operation.
Contact your supplier or OMEGA ENGINEERING. Please send the device for repair with a
brief description of the problem, the environmental conditions and the length of time the
device was operational before the problem occurred.
7.1 Return shipment to the manufacturer
Due to legal requirements placed on environmental protection and occupational safety and
health and to maintain the health and safety of our employees, all units returned to OMEGA
ENGINEERING for repair must be free of toxins and hazardous substances. That also
applies to cavities in the devices. If necessary, the customer must neutralize or purge the unit
before return to OMEGA ENGINEERING.
Costs incurred due to inadequate cleaning of the device and possible costs for disposal
and/or personal injuries will be billed to the operating company.
WARNING! Risk of injury due to insufficient cleaning!
The operating company is responsible for all damages and harm of any kind, in particular
physical injuries (e.g. caustic burns or toxic contaminations), decontamination measures,
disposal etc. that can be attributed to insufficient cleaning of the measuring instrument.
 Comply with the instructions below before returning the unit.
- 48 -
Series TCL-3M165E2
Problems
The following measures must be taken before you send the unit to OMEGA ENGINEERING
for repair:
 Clean the device thoroughly. This is of extreme importance if the medium is hazardous to
health, i.e. caustic, toxic, carcinogenic or radioactive etc.
 Remove all residues of the media and pay special attention to sealing grooves and slits.
 Attach a note describing the malfunction, state the application field and the
chemical/physical properties of the media.
 Please follow the instructions on the procedure for sending returns which are on our
website and please specify a point of contact in case our service department has any
questions.
The customer must confirm that the measures were taken by filling out the declaration of
decontamination. It can be found on our website as a download:
www.omega.com/cservice/wettedproductreturnform.pdf
Technical changes reserved
- 49 -
Maintenance and cleaning
8
Series TCL-3M165E2
Maintenance and cleaning
Before maintenance and cleaning, check whether
 the calibrator has cooled sufficiently ( § 4.4 "Cool down").
 the calibrator has been switched off and disconnected from the power.
8.1 Maintenance
The TCL-3M165E2 itself is maintenance-free and cannot be repaired by the user. In case of
a defect, the device must be returned to the manufacturer for repair.
CAUTION! Material damage!
When opening the device, critical parts or components can be damaged.
 Never open the device and perform any repair yourself.
For safe operation of the calibrator, the following checks must be carried out at regular
intervals:
Before use:
 Check the calibrator for damage.
 For micro bath calibrators, check the filled height of the calibration liquid. When doing so,
note the Information on filling amounts ( § 4.3.4.2).
Annually:
 Subject all the parts of the calibrator to a visual inspection for corrosion, wear and
damage.
 Have a trained technical person carry out a safety inspection of all the electrical parts.
Recalibration:
 The recommended re-calibration interval ( § 8.3) is 36 months or after a maximum of
500 operating hours.
Calibration liquid:
Calibration liquids get fouled or age with time. This depends to a great extent on the type of
liquid and the usage behavior.
 Replace the fouled or aged calibration liquid.
Magnetic stirrer:
The magnetic stirrer is a limited life part. The fillet in the middle reduces the friction during the
rotary movement. Once the fillet has worn, the stirring function can no longer be guaranteed
because of the increased friction.
 Check the fillet of the magnetic stirrer for wear and replace it in time.
- 50 -
Series TCL-3M165E2
Maintenance and cleaning
Thermal fuse:
The thermal fuses of the calibrator are located on the front side and are integrated in the
power connection. If there is a power voltage present, but the screen is dark and the fan is
not running, you should check the fuses and replace them if required.




Pull the power connection cable from the calibrator.
Open the fuse compartment from the bottom with a flat screwdriver.
Remove the compartment with the fuses.
Check the fuses and replace the faulty fuses.
IMPORTANT! Only use fuses of the same type (T6.3AH 250V).
Always replace both fuses, even if only one is defective.
 Fit the fuse compartment back in place and connect the main connecting cable ( § 4.2
"Electrical connection").
Should the fuses blow repeatedly, there is probably a fault in the calibrator.
In this case, send the calibrator to OMEGA ENGINEERING for repairs ( § 7+§ 7.1).
8.2 Cleaning
External cleaning:
Clean the TCL-3M165E2 with a dry or slightly damp lint-free cloth. Do not use sharp objects
or aggressive agents for cleaning.
