Honeywell | G Pressure Transducer | Datasheet | Honeywell G Pressure Transducer Datasheet

Honeywell G Pressure Transducer Datasheet
Model G
Subminiature, Flush Diaphragm
Pressure Transducer
DESCRIPTION
Honeywell’s full line of subminiature pressure transducers
the same temperature as the transducers. All transducers have
accurately measure pressure ranges from 150 psi to 20000 psi.
four (4) active bonded strain gages arranged in a Wheatstone-
These gage-only subminiature pressure transducers have a high
bridge configuration.
natural frequency and utilize a flush diaphragm.
Most are manfuactured with a unitized stainless steel diaphragm.
Temperature compensation is accomplished by using
The advantage of this type of design is that a thin diaphragm
temperature-sensitive components located inside the
and heavy sidewalls are made from one piece of stainless steel.
transducers. These transducers have a small electrical zero
This unitized diaphragm is rugged, but at the same time it is thin
balance circuit board which is in the lead wire (approximately 1
enough to measure low pressures.
in x 0.087 in thick). This balance board does not have to be in
FEATURES
•
150 psi to 20000 psi range
•
Flush mount design
•
High frequency
•
No internal dead volume
•
mV/V output
•
Subminiature footprint
•
1 % accuracy
•
CE approved
Model G
PERFORMANCE SPECIFICATIONS
RANGE CODES
Characteristic
Measure
Range Code
Available ranges
Pressure ranges
150, 200, 250, 300, 400, 500, 600, 750,
1000, 1500, 2000, 2500, 3000, 4000,
5000, 6000, 7500, 10000, 15000, 20000
psig
CJ
150 psig
CL
200 psig
CN
250 psig
Accuracy1
1.0 % full scale
CP
300 psig
Linearity
±1.0 % full scale
CQ
400 psig
Hysteresis
±1.0 % full scale
CR
500 psig
Non-repeatability
±0.1 % full scale
CS
600 psig
Output
2 mV/V
CT
750 psig
Resolution
Infinite
CV
1000 psig
DJ
1500 psig
DL
2000 psig
DM
2500 psig
DN
3000 psig
DR
5000 psig
DS
6000 psig
DT
7500 psig
DV
10000 psig
EJ
15000 psig
EL
20000 psig
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Measure
Temperature, operating
-54 °C to 149 °C [-65 °F to 300 °F]
Temperature, compensated
16 °C to 71 °C [60 °F to 160 °F]
Temperature effect, zero
0.01 % full scale/°F
Temperature effect, span
0.02 % reading/°F
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Strained gage type
Bonded foil
Excitation (calibration)
5 Vdc
Bridge resistance
350 ohm
Shunt calibration data
Included
Electrical termination (std)
Four twisted leads (1.83 m [5 ft]) with
external balance board
MECHANICAL SPECIFICATIONS
Characteristic
Measure
Media
Gases, liquids compatible with wetted
material
Overload, safe
50 % over capacity
Wetted parts material
17-4 PH stainless steel
Case material
Stainless steel
OPTION CODES
Range Code
Many range/option combinations are available in
our quick-ship and fast-track manufacture programs. Please see http://sensing.honeywell.com/
TMsensor-ship for updated listings.
Pressure ranges
150, 200, 250, 300, 400, 500, 600, 750, 1000,
1500, 2000, 2500, 3000, 4000, 5000, 6000,
7500, 10000, 15000, 20000 psig
Temperature
compensation
1a. 60 °F to 160 °F
1b. 30 °F to 130 °F
1c. 0 °F to 185 °F
1d. -20 °F to 130 °F
1e. -20 °F to 200 °F
Internal
amplifiers
2u. Unamp., mV/V output
Electrical
termination
Four twisted leads 1.83 m [5 ft]
6d. Microtec DR-4S-4H 4-pin (max 250 °F)2
WIRING CODES
2 Honeywell • Sensing and Control
Wire
Cable/unamplified
Red
(+) excitation
Black
(-) excitation
Green
(-) output
White
(+) output
1f. 70 °F to 250 °F
1g. 70 °F to 325 °F
1h. 70 °F to 400 °F
1i. -65 °F to 250 °F
Subminiature, Flush Diaphragm Pressure Transducer
MOUNTING DIMENSIONS AND characteristics
Order
code
Range
(psig)
ØA
B
C
D
Ø
E
Diaphragm
BP386
150
psig to
15000
psig
7,87
mm
[0.310
in]
3,81
mm
[0.15
in]
11,94
mm
[0.47
in]
11,18
mm
[0.44
in]
3,81
mm
[0.15
in]
Welded
BP387
150
psig to
20000
psig
9,53
mm
[0.375
in]
3,81
mm
[0.15
in]
11,94
mm
[0.47
in]
12,7
mm
[0.50
in]
3,81
mm
[0.15
in]
Welded
For reference only
TYPICAL SYSTEM DIAGRAM
Transducers have a small electrical balance circuit board
(approximately 1.0 in long by 0.38 in wide) located in the lead
wire that is two feet from the transducer body.
Special requirements (consult factory)
Have a special requirement? New case pressure, different cable lengths,
electrical connectors, or materials? Consult our factory by calling +1
614-850-5000 (800-848-6564). Customization is key to our test and
measurement business. Special outputs, wiring codes, and calibrations
are all standard to us.
Honeywell • Sensing and Control 3
Model G
NOTES
1.
2.
Subminiature, Flush Diaphragm Pressure Transducer
Accuracies stated are expected for best-fit straight line for all errors, including linearity, hysteresis & non-repeatability thru zero.
Only available on certain models, consult factory.
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to info.sc@honeywell.com
WARNING
PERSONAL INJURY
• DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
• The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
• Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008603-1-EN IL50 GLO
May 2008
Copyright © 2008 Honeywell International Inc. All rights reserved.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertising