Peak Scientific Genius ABN2ZA User Manual
Advertisement
Advertisement
Genius ABN2ZA
User Manual
GeniusABN2ZA User Manual Rev 4 PN UM-0004 23/03/18
Contents
Change History
How to use this Manual
Introduction 4
Warranties and Liabilities 5
3
3
Safety Notices 6
Symbols 6
Safety Notice to Users
Declaration of Conformity
6
7
Environmental Declaration
Technical Specification
Genius ABN2ZA
8
9
9
Unpacking 10
Fittings Kit Contents 11
Installation 12
Generator Environment 12
Generator Overview
General Dimensions
Rear Connections
Unit Controls
13
13
14
14
Drain Connection
Electrical Connection
Start-Up Sequence
Connecting to the application
Tubing Lengths
Normal Operation
On Demand Gas
Generator Cycling
Unusual Operation
Service Requirements
18
19
19
19
15
16
17
18
19
20
Service Schedule
Service Indication
Stage 1
Stage 2
Service Indication Reset
Peak Protected
21
22
Cleaning 23
Run Time Alarm 24
20
21
21
21
Stage 1
Stage 2
24
24
Stage 3 24
Troubleshooting 25
Page 2
Change History
Rev
2
Comment
Service Kit Ordering Update
Name
Liam Couttie
Date
23/03/2018
How to use this Manual
This manual is intended for end users and has been written as a reference document where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Page 3
Introduction
The Genius ABN2ZA has been developed to cater for the requirements of the AB Sciex range of LCMS instruments.
This model provides a source of Nitrogen gas as well as two independent supplies of clean, dry air; other features include:
•
•
•
•
Small in size – fits under a standard lab bench
Anti- vibration – maximum reduction of vibration
•
•
Service indication – allowing you to plan your maintenance and keep your LCMS uptime at a maximum
Improved drainage – reduction of moisture carry over and thus increased
reliability
Re- heat technology – improves membrane performance and reliability
Robust control system – improves safety and reliability of units
With the ABN2ZA based on proven technology, it selectively removes oxygen, moisture and other gases to leave clean, dry, phthalate free Nitrogen and Dry Air. Two internal air compressors make this unit independent from in- house air supplies and fitted castors allow the user to easily position the unit in the lab. The unit has further been approved by AB SCIEX to supply curtain, source and exhaust gas to the vast majority their LCMS applications.
To ensure this Generator model meets our high expectations with regards to reliability and performance, we have tested this new model extensively at our manufacturing plant and with end users around the world to ensure reliability and longevity of the system.
Page 4
Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed, commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in sub- clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim for death or personal injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect of any claim by the Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party.
Page 5
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may represent a potential hazard. The warnings detailed within this manual refer to the most likely potential hazards, but by definition cannot be all inclusive. If the user employs an operating procedure, item of equipment or a method of working which is not specifically recommended by Peak Scientific, the user must ensure that the equipment will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause personal injury or in the worst case death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause damage to the generator or the application. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed before proceeding.
Safety Notice to Users
WARNING
These instructions must be read thoroughly and understood before installation and operation of your Peak Generator. Use of the generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment.
WARNING
When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment maybe impaired.
WARNING
Page 6
Declaration of Conformity
We
Of
Declare that:
Equipment:
Peak Scientific Instruments Ltd.
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Nitrogen & Dry Air Generator
To which this declaration relates, is in conformity with the applicable EC directives, harmonized standards, and other normative requirements.
• Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Electrical Equipment for measurement, control and laboratory use.
• Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for measurement, control and laboratory use.
• FCC47 CFR Part 15 Class B
Unintentional radiators; Conducted and Radiated emissions limits.
All evaluation, testing and certification issued by:
Nemko Canada Inc.
303 River Road, Ottawa
K1V 1H2
Date: 20/05/2016 Position: Engineering & Operations Director
Page 7
Environmental Declaration
We
Of
Declare that:
Equipment:
Peak Scientific Instruments Ltd.
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Nitrogen & Dry Air Generator
Is fully compliant with the following Directives
• 2012/19/EU WEEE (Waste of Electrical and Electronic Equipment)
• 2011/65/EU RoHS 2 (Restriction of Hazardous Substance)
Peak Scientific Instruments Ltd fully complies with its obligations towards the European
WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These obligations are being met within the B2B compliance group.
Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to ensure that our products comply with the principles and requirements of the European recast RoHS (Restriction of Hazardous Substances) Directive 2011/65/EU. Similar directives in the United States and China, for example, have also been captured within this program.
