MICRO-EPSILON optoNCDT 1750 Housing M User manual

MICRO-EPSILON optoNCDT 1750 Housing M User manual
Sensor Mounting, Dimensions
Drawings, Free Space
150 (5.91)
140 (5.51)
5 (.20)
5
The optoNCDT 1750 sensor is an optical system for measurements with micrometer accuracy.
Pay attention to careful handling during mounting and operation.
35 (1.38)
98 (3.86)
3x Mounting
holes ø4.5 for
M4 screws
6
6 (.24)
17.5
(.69)
SMR
15 (.59)
Housing size M
mm (Inch)
sensor off
RS422 on
output off
analog on
optoNCDT
Limits for
free space
MR
Voltage
3 mA
5.2 V / 10.2 V
4 mA (SMR)
0V
Digital value 1
12 mA (MMR) 2.5 V / 5 V
20 mA (EMR) 5 V / 10 V
3 mA
MR
SMR
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
IEC 60825-1: 20xx-xx
P ≤1mW; =670nm
Current
5.2 V / 10.2 V
262077
98232
131000
163768
262078
Sensor Arrangement for Holes and Edges
MR
= Measuring range
SMR
= Start of measuring range
MMR = Midrange
EMR
= End of measuring range
1) For displacement values
without zero setting or mastering.
2 (.08)
Y
i
Keep this area free
from other light
sources and/or
their reflections
Turning
object
Indentation
Color
change
+U B
5
Supply voltage (11 ... 30 VDC)
GND
6
System ground supply,
switch signals (Laser on/off, Zero, Limits)
Analog output
13
Voltage 0 ... 5 VDC
Voltage 0 ... 10 VDC (Ri = 50 Ohm, I max = 5 mA)
AGND
14
Reference potential for analog output
9
Switching input,
Laser operates when pin 9 is connected to GND
Laser on/off
black
Screening,
black
red and blue
Switching input,
TrigIn, Zero/Master, TeachIn, SlaveIn
Switching output 1 8
Error/Limit 1
Switching output 2 7
Limit 2, programmable switching characteristic: (NPN,
PNP, Push-Pull)
Sync +
3
Sync -
4
Tx +
1
Tx ­-
2
Rx +
12
Rx ­-
11
white and
green
gray and pink
Symmetrical synchronous output (Master) or input (Slave)
RS422 level, terminating resistor 120 Ohm switchable, input or output depends on selected synchronization mode
RS422 - Output
(symmetric) terminate with 120 Ohm receive-site
RS422 - Input
(symmetric) internally terminated with 120 Ohm
1
2
12
3
9
11
14
13
4
5
violet
blue
pink
green
brown
gray
yellow
The PC1700 sensor cable is
qualified for drag chain use. One
end of the cable has a molded
cable connector, the other end
has braids with ferrules. Connector: ODU MINI-SNAP, 14 poles, B
series, size 2, coding 0, IP 68
10
8
7
6
Supply Voltage, Nominal value: 24 V DC (11 ... 30 V, P < 3 W)
min 5
Washer A4.3; ISO 7089 - A2
M4 x 35; ISO 4762-A2
Bolt connection
MR
500
750
SMR
200
200
Y
180
270
11 ...
30 VDC
5
ILD1750
6
Sensor PC1700-x/Y
Pin
Color
Supply
5
red
+UB
6
black
Ground
MICRO-EPSILON MESSTECHNIK
GmbH & Co. KG
Koenigbacher Str. 15 · 94496 Ortenburg
www.micro-epsilon.com
Assembly Instructions
optoNCDT 1750
Coaxial inner
conductor
Multi-function input 10
View: Solder-pin side
male cable connector,
insulator
Mounting
Alignment
to a wall
Pin Description
Mount the sensor only to the existing holes on a flat surface. Clamps of any kind are not
permitted. Do not exceed torques. The laser beam must be directed perpendicularly onto
the surface of the target. In case of misalignment it is possible that the measurement
results will not always be accurate.
