About the MC series balance function settings
‫ ڧ‬Factory settings have been changed for stable weighing
The factory settings described on page 31 of the MC series instruction manual have been changed.
The MC series are high resolution balances. Therefore, external disturbances such as breezes or vibrations
may cause unstable weighing values.
For stable weighing, the factory settings of the following functions have been changed to those more
resistant to external disturbances, as shown below.
New factory settings
Class
Item and Parameter
Bba5fnc
Environment
Display
Description
Cond Condition
2
Slow response rate, stable value “SLOW”
fil Filter
1
Used
Note: When the balance initialization is performed, the function settings will be changed as shown
below. After initialization, confirm that the settings are appropriate before weighing.
Settings after initialization
Class
Item and Parameter
Bba5fnc
Environment
Display
Description
Cond Condition
1
“MID.”
fil Filter
0
Not used
‫ ڧ‬For stable weighing
When weighing values are unstable even with the functions set to those more resistant to external
disturbances (“Cond” to “1” and “fil” to “1), use the AD-1672 Tabletop Breeze Break to protect the
balance from breezes.
Usage example: AD-1672 Tabletop Breeze Break and MC-1000 Balance