Tripp Lite | N234-MI-1005 | Owner's Manual | Tripp Lite N234-MI-1005 Owner's Manual

Tripp Lite N234-MI-1005 Owner's Manual
Owner’s Manual
Medical-Grade
(IEC 60601-1) Cat6 Isolator
Model: N234-MI-1005
Este manual esta disponible en español en la
página de Tripp Lite: www.tripplite.com/support
Ce manuel est disponible en français sur le site
Web de Tripp Lite : www.tripplite.com/support
Русскоязычная версия настоящего руководства
представлена на веб-сайте компании Tripp Lite по
адресу: www.tripplite.com/support
PROTECT YOUR INVESTMENT!
Register your product for quicker service and ultimate
peace of mind. You could also win an ISOBAR6ULTRA
surge protector—a $100 value!
www.tripplite.com/warranty
1111 W. 35th Street, Chicago, IL 60609 USA
www.tripplite.com/support
Copyright © 2018 Tripp Lite. All rights reserved.
18-01-295-9337D2-EN.indd 1
1
2/6/2018 8:53:15 AM
Package Contents
• N234-MI-1005
• Quick Start Guide
Optional Accessories
• N201-Series Cat6 Gigabit Snagless Patch
Cable
• N202-Series Cat6 24 AWG Solid Wire Patch
Cable
• N125-Series Gigabit Molded Shielded Patch
Cable
General Safety Instructions
• Before use, the following section should be
read carefully and in its entirety.
• The N234-MI-1005 should only be installed
and put into operation by qualified and
trained personnel.
• Do not open the housing. There are no
serviceable parts inside. Faulty or defective
devices must be exchanged and returned to
the manufacturer.
18-01-295-9337D2-EN.indd 2
2
2/6/2018 8:53:15 AM
General Safety Instructions
• Before starting the N234-MI-1005, make
sure your medical appliance/systems permit
coupling with sufficient galvanic isolation
to the local network (LAN) or to another
computer.
Location
• During installation, ensure the
N234-MI-1005 is protected from moisture,
high humidity and mechanical damage.
The unit should not be operated in close
proximity to atmospheres with combustible
composites.
• The N234-MI-1005 should also not be
operated in close proximity to devices that
emit electromagnetic radiation so as to
avoid electromagnetic interference (e.g.
mobile phones). Avoid potential interference
by operating the N234-MI-1005 outside of
environments with electromagnetism.
• Do not use this equipment within oxygenenriched atmospheres, or within 0.3 m
(1 ft.) of a point at which an oxygenenriched atmosphere is intentionally vented.
18-01-295-9337D2-EN.indd 3
3
2/6/2018 8:53:15 AM
General Safety Instructions
Connection
When connecting the N234-MI-1005 Medical
Isolator, note that:
• You cannot connect any telecommunication
appliances, voltage supplies or similar to
the N234-MI-1005. By doing so, both the
N234-MI-1005 and personnel could be
harmed.
• You can only connect network components
complying with IEEE 802.3 or IEEE 802.3ab
(10/100/1000BASE-T, Twisted-Pair) to the
N234-MI-1005.
• You use, without exception, tested
and approved Cat6 network cable with
appropriate test mark.
Environmental Protection
For disposal of the N234-MI-1005, stipulations
subject to public law may involve special
instructions. Before final decommissioning of
the appliance, contact the manufacturer.
18-01-295-9337D2-EN.indd 4
4
2/6/2018 8:53:15 AM
Intended Use
The N234-MI-1005 is an appliance specially
developed for use in medical technology. The
purpose of the N234-MI-1005 is to connect in
compliance with electrical safety standards
IEC 60601-1 and EMC/IEC 60601-1-2 a
medical appliance or system equipped with a
network interface to a non-medical appliance
or system (single PC or local network).
Examples for using the N234-MI-1005 include:
• Connection to the practice or clinic networks
with the PC of a prolonged EEG system.
• Connection to an analysis computer in the
doctor’s or consulting room with an EMG
measuring station.
The last bullet point in the General Safety
Instructions section is to be observed for this.
18-01-295-9337D2-EN.indd 5
5
2/6/2018 8:53:15 AM
Installation
Only qualified and trained personnel should
connect the N234-MI-1005. Please instruct
your network systems administrator or
medicine technology department for this
purpose. The installation of the N234-MI-1005
is performed by way of the two pluggable
connections (RJ45 jacks). Plug the network
connection cable into the socket of the
N234-MI-1005 until the lock engages with
an audible click sound. Installation of the
N234-MI-1005 should be performed near
the medical appliance/system. Make sure the
safety information in the Connection section
of General Safety Instructions is followed.
18-01-295-9337D2-EN.indd 6
6
2/6/2018 8:53:15 AM
Maintenance
To maintain a safe technical condition, an
annual inspection is recommended. Review
and perform the following checks:
1. Examine unit for external damage (housing,
network connections, legibility of labeling,
dirt, etc.).