2 Nm
Cable
PC1700-x
red
Signal
Current 4 ... 20 mA (RB < (UB - 6 V) / 20 mA)
10 (.39)
5
Mount the sensor only to the existing holes on a flat surface. Clamps of any kind are not
permitted.
Use three M4 screws to mount the sensors. The bearing surfaces surrounding the fastening
holes (through-holes) are slightly raised.
Measuring range, Start of Measuring range
Pin Assignment
70 (2.76)
35 (1.38)
-- Protection class: IP 65 (applies only when the sensor cable is plugged in)
Optical inputs are excluded from protection class. Contamination leads to impairment or failure
of the function.
-- Operating temperature: 0 °C ... 50 °C (+32 up to +104 °F)
-- Storage temperature: -20 °C ... 70 °C (-4 up to +158 °F)
-- Humidity: 5 - 95 % (non-condensing)
-- Ambient pressure:
Atmospheric pressure
80 (3.15)
Proper Environment
Use supply voltage for measurement instruments only. MICRO-EPSILON recommends using an optional available power
supply unit PS2020 for the sensor.
Proper Use
The optoNCDT 1750 system is designed for use in industrial and laboratory areas. It is used for
measuring displacement, distance and position as well as in in-process quality control and dimensional testing.
The sensor may only be operated within the limits specified in the technical data, see instruction
manual, Chap. 3.3. The sensor must be used in such a way that no persons are endangered or
machines are damaged in case of malfunctions or total failure of the sensor. Take additional precautions for safety and damage prevention for safety-related applications.
Warnings
Avoid unnecessary laser radiation to be exposed to the human body. Switch off the sensor for cleaning and maintenance, for system maintenance and repair if the sensor is integrated into a system.
Caution - use of controls or adjustments or performance of procedures other than those specified
may cause harm.
Connect the power supply and the display-/output device in accordance with the safety regulations
for electrical equipment. The power supply may not exceed the specified limits.
>> Risk of injury. Damage to or destruction of the sensor.
Avoid continuous exposure to splashing water on the sensor and the controller. Avoid exposure to
aggressive materials (washing agent, cooling emulsions) on the sensor.
>> Damage to or destruction of the sensor.
Avoid shock and vibration to the sensor. Protect the sensor cable against damage.
>> Damage to or destruction of the sensor , failure of the measuring device.
Laser Class
The optoNCDT 1750 sensors operate with a semiconductor laser with a wavelength of 670 nm
(visible/red) or 405 nm (visible/blue).
The following warning labels are attached to the cover (front and rear side) of the sensor housing:
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
IEC 60825-1: 2014
P≤1mW; =670nm
COMPLIES WITH 21 CFR 1040.10 AND 1040.11
EXCEPT FOR CONFORMANCE WITH
IEC 60825-1 ED. 3., AS DESCRIBED IN
LASER NOTICE NO. 56, DATED MAY 8, 2019
X9771376.01-A031119SWE
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
IEC 60825-1: 2014
P≤1mW; =405nm
Only for USA
Never deliberately look into the laser beam! Consciously close your eyes or
turn away immediately if ever the laser beam should hit your eyes.
ILD1750
+U H
1
red-blue
GND 6
i
Current output 4 ... 20 mA or
Voltage output 0 ... 5 V or 0 ... 10 V
11...
30 VDC
RB < (UB -6 V) / 20 mA;
CI
RB max. = 250 Ohm at UB = 11 V
CI ≤ 33 nF
14
6
ILD1750
13
14-pin cable
connector
Sensor
cable
Tx + (Pin 1)
green
Rx + (Pin 3)
Tx -(Pin 2)
brown
Rx -(Pin 4)
Rx + (Pin 12)
gray
Tx + (Pin 1)
Rx -(Pin 11)
yellow
Tx -(Pin 2)
GND (Pin 6)
black
GND (Pin 9)
Type 2
Type 3
PC1700-x
ILD1750
+U H
1
white-green
black
Cable/Supply
10
GND 6
24 V logic (HTL):
Low level ≤ 3 V;
High level ≥ 8 V (max 30 V)
5 V logic (TTL):
Low level ≤ 0.8 V;
High level ≥ 2 V
internal pull-up resistor, an
open input is detected as High.