Where a specific certificate of compliance is required, this can be requested, on a product serial number basis, directly from Peak Scientific Instruments Ltd, by contacting us through our website on www.peakscientific.com
Date: 20/05/2016 Position: Engineering & Operations Director
Page 8
Technical Specification
Genius ABN2ZA
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
5°C (41°F)
30°C (86°F)
2000 m
80% Non-Condensing
-20°C (-4°F)
60°C (140°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature for a minimum of 3 hours before operation.
Generator Outlets
Curtain Maximum Flow
Source Maximum Flow
Exhaust Maximum Flow
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlet
Pressure Gauges
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
12 L/min @ 5.5 bar (0.42 cfm @ 80 psi)
24 L/min @ 7.6 bar (0.85 cfm @ 110 psi)
8 L/min @ 4.2 bar (0.28 cfm @ 60 psi)
<0.01µm
NONE
NONE
3 x ¼” BSPP
1 x ¼” BSPP
3
230 VAC ± 10%
50/60 Hz
7 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
60 x 75 x 71.2 (23.6 x 29.5 x 28)
108.5 (239)
134 (295)
Page 9
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to unpacking.
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two people to remove the unit from the shipping crate and to manoeuvre the generator to the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful not to discard these with the packaging.
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the application. The contents of the Fittings Kit are as follows:
4. 6mm Push-Fit Fitting
Hex 1
8. US Mains Power Cable 230v x 1
All of the generators output ports are located on the output panel at the rear of the unit.
Page 11
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to the application it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe.
Performance of the generator (like all sophisticated equipment) is affected by ambient conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity. Operation of the unit within such a pocket could adversely affect its performance. Consideration should also be given to the air flow around the unit. It is recommended that an air gap of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please refer to the drawing on the following page for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
Maximum Operating Ambient Temperature:
5 °C (41 °F)
30 °C (86q °F)
Page 12
Generator Overview
General Dimensions
600 mm / 23.6” 750 mm / 29.5”
WARNING
The generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the generator.
Page 13
Rear Connections
6mm Push-Fit Fitting
¼" Compression Fitting
Mains Cable
Unit Controls
Power Switch
Pressure Gauges
Alarm LED
Service LED
Mute Button
CURTAIN
SOURCE
EXHAUST
Page 14
Drain Connection
Fit the 6mm push fit fitting to the drain port located on the output panel. Tighten using a 16mm or 5/8” spanner. Use the 6mm tubing to connect this to a suitable drain connection or container. It should be noted that the Generator can expel a considerable amount of water from this (dependant on ambient humidity).
CAUTION
If a container is used it should be emptied at regular intervals. The container must NOT have an air tight seal as water and air are expelled at pressure.
Page 15
Electrical Connection
Connect the generator to an appropriate 230 volt single-phase supply, refer to the generator serial plate for input specification and ensure your supply matches the requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can be fitted by a qualified electrician.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
WARNING
EARTH/GROUND (E):- Green & Yellow or Green
LIVE (L):- Brown or Black
Neutral (N):- Blue or White
Electrical requirements are 230VAC nominal +/- 10%. However, running continuously at voltages outwith this is not recommended. Extended periods at extremes can have a detrimental effect on the operation and life of the generator.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment maybe impaired.
To ensure that the correct voltage is being supplied to the generator, the generator comes equipped with an inbuilt voltmeter. This is located on the underside of the compressor compartment, and will measure and display the mains voltage that is being supplied to the generator.
The voltmeter should be checked prior to the initial purge of the system.
If the reading is 219V or less, then we would highly recommend fitting a Dual Tap
Transformer 06-3200 . This can be ordered directly from Peak Scientific.
Page 16
Start-Up Sequence
Before the Generator is connected to the application, the Generator should be operated in isolation (i.e. not connected to the application) for thirty minutes. This is to ensure any impurities present are purged from the system. Failure to do this may harm the application.
CAUTION
Before re-connecting the Generator to the mains and switching it on for the isolation run, it is necessary to fit the silencer to the outlet port of the output panel.
Once this is done, the Generator can be re-connected to the mains and switched on.
Pressure will start to build in the internal storage tanks which can be monitored by watching the output pressure gauge on the front panel. This will climb to the factory set pressure as noted in the specifications.
Once this pressure is reached, the compressors will continue to run until the internal tank upper pressure limit is reached and the compressor has run for a period of at least two and a half minutes.
The compressors will then rest until the internal tank lower pressure limits is reached.
Once this limit is reached the compressors will switch back on again. This compressor cycling is normal and will continue throughout the operation of the Generator.
When the system has been operated for a period of 30 minutes, all the internal pipework and storage tanks will have been purged with Nitrogen.
The Generator is now purged and the tubes can be connected at the rear of the unit.
Page 17
Connecting to the application
Once the initial purge run of 30 minutes has completed, and the generator has been running for 1 hour, (required to reach purity for the application) it is ready to be connected to the application(s)
The pressure in the internal storage tanks must be allowed to dissipate before connecting the generator to the application(s)
CAUTION
Attach the ¼” compression fittings to the outlets of the generator. Using the ¼” tubing supplied, connect the outlets of the generator to the inlets on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.
CAUTION
Tubing Lengths
Once the tubing is connected to the application, please ensure that it is thoroughly checked for being leak-tight. Even the slightest leak in the gas supply between the generator and the application can lead to a reduction in efficiency.
CAUTION
The diameter of the tubing which will be connected to the gas outlets is important and is determined by the length of tubing required. Failure to follow these recommendations could lead to pressure between generator and application.
< 10 meters: Use ¼”/ 3 ⁄
16
” (¼” O/D, 3 ⁄
16
” I/D) P.T.F.E. tubing.
> 10 - 40 meters: Use 3 ⁄
8
”/ 5 ⁄
16
in
” ( 3 ⁄
8
” O/D, 5 ⁄
16
” I/D). Tubing and fittings not supplied
> 40 metres: Please contact Peak Scientific with the relevant distance and we will calculate the flow resistance and the tubing size required.
A combination of ¼”/ 3 ⁄
16
” and 3 ⁄
8
”/ 5 ⁄
16
” tubing may be used to ensure that there is no large diameter tubing within the lab (i.e. for the first 20 meters from the generator use 3 ⁄ and the final 10 meters to the application use ¼”/ bends to a minimum.
3 ⁄
16
8
”/ 5 ⁄
16
”
” tubing). Keep the connections and
Page 18
Normal Operation
The generator is designed specifically to minimize operator involvement. Given that the system is installed as described in earlier sections and is serviced in accordance with the specified maintenance recommendations (see Service Requirements), then it should simply be a matter of turning the generator on when it is required.
The generator will automatically produce the factory set flow and pressure as detailed in the Technical Specifications.
On Demand Gas
The generator produces gas on demand. If the application is operating and requires a gas flow, the generator will supply this to suit the requirements of the application. If the application requirement for gas stops, the generator will also stop, once it has reached its upper set limit in the internal storage tanks. If the demand from the application starts again, the Generator will detect the demand for gas and will automatically start again to suit the demand.
Generator Cycling
The generator is designed for the internal compressors to cycle. This cycling reduces the duty (run time) on the compressors. The rate at which they cycle will be dependent on the gas required to satisfy the demand of the application. If the application demands the maximum gas flow of the generator, the compressor duty will be higher, (the rest period in the compressor cycle will be shorter). If the application demand is lower than the maximum gas flow, then the duty on the compressors will decrease, (the rest period in the compressor cycle will be longer). If the generator is installed in an extreme environment or is subjected to low supply voltage or high altitude the compressors may undergo periods where they do not cycle (see High Duty Indication).
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact Peak Scientific or the Peak Partner from which the generator has been purchased.
Page 19
Service Requirements
Service Schedule
Purchase Interval Component
ABN2ZA (Duplex Filter)
Annual Filter Service Kit*
12 Months
ABN2ZA Annual Filter
Service Kit*
Part No.
www.peakscientific.com/ordering
Qty.
1
1
Compressor
Assembly **
Compressor Assembly
ABN2ZA
2
* Please quote your generator serial number when ordering your Annual Maintenance
Kit to ensure selection of the appropriate kit for your generator.
** If your compressors have been refitted 3 times they must now be replaced. Please note that when replacing the compressor assembly, the full Annual Maintenance Kit will not be required. Please contact your service provider for more information.
Page 20
Service Indication
The generator has the following Service Indication Stages:-
Stage 1
Once either compressor requires a service the LED indicator (yellow) on the front of the generator will illuminate.
This is to make the user aware that a service of the generator is due and should be planned at the earliest convenience. The generator will continue to operate as normal with the LED on.
Stage 2
If the service is not completed the generator will continue to run. After 2 weeks, the service LED indicator (yellow) will start to flash.
This is to make the user aware that the service of the generator is now overdue and must be completed immediately to ensure the continuous trouble free operation of the generator.
Service Indication Reset
Once the service has been completed the Service Indication LED can be reset in the main control PLC. This will be performed by the Peak Service Engineer or trained service representative that completes the service operation.
Page 21
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of certified Peak engineers stationed throughout the globe, Peak’s rapid response team are never far away and our commitment is to keep your generator running day in, day out, protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit your lab to install and setup your generator
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned preventative maintenance
Premium Protected
Guaranteed rapid on-site response within 24 hours & planned preventative maintenance
IQ/OQ
Certified assurance for applications requiring documented qualification
Spares
Genuine Peak parts with express delivery, ensuring optimal performance and lifetime
Technical Support Hotline
Around the clock support by phone or online with our global technical helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 22
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
CAUTION
Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator.
Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the generator.
Care should be taken with Leak Detections Liquids.
CAUTION
Page 23
Run Time Alarm
The ABN2ZA monitors the operating tank pressures, in this case the N2. When this is at the set point, there is no alarm. Only when the pressure drops to a pre-determined set point will the alarm be activated, indicating LOW pressure.
There are three Alarm Stages as follows:-
Stage 1
If the control system detects that a compressor has not cycled for a period of 8 hours, then the Alarm LED indicator (red) on the front of the Generator will light. In addition to this a buzzer will sound constantly.
If the Generator returns to a cycling operation after the 8 hours, then the alarm will cancel, the Alarm LED indicator will go out and the buzzer will stop sounding.
Stage 2
If the control system detects that a compressor has not cycled for a period of 12 hours, then the Alarm will progress to Stage 2. Once at Stage 2 the Alarm LED indicator will start to flash and the buzzer will start to pulse.
If the Generator returns to a cycling operation after the 12 hours, then the Alarm will cancel, the Alarm LED indicator will go out and the buzzer will stop sounding.
Stage 3
If the control system detects that a compressor has not cycled for a period of 72 hours, then the alarm LED will continue to flash and the buzzer will continue to pulse, however the alarm will NOT reset itself. The Generator will have to be addressed by a Service
Representative and the alarm reset manually.
It should be noted that the Generator will continue to operate so long as it can maintain pressure during these alarms.
Note: the buzzer may be temporarily silenced for approximately 90 minutes by pressing the “MUTE” button on the front facia; the Alarm LED will remain lit.
Page 24
Troubleshooting
Problem
The Generator will not switch on and the power switch does not illuminate.
Possible Solution
• Ensure power cord is plugged into the Generator and that the power socket is turned on.
• Check the fuse in the power cord plug.
• Contact your service provider.
Compressors are running but pressure is not building.
• Contact your service provider.
The mass spec is reporting low pressure.
• Check pressure gauges are showing normal pressure.
• Contact your service provider.
Yellow “SERVICE” LED on front panel is on constantly.
Yellow “SERVICE” LED on front panel is flashing.
Red “RUN TIME ALARM” LED on front panel is on constantly.
• A compressor(s) due for service. Contact your service provider.
• Refer to Service Indication section of this manual for further information.
• A compressor(s) has not cycled for 8 hours. Refer to the Run Time Alarm section of this manual.
• Ensure ambient temperatures are within the specification.
• Turn the Generator off and on to reset the alarm and contact your service provider.
• A compressor(s) has not cycled for a period of 8 hours. Refer to the High Duty Indication section of this manual.
• Ensure ambient temperatures are within the specification.
• Ensure there are no leaks between the Generator and the mass spec.
Red “RUN TIME ALARM” LED on front panel is slowly flashing.
Red “RUN TIME ALARM” LED on front panel is rapidly flashing constantly and the buzzer will not stop even after power cycling the unit
• A compressor(s) has not cycled for 12 hours. Refer to the Run Time Alarm section of this manual.
• Ensure ambient temperatures are within the specification.
• Turn the Generator off and on to reset the alarm and contact your service provider.
• A compressor(s) has not cycled for 72 hours. Refer to the Run Time Alarm section of this manual.
• Ensure ambient temperatures are within the specification.
• The alarm must be reset by a qualified service technician. Contact your service provider.
Page 25
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the first thing on your mind- but it is very important to both of us. Not all warranties are alike and Peak Scientific stand out against other gas suppliers as we offer a comprehensive, quick response, on-site warranty. This means that in the very unlikely case that your gas generator develops a fault we have rapid support teams on-hand around the world who are able to come to your lab and get you back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or email at [email protected].
Page 26
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at www.peakscientific/protected .
Alternatively, you can send the completed form to Peak Scientific by post or email at [email protected]
.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an alternative gas solution i.e. cylinders or bulk liquid?
Yes No
What gas requirements do you have in your lab?
Hydrogen Nitrogen Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify Peak Scientific immediately by emailing [email protected].
For generators that remain unregistered after 1 month from the shipment date, the warranty will be considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing [email protected]
. For generators that remain unregistered after 1 month from the shipment date, the warranty will be considered active from the date of factory dispatch.
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries across every continent around the world. This allows us to provide an industry-leading rapid response service to our customers. With [Peak Protected] , your laboratory’s productivity becomes our top priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: [email protected]
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our generator products please contact [email protected]
GeniusABN2ZA User Manual Rev 4 PN UM-0004 23/03/18

Public link updated
The public link to your chat has been updated.
Advertisement