Ensure that your cleaning agent cannot be a source of danger from a reaction with parts of
the calibrator or the materials inside it.
If you have any questions regarding compatibility, please contact our customer service.
Venting grilles for inlet air:
The grille openings
+
in the base of the calibrator must be cleaned
at regular intervals.
The cleaning intervals depend very closely on the air pollution at the
installation site and the daily operation duration.
 Clean the grille openings by vacuuming or brushing off.
 Please keep in mind the following instruction regarding the
temperature fuse.
IMPORTANT! Temperature fuse at over-temperature!
An air flow that is too low can result in the temperature fuse getting triggered. The
calibrator is then rendered operationally disabled, and must be sent to OMEGA
ENGINEERING.
 Ensure that the grille openings are always clear.
Technical changes reserved
- 51 -
Maintenance and cleaning
Series TCL-3M165E2
Measuring insert (Adapter sleeves and calibration inserts):
During operation, small quantities of metal dust get created. It can result in the measuring
insert getting stuck in the calibration block.
 Pull the measuring insert out of the calibration block with the help of the sleeve exchange
tool.
 Clean the measuring insert and the calibration block at regular intervals.
Before a prolonged shutdown of the calibrator, remove the measuring insert from the
calibration block.
Tank (Micro bath calibrator):
Before cleaning, the tank must be drained as much as possible with the drain syringe.
When doing so, follow the corresponding instructions in the safety data sheet of the
calibration liquid used.
Distilled Water:
 Remove the sensor cage from the tank.
 Remove the magnetic stirrer with the help of the magnetic lifter.
 Dry the tank, the sensor cage, the magnetic stirrer and the drain syringe thoroughly.
Silicone oil:
 Remove the sensor cage from the tank.
 Remove the magnetic stirrer with the help of the magnetic lifter
 Clean the cage, the magnetic stirrer and the tank with water to which a generous amount
of rinsing liquid has been added.
 Remove the cleaning water to the maximum extent possible using the drain syringe.
 Dry the tank, the sensor cage, the magnetic stirrer and the drain syringe thoroughly.
8.3 Recalibration
The calibrator is adjusted and tested with measuring equipment in accordance with
recognized national standards prior to delivery.
The calibrator should, depending on the application situation, be inspected at appropriate
intervals on the basis of ISO 10012. We recommend you to return the calibrator to OMEGA
ENGINEERING at intervals of max. 12 months or approx. 500 operating hours for
recalibration and readjustment.
- 52 -
Series TCL-3M165E2
Maintenance and cleaning
8.4 Adjustment
In practice, the usage conditions during the testing process can vary from the measurement
conditions prevailing at the time of calibration of the calibrator.
For this case, the TCL-3M165E2 offers the facility to store your own adjustment values. This
is possible for your own functions of the calibrator and for your test specimens.
The number of adjustment values can be freely selected. You can equalize the calibrator
directly on-site with your standard thermometer and thus achieve a maximum of display
accuracy. It is not necessary to send it to the manufacturer.
The calibration values set at the factory are retained. You can reset the calibrator to these
values at any time.
From the configuration range of the main window ( § 5.2), or in the "Configure DUT"
window ( § 5.5.3), you can input your own adjustment values.
IMPORTANT!
Only functions and test specimens that are marked with the
icon can be adjusted.
All test tasks that use adjusted functions or test specimens are affected by the
adjustments.
Technical changes reserved
- 53 -
Decommissioning and disposal
9
Series TCL-3M165E2
Decommissioning and disposal
CAUTION! Risk of injury!
Never remove the device from a measurement set up in operation.
 Make sure that the measurement set up is shut down professionally.
Before decommissioning:
Prior to decommissioning, ensure that
 the measurement set up is switched off and is in a safe and de-energized state.
 the calibrator and the accessories have completely cooled down ( § 4.4 "Switching on,
cool down and").
Decommissioning:
 Remove all connected sensors and devices.
 Switch off the calibrator and disconnect the power plug
 Empty the calibrator tank in case there is residual calibration liquid ( § 8.2 "Cleaning").
Disposal calibration liquid:
IMPORTANT! Disposal calibration liquid!
Dispose of the calibration liquid in accordance with the Technical Safety Data Sheet.
Calibrator Disposal:
NO HOUSEHOLD WASTE!