50
PC1700-x/OE
PS 2020
0
Interactive web pages for programming the sensor now appear in the web browser. The sensor is
active and supplies measurement values. The ongoing measurement can be operated by means
of function buttons in the area Measurement chart.
Interface
SPS
PC1700-x/OE
Connect the input to GND to trigger the function.
Defines which interface is used for output of measured values. A parallel output o measured values via multiple channels is not possible. RS422 and analog output cannot be operated simultaneously. While using the web interface, the output is switched off via RS422.
Access via Web Interface
CU ≤ 100 nF
The inputs are not electrically isolated. The maximum switching frequency is 10 kHz.
Go to the menu Settings > Output > Output interface.
100 %
Mount the sensor and connect the components.
Source
The multi-function input enables triggering, zero setting/mastering and teaching. The function
depends on the programming of the input and on the timing of the input signal.
Select an Interface
Position the target (measurement object) as much as possible in the midrange.
RL > 20 MOhm
14
Start with a medium measuring rate. Select a measuring rate from the list. Confirm with Apply.
Place target
Components
Short circuit protection 7 mA
CU
Start the program SensorFinder Vx.x.x. You will find
this program on the provided CD.
Go to the menu Settings > Data recording > Measuring rate.
Quick Guide
Ri = 50 Ohm, I max = 5 mA,
RL
Connect the sensor to a PC/notebook via a RS422 connector. Connect the supply voltage.
Symmetric differential signals acc. to EIA-422, not electrically isolated from supply voltage. Use a
shielded cable with twisted cores e.g. PC1700-x.
Multi-Function Input
Type 1
End device (converter)
Type IF2001/USB
from MICRO-EPSILON
Voltage output
Uout
Ri
6
The current output may not be continuously operated in short-circuit operation without load resistor. This would
lead to thermal overload and thus to
the automatic overload cut-off of the
output.
Current output
Iout
RB
5
11...
30 VDC
13
Disconnect or connect the D-sub connection between RS422 and USB converter when the
sensor is disconnected from power supply only.
Sensor
Analog Output
ILD1750
i
9
black
5
i
Signal
PC1700-x
IF2004/USB
Sensor supply
is done by peripheral.
EMR
Measuring range
The State LED on the sensor indicates the position of the target to the sensor.
LED
State
USB
PC1700-x/IF2008
and IF2008-Y
adapter cable
MMR
Measuring object
SMR
IF2030/PNET
IF2001/USB
PC1700-x/IF2008 (IF2008-Y)
PC1750-x/C-Box/RJ45
In the top navigation bar other
auxiliary functions (settings,
measurement
chart etc.) are
available.
Displacement
SMR
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
IEC 60825-1: 20xx-xx
P ≤1mW; =670nm
Type 3
Select a Measuring Rate
You need a web browser (e. g.
Mozilla Firefox or Internet Explorer) on a PC/notebook.
Select the desired sensor.
Click on the button Open
WebPage.
optoNCDT
Type 2
Commissioning
Cross the lines for connections between sensor and PC.
PC1700-x/OE
Type 1
RS422 Connection with USB Converter IF2001/USB
If pin 9 is not connected
with pin 6, the laser is off.
sensor off
RS422 on
output off
analog on
Laser On
Color
Labeling
Meaning
off
Laser off
Laser beam is switched off
green
In range
Target within measuring range
yellow Midrange
Target within the midrange
red
Target outside the measuring range,
too low reflection
Error
Store the Settings
IF2008/PCIE
USB
Ethernet
C-Box/2A
Go to the menu Settings > System settings > Load & Stores or click the Save
settings button.
PC
The appearance of the websites can change dependent of the functions. Each page contains
descriptions of parameters and so tips for filling the website.
Read the detailed instruction manual before using the sensor. The manual is available online on
www.micro-epsilon.com/download/manuals/man--optoNCDT-1750--en.pdf or on the supplied
CD.
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement