Midmark Wall Hugger Exam Table Installation guide

Midmark Wall Hugger Exam Table Installation guide
Wall Hugger Exam Table Setup Manual
Preparation of wood stud walls
The instructions below show how to properly brace a wood stud wall to securely hang a Wall Hugger
Exam Table. Cross braces need to be installed to minimize flexing. Mounting holes are pre-drilled into the
table mounting brackets 16" apart to correspond with standard wall stud placement. If placement of the
exam table does not match existing wall studs, modifications to the wall will be necessary. Reinforcing the
wall by securing a board between the studs is recommended for both installation on studs or between
studs.
1/8" Diameter starter holes
for lag bolts (6 plc's)
16" Standard
Stud Spacing
Determine where the table will be
installed. To mount the table at standard
36" counter top height, drill the three
bottom holes 28" above the floor per the
diagram at left. Six 3/8" diameter x 2"
long lag bolts and six flat washers have
been included with the exam table for
installation to wood stud walls. When
bolting the table to the wall, place the
flat washer between the bolt head and
table mounting bracket. See page 2.
16" Standard
Stud Spacing
6 5/8"
Drill Bottom
holes 28" from
floor for 36"
counter top.
Secure table properly. The wall hugger exam table is only as sturdy as
its mounting.
Reinforced wall studs with 2" x 10 " board secured between
studs
PREPARATION OF MASONRY OR METAL STUD WALLS
To mount Wall Hugger Exam Table on masonry or metal stud walls, a 3/4" x 20" x 50" plywood mounting
plate is recommended. Anchor hardware should be appropriate to the wall construction material. On metal
stud walls, mounting plate should span four studs.
1
1/2"
hole for
anchors/
bolts
(10x)
9
17
25
33
41
50
49
Wall Hugger mount
plate/brackets
1
3.5
10 10 1/8
19 20
Plywood
Metal Studs
Bolt Locations.
You must drill
holes through
the metal studs
to allow use of
anchors.
3/4" Plywood
50" x 20"
3/8" DIA hole for carriage bolts
To mount table (6x)
3/4" x 20" x 50" Mounting plate detail
continued
Rev. AA1 (4/24/18)
Installing the Wall Hugger Exam Table
Anchoring Wall Hugger
Exam Table to interior
metal stud wall
Anchoring Wall Hugger Exam
Table to 4" masonry wall
Metal Stud
3/4" Plywood (50" x 20")
Interior Concrete wall
3/4" Plywood (50" x 20")
When mounting,
locate bottom holes
28" from floor for
36" counter top
When mounting,
locate bottom holes
28" from floor for
36" counter top
Side view of stud wall
On metal stud walls, anchor the mounting
plywood through the metal studs. Mount
the Wall Hugger Exam Table to the
plywood mounting plate.
Side view of masonry wall
On masonry walls, anchor the mounting plywood
to the wall on side opposite of the side the Wall
Hugger Exam Table is to be mounted. Anchor the
Wall Hugger Exam Table with six lag bolts through
the wall and plywood and fasten to plywood
mounting plate with flat washers and nuts.
When bolting table to wall, place large
flat washer between the bolt head and
the table mounting plate.
Leveling the Wall Hugger Exam Table
1
Drill and insert screw as shown to insure that the table remains level
If top is not level after
mounting, loosen
screws at each end of
the slide bracket.
2
Reposition the slide
bracket until the table
top is level, then tighten
screws.
Midmark Corporation - Carthage Facilities
1000 Civil War Road
Area 5
Carthage, MO 64836
1-800-643-6275
www.midmark.com
3
Drill a pilot hole 1/8"
diameter, 1/2" deep in
the wood through the
center hole of the
bracket.
4
Install the 1/2"
screw (included) into
the pilot hole to hold
the bracket in place.