The TCL-3M165E2 series of calibrators consists of various different materials. It must not
be disposed of with household waste.
 Take the TCL-3M165E2 to your local recycling plant
- 54 -
Series TCL-3M165E2
Technical data
10 Technical data
The technical data of customized versions may differ from the data in these instructions.
Please observe the information specified on the type plate.
10.1 Characteristics TCL-3M165E2
Characteristics
Type
TCL-3M165E2
Calibrator
Temperature range:
- Dry block
- Infrared
- Surface
- Micro bath (Silicone oil)
- Micro bath (Water)
Tolerance:
- Dry block
- Infrared
- Surface
- Micro bath
Emission degree:
- Infrared
Stability (of the control):
- Dry block
- Infrared
- Surface
- Micro bath
Measurement zone:
- Dry block
- Infrared
- Micro bath
Detection speed
Control sensor
-35…165 °C (-31…329 °F)
-35…165 °C -31…329 °F
-25…150 °C (-13…302 °F)
-35…165 °C (-31…329 °F)
2…95 °C (36…203 °F)
±0.3 °C (±0.54 °F)
±0.5 °C (±0.9 °F)
±1 °C (±1.8 °F)
±0.1 °C (±0.18 °F)
0.9994
±0.010 °C (±0.018 °F)
±0.020 °C (±0.018 °F)
±0.150 °C (±0.18 °F)
±0.010 °C (±0.018 °F)
123…163 mm (4.84...6.42 in)
110 mm (4.33 in)
110…150 mm (4.33...5.91 in)
130 ms
Internal • External
Display
Properties:
- Kind • Size
- Viewing angle
- Brightness (adjustable)
Display range
Resolution
Unit
Sensor break:
- Display
- Behavior
Behavior when temperature is
exceeded in the housing
Technical changes reserved
Color Touch screen • 7"
120…140°
≤ 400 cd/m2
-50…165 °C (-58...329 °F)
0.1 / 0.01 / 0.001 °C / °F / K
Unit °C / °F / K
Text message
Turn off the controller
Temperature fuses switch off the heating
when the limit values are exceeded.
- 55 -
Technical data
Characteristics
Series TCL-3M165E2
Type
TCL-3M165E2
Electrical characteristics
Supply voltage
Power consumption
Fuse
Interfaces:
- Network / PC
- Service
- external devices
Connecting cable
100…240 VAC • 50/60 Hz
~ 400 VA
100…240 VAC • 6.3 A slow-blow
Ethernet: RJ45
USB OTG: USB-jack Type B
USB 2.0 Host: USB-jack Type A • 12 Mbit/s Full-Speed • max 100
mA
2
H05VV-F 3 G 0.75 mm with angled protective contact plug and
cold equipment plug • Length ~ 2 m (~ 6.6 ft.)
Process variables
Metal block /
Tank opening:
- Borehole
- Depth
Dimensions:
- Width
- Height
- Depth
Weight
Sensor cage working depth
∅ 60 mm (∅ 2.36 in)
170 mm (6.69 in)
210 mm (8.27 in)
380 + 50 mm (14.96 + 1.97 in)
300 mm (11.81 in)
~ 13 kg (~ 28.66 lbs.)
150 mm (5.91 in)
Ambient conditions
Operating temperature
Transport and storage
temperature
Humidity (RH)
Operating conditions:
- Location • Altitude
- Operating position
Application area
- 56 -
0…50 °C (32…122 °F)
-10...60 °C (-14...140 °F)
< 80% to 31 °C / 87.8 °F, reducing linearly up to 50% at 40 °C /
104 °F
(non-condensing environment)
Interiors • up to 2000 m (6562 ft.)
Standing upright/vertically
tested according to EN 61326-1, class A (industrial sector)
Series TCL-3M165E2
10.2
Technical data
Heating up and cooling times
The heating up and cooling times are dependent on the parameters of the test task used and
the ambient conditions. The following times are guide values for a room temperature of 23 °C
(73.4 °F) without settling processes.
10.2.1 TCL-3M165E2
Dry block, infrared and surface
Micro bath: Water and silicone oil
Technical changes reserved
- 57 -
For your notes
For your notes
- 58 -
Series TCL-3M165E2
Technical changes reserved
- 59 -
Series TCL-3M165E2
M-5663/0318
- 60 -
Was this manual useful for you? yes no
Thank you for your participation!