2. Examine unit for any substance penetration
inside of the unit (in particular, moisture).
3. Perform a function test.
4. Perform an isolation test. The isolation
resistance of the N234-MI-1005 should not
exceed 20 MOhms. During testing, all data
cables at both connections should be shut
down. The screen also should be shut down.
The isolation resistance should be measured
between signal entry and signal exit.
5. Review the availability and sufficiency of the
documents.
Should any deficiency be found during a
test, the N234-MI-1005 should not be used
any longer, as the safety of the patients
and the operator cannot be guaranteed.
If in doubt, please contact your supplier or the
manufacturer.
7
18-01-295-9337D2-EN.indd 7
2/6/2018 8:53:15 AM
Cleaning
Use a dry or a moderately damp cloth to clean
the N234-MI-1005. Make sure no fluids or
moisture penetrates the device.
18-01-295-9337D2-EN.indd 8
8
2/6/2018 8:53:15 AM
Specifications
Type Designation
Operating Mode
Operating Temperature
Operating Humidity
Operating Air Pressure
Storage Temperature
Storage Humidity
Storage Air Pressure
Rated for Altitude
Dimensions (H x W x D)
Weight
Pollution Degree
Max Isolation Voltage
between Ethernet
Connectors
Connection Sockets
Network Specifications
Material
MTBF
Product Rating
18-01-295-9337D2-EN.indd 9
MI 1005
Continuous Operation
14 to 185°F / -10 to 85°C
10% to 90% RH, Non-Condensing
700 hPA to 1060 hPA
-13 to 185°F / -25 to 85°C
10% to 95% RH, Non-Condensing
500 hPA to 1060 hPA
Up to 9,800 ft. / 3000 m
0.9 x 2.5 x 1.1 in. / 23 x 65 x 29 mm
0.1 lb. / 50 g
2
5kV AC (50/60 Hz, 1 Minute)
2x RJ45 Female
IEEE 803.2ab 10/100/1000BASE-T,
Twisted-Pair, auto-conf (completely
transparent in Ethernet network, no
drivers required)
UL94V-0 Complying Materials/RoHS
Compliant
0,21 X 10^8
(1 Fault in 21,000,000 hrs.)
Passive Ethernet Isolating Device,
Insulation Rating DI (250 V AC /
300V DC) tested at 5kV AC
9
2/6/2018 8:53:15 AM
Warranty and
Product Registration
1-Year Limited Warranty
Tripp Lite warrants its products to be free from defects in
materials and workmanship for a period of one (1) year from
the date of initial purchase. Tripp Lite’s obligation under this
warranty is limited to repairing or replacing (at its sole option)
any such defective products. To obtain service under this
warranty, you must obtain a Returned Material Authorization
(RMA) number from Tripp Lite or an authorized Tripp Lite
service center. Products must be returned to Tripp Lite or an
authorized Tripp Lite service center with transportation charges
prepaid and must be accompanied by a brief description of the
problem encountered and proof of date and place of purchase.
This warranty does not apply to equipment, which has been
damaged by accident, negligence or misapplication or has
been altered or modified in any way.
EXCEPT AS PROVIDED HEREIN, Tripp Lite MAKES NO
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. Some states do not permit limitation or exclusion
of implied warranties; therefore, the aforesaid limitation(s) or
exclusion(s) may not apply to the purchaser.
EXCEPT AS PROVIDED ABOVE, IN NO EVENT WILL Tripp Lite
BE LIABLE FOR DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR
CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF THIS
PRODUCT, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH
DAMAGE. Specifically, Tripp Lite is not liable for any costs, such
as lost profits or revenue, loss of equipment, loss of use of
equipment, loss of software, loss of data, costs of substitutes,
claims by third parties, or otherwise.
18-01-295-9337D2-EN.indd 10
10
2/6/2018 8:53:15 AM
Warranty and
Product Registration
PRODUCT REGISTRATION
Visit www.tripplite.com/warranty today to register your new
Tripp Lite product. You’ll be automatically entered into a
drawing for a chance to win a FREE Tripp Lite product!*
* No purchase necessary. Void where prohibited. Some
restrictions apply. See website for details.
WEEE Compliance Information for Tripp Lite Customers
and Recyclers (European Union)
Under the Waste Electrical and Electronic Equipment
(WEEE) Directive and implementing regulations, when
customers buy new electrical and electronic equipment
from Tripp Lite they are entitled to:
• Send old equipment for recycling on a one-for-one,
like-for-like basis (this varies depending on the
country)
• Send the new equipment back for recycling when this
ultimately becomes waste
Tripp Lite has a policy of continuous improvement.
Specifications are subject to change without notice.
18-01-295-9337D2-EN.indd 11
11
2/6/2018 8:53:15 AM
1111 W. 35th Street, Chicago, IL 60609 USA
www.tripplite.com/support
18-01-295-9337D2-EN.indd 12
12
18-01-295 93-37D2_RevA
2/6/2018 8:53:15 AM