Honeywell | 101 Load Cell | Datasheet | Honeywell 101 Load Cell Datasheet

Honeywell 101 Load Cell Datasheet
Model 101
Load Cell
DESCRIPTION
Model 101 load cells is engineered to measure bending
forces and still achieve an impressive 0.02 % accuracy. Small
dimensions allow easy unit integration into existing systems.
Model 101 uses foil gages and is available for load ranges from
50 grams to 1000 pounds.
FEATURES
•
50 kg to 1000 kg range
•
±0.02 % accuracy
•
Integral cable
•
Cost efficient
•
One-piece, nickel-plated alloy steel construction
Model 101
PERFORMANCE SPECIFICATIONS
RANGE CODES
Characteristic
Measure
Range Codes
Range
Load ranges
50, 100, 250, 500, 1000 kg
LF
50 kg
Accuracy
±0.02 % full scale
LH
100 kg
Output
2 mV/V (nominal)
LL
250 kg
Creep (max.)
±0.02 % full scale (30 minutes)
LN
500 kg
LQ
1000 kg
3
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Measure
Temperature, operating
-30 °C to 70 °C [-22 °F to 158 °F]
Temperature, compensated
-20 °C to 55 °C [-4 °F to 130 °F]
Temperature effect, zero
0.02 % full scale/10 °C
Temperature effect, span
0.02 % full scale/10 °C
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Strain gage type
Bonded foil
Excitation (calibration)
10 Vdc to 12 Vdc
Excitation (max.)
15 Vdc
Insulation resistance
>5000 mOhm
Bridge resistance (tolerance) - input resistance
400 ohm ±10 ohm (nominal)
Bridge resistance (tolerance) - output resistance
352 ohm ±2 ohm (nominal)
Zero balance (tolerance)
±1 % full scale
Electrical termination (std)
3 m flying leads
MECHANICAL SPECIFICATIONS
Characteristic
Measure
Maximum allowable load
150 % full scale1
Material
Nickel plated alloy steel
Deflection
< 2 mm
WIRING CODES
Cable
Red
(+) supply
Black
(-) supply
Green
(+) output
White
(-) output
Large black
shield
2 Honeywell • Sensing and Control
MOUNTING DIMENSIONS
Load (kg) H mm [in]
Recommended bolt
torque (N-m)
50 to 2250
18 [0.71]
60
500, 1000
24 [0.94]
60
Model 101
Load Cell
NOTES
1.
2.
3.
Allowable maximum loads – maximum load to be applied without
damage2.
Without damage - loading to this level will not cause excessive zero
shift of performance degradation. The user must consider fatigue
life for long term use and structural integrity. All structurally critical
applications (overhead loading, etc.) should always be designed
with safety redundant load paths.
This unit is calibrated to Metric (non-Imperial) units.
TYPICAL SYSTEM DIAGRAM
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to info.sc@honeywell.com
WARNING
PERSONAL INJURY
• DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
• The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
• Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008674-1-EN IL50 GLO
May 2008
Copyright © 2008 Honeywell International Inc. All rights reserved.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF