Peak Scientific Infinity 1030 Installation Guide
Peak Scientific Infinity 1030 is a gas generator designed to provide a continuous and reliable supply of gas to analytical instruments. It is a compact and lightweight generator that is easy to install and operate. The Infinity 1030 can be used to generate a variety of gases, including nitrogen, hydrogen, and zero air. It is ideal for use in a variety of applications, including chromatography, mass spectrometry, and environmental monitoring.
The Infinity 1030 features a number of advanced technologies that make it one of the most efficient and reliable gas generators on the market. These technologies include:
Advertisement
Advertisement
Installation Guide – Infinity 1032 & 1033 Generator
FITTINGS KIT
Fountain Crescent
Inchinnan Business Park F: +44 (0)141 812 8200
Inchinnan, PA4 9RE W: www.peakscientific.com
Scotland, UK
T: +44 (0)141 812 8100 Part No.
Rev. 1
Page
IG-Infinity1032&33
Page 1 of 5
1. 2.
INSTALLATION
3.
Unpack the generator from the shipping crate and place on a flat surface.
4.
The generator can be wall mounted, shown above are the distances required between fittings.
5.
Using the wall plugs and screws provided, mount the generator to the wall in the desired location.
¼” Teflon
Select the ¼” compression fittings and attach to the ‘CURTAIN’ ’SOURCE’ and ‘EXHAUST’ ports. It is recommended that PTFE tape is applied to the threads to ensure a leak-tight fit.
Select the ¼” Teflon Tube and connect to the ‘CURTAIN’ ’SOURCE’ and
‘EXHAUST’ ports. If using the compression fitting ensure the internal olive is crimped securely.
Fountain Crescent
Inchinnan Business Park F: +44 (0)141 812 8200
Inchinnan, PA4 9RE W: www.peakscientific.com
Scotland, UK
T: +44 (0)141 812 8100 Part No.
Rev. 1
Page
IG-Infinity1032&33
Page 2 of 5
6. 7.
Connect the other end of the Teflon Tube to the instrument gas inputs; again ensure the fittings are securely fastened.
8.
Attach house air supply to BA filter at left side of generator. Ensure valve on filter is at the open position.
CONGRATULATIONS
Your PEAK SCIENTIFIC gas generator is now fully installed, operational and ready to supply gas on demand to your instrument.
Fountain Crescent
Inchinnan Business Park F: +44 (0)141 812 8200
Inchinnan, PA4 9RE W: www.peakscientific.com
Scotland, UK
T: +44 (0)141 812 8100 Part No.
Rev. 1
Page
IG-Infinity1032&33
Page 3 of 5
General Notes
* The 12m length of Teflon tube supplied in the fittings kit is to allow the generator to be installed adjacent to the instrument it is supplying. If the generator is to be installed further away from the instrument then attention needs to be paid to pressure drops that could occur. The maximum distance away from the instrument the generator can be placed still using 6mm (4mm internal diameter) or ¼” (3/16” internal diameter) tubing is 10m. For further information relating to tubing lengths please refer to the user manual.
For all other technical specifications, operating instructions, service requirements, contact details and trouble shooting, please refer to the user manual contained on the CD supplied in the fittings kit.
Please keep this for future reference.
28.
It is very important to register your generator with PEAK SCIENTIFIC. This will initiate your warranty entitlement. Please use the form on the next page to register your generator. You will need the generators serial number which can be found on the serial label on the rear of the generator.
Fountain Crescent
Inchinnan Business Park F: +44 (0)141 812 8200
Inchinnan, PA4 9RE W: www.peakscientific.com
Scotland, UK
T: +44 (0)141 812 8100 Part No.
Rev. 1
Page
IG-Infinity1032&33
Page 4 of 5