Midmark Digital Vital Signs Device
TECHNICAL SPECIFICATIONS
Features
Usage
Non-invasive, multi-parameter vital signs device
Data
Quick, easy entry and viewing of vital signs data in one location
Windows® based platform
Flexible, easy-to-upgrade software for the life of the product
Display
High resolution for optimal viewing and control; 7" VGA TFT LCD display with LED backlight and resistive touch screen
Internal Battery
Vital signs can be captured continuously and quickly without plugging in
100-Reading Memory
Easy for recall, printing or download of patient data
Multiple Mounting Options
Flexibility to meet different workflow needs: mobile cart, wall mount, tabletop mount, IV pole mount
Connection
USB
Digital Scale
Supports the Midmark 626 Barrier-Free® Exam Chair with Midmark Digital Scale, Fairbanks® TeleWeigh™ scale or Health o meter® digital scales
(Note: Health o meter digital scales require connection to a computer with IQmanager® software)
Limited Warranty
Two (2) years for Midmark Digital Vital Signs Device, one (1) year for blood pressure cuffs and SpO2 sensor, 90 days for other accessories included with the kit
Safety Parameters
Type of Protection (Electrical)
Class I
Degree of Protection (Electrical)
Type BF
Degree of Protection (Water)
IPX1; protection against dripping water
Degree of Safety (Flammable Anesthetic Mixture)
Not suitable for use in the presence of a flammable anesthetic mixture
Disinfecting Method
Per the instructions in the cleaning section of the operation manual
Environmental Specifications
Operating Temperature
32°F to 104°F (0°C to 40°C) For patient temperature measurement: 61°F to 91°F (16°C to 33°C)
Storage Temperature
-4°F to 140°F (-20°C to 60°C)
Operating Altitude
0 to 15,000 feet
Storage Altitude
0 to 40,000 feet
Operating Humidity
15 to 95% non-condensing
Storage Humidity
15% to 95% non-condensing
EMC Standard
Per IEC 60601-1-2 (Emissions Class B)
Physical Characteristics
Weight
3.9 lb (1.77 kg)
Dimensions
4" W x 10.5" L x 7" H, 10.16 cm x 17.78 cm x 26.67 cm
Power (unit)
Simultaneously operates and charges device when AC power is plugged in;
low power indicator; automatic shutdown on low power
Power (battery)
Battery: rechargeable 10.8 V lithium ion battery; 4 hours to full charge, 3 hours for
95% charge; operates 8 hours on full charge
Non-Invasive Blood Pressure
Temperature
Pulse Oximetry (SpO2) Specifications
Method
Oscillometric
Probe Type
Alaris® Turbo Temp®
Cuff
Infant, Child, Small Adult, Adult, Adult Long, Large Adult, Large Adult Long and Thigh
Measurement Scale
Fahrenheit (F), Celsius (C)
Derived Parameters
Systolic and Diastolic Pressures and Mean Arterial Pressure (MAP)
Measurement Range
Systolic: 30 to 250 mmHg, Mean: 20 to 230 mmHg, Diastolic: 10 to 210 mmHg
Measurement Range
Oral/Axillary: 95°F to 107.6°F
(35°C to 42°C)
Measurement Accuracy
Systolic: ±5 mmHg, Mean: ±5 mmHg, Diastolic: ±5 mmHg
Probe Measurement Accuracy
±0.2°F (±0.1°C)
Pulse Rate Accuracy
±5% or ±2 BPM, whichever is greater
Predictive Measurement
Pulse Rate Range
30 to 240 BPM
Time (seconds)
Measurement Time
Oral: 8–10, Axillary: 13–20
Predictive Measurement
Accuracy
Oral/Axillary: Mean Difference: 0.06°F, Standard deviation: 0.38°F
(relative to steady-state temperature)
Initial Cuff Pressure
User-Selectable
Method Absorption
Spectrophotometric
(dual wavelength)
Report Interval
1 second
Measurement Range
SpO2: 20% to 100%
PR: 30 to 240 BPM
Measurement Accuracy
SpO2: 70% to 100%,
±2%, PR: ±5%
Alaris and Turbo Temp are registered trademarks of CareFusion Corp. - Windows is a registered trademark of Microsoft Corporation. - Fairbanks and TeleWeighTM are registered trademarks of Fairbanks Scales,
Inc. Health o meter is a registered trademark of Sunbeam Products, Inc.
For additional product information, contact Midmark Corporation at 1.844.856.1230, option 2.
Midmark Corporation, 690 Knox Street, Suite 100, Torrance, California 90502 USA