Peak Scientific MS Bench SCI User Manual
Peak Scientific MS Bench SCI is a support cabinet with in-built cooling fans. Its top surface and slide-out tray can be used as required for various applications, such as housing laboratory equipment or storing supplies. The bench is spacious and can support a combined weight of up to 272KG, ensuring stability even when fully loaded. Its compact design allows for easy integration into existing laboratory setups, making it a versatile and practical addition to any workspace.
Advertisement
Advertisement
MS Bench (All Models)
User Manual
MS Bench (All Models) User Manual Rev 3 RSID 3023 EN 26/06/19
Contents
Change History
How to use this Manual
Intended Use of Equipment
Warranties and Liabilities
Safety Notices 5
Symbols 5
3
4
Safety Notice to Users
Weight Loading & Stability
Declaration of Conformity
Environmental Declaration
7
8
5
6
Technical Specification 9
Unpacking 10
Fittings Kit Contents 11
Installation 12
Generator Environment
Generator Overview
General Dimensions
MS Bench G SCI 1 Rear Connections
MS Bench G SCI 1 Unit Controls
MS Bench SCI 1 Rear Connections
14
15
12
13
13
14
3
3
MS Bench SCI 1 Unit Controls
Drain Connection (MS Bench G SCI 1 Only)
Electrical Connection
Start-Up Sequence (MS Bench G SCI 1 Only)
Connecting to the application (MS Bench G SCI 1 Only)
Tubing Lengths
Normal Operation (MS Bench G SCI 1 Only)
On Demand Gas (MS Bench G SCI 1 Only)
Generator Cycling (MS Bench G SCI 1 Only)
Unusual Operation (MS Bench G SCI 1 Only)
Fan Power LED (MS Bench SCI 1 Only)
Safe Operation of Pump Tray (MS Bench SCI 1 Only)
Service Requirements (MS Bench G SCI 1 Only)
Service Schedule
Service Indication (MS Bench G SCI 1 Only)
Stage 1
Stage 2
Service Indication Reset
Peak Protected 25
Cleaning 26
Troubleshooting (MS Bench G SCI 1 Only) 27
23
23
24
24
20
20
21
22
24
24
19
19
20
20
15
16
17
18
Page 2
Change History
2
3
1
Rev Comment
Initial Release
CSA Feedback Updates
Weight Loading Update
Name
L. Couttie
L. Couttie
L. Couttie
Date
06/06/2019
20/06/2019
26/06/2019
How to use this Manual
This manual is intended for end users and has been written as a reference document where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Intended Use of Equipment
MS Bench SCI 1, is a support cabinet with in-built cooling provided by fans, its top surface and slide out tray are available for use by the customer as require for their planned application.
MS Bench G SCI 1 is a support cabinet with an in-built Nitrogen generator. Its top surface and internal storage space are available for use by the customer as required.
The inbuilt generator provides an output of both compressed air and nitrogen for supply to customer application on demand via its compressors.
Page 3
Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed, commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in sub- clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim for death or personal injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect of any claim by the Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party.
Page 4
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may represent a potential hazard. The warnings detailed within this manual refer to the most likely potential hazards, but by definition cannot be all inclusive. If the user employs an operating procedure, item of equipment or a method of working which is not specifically recommended by Peak Scientific, the user must ensure that the equipment will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause personal injury or in the worst case death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause damage to the generator or the application. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed before proceeding.
Safety Notice to Users
WARNING
These instructions must be read thoroughly and understood before installation and operation of your Peak Generator. Use of the generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment.
WARNING
When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment maybe impaired.
WARNING
Page 5
Weight Loading & Stability
Additional weight added to product either on top surface or within cabinet must not exceed the following maximum combined total:
MS Bench SCI 1 : 272KG
MS Bench G SCI 1: 215KG
Failure to adhere to these values could lead to injury and or damage to the equipment.
CAUTION
Ensure placement of additional equipment / weight on top surface or with in cabinet does not cause overall Bench to become unstable, either in use or when sliding tray is extended and loaded on MS Bench SCI 1.
Failure to ensure overall product is stable could lead to injury and or damage to the equipment.
CAUTION
Page 6
Declaration of Conformity
We
Of
Declare that:
Peak Scientific Instruments Ltd.
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Equipment:
Models:
Bench and Nitrogen Generator Bench
MS Bench SCI 1 & MS Bench (G) SCI 1
To which this declaration relates, is in conformity with the applicable EC directives, harmonized standards, and other normative requirements.
• Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use.
• Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for Measurement, Control and Laboratory Use
- EMC Requirements.
• FCC47 CFR Part 15 Class B
Unintentional radiators; Conducted and Radiated emissions limits.
All evaluation, testing and certification issued by:
Nemko Canada Inc.
303 River Road
Canada, K1V 1H2
Peak Scientific Instruments Ltd.
Fountain Crescent, Inchinnan Business Park
PA4 9RE
Renfrewshire,
Page 7
Environmental Declaration
We
Of
Declare that:
Equipment:
Models:
Peak Scientific Instruments Ltd.
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Bench and Nitrogen Generator Bench
MS Bench SCI 1 & MS Bench (G) SCI 1
Is fully compliant with the following Directives.
• 2012/19/EU WEEE (Waste of Electrical and Electronic Equipment)
• 2011/65/EU & 2015/863 RoHS (Restriction of Hazardous Substance)
Peak Scientific Instruments Ltd fully complies with its obligations towards the European
WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These obligations are being met within the B2B compliance group.
Peak Scientific Instruments Ltd, herewith confirm that as a manufacturer of articles we conform to the European Parliament and Council Directives, 2011/65/EU for the original
6 restricted substances listed in the Restriction of Hazardous Substances (RoHS)
Directive and Commision Delegated Directive (EU) 2015/863 to include 4 additional restricted substances. Peak Scientific Instruments LTD have developed all reasonable
'due diligence' controls to ensure that our products comply with the principles and requirements of the RoHS (Restriction of Hazardous Substances) Directives.
Where a specific certificate of compliance is required, this can be requested, on a product serial number basis, directly from Peak Scientific Instruments Ltd, by contacting us through our website at www.peakscientific.com
Page 8
Technical Specification
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
MS Bench SCI 1 MS Bench G SCI 1
5°C (41°F)
30°C (86°F)
3000 m
80% Non-Condensing
-20°C (-4°F)
60°C (140°F)
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature for a minimum of 3 hours before operation.
Generator Outlets
Curtain Maximum Flow^
Source Maximum Flow^
Exhaust Maximum Flow^
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlet
N/A
19 L/min @ 4.48 bar (0.670 cfm @ 65 psi)
26 L/min @ 6.89 bar (0.918 cfm @ 100 psi)
25 L/min @ 4.13 bar (0.882 cfm @ 60 psi)
<0.01µm
NONE
NONE
3 x ¼” BSPP
1 x ¼” BSPP
^ NOTE – These performance characteristics are valid only when paired with an approved Sciex application. Please refer to current application matrix for a list of supported products.
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Installation / Over Voltage Category
100 - 240v ±10% 220 - 240v ±10%
50/60 Hz
2 Amps 6 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
91.4 x 78.7 x 76.2 (36 x 31 x 30)
103 (227)
118 (260)
160 (352)
175 (385)
Page 9
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to unpacking.
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two people to remove the unit from the shipping crate and to manoeuvre the generator to the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful not to discard these with the packaging.
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the application. The contents of the Fittings Kit are as follows:
1. ¼ ” Teflon Tubing x 3m*
2. 6mm PE Tubing x 3m*
3. ¼ ”
4. 6mm Push-Fit Fitting* x 3 x 1
6. UK Mains Power Cable 230v
7. EU Mains Power Cable 230v x 1 x 1
8. US Mains Power Cable 230v
* MS Bench G SCI 1 only x 1
All of the generators output ports are located on the output panel at the rear of the unit.
Page 11
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to the application it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket could adversely affect its performance. Consideration should also be given to the air flow around the unit. It is recommended that an air gap of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please refer to the drawing on the following page for the general dimensions of the unit.
Minimum Operating Ambient Temperature: 5 °C (41 °F)
Maximum Operating Ambient Temperature: 30 °C (86 °F)
Page 12
Generator Overview
General Dimensions
914 mm / 36” 787 mm / 31”
WARNING
The generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the generator.
Page 13
MS Bench G SCI 1 Rear Connections
IEC Power Inlet
Power Switch
Drain Outlet
Curtain Outlet
Source Outlet
Exhaust Outlet
MS Bench G SCI 1 Unit Controls
Power LED (Green)
Service/High Duty LED (Yellow)
Page 14
MS Bench SCI 1 Rear Connections
IEC Power Inlet
Power Switch
MS Bench SCI 1 Unit Controls
Fan Power LED
Page 15
Drain Connection
(MS Bench G SCI 1 Only)
Fit the 6mm push fit fitting to the drain port located on the output panel. Tighten using a 16mm or 5/8” spanner. Use the 6mm tubing to connect this to a suitable drain connection or container. It should be noted that the Generator can expel a considerable amount of water from this (dependant on ambient humidity).
If a container is used it should be emptied at regular intervals. The container must NOT have an air tight seal as water and air are expelled at pressure.
CAUTION
Page 16
Electrical Connection
Connect the generator to an appropriate single-phase supply, refer to the generator serial plate for input specification and ensure your supply matches the requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can be fitted by a qualified electrician.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
WARNING
EARTH/GROUND (E):- Green & Yellow or Green
LIVE (L):-
Neutral (N):-
Brown
Blue
or or
Black
White
Electrical requirements are 220 - 240v ±10% (MS Bench G SCI 1) & 100 - 240v ±10% (MS
Bench SCI 1). Running at voltages outwith this is not recommended. Extended periods at extremes can have a detrimental effect on the operation and life of the generator.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment maybe impaired.
WARNING
Page 17
Start-Up Sequence
(MS Bench G SCI 1 Only)
Before the Generator is connected to the application, the Generator should be operated in isolation (i.e. not connected to the application) for thirty minutes. This is to ensure any impurities present are purged from the system. Failure to do this may harm the application.
CAUTION
Once this is done, the Generator can be re-connected to the mains and switched on.
Pressure will start to build in the internal storage tanks. Once this pressure is reached, the compressors will continue to run until the internal tank upper pressure limit is reached and the compressor has run for a period of at least two minutes.
The compressors will then rest until the internal tank lower pressure limits is reached.
Once this limit is reached the compressors will switch back on again. This compressor cycling is normal and will continue throughout the operation of the Generator.
When the system has been operated for a period of 30 minutes, all the internal pipework and storage tanks will have been purged with Nitrogen.
The Generator is now purged and the tubes can be connected at the rear of the unit.
Page 18
Connecting to the application
(MS Bench G SCI 1 Only)
Once the initial purge run of 30 minutes has completed, and the generator has been running for 15 mins, (required to reach purity for the application) it is ready to be connected to the application(s)
The pressure in the internal storage tanks must be allowed to dissipate before connecting the generator to the application(s)
CAUTION
Attach the ¼ ” compression fittings to the outlets of the generator. Using the ¼ ” tubing supplied, connect the outlets of the generator to the inlets on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.
Once the tubing is connected to the application, please ensure that it is thoroughly checked for being leak-tight. Even the slightest leak in the gas supply between the generator and the application can lead to a reduction in efficiency.
CAUTION
Tubing Lengths
CAUTION
The diameter of the tubing which will be connected to the gas outlets is important and is determined by the length of tubing required. Failure to follow these recommendations could lead to pressure between generator and application.
< 10 meters: Use ¼ ”/ 3 ⁄
16
” ( ¼ ” O/D, 3 ⁄
16
” I/D) P.T.F.E. tubing.
> 10 - 40 meters: Use 3 ⁄
8
”/ 5 ⁄
16
in
” ( 3 ⁄
8
” O/D, 5 ⁄
16
” I/D). Tubing and fittings not supplied
> 40 metres: Please contact Peak Scientific with the relevant distance and we will calculate the flow resistance and the tubing size required.
A combination of ¼ ”/ 3 ⁄
16
” and 3 ⁄
8
”/ 5 ⁄
16
” tubing may be used to ensure that there is no large diameter tubing within the lab (i.e. for the first 20 meters from the generator use 3 ⁄ and the final 10 meters to the application use bends to a minimum.
¼ ”/ 3 ⁄
16
8
”/ 5 ⁄
16
”
” tubing). Keep the connections and
Page 19
Normal Operation
(MS Bench G SCI 1 Only)
The generator is designed specifically to minimize operator involvement. Given that the system is installed as described in earlier sections and is serviced in accordance with the specified maintenance recommendations (see Service Requirements), then it should simply be a matter of turning the generator on when it is required.
The generator will automatically produce the factory set flow and pressure as detailed in the Technical Specifications.
On Demand Gas
(MS Bench G SCI 1 Only)
When the application requires a gas flow, the generator will produce gas on demand.
The generator will continue to supply gas whilst the application is in a ‘Standby’ mode.
As soon as a demand is detected from the application, the generator will match the applications requirements.
Generator Cycling
(MS Bench G SCI 1 Only)
The generator is designed to operate in either a cyclic mode or run continuously, this will be dependent on the output flow.
The cyclic rate of the compressors will depend on the gas demand of the application; the generator will cycle when the upper tank pressure set limits are achieved within a
3-minute time period.
When the application demands increase, the compressor duty will be higher and the rest period of the compressor cycle will be shorter. If operated at maximum flow, the generator will run continuously to maintain the upper tank pressures. Another cause for continuous operation would be if the generator is installed at high altitude, in an extreme environment or subjected to low voltage supply. The compressors will undergo periods where they do not cycle.
Unusual Operation
(MS Bench G SCI 1 Only)
If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact Peak Scientific or the Peak Partner from which the generator has been purchased.
Page 20
Fan Power LED
(MS Bench SCI 1 Only)
The MS Bench Bench is fitted with a fan power LED on the front panel.
The LED should be illuminated when the table is supplied with power. This is indication that the internal cooling fans are being supplied with power, and are running correctly.
Page 21
Safe Operation of Pump Tray
(MS Bench SCI 1 Only)
For safe installation or removal of roughing pumps, we recommend use of the sliding function of the pump tray, as shown below.
WARNING
This should only be attempted once the Mass Spec has been positioned on top of the table.
Remove the front panel, ensuring that the fan power is disconnected.
Depress both yellow tabs simultaneously and pull tray outwards until it clicks into position.
Once pumps have been installed, ensure correct reconnection of the fan power lead.
Page 22
Service Requirements
(MS Bench G SCI 1 Only)
Service Schedule
Purchase Interval Component Visit
12 Months MS Bench G SCI 1 Annual Maintenance Kit* www.peakscientific.com/ordering
4 Years MS Bench G SCI 1 Year 4 Service Kit* www.peakscientific.com/ordering
* Please quote your generator serial number when ordering your Maintenance Kit to ensure selection of the appropriate kit for your generator.
Page 23
Service Indication
(MS Bench G SCI 1 Only)
The generator has the following Service Indication Stages:-
Stage 1
Once either compressor requires a service the LED indicator (yellow) on the front of the generator will illuminate.
This is to make the user aware that a service of the generator is due and should be planned at the earliest convenience. The generator will continue to operate as normal with the LED on.
Stage 2
If the service is not completed the generator will continue to run. After 2 weeks, the service LED indicator (yellow) will start to flash.
This is to make the user aware that the service of the generator is now overdue and must be completed immediately to ensure the continuous trouble free operation of the generator.
Service Indication Reset
Once the service has been completed the Service Indication LED can be reset in the main control PCB. This will be performed by the Peak Service Engineer or trained service representative that completes the service operation.
Page 24
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of certified Peak engineers stationed throughout the globe, Peak’s rapid response team are never far away and our commitment is to keep your generator running day in, day out, protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit your lab to install and setup your generator
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned preventative maintenance
Premium Protected
Guaranteed rapid on-site response within 24 hours & planned preventative maintenance
IQ/OQ
Certified assurance for applications requiring documented qualification
Spares
Genuine Peak parts with express delivery, ensuring optimal performance and lifetime
Technical Support Hotline
Around the clock support by phone or online with our global technical helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 25
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator.
Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the generator.
CAUTION
Care should be taken with Leak Detections Liquids.
CAUTION
Page 26
Troubleshooting
(MS Bench G SCI 1 Only)
Problem
The Generator will not switch on and the power switch does not illuminate.
Possible Solution
• Ensure power cord is plugged into the Generator and that the power socket is turned on.
• Check the fuse in the power cord plug.
• Contact your service provider.
Compressors are running but pressure is not building.
• Contact your service provider.
The mass spec is reporting low pressure.
• Contact your service provider.
Yellow “SERVICE” LED on front panel is on constantly.
Yellow “SERVICE” LED on front panel is flashing.
• A compressor(s) due for service. Contact your service provider.
• Refer to Service Indication section of this manual for further information.
• A compressor(s) has not cycled for 8 hours. Refer to the Run Time Alarm section of this manual.
• Ensure ambient temperatures are within the specification.
• Turn the Generator off and on to reset the alarm and contact your service provider.
Page 27
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the first thing on your mind- but it is very important to both of us. Not all warranties are alike and Peak Scientific stand out against other gas suppliers as we offer a comprehensive, quick response, on-site warranty. This means that in the very unlikely case that your gas generator develops a fault we have rapid support teams on-hand around the world who are able to come to your lab and get you back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or email at [email protected].
Page 28
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at www.peakscientific/protected .
Alternatively, you can send the completed form to Peak Scientific by post or email at [email protected]
.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an alternative gas solution i.e. cylinders or bulk liquid?
Yes No
What gas requirements do you have in your lab?
Hydrogen Nitrogen Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify Peak Scientific immediately by emailing [email protected].
For generators that remain unregistered after 1 month from the shipment date, the warranty will be considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing [email protected]
. For generators that remain unregistered after 1 month from the shipment date, the warranty will be considered active from the date of factory dispatch.
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries across every continent around the world. This allows us to provide an industry-leading rapid response service to our customers. With [Peak Protected] , your laboratory’s productivity becomes our top priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: [email protected]
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our generator products please contact [email protected]
MS Bench (All Models) User Manual Rev 3 RSID 3023 EN 26/